Erschienen in: Clinical Oral Investigations 6/2017

08.11.2016 | Original Article

CAD/CAM milled removable complete dentures: an in vitro evaluation of trueness

verfasst von: Murali Srinivasan, Yoann Cantin, Albert Mehl, Harald Gjengedal, Frauke Müller, Martin Schimmel

Erschienen in: Clinical Oral Investigations | Ausgabe 6/2017

Einloggen, um Zugang zu erhalten

Abstract

Objectives

This study aimed to compare the trueness of one type of CAD/CAM milled complete removable dental prostheses (CRDPs) with injection-molding and conventionally manufactured CRDPs.

Materials and methods

Thirty-three CRDPs were fabricated by three different manufacturing techniques (group CAD/CAM (AvaDent™): n = 11; group injection molding (Ivocap™): n = 11; group flask-pack-press: n = 11) using a single master reference model and incubated in artificial saliva for 21 days. The trueness of the entire intaglio surface along with five specific regions of interest (vestibular-flange, palate, tuberosities, alveolar crest, and post-dam areas) was compared. Non-parametric tests were used with a level of significance set at p < 0.05.

Results

At baseline, there was no difference in the trueness of the total intaglio surfaces between the groups. After incubation, only the conventional CRDPs showed a significant improvement in trueness of the entire intaglio surface (p = 0.0044), but improved trueness was confirmed for all three techniques in most individual regions of interest. The 80–20 % /2 median quantile of the CAD/CAM group demonstrated the highest variability of individual readings, probably due to the size of the milling instrument. However, for all three techniques, 80 % of all deviations of the complete intaglio surface after incubation in saliva were below 0.1 mm.

Conclusions

In this in vitro study, the trueness of the intaglio surface of all three investigated techniques seems to remain within a clinically acceptable range. Additional research is warranted on material-related aspects, cost-effectiveness, clinical performance, patient-centered outcomes, as well as other CAD/CAM techniques for CRDP fabrication.

Clinical relevance

The intaglio surface trueness is an essential aspect in the clinical performance of CRDPs.
Literatur
1.
Zurück zum Zitat Jacob RF (1998) The traditional therapeutic paradigm: complete denture therapy. J Prosthet Dent 79:6–13CrossRefPubMed Jacob RF (1998) The traditional therapeutic paradigm: complete denture therapy. J Prosthet Dent 79:6–13CrossRefPubMed
2.
Zurück zum Zitat Murray MD, Darvell BW (1993) The evolution of the complete denture base. Theories of complete denture retention--a review. Part 1. Aust Dent J 38:216–219CrossRefPubMed Murray MD, Darvell BW (1993) The evolution of the complete denture base. Theories of complete denture retention--a review. Part 1. Aust Dent J 38:216–219CrossRefPubMed
3.
Zurück zum Zitat Busch M, Kordass B (2006) Concept and development of a computerized positioning of prosthetic teeth for complete dentures. Int J Comput Dent 9:113–120PubMed Busch M, Kordass B (2006) Concept and development of a computerized positioning of prosthetic teeth for complete dentures. Int J Comput Dent 9:113–120PubMed
4.
5.
Zurück zum Zitat Kanazawa M, Inokoshi M, Minakuchi S, Ohbayashi N (2011) Trial of a CAD/CAM system for fabricating complete dentures. Dent Mater J 30:93–96CrossRefPubMed Kanazawa M, Inokoshi M, Minakuchi S, Ohbayashi N (2011) Trial of a CAD/CAM system for fabricating complete dentures. Dent Mater J 30:93–96CrossRefPubMed
6.
Zurück zum Zitat Kawahata N, Ono H, Nishi Y, Hamano T, Nagaoka E (1997) Trial of duplication procedure for complete dentures by CAD/CAM. J Oral Rehabil 24:540–548CrossRefPubMed Kawahata N, Ono H, Nishi Y, Hamano T, Nagaoka E (1997) Trial of duplication procedure for complete dentures by CAD/CAM. J Oral Rehabil 24:540–548CrossRefPubMed
7.
Zurück zum Zitat Maeda Y, Minoura M, Tsutsumi S, Okada M, Nokubi T (1994) A CAD/CAM system for removable denture. Part I: fabrication of complete dentures. Int J Prosthodont 7:17–21PubMed Maeda Y, Minoura M, Tsutsumi S, Okada M, Nokubi T (1994) A CAD/CAM system for removable denture. Part I: fabrication of complete dentures. Int J Prosthodont 7:17–21PubMed
8.
Zurück zum Zitat Inokoshi M, Kanazawa M, Minakuchi S (2012) Evaluation of a complete denture trial method applying rapid prototyping. Dent Mater J 31:40–46CrossRefPubMed Inokoshi M, Kanazawa M, Minakuchi S (2012) Evaluation of a complete denture trial method applying rapid prototyping. Dent Mater J 31:40–46CrossRefPubMed
10.
12.
Zurück zum Zitat Kattadiyil MT, Goodacre CJ, Lozada JL, Garbacea A (2015) Digitally planned and fabricated mandibular fixed complete dentures. Part 2. Prosthodontic phase. Int J Prosthodont 28:119–123. doi:10.11607/ijp.4181 CrossRefPubMed Kattadiyil MT, Goodacre CJ, Lozada JL, Garbacea A (2015) Digitally planned and fabricated mandibular fixed complete dentures. Part 2. Prosthodontic phase. Int J Prosthodont 28:119–123. doi:10.​11607/​ijp.​4181 CrossRefPubMed
14.
15.
Zurück zum Zitat Papaspyridakos P, Gallucci GO, Chen CJ, Hanssen S, Naert I, Vandenberghe B (2016) Digital versus conventional implant impressions for edentulous patients: accuracy outcomes. Clin Oral Implants Res 27:465–472. doi:10.1111/clr.12567 CrossRefPubMed Papaspyridakos P, Gallucci GO, Chen CJ, Hanssen S, Naert I, Vandenberghe B (2016) Digital versus conventional implant impressions for edentulous patients: accuracy outcomes. Clin Oral Implants Res 27:465–472. doi:10.​1111/​clr.​12567 CrossRefPubMed
16.
Zurück zum Zitat Srinivasan M, Schimmel M, Badoud I, Ammann P, Herrmann FR, Muller F (2015) Influence of implant angulation and cyclic dislodging on the retentive force of two different overdenture attachments - an in vitro study. Clin Oral Implants Res. doi:10.1111/clr.12643 Srinivasan M, Schimmel M, Badoud I, Ammann P, Herrmann FR, Muller F (2015) Influence of implant angulation and cyclic dislodging on the retentive force of two different overdenture attachments - an in vitro study. Clin Oral Implants Res. doi:10.​1111/​clr.​12643
17.
Zurück zum Zitat Srinivasan M, Schimmel M, Kobayashi M, Badoud I, Ammann P, Herrmann FR, Muller F (2015) Influence of different lubricants on the retentive force of LOCATOR attachments - an in vitro pilot study. Clin Oral Implants Res. doi:10.1111/clr.12680 Srinivasan M, Schimmel M, Kobayashi M, Badoud I, Ammann P, Herrmann FR, Muller F (2015) Influence of different lubricants on the retentive force of LOCATOR attachments - an in vitro pilot study. Clin Oral Implants Res. doi:10.​1111/​clr.​12680
18.
Zurück zum Zitat Mehl A, Koch R, Zaruba M, Ender A (2013) 3D monitoring and quality control using intraoral optical camera systems. Int J Comput Dent 16:23–36PubMed Mehl A, Koch R, Zaruba M, Ender A (2013) 3D monitoring and quality control using intraoral optical camera systems. Int J Comput Dent 16:23–36PubMed
19.
Zurück zum Zitat Flugge TV, Schlager S, Nelson K, Nahles S, Metzger MC (2013) Precision of intraoral digital dental impressions with iTero and extraoral digitization with the iTero and a model scanner. Am J Orthod Dentofac Orthop 144:471–478. doi:10.1016/j.ajodo.2013.04.017 CrossRef Flugge TV, Schlager S, Nelson K, Nahles S, Metzger MC (2013) Precision of intraoral digital dental impressions with iTero and extraoral digitization with the iTero and a model scanner. Am J Orthod Dentofac Orthop 144:471–478. doi:10.​1016/​j.​ajodo.​2013.​04.​017 CrossRef
20.
Zurück zum Zitat Harrell FE Jr, Lee KL, Califf RM, Pryor DB, Rosati RA (1984) Regression modelling strategies for improved prognostic prediction. Stat Med 3:143–152CrossRefPubMed Harrell FE Jr, Lee KL, Califf RM, Pryor DB, Rosati RA (1984) Regression modelling strategies for improved prognostic prediction. Stat Med 3:143–152CrossRefPubMed
22.
Zurück zum Zitat Gal JY, Fovet Y, Adib-Yadzi M (2001) About a synthetic saliva for in vitro studies. Talanta 53:1103–1115CrossRefPubMed Gal JY, Fovet Y, Adib-Yadzi M (2001) About a synthetic saliva for in vitro studies. Talanta 53:1103–1115CrossRefPubMed
24.
Zurück zum Zitat Besimo CE, Guarneri A (2003) In vitro retention force changes of prefabricated attachments for overdentures. J Oral Rehabil 30:671–678CrossRefPubMed Besimo CE, Guarneri A (2003) In vitro retention force changes of prefabricated attachments for overdentures. J Oral Rehabil 30:671–678CrossRefPubMed
26.
Zurück zum Zitat Goiato MC, Dos Santos DM, Andreotti AM, Nobrega AS, Moreno A, Haddad MF, Pesqueira AA (2014) Effect of beverages and mouthwashes on the hardness of polymers used in intraoral prostheses. Journal of prosthodontics : official journal of the American College of Prosthodontists 23:559–564. doi:10.1111/jopr.12151 CrossRef Goiato MC, Dos Santos DM, Andreotti AM, Nobrega AS, Moreno A, Haddad MF, Pesqueira AA (2014) Effect of beverages and mouthwashes on the hardness of polymers used in intraoral prostheses. Journal of prosthodontics : official journal of the American College of Prosthodontists 23:559–564. doi:10.​1111/​jopr.​12151 CrossRef
27.
Zurück zum Zitat Goiato MC, Dos Santos DM, Baptista GT, Moreno A, Andreotti AM, Dekon SF (2013) Effect of thermal cycling and disinfection on microhardness of acrylic resin denture base. Journal of medical engineering & technology 37:203–207. doi:10.3109/03091902.2013.774444 CrossRef Goiato MC, Dos Santos DM, Baptista GT, Moreno A, Andreotti AM, Dekon SF (2013) Effect of thermal cycling and disinfection on microhardness of acrylic resin denture base. Journal of medical engineering & technology 37:203–207. doi:10.​3109/​03091902.​2013.​774444 CrossRef
30.
Zurück zum Zitat Firtell DN, Green AJ, Elahi JM (1981) Posterior peripheral seal distortion related to processing temperature. J Prosthet Dent 45:598–601CrossRefPubMed Firtell DN, Green AJ, Elahi JM (1981) Posterior peripheral seal distortion related to processing temperature. J Prosthet Dent 45:598–601CrossRefPubMed
31.
Zurück zum Zitat Sanders JL, Levin B, Reitz PV (1991) Comparison of the adaptation of acrylic resin cured by microwave energy and conventional water bath. Quintessence Int 22:181–186PubMed Sanders JL, Levin B, Reitz PV (1991) Comparison of the adaptation of acrylic resin cured by microwave energy and conventional water bath. Quintessence Int 22:181–186PubMed
32.
Zurück zum Zitat Sykora O, Sutow EJ (1993) Posterior palatal seal adaptation: influence of processing technique, palate shape and immersion. J Oral Rehabil 20:19–31CrossRefPubMed Sykora O, Sutow EJ (1993) Posterior palatal seal adaptation: influence of processing technique, palate shape and immersion. J Oral Rehabil 20:19–31CrossRefPubMed
33.
Zurück zum Zitat Woelfel JB, Paffenbarger GC, Sweeney WT (1960) Dimensional changes occurring in dentures during processing. J Am Dent Assoc 61:413–430CrossRefPubMed Woelfel JB, Paffenbarger GC, Sweeney WT (1960) Dimensional changes occurring in dentures during processing. J Am Dent Assoc 61:413–430CrossRefPubMed
34.
Zurück zum Zitat Wong DM, Cheng LY, Chow TW, Clark RK (1999) Effect of processing method on the dimensional accuracy and water sorption of acrylic resin dentures. J Prosthet Dent 81:300–304CrossRefPubMed Wong DM, Cheng LY, Chow TW, Clark RK (1999) Effect of processing method on the dimensional accuracy and water sorption of acrylic resin dentures. J Prosthet Dent 81:300–304CrossRefPubMed
Metadaten
Titel
CAD/CAM milled removable complete dentures: an in vitro evaluation of trueness
verfasst von
Murali Srinivasan
Yoann Cantin
Albert Mehl
Harald Gjengedal
Frauke Müller
Martin Schimmel
Publikationsdatum
08.11.2016
Verlag
Springer Berlin Heidelberg
Erschienen in
Clinical Oral Investigations / Ausgabe 6/2017
Print ISSN: 1432-6981
Elektronische ISSN: 1436-3771
DOI
https://doi.org/10.1007/s00784-016-1989-7

Weitere Artikel der Ausgabe 6/2017

Clinical Oral Investigations 6/2017 Zur Ausgabe

Newsletter

Bestellen Sie unseren kostenlosen Newsletter Update Zahnmedizin und bleiben Sie gut informiert – ganz bequem per eMail.