Strict component positioning is necessary in hip resurfacing

verfasst von: Yoshitomo Kajino, Tamon Kabata, Toru Maeda, Shintaro Iwai, Kazunari Kuroda, Kenji Fujita, Hiroyuki Tsuchiya

Erschienen in: Journal of Orthopaedic Science | Ausgabe 2/2013

Einloggen, um Zugang zu erhalten

Abstract

Background

Hip resurfacing arthroplasty has some advantages, including improved metal-on-metal articulation, a lower dislocation rate and preserved femoral bone. This procedure is a surgical option for younger and more active patients with osteoarthritis and osteonecrosis of the femoral head. Although there have been some reports about the efficacy of this technique, others report serious complications caused by metal debris. Additionally, femoral neck preservation adversely decreases the head–neck ratio and results in postoperative impingement.

Methods

We evaluated the range of motion after hip resurfacing with various component orientations and optimal component orientations to avoid postoperative impingement using computer simulations in 10 male patients with osteonecrosis.

Results

The mean ranges of motion in flexion, extension, abduction, adduction and internal rotation at 90° of flexion were 92.4° ± 13.8°, 25.7° ± 13.8°, 38.0° ± 11.1°, 29.1° ± 10.0° and 20.9° ± 11.5°, respectively. The oscillation angle in flexion and extension motion was 118.1° ± 10.3°. More than 100° of flexion was acquired in 79 of 240 simulations (32.9 %), and more than 20° extension was acquired in 142 simulations (59.2 %). Combined anteversion was significantly correlated with maximal flexion and extension angles. The component safe zone to fulfill the range of motion criteria varied among patients, and 4 of 10 patients had no safe zone.

Conclusions

Postoperative impingement occurs relatively frequently in hip resurfacing because of preservation of the femoral neck and component malpositioning. The safe zone of the acetabular component to avoid postoperative impingement is very narrow. Greater care should be taken regarding patient selection, rigorous preoperative planning and accurate component positioning.
Anhänge
Nur mit Berechtigung zugänglich
Literatur
1.
Zurück zum Zitat Vendittoli PA, Lavigne M, Girard J, Roy AG. A randomised study comparing resection of acetabular bone at resurfacing and total hip replacement. J Bone Joint Surg Br. 2006;88:997–1002.PubMedCrossRef Vendittoli PA, Lavigne M, Girard J, Roy AG. A randomised study comparing resection of acetabular bone at resurfacing and total hip replacement. J Bone Joint Surg Br. 2006;88:997–1002.PubMedCrossRef
2.
Zurück zum Zitat Bengs BC, Sangiorgio SN, Ebramzadeh E. Less range of motion with resurfacing arthroplasty than with total hip arthroplasty: in vitro examination of 8 designs. Acta Orthop. 2008;79:755–62.PubMedCrossRef Bengs BC, Sangiorgio SN, Ebramzadeh E. Less range of motion with resurfacing arthroplasty than with total hip arthroplasty: in vitro examination of 8 designs. Acta Orthop. 2008;79:755–62.PubMedCrossRef
3.
Zurück zum Zitat Kluess D, Zietz C, Lindner T, Mittelmeier W, Schmitz KP, Bader R. Limited range of motion of hip resurfacing arthroplasty due to unfavorable ratio of prosthetic head size and femoral neck diameter. Acta Orthop. 2008;79:748–54.PubMedCrossRef Kluess D, Zietz C, Lindner T, Mittelmeier W, Schmitz KP, Bader R. Limited range of motion of hip resurfacing arthroplasty due to unfavorable ratio of prosthetic head size and femoral neck diameter. Acta Orthop. 2008;79:748–54.PubMedCrossRef
4.
Zurück zum Zitat Quesada MJ, Marker DR, Mont MA. Metal-on-metal hip resurfacing. J Arthroplasty. 2008;23(Suppl):69–73.PubMedCrossRef Quesada MJ, Marker DR, Mont MA. Metal-on-metal hip resurfacing. J Arthroplasty. 2008;23(Suppl):69–73.PubMedCrossRef
5.
Zurück zum Zitat Vail TP. Hip resurfacing. J Am Acad Orthop Surg. 2011;19(4):236–44.PubMed Vail TP. Hip resurfacing. J Am Acad Orthop Surg. 2011;19(4):236–44.PubMed
6.
Zurück zum Zitat McMinn DJ, Daniel J, Ziaee H, Pradhan C. Indications and results of hip resurfacing. Int Orthop. 2011;35:231–7.PubMedCrossRef McMinn DJ, Daniel J, Ziaee H, Pradhan C. Indications and results of hip resurfacing. Int Orthop. 2011;35:231–7.PubMedCrossRef
7.
Zurück zum Zitat Amstutz HC, Le Duff MJ. Eleven years of experience with metal-on-metal hybrid hip resurfacing. J Arthroplasty. 2008;23(Suppl 1):36–43.PubMedCrossRef Amstutz HC, Le Duff MJ. Eleven years of experience with metal-on-metal hybrid hip resurfacing. J Arthroplasty. 2008;23(Suppl 1):36–43.PubMedCrossRef
8.
Zurück zum Zitat Vendittoli PA, Ganapathi M, Nuño N, Plamondon D, Lavigne M. Factors affecting hip range of motion in surface replacement arthroplasty. Clin Biomech (Bristol, Avon). 2007;22:1004–12.CrossRef Vendittoli PA, Ganapathi M, Nuño N, Plamondon D, Lavigne M. Factors affecting hip range of motion in surface replacement arthroplasty. Clin Biomech (Bristol, Avon). 2007;22:1004–12.CrossRef
9.
Zurück zum Zitat Malviya A, Lingard EA, Malik A, Bowman R, Holland JP. Hip flexion after Birmingham hip resurfacing. J Arthroplasty. 2010;25:387–91.PubMedCrossRef Malviya A, Lingard EA, Malik A, Bowman R, Holland JP. Hip flexion after Birmingham hip resurfacing. J Arthroplasty. 2010;25:387–91.PubMedCrossRef
10.
Zurück zum Zitat Gruen TA, Le Duff MJ, Wisk LE, Amstutz HC. Prevalence and clinical relevance of radiographic signs of impingement in metal-on-metal hybrid hip resurfacing. J Bone Joint Surg Am. 2011;93:1519–26.PubMedCrossRef Gruen TA, Le Duff MJ, Wisk LE, Amstutz HC. Prevalence and clinical relevance of radiographic signs of impingement in metal-on-metal hybrid hip resurfacing. J Bone Joint Surg Am. 2011;93:1519–26.PubMedCrossRef
11.
Zurück zum Zitat Widmer KH, Zurfluh B. Compliant positioning of total hip components for optimal range of motion. J Orthop Res. 2004;22:815–21.PubMedCrossRef Widmer KH, Zurfluh B. Compliant positioning of total hip components for optimal range of motion. J Orthop Res. 2004;22:815–21.PubMedCrossRef
12.
Zurück zum Zitat Lewinnek GE, Lewis JL, Tarr R, Compere CL, Zimmerman JR. Dislocations after total hip replacement arthroplasties. J Bone Joint Surg Am. 1978;60:217–20.PubMed Lewinnek GE, Lewis JL, Tarr R, Compere CL, Zimmerman JR. Dislocations after total hip replacement arthroplasties. J Bone Joint Surg Am. 1978;60:217–20.PubMed
13.
Zurück zum Zitat Tannast M, Siebenrock KA, Anderson SE. Femoroacetabular impingement. AJR Am J Roentgenol. 2007;188:1540–52.PubMedCrossRef Tannast M, Siebenrock KA, Anderson SE. Femoroacetabular impingement. AJR Am J Roentgenol. 2007;188:1540–52.PubMedCrossRef
14.
Zurück zum Zitat Murray DW. The definition and measurement of acetabular orientation. J Bone Joint Surg Br. 1993;75:228–32.PubMed Murray DW. The definition and measurement of acetabular orientation. J Bone Joint Surg Br. 1993;75:228–32.PubMed
15.
Zurück zum Zitat Davis ET, Olsen M, Zdero R, Waddell JP, Schemitsch EH. Femoral neck fracture following hip resurfacing. J Bone Joint Surg Br. 2008;90:1522–7.PubMedCrossRef Davis ET, Olsen M, Zdero R, Waddell JP, Schemitsch EH. Femoral neck fracture following hip resurfacing. J Bone Joint Surg Br. 2008;90:1522–7.PubMedCrossRef
16.
Zurück zum Zitat Sakagoshi D, Kabata T, Umemoto Y, Sakamoto J, Tomita K. A mechanical analysis of femoral resurfacing implantation for osteonecrosis of the femoral head. J Arthroplasty. 2010;25:1282–9.PubMedCrossRef Sakagoshi D, Kabata T, Umemoto Y, Sakamoto J, Tomita K. A mechanical analysis of femoral resurfacing implantation for osteonecrosis of the femoral head. J Arthroplasty. 2010;25:1282–9.PubMedCrossRef
17.
Zurück zum Zitat Yamamura M, Miki H, Nakamura N, Murai M, Yoshikawa H, Sugano N. Open-configuration MRI study of femoro-acetabular impingement. J Orthop Res. 2007;25:1582–8.PubMedCrossRef Yamamura M, Miki H, Nakamura N, Murai M, Yoshikawa H, Sugano N. Open-configuration MRI study of femoro-acetabular impingement. J Orthop Res. 2007;25:1582–8.PubMedCrossRef
18.
Zurück zum Zitat Miki H, Yamanashi W, Nishii T, Sato Y, Yoshikawa H, Sugano N. Anatomic hip range of motion after implantation during total hip arthroplasty as measured by a navigation system. J Arthroplasty. 2007;22:946–52.PubMedCrossRef Miki H, Yamanashi W, Nishii T, Sato Y, Yoshikawa H, Sugano N. Anatomic hip range of motion after implantation during total hip arthroplasty as measured by a navigation system. J Arthroplasty. 2007;22:946–52.PubMedCrossRef
19.
Zurück zum Zitat Hagio K, Sugano N, Nishii T, Miki H, Otake Y, Hattori A, Suzuki N, Yonenobu K, Yoshikawa H, Ochi T. A novel system of four-dimensional motion analysis after total hip arthroplasty. J Orthop Res. 2004;22:665–70.PubMedCrossRef Hagio K, Sugano N, Nishii T, Miki H, Otake Y, Hattori A, Suzuki N, Yonenobu K, Yoshikawa H, Ochi T. A novel system of four-dimensional motion analysis after total hip arthroplasty. J Orthop Res. 2004;22:665–70.PubMedCrossRef
20.
Zurück zum Zitat Incavo SJ, Thompson MT, Gold JE, Patel RV, Icenogle KD, Noble PC. Which procedure better restores intact hip range of motion. J Arthroplasty. 2011;26:391–7.PubMedCrossRef Incavo SJ, Thompson MT, Gold JE, Patel RV, Icenogle KD, Noble PC. Which procedure better restores intact hip range of motion. J Arthroplasty. 2011;26:391–7.PubMedCrossRef
21.
Zurück zum Zitat Lavigne M, Rama KR, Roy A, Vendittoli PA. Painful impingement of the hip joint after total hip resurfacing. J Arthroplasty. 2008;23:1074–9.PubMedCrossRef Lavigne M, Rama KR, Roy A, Vendittoli PA. Painful impingement of the hip joint after total hip resurfacing. J Arthroplasty. 2008;23:1074–9.PubMedCrossRef
22.
Zurück zum Zitat Le Duff MJ, Wisk LE, Amstutz HC. Range of motion after stemmed total hip arthroplasty and hip resurfacing. Bull NYU Hosp Jt Dis. 2009;67:177–81.PubMed Le Duff MJ, Wisk LE, Amstutz HC. Range of motion after stemmed total hip arthroplasty and hip resurfacing. Bull NYU Hosp Jt Dis. 2009;67:177–81.PubMed
23.
Zurück zum Zitat Langton DJ, Sprowson AP, Joyce TJ, Reed M, Carluke I, Partington P, Nargol AV. Blood metal ion concentrations after hip resurfacing arthroplasty. J Bone Joint Surg Br. 2009;91:1287–95.PubMedCrossRef Langton DJ, Sprowson AP, Joyce TJ, Reed M, Carluke I, Partington P, Nargol AV. Blood metal ion concentrations after hip resurfacing arthroplasty. J Bone Joint Surg Br. 2009;91:1287–95.PubMedCrossRef
24.
Zurück zum Zitat Grammatopoulos G, Pandit H, Glyn-Jones S, McLardy-Smith P, Gundle R, Whitwell D, Gill HS, Murray DW. Optimal acetabular orientation for hip resurfacing. J Bone Joint Surg Br. 2010;92:1072–8.PubMedCrossRef Grammatopoulos G, Pandit H, Glyn-Jones S, McLardy-Smith P, Gundle R, Whitwell D, Gill HS, Murray DW. Optimal acetabular orientation for hip resurfacing. J Bone Joint Surg Br. 2010;92:1072–8.PubMedCrossRef
25.
Zurück zum Zitat Beaulé PE, Harvey N, Zaragoza E, Le Duff MJ, Dorey FJ. The femoral head/neck offset and hip resurfacing. J Bone Joint Surg Br. 2007;89:9–15.PubMedCrossRef Beaulé PE, Harvey N, Zaragoza E, Le Duff MJ, Dorey FJ. The femoral head/neck offset and hip resurfacing. J Bone Joint Surg Br. 2007;89:9–15.PubMedCrossRef
Metadaten
Titel
Strict component positioning is necessary in hip resurfacing
verfasst von
Yoshitomo Kajino
Tamon Kabata
Toru Maeda
Shintaro Iwai
Kazunari Kuroda
Kenji Fujita
Hiroyuki Tsuchiya
Publikationsdatum
01.03.2013
Verlag
Springer Japan
Erschienen in
Journal of Orthopaedic Science / Ausgabe 2/2013
Print ISSN: 0949-2658
Elektronische ISSN: 1436-2023
DOI
https://doi.org/10.1007/s00776-012-0351-4

Weitere Artikel der Ausgabe 2/2013

Journal of Orthopaedic Science 2/2013 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.