Skip to main content
Erschienen in: Magnetic Resonance Materials in Physics, Biology and Medicine 2/2012

01.04.2012 | Research Article

Repeatability of renal arterial spin labelling MRI in healthy subjects

verfasst von: Marica Cutajar, David L. Thomas, Tina Banks, Christopher A. Clark, Xavier Golay, Isky Gordon

Erschienen in: Magnetic Resonance Materials in Physics, Biology and Medicine | Ausgabe 2/2012

Einloggen, um Zugang zu erhalten

Abstract

Object

Arterial spin labelling (ASL) can be used to measure renal perfusion non-invasively. The aim of this study was to determine the repeatability of this technique in healthy kidneys to vindicate its use in clinic.

Materials and methods

Two groups of healthy volunteers were imaged two different days to assess intra- and inter-session repeatability. Oblique-coronal data volumes were acquired on a 1.5 T scanner with a dedicated abdominal 32-channel body phased array coil. ASL was performed using a multi-TI FAIR labelling scheme and 3D GRASE imaging module. Background suppression and respiratory triggering were used. T1 maps of the kidney were acquired using the same sequence with background suppression disabled.

Results

For the group with multiple intra-session ASL measurements, the average cortical perfusion was 197 mL min−1100 g−1 and average cortical T1 was 1265 ms. For both perfusion and T1 the variation shown by the within-subject standard deviation (SDws) (14.6 mL min−1100 g−1 and 33.4 ms) and coefficient of variation (CVws) (7.52 and 2.69%, respectively) was small for all the analyses carried out. Bland–Altman plots were also used to visualise the variation between the same parameters collected from the different scanning sessions in both groups, and demonstrated good reproducibility.

Conclusion

We have shown that in healthy volunteers, ASL parameters are repeatable over a short and long period. This supports the overall aim of using ASL in the clinic to assess longitudinal renal perfusion changes in patients.
Literatur
1.
Zurück zum Zitat Neimatallah MA, Dong Q, Schoenberg SO, Cho KJ, Prince MR (1999) Magnetic resonance imaging in renal transplantation. J Magn Reson Imaging 10(3):357–368PubMedCrossRef Neimatallah MA, Dong Q, Schoenberg SO, Cho KJ, Prince MR (1999) Magnetic resonance imaging in renal transplantation. J Magn Reson Imaging 10(3):357–368PubMedCrossRef
2.
Zurück zum Zitat Szolar DH, Preidler K, Ebner F, Kammerhuber F, Horn S, Ratschek M, Ranner G, Petritsch P, Horina JH (1997) Functional magnetic resonance imaging of human renal allografts during the post-transplant period: preliminary observations. Magn Reson Imaging 15(7):727–735PubMedCrossRef Szolar DH, Preidler K, Ebner F, Kammerhuber F, Horn S, Ratschek M, Ranner G, Petritsch P, Horina JH (1997) Functional magnetic resonance imaging of human renal allografts during the post-transplant period: preliminary observations. Magn Reson Imaging 15(7):727–735PubMedCrossRef
3.
Zurück zum Zitat Haufe SE, Riedmuller K, Haberkorn U (2006) Nuclear medicine procedures for the diagnosis of acute and chronic renal failure. Nephron Clin Pract 103(2):c77–c84PubMedCrossRef Haufe SE, Riedmuller K, Haberkorn U (2006) Nuclear medicine procedures for the diagnosis of acute and chronic renal failure. Nephron Clin Pract 103(2):c77–c84PubMedCrossRef
4.
Zurück zum Zitat Peters AM, Brown J, Hartnell GG, Myers MJ, Haskell C, Lavender JP (1987) Non-invasive measurement of renal blood flow with 99mTc DTPA: comparison with radiolabelled microspheres. Cardiovasc Res 21(11):830–834PubMedCrossRef Peters AM, Brown J, Hartnell GG, Myers MJ, Haskell C, Lavender JP (1987) Non-invasive measurement of renal blood flow with 99mTc DTPA: comparison with radiolabelled microspheres. Cardiovasc Res 21(11):830–834PubMedCrossRef
5.
Zurück zum Zitat Golay X, Hendrikse J, Lim TC (2004) Perfusion imaging using arterial spin labeling. Top Magn Reson Imaging 15(1):10–27PubMedCrossRef Golay X, Hendrikse J, Lim TC (2004) Perfusion imaging using arterial spin labeling. Top Magn Reson Imaging 15(1):10–27PubMedCrossRef
6.
Zurück zum Zitat Petersen ET, Zimine I, Ho YC, Golay X (2006) Non-invasive measurement of perfusion: a critical review of arterial spin labelling techniques. Br J Radiol 79(944):688–701PubMedCrossRef Petersen ET, Zimine I, Ho YC, Golay X (2006) Non-invasive measurement of perfusion: a critical review of arterial spin labelling techniques. Br J Radiol 79(944):688–701PubMedCrossRef
7.
Zurück zum Zitat Buxton RB (2002) Introduction to functional magnetic resonance imaging. Cambridge University Press, Cambridge Buxton RB (2002) Introduction to functional magnetic resonance imaging. Cambridge University Press, Cambridge
8.
Zurück zum Zitat Martirosian P, Boss A, Schraml C, Schwenzer NF, Graf H, Claussen CD, Schick F (2010) Magnetic resonance perfusion imaging without contrast media. Eur J Nucl Med Mol Imaging 37(Suppl 1):S52–S64PubMedCrossRef Martirosian P, Boss A, Schraml C, Schwenzer NF, Graf H, Claussen CD, Schick F (2010) Magnetic resonance perfusion imaging without contrast media. Eur J Nucl Med Mol Imaging 37(Suppl 1):S52–S64PubMedCrossRef
9.
Zurück zum Zitat Roberts DA, Detre JA, Bolinger L, Insko EK, Lenkinski RE, Pentecost MJ, Leigh JS Jr (1995) Renal perfusion in humans: MR imaging with spin tagging of arterial water. Radiology 196(1):281–286PubMed Roberts DA, Detre JA, Bolinger L, Insko EK, Lenkinski RE, Pentecost MJ, Leigh JS Jr (1995) Renal perfusion in humans: MR imaging with spin tagging of arterial water. Radiology 196(1):281–286PubMed
10.
Zurück zum Zitat Chen Q, Siewert B, Bly BM, Warach S, Edelman RR (1997) STAR-HASTE: perfusion imaging without magnetic susceptibility artifact. Magn Reson Med 38:404–408PubMedCrossRef Chen Q, Siewert B, Bly BM, Warach S, Edelman RR (1997) STAR-HASTE: perfusion imaging without magnetic susceptibility artifact. Magn Reson Med 38:404–408PubMedCrossRef
11.
Zurück zum Zitat Karger N, Biederer J, Lusse S, Grimm J, Steffens J, Heller M, Gluer C (2000) Quantitation of renal perfusion using arterial spin labeling with FAIR-UFLARE. Magn Reson Imaging 18(6):641–647PubMedCrossRef Karger N, Biederer J, Lusse S, Grimm J, Steffens J, Heller M, Gluer C (2000) Quantitation of renal perfusion using arterial spin labeling with FAIR-UFLARE. Magn Reson Imaging 18(6):641–647PubMedCrossRef
12.
Zurück zum Zitat Martirosian P, Klose U, Mader I, Schick F (2004) FAIR True-FISP perfusion imaging of the kidneys. Magn Reson Med 51:353–361PubMedCrossRef Martirosian P, Klose U, Mader I, Schick F (2004) FAIR True-FISP perfusion imaging of the kidneys. Magn Reson Med 51:353–361PubMedCrossRef
13.
Zurück zum Zitat De Bazelaire C, Rofsky NM, Duhamel G, Michaelson MD, George D, Alsop DC (2005) Arterial spin labeling blood flow magnetic resonance imaging for the characterization of metastatic renal cell carcinoma(1). Acad Radiol 12(3):347–357PubMedCrossRef De Bazelaire C, Rofsky NM, Duhamel G, Michaelson MD, George D, Alsop DC (2005) Arterial spin labeling blood flow magnetic resonance imaging for the characterization of metastatic renal cell carcinoma(1). Acad Radiol 12(3):347–357PubMedCrossRef
14.
Zurück zum Zitat Fenchel M, Martirosian P, Langanke J, Giersch J, Miller S, Stauder NI, Kramer U, Claussen CD, Schick F (2006) Perfusion MR imaging with FAIR true FISP spin labeling in patients with and without renal artery stenosis: initial experience. Radiology 238(3):1013–1021PubMedCrossRef Fenchel M, Martirosian P, Langanke J, Giersch J, Miller S, Stauder NI, Kramer U, Claussen CD, Schick F (2006) Perfusion MR imaging with FAIR true FISP spin labeling in patients with and without renal artery stenosis: initial experience. Radiology 238(3):1013–1021PubMedCrossRef
15.
Zurück zum Zitat Artz NS, Sadowski EA, Wentland AL, Grist TM, Seo S, Djamali A, Fain SB (2011) Arterial spin labeling MRI for assessment of perfusion in native and transplanted kidneys. Magn Reson Imaging 29(7):74–82PubMedCrossRef Artz NS, Sadowski EA, Wentland AL, Grist TM, Seo S, Djamali A, Fain SB (2011) Arterial spin labeling MRI for assessment of perfusion in native and transplanted kidneys. Magn Reson Imaging 29(7):74–82PubMedCrossRef
16.
Zurück zum Zitat Lanzman RS, Wittsack HJ, Martirosian P, Zgoura P, Bilk P, Kropil P, Schick F, Voiculescu A, Blondin D (2010) Quantification of renal allograft perfusion using arterial spin labeling MRI: initial results. Eur Radiol 20(6):1485–1491PubMedCrossRef Lanzman RS, Wittsack HJ, Martirosian P, Zgoura P, Bilk P, Kropil P, Schick F, Voiculescu A, Blondin D (2010) Quantification of renal allograft perfusion using arterial spin labeling MRI: initial results. Eur Radiol 20(6):1485–1491PubMedCrossRef
17.
Zurück zum Zitat Pedrosa I, Alsop DC, Rofsky NM (2009) Magnetic resonance imaging as a biomarker in renal cell carcinoma. Cancer 115(10 Suppl):2334–2345PubMedCrossRef Pedrosa I, Alsop DC, Rofsky NM (2009) Magnetic resonance imaging as a biomarker in renal cell carcinoma. Cancer 115(10 Suppl):2334–2345PubMedCrossRef
18.
Zurück zum Zitat Kim SG (1995) Quantification of relative cerebral blood flow change by flow-sensitive alternating inversion recovery (FAIR) technique: application to functional mapping. Magn Reson Med 34:293–301PubMedCrossRef Kim SG (1995) Quantification of relative cerebral blood flow change by flow-sensitive alternating inversion recovery (FAIR) technique: application to functional mapping. Magn Reson Med 34:293–301PubMedCrossRef
19.
Zurück zum Zitat Song R, Loeffler RB, Hillenbrand CM (2010) Improved renal perfusion measurement with a dual navigator-gated Q2TIPS fair technique. Magn Reson Med 64(5):1352–1359PubMedCrossRef Song R, Loeffler RB, Hillenbrand CM (2010) Improved renal perfusion measurement with a dual navigator-gated Q2TIPS fair technique. Magn Reson Med 64(5):1352–1359PubMedCrossRef
20.
Zurück zum Zitat Wong EC, Buxton RB, Frank LR (1998) Quantitative imaging of perfusion using a single subtraction (QUIPSS and QUIPSS II). Magn Reson Med 39:702–708PubMedCrossRef Wong EC, Buxton RB, Frank LR (1998) Quantitative imaging of perfusion using a single subtraction (QUIPSS and QUIPSS II). Magn Reson Med 39:702–708PubMedCrossRef
21.
Zurück zum Zitat Dobbs MS, Woodhouse N, Parker GJ, Naish JH (2010) Renal ASL using multiple inversion time, free breathing STAR-HASTE technique at 3T [abstract]. In: Proceedings of 18th Meeting of ISMRM, p 2676 Dobbs MS, Woodhouse N, Parker GJ, Naish JH (2010) Renal ASL using multiple inversion time, free breathing STAR-HASTE technique at 3T [abstract]. In: Proceedings of 18th Meeting of ISMRM, p 2676
22.
Zurück zum Zitat Buxton RB, Frank LR, Wong EC, Siewert B, Warach S, Edelman RR (1998) A general kinetic model for quantitative perfusion imaging with arterial spin labeling. Magn Reson Med 40(3):383–396PubMedCrossRef Buxton RB, Frank LR, Wong EC, Siewert B, Warach S, Edelman RR (1998) A general kinetic model for quantitative perfusion imaging with arterial spin labeling. Magn Reson Med 40(3):383–396PubMedCrossRef
23.
Zurück zum Zitat Gardener AG, Francis ST (2010) Multislice perfusion of the kidneys using parallel imaging: image acquisition and analysis strategies. Magn Reson Imaging 63(6):1627–1636 Gardener AG, Francis ST (2010) Multislice perfusion of the kidneys using parallel imaging: image acquisition and analysis strategies. Magn Reson Imaging 63(6):1627–1636
24.
Zurück zum Zitat Gunther M, Oshio K, Feinberg DA (2005) Single-shot 3D imaging techniques improve arterial spin labeling perfusion measurements. Magn Reson Med 54(2):491–498PubMedCrossRef Gunther M, Oshio K, Feinberg DA (2005) Single-shot 3D imaging techniques improve arterial spin labeling perfusion measurements. Magn Reson Med 54(2):491–498PubMedCrossRef
25.
Zurück zum Zitat Freeman R (1997) A handbook of nuclear magnetic resonance. Longman Limited, London, p 275 Freeman R (1997) A handbook of nuclear magnetic resonance. Longman Limited, London, p 275
26.
Zurück zum Zitat Regan MC, Young LS, Geraghty J, Fitzpatrick JM (1995) Regional renal blood flow in normal and disease states. Urol Res 23(1):1–10PubMedCrossRef Regan MC, Young LS, Geraghty J, Fitzpatrick JM (1995) Regional renal blood flow in normal and disease states. Urol Res 23(1):1–10PubMedCrossRef
27.
Zurück zum Zitat Cheong B, Muthupillai R, Rubin MF, Flamm SD (2007) Normal values for renal length and volume as measured by magnetic resonance imaging. Clin J Am Soc Nephrol 2(1):38–45PubMedCrossRef Cheong B, Muthupillai R, Rubin MF, Flamm SD (2007) Normal values for renal length and volume as measured by magnetic resonance imaging. Clin J Am Soc Nephrol 2(1):38–45PubMedCrossRef
28.
Zurück zum Zitat Ye FQ, Frank JA, Weinberger DR, McLaughlin AC (2000) Noise reduction in 3D perfusion imaging by attenuating the static signal in arterial spin tagging (ASSIST). Magn Reson Med 44(1):92–100PubMedCrossRef Ye FQ, Frank JA, Weinberger DR, McLaughlin AC (2000) Noise reduction in 3D perfusion imaging by attenuating the static signal in arterial spin tagging (ASSIST). Magn Reson Med 44(1):92–100PubMedCrossRef
29.
30.
Zurück zum Zitat Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1(8476):307–310PubMedCrossRef Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1(8476):307–310PubMedCrossRef
31.
Zurück zum Zitat Chasis H, Ranges HA, Goldring W, Smith HW (1938) The control of renal blood flow and glomerular filtration in normal man. J Clin Invest 17(5):683–697PubMedCrossRef Chasis H, Ranges HA, Goldring W, Smith HW (1938) The control of renal blood flow and glomerular filtration in normal man. J Clin Invest 17(5):683–697PubMedCrossRef
32.
Zurück zum Zitat Ladefoged J (1966) Measurements of the renal blood flow in man with the 133 xenon wash-out technique. A description of the method. Scand J Clin Lab Invest 18(3):299–315PubMedCrossRef Ladefoged J (1966) Measurements of the renal blood flow in man with the 133 xenon wash-out technique. A description of the method. Scand J Clin Lab Invest 18(3):299–315PubMedCrossRef
33.
Zurück zum Zitat Miles KA (1991) Measurement of tissue perfusion by dynamic computed tomography. Br J Radiol 64(761):409–412PubMedCrossRef Miles KA (1991) Measurement of tissue perfusion by dynamic computed tomography. Br J Radiol 64(761):409–412PubMedCrossRef
34.
Zurück zum Zitat Kiefer C, Schroth G, Gralla J, Diehm N, Baumgartner I, Husmann M (2009) A feasibility study on model-based evaluation of kidney perfusion measured by means of FAIR prepared true-FISP arterial spin labeling (ASL) on a 3-T MR scanner. Acad Radiol 16(1):79–87PubMedCrossRef Kiefer C, Schroth G, Gralla J, Diehm N, Baumgartner I, Husmann M (2009) A feasibility study on model-based evaluation of kidney perfusion measured by means of FAIR prepared true-FISP arterial spin labeling (ASL) on a 3-T MR scanner. Acad Radiol 16(1):79–87PubMedCrossRef
35.
Zurück zum Zitat Song R, Loeffler RB, Hillenbrand CM (2011) QUIPSS II with window-sliding saturation sequence (Q2WISE). Magn Reson Med EPub ahead of print. doi:10.1002/mrm.23093 Song R, Loeffler RB, Hillenbrand CM (2011) QUIPSS II with window-sliding saturation sequence (Q2WISE). Magn Reson Med EPub ahead of print. doi:10.​1002/​mrm.​23093
36.
Zurück zum Zitat Robson PM, Madhuranthakam AJ, Dai W, Pedrosa I, Rofsky NM, Alsop DC (2009) Strategies for reducing respiratory motion artifacts in renal perfusion imaging with arterial spin labeling. Magn Reson Med 61(6):1374–1387PubMedCrossRef Robson PM, Madhuranthakam AJ, Dai W, Pedrosa I, Rofsky NM, Alsop DC (2009) Strategies for reducing respiratory motion artifacts in renal perfusion imaging with arterial spin labeling. Magn Reson Med 61(6):1374–1387PubMedCrossRef
37.
Zurück zum Zitat Artz NS, Sadowski EA, Wentland AL, Djamali A, Grist TM, Seo S, Fain SB (2011) Reproducibility of renal perfusion MR imaging in native and transplanted kidneys using non-contrast arterial spin labeling. J Magn Reson Imaging 33(6):1414–1421PubMedCrossRef Artz NS, Sadowski EA, Wentland AL, Djamali A, Grist TM, Seo S, Fain SB (2011) Reproducibility of renal perfusion MR imaging in native and transplanted kidneys using non-contrast arterial spin labeling. J Magn Reson Imaging 33(6):1414–1421PubMedCrossRef
Metadaten
Titel
Repeatability of renal arterial spin labelling MRI in healthy subjects
verfasst von
Marica Cutajar
David L. Thomas
Tina Banks
Christopher A. Clark
Xavier Golay
Isky Gordon
Publikationsdatum
01.04.2012
Verlag
Springer-Verlag
Erschienen in
Magnetic Resonance Materials in Physics, Biology and Medicine / Ausgabe 2/2012
Print ISSN: 0968-5243
Elektronische ISSN: 1352-8661
DOI
https://doi.org/10.1007/s10334-011-0300-9

Weitere Artikel der Ausgabe 2/2012

Magnetic Resonance Materials in Physics, Biology and Medicine 2/2012 Zur Ausgabe

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.