Digital method for quantification of circumferential periodontal bone level using cone beam CT

verfasst von: Jonathan Fleiner, Christian Hannig, Dirk Schulze, Andres Stricker, Reinhilde Jacobs

Erschienen in: Clinical Oral Investigations | Ausgabe 2/2013

Einloggen, um Zugang zu erhalten

Abstract

Objectives

The objective of this study is to develop a new approach for radiographically measuring circumferential periodontal bone level using cone beam CT (CBCT) data. Accuracy and precision were assessed using direct probe measurements on a human skull as a reference.

Materials and methods

Digital quantification of circumferential periodontal bone levels was conducted considering bone level measurements, infrabony crater, and furcation detection. For this purpose, a human bony cadaver skull with a restoration free dentition was used, showing periodontal bony defects of teeth 15-17,25-27,35-37,45-47 (FDI classification). Image datasets were acquired using a Promax 3D CBCT device (Planmeca Oy, Helsinki, Finland) at 80 kV and 8 mA, 160 μm voxel size. Circumferential radiographic measurements between cemento-enamel junction and the alveolar crest for the mesial, central, and distal bone levels on the oral and vestibular sides of the examined teeth were carried out based on a prototype of specifically developed software. The measurements were performed by an expert panel of three independent, calibrated, and blinded observers. Manual probe measurements of the periodontal bone loss served as reference standard.

Results

The adopted software allowed the quantification of periodontal bone loss at all examined teeth. Overall deviation between radiographic and manual measurements of the observers ranged between 0.36 and 0.69 mm; hereby, 83 % of all results were <0.5 mm. Comparing overall accuracy between the ten turns of radiological measurements, accuracy for all observers ranged from 0.29 to 0.46 mm. The present study design showed a 100 % detection of furcation involvement for radiographic evaluation.

Conclusions

The adoption of a special measurement procedure in terms of a 3D coordinate system, which is placed through and perpendicular to the long axis of the tooth, allows consistent measurement positions of the mesial, central, and distal bone levels both for the oral and vestibular sides of the alveolar crest. In this way, reliable and reproducible quantification of circumferential periodontal bone loss using CBCT data with standardized resolution of 160 μm can be performed in all three dimensions.

Clinical relevance

This new approach of radiographically assessing circumferential periodontal bone level using CBCT data shows a first promising attempt of accurate detection of periodontal bony defects. Yet, possible negative impact of further clinical parameters in terms of artifact occurrence will have to be furthermore carefully investigated.
Literatur
1.
Zurück zum Zitat Eley BM, Cox SW (1998) Advances in periodontal diagnosis. 1. Traditional clinical methods of diagnosis. Br Dent J 184:12–16PubMedCrossRef Eley BM, Cox SW (1998) Advances in periodontal diagnosis. 1. Traditional clinical methods of diagnosis. Br Dent J 184:12–16PubMedCrossRef
2.
Zurück zum Zitat Pepelassi EA, Diamanti-Kipioti A (1997) Selection of the most accurate method of conventional radiography for the assessment of periodontal osseous destruction. J Clin Periodontol 24:557–567PubMedCrossRef Pepelassi EA, Diamanti-Kipioti A (1997) Selection of the most accurate method of conventional radiography for the assessment of periodontal osseous destruction. J Clin Periodontol 24:557–567PubMedCrossRef
4.
Zurück zum Zitat Hashimoto K, Arai Y, Iwai K, Araki M, Kawashima S, Terakado M (2003) A comparison of a new limited cone beam computed tomography machine for dental use with a multidetector row helical CT machine. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 95:371–377PubMedCrossRef Hashimoto K, Arai Y, Iwai K, Araki M, Kawashima S, Terakado M (2003) A comparison of a new limited cone beam computed tomography machine for dental use with a multidetector row helical CT machine. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 95:371–377PubMedCrossRef
5.
Zurück zum Zitat Scarfe WC, Farman AG, Sukovic P (2006) Clinical applications of cone-beam computed tomography in dental practice. J Can Dent Assoc 72:75–80PubMed Scarfe WC, Farman AG, Sukovic P (2006) Clinical applications of cone-beam computed tomography in dental practice. J Can Dent Assoc 72:75–80PubMed
6.
Zurück zum Zitat Vandenberghe B, Jacobs R, Yang J (2008) Detection of periodontal bone loss using digital intraoral and cone beam computed tomography images: an in vitro assessment of bony and/or infrabony defects. Dentomaxillofac Radiol 37:252–260PubMedCrossRef Vandenberghe B, Jacobs R, Yang J (2008) Detection of periodontal bone loss using digital intraoral and cone beam computed tomography images: an in vitro assessment of bony and/or infrabony defects. Dentomaxillofac Radiol 37:252–260PubMedCrossRef
7.
Zurück zum Zitat Akesson L, Håkansson J, Rohlin M (1992) Comparison of panoramic and intraoral radiography and pocket probing for the measurement of the marginal bone level. J Clin Periodontol 19:326–332PubMedCrossRef Akesson L, Håkansson J, Rohlin M (1992) Comparison of panoramic and intraoral radiography and pocket probing for the measurement of the marginal bone level. J Clin Periodontol 19:326–332PubMedCrossRef
8.
Zurück zum Zitat Vandenberghe B, Jacobs R, Yang J (2007) Diagnostic validity (or acuity) of 2D CCD versus 3D CBCT-images for assessing periodontal breakdown. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 104:395–401PubMedCrossRef Vandenberghe B, Jacobs R, Yang J (2007) Diagnostic validity (or acuity) of 2D CCD versus 3D CBCT-images for assessing periodontal breakdown. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 104:395–401PubMedCrossRef
9.
Zurück zum Zitat Misch KA, Yi ES, Sarment DP (2006) Accuracy of cone beam computed tomography for periodontal defect measurements. J Periodontol 77:1261–1266PubMedCrossRef Misch KA, Yi ES, Sarment DP (2006) Accuracy of cone beam computed tomography for periodontal defect measurements. J Periodontol 77:1261–1266PubMedCrossRef
10.
Zurück zum Zitat Mengel R, Candir M, Shiratori K, Flores-de-Jacoby L (2005) Digital volume tomography in the diagnosis of periodontal defects: an in vitro study on native pig and human mandibles. J Periodontol 76:665–673PubMedCrossRef Mengel R, Candir M, Shiratori K, Flores-de-Jacoby L (2005) Digital volume tomography in the diagnosis of periodontal defects: an in vitro study on native pig and human mandibles. J Periodontol 76:665–673PubMedCrossRef
11.
Zurück zum Zitat Farman AG, Farman TT (2005) A comparison of 18 different X-ray detectors currently used in dentistry. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 99:485–489PubMedCrossRef Farman AG, Farman TT (2005) A comparison of 18 different X-ray detectors currently used in dentistry. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 99:485–489PubMedCrossRef
12.
Zurück zum Zitat Yamamoto K, Ueno K, Seo K, Shinohara D (2003) Development of dento-maxillofacial cone beam X-ray computed tomography system. Orthod Craniofac Res 6(Suppl 1):160–162PubMedCrossRef Yamamoto K, Ueno K, Seo K, Shinohara D (2003) Development of dento-maxillofacial cone beam X-ray computed tomography system. Orthod Craniofac Res 6(Suppl 1):160–162PubMedCrossRef
13.
Zurück zum Zitat Kasaj A, Willershausen B (2007) Digital volume tomography for diagnostics in periodontology. Int J Comput Dent 10:155–168PubMed Kasaj A, Willershausen B (2007) Digital volume tomography for diagnostics in periodontology. Int J Comput Dent 10:155–168PubMed
14.
Zurück zum Zitat Mora MA, Mol A, Tyndall DA, Rivera EM (2007) In vitro assessment of local computed tomography for the detection of longitudinal tooth fractures. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 103:825–829PubMedCrossRef Mora MA, Mol A, Tyndall DA, Rivera EM (2007) In vitro assessment of local computed tomography for the detection of longitudinal tooth fractures. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 103:825–829PubMedCrossRef
15.
Zurück zum Zitat Soğur E, Baksi BG, Gröndahl HG (2007) Imaging of root canal fillings: a comparison of subjective image quality between limited cone-beam CT, storage phosphor and film radiography. Int Endod J 40:179–185PubMedCrossRef Soğur E, Baksi BG, Gröndahl HG (2007) Imaging of root canal fillings: a comparison of subjective image quality between limited cone-beam CT, storage phosphor and film radiography. Int Endod J 40:179–185PubMedCrossRef
16.
Zurück zum Zitat Lofthag-Hansen S, Huumonen S, Gröndahl K, Gröndahl HG (2007) Limited cone-beam CT and intraoral radiography for the diagnosis of periapical pathology. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 103:114–119PubMedCrossRef Lofthag-Hansen S, Huumonen S, Gröndahl K, Gröndahl HG (2007) Limited cone-beam CT and intraoral radiography for the diagnosis of periapical pathology. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 103:114–119PubMedCrossRef
17.
Zurück zum Zitat Ludlow JB, Davies-Ludlow LE, Brooks SL (2003) Dosimetry of two extraoral direct digital imaging devices: NewTom cone beam CT and Orthophos Plus DS panoramic unit. Dentomaxillofac Radiol 32:229–234PubMedCrossRef Ludlow JB, Davies-Ludlow LE, Brooks SL (2003) Dosimetry of two extraoral direct digital imaging devices: NewTom cone beam CT and Orthophos Plus DS panoramic unit. Dentomaxillofac Radiol 32:229–234PubMedCrossRef
18.
Zurück zum Zitat Hirsch E, Wolf U, Heinicke F, Silva M (2008) Dosimetry of the cone beam computed tomography Veraviewepocs 3D compared with the 3D Accuitomo in different fields of view. Dentomaxillofac Radiol 37:268–273PubMedCrossRef Hirsch E, Wolf U, Heinicke F, Silva M (2008) Dosimetry of the cone beam computed tomography Veraviewepocs 3D compared with the 3D Accuitomo in different fields of view. Dentomaxillofac Radiol 37:268–273PubMedCrossRef
19.
Zurück zum Zitat Schulze D, Heiland M, Thurmann H, Adam G (2004) Radiation exposure during midfacial imaging using 4- and 16-slice computed tomography, cone beam computed tomography systems and conventional radiography. Dentomaxillofac Radiol 33:83–86PubMedCrossRef Schulze D, Heiland M, Thurmann H, Adam G (2004) Radiation exposure during midfacial imaging using 4- and 16-slice computed tomography, cone beam computed tomography systems and conventional radiography. Dentomaxillofac Radiol 33:83–86PubMedCrossRef
20.
Zurück zum Zitat Ludlow JB, Ivanovic M (2008) Comparative dosimetry of dental CBCT devices and 64-slice CT for oral and maxillofacial radiology. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 106:106–114PubMedCrossRef Ludlow JB, Ivanovic M (2008) Comparative dosimetry of dental CBCT devices and 64-slice CT for oral and maxillofacial radiology. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 106:106–114PubMedCrossRef
21.
Zurück zum Zitat Tsiklakis K, Donta C, Gavala S, Karayianni K, Hourdakis CJ (2005) Dose reduction in maxillofacial imaging using low dose cone beam CT. Eur J Radiol 56:413–417PubMedCrossRef Tsiklakis K, Donta C, Gavala S, Karayianni K, Hourdakis CJ (2005) Dose reduction in maxillofacial imaging using low dose cone beam CT. Eur J Radiol 56:413–417PubMedCrossRef
22.
Zurück zum Zitat Noujeim M, Prihoda T, Langlais R, Nummikoski P (2009) Evaluation of high-resolution cone beam computed tomography in the detection of simulated interradicular bone lesions. Dentomaxillofac Radiol 38:156–162PubMedCrossRef Noujeim M, Prihoda T, Langlais R, Nummikoski P (2009) Evaluation of high-resolution cone beam computed tomography in the detection of simulated interradicular bone lesions. Dentomaxillofac Radiol 38:156–162PubMedCrossRef
23.
Zurück zum Zitat Stratemann SA, Huang JC, Maki K, Miller AJ, Hatcher DC (2008) Comparison of cone beam computed tomography imaging with physical measures. Dentomaxillofac Radiol 37:80–93PubMedCrossRef Stratemann SA, Huang JC, Maki K, Miller AJ, Hatcher DC (2008) Comparison of cone beam computed tomography imaging with physical measures. Dentomaxillofac Radiol 37:80–93PubMedCrossRef
24.
Zurück zum Zitat Brown AA, Scarfe WC, Scheetz JP, Silveira AM, Farman AG (2009) Linear accuracy of cone beam CT derived 3D images. Angle Orthod 79:150–157PubMedCrossRef Brown AA, Scarfe WC, Scheetz JP, Silveira AM, Farman AG (2009) Linear accuracy of cone beam CT derived 3D images. Angle Orthod 79:150–157PubMedCrossRef
26.
Zurück zum Zitat Osborn JB, Stoltenberg JL, Huso BA, Aeppli DM, Pihlstrom BL (1992) Comparison of measurement variability in subjects with moderate periodontitis using a conventional and constant force periodontal probe. J Periodontol 63:283–289PubMedCrossRef Osborn JB, Stoltenberg JL, Huso BA, Aeppli DM, Pihlstrom BL (1992) Comparison of measurement variability in subjects with moderate periodontitis using a conventional and constant force periodontal probe. J Periodontol 63:283–289PubMedCrossRef
Metadaten
Titel
Digital method for quantification of circumferential periodontal bone level using cone beam CT
verfasst von
Jonathan Fleiner
Christian Hannig
Dirk Schulze
Andres Stricker
Reinhilde Jacobs
Publikationsdatum
01.03.2013
Verlag
Springer-Verlag
Erschienen in
Clinical Oral Investigations / Ausgabe 2/2013
Print ISSN: 1432-6981
Elektronische ISSN: 1436-3771
DOI
https://doi.org/10.1007/s00784-012-0715-3

Weitere Artikel der Ausgabe 2/2013

Clinical Oral Investigations 2/2013 Zur Ausgabe

Newsletter

Bestellen Sie unseren kostenlosen Newsletter Update Zahnmedizin und bleiben Sie gut informiert – ganz bequem per eMail.