Marginal and internal fit of posterior three-unit fixed zirconia dental prostheses fabricated with two different CAD/CAM systems and materials

verfasst von: Joana Schönberger, Kurt-Jürgen Erdelt, Daniel Bäumer, Florian Beuer

Erschienen in: Clinical Oral Investigations | Ausgabe 8/2017

Einloggen, um Zugang zu erhalten

Abstract

Objectives

The purpose of this in vitro study was to compare the precision of fit of frameworks milled from semi-sintered regular zirconia and high-translucent (HT) zirconia blanks, fabricated with two different CAD/CAM systems.

Material and methods

Three-unit, posterior fixed dental prostheses (FDP) frameworks were fabricated for standardized dies (n = 11) with two different laboratory computer-aided design (CAD)/computer-aided manufacturing (CAM) systems (Cercon/Ceramill). The replica technique was used to evaluate the marginal and internal fit under an optical microscope. Evaluation of the data was performed according to prior studies at a level of significance of 5%.

Results

The systems showed a statistically significant influence on the internal fit of the frameworks (p ≤ 0.001) and on the marginal fit (p < 0.001). The type of material showed no influence on the marginal fit for the Cercon system (p = 0.636) and on the marginal fit (p = 0.064) and the internal fit (p = 0.316) for the Ceramill system, while regular zirconia from Cercon showed higher internal values than HT zirconia (p = 0.016).

Conclusions and clinical relevance

Both investigated systems showed clinically acceptable values within the limitations of this in vitro study. However, one showed less internal accuracy when regular zirconia was used.
Literatur
1.
Zurück zum Zitat Tinschert J, Natt G, Mohrbotter N, Spiekermann H, Schulze KA (2007) Lifetime of alumina- and zirconia ceramics used for crown and bridge restorations. J Biomed Mater Res B Appl Biomater 80(2):317–321. doi:10.1002/jbm.b.30599 CrossRefPubMed Tinschert J, Natt G, Mohrbotter N, Spiekermann H, Schulze KA (2007) Lifetime of alumina- and zirconia ceramics used for crown and bridge restorations. J Biomed Mater Res B Appl Biomater 80(2):317–321. doi:10.​1002/​jbm.​b.​30599 CrossRefPubMed
2.
Zurück zum Zitat Sailer I, Feher A, Filser F, Gauckler LJ, Luthy H, Hammerle CH (2007) Five-year clinical results of zirconia frameworks for posterior fixed partial dentures. Int J Prosthodont 20(4):383–388PubMed Sailer I, Feher A, Filser F, Gauckler LJ, Luthy H, Hammerle CH (2007) Five-year clinical results of zirconia frameworks for posterior fixed partial dentures. Int J Prosthodont 20(4):383–388PubMed
3.
Zurück zum Zitat Tinschert J, Natt G, Mautsch W, Augthun M, Spiekermann H (2001) Fracture resistance of lithium disilicate-, alumina-, and zirconia-based three-unit fixed partial dentures: a laboratory study. Int J Prosthodont 14(3):231–238PubMed Tinschert J, Natt G, Mautsch W, Augthun M, Spiekermann H (2001) Fracture resistance of lithium disilicate-, alumina-, and zirconia-based three-unit fixed partial dentures: a laboratory study. Int J Prosthodont 14(3):231–238PubMed
4.
Zurück zum Zitat Stawarczyk B, Emslander A, Roos M, Sener B, Noack F, Keul C (2014) Zirconia ceramics, their contrast ratio and grain size depending on sintering parameters. Dent Mater J 33(5):591–598CrossRefPubMed Stawarczyk B, Emslander A, Roos M, Sener B, Noack F, Keul C (2014) Zirconia ceramics, their contrast ratio and grain size depending on sintering parameters. Dent Mater J 33(5):591–598CrossRefPubMed
7.
Zurück zum Zitat Edelhoff D, Florian B, Florian W, Johnen C (2008) HIP zirconia fixed partial dentures—clinical results after 3 years of clinical service. Quintessence Int 39(6):459–471PubMed Edelhoff D, Florian B, Florian W, Johnen C (2008) HIP zirconia fixed partial dentures—clinical results after 3 years of clinical service. Quintessence Int 39(6):459–471PubMed
8.
Zurück zum Zitat Sachs C, Groesser J, Stadelmann M, Schweiger J, Erdelt K, Beuer F (2014) Full-arch prostheses from translucent zirconia: accuracy of fit. Dental materials: official publication of the Academy of Dental Materials 30(8):817–823. doi:10.1016/j.dental.2014.05.001 CrossRef Sachs C, Groesser J, Stadelmann M, Schweiger J, Erdelt K, Beuer F (2014) Full-arch prostheses from translucent zirconia: accuracy of fit. Dental materials: official publication of the Academy of Dental Materials 30(8):817–823. doi:10.​1016/​j.​dental.​2014.​05.​001 CrossRef
9.
Zurück zum Zitat Stawarczyk B, Ozcan M, Hallmann L, Ender A, Mehl A, Hammerlet CH (2013) The effect of zirconia sintering temperature on flexural strength, grain size, and contrast ratio. Clinical oral investigations 17(1):269–274. doi:10.1007/s00784-012-0692-6 CrossRefPubMed Stawarczyk B, Ozcan M, Hallmann L, Ender A, Mehl A, Hammerlet CH (2013) The effect of zirconia sintering temperature on flexural strength, grain size, and contrast ratio. Clinical oral investigations 17(1):269–274. doi:10.​1007/​s00784-012-0692-6 CrossRefPubMed
12.
Zurück zum Zitat Tinschert J, Schulze KA, Natt G, Latzke P, Heussen N, Spiekermann H (2008) Clinical behavior of zirconia-based fixed partial dentures made of DC-Zirkon: 3-year results. Int J Prosthodont 21(3):217–222PubMed Tinschert J, Schulze KA, Natt G, Latzke P, Heussen N, Spiekermann H (2008) Clinical behavior of zirconia-based fixed partial dentures made of DC-Zirkon: 3-year results. Int J Prosthodont 21(3):217–222PubMed
13.
Zurück zum Zitat Belser UC, MacEntee MI, Richter WA (1985) Fit of three porcelain-fused-to-metal marginal designs in vivo: a scanning electron microscope study. J Prosthet Dent 53(1):24–29CrossRefPubMed Belser UC, MacEntee MI, Richter WA (1985) Fit of three porcelain-fused-to-metal marginal designs in vivo: a scanning electron microscope study. J Prosthet Dent 53(1):24–29CrossRefPubMed
14.
Zurück zum Zitat Karlsson S (1993) The fit of Procera titanium crowns. An in vitro and clinical study. Acta Odontol Scand 51(3):129–134CrossRefPubMed Karlsson S (1993) The fit of Procera titanium crowns. An in vitro and clinical study. Acta Odontol Scand 51(3):129–134CrossRefPubMed
15.
Zurück zum Zitat McLean JW, von Fraunhofer JA (1971) The estimation of cement film thickness by an in vivo technique. Br Dent J 131(3):107–111CrossRefPubMed McLean JW, von Fraunhofer JA (1971) The estimation of cement film thickness by an in vivo technique. Br Dent J 131(3):107–111CrossRefPubMed
16.
Zurück zum Zitat Sulaiman F, Chai J, Jameson LM, Wozniak WT (1997) A comparison of the marginal fit of In-Ceram, IPS Empress, and Procera crowns. Int J Prosthodont 10(5):478–484PubMed Sulaiman F, Chai J, Jameson LM, Wozniak WT (1997) A comparison of the marginal fit of In-Ceram, IPS Empress, and Procera crowns. Int J Prosthodont 10(5):478–484PubMed
17.
Zurück zum Zitat Hung SH, Hung KS, Eick JD, Chappell RP (1990) Marginal fit of porcelain-fused-to-metal and two types of ceramic crown. J Prosthet Dent 63(1):26–31CrossRefPubMed Hung SH, Hung KS, Eick JD, Chappell RP (1990) Marginal fit of porcelain-fused-to-metal and two types of ceramic crown. J Prosthet Dent 63(1):26–31CrossRefPubMed
18.
Zurück zum Zitat Weaver JD, Johnson GH, Bales DJ (1991) Marginal adaptation of castable ceramic crowns. J Prosthet Dent 66(6):747–753CrossRefPubMed Weaver JD, Johnson GH, Bales DJ (1991) Marginal adaptation of castable ceramic crowns. J Prosthet Dent 66(6):747–753CrossRefPubMed
20.
Zurück zum Zitat Holmes JR, Bayne SC, Holland GA, Sulik WD (1989) Considerations in measurement of marginal fit. J Prosthet Dent 62(4):405–408CrossRefPubMed Holmes JR, Bayne SC, Holland GA, Sulik WD (1989) Considerations in measurement of marginal fit. J Prosthet Dent 62(4):405–408CrossRefPubMed
21.
Zurück zum Zitat Schaerer P, Sato T, Wohlwend A (1988) A comparison of the marginal fit of three cast ceramic crown systems. J Prosthet Dent 59(5):534–542CrossRefPubMed Schaerer P, Sato T, Wohlwend A (1988) A comparison of the marginal fit of three cast ceramic crown systems. J Prosthet Dent 59(5):534–542CrossRefPubMed
22.
Zurück zum Zitat Felton DA, Kanoy BE, Bayne SC, Wirthman GP (1991) Effect of in vivo crown margin discrepancies on periodontal health. J Prosthet Dent 65(3):357–364CrossRefPubMed Felton DA, Kanoy BE, Bayne SC, Wirthman GP (1991) Effect of in vivo crown margin discrepancies on periodontal health. J Prosthet Dent 65(3):357–364CrossRefPubMed
23.
Zurück zum Zitat Sorensen JA (1989) A rationale for comparison of plaque-retaining properties of crown systems. J Prosthet Dent 62(3):264–269CrossRefPubMed Sorensen JA (1989) A rationale for comparison of plaque-retaining properties of crown systems. J Prosthet Dent 62(3):264–269CrossRefPubMed
24.
Zurück zum Zitat Sorensen SE, Larsen IB, Jorgensen KD (1986) Gingival and alveolar bone reaction to marginal fit of subgingival crown margins. Scand J Dent Res 94(2):109–114PubMed Sorensen SE, Larsen IB, Jorgensen KD (1986) Gingival and alveolar bone reaction to marginal fit of subgingival crown margins. Scand J Dent Res 94(2):109–114PubMed
25.
Zurück zum Zitat Tuntiprawon M, Wilson PR (1995) The effect of cement thickness on the fracture strength of all-ceramic crowns. Aust Dent J 40(1):17–21CrossRefPubMed Tuntiprawon M, Wilson PR (1995) The effect of cement thickness on the fracture strength of all-ceramic crowns. Aust Dent J 40(1):17–21CrossRefPubMed
27.
Zurück zum Zitat Beuer F, Korczynski N, Rezac A, Naumann M, Gernet W, Sorensen JA (2010) Marginal and internal fit of zirconia based fixed dental protheses fabricated with different concepts. Clin Cosmet Investig Dent 2:5–11CrossRefPubMedPubMedCentral Beuer F, Korczynski N, Rezac A, Naumann M, Gernet W, Sorensen JA (2010) Marginal and internal fit of zirconia based fixed dental protheses fabricated with different concepts. Clin Cosmet Investig Dent 2:5–11CrossRefPubMedPubMedCentral
30.
Zurück zum Zitat Molin M, Karlsson S (1993) The fit of gold inlays and three ceramic inlay systems. A clinical and in vitro study. Acta Odontol Scand 51(4):201–206CrossRefPubMed Molin M, Karlsson S (1993) The fit of gold inlays and three ceramic inlay systems. A clinical and in vitro study. Acta Odontol Scand 51(4):201–206CrossRefPubMed
31.
Zurück zum Zitat Beuer F, Aggstaller H, Edelhoff D, Gernet W, Sorensen J (2009) Marginal and internal fits of fixed dental prostheses zirconia retainers. Dental materials: official publication of the Academy of Dental Materials 25(1):94–102. doi:10.1016/j.dental.2008.04.018 CrossRef Beuer F, Aggstaller H, Edelhoff D, Gernet W, Sorensen J (2009) Marginal and internal fits of fixed dental prostheses zirconia retainers. Dental materials: official publication of the Academy of Dental Materials 25(1):94–102. doi:10.​1016/​j.​dental.​2008.​04.​018 CrossRef
32.
Zurück zum Zitat Tinschert J, Natt G, Mautsch W, Spiekermann H, Anusavice KJ (2001) Marginal fit of alumina- and zirconia-based fixed partial dentures produced by a CAD/CAM system. Oper Dent 26(4):367–374PubMed Tinschert J, Natt G, Mautsch W, Spiekermann H, Anusavice KJ (2001) Marginal fit of alumina- and zirconia-based fixed partial dentures produced by a CAD/CAM system. Oper Dent 26(4):367–374PubMed
33.
Zurück zum Zitat Coli P, Karlsson S (2004) Precision of a CAD/CAM technique for the production of zirconium dioxide copings. Int J Prosthodont 17(5):577–580PubMed Coli P, Karlsson S (2004) Precision of a CAD/CAM technique for the production of zirconium dioxide copings. Int J Prosthodont 17(5):577–580PubMed
34.
Zurück zum Zitat Rekow ED, Harsono M, Janal M, Thompson VP, Zhang G (2006) Factorial analysis of variables influencing stress in all-ceramic crowns. Dental materials: official publication of the Academy of Dental Materials 22(2):125–132. doi:10.1016/j.dental.2005.04.010 CrossRef Rekow ED, Harsono M, Janal M, Thompson VP, Zhang G (2006) Factorial analysis of variables influencing stress in all-ceramic crowns. Dental materials: official publication of the Academy of Dental Materials 22(2):125–132. doi:10.​1016/​j.​dental.​2005.​04.​010 CrossRef
35.
Zurück zum Zitat Rekow D, Thompson VP (2005) Near-surface damage—a persistent problem in crowns obtained by computer-aided design and manufacturing. Proc Inst Mech Eng H J Eng Med 219(4):233–243CrossRef Rekow D, Thompson VP (2005) Near-surface damage—a persistent problem in crowns obtained by computer-aided design and manufacturing. Proc Inst Mech Eng H J Eng Med 219(4):233–243CrossRef
Metadaten
Titel
Marginal and internal fit of posterior three-unit fixed zirconia dental prostheses fabricated with two different CAD/CAM systems and materials
verfasst von
Joana Schönberger
Kurt-Jürgen Erdelt
Daniel Bäumer
Florian Beuer
Publikationsdatum
04.02.2017
Verlag
Springer Berlin Heidelberg
Erschienen in
Clinical Oral Investigations / Ausgabe 8/2017
Print ISSN: 1432-6981
Elektronische ISSN: 1436-3771
DOI
https://doi.org/10.1007/s00784-017-2064-8

Weitere Artikel der Ausgabe 8/2017

Clinical Oral Investigations 8/2017 Zur Ausgabe

Newsletter

Bestellen Sie unseren kostenlosen Newsletter Update Zahnmedizin und bleiben Sie gut informiert – ganz bequem per eMail.