Erschienen in:
01.11.2012 | How I Do it
Robotic implantation of deep brain stimulation leads, assisted by intra-operative, flat-panel CT
verfasst von:
M. Lefranc, D. Le Gars
Erschienen in:
Acta Neurochirurgica
|
Ausgabe 11/2012
Einloggen, um Zugang zu erhalten
Abstract
Background
There are two mandatory skills in deep brain stimulation (DBS) neurosurgery: accuracy and control.
Method
Frame-based robotic registration was performed. Prior to insertion into the skull, the guide tube’s position was checked with flat-panel computed tomography (fpCT). After registration against the pre-operative plan, we measured and corrected the robotic arm’s position so that the guide tube with the micro-electrode would follow the planned trajectory exactly. We then used fpCT again to check the DBS lead’s final position.
Conclusion
The combination of intra-operative fpCT with robotised surgery provides an appropriate, user-friendly solution to the key technical challenges in DBS lead implantation.