Initial evaluation of coronary images from 320-detector row computed tomography

verfasst von: Frank J. Rybicki, Hansel J. Otero, Michael L. Steigner, Gabriel Vorobiof, Leelakrishna Nallamshetty, Dimitrios Mitsouras, Hale Ersoy, Richard T. Mather, Philip F. Judy, Tianxi Cai, Karl Coyner, Kurt Schultz, Amanda G. Whitmore, Marcelo F. Di Carli

Erschienen in: The International Journal of Cardiovascular Imaging | Ausgabe 5/2008

Einloggen, um Zugang zu erhalten

Abstract

Purpose To evaluate image quality and contrast opacification from coronary images acquired from 320-detector row computed tomography (CT). Patient dose is estimated for prospective and retrospective ECG-gating; initial correlation between 320-slice CT and coronary catheterization is illustrated. Methods Retrospective image evaluation from forty consecutive patients included subjective assessment of image quality and contrast opacification (80 ml iopamidol 370 mg I/ml followed by 40 ml saline). Region of interest opacification measurements at the ostium and at 2.5 mm diameter were used to determine the gradient of contrast opacification (defined as the proximal minus distal HU measurements) in coronary arteries imaged in a single heartbeat. Estimated effective dose was compared for prospective versus retrospective ECG-gating, two body mass index categories (30 kg/m2 cutoff), and single versus two heartbeat acquisition. When available, CT findings were correlated with those from coronary catheterization. Results Over 89% of arterial segments (15 segment model) had excellent image quality. The most common reason for image degradation was cardiac motion. One segment in one patient was considered unevaluable. Contrast opacification was almost universally considered excellent. The mean Hounsfield units (HU) was greater than 350; the coronary contrast opacification gradient was 30–50 HU. Patient doses were greater for retrospective ECG-gating, larger patients, and those imaged with two heartbeats. For the most common (n = 25) protocol (120 kV, 400 mA, prospective ECG-gating, 60–100% phase window, 16 cm craniocaudal coverage, single heartbeat), the mean dose was 6.8 ± 1.4 mSv. All CT findings were confirmed in the four patients who underwent coronary catheterization. Conclusion Initial 320-detector row coronary CT images have consistently excellent quality and iodinated contrast opacification. These patients were scanned with conservative protocols with respect to iodine load, prospective ECG-gating phase window, and craniocaudal coverage. Future work will focus on lowering contrast and radiation dose while maintaining image quality.
Literatur
1.
Zurück zum Zitat Gershlick AH, de Belder M, Chambers J, Hackett D, Keal R, Kelion A et al (2007) Role of non-invasive imaging in the management of coronary artery disease: an assessment of likely change over the next 10 years. A report from the British Cardiovascular Society Working Group. Heart 93(4):423–431PubMedCrossRef Gershlick AH, de Belder M, Chambers J, Hackett D, Keal R, Kelion A et al (2007) Role of non-invasive imaging in the management of coronary artery disease: an assessment of likely change over the next 10 years. A report from the British Cardiovascular Society Working Group. Heart 93(4):423–431PubMedCrossRef
2.
Zurück zum Zitat Hoe JW, Toh KH (2007) A practical guide to reading CT coronary angiograms-how to avoid mistakes when assessing for coronary stenoses. Int J Cardiovasc Imaging 23(5):617–633PubMedCrossRef Hoe JW, Toh KH (2007) A practical guide to reading CT coronary angiograms-how to avoid mistakes when assessing for coronary stenoses. Int J Cardiovasc Imaging 23(5):617–633PubMedCrossRef
3.
Zurück zum Zitat Limkakeng AT, Halpern E, Takakuwa KM (2007) Sixty-four-slice multidetector computed tomography: the future of ED cardiac care. Am J Emerg Med 25(4):450–458PubMedCrossRef Limkakeng AT, Halpern E, Takakuwa KM (2007) Sixty-four-slice multidetector computed tomography: the future of ED cardiac care. Am J Emerg Med 25(4):450–458PubMedCrossRef
4.
Zurück zum Zitat Ramos JJ, Williams M, Synetos A, Lerakis S (2007) Clinical utility of cardiac computed tomography. Am J Med Sci 334(5):350–355PubMedCrossRef Ramos JJ, Williams M, Synetos A, Lerakis S (2007) Clinical utility of cardiac computed tomography. Am J Med Sci 334(5):350–355PubMedCrossRef
5.
Zurück zum Zitat Baur LH (2008) Cardiac imaging at the emergency department is a must! The role of cardiac computed tomography and magnetic resonance imaging in the evaluation of acute chest pain in the emergency department. Int J Cardiovasc Imaging 24(3):343–344PubMedCrossRef Baur LH (2008) Cardiac imaging at the emergency department is a must! The role of cardiac computed tomography and magnetic resonance imaging in the evaluation of acute chest pain in the emergency department. Int J Cardiovasc Imaging 24(3):343–344PubMedCrossRef
6.
Zurück zum Zitat Vanhoenacker PK, Heijenbrok-Kal MH, Van Heste R, Decramer I, Van Hoe LR, Wijns W et al (2007) Diagnostic performance of multidetector CT angiography for assessment of coronary artery disease: meta-analysis. Radiology 244(2):419–428PubMedCrossRef Vanhoenacker PK, Heijenbrok-Kal MH, Van Heste R, Decramer I, Van Hoe LR, Wijns W et al (2007) Diagnostic performance of multidetector CT angiography for assessment of coronary artery disease: meta-analysis. Radiology 244(2):419–428PubMedCrossRef
7.
Zurück zum Zitat Scheffel H, Alkadhi H, Plass A, Vachenauer R, Desbiolles L, Gaemperli O et al (2006) Accuracy of dual-source CT coronary angiography: first experience in a high pre-test probability population without heart rate control. Eur Radiol 16(12):2739–2747PubMedCrossRef Scheffel H, Alkadhi H, Plass A, Vachenauer R, Desbiolles L, Gaemperli O et al (2006) Accuracy of dual-source CT coronary angiography: first experience in a high pre-test probability population without heart rate control. Eur Radiol 16(12):2739–2747PubMedCrossRef
8.
Zurück zum Zitat Raff GL, Gallagher MJ, O’Neill WW, Goldstein JA (2005) Diagnostic accuracy of noninvasive coronary angiography using 64-slice spiral computed tomography. J Am Coll Cardiol 46(3):552–557PubMedCrossRef Raff GL, Gallagher MJ, O’Neill WW, Goldstein JA (2005) Diagnostic accuracy of noninvasive coronary angiography using 64-slice spiral computed tomography. J Am Coll Cardiol 46(3):552–557PubMedCrossRef
9.
Zurück zum Zitat Mollet NR, Cademartiri F, van Mieghem CA, Runza G, McFadden EP, Baks T et al (2005) High-resolution spiral computed tomography coronary angiography in patients referred for diagnostic conventional coronary angiography. Circulation 112(15):2318–23PubMedCrossRef Mollet NR, Cademartiri F, van Mieghem CA, Runza G, McFadden EP, Baks T et al (2005) High-resolution spiral computed tomography coronary angiography in patients referred for diagnostic conventional coronary angiography. Circulation 112(15):2318–23PubMedCrossRef
10.
Zurück zum Zitat Hoffmann MH, Shi H, Schmitz BL, Schmid FT, Lieberknecht M, Schulze R et al (2005) Noninvasive coronary angiography with multislice computed tomography. JAMA 293(20):2471–2478PubMedCrossRef Hoffmann MH, Shi H, Schmitz BL, Schmid FT, Lieberknecht M, Schulze R et al (2005) Noninvasive coronary angiography with multislice computed tomography. JAMA 293(20):2471–2478PubMedCrossRef
11.
Zurück zum Zitat Bordeleau E, Lamonde A, Prenovault J, Belblidia A, Cote G, Lesperance J et al (2007) Accuracy and rate of coronary artery segment visualization with CT angiography for the non-invasive detection of coronary artery stenoses. Int J Cardiovasc Imaging 23(6):771–780PubMedCrossRef Bordeleau E, Lamonde A, Prenovault J, Belblidia A, Cote G, Lesperance J et al (2007) Accuracy and rate of coronary artery segment visualization with CT angiography for the non-invasive detection of coronary artery stenoses. Int J Cardiovasc Imaging 23(6):771–780PubMedCrossRef
12.
Zurück zum Zitat Hamon M, Morello R, Riddell JW (2007) Coronary arteries: diagnostic performance of 16- versus 64-section spiral CT compared with invasive coronary angiography−meta-analysis. Radiology 245(3):720–731PubMedCrossRef Hamon M, Morello R, Riddell JW (2007) Coronary arteries: diagnostic performance of 16- versus 64-section spiral CT compared with invasive coronary angiography−meta-analysis. Radiology 245(3):720–731PubMedCrossRef
13.
Zurück zum Zitat Pugliese F, Mollet NR, Hunink MG, Cademartiri F, Nieman K, van Domburg RT et al (2008) Diagnostic performance of coronary CT angiography by using different generations of multisection scanners: single-center experience. Radiology 246(2):384–393PubMedCrossRef Pugliese F, Mollet NR, Hunink MG, Cademartiri F, Nieman K, van Domburg RT et al (2008) Diagnostic performance of coronary CT angiography by using different generations of multisection scanners: single-center experience. Radiology 246(2):384–393PubMedCrossRef
14.
Zurück zum Zitat Jakobs TF, Wintersperger BJ, Herzog P, Flohr T, Suess C, Knez A et al (2003) Ultra-low-dose coronary artery calcium screening using multislice CT with retrospective ECG gating. Eur Radiol 13(8):1923–1930PubMedCrossRef Jakobs TF, Wintersperger BJ, Herzog P, Flohr T, Suess C, Knez A et al (2003) Ultra-low-dose coronary artery calcium screening using multislice CT with retrospective ECG gating. Eur Radiol 13(8):1923–1930PubMedCrossRef
15.
Zurück zum Zitat d’Agostino AG, Remy-Jardin M, Khalil C, Delannoy-Deken V, Flohr T, Duhamel A et al (2006) Low-dose ECG-gated 64-slices helical CT angiography of the chest: evaluation of image quality in 105 patients. Eur Radiol 16(10):2137–2146PubMedCrossRef d’Agostino AG, Remy-Jardin M, Khalil C, Delannoy-Deken V, Flohr T, Duhamel A et al (2006) Low-dose ECG-gated 64-slices helical CT angiography of the chest: evaluation of image quality in 105 patients. Eur Radiol 16(10):2137–2146PubMedCrossRef
16.
Zurück zum Zitat Gerber TC, Kuzo RS, Morin RL (2005) Techniques and parameters for estimating radiation exposure and dose in cardiac computed tomography. Int J Cardiovasc Imaging 21(1):165–176PubMedCrossRef Gerber TC, Kuzo RS, Morin RL (2005) Techniques and parameters for estimating radiation exposure and dose in cardiac computed tomography. Int J Cardiovasc Imaging 21(1):165–176PubMedCrossRef
17.
Zurück zum Zitat Husmann L, Valenta I, Gaemperli O, Adda O, Treyer V, Wyss CA et al (2008) Feasibility of low-dose coronary CT angiography: first experience with prospective ECG-gating. Eur Heart J 29(2):191–197PubMedCrossRef Husmann L, Valenta I, Gaemperli O, Adda O, Treyer V, Wyss CA et al (2008) Feasibility of low-dose coronary CT angiography: first experience with prospective ECG-gating. Eur Heart J 29(2):191–197PubMedCrossRef
18.
Zurück zum Zitat Horiguchi J, Kiguchi M, Fujioka C, Shen Y, Arie R, Sunasaka K et al (2008) Radiation dose, image quality, stenosis measurement, and CT densitometry using ECG-triggered coronary 64-MDCT angiography: a phantom study. AJR Am J Roentgenol 190(2):315–320PubMedCrossRef Horiguchi J, Kiguchi M, Fujioka C, Shen Y, Arie R, Sunasaka K et al (2008) Radiation dose, image quality, stenosis measurement, and CT densitometry using ECG-triggered coronary 64-MDCT angiography: a phantom study. AJR Am J Roentgenol 190(2):315–320PubMedCrossRef
19.
Zurück zum Zitat Earls JP, Berman EL, Urban BA, Curry CA, Lane JL, Jennings RS et al (2008) Prospectively gated transverse coronary CT angiography versus retrospectively gated helical technique: improved image quality and reduced radiation dose. Radiology 246(3):742–753PubMedCrossRef Earls JP, Berman EL, Urban BA, Curry CA, Lane JL, Jennings RS et al (2008) Prospectively gated transverse coronary CT angiography versus retrospectively gated helical technique: improved image quality and reduced radiation dose. Radiology 246(3):742–753PubMedCrossRef
20.
Zurück zum Zitat Brodoefel H, Kramer U, Reimann A, Burgstahler C, Schroeder S, Kopp A et al (2007) Dual-source CT with improved temporal resolution in assessment of left ventricular function: a pilot study. AJR Am J Roentgenol 189(5):1064–1070PubMedCrossRef Brodoefel H, Kramer U, Reimann A, Burgstahler C, Schroeder S, Kopp A et al (2007) Dual-source CT with improved temporal resolution in assessment of left ventricular function: a pilot study. AJR Am J Roentgenol 189(5):1064–1070PubMedCrossRef
21.
Zurück zum Zitat Stolzmann P, Scheffel H, Schertler T, Frauenfelder T, Leschka S, Husmann L et al (2007) Radiation dose estimates in dual-source computed tomography coronary angiography. Eur Radiol 18(3):592–599PubMedCrossRef Stolzmann P, Scheffel H, Schertler T, Frauenfelder T, Leschka S, Husmann L et al (2007) Radiation dose estimates in dual-source computed tomography coronary angiography. Eur Radiol 18(3):592–599PubMedCrossRef
22.
Zurück zum Zitat Reimann AJ, Rinck D, Birinci-Aydogan A, Scheuering M, Burgstahler C, Schroeder S et al (2007) Dual-source computed tomography: advances of improved temporal resolution in coronary plaque imaging. Invest Radiol 42(3):196–203PubMedCrossRef Reimann AJ, Rinck D, Birinci-Aydogan A, Scheuering M, Burgstahler C, Schroeder S et al (2007) Dual-source computed tomography: advances of improved temporal resolution in coronary plaque imaging. Invest Radiol 42(3):196–203PubMedCrossRef
23.
Zurück zum Zitat McCollough CH, Primak AN, Saba O, Bruder H, Stierstorfer K, Raupach R et al (2007) Dose performance of a 64-channel dual-source CT scanner. Radiology 243(3):775–784PubMedCrossRef McCollough CH, Primak AN, Saba O, Bruder H, Stierstorfer K, Raupach R et al (2007) Dose performance of a 64-channel dual-source CT scanner. Radiology 243(3):775–784PubMedCrossRef
24.
Zurück zum Zitat Matt D, Scheffel H, Leschka S, Flohr TG, Marincek B, Kaufmann PA et al (2007) Dual-source CT coronary angiography: image quality, mean heart rate, and heart rate variability. AJR Am J Roentgenol 189(3):567–573PubMedCrossRef Matt D, Scheffel H, Leschka S, Flohr TG, Marincek B, Kaufmann PA et al (2007) Dual-source CT coronary angiography: image quality, mean heart rate, and heart rate variability. AJR Am J Roentgenol 189(3):567–573PubMedCrossRef
25.
Zurück zum Zitat Johnson TR, Nikolaou K, Wintersperger BJ, Leber AW, von Ziegler F, Rist C et al (2006) Dual-source CT cardiac imaging: initial experience. Eur Radiol 16(7):1409–1415PubMedCrossRef Johnson TR, Nikolaou K, Wintersperger BJ, Leber AW, von Ziegler F, Rist C et al (2006) Dual-source CT cardiac imaging: initial experience. Eur Radiol 16(7):1409–1415PubMedCrossRef
26.
Zurück zum Zitat Mizuno N, Funabashi N, Imada M, Tsunoo T, Endo M, Komuro I (2007) Utility of 256-slice cone beam tomography for real four-dimensional volumetric analysis without electrocardiogram gated acquisition. Int J Cardiol 120(2):262–267PubMedCrossRef Mizuno N, Funabashi N, Imada M, Tsunoo T, Endo M, Komuro I (2007) Utility of 256-slice cone beam tomography for real four-dimensional volumetric analysis without electrocardiogram gated acquisition. Int J Cardiol 120(2):262–267PubMedCrossRef
27.
Zurück zum Zitat Funabashi N, Mizuno N, Yoshida K, Tsunoo T, Mori S, Tanada S et al (2007) Superiority of synchrony of 256-slice cone beam computed tomography for acquiring pulsating objects. Comparison with conventional multislice computed tomography. Int J Cardiol 118(3):400–405PubMedCrossRef Funabashi N, Mizuno N, Yoshida K, Tsunoo T, Mori S, Tanada S et al (2007) Superiority of synchrony of 256-slice cone beam computed tomography for acquiring pulsating objects. Comparison with conventional multislice computed tomography. Int J Cardiol 118(3):400–405PubMedCrossRef
28.
Zurück zum Zitat Endo M, Mori S, Tsunoo T, Miyazaki H (2006) Magnitude and effects of x-ray scatter in a 256-slice CT scanner. Med Phys 33(9):3359–3368PubMedCrossRef Endo M, Mori S, Tsunoo T, Miyazaki H (2006) Magnitude and effects of x-ray scatter in a 256-slice CT scanner. Med Phys 33(9):3359–3368PubMedCrossRef
29.
Zurück zum Zitat Mori S, Nishizawa K, Ohno M, Endo M (2006) Conversion factor for CT dosimetry to assess patient dose using a 256-slice CT scanner. Br J Radiol 79(947):888–892PubMedCrossRef Mori S, Nishizawa K, Ohno M, Endo M (2006) Conversion factor for CT dosimetry to assess patient dose using a 256-slice CT scanner. Br J Radiol 79(947):888–892PubMedCrossRef
30.
Zurück zum Zitat Mori S, Endo M, Nishizawa K, Murase K, Fujiwara H, Tanada S (2006) Comparison of patient doses in 256-slice CT and 16-slice CT scanners. Br J Radiol 79(937):56–61PubMedCrossRef Mori S, Endo M, Nishizawa K, Murase K, Fujiwara H, Tanada S (2006) Comparison of patient doses in 256-slice CT and 16-slice CT scanners. Br J Radiol 79(937):56–61PubMedCrossRef
31.
Zurück zum Zitat Mori S, Nishizawa K, Kondo C, Ohno M, Akahane K, Endo M (2007) Effective doses in subjects undergoing computed tomography cardiac imaging with the 256-multislice CT scanner. Eur J Radiol 65(3):442–448PubMedCrossRef Mori S, Nishizawa K, Kondo C, Ohno M, Akahane K, Endo M (2007) Effective doses in subjects undergoing computed tomography cardiac imaging with the 256-multislice CT scanner. Eur J Radiol 65(3):442–448PubMedCrossRef
32.
Zurück zum Zitat Austen WG, Edwards JE, Frye RL, Gensini GG, Gott VL, Griffith LS et al (1975) A reporting system on patients evaluated for coronary artery disease. Report of the Ad Hoc Committee for Grading of Coronary Artery Disease, Council on Cardiovascular Surgery, American Heart Association. Circulation 51(4 Suppl):5–40PubMed Austen WG, Edwards JE, Frye RL, Gensini GG, Gott VL, Griffith LS et al (1975) A reporting system on patients evaluated for coronary artery disease. Report of the Ad Hoc Committee for Grading of Coronary Artery Disease, Council on Cardiovascular Surgery, American Heart Association. Circulation 51(4 Suppl):5–40PubMed
33.
Zurück zum Zitat European Commission (2000) European guidelines on quality criteria for computed tomography. In. Luxembourg: Office for Official Publications of the European Communities European Commission (2000) European guidelines on quality criteria for computed tomography. In. Luxembourg: Office for Official Publications of the European Communities
34.
Zurück zum Zitat Rybicki FJ, Sheth T (2004) Imaging protocol for cardiac CT. In: Siani S, Rubin G, Kalra M (eds) Multi detector CT protocols: a practical approach. Springer-Verlag, Milan, pp 97–110 Rybicki FJ, Sheth T (2004) Imaging protocol for cardiac CT. In: Siani S, Rubin G, Kalra M (eds) Multi detector CT protocols: a practical approach. Springer-Verlag, Milan, pp 97–110
35.
Zurück zum Zitat Delhaye D, Remy-Jardin M, Rozel C, Dusson C, Wurtz A, Delannoy-Deken V et al (2007) Coronary artery imaging during preoperative CT staging: preliminary experience with 64-slice multidetector CT in 99 consecutive patients. Eur Radiol 17(3):591–602PubMedCrossRef Delhaye D, Remy-Jardin M, Rozel C, Dusson C, Wurtz A, Delannoy-Deken V et al (2007) Coronary artery imaging during preoperative CT staging: preliminary experience with 64-slice multidetector CT in 99 consecutive patients. Eur Radiol 17(3):591–602PubMedCrossRef
36.
Zurück zum Zitat Muhlenbruch G, Seyfarth T, Soo CS, Pregalathan N, Mahnken AH (2007) Diagnostic value of 64-slice multi-detector row cardiac CTA in symptomatic patients. Eur Radiol 17(3):603–609PubMedCrossRef Muhlenbruch G, Seyfarth T, Soo CS, Pregalathan N, Mahnken AH (2007) Diagnostic value of 64-slice multi-detector row cardiac CTA in symptomatic patients. Eur Radiol 17(3):603–609PubMedCrossRef
37.
Zurück zum Zitat Delhaye D, Remy-Jardin M, Salem R, Teisseire A, Khalil C, Delannoy-Deken V et al (2007) Coronary imaging quality in routine ECG-gated multidetector CT examinations of the entire thorax: preliminary experience with a 64-slice CT system in 133 patients. Eur Radiol 17(4):902–910PubMedCrossRef Delhaye D, Remy-Jardin M, Salem R, Teisseire A, Khalil C, Delannoy-Deken V et al (2007) Coronary imaging quality in routine ECG-gated multidetector CT examinations of the entire thorax: preliminary experience with a 64-slice CT system in 133 patients. Eur Radiol 17(4):902–910PubMedCrossRef
38.
Zurück zum Zitat Leschka S, Husmann L, Desbiolles LM, Gaemperli O, Schepis T, Koepfli P et al (2006) Optimal image reconstruction intervals for non-invasive coronary angiography with 64-slice CT. Eur Radiol 16(9):1964–1972PubMedCrossRef Leschka S, Husmann L, Desbiolles LM, Gaemperli O, Schepis T, Koepfli P et al (2006) Optimal image reconstruction intervals for non-invasive coronary angiography with 64-slice CT. Eur Radiol 16(9):1964–1972PubMedCrossRef
39.
Zurück zum Zitat Schueler BA (2000) The AAPM/RSNA physics tutorial for residents: general overview of fluoroscopic imaging. Radiographics 20(4):1115–1126PubMed Schueler BA (2000) The AAPM/RSNA physics tutorial for residents: general overview of fluoroscopic imaging. Radiographics 20(4):1115–1126PubMed
40.
Zurück zum Zitat Ropers U, Ropers D, Pflederer T, Anders K, Kuettner A, Stilianakis NI et al (2007) Influence of heart rate on the diagnostic accuracy of dual-source computed tomography coronary angiography. J Am Coll Cardiol 50(25):2393–2398PubMedCrossRef Ropers U, Ropers D, Pflederer T, Anders K, Kuettner A, Stilianakis NI et al (2007) Influence of heart rate on the diagnostic accuracy of dual-source computed tomography coronary angiography. J Am Coll Cardiol 50(25):2393–2398PubMedCrossRef
41.
Zurück zum Zitat Woodhouse CE, Janowitz WR, Viamonte M Jr (1997) Coronary arteries: retrospective cardiac gating technique to reduce cardiac motion artifact at spiral CT. Radiology 204(2):566–569PubMed Woodhouse CE, Janowitz WR, Viamonte M Jr (1997) Coronary arteries: retrospective cardiac gating technique to reduce cardiac motion artifact at spiral CT. Radiology 204(2):566–569PubMed
42.
Zurück zum Zitat Herzog C, Nguyen SA, Savino G, Zwerner PL, Doll J, Nielsen CD et al (2007) Does two-segment image reconstruction at 64-section CT coronary angiography improve image quality and diagnostic accuracy? Radiology 244(1):121–129PubMedCrossRef Herzog C, Nguyen SA, Savino G, Zwerner PL, Doll J, Nielsen CD et al (2007) Does two-segment image reconstruction at 64-section CT coronary angiography improve image quality and diagnostic accuracy? Radiology 244(1):121–129PubMedCrossRef
43.
Zurück zum Zitat Shu KM, MacKenzie JD, Smith JB, Blinder EM, Bourgeois LM, Ledbetter S et al (2006) Lowering the thyroid dose in screening examinations of the cervical spine. Emerg Radiol 12(3):133–136PubMedCrossRef Shu KM, MacKenzie JD, Smith JB, Blinder EM, Bourgeois LM, Ledbetter S et al (2006) Lowering the thyroid dose in screening examinations of the cervical spine. Emerg Radiol 12(3):133–136PubMedCrossRef
44.
Zurück zum Zitat Einstein AJ, Henzlova MJ, Rajagopalan S (2007) Estimating risk of cancer associated with radiation exposure from 64-slice computed tomography coronary angiography. JAMA 298(3):317–323PubMedCrossRef Einstein AJ, Henzlova MJ, Rajagopalan S (2007) Estimating risk of cancer associated with radiation exposure from 64-slice computed tomography coronary angiography. JAMA 298(3):317–323PubMedCrossRef
45.
Zurück zum Zitat Jakobs TF, Becker CR, Ohnesorge B, Flohr T, Suess C, Schoepf UJ et al (2002) Multislice helical CT of the heart with retrospective ECG gating: reduction of radiation exposure by ECG-controlled tube current modulation. Eur Radiol 12(5):1081–1086PubMedCrossRef Jakobs TF, Becker CR, Ohnesorge B, Flohr T, Suess C, Schoepf UJ et al (2002) Multislice helical CT of the heart with retrospective ECG gating: reduction of radiation exposure by ECG-controlled tube current modulation. Eur Radiol 12(5):1081–1086PubMedCrossRef
46.
Zurück zum Zitat Paul JF, Abada HT (2007) Strategies for reduction of radiation dose in cardiac multislice CT. Eur Radiol 17(8):2028–2037PubMedCrossRef Paul JF, Abada HT (2007) Strategies for reduction of radiation dose in cardiac multislice CT. Eur Radiol 17(8):2028–2037PubMedCrossRef
47.
Zurück zum Zitat Weinreb JC, Larson PA, Woodard PK, Stanford W, Rubin GD, Stillman AE et al (2005) ACR clinical statement on noninvasive cardiac imaging. J Am Coll Radiol 2(6):471–477PubMedCrossRef Weinreb JC, Larson PA, Woodard PK, Stanford W, Rubin GD, Stillman AE et al (2005) ACR clinical statement on noninvasive cardiac imaging. J Am Coll Radiol 2(6):471–477PubMedCrossRef
Metadaten
Titel
Initial evaluation of coronary images from 320-detector row computed tomography
verfasst von
Frank J. Rybicki
Hansel J. Otero
Michael L. Steigner
Gabriel Vorobiof
Leelakrishna Nallamshetty
Dimitrios Mitsouras
Hale Ersoy
Richard T. Mather
Philip F. Judy
Tianxi Cai
Karl Coyner
Kurt Schultz
Amanda G. Whitmore
Marcelo F. Di Carli
Publikationsdatum
01.06.2008
Verlag
Springer Netherlands
Erschienen in
The International Journal of Cardiovascular Imaging / Ausgabe 5/2008
Print ISSN: 1569-5794
Elektronische ISSN: 1875-8312
DOI
https://doi.org/10.1007/s10554-008-9308-2

Weitere Artikel der Ausgabe 5/2008

The International Journal of Cardiovascular Imaging 5/2008 Zur Ausgabe

Update Kardiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.