01.11.2014 | Original Article

Biomechanical comparison of the strength of adhesion of polymethylmethacrylate cement to zirconia ceramic and cobalt–chromium alloy components in a total knee arthroplasty

verfasst von: Nobuyuki Kumahashi, Yuji Uchio, Nobuto Kitamura, Shigeru Satake, Mikio Iwamoto, Kazunori Yasuda

Erschienen in: Journal of Orthopaedic Science | Ausgabe 6/2014

Einloggen, um Zugang zu erhalten

Abstract

Purpose

The purpose of this study was to clarify the biomechanical characteristics of cement–material interfaces for the zirconia ceramic and cobalt–chromium (Co–Cr) alloy femoral components used for total knee arthroplasty.

Methods

In the first sub-study, we compared the strength of adhesion of the cement to flat plates, by tensile testing under dry and moistened conditions. In the second sub-study, we compared the maximum load of the cement–component complex by tensile testing. In the third sub-study, we compared the fatigue characteristics of the cement–component complex by use of a dynamic tensile testing machine.

Results

Under dry conditions, the maximum strength of adhesion to the zirconia ceramic plate was the same as that to the Co–Cr alloy plate. Under moistened conditions, however, the strength of adhesion to the zirconia ceramic plate was significantly lower (p = 0.0017) whereas the strength of adhesion to the Co–Cr alloy plate was not reduced. Maximum load for the cement–component complexes for zirconia ceramic and Co–Cr alloy was no different under both dry and moistened conditions. Fatigue testing showed that cement–zirconia adhesion was stronger than cement–Co–Cr alloy adhesion (p = 0.0161).

Conclusions

The strength of adhesion of cement to zirconia ceramic is substantially weaker under wet conditions than under dry conditions. The mechanical properties of cement–zirconia ceramic component complexes and cement–Co–Cr alloy component complexes are equivalent.
Literatur
1.
Zurück zum Zitat Christel PS. Biocompatibility of surgical-grade dense polycrystalline alumina. Clin Orthop Relat Res. 1992;282:10–8.PubMed Christel PS. Biocompatibility of surgical-grade dense polycrystalline alumina. Clin Orthop Relat Res. 1992;282:10–8.PubMed
2.
Zurück zum Zitat Oonishi H, Aono M, Murata N, Kushitani S. Alumina versus polyethylene in total knee arthroplasty. Clin Orthop Relat Res. 1992;282:95–104.PubMed Oonishi H, Aono M, Murata N, Kushitani S. Alumina versus polyethylene in total knee arthroplasty. Clin Orthop Relat Res. 1992;282:95–104.PubMed
3.
Zurück zum Zitat Semlitsch M, Lehmann M, Weber H, Doerre E, Willert HG. New prospects for a prolonged functional life-span of artificial hip joint by using the material combination polyethylene/aluminum oxide ceramic/metal. J Biomed Mater Res. 1977;11:537–52.PubMedCrossRef Semlitsch M, Lehmann M, Weber H, Doerre E, Willert HG. New prospects for a prolonged functional life-span of artificial hip joint by using the material combination polyethylene/aluminum oxide ceramic/metal. J Biomed Mater Res. 1977;11:537–52.PubMedCrossRef
4.
Zurück zum Zitat Yasuda K, Miyagi N, Kaneda K. Low friction total knee arthroplasty with the alumina ceramic condylar prosthesis. Bull Hosp Jt Dis. 1993;53:15–21.PubMed Yasuda K, Miyagi N, Kaneda K. Low friction total knee arthroplasty with the alumina ceramic condylar prosthesis. Bull Hosp Jt Dis. 1993;53:15–21.PubMed
5.
Zurück zum Zitat Dorlot JM. Long-term effects of alumina components in total hip prostheses. Clin Orthop Relat Res. 1992;282:47–52.PubMed Dorlot JM. Long-term effects of alumina components in total hip prostheses. Clin Orthop Relat Res. 1992;282:47–52.PubMed
6.
Zurück zum Zitat Mittelmeier H, Heisel J. Sixteen-years’ experience with ceramic hip prostheses. Clin Orthop Relat Res. 1992;282:64–72.PubMed Mittelmeier H, Heisel J. Sixteen-years’ experience with ceramic hip prostheses. Clin Orthop Relat Res. 1992;282:64–72.PubMed
7.
Zurück zum Zitat Nizard RS, Sedel L, Christel P, Meunier A, Soudry M, Witvoet J. Ten-year survivorship of cemented ceramic-ceramic total hip prosthesis. Clin Orthop Relat Res. 1992;282:53–63.PubMed Nizard RS, Sedel L, Christel P, Meunier A, Soudry M, Witvoet J. Ten-year survivorship of cemented ceramic-ceramic total hip prosthesis. Clin Orthop Relat Res. 1992;282:53–63.PubMed
8.
Zurück zum Zitat Winter M, Griss P, Scheller G, Moser T. Ten- to 14-year results of a ceramic hip prosthesis. Clin Orthop Relat Res. 1992;282:73–80.PubMed Winter M, Griss P, Scheller G, Moser T. Ten- to 14-year results of a ceramic hip prosthesis. Clin Orthop Relat Res. 1992;282:73–80.PubMed
9.
Zurück zum Zitat Akagi M, Nakamura T, Matsusue Y, Ueo T, Nishijyo K, Ohnishi E. The bisurface total knee replacement: a unique design for flexion. Four-to-nine-year follow-up study. J Bone Jt Surg Am. 2000;82:1626–33. Akagi M, Nakamura T, Matsusue Y, Ueo T, Nishijyo K, Ohnishi E. The bisurface total knee replacement: a unique design for flexion. Four-to-nine-year follow-up study. J Bone Jt Surg Am. 2000;82:1626–33.
10.
Zurück zum Zitat Bal BS, Greenberg DD, Buhrmester L, Aleto TJ. Primary TKA with a zirconia ceramic femoral component. J Knee Surg. 2006;19:89–93.PubMed Bal BS, Greenberg DD, Buhrmester L, Aleto TJ. Primary TKA with a zirconia ceramic femoral component. J Knee Surg. 2006;19:89–93.PubMed
11.
Zurück zum Zitat Majima T, Yasuda K, Tago H, Aoki Y, Minami A. Clinical results of posterior cruciate ligament retaining TKA with alumina ceramic condular prosthesis: comparison to Co–Cr alloy prosthesis. Knee Surg Sports Traumatol Arthrosc. 2008;16:152–6.PubMedCrossRef Majima T, Yasuda K, Tago H, Aoki Y, Minami A. Clinical results of posterior cruciate ligament retaining TKA with alumina ceramic condular prosthesis: comparison to Co–Cr alloy prosthesis. Knee Surg Sports Traumatol Arthrosc. 2008;16:152–6.PubMedCrossRef
12.
Zurück zum Zitat Yasuda K, Onodera S, Kondo E, Kitamura N, Inoue M. Total knee arthroplasty with cruciate-retention type alumina ceramic condylar prosthesis. Techn Knee Surg. 2007;6:213–9.CrossRef Yasuda K, Onodera S, Kondo E, Kitamura N, Inoue M. Total knee arthroplasty with cruciate-retention type alumina ceramic condylar prosthesis. Techn Knee Surg. 2007;6:213–9.CrossRef
13.
Zurück zum Zitat Kamishima T, Kitamura N, Amemiya M, Ishizaka K, Kato F, Yasuda K, Shirato H, Terae S. Experimental MR imaging of zirconia ceramic joint implants at 1.5 and 3 T. Clin Radiol. 2010;65:387–90.PubMedCrossRef Kamishima T, Kitamura N, Amemiya M, Ishizaka K, Kato F, Yasuda K, Shirato H, Terae S. Experimental MR imaging of zirconia ceramic joint implants at 1.5 and 3 T. Clin Radiol. 2010;65:387–90.PubMedCrossRef
14.
Zurück zum Zitat Kasahara T, Iwasaki K, Kawamoto M, Watanabe K, Kobayashi D. A case of loosening of KU type after total knee arthroplasty. Chubu-seisaishi. 2005;48:647–8. Kasahara T, Iwasaki K, Kawamoto M, Watanabe K, Kobayashi D. A case of loosening of KU type after total knee arthroplasty. Chubu-seisaishi. 2005;48:647–8.
15.
Zurück zum Zitat Han HS, Kang SB, Yoon KS. High incidence of loosening of the femoral component in legacy posterior stabilised-flex total knee replacement. J Bone Joint Surg Br. 2007;89:1457–61.PubMedCrossRef Han HS, Kang SB, Yoon KS. High incidence of loosening of the femoral component in legacy posterior stabilised-flex total knee replacement. J Bone Joint Surg Br. 2007;89:1457–61.PubMedCrossRef
16.
Zurück zum Zitat Agathopoulos S, Nikolopoulos P. Wettability and interfacial interactions in bioceramic-body-liquid systems. J Biomed Mater Res. 1995;29:421–9.PubMedCrossRef Agathopoulos S, Nikolopoulos P. Wettability and interfacial interactions in bioceramic-body-liquid systems. J Biomed Mater Res. 1995;29:421–9.PubMedCrossRef
17.
Zurück zum Zitat Davidson JA. Characteristics of metal and ceramic total hip bearing surfaces and their effect on long-term ultra high molecular weight polyethylene wear. Clin Orthop Relat Res. 1993;294:361–78.PubMed Davidson JA. Characteristics of metal and ceramic total hip bearing surfaces and their effect on long-term ultra high molecular weight polyethylene wear. Clin Orthop Relat Res. 1993;294:361–78.PubMed
18.
Zurück zum Zitat Rahaman MN, Yao A. Ceramics for prosthetic hip and knee joint replacement. J Am Ceram Soc. 2007;90:1965–88.CrossRef Rahaman MN, Yao A. Ceramics for prosthetic hip and knee joint replacement. J Am Ceram Soc. 2007;90:1965–88.CrossRef
19.
Zurück zum Zitat Crookshank M, Coquim J, Olsen M, Schemitsch EH, Bougherara H, Zdero R. Biomechanical measurements of axial crush injury to the distal condyles of human and synthetic femurs. Proc Inst Mech Eng H. 2012;226:320–9.PubMedCrossRef Crookshank M, Coquim J, Olsen M, Schemitsch EH, Bougherara H, Zdero R. Biomechanical measurements of axial crush injury to the distal condyles of human and synthetic femurs. Proc Inst Mech Eng H. 2012;226:320–9.PubMedCrossRef
20.
Zurück zum Zitat Vicente A, Toledano M, Bravo LA, Romeo A, de la Higuera B, Osorio R. Effect of water contamination on the shear bond strength of five orthodontic adhesives. Med Oral Patol Oral Cir Bucal. 2010;15:e820–6.PubMedCrossRef Vicente A, Toledano M, Bravo LA, Romeo A, de la Higuera B, Osorio R. Effect of water contamination on the shear bond strength of five orthodontic adhesives. Med Oral Patol Oral Cir Bucal. 2010;15:e820–6.PubMedCrossRef
21.
Zurück zum Zitat Cacciafesta V, Jost-Brinkmann PG, Süssenberger U, Miethke RR. Effects of saliva and water contamination on the enamel shear bond strength of a light-cured glass ionomer cement. Am J Orthod Dentofacial Orthop. 1998;113:402–7.PubMed Cacciafesta V, Jost-Brinkmann PG, Süssenberger U, Miethke RR. Effects of saliva and water contamination on the enamel shear bond strength of a light-cured glass ionomer cement. Am J Orthod Dentofacial Orthop. 1998;113:402–7.PubMed
Metadaten
Titel
Biomechanical comparison of the strength of adhesion of polymethylmethacrylate cement to zirconia ceramic and cobalt–chromium alloy components in a total knee arthroplasty
verfasst von
Nobuyuki Kumahashi
Yuji Uchio
Nobuto Kitamura
Shigeru Satake
Mikio Iwamoto
Kazunori Yasuda
Publikationsdatum
01.11.2014
Verlag
Springer Japan
Erschienen in
Journal of Orthopaedic Science / Ausgabe 6/2014
Print ISSN: 0949-2658
Elektronische ISSN: 1436-2023
DOI
https://doi.org/10.1007/s00776-014-0639-7

Weitere Artikel der Ausgabe 6/2014

Journal of Orthopaedic Science 6/2014 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.