Novel control system to prevent right ventricular failure induced by rotary blood pump

verfasst von: Mamoru Arakawa, Takashi Nishimura, Yoshiaki Takewa, Akihide Umeki, Masahiko Ando, Yuichiro Kishimoto, Yutaka Fujii, Shunei Kyo, Hideo Adachi, Eisuke Tatsumi

Erschienen in: Journal of Artificial Organs | Ausgabe 2/2014

Einloggen, um Zugang zu erhalten

Abstract

Right ventricular (RV) failure is a potentially fatal complication after treatment with a left ventricular assist device (LVAD). Ventricular septal shift caused by such devices is an important factor in the progress of RV dysfunction. We developed a control system for a rotary blood pump that can change rotational speed (RS) in synchrony with the cardiac cycle. We postulated that decreasing systolic RS using this system would alter ventricular septal movement and thus prevent RV failure. We implanted the EVAHEART ventricular assist device into seven adult goats weighing 54.1 ± 2.1 kg and induced acute bi-ventricular dysfunction by coronary embolization. Left and RV pressure was monitored, and ventricular septal movement was echocardiographically determined. We evaluated circuit-clamp mode as the control condition, as well as continuous and counter-pulse modes, both with full bypass. As a result, a leftward ventricular septal shift occurred in continuous and counter-pulse modes. The septal shift was corrected as a result of decreased RS during the systolic phase in counter-pulse mode. RV fractional area change improved in counter-pulse (59.0 ± 4.6 %) compared with continuous (44.7 ± 4.0 %) mode. In conclusion, decreased RS delivered during the systolic phase using the counter-pulse mode of our new system holds promise for the clinical correction of ventricular septal shift resulting from a LVAD and might confer a benefit upon RV function.
Literatur
1.
Zurück zum Zitat Miller LW, Pagani FD, Russell SD, John R, Boyle AJ, Aaronson KD, Conte JV, Naka Y, Mancini D, Delgado RM, MacGillivray TE, Farrar DJ, Frazier OH, HeartMate II, Clinical investigators. Use of a continuous-flow device in patients awaiting heart transplantation. N Engl J Med. 2007;357:885–96.PubMedCrossRef Miller LW, Pagani FD, Russell SD, John R, Boyle AJ, Aaronson KD, Conte JV, Naka Y, Mancini D, Delgado RM, MacGillivray TE, Farrar DJ, Frazier OH, HeartMate II, Clinical investigators. Use of a continuous-flow device in patients awaiting heart transplantation. N Engl J Med. 2007;357:885–96.PubMedCrossRef
2.
Zurück zum Zitat Lietz K, Long JW, Kfoury AG, Slaughter MS, Silver MA, Milano CA, Rogers JG, Naka Y, Mancini D, Miller LW. Outcomes of left ventricular assist device implantation as destination therapy in the post-REMATCH era: implications for patient selection. Circulation. 2007;116:497–505.PubMedCrossRef Lietz K, Long JW, Kfoury AG, Slaughter MS, Silver MA, Milano CA, Rogers JG, Naka Y, Mancini D, Miller LW. Outcomes of left ventricular assist device implantation as destination therapy in the post-REMATCH era: implications for patient selection. Circulation. 2007;116:497–505.PubMedCrossRef
3.
Zurück zum Zitat Slaughter MS, Rogers JG, Milano CA, Russell SD, Conte JV, Feldman D, Sun B, Tatooles AJ, Delgado RM 3rd, Long JW, Wozniak TC, Ghumman W, Farrar DJ, Frazier OH, HeartMate II Investigators. Advanced heart failure treated with continuous-flow left ventricular assist device. N Engl J Med. 2009;361:2241–51.PubMedCrossRef Slaughter MS, Rogers JG, Milano CA, Russell SD, Conte JV, Feldman D, Sun B, Tatooles AJ, Delgado RM 3rd, Long JW, Wozniak TC, Ghumman W, Farrar DJ, Frazier OH, HeartMate II Investigators. Advanced heart failure treated with continuous-flow left ventricular assist device. N Engl J Med. 2009;361:2241–51.PubMedCrossRef
4.
Zurück zum Zitat Garatti A, Bruschi G, Colombo T, Russo C, Lanfranconi M, Milazzo F, Frigerio M, Vitali E. Clinical outcome and bridge to transplant rate of left ventricular assist device recipient patients: comparison between continuous-flow and pulsatile-flow devices. Eur J Cardio Thorac Surg. 2008;34:275–80.CrossRef Garatti A, Bruschi G, Colombo T, Russo C, Lanfranconi M, Milazzo F, Frigerio M, Vitali E. Clinical outcome and bridge to transplant rate of left ventricular assist device recipient patients: comparison between continuous-flow and pulsatile-flow devices. Eur J Cardio Thorac Surg. 2008;34:275–80.CrossRef
5.
Zurück zum Zitat Koenig SC, Pantalos GM, Gillars KJ, Ewert DL, Litwak KN, Etoch SW. Hemodynamic and pressure-volume responses to continuous and pulsatile ventricular assist in an adult mock circulation. ASAIO J. 2004;50:15–24.PubMedCrossRef Koenig SC, Pantalos GM, Gillars KJ, Ewert DL, Litwak KN, Etoch SW. Hemodynamic and pressure-volume responses to continuous and pulsatile ventricular assist in an adult mock circulation. ASAIO J. 2004;50:15–24.PubMedCrossRef
6.
Zurück zum Zitat Nishinaka T, Tatsumi E, Nishimura T, Taenaka Y, Masuzawa T, Nakata M, Takano H, Koyanagi H. Cardiac autonomic nervous function during long-term nonpulsatile left heart bypass. Artif Organs. 1999;23:500–3.PubMedCrossRef Nishinaka T, Tatsumi E, Nishimura T, Taenaka Y, Masuzawa T, Nakata M, Takano H, Koyanagi H. Cardiac autonomic nervous function during long-term nonpulsatile left heart bypass. Artif Organs. 1999;23:500–3.PubMedCrossRef
7.
Zurück zum Zitat Patel ND, Weiss ES, Schaffer J, Ullrich SL, Rivard DC, Shah AS, Russell SD, Conte JV. Right heart dysfunction after left ventricular assist device implantation: a comparison of the pulsatile HeartMate I and axial-flow HeartMate II devices. Ann Thorac Surg. 2008;86:832–40.PubMedCrossRef Patel ND, Weiss ES, Schaffer J, Ullrich SL, Rivard DC, Shah AS, Russell SD, Conte JV. Right heart dysfunction after left ventricular assist device implantation: a comparison of the pulsatile HeartMate I and axial-flow HeartMate II devices. Ann Thorac Surg. 2008;86:832–40.PubMedCrossRef
8.
Zurück zum Zitat Baumwol J, Macdonald PS, Keogh AM, Kotlyar E, Spratt P, Jansz P, Hayward CS. Right heart failure and “failure to thrive” after left ventricular assist device: clinical predictors and outcomes. J Heart Lung Transplant. 2011;30:888–95.PubMed Baumwol J, Macdonald PS, Keogh AM, Kotlyar E, Spratt P, Jansz P, Hayward CS. Right heart failure and “failure to thrive” after left ventricular assist device: clinical predictors and outcomes. J Heart Lung Transplant. 2011;30:888–95.PubMed
9.
Zurück zum Zitat Omoto T, Tanabe H, LaRia PJ, Guererro J, Vlahakes GJ. Right ventricular performance during left ventricular unloading conditions: the contribution of the right ventricular free wall. Thorac Cardiovasc Surg. 2002;50:16–20.PubMedCrossRef Omoto T, Tanabe H, LaRia PJ, Guererro J, Vlahakes GJ. Right ventricular performance during left ventricular unloading conditions: the contribution of the right ventricular free wall. Thorac Cardiovasc Surg. 2002;50:16–20.PubMedCrossRef
10.
Zurück zum Zitat Park CH, Nishimura K, Kitano M, Okamoto Y, Ban T. Right ventricular performance is impaired by full assist of left heart bypass. Analysis of right ventricular performance against change in afterload in heart failure models. ASAIO J. 1994;40:M303–8.PubMedCrossRef Park CH, Nishimura K, Kitano M, Okamoto Y, Ban T. Right ventricular performance is impaired by full assist of left heart bypass. Analysis of right ventricular performance against change in afterload in heart failure models. ASAIO J. 1994;40:M303–8.PubMedCrossRef
11.
Zurück zum Zitat Kitano M, Nishimura K, Hee PC, Okamoto Y, Ban T. Right ventricular function evaluated by volumetric analysis during left heart bypass in a canine model of postischemic cardiac dysfunction. J Thorac Cardiovasc Surg. 1995;109:796–803.PubMedCrossRef Kitano M, Nishimura K, Hee PC, Okamoto Y, Ban T. Right ventricular function evaluated by volumetric analysis during left heart bypass in a canine model of postischemic cardiac dysfunction. J Thorac Cardiovasc Surg. 1995;109:796–803.PubMedCrossRef
12.
Zurück zum Zitat Daly RC, Chandrasekaran K, Cavarocchi NC, Tajik AJ, Schaff HV. Ischemia of the interventricular septum. A mechanism of right ventricular failure during mechanical left ventricular assist. J Thorac Cardiovasc Surg. 1992;103:1186–91.PubMed Daly RC, Chandrasekaran K, Cavarocchi NC, Tajik AJ, Schaff HV. Ischemia of the interventricular septum. A mechanism of right ventricular failure during mechanical left ventricular assist. J Thorac Cardiovasc Surg. 1992;103:1186–91.PubMed
13.
Zurück zum Zitat Yoshioka D, Sakaguchi T, Saito S, Miyagawa S, Nishi H, Yoshikawa Y, Fukushima S, Saito T, Daimon T, Ueno T, Kuratani T, Sawa Y. Predictor of early mortality for severe heart failure patients with left ventricular assist device implantation: significance of INTERMACS level and renal function. Circ J. 2012;76:1631–8.PubMedCrossRef Yoshioka D, Sakaguchi T, Saito S, Miyagawa S, Nishi H, Yoshikawa Y, Fukushima S, Saito T, Daimon T, Ueno T, Kuratani T, Sawa Y. Predictor of early mortality for severe heart failure patients with left ventricular assist device implantation: significance of INTERMACS level and renal function. Circ J. 2012;76:1631–8.PubMedCrossRef
14.
Zurück zum Zitat Yamazaki K, Kihara S, Akimoto T, Tagusari O, Kawai A, Umezu M, Tomioka J, Kormos RL, Griffith BP, Kurosawa H. EVAHEART: an implantable centrifugal blood pump for long-term circulatory support. Jpn J Thorac Cardiovasc Surg. 2002;50:461–5.PubMedCrossRef Yamazaki K, Kihara S, Akimoto T, Tagusari O, Kawai A, Umezu M, Tomioka J, Kormos RL, Griffith BP, Kurosawa H. EVAHEART: an implantable centrifugal blood pump for long-term circulatory support. Jpn J Thorac Cardiovasc Surg. 2002;50:461–5.PubMedCrossRef
15.
Zurück zum Zitat Ando M, Nishimura T, Takewa Y, Ogawa D, Yamazaki K, Kashiwa K, Kyo S, Ono M, Taenaka Y, Tatsumi E. A novel counterpulse drive mode of continuous-flow left ventricular assist devices can minimize intracircuit backward flow during pump weaning. J Artif Organs. 2011;14:74–9.PubMedCrossRef Ando M, Nishimura T, Takewa Y, Ogawa D, Yamazaki K, Kashiwa K, Kyo S, Ono M, Taenaka Y, Tatsumi E. A novel counterpulse drive mode of continuous-flow left ventricular assist devices can minimize intracircuit backward flow during pump weaning. J Artif Organs. 2011;14:74–9.PubMedCrossRef
16.
Zurück zum Zitat Ando M, Nishimura T, Takewa Y, et al. Electrocardiogram-synchronized rotational speed change mode in rotary pumps could improve pulsatility. Artif Organs. 2011;35:941–7.PubMedCrossRef Ando M, Nishimura T, Takewa Y, et al. Electrocardiogram-synchronized rotational speed change mode in rotary pumps could improve pulsatility. Artif Organs. 2011;35:941–7.PubMedCrossRef
17.
Zurück zum Zitat Ando M, Takewa Y, Nishimura T, Yamazaki K, Kyo S, Ono M, Tsukiya T, Mizuno T, Taenaka Y, Tatsumi E. A novel counterpulsation mode of rotary left ventricular assist devices can enhance myocardial perfusion. J Artif Organs. 2011;14:185–91.PubMedCrossRef Ando M, Takewa Y, Nishimura T, Yamazaki K, Kyo S, Ono M, Tsukiya T, Mizuno T, Taenaka Y, Tatsumi E. A novel counterpulsation mode of rotary left ventricular assist devices can enhance myocardial perfusion. J Artif Organs. 2011;14:185–91.PubMedCrossRef
18.
Zurück zum Zitat Umeki A, Nishimura T, Ando M, Takewa Y, Yamazaki K, Kyo S, Ono M, Tsukiya T, Mizuno T, Taenaka Y, Tatsumi E. Alteration of LV end-diastolic volume by controlling the power of the continuous-flow LVAD, so it is synchronized with cardiac beat: development of a native heart load control system (NHLCS). J Artif Organs. 2012;15:128–33.PubMedCrossRef Umeki A, Nishimura T, Ando M, Takewa Y, Yamazaki K, Kyo S, Ono M, Tsukiya T, Mizuno T, Taenaka Y, Tatsumi E. Alteration of LV end-diastolic volume by controlling the power of the continuous-flow LVAD, so it is synchronized with cardiac beat: development of a native heart load control system (NHLCS). J Artif Organs. 2012;15:128–33.PubMedCrossRef
19.
Zurück zum Zitat Kishimoto Y, Takewa Y, Arakawa M, Umeki A, Ando M, Nishimura T, Fujii Y, Mizuno T, Nishimura M, Tatsumi E. Development of a novel drive mode to prevent aortic insufficiency during continuous-flow LVAD support by synchronizing rotational speed with heartbeat. J Artif Organs. 2013;16:129–37.PubMedCrossRef Kishimoto Y, Takewa Y, Arakawa M, Umeki A, Ando M, Nishimura T, Fujii Y, Mizuno T, Nishimura M, Tatsumi E. Development of a novel drive mode to prevent aortic insufficiency during continuous-flow LVAD support by synchronizing rotational speed with heartbeat. J Artif Organs. 2013;16:129–37.PubMedCrossRef
Metadaten
Titel
Novel control system to prevent right ventricular failure induced by rotary blood pump
verfasst von
Mamoru Arakawa
Takashi Nishimura
Yoshiaki Takewa
Akihide Umeki
Masahiko Ando
Yuichiro Kishimoto
Yutaka Fujii
Shunei Kyo
Hideo Adachi
Eisuke Tatsumi
Publikationsdatum
01.06.2014
Verlag
Springer Japan
Erschienen in
Journal of Artificial Organs / Ausgabe 2/2014
Print ISSN: 1434-7229
Elektronische ISSN: 1619-0904
DOI
https://doi.org/10.1007/s10047-014-0757-1

Weitere Artikel der Ausgabe 2/2014

Journal of Artificial Organs 2/2014 Zur Ausgabe

Update Chirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.

S3-Leitlinie „Diagnostik und Therapie des Karpaltunnelsyndroms“

CME: 2 Punkte

Prof. Dr. med. Gregor Antoniadis Das Karpaltunnelsyndrom ist die häufigste Kompressionsneuropathie peripherer Nerven. Obwohl die Anamnese mit dem nächtlichen Einschlafen der Hand (Brachialgia parästhetica nocturna) sehr typisch ist, ist eine klinisch-neurologische Untersuchung und Elektroneurografie in manchen Fällen auch eine Neurosonografie erforderlich. Im Anfangsstadium sind konservative Maßnahmen (Handgelenksschiene, Ergotherapie) empfehlenswert. Bei nicht Ansprechen der konservativen Therapie oder Auftreten von neurologischen Ausfällen ist eine Dekompression des N. medianus am Karpaltunnel indiziert.

Prof. Dr. med. Gregor Antoniadis
Berufsverband der Deutschen Chirurgie e.V.

S2e-Leitlinie „Distale Radiusfraktur“

CME: 2 Punkte

Dr. med. Benjamin Meyknecht, PD Dr. med. Oliver Pieske Das Webinar S2e-Leitlinie „Distale Radiusfraktur“ beschäftigt sich mit Fragen und Antworten zu Diagnostik und Klassifikation sowie Möglichkeiten des Ausschlusses von Zusatzverletzungen. Die Referenten erläutern, welche Frakturen konservativ behandelt werden können und wie. Das Webinar beantwortet die Frage nach aktuellen operativen Therapiekonzepten: Welcher Zugang, welches Osteosynthesematerial? Auf was muss bei der Nachbehandlung der distalen Radiusfraktur geachtet werden?

PD Dr. med. Oliver Pieske
Dr. med. Benjamin Meyknecht
Berufsverband der Deutschen Chirurgie e.V.

S1-Leitlinie „Empfehlungen zur Therapie der akuten Appendizitis bei Erwachsenen“

CME: 2 Punkte

Dr. med. Mihailo Andric
Inhalte des Webinars zur S1-Leitlinie „Empfehlungen zur Therapie der akuten Appendizitis bei Erwachsenen“ sind die Darstellung des Projektes und des Erstellungswegs zur S1-Leitlinie, die Erläuterung der klinischen Relevanz der Klassifikation EAES 2015, die wissenschaftliche Begründung der wichtigsten Empfehlungen und die Darstellung stadiengerechter Therapieoptionen.

Dr. med. Mihailo Andric
Berufsverband der Deutschen Chirurgie e.V.