Skip to main content
Erschienen in: Odontology 2/2014

01.07.2014 | Original Article

Canal shaping with WaveOne reciprocating files: influence of operator experience on instrument breakage and canal preparation time

verfasst von: Luigi Generali, Elena Righi, Maria Vittoria Todesca, Ugo Consolo

Erschienen in: Odontology | Ausgabe 2/2014

Einloggen, um Zugang zu erhalten

Abstract

The aim of this study was to determine the fracture incidence of WaveOne Primary reciprocating files and the time required for shaping of curved canals based on the experience of the operator. A total of 109 mesiobuccal canals of permanent molars extracted with an angle of curvature of >25°–45° according to Schneider were randomly assigned to four groups. An experienced operator (endodontist) and an inexperienced operator (student) each shaped one of two groups: one with the instrument WaveOne Primary to WL and the other after creation of a glide path with PathFile 1, 2 and 3 at the WL. Any fractures or visible deformations of the instruments during the shaping phase and the effective time required to prepare the canals for each instrument were recorded. No visible deformation or fracture was observed. The experienced operator tended to finish their shape faster than the inexperienced operator regardless of the technique applied. For the inexperienced operator, the usage time with only WaveOne Primary was significantly lower when the canals were preliminarily instrumented with the PathFile than when these instruments were not used (average time, 22.03 vs. 36.22 s, respectively; p < 0.001). The experience of the operator did not influence fracture of the WaveOne Primary instruments. The time required to prepare the canals was instead inversely proportional to the experience of the operator. However, the creation of a glide path with PathFile instruments reduced the time required by the inexperienced operator to prepare the canal.
Literatur
1.
Zurück zum Zitat Berutti E, Cantatore G, Castellucci A, Chiandussi G, Pera F, Migliaretti G, Pasqualini D. Use of nickel titanium rotary PathFile to create the glide path: comparison with manual preflaring in simulated root canals. J Endod. 2009;35:408–12.CrossRef Berutti E, Cantatore G, Castellucci A, Chiandussi G, Pera F, Migliaretti G, Pasqualini D. Use of nickel titanium rotary PathFile to create the glide path: comparison with manual preflaring in simulated root canals. J Endod. 2009;35:408–12.CrossRef
2.
Zurück zum Zitat Pasqualini D, Bianchi CC, Paolino DS, Mancini L, Cemenasco A, Cantatore G, Castellucci A, Berutti E. Computed micro-tomographic evaluation of glide path with nickel-titanium rotary PathFile in maxillary first molars curved canals. J Endod. 2012;38:389–93.CrossRef Pasqualini D, Bianchi CC, Paolino DS, Mancini L, Cemenasco A, Cantatore G, Castellucci A, Berutti E. Computed micro-tomographic evaluation of glide path with nickel-titanium rotary PathFile in maxillary first molars curved canals. J Endod. 2012;38:389–93.CrossRef
3.
Zurück zum Zitat Glossen CR, Haller RH, Dove SB, del Rio CE. A comparison of root canal preparations using NiTi hand, NiTi engine-driven, and K-Flex endodontic instruments. J Endod. 1995;25:434–40. Glossen CR, Haller RH, Dove SB, del Rio CE. A comparison of root canal preparations using NiTi hand, NiTi engine-driven, and K-Flex endodontic instruments. J Endod. 1995;25:434–40.
4.
Zurück zum Zitat Setzer FC, Kwon TK, Karabucak B. Comparison of apical transportation between two rotary file systems and two hybrid rotary instrumentation sequences. J Endod. 2010;36:1991–4.CrossRef Setzer FC, Kwon TK, Karabucak B. Comparison of apical transportation between two rotary file systems and two hybrid rotary instrumentation sequences. J Endod. 2010;36:1991–4.CrossRef
5.
Zurück zum Zitat Cheung GSP, Liu CSY. A retrospective study of endodontic treatment outcomes between nickel-titanium rotary and stainless steel hand filing techniques. J Endod. 2009;35:938–43.CrossRef Cheung GSP, Liu CSY. A retrospective study of endodontic treatment outcomes between nickel-titanium rotary and stainless steel hand filing techniques. J Endod. 2009;35:938–43.CrossRef
6.
Zurück zum Zitat Sattapan B, Nervo GJ, Palamara JE, Messer HH. Defects in rotary nickel-titanium files after clinical use. J Endod. 2000;26:161–5.CrossRef Sattapan B, Nervo GJ, Palamara JE, Messer HH. Defects in rotary nickel-titanium files after clinical use. J Endod. 2000;26:161–5.CrossRef
7.
Zurück zum Zitat Parashos P, Messer HH. Rotary NiTi instruments fracture and its conseguences. J Endod. 2006;32:1031–43.CrossRef Parashos P, Messer HH. Rotary NiTi instruments fracture and its conseguences. J Endod. 2006;32:1031–43.CrossRef
8.
Zurück zum Zitat Jamleh A, Kobayashi C, Yahata Y, Ebihara A, Suda H. Deflecting load of nickel titanium rotary instruments during cyclic fatigue. Dent Mater J. 2012;31:389–93.CrossRef Jamleh A, Kobayashi C, Yahata Y, Ebihara A, Suda H. Deflecting load of nickel titanium rotary instruments during cyclic fatigue. Dent Mater J. 2012;31:389–93.CrossRef
9.
Zurück zum Zitat Martin B, Zelada G, Varela P, Bahillo JG, Magan F, Ahn S, Rodriguez C. Factors influencing the fracture of nickel-titanium rotary instruments. Int Endod J. 2003;36:262–6. Martin B, Zelada G, Varela P, Bahillo JG, Magan F, Ahn S, Rodriguez C. Factors influencing the fracture of nickel-titanium rotary instruments. Int Endod J. 2003;36:262–6.
10.
Zurück zum Zitat Cheung GSP, Peng B, Bian Z, Shen Y, Darvell BW. Defects in ProTaper S1 instruments after clinical use: fractographic examination. Int Endod J. 2005;38:802–9. Cheung GSP, Peng B, Bian Z, Shen Y, Darvell BW. Defects in ProTaper S1 instruments after clinical use: fractographic examination. Int Endod J. 2005;38:802–9.
11.
Zurück zum Zitat Kramkowski TR, Bahcall J. An in vitro comparison of torsional stress and cyclic fatigue resistance of ProFile GT and ProFile GT series X rotary nickel-titanium files. J Endod. 2009;35:404–7.CrossRef Kramkowski TR, Bahcall J. An in vitro comparison of torsional stress and cyclic fatigue resistance of ProFile GT and ProFile GT series X rotary nickel-titanium files. J Endod. 2009;35:404–7.CrossRef
12.
Zurück zum Zitat Blum JY, Machtou P, Ruddle C, Micallef JP. Analysis of mechanical preparations in extracted teeth using ProTaper rotary instruments: value of the safety quotient. J Endod. 2003;29:567–75.CrossRef Blum JY, Machtou P, Ruddle C, Micallef JP. Analysis of mechanical preparations in extracted teeth using ProTaper rotary instruments: value of the safety quotient. J Endod. 2003;29:567–75.CrossRef
13.
Zurück zum Zitat Patino PV, Biedma BM, Liebana CR, Cantatore G, Bahilo JG. The influence of a manual glide path on the separation rate of NiTi rotary instruments. J Endod. 2005;31:114–6.CrossRef Patino PV, Biedma BM, Liebana CR, Cantatore G, Bahilo JG. The influence of a manual glide path on the separation rate of NiTi rotary instruments. J Endod. 2005;31:114–6.CrossRef
14.
Zurück zum Zitat Berutti E, Paolino DS, Chiandussi G, Alovisi M, Cantatore G, Castellucci A, Pasqualini D. Root canal anatomy preservation of WaveOne reciprocating files with or without glide path. J Endod. 2012;38:101–4.CrossRef Berutti E, Paolino DS, Chiandussi G, Alovisi M, Cantatore G, Castellucci A, Pasqualini D. Root canal anatomy preservation of WaveOne reciprocating files with or without glide path. J Endod. 2012;38:101–4.CrossRef
15.
Zurück zum Zitat Berutti E, Negro AR, Lendini M, Pasqualini D. Influence of manual preflaring and torque on the failure rate of ProTaper rotary instruments. J Endod. 2004;30:228–30.CrossRef Berutti E, Negro AR, Lendini M, Pasqualini D. Influence of manual preflaring and torque on the failure rate of ProTaper rotary instruments. J Endod. 2004;30:228–30.CrossRef
16.
Zurück zum Zitat Wan J, Rasimick BJ, Musikant BL, Deutsch AS. A comparison of cyclic fatigue resistance in reciprocating and rotary nickel-titanium instruments. Aust Endod J. 2011;37:122–7.CrossRef Wan J, Rasimick BJ, Musikant BL, Deutsch AS. A comparison of cyclic fatigue resistance in reciprocating and rotary nickel-titanium instruments. Aust Endod J. 2011;37:122–7.CrossRef
17.
Zurück zum Zitat Gambarini G, Gergi R, Naaman A, Osta N, Al Sudani D. Cyclic fatigue analysis of twisted file rotary NiTi instruments used in reciprocating motion. Int Endod J. 2012;45:802–6. Gambarini G, Gergi R, Naaman A, Osta N, Al Sudani D. Cyclic fatigue analysis of twisted file rotary NiTi instruments used in reciprocating motion. Int Endod J. 2012;45:802–6.
18.
Zurück zum Zitat You SY, Bae KS, Baek SH, Kum KY, Shon WJ, Lee W. Lifespan of one nickel-titanium rotary file with reciprocating motion in curved root canals. J Endod. 2010;36:1991–4.CrossRef You SY, Bae KS, Baek SH, Kum KY, Shon WJ, Lee W. Lifespan of one nickel-titanium rotary file with reciprocating motion in curved root canals. J Endod. 2010;36:1991–4.CrossRef
19.
Zurück zum Zitat Schäfer E, Tepel J. Relationship between design features of endodontic instruments and their properties. Part 3 Resistance to bending and fracture. J Endod. 2001;27:299–303.CrossRef Schäfer E, Tepel J. Relationship between design features of endodontic instruments and their properties. Part 3 Resistance to bending and fracture. J Endod. 2001;27:299–303.CrossRef
20.
Zurück zum Zitat Berutti E, Chiandussi G, Gaviglio I, Ibba A. Comparative analysis of torsional an bending stresses in two mathematical models of nickel-titanium rotary instruments: ProTaper versus ProFile. J Endod. 2003;29:15–9.CrossRef Berutti E, Chiandussi G, Gaviglio I, Ibba A. Comparative analysis of torsional an bending stresses in two mathematical models of nickel-titanium rotary instruments: ProTaper versus ProFile. J Endod. 2003;29:15–9.CrossRef
21.
Zurück zum Zitat Mandel E, Adib-Yazdi M, Benhamou LM, Lachkar T, Mesgouez C, Sobel M. Rotary Ni-Ti profile systems for preparing curved canals in resin blocks: influence of operator on instrument breakage. Int Endod J. 1999;32:436–43. Mandel E, Adib-Yazdi M, Benhamou LM, Lachkar T, Mesgouez C, Sobel M. Rotary Ni-Ti profile systems for preparing curved canals in resin blocks: influence of operator on instrument breakage. Int Endod J. 1999;32:436–43.
22.
Zurück zum Zitat Yared GM, Bou Dagher FE, Machtou P. Influence of rotational speed, torque and operator proficiency on ProFile failures. Int Endod J. 2001;34:47–53. Yared GM, Bou Dagher FE, Machtou P. Influence of rotational speed, torque and operator proficiency on ProFile failures. Int Endod J. 2001;34:47–53.
23.
Zurück zum Zitat Yared GM, Bou Dagher FE, Machtou P, Kulkarni GK. Influence of rotational speed, torque and operator proficiency on failure of Greater Taper files. Int Endod J. 2002;35:7–12. Yared GM, Bou Dagher FE, Machtou P, Kulkarni GK. Influence of rotational speed, torque and operator proficiency on failure of Greater Taper files. Int Endod J. 2002;35:7–12.
24.
Zurück zum Zitat Ye J, Gao Y. Metallurgical characterization of M-Wire nickel-titanium shape memory alloy used for endodontic rotary instruments during low-cycle fatigue. J Endod. 2012;38:105–7.CrossRef Ye J, Gao Y. Metallurgical characterization of M-Wire nickel-titanium shape memory alloy used for endodontic rotary instruments during low-cycle fatigue. J Endod. 2012;38:105–7.CrossRef
25.
Zurück zum Zitat Webber J, Machtou P, Pertot W, Kuttler S, Ruddle C, West J. The WaveOne single-file reciprocating system. Roots. 2011;1:28–33. Webber J, Machtou P, Pertot W, Kuttler S, Ruddle C, West J. The WaveOne single-file reciprocating system. Roots. 2011;1:28–33.
26.
Zurück zum Zitat Berutti E, Chiandussi G, Paolino DS, Scotti N, Cantatore G, Castellucci A, Pasqualini D. Effect of canal lenght and curvature on working lenght alteration with WaveOne reciprocating file. J Endod. 2011;37:1687–90.CrossRef Berutti E, Chiandussi G, Paolino DS, Scotti N, Cantatore G, Castellucci A, Pasqualini D. Effect of canal lenght and curvature on working lenght alteration with WaveOne reciprocating file. J Endod. 2011;37:1687–90.CrossRef
27.
Zurück zum Zitat Schneider SW. A comparison of canal preparations in straight and curved root canals. Oral Surg Oral Med Oral Pathol. 1971;32:271–5.CrossRef Schneider SW. A comparison of canal preparations in straight and curved root canals. Oral Surg Oral Med Oral Pathol. 1971;32:271–5.CrossRef
28.
Zurück zum Zitat Roane JB, Sabala CL, Duncanson MG Jr. The “balanced force” concept for instrumentation of curved canals. J Endod. 1985;11:203–11.CrossRef Roane JB, Sabala CL, Duncanson MG Jr. The “balanced force” concept for instrumentation of curved canals. J Endod. 1985;11:203–11.CrossRef
29.
Zurück zum Zitat Varela-Patino P, Ibanez-Parraga A, Rivas-Mundina B, Cantatore G, Otero XL, Martin-Biedma B. Alternating versus continuous rotation: a comparative study of the effect on instrument life. J Endod. 2010;36:157–9.CrossRef Varela-Patino P, Ibanez-Parraga A, Rivas-Mundina B, Cantatore G, Otero XL, Martin-Biedma B. Alternating versus continuous rotation: a comparative study of the effect on instrument life. J Endod. 2010;36:157–9.CrossRef
30.
Zurück zum Zitat Gavini G, Caldeira CL, Akisue E, Candeiro GT, Kawakami DA. Resistance to flexural fatigue of Reciproc R25 files under continuous rotation and reciprocating motion. J Endod. 2012;38:684–7.CrossRef Gavini G, Caldeira CL, Akisue E, Candeiro GT, Kawakami DA. Resistance to flexural fatigue of Reciproc R25 files under continuous rotation and reciprocating motion. J Endod. 2012;38:684–7.CrossRef
31.
Zurück zum Zitat Johnson E, Lloyd A, Kuttler S, Namerow K. Comparison between a novel nickel-titanium alloy and 508 nitinol on the cyclic fatigue life of ProFile 25/.04 rotary instruments. J Endod. 2008;34:1406–9.CrossRef Johnson E, Lloyd A, Kuttler S, Namerow K. Comparison between a novel nickel-titanium alloy and 508 nitinol on the cyclic fatigue life of ProFile 25/.04 rotary instruments. J Endod. 2008;34:1406–9.CrossRef
32.
Zurück zum Zitat Bouska J, Justman B, Williamson A, DeLong C, Qian F. Resistance to cyclic fatigue of a new endodontic rotary file. J Endod. 2012;38:667–9.CrossRef Bouska J, Justman B, Williamson A, DeLong C, Qian F. Resistance to cyclic fatigue of a new endodontic rotary file. J Endod. 2012;38:667–9.CrossRef
33.
Zurück zum Zitat Yared G. Canal preparation using only one Ni-Ti rotary instruments: preliminary observations. Int Endod J. 2008;41:339–44. Yared G. Canal preparation using only one Ni-Ti rotary instruments: preliminary observations. Int Endod J. 2008;41:339–44.
34.
Zurück zum Zitat Bürklein S, Hinschitza K, Dammaschke T, Schäfer E. Shaping ability and cleaning effectiveness of two single-file systems in severely curved root canals of extracted teeth: reciproc and WaveOne versus Mtwo and ProTaper. Int Endod J. 2012;45:449–61. Bürklein S, Hinschitza K, Dammaschke T, Schäfer E. Shaping ability and cleaning effectiveness of two single-file systems in severely curved root canals of extracted teeth: reciproc and WaveOne versus Mtwo and ProTaper. Int Endod J. 2012;45:449–61.
35.
Zurück zum Zitat Baumann MA. WaveOne—First experiences of third-year students. Roots. 2012;2:20–3. Baumann MA. WaveOne—First experiences of third-year students. Roots. 2012;2:20–3.
36.
Zurück zum Zitat Plotino G, Grande NM, Testarelli L, Gambarini G. Cyclic fatigue of Reciproc and WaveOne reciprocating instruments. Int Endod J. 2012;45:614–8. Plotino G, Grande NM, Testarelli L, Gambarini G. Cyclic fatigue of Reciproc and WaveOne reciprocating instruments. Int Endod J. 2012;45:614–8.
37.
Zurück zum Zitat Mesgouez C, Rilliard F, Matossian L, Nassiri K, Mandel E. Influence of operator experience on canal preparation time when using the rotary Ni-Ti ProFile system in simulated curved canals. Int Endod J. 2003;36:161–5. Mesgouez C, Rilliard F, Matossian L, Nassiri K, Mandel E. Influence of operator experience on canal preparation time when using the rotary Ni-Ti ProFile system in simulated curved canals. Int Endod J. 2003;36:161–5.
38.
Zurück zum Zitat Bürklein S, Schäfer E. Apically extruded debris with reciprocating single-file and full-sequence rotary instrumentation systems. J Endod. 2012;38:850–2.CrossRef Bürklein S, Schäfer E. Apically extruded debris with reciprocating single-file and full-sequence rotary instrumentation systems. J Endod. 2012;38:850–2.CrossRef
39.
Zurück zum Zitat Berutti E, Chiandussi G, Paolino DS, Scotti N, Cantatore G, Castellucci A, Pasqualini D. Canal shaping with WaveOne primary reciprocating files and ProTaper system: a comparative study. J Endod. 2012;38:505–9.CrossRef Berutti E, Chiandussi G, Paolino DS, Scotti N, Cantatore G, Castellucci A, Pasqualini D. Canal shaping with WaveOne primary reciprocating files and ProTaper system: a comparative study. J Endod. 2012;38:505–9.CrossRef
Metadaten
Titel
Canal shaping with WaveOne reciprocating files: influence of operator experience on instrument breakage and canal preparation time
verfasst von
Luigi Generali
Elena Righi
Maria Vittoria Todesca
Ugo Consolo
Publikationsdatum
01.07.2014
Verlag
Springer Japan
Erschienen in
Odontology / Ausgabe 2/2014
Print ISSN: 1618-1247
Elektronische ISSN: 1618-1255
DOI
https://doi.org/10.1007/s10266-013-0118-1

Weitere Artikel der Ausgabe 2/2014

Odontology 2/2014 Zur Ausgabe

Newsletter

Bestellen Sie unseren kostenlosen Newsletter Update Zahnmedizin und bleiben Sie gut informiert – ganz bequem per eMail.