Skip to main content
Erschienen in: Acta Neurochirurgica 4/2014

01.04.2014 | Experimental research - Neurosurgery Training

Non-assisted versus neuro-navigated and XperCT-guided external ventricular catheter placement: a comparative cadaver study

verfasst von: Oliver P. Gautschi, N. R. Smoll, M. Kotowski, B. Schatlo, M. Tosic, B. Stimec, J. Fasel, K. Schaller, P. Bijlenga

Erschienen in: Acta Neurochirurgica | Ausgabe 4/2014

Einloggen, um Zugang zu erhalten

Abstract

Background and purpose

Accurate placement of an external ventricular drain (EVD) for the treatment of hydrocephalus is of paramount importance for its functionality and in order to minimize morbidity and complications. The aim of this study was to compare two different drain insertion assistance tools with the traditional free-hand anatomical landmark method, and to measure efficacy, safety and precision.

Methods

Ten cadaver heads were prepared by opening large bone windows centered on Kocher’s points on both sides. Nineteen physicians, divided in two groups (trainees and board certified neurosurgeons) performed EVD insertions. The target for the ventricular drain tip was the ipsilateral foramen of Monro. Each participant inserted the external ventricular catheter in three different ways: 1) free-hand by anatomical landmarks, 2) neuronavigation-assisted (NN), and 3) XperCT-guided (XCT). The number of ventricular hits and dangerous trajectories; time to proceed; radiation exposure of patients and physicians; distance of the catheter tip to target and size of deviations projected in the orthogonal plans were measured and compared.

Results

Insertion using XCT increased the probability of ventricular puncture from 69.2 to 90.2 % (p = 0.02). Non-assisted placements were significantly less precise (catheter tip to target distance 14.3 ± 7.4 mm versus 9.6 ± 7.2 mm, p = 0.0003). The insertion time to proceed increased from 3.04 ± 2.06 min. to 7.3 ± 3.6 min. (p < 0.001). The X-ray exposure for XCT was 32.23 mSv, but could be reduced to 13.9 mSv if patients were initially imaged in the hybrid-operating suite. No supplementary radiation exposure is needed for NN if patients are imaged according to a navigation protocol initially.

Conclusion

This ex vivo study demonstrates a significantly improved accuracy and safety using either NN or XCT-assisted methods. Therefore, efforts should be undertaken to implement these new technologies into daily clinical practice. However, the accuracy versus urgency of an EVD placement has to be balanced, as the image-guided insertion technique will implicate a longer preparation time due to a specific image acquisition and trajectory planning.
Literatur
1.
Zurück zum Zitat Azeem SS, Origitano TC (2007) Ventricular catheter placement with a frameless neuronavigational system: a 1-year experience. Neurosurgery 60:243–248PubMedCrossRef Azeem SS, Origitano TC (2007) Ventricular catheter placement with a frameless neuronavigational system: a 1-year experience. Neurosurgery 60:243–248PubMedCrossRef
2.
Zurück zum Zitat Bauer DF, Razdan SN, Bartolucci AA, Markert JM (2011) Meta-analysis of hemorrhagic complications from ventriculostomy placement by neurosurgeons. Neurosurgery 69:255–260PubMedCrossRef Bauer DF, Razdan SN, Bartolucci AA, Markert JM (2011) Meta-analysis of hemorrhagic complications from ventriculostomy placement by neurosurgeons. Neurosurgery 69:255–260PubMedCrossRef
3.
Zurück zum Zitat Binz DD, Toussaint LG 3rd, Friedman JA (2009) Hemorrhagic complications of ventriculostomy placement: a meta-analysis. Neurocrit Care 10:253–256PubMedCrossRef Binz DD, Toussaint LG 3rd, Friedman JA (2009) Hemorrhagic complications of ventriculostomy placement: a meta-analysis. Neurocrit Care 10:253–256PubMedCrossRef
4.
Zurück zum Zitat Clark S, Sangra M, Hayhurst C, Kandasamy J, Jenkinson M, Lee M, Mallucci C (2008) The use of noninvasive electromagnetic neuronavigation for slit ventricle syndrome and complex hydrocephalus in a pediatric population. J Neurosurg Pediatr 2:430–434PubMedCrossRef Clark S, Sangra M, Hayhurst C, Kandasamy J, Jenkinson M, Lee M, Mallucci C (2008) The use of noninvasive electromagnetic neuronavigation for slit ventricle syndrome and complex hydrocephalus in a pediatric population. J Neurosurg Pediatr 2:430–434PubMedCrossRef
5.
Zurück zum Zitat Dey M, Jaffe J, Stadnik A, Awad IA (2012) External ventricular drainage for intraventricular hemorrhage. Curr Neurol Neurosci Rep 12:24–33PubMedCrossRef Dey M, Jaffe J, Stadnik A, Awad IA (2012) External ventricular drainage for intraventricular hemorrhage. Curr Neurol Neurosci Rep 12:24–33PubMedCrossRef
6.
Zurück zum Zitat Dickerman RD, McConathy WJ, Morgan J, Stevens QE, Jolley JT, Schneider S, Mittler MA (2005) Failure rate of frontal versus parietal approaches for proximal catheter placement in ventriculoperitoneal shunts: revisited. J Clin Neurosci 12:781–783PubMedCrossRef Dickerman RD, McConathy WJ, Morgan J, Stevens QE, Jolley JT, Schneider S, Mittler MA (2005) Failure rate of frontal versus parietal approaches for proximal catheter placement in ventriculoperitoneal shunts: revisited. J Clin Neurosci 12:781–783PubMedCrossRef
7.
Zurück zum Zitat Friedman WA, Vries JK (1980) Percutaneous tunnel ventriculostomy. Summary of 100 procedures. J Neurosurg 53:662–665PubMedCrossRef Friedman WA, Vries JK (1980) Percutaneous tunnel ventriculostomy. Summary of 100 procedures. J Neurosurg 53:662–665PubMedCrossRef
8.
Zurück zum Zitat Gil Z, Siomin V, Beni-Adani L, Sira B, Constantini S (2002) Ventricular catheter placement in children with hydrocephalus and small ventricles: the use of a frameless neuronavigation system. Childs Nerv Syst 18:26–29PubMedCrossRef Gil Z, Siomin V, Beni-Adani L, Sira B, Constantini S (2002) Ventricular catheter placement in children with hydrocephalus and small ventricles: the use of a frameless neuronavigation system. Childs Nerv Syst 18:26–29PubMedCrossRef
9.
Zurück zum Zitat Hermann EJ, Capelle HH, Tschan CA, Krauss JK (2012) Electromagnetic-guided neuronavigation for safe placement of intraventricular catheters in pediatric neurosurgery. J Neurosurg Pediatr 10:327–333PubMedCrossRef Hermann EJ, Capelle HH, Tschan CA, Krauss JK (2012) Electromagnetic-guided neuronavigation for safe placement of intraventricular catheters in pediatric neurosurgery. J Neurosurg Pediatr 10:327–333PubMedCrossRef
10.
Zurück zum Zitat Huyette DR, Turnbow BJ, Kaufman C, Vaslow DF, Whiting BB, Oh MY (2008) Accuracy of the freehand pass technique for ventriculostomy catheter placement: retrospective assessment using computed tomography scans. J Neurosurg 108:88–91PubMedCrossRef Huyette DR, Turnbow BJ, Kaufman C, Vaslow DF, Whiting BB, Oh MY (2008) Accuracy of the freehand pass technique for ventriculostomy catheter placement: retrospective assessment using computed tomography scans. J Neurosurg 108:88–91PubMedCrossRef
11.
Zurück zum Zitat Jakola AS, Reinertsen I, Selbekk T, Solheim O, Lindseth F, Gulati S, Unsgard G (2013) Three-dimensional ultrasound-guided placement of ventricular catheters. World Neurosurg S1878–8750(13)01008–5. doi:10.1016/j.wneu.2013.08.021 Jakola AS, Reinertsen I, Selbekk T, Solheim O, Lindseth F, Gulati S, Unsgard G (2013) Three-dimensional ultrasound-guided placement of ventricular catheters. World Neurosurg S1878–8750(13)01008–5. doi:10.​1016/​j.​wneu.​2013.​08.​021
12.
Zurück zum Zitat Kakarla UK, Kim LJ, Chang SW, Theodore N, Spetzler RF (2008) Safety and accuracy of bedside external ventricular drain placement. Neurosurgery 63:ONS162–ONS167PubMedCrossRef Kakarla UK, Kim LJ, Chang SW, Theodore N, Spetzler RF (2008) Safety and accuracy of bedside external ventricular drain placement. Neurosurgery 63:ONS162–ONS167PubMedCrossRef
13.
Zurück zum Zitat Keric N, Neulen A, Kantelhardt SR, Giese A (2013) Image-guided implantation of pre-calibrated catheters in the ICU: a feasibility study. Acta Neurochir (Wien) 155:1781–1786CrossRef Keric N, Neulen A, Kantelhardt SR, Giese A (2013) Image-guided implantation of pre-calibrated catheters in the ICU: a feasibility study. Acta Neurochir (Wien) 155:1781–1786CrossRef
14.
Zurück zum Zitat Kim JH, Desai NS, Ricci J, Stieg PE, Rosengart AJ, Hartl R, Fraser JF (2012) Factors contributing to ventriculostomy infection. World Neurosurg 77:135–140PubMedCrossRef Kim JH, Desai NS, Ricci J, Stieg PE, Rosengart AJ, Hartl R, Fraser JF (2012) Factors contributing to ventriculostomy infection. World Neurosurg 77:135–140PubMedCrossRef
15.
Zurück zum Zitat Krombach G, Ganser A, Fricke C, Rohde V, Reinges M, Gilsbach J, Spetzger U (2000) Virtual placement of frontal ventricular catheters using frameless neuronavigation: an “unbloody training” for young neurosurgeons. Minim Invasive Neurosurg 43:171–175PubMedCrossRef Krombach G, Ganser A, Fricke C, Rohde V, Reinges M, Gilsbach J, Spetzger U (2000) Virtual placement of frontal ventricular catheters using frameless neuronavigation: an “unbloody training” for young neurosurgeons. Minim Invasive Neurosurg 43:171–175PubMedCrossRef
16.
Zurück zum Zitat Levitt MR, O’Neill BR, Ishak GE, Khanna PC, Temkin NR, Ellenbogen RG, Ojemann JG, Browd SR (2012) Image-guided cerebrospinal fluid shunting in children: catheter accuracy and shunt survival. J Neurosurg Pediatr 10:112–117PubMedCrossRef Levitt MR, O’Neill BR, Ishak GE, Khanna PC, Temkin NR, Ellenbogen RG, Ojemann JG, Browd SR (2012) Image-guided cerebrospinal fluid shunting in children: catheter accuracy and shunt survival. J Neurosurg Pediatr 10:112–117PubMedCrossRef
17.
Zurück zum Zitat Mahan M, Spetzler RF, Nakaji P (2013) Electromagnetic stereotactic navigation for external ventricular drain placement in the intensive care unit. J Clin Neurosci 20:1718–1722PubMedCrossRef Mahan M, Spetzler RF, Nakaji P (2013) Electromagnetic stereotactic navigation for external ventricular drain placement in the intensive care unit. J Clin Neurosci 20:1718–1722PubMedCrossRef
18.
Zurück zum Zitat Morgan T, Awad I, Keyl P, Lane K, Hanley D (2008) Preliminary report of the clot lysis evaluating accelerated resolution of intraventricular hemorrhage (CLEAR-IVH) clinical trial. Acta Neurochir Suppl 105:217–220PubMedCrossRef Morgan T, Awad I, Keyl P, Lane K, Hanley D (2008) Preliminary report of the clot lysis evaluating accelerated resolution of intraventricular hemorrhage (CLEAR-IVH) clinical trial. Acta Neurochir Suppl 105:217–220PubMedCrossRef
19.
Zurück zum Zitat O’Leary ST, Kole MK, Hoover DA, Hysell SE, Thomas A, Shaffrey CI (2000) Efficacy of the Ghajar Guide revisited: a prospective study. J Neurosurg 92:801–803PubMedCrossRef O’Leary ST, Kole MK, Hoover DA, Hysell SE, Thomas A, Shaffrey CI (2000) Efficacy of the Ghajar Guide revisited: a prospective study. J Neurosurg 92:801–803PubMedCrossRef
20.
Zurück zum Zitat O’Neill BR, Velez DA, Braxton EE, Whiting D, Oh MY (2008) A survey of ventriculostomy and intracranial pressure monitor placement practices. Surg Neurol 70:268–273PubMedCrossRef O’Neill BR, Velez DA, Braxton EE, Whiting D, Oh MY (2008) A survey of ventriculostomy and intracranial pressure monitor placement practices. Surg Neurol 70:268–273PubMedCrossRef
21.
Zurück zum Zitat Patil V, Lacson R, Vosburgh KG, Wong JM, Prevedello L, Andriole K, Mukundan S, Popp AJ, Khorasani R (2013) Factors associated with external ventricular drain placement accuracy: data from an electronic health record repository. Acta Neurochir 155:1773–1779PubMedCrossRef Patil V, Lacson R, Vosburgh KG, Wong JM, Prevedello L, Andriole K, Mukundan S, Popp AJ, Khorasani R (2013) Factors associated with external ventricular drain placement accuracy: data from an electronic health record repository. Acta Neurochir 155:1773–1779PubMedCrossRef
22.
Zurück zum Zitat Reinertsen I, Jakola AS, Friderichsen P, Lindseth F, Solheim O, Selbekk T, Unsgard G (2012) A new system for 3D ultrasound-guided placement of cerebral ventricle catheters. Int J Comput Assist Radiol Surg 7:151–157PubMedCrossRef Reinertsen I, Jakola AS, Friderichsen P, Lindseth F, Solheim O, Selbekk T, Unsgard G (2012) A new system for 3D ultrasound-guided placement of cerebral ventricle catheters. Int J Comput Assist Radiol Surg 7:151–157PubMedCrossRef
23.
Zurück zum Zitat Sainte-Rose C, Piatt JH, Renier D, Pierre-Kahn A, Hirsch JF, Hoffman HJ, Humphreys RP, Hendrick EB (1991) Mechanical complications in shunts. Pediatr Neurosurg 17:2–9PubMedCrossRef Sainte-Rose C, Piatt JH, Renier D, Pierre-Kahn A, Hirsch JF, Hoffman HJ, Humphreys RP, Hendrick EB (1991) Mechanical complications in shunts. Pediatr Neurosurg 17:2–9PubMedCrossRef
24.
Zurück zum Zitat Sampath R, Wadhwa R, Tawfik T, Nanda A, Guthikonda B (2012) Stereotactic placement of ventricular catheters: does it affect proximal malfunction rates? Stereotact Funct Neurosurg 90:97–103PubMedCrossRef Sampath R, Wadhwa R, Tawfik T, Nanda A, Guthikonda B (2012) Stereotactic placement of ventricular catheters: does it affect proximal malfunction rates? Stereotact Funct Neurosurg 90:97–103PubMedCrossRef
25.
Zurück zum Zitat Stieglitz LH, Giordano M, Samii M, Luedemann WO (2010) A new tool for frameless stereotactic placement of ventricular catheters. Neurosurgery 67:ons131–ons135PubMed Stieglitz LH, Giordano M, Samii M, Luedemann WO (2010) A new tool for frameless stereotactic placement of ventricular catheters. Neurosurgery 67:ons131–ons135PubMed
26.
Zurück zum Zitat Thomale UW, Knitter T, Schaumann A, Ahmadi SA, Ziegler P, Schulz M, Miethke C (2013) Smartphone-assisted guide for the placement of ventricular catheters. Childs Nerv Syst 29:131–139PubMedCrossRef Thomale UW, Knitter T, Schaumann A, Ahmadi SA, Ziegler P, Schulz M, Miethke C (2013) Smartphone-assisted guide for the placement of ventricular catheters. Childs Nerv Syst 29:131–139PubMedCrossRef
27.
Zurück zum Zitat Toma AK, Camp S, Watkins LD, Grieve J, Kitchen ND (2009) External ventricular drain insertion accuracy: is there a need for change in practice? Neurosurgery 65:1197–1200PubMedCrossRef Toma AK, Camp S, Watkins LD, Grieve J, Kitchen ND (2009) External ventricular drain insertion accuracy: is there a need for change in practice? Neurosurgery 65:1197–1200PubMedCrossRef
28.
Zurück zum Zitat Villavicencio AT, Leveque JC, McGirt MJ, Hopkins JS, Fuchs HE, George TM (2003) Comparison of revision rates following endoscopically versus nonendoscopically placed ventricular shunt catheters. Surg Neurol 59:375–380PubMedCrossRef Villavicencio AT, Leveque JC, McGirt MJ, Hopkins JS, Fuchs HE, George TM (2003) Comparison of revision rates following endoscopically versus nonendoscopically placed ventricular shunt catheters. Surg Neurol 59:375–380PubMedCrossRef
29.
Zurück zum Zitat Whitehead WE, Jea A, Vachhrajani S, Kulkarni AV, Drake JM (2007) Accurate placement of cerebrospinal fluid shunt ventricular catheters with real-time ultrasound guidance in older children without patent fontanelles. J Neurosurg 107:406–410PubMed Whitehead WE, Jea A, Vachhrajani S, Kulkarni AV, Drake JM (2007) Accurate placement of cerebrospinal fluid shunt ventricular catheters with real-time ultrasound guidance in older children without patent fontanelles. J Neurosurg 107:406–410PubMed
30.
Zurück zum Zitat Wilson TJ, Stetler WR Jr, Al-Holou WN, Sullivan SE (2013) Comparison of the accuracy of ventricular catheter placement using freehand placement, ultrasonic guidance, and stereotactic neuronavigation. J Neurosurg 119:66–70PubMedCrossRef Wilson TJ, Stetler WR Jr, Al-Holou WN, Sullivan SE (2013) Comparison of the accuracy of ventricular catheter placement using freehand placement, ultrasonic guidance, and stereotactic neuronavigation. J Neurosurg 119:66–70PubMedCrossRef
31.
Zurück zum Zitat Woernle CM, Burkhardt JK, Bellut D, Krayenbuehl N, Bertalanffy H (2011) Do iatrogenic factors bias the placement of external ventricular catheters?–a single institute experience and review of the literature. Neurol Med Chir (Tokyo) 51:180–186CrossRef Woernle CM, Burkhardt JK, Bellut D, Krayenbuehl N, Bertalanffy H (2011) Do iatrogenic factors bias the placement of external ventricular catheters?–a single institute experience and review of the literature. Neurol Med Chir (Tokyo) 51:180–186CrossRef
32.
Zurück zum Zitat Woodworth GF, McGirt MJ, Elfert P, Sciubba DM, Rigamonti D (2005) Frameless stereotactic ventricular shunt placement for idiopathic intracranial hypertension. Stereotact Funct Neurosurg 83:12–16PubMedCrossRef Woodworth GF, McGirt MJ, Elfert P, Sciubba DM, Rigamonti D (2005) Frameless stereotactic ventricular shunt placement for idiopathic intracranial hypertension. Stereotact Funct Neurosurg 83:12–16PubMedCrossRef
Metadaten
Titel
Non-assisted versus neuro-navigated and XperCT-guided external ventricular catheter placement: a comparative cadaver study
verfasst von
Oliver P. Gautschi
N. R. Smoll
M. Kotowski
B. Schatlo
M. Tosic
B. Stimec
J. Fasel
K. Schaller
P. Bijlenga
Publikationsdatum
01.04.2014
Verlag
Springer Vienna
Erschienen in
Acta Neurochirurgica / Ausgabe 4/2014
Print ISSN: 0001-6268
Elektronische ISSN: 0942-0940
DOI
https://doi.org/10.1007/s00701-014-2026-8

Weitere Artikel der Ausgabe 4/2014

Acta Neurochirurgica 4/2014 Zur Ausgabe

Leitlinien kompakt für die Neurologie

Mit medbee Pocketcards sicher entscheiden.

Seit 2022 gehört die medbee GmbH zum Springer Medizin Verlag

Update Neurologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.