Mechanical circulatory support: devices, outcomes and complications

verfasst von: Carmelo A. Milano, Alan A. Simeone

Erschienen in: Heart Failure Reviews | Ausgabe 1/2013

Einloggen, um Zugang zu erhalten

Abstract

Systolic heart failure is a problem of substantial magnitude worldwide. Over the last 25 years great progress has been made in the medical management of heart failure with the recognition of the benefits of beta-adrenergic blockade, modulation of the renin-angiotensin and mineralocorticoid axes and judicious diuretic therapy. In addition, cardiac resynchronization therapy and prophylactic implantation of cardiac defibrillators have been responsible for measurable benefits in terms of functional status and dysrhythmia-related mortality, respectively. Unfortunately, progressive cardiac dysfunction often results in activity limitation, symptoms at rest, hospital admission, end-organ dysfunction and death despite maximal implementation of standard therapies. Heart transplantation has been a dramatic and effective therapy for end-stage heart failure, but it remains limited by a shortage of donor organs, strict criteria defining acceptable recipients and often unsatisfactory long-term success. Mechanical alternatives to support the failing circulation have been sought for the last 50 years. The history of device development has been marked in general by the slow progress achieved by a few dedicated and persevering pioneers. In the past decade, however, evolving technology has dramatically changed the field and broadened the options for the treatment of advanced heart failure. This review will detail the important milestones and the current state of the art, with an emphasis on implantable devices for intermediate to long term support.
Literatur
1.
Zurück zum Zitat Liotta D, Hall CW, Henly WS, Cooley DA, Crawford ES, DeBakey ME (1963) Prolonged assisted circulation during and after cardiac or aortic surgery: prolonged partial left ventricular bypass by means of intracorporeal circulation. Am J Cardiol 12(3):399–405PubMedCrossRef Liotta D, Hall CW, Henly WS, Cooley DA, Crawford ES, DeBakey ME (1963) Prolonged assisted circulation during and after cardiac or aortic surgery: prolonged partial left ventricular bypass by means of intracorporeal circulation. Am J Cardiol 12(3):399–405PubMedCrossRef
2.
Zurück zum Zitat DeBakey ME (1971) Left ventricular bypass pump for cardiac assistance. Clinical experience. Am J Cardiol 27(1):3–11PubMedCrossRef DeBakey ME (1971) Left ventricular bypass pump for cardiac assistance. Clinical experience. Am J Cardiol 27(1):3–11PubMedCrossRef
3.
Zurück zum Zitat Cooley DA, Liotta D, Hallman GL, Bloodwell RD, Leachman RD, Milam JD (1969) Orthotopic cardiac prosthesis for two-staged cardiac replacement. Am J Cardiol 24(5):723–730PubMedCrossRef Cooley DA, Liotta D, Hallman GL, Bloodwell RD, Leachman RD, Milam JD (1969) Orthotopic cardiac prosthesis for two-staged cardiac replacement. Am J Cardiol 24(5):723–730PubMedCrossRef
4.
Zurück zum Zitat Gaitan BD, Thunberg CA, Stansbury LG, Jaroszewski DE, Arabia FA, Griffith BP, Grigore AM (2011) Development, current status, and anesthetic management of the implanted artificial heart. J Cardiothorac Vasc Anesth 25(6):1179–1192PubMedCrossRef Gaitan BD, Thunberg CA, Stansbury LG, Jaroszewski DE, Arabia FA, Griffith BP, Grigore AM (2011) Development, current status, and anesthetic management of the implanted artificial heart. J Cardiothorac Vasc Anesth 25(6):1179–1192PubMedCrossRef
5.
Zurück zum Zitat Hill JD (1986) Use of a prosthetic ventricle as a bridge to cardiac transplantation for postinfarction cardiogenic shock. N Engl J Med 314(10):626–628PubMedCrossRef Hill JD (1986) Use of a prosthetic ventricle as a bridge to cardiac transplantation for postinfarction cardiogenic shock. N Engl J Med 314(10):626–628PubMedCrossRef
6.
Zurück zum Zitat Pierce WS, Parr GVS, Myers JL, Pae WE, Bull AP, Waldhausen JA (1981) Ventricular-assist pumping in patients with cardiogenic shock after cardiac operations. N Engl J Med 305(27):1606–1610PubMedCrossRef Pierce WS, Parr GVS, Myers JL, Pae WE, Bull AP, Waldhausen JA (1981) Ventricular-assist pumping in patients with cardiogenic shock after cardiac operations. N Engl J Med 305(27):1606–1610PubMedCrossRef
7.
Zurück zum Zitat McBride LR, Naunheim KS, Fiore AC, Moroney DA, Swartz MT (1999) Clinical experience with 111 thoratec ventricular assist devices. Ann Thorac Surg 67(5):1233–1238PubMedCrossRef McBride LR, Naunheim KS, Fiore AC, Moroney DA, Swartz MT (1999) Clinical experience with 111 thoratec ventricular assist devices. Ann Thorac Surg 67(5):1233–1238PubMedCrossRef
8.
Zurück zum Zitat Korfer R, El-Banayosy A, Arusoglu L, Minami K, Korner MM, Kizner L, Fey O, Schutt U, Morshuis M, Posival H (2000) Single center experience with the thoratec ventricular assist device. J Thorac Cardiovasc Surg 119(3):596–600PubMedCrossRef Korfer R, El-Banayosy A, Arusoglu L, Minami K, Korner MM, Kizner L, Fey O, Schutt U, Morshuis M, Posival H (2000) Single center experience with the thoratec ventricular assist device. J Thorac Cardiovasc Surg 119(3):596–600PubMedCrossRef
9.
Zurück zum Zitat Schoeb R (2000) A bearingless motor for a left ventricular assist device. Paper presented at the International Symposium on Magnetic Bearings, Zurich, Switzerland, August 23–25, 2000 Schoeb R (2000) A bearingless motor for a left ventricular assist device. Paper presented at the International Symposium on Magnetic Bearings, Zurich, Switzerland, August 23–25, 2000
10.
Zurück zum Zitat McCormick AJ, Reid BB, Budge D, Clayson SE, Revenaugh JR, Stoker S, Raymond AL, Smith HK, Doty JR, Caine WT, Alharethi R, Kfoury AG (2010) 296: versatility and patient outcomes with the levitronix centrimag temporary ventricular assist device. J Heart Lung Transpl 29(2, Supplement 1):S100 McCormick AJ, Reid BB, Budge D, Clayson SE, Revenaugh JR, Stoker S, Raymond AL, Smith HK, Doty JR, Caine WT, Alharethi R, Kfoury AG (2010) 296: versatility and patient outcomes with the levitronix centrimag temporary ventricular assist device. J Heart Lung Transpl 29(2, Supplement 1):S100
11.
Zurück zum Zitat De Robertis F, Birks EJ, Rogers P, Dreyfus G, Pepper JR, Khaghani A (2006) Clinical performance with the Levitronix Centrimag short-term ventricular assist device. J Heart Lung Transpl 25(2):181–186CrossRef De Robertis F, Birks EJ, Rogers P, Dreyfus G, Pepper JR, Khaghani A (2006) Clinical performance with the Levitronix Centrimag short-term ventricular assist device. J Heart Lung Transpl 25(2):181–186CrossRef
12.
Zurück zum Zitat John R, Liao K, Lietz K, Kamdar F, Colvin-Adams M, Boyle A, Miller L, Joyce L (2007) Experience with the Levitronix CentriMag circulatory support system as a bridge to decision in patients with refractory acute cardiogenic shock and multisystem organ failure. J Thorac Cardiovasc Surg 134(2):351–358PubMedCrossRef John R, Liao K, Lietz K, Kamdar F, Colvin-Adams M, Boyle A, Miller L, Joyce L (2007) Experience with the Levitronix CentriMag circulatory support system as a bridge to decision in patients with refractory acute cardiogenic shock and multisystem organ failure. J Thorac Cardiovasc Surg 134(2):351–358PubMedCrossRef
13.
Zurück zum Zitat Worku B, Takayama H, Van Patten D, Pak P, Naka Y (2010) The Columbia experience with the Levitronix CentriMag as a rescue device in high risk patients. J Heart Lung Transpl 29(2, Supplement 1):S9 Worku B, Takayama H, Van Patten D, Pak P, Naka Y (2010) The Columbia experience with the Levitronix CentriMag as a rescue device in high risk patients. J Heart Lung Transpl 29(2, Supplement 1):S9
14.
Zurück zum Zitat Bhama JK, Kormos RL, Toyoda Y, Teuteberg JJ, McCurry KR, Siegenthaler MP (2009) Clinical experience using the Levitronix Centrimag system for temporary right ventricular mechanical circulatory support. J Heart Lung Transpl 28(9):971–976CrossRef Bhama JK, Kormos RL, Toyoda Y, Teuteberg JJ, McCurry KR, Siegenthaler MP (2009) Clinical experience using the Levitronix Centrimag system for temporary right ventricular mechanical circulatory support. J Heart Lung Transpl 28(9):971–976CrossRef
15.
Zurück zum Zitat Loforte A, Montalto A, Lilla Della Monica P, Musumeci F (2010) Simultaneous temporary CentriMag right ventricular assist device placement in HeartMate II left ventricular assist system recipients at high risk of right ventricular failure. Interact CardioVasc Thorac Surg 10(6):847–850PubMedCrossRef Loforte A, Montalto A, Lilla Della Monica P, Musumeci F (2010) Simultaneous temporary CentriMag right ventricular assist device placement in HeartMate II left ventricular assist system recipients at high risk of right ventricular failure. Interact CardioVasc Thorac Surg 10(6):847–850PubMedCrossRef
16.
Zurück zum Zitat Stepanenko A, Loforte A, Dranishnikov N, Potapov EV, Montalto A, Pasic M, Weng YG, Dandel M, Siniawski H, Kukucka M, Krabatsch T, Musumeci F, Hetzer R (2011) 468 temporary right ventricular support in left ventricular assist device recipients. J Heart Lung Transpl 30(4, Supplement 1):S159–S160 Stepanenko A, Loforte A, Dranishnikov N, Potapov EV, Montalto A, Pasic M, Weng YG, Dandel M, Siniawski H, Kukucka M, Krabatsch T, Musumeci F, Hetzer R (2011) 468 temporary right ventricular support in left ventricular assist device recipients. J Heart Lung Transpl 30(4, Supplement 1):S159–S160
17.
Zurück zum Zitat Mehra MR, Kobashigawa J, Starling R, Russell S, Uber PA, Parameshwar J, Mohacsi P, Augustine S, Aaronson K, Barr M (2006) Listing criteria for heart transplantation: International society for heart and lung transplantation guidelines for the care of cardiac transplant candidates–2006. J Heart Lung Transpl 25(9):1024–1042CrossRef Mehra MR, Kobashigawa J, Starling R, Russell S, Uber PA, Parameshwar J, Mohacsi P, Augustine S, Aaronson K, Barr M (2006) Listing criteria for heart transplantation: International society for heart and lung transplantation guidelines for the care of cardiac transplant candidates–2006. J Heart Lung Transpl 25(9):1024–1042CrossRef
18.
Zurück zum Zitat Rose EA, Gelijns AC, Moskowitz AJ, Heitjan DF, Stevenson LW, Dembitsky W, Long JW, Ascheim DD, Tierney AR, Levitan RG, Watson JT, Meier P, Ronan NS, Shapiro PA, Lazar RM, Miller LW, Gupta L, Frazier OH, Desvigne-Nickens P, Oz MC, Poirier VL, Randomized Evaluation of Mechanical Assistance for the Treatment of Congestive Heart Failure Study G (2001) Long-term use of a left ventricular assist device for end-stage heart failure. N Engl J Med 345(20):1435–1443PubMedCrossRef Rose EA, Gelijns AC, Moskowitz AJ, Heitjan DF, Stevenson LW, Dembitsky W, Long JW, Ascheim DD, Tierney AR, Levitan RG, Watson JT, Meier P, Ronan NS, Shapiro PA, Lazar RM, Miller LW, Gupta L, Frazier OH, Desvigne-Nickens P, Oz MC, Poirier VL, Randomized Evaluation of Mechanical Assistance for the Treatment of Congestive Heart Failure Study G (2001) Long-term use of a left ventricular assist device for end-stage heart failure. N Engl J Med 345(20):1435–1443PubMedCrossRef
19.
Zurück zum Zitat Nativi JN, Drakos SG, Kucheryavaya AY, Edwards LB, Selzman CH, Taylor DO, Hertz MI, Kfoury AG, Stehlik J (2011) Changing outcomes in patients bridged to heart transplantation with continuous—versus pulsatile-flow ventricular assist devices: an analysis of the registry of the International Society for Heart and Lung Transplantation. J Heart Lung Transpl 30(8):854–861 Nativi JN, Drakos SG, Kucheryavaya AY, Edwards LB, Selzman CH, Taylor DO, Hertz MI, Kfoury AG, Stehlik J (2011) Changing outcomes in patients bridged to heart transplantation with continuous—versus pulsatile-flow ventricular assist devices: an analysis of the registry of the International Society for Heart and Lung Transplantation. J Heart Lung Transpl 30(8):854–861
20.
Zurück zum Zitat Long JW, Healy AH, Rasmusson BY, Cowley CG, Nelson KE, Kfoury AG, Clayson SE, Reid BB, Moore SA, Blank DU, Renlund DG (2008) Improving outcomes with long-term “destination” therapy using left ventricular assist devices. J Thorac Cardiovasc Surg 135(6):1353–1360; discussion 1360–1351 Long JW, Healy AH, Rasmusson BY, Cowley CG, Nelson KE, Kfoury AG, Clayson SE, Reid BB, Moore SA, Blank DU, Renlund DG (2008) Improving outcomes with long-term “destination” therapy using left ventricular assist devices. J Thorac Cardiovasc Surg 135(6):1353–1360; discussion 1360–1351
21.
Zurück zum Zitat Kamdar F, Boyle A, Liao K, Colvin-adams M, Joyce L, John R (2009) Effects of centrifugal, axial, and pulsatile left ventricular assist device support on end-organ function in heart failure patients. J Heart Lung Transpl 28(4):352–359CrossRef Kamdar F, Boyle A, Liao K, Colvin-adams M, Joyce L, John R (2009) Effects of centrifugal, axial, and pulsatile left ventricular assist device support on end-organ function in heart failure patients. J Heart Lung Transpl 28(4):352–359CrossRef
22.
Zurück zum Zitat Russell SD, Rogers JG, Milano CA, Dyke DB, Pagani FD, Aranda JM, Klodell CT Jr, Boyle AJ, John R, Chen L, Massey HT, Farrar DJ, Conte JV (2009) Renal and hepatic function improve in advanced heart failure patients during continuous-flow support with the HeartMate II left ventricular assist device. Circulation 120(23):2352–2357PubMedCrossRef Russell SD, Rogers JG, Milano CA, Dyke DB, Pagani FD, Aranda JM, Klodell CT Jr, Boyle AJ, John R, Chen L, Massey HT, Farrar DJ, Conte JV (2009) Renal and hepatic function improve in advanced heart failure patients during continuous-flow support with the HeartMate II left ventricular assist device. Circulation 120(23):2352–2357PubMedCrossRef
23.
Zurück zum Zitat Drews T, Jurmann M, Michael D, Miralem P, Weng Y, Hetzer R (2008) Differences in pulsatile and non-pulsatile mechanical circulatory support in long-term use. J Heart Lung Transpl 27(10):1096–1101CrossRef Drews T, Jurmann M, Michael D, Miralem P, Weng Y, Hetzer R (2008) Differences in pulsatile and non-pulsatile mechanical circulatory support in long-term use. J Heart Lung Transpl 27(10):1096–1101CrossRef
24.
Zurück zum Zitat Radovancevic B, Vrtovec B, de Kort E, Radovancevic R, Gregoric ID, Frazier OH (2007) End-organ function in patients on long-term circulatory support with continuous- or pulsatile-flow assist devices. J Heart Lung Transpl 26(8):815–818CrossRef Radovancevic B, Vrtovec B, de Kort E, Radovancevic R, Gregoric ID, Frazier OH (2007) End-organ function in patients on long-term circulatory support with continuous- or pulsatile-flow assist devices. J Heart Lung Transpl 26(8):815–818CrossRef
25.
Zurück zum Zitat Rogers JG, Aaronson KD, Boyle AJ, Russell SD, Milano CA, Pagani FD, Edwards BS, Park S, John R, Conte JV, Farrar DJ, Slaughter MS (2010) Continuous flow left ventricular assist device improves functional capacity and quality of life of advanced heart failure patients. J Am Coll Cardiol 55(17):1826–1834PubMedCrossRef Rogers JG, Aaronson KD, Boyle AJ, Russell SD, Milano CA, Pagani FD, Edwards BS, Park S, John R, Conte JV, Farrar DJ, Slaughter MS (2010) Continuous flow left ventricular assist device improves functional capacity and quality of life of advanced heart failure patients. J Am Coll Cardiol 55(17):1826–1834PubMedCrossRef
26.
Zurück zum Zitat Aranda Jr JM, Rogers JG, Aronson KD, Boyle AJ, Russell SD, Edwards B, Zucker MJ, Klodell Jr CT, Sundareswaran KS, Farrar DJ, Slaughter MS, HeartMate IICI (2010) Quality of life improvements are greater in destination therapy than bridge to transplant patients with a continuous flow left ventricular assist device. J Am Coll Cardiol 55(10, Supplement 1):A22.E210–A222.E210 Aranda Jr JM, Rogers JG, Aronson KD, Boyle AJ, Russell SD, Edwards B, Zucker MJ, Klodell Jr CT, Sundareswaran KS, Farrar DJ, Slaughter MS, HeartMate IICI (2010) Quality of life improvements are greater in destination therapy than bridge to transplant patients with a continuous flow left ventricular assist device. J Am Coll Cardiol 55(10, Supplement 1):A22.E210–A222.E210
27.
Zurück zum Zitat Fabienne D, Hilde B, Bart M (2011) To “pump up” quality of life and exercise performance studies in patients with left ventricular assist devices. J Heart Lung Transpl 30(2):141–143CrossRef Fabienne D, Hilde B, Bart M (2011) To “pump up” quality of life and exercise performance studies in patients with left ventricular assist devices. J Heart Lung Transpl 30(2):141–143CrossRef
28.
Zurück zum Zitat Kugler C, Malehsa D, Tegtbur U, Guetzlaff E, Meyer AL, Bara C, Haverich A, Strueber M (2011) Health-related quality of life and exercise tolerance in recipients of heart transplants and left ventricular assist devices: a prospective, comparative study. J Heart Lung Transpl 30(2):204–210CrossRef Kugler C, Malehsa D, Tegtbur U, Guetzlaff E, Meyer AL, Bara C, Haverich A, Strueber M (2011) Health-related quality of life and exercise tolerance in recipients of heart transplants and left ventricular assist devices: a prospective, comparative study. J Heart Lung Transpl 30(2):204–210CrossRef
29.
Zurück zum Zitat Strueber M, O’Driscoll G, Jansz P, Khaghani A, Levy WC, Wieselthaler GM, Heart Ware Investigators (2011) Multicenter evaluation of an intrapericardial left ventricular assist system. J Am Coll Cardiol 57(12):1375–1382PubMedCrossRef Strueber M, O’Driscoll G, Jansz P, Khaghani A, Levy WC, Wieselthaler GM, Heart Ware Investigators (2011) Multicenter evaluation of an intrapericardial left ventricular assist system. J Am Coll Cardiol 57(12):1375–1382PubMedCrossRef
30.
Zurück zum Zitat Lietz K, Long JW, Kfoury AG, Slaughter MS, Silver MA, Milano CA, Rogers JG, Naka Y, Mancini D, Miller LW (2007) Outcomes of left ventricular assist device implantation as destination therapy in the post-REMATCH era. Circulation 116(5):497–505PubMedCrossRef Lietz K, Long JW, Kfoury AG, Slaughter MS, Silver MA, Milano CA, Rogers JG, Naka Y, Mancini D, Miller LW (2007) Outcomes of left ventricular assist device implantation as destination therapy in the post-REMATCH era. Circulation 116(5):497–505PubMedCrossRef
31.
Zurück zum Zitat Boyle AJ, Russell SD, Teuteberg JJ, Slaughter MS, Moazami N, Pagani FD, Frazier OH, Heatley G, Farrar DJ, John R (2009) Low thromboembolism and pump thrombosis with the HeartMate II left ventricular assist device: analysis of outpatient anti-coagulation. J Heart Lung Transpl 28(9):881–887CrossRef Boyle AJ, Russell SD, Teuteberg JJ, Slaughter MS, Moazami N, Pagani FD, Frazier OH, Heatley G, Farrar DJ, John R (2009) Low thromboembolism and pump thrombosis with the HeartMate II left ventricular assist device: analysis of outpatient anti-coagulation. J Heart Lung Transpl 28(9):881–887CrossRef
32.
Zurück zum Zitat Miller LW, Pagani FD, Russell SD, John R, Boyle AJ, Aaronson KD, Conte JV, Naka Y, Mancini D, Delgado RM, MacGillivray TE, Farrar DJ, Frazier OH (2007) Use of a continuous-flow device in patients awaiting heart transplantation. N Engl J Med 357(9):885–896PubMedCrossRef Miller LW, Pagani FD, Russell SD, John R, Boyle AJ, Aaronson KD, Conte JV, Naka Y, Mancini D, Delgado RM, MacGillivray TE, Farrar DJ, Frazier OH (2007) Use of a continuous-flow device in patients awaiting heart transplantation. N Engl J Med 357(9):885–896PubMedCrossRef
33.
Zurück zum Zitat Pagani FD, Miller LW, Russell SD, Aaronson KD, John R, Boyle AJ, Conte JV, Bogaev RC, MacGillivray TE, Naka Y, Mancini D, Massey HT, Chen L, Klodell CT, Aranda JM, Moazami N, Ewald GA, Farrar DJ, Frazier OH (2009) Extended mechanical circulatory support with a continuous-flow rotary left ventricular assist device. J Am Coll Cardiol 54(4):312–321PubMedCrossRef Pagani FD, Miller LW, Russell SD, Aaronson KD, John R, Boyle AJ, Conte JV, Bogaev RC, MacGillivray TE, Naka Y, Mancini D, Massey HT, Chen L, Klodell CT, Aranda JM, Moazami N, Ewald GA, Farrar DJ, Frazier OH (2009) Extended mechanical circulatory support with a continuous-flow rotary left ventricular assist device. J Am Coll Cardiol 54(4):312–321PubMedCrossRef
34.
Zurück zum Zitat Starling RC, Naka Y, Boyle AJ, Gonzalez-Stawinski G, John R, Jorde U, Russell SD, Conte JV, Aaronson KD, McGee EC Jr, Cotts WG, DeNofrio D, Pham DT, Farrar DJ, Pagani FD (2011) Results of the post-U.S. Food and Drug Administration-approval study with a continuous flow left ventricular assist device as a bridge to heart transplantation: a prospective study using the INTERMACS (Interagency Registry for Mechanically Assisted Circulatory Support). J Am Coll Cardiol 57(19):1890–1898 Starling RC, Naka Y, Boyle AJ, Gonzalez-Stawinski G, John R, Jorde U, Russell SD, Conte JV, Aaronson KD, McGee EC Jr, Cotts WG, DeNofrio D, Pham DT, Farrar DJ, Pagani FD (2011) Results of the post-U.S. Food and Drug Administration-approval study with a continuous flow left ventricular assist device as a bridge to heart transplantation: a prospective study using the INTERMACS (Interagency Registry for Mechanically Assisted Circulatory Support). J Am Coll Cardiol 57(19):1890–1898
35.
Zurück zum Zitat John R, Naka Y, Smedira N (2011) Outcomes with a continuous flow left ventricular assist device in over 1,000 patients in commercial use compared to clinical trial experience. Paper presented at the Society of Thoracic Surgeons, San Diego, CA John R, Naka Y, Smedira N (2011) Outcomes with a continuous flow left ventricular assist device in over 1,000 patients in commercial use compared to clinical trial experience. Paper presented at the Society of Thoracic Surgeons, San Diego, CA
36.
Zurück zum Zitat Slaughter MS, Rogers JG, Milano CA, Russell SD, Conte JV, Feldman D, Sun B, Tatooles AJ, Delgado RM III, Long JW, Wozniak TC, Ghumman W, Farrar DJ, Frazier OH (2009) Advanced heart failure treated with continuous-flow left ventricular assist device. N Engl J Med 361(23):2241–2251PubMedCrossRef Slaughter MS, Rogers JG, Milano CA, Russell SD, Conte JV, Feldman D, Sun B, Tatooles AJ, Delgado RM III, Long JW, Wozniak TC, Ghumman W, Farrar DJ, Frazier OH (2009) Advanced heart failure treated with continuous-flow left ventricular assist device. N Engl J Med 361(23):2241–2251PubMedCrossRef
37.
Zurück zum Zitat Park SJ, Milano CA, Tatooles AJ, Rogers JG, Adamson RM, Steidley DE, Ewald GA, Farrar DJ, Slaughter MS, The HeartMate II Clinical Investigators (2010) Abstract 14430: outcomes in advanced heart failure with left ventricular assist devices for destination therapy. Circulation 122(21 Supplement):A14430 Park SJ, Milano CA, Tatooles AJ, Rogers JG, Adamson RM, Steidley DE, Ewald GA, Farrar DJ, Slaughter MS, The HeartMate II Clinical Investigators (2010) Abstract 14430: outcomes in advanced heart failure with left ventricular assist devices for destination therapy. Circulation 122(21 Supplement):A14430
38.
Zurück zum Zitat Goldstein DJ (2003) Worldwide experience with the MicroMed DeBakey Ventricular Assist Device® as a bridge to transplantation. Circulation 108(10 suppl 1):II-272–II-277 Goldstein DJ (2003) Worldwide experience with the MicroMed DeBakey Ventricular Assist Device® as a bridge to transplantation. Circulation 108(10 suppl 1):II-272–II-277
39.
Zurück zum Zitat Goldstein DJ, Zucker M, Arroyo L, McCarthy P, Noon G (2003) The Micromed-Debakey VAD: initial American experience. J Heart Lung Transpl 22(1, Supplement 1):S218 Goldstein DJ, Zucker M, Arroyo L, McCarthy P, Noon G (2003) The Micromed-Debakey VAD: initial American experience. J Heart Lung Transpl 22(1, Supplement 1):S218
40.
Zurück zum Zitat Rothenburger M, Wilhelm MJ, Hammel D, Schmidt C, Tjan TDT, Böcker D, Scheld HH, Schmid C (2002) Treatment of thrombus formation associated with the MicroMed DeBakey VAD using recombinant tissue plasminogen activator. Circulation 106(12 suppl 1):I-189–I-192 Rothenburger M, Wilhelm MJ, Hammel D, Schmidt C, Tjan TDT, Böcker D, Scheld HH, Schmid C (2002) Treatment of thrombus formation associated with the MicroMed DeBakey VAD using recombinant tissue plasminogen activator. Circulation 106(12 suppl 1):I-189–I-192
41.
Zurück zum Zitat Loebe M, Noon GP, Benkowski R, Gruber P, Richards T (2007) 103: Improved clinical results with the redesigned MicroMed DeBakey VAD. J Heart Lung Transpl 26(2, Supplement 1):S96–S97 Loebe M, Noon GP, Benkowski R, Gruber P, Richards T (2007) 103: Improved clinical results with the redesigned MicroMed DeBakey VAD. J Heart Lung Transpl 26(2, Supplement 1):S96–S97
42.
Zurück zum Zitat Haj-Yahia S, Birks EJ, Rogers P, Bowles C, Hipkins M, George R, Amrani M, Petrou M, Pepper J, Dreyfus G, Khaghani A (2007) Midterm experience with the Jarvik 2000 axial flow left ventricular assist device. J Thorac Cardiovasc Surg 134(1):199–203PubMedCrossRef Haj-Yahia S, Birks EJ, Rogers P, Bowles C, Hipkins M, George R, Amrani M, Petrou M, Pepper J, Dreyfus G, Khaghani A (2007) Midterm experience with the Jarvik 2000 axial flow left ventricular assist device. J Thorac Cardiovasc Surg 134(1):199–203PubMedCrossRef
43.
Zurück zum Zitat Aronson K, Slaughter M, McGee E (2010) Evaluation of the HeartWare HVAD left ventricular assist system for the treatment of advanced heart failure: results of the ADVANCE Bridge to transplant trial. Paper presented at the American Heart Association Scientific Sessions 2010, Chicago, Il, November 2010 Aronson K, Slaughter M, McGee E (2010) Evaluation of the HeartWare HVAD left ventricular assist system for the treatment of advanced heart failure: results of the ADVANCE Bridge to transplant trial. Paper presented at the American Heart Association Scientific Sessions 2010, Chicago, Il, November 2010
44.
Zurück zum Zitat Slaughter M, Aronson K, Boyce S, Miller L (2011) Evaluation of the HeartWare HVAD left ventricular assist system for the treatment of advanced heart failure: results of the ADVANCE bridge to transplant trial with continued access patients. Paper presented at the International Society of Heart and Lung Transplantation, San Diego, CA, April 2011 Slaughter M, Aronson K, Boyce S, Miller L (2011) Evaluation of the HeartWare HVAD left ventricular assist system for the treatment of advanced heart failure: results of the ADVANCE bridge to transplant trial with continued access patients. Paper presented at the International Society of Heart and Lung Transplantation, San Diego, CA, April 2011
45.
Zurück zum Zitat Morshuis M, El-Banayosy A, Arusoglu L, Koerfer R, Hetzer R, Wieselthaler G, Pavie A, Nojiri C (2009) European experience of DuraHeartTM magnetically levitated centrifugal left ventricular assist system. Eur J Cardiothorac Surg 35(6):1020–1028PubMedCrossRef Morshuis M, El-Banayosy A, Arusoglu L, Koerfer R, Hetzer R, Wieselthaler G, Pavie A, Nojiri C (2009) European experience of DuraHeartTM magnetically levitated centrifugal left ventricular assist system. Eur J Cardiothorac Surg 35(6):1020–1028PubMedCrossRef
46.
Zurück zum Zitat Loree HM, Bourque K, Gernes DB, Richardson JS, Poirier VL, Barletta N, Fleischli A, Foiera G, Gempp TM, Schoeb R, Litwak KN, Akimoto T, Kameneva M, Watach MJ, Litwak P (2001) The HeartMate III: design and in vivo studies of a maglev centrifugal left ventricular assist device. Artif Organs 25(5):386–391PubMedCrossRef Loree HM, Bourque K, Gernes DB, Richardson JS, Poirier VL, Barletta N, Fleischli A, Foiera G, Gempp TM, Schoeb R, Litwak KN, Akimoto T, Kameneva M, Watach MJ, Litwak P (2001) The HeartMate III: design and in vivo studies of a maglev centrifugal left ventricular assist device. Artif Organs 25(5):386–391PubMedCrossRef
47.
Zurück zum Zitat Jaroszewski DE, Anderson EM, Pierce CN, Arabia FA (2011) The SynCardia freedom driver: a portable driver for discharge home with the total artificial heart. J Heart Lung Transplant 30(7):844–845PubMedCrossRef Jaroszewski DE, Anderson EM, Pierce CN, Arabia FA (2011) The SynCardia freedom driver: a portable driver for discharge home with the total artificial heart. J Heart Lung Transplant 30(7):844–845PubMedCrossRef
48.
Zurück zum Zitat Slepian MJ (2011) The SynCardia temporary total artificial heart—evolving clinical role and future status. US Cardiol 8(1):39–46 Slepian MJ (2011) The SynCardia temporary total artificial heart—evolving clinical role and future status. US Cardiol 8(1):39–46
49.
Zurück zum Zitat Copeland JG, Smith RG, Arabia FA, Nolan PE, Sethi GK, Tsau PH, McClellan D, Slepian MJ (2004) Cardiac replacement with a total artificial heart as a bridge to transplantation. N Engl J Med 351(9):859–867PubMedCrossRef Copeland JG, Smith RG, Arabia FA, Nolan PE, Sethi GK, Tsau PH, McClellan D, Slepian MJ (2004) Cardiac replacement with a total artificial heart as a bridge to transplantation. N Engl J Med 351(9):859–867PubMedCrossRef
50.
Zurück zum Zitat Roussel JC, Sénage T, Baron O, Périgaud C, Habash O, Rigal JC, Treilhaud M, Trochu JN, Despins P, Duveau D (2009) CardioWest (Jarvik) total artificial heart: a single-center experience with 42 patients. Ann Thorac Surg 87(1):124–130PubMedCrossRef Roussel JC, Sénage T, Baron O, Périgaud C, Habash O, Rigal JC, Treilhaud M, Trochu JN, Despins P, Duveau D (2009) CardioWest (Jarvik) total artificial heart: a single-center experience with 42 patients. Ann Thorac Surg 87(1):124–130PubMedCrossRef
51.
Zurück zum Zitat Dowling RD, Gray LA, Etoch SW, Laks H, Marelli D, Samuels L, Entwistle J, Couper G, Vlahakes GJ, Frazier OH (2004) Initial experience with the AbioCor implantable replacement heart system. J Thorac Cardiovasc Surg 127(1):131–141PubMedCrossRef Dowling RD, Gray LA, Etoch SW, Laks H, Marelli D, Samuels L, Entwistle J, Couper G, Vlahakes GJ, Frazier OH (2004) Initial experience with the AbioCor implantable replacement heart system. J Thorac Cardiovasc Surg 127(1):131–141PubMedCrossRef
52.
Zurück zum Zitat Topkara VK, Kondareddy S, Malik F, Wang IW, Mann DL, Ewald GA, Moazami N (2010) Infectious complications in patients with left ventricular assist device: etiology and outcomes in the continuous-flow era. Ann Thorac Surg 90(4):1270–1277PubMedCrossRef Topkara VK, Kondareddy S, Malik F, Wang IW, Mann DL, Ewald GA, Moazami N (2010) Infectious complications in patients with left ventricular assist device: etiology and outcomes in the continuous-flow era. Ann Thorac Surg 90(4):1270–1277PubMedCrossRef
53.
Zurück zum Zitat Goldstein DJ, Naftel DC, Holman WL, Bellumkonda L, Pamboukian SV, Pagani FD, Kirklin JK (2011) 7 Driveline Infections in LVADs: is it the pump or the patient? J Heart Lung Transpl 30(4, Supplement 1):S10 Goldstein DJ, Naftel DC, Holman WL, Bellumkonda L, Pamboukian SV, Pagani FD, Kirklin JK (2011) 7 Driveline Infections in LVADs: is it the pump or the patient? J Heart Lung Transpl 30(4, Supplement 1):S10
54.
Zurück zum Zitat Bomholt T, Moser C, Sander K, Boesgaard S, Kober L, Olsen PS, Hansen PB, Mortensen SA, Gustafsson F (2011) Driveline infections in patients supported with a HeartMate II: incidence, aetiology and outcome. Scand Cardiovasc J 45(5):273–278PubMedCrossRef Bomholt T, Moser C, Sander K, Boesgaard S, Kober L, Olsen PS, Hansen PB, Mortensen SA, Gustafsson F (2011) Driveline infections in patients supported with a HeartMate II: incidence, aetiology and outcome. Scand Cardiovasc J 45(5):273–278PubMedCrossRef
55.
Zurück zum Zitat Baumblatt JA, Pauwaa S, Treitman A, Gallagher C, Blair C, Pappas P, Tatooles AJ, Bhat G (2011) 6 are blood stream infections associated with increased risk of stroke in patients with a continuous flow left ventricular assist device? J Heart Lung Transplant 30(4, Supplement 1):S9–S10 Baumblatt JA, Pauwaa S, Treitman A, Gallagher C, Blair C, Pappas P, Tatooles AJ, Bhat G (2011) 6 are blood stream infections associated with increased risk of stroke in patients with a continuous flow left ventricular assist device? J Heart Lung Transplant 30(4, Supplement 1):S9–S10
56.
Zurück zum Zitat Hannan MM, Husain S, Mattner F, Danziger-Isakov L, Drew RJ, Corey GR, Schueler S, Holman WL, Lawler LP, Gordon SM, Mahon NG, Herre JM, Gould K, Montoya JG, Padera RF, Kormos RL, Conte JV, Mooney ML (2011) Working formulation for the standardization of definitions of infections in patients using ventricular assist devices. J Heart Lung Transpl 30(4):375–384CrossRef Hannan MM, Husain S, Mattner F, Danziger-Isakov L, Drew RJ, Corey GR, Schueler S, Holman WL, Lawler LP, Gordon SM, Mahon NG, Herre JM, Gould K, Montoya JG, Padera RF, Kormos RL, Conte JV, Mooney ML (2011) Working formulation for the standardization of definitions of infections in patients using ventricular assist devices. J Heart Lung Transpl 30(4):375–384CrossRef
57.
Zurück zum Zitat John R, Kamdar F, Liao K, Colvin-Adams M, Boyle A, Joyce L (2008) Improved survival and decreasing incidence of adverse events with the HeartMate II left ventricular assist device as bridge-to-transplant therapy. Ann Thorac Surg 86(4):1227–1234; discussion 1234–1225 John R, Kamdar F, Liao K, Colvin-Adams M, Boyle A, Joyce L (2008) Improved survival and decreasing incidence of adverse events with the HeartMate II left ventricular assist device as bridge-to-transplant therapy. Ann Thorac Surg 86(4):1227–1234; discussion 1234–1225
58.
Zurück zum Zitat Geisen U, Heilmann C, Beyersdorf F, Benk C, Berchtold-Herz M, Schlensak C, Budde U, Zieger B (2008) Non-surgical bleeding in patients with ventricular assist devices could be explained by acquired von Willebrand disease. Eur J Cardiothorac Surg 33(4):679–684PubMedCrossRef Geisen U, Heilmann C, Beyersdorf F, Benk C, Berchtold-Herz M, Schlensak C, Budde U, Zieger B (2008) Non-surgical bleeding in patients with ventricular assist devices could be explained by acquired von Willebrand disease. Eur J Cardiothorac Surg 33(4):679–684PubMedCrossRef
59.
Zurück zum Zitat Crow S, Chen D, Milano C, Thomas W, Joyce L, Piacentino V III, Sharma R, Wu J, Arepally G, Bowles D, Rogers J, Villamizar-Ortiz N (2010) Acquired von Willebrand syndrome in continuous-flow ventricular assist device recipients. Ann Thorac Surg 90(4):1263–1269; discussion 1269 Crow S, Chen D, Milano C, Thomas W, Joyce L, Piacentino V III, Sharma R, Wu J, Arepally G, Bowles D, Rogers J, Villamizar-Ortiz N (2010) Acquired von Willebrand syndrome in continuous-flow ventricular assist device recipients. Ann Thorac Surg 90(4):1263–1269; discussion 1269
60.
Zurück zum Zitat Stern DR, Kazam J, Edwards P, Maybaum S, Bello RA, D’Alessandro DA, Goldstein DJ (2010) Increased incidence of gastrointestinal bleeding following implantation of the HeartMate II LVAD. J Card Surg 25(3):352–356PubMedCrossRef Stern DR, Kazam J, Edwards P, Maybaum S, Bello RA, D’Alessandro DA, Goldstein DJ (2010) Increased incidence of gastrointestinal bleeding following implantation of the HeartMate II LVAD. J Card Surg 25(3):352–356PubMedCrossRef
61.
Zurück zum Zitat Uriel N, Pak S-W, Jorde UP, Jude B, Susen S, Vincentelli A, Ennezat P-V, Cappleman S, Naka Y, Mancini D (2010) Acquired von Willebrand syndrome after continuous-flow mechanical device support contributes to a high prevalence of bleeding during long-term support and at the time of transplantation. J Am Coll Cardiol 56(15):1207–1213PubMedCrossRef Uriel N, Pak S-W, Jorde UP, Jude B, Susen S, Vincentelli A, Ennezat P-V, Cappleman S, Naka Y, Mancini D (2010) Acquired von Willebrand syndrome after continuous-flow mechanical device support contributes to a high prevalence of bleeding during long-term support and at the time of transplantation. J Am Coll Cardiol 56(15):1207–1213PubMedCrossRef
62.
Zurück zum Zitat Crow S, John R, Boyle A, Shumway S, Liao K, Colvin-Adams M, Toninato C, Missov E, Pritzker M, Martin C, Garry D, Thomas W, Joyce L (2009) Gastrointestinal bleeding rates in recipients of nonpulsatile and pulsatile left ventricular assist devices. J Thorac Cardiovasc Surg 137(1):208–215PubMedCrossRef Crow S, John R, Boyle A, Shumway S, Liao K, Colvin-Adams M, Toninato C, Missov E, Pritzker M, Martin C, Garry D, Thomas W, Joyce L (2009) Gastrointestinal bleeding rates in recipients of nonpulsatile and pulsatile left ventricular assist devices. J Thorac Cardiovasc Surg 137(1):208–215PubMedCrossRef
63.
Zurück zum Zitat Strueber M, Malehsa D, Avsar M, Jahr N, Bara C, Haverich A (2011) 453 gastrointestinal bleeding after implantation of the HeartWare HVAD left ventricular assist device. J Heart Lung Transpl 30(4):S154CrossRef Strueber M, Malehsa D, Avsar M, Jahr N, Bara C, Haverich A (2011) 453 gastrointestinal bleeding after implantation of the HeartWare HVAD left ventricular assist device. J Heart Lung Transpl 30(4):S154CrossRef
64.
Zurück zum Zitat Meyer AL, Malehsa D, Bara C, Budde U, Slaughter MS, Haverich A, Strueber M (2010) Acquired von Willebrand syndrome in patients with an axial flow left ventricular assist device/clinical perspective. Circ Heart Fail 3(6):675–681PubMedCrossRef Meyer AL, Malehsa D, Bara C, Budde U, Slaughter MS, Haverich A, Strueber M (2010) Acquired von Willebrand syndrome in patients with an axial flow left ventricular assist device/clinical perspective. Circ Heart Fail 3(6):675–681PubMedCrossRef
65.
Zurück zum Zitat Azzouz M, Young SD, Meyer MM, Sun B, Firstenberg MS (2011) Gastrointestinal bleeding in patients with ventricular assist devices: a diagnostic and treatment algorithm. J Heart Lung Transpl 30(4):S154CrossRef Azzouz M, Young SD, Meyer MM, Sun B, Firstenberg MS (2011) Gastrointestinal bleeding in patients with ventricular assist devices: a diagnostic and treatment algorithm. J Heart Lung Transpl 30(4):S154CrossRef
66.
Zurück zum Zitat Demirozu ZT, Radovancevic R, Hochman LF, Gregoric ID, Letsou GV, Kar B, Bogaev RC, Frazier OH (2011) Arteriovenous malformation and gastrointestinal bleeding in patients with the HeartMate II left ventricular assist device. J Heart Lung Transpl 30(8):849–853 Demirozu ZT, Radovancevic R, Hochman LF, Gregoric ID, Letsou GV, Kar B, Bogaev RC, Frazier OH (2011) Arteriovenous malformation and gastrointestinal bleeding in patients with the HeartMate II left ventricular assist device. J Heart Lung Transpl 30(8):849–853
67.
Zurück zum Zitat Romano MA, Haft JW, Stulak JM, Aaronson KD, Pagani FD, Cowger J (2011) 9 hemolysis predicts poor outcome following LVAD implant. J Heart Lung Transpl 30(4, Supplement):S11 Romano MA, Haft JW, Stulak JM, Aaronson KD, Pagani FD, Cowger J (2011) 9 hemolysis predicts poor outcome following LVAD implant. J Heart Lung Transpl 30(4, Supplement):S11
68.
Zurück zum Zitat Slaughter MS, Pagani FD, Rogers JG, Miller LW, Sun B, Russell SD, Starling RC, Chen L, Boyle AJ, Chillcott S, Adamson RM, Blood MS, Camacho MT, Idrissi KA, Petty M, Sobieski M, Wright S, Myers TJ, Farrar DJ (2010) Clinical management of continuous-flow left ventricular assist devices in advanced heart failure. J Heart Lung Transpl 29(4 Suppl):S1–S39CrossRef Slaughter MS, Pagani FD, Rogers JG, Miller LW, Sun B, Russell SD, Starling RC, Chen L, Boyle AJ, Chillcott S, Adamson RM, Blood MS, Camacho MT, Idrissi KA, Petty M, Sobieski M, Wright S, Myers TJ, Farrar DJ (2010) Clinical management of continuous-flow left ventricular assist devices in advanced heart failure. J Heart Lung Transpl 29(4 Suppl):S1–S39CrossRef
69.
Zurück zum Zitat Kiernan MS, Pham DT, Denofrio D, Kapur NK (2011) Management of HeartWare left ventricular assist device thrombosis using intracavitary thrombolytics. J Thorac Cardiovasc Surg 142(3):712–714PubMedCrossRef Kiernan MS, Pham DT, Denofrio D, Kapur NK (2011) Management of HeartWare left ventricular assist device thrombosis using intracavitary thrombolytics. J Thorac Cardiovasc Surg 142(3):712–714PubMedCrossRef
70.
Zurück zum Zitat Paluszkiewicz L, Gursoy D, Spiliopoulos S, Dogan G, Daliakopoulos S, Tenderich M, Korfer R, Tenderich G (2011) HeartMate II ventricular assist device thrombosis-an echocardiographic approach to diagnosis: can Doppler evaluation of flow be useful? J Am Soc Echocardiogr 24(3):350 e351–e354 Paluszkiewicz L, Gursoy D, Spiliopoulos S, Dogan G, Daliakopoulos S, Tenderich M, Korfer R, Tenderich G (2011) HeartMate II ventricular assist device thrombosis-an echocardiographic approach to diagnosis: can Doppler evaluation of flow be useful? J Am Soc Echocardiogr 24(3):350 e351–e354
71.
Zurück zum Zitat Thomas MD, Wood C, Lovett M, Dembo L, O’Driscoll G (2008) Successful treatment of rotary pump thrombus with the glycoprotein IIb/IIIa inhibitor tirofiban. J Heart Lung Transpl 27(8):925–927CrossRef Thomas MD, Wood C, Lovett M, Dembo L, O’Driscoll G (2008) Successful treatment of rotary pump thrombus with the glycoprotein IIb/IIIa inhibitor tirofiban. J Heart Lung Transpl 27(8):925–927CrossRef
72.
Zurück zum Zitat Delgado Iii R, Frazier OH, Myers TJ, Gregoric ID, Robertson K, Shah NA, Kar B (2005) Direct thrombolytic therapy for intraventricular thrombosis in patients with the Jarvik 2000 left ventricular assist device. J Heart Lung Transpl 24(2):231–233CrossRef Delgado Iii R, Frazier OH, Myers TJ, Gregoric ID, Robertson K, Shah NA, Kar B (2005) Direct thrombolytic therapy for intraventricular thrombosis in patients with the Jarvik 2000 left ventricular assist device. J Heart Lung Transpl 24(2):231–233CrossRef
73.
Zurück zum Zitat Krabatsch T, Schweiger M, Stepanenko A, Kukucka M, Vierecke J, Lehmkuhl HB, Huebler M, Hennig E, Potapov E, Hetzer R (2011) Mechanical circulatory support-results, developments and trends. J Cardiovasc Transl Res 4(3):332–339PubMedCrossRef Krabatsch T, Schweiger M, Stepanenko A, Kukucka M, Vierecke J, Lehmkuhl HB, Huebler M, Hennig E, Potapov E, Hetzer R (2011) Mechanical circulatory support-results, developments and trends. J Cardiovasc Transl Res 4(3):332–339PubMedCrossRef
74.
Zurück zum Zitat Matthews JC, Koelling TM, Pagani FD, Aaronson KD (2008) The right ventricular failure risk score a pre-operative tool for assessing the risk of right ventricular failure in left ventricular assist device candidates. J Am Coll Cardiol 51(22):2163–2172PubMedCrossRef Matthews JC, Koelling TM, Pagani FD, Aaronson KD (2008) The right ventricular failure risk score a pre-operative tool for assessing the risk of right ventricular failure in left ventricular assist device candidates. J Am Coll Cardiol 51(22):2163–2172PubMedCrossRef
75.
Zurück zum Zitat Lee S, Kamdar F, Madlon-Kay R, Boyle A, Colvin-Adams M, Pritzker M, John R (2010) Effects of the HeartMate II continuous-flow left ventricular assist device on right ventricular function. J Heart Lung Transpl 29(2):209–215CrossRef Lee S, Kamdar F, Madlon-Kay R, Boyle A, Colvin-Adams M, Pritzker M, John R (2010) Effects of the HeartMate II continuous-flow left ventricular assist device on right ventricular function. J Heart Lung Transpl 29(2):209–215CrossRef
76.
Zurück zum Zitat Kukucka M, Stepanenko A, Potapov E, Krabatsch T, Redlin M, Mladenow A, Kuppe H, Hetzer R, Habazettl H (2011) Right-to-left ventricular end-diastolic diameter ratio and prediction of right ventricular failure with continuous-flow left ventricular assist devices. J Heart Lung Transpl 30(1):64–69CrossRef Kukucka M, Stepanenko A, Potapov E, Krabatsch T, Redlin M, Mladenow A, Kuppe H, Hetzer R, Habazettl H (2011) Right-to-left ventricular end-diastolic diameter ratio and prediction of right ventricular failure with continuous-flow left ventricular assist devices. J Heart Lung Transpl 30(1):64–69CrossRef
77.
Zurück zum Zitat Kormos RL, Teuteberg JJ, Pagani FD, Russell SD, John R, Miller LW, Massey T, Milano CA, Moazami N, Sundareswaran KS, Farrar DJ (2010) Right ventricular failure in patients with the HeartMate II continuous-flow left ventricular assist device: incidence, risk factors, and effect on outcomes. J Thorac Cardiovasc Surg 139(5):1316–1324PubMedCrossRef Kormos RL, Teuteberg JJ, Pagani FD, Russell SD, John R, Miller LW, Massey T, Milano CA, Moazami N, Sundareswaran KS, Farrar DJ (2010) Right ventricular failure in patients with the HeartMate II continuous-flow left ventricular assist device: incidence, risk factors, and effect on outcomes. J Thorac Cardiovasc Surg 139(5):1316–1324PubMedCrossRef
78.
Zurück zum Zitat Baumwol J, Macdonald PS, Keogh AM, Kotlyar E, Spratt P, Jansz P, Hayward CS (2011) Right heart failure and “failure to thrive” after left ventricular assist device: clinical predictors and outcomes. J Heart Lung Transpl 30(8):888–895 Baumwol J, Macdonald PS, Keogh AM, Kotlyar E, Spratt P, Jansz P, Hayward CS (2011) Right heart failure and “failure to thrive” after left ventricular assist device: clinical predictors and outcomes. J Heart Lung Transpl 30(8):888–895
79.
Zurück zum Zitat Iribarne A, Pak S-W, Russo MJ, Easterwood R, Yang J, Takayama H, Mancini D, Naka Y (2011) Incremental cost of right ventricular failure after left ventricular assist device placement. J Am Coll Cardiol 57(14, Supplement 1):E313 Iribarne A, Pak S-W, Russo MJ, Easterwood R, Yang J, Takayama H, Mancini D, Naka Y (2011) Incremental cost of right ventricular failure after left ventricular assist device placement. J Am Coll Cardiol 57(14, Supplement 1):E313
80.
Zurück zum Zitat Ochiai Y, McCarthy PM, Smedira NG, Banbury MK, Navia JL, Feng J, Hsu AP, Yeager ML, Buda T, Hoercher KJ, Howard MW, Takagaki M, Doi K, Fukamachi K (2002) Predictors of severe right ventricular failure after implantable left ventricular assist device insertion: analysis of 245 patients. Circulation 106(12 suppl 1):I-198–I-202 Ochiai Y, McCarthy PM, Smedira NG, Banbury MK, Navia JL, Feng J, Hsu AP, Yeager ML, Buda T, Hoercher KJ, Howard MW, Takagaki M, Doi K, Fukamachi K (2002) Predictors of severe right ventricular failure after implantable left ventricular assist device insertion: analysis of 245 patients. Circulation 106(12 suppl 1):I-198–I-202
81.
Zurück zum Zitat Puwanant S, Hamilton KK, Klodell CT, Hill JA, Schofield RS, Cleeton TS, Pauly DF, Aranda JM Jr (2008) Tricuspid annular motion as a predictor of severe right ventricular failure after left ventricular assist device implantation. J Heart Lung Transpl 27(10):1102–1107CrossRef Puwanant S, Hamilton KK, Klodell CT, Hill JA, Schofield RS, Cleeton TS, Pauly DF, Aranda JM Jr (2008) Tricuspid annular motion as a predictor of severe right ventricular failure after left ventricular assist device implantation. J Heart Lung Transpl 27(10):1102–1107CrossRef
82.
Zurück zum Zitat Fitzpatrick JR III, Frederick JR, Hiesinger W, Hsu VM, McCormick RC, Kozin ED, Laporte CM, O’Hara ML, Howell E, Dougherty D, Cohen JE, Southerland KW, Howard JL, Paulson EC, Acker MA, Morris RJ, Woo YJ (2009) Early planned institution of biventricular mechanical circulatory support results in improved outcomes compared with delayed conversion of a left ventricular assist device to a biventricular assist device. J Thorac Cardiovasc Surg 137(4):971–977PubMedCrossRef Fitzpatrick JR III, Frederick JR, Hiesinger W, Hsu VM, McCormick RC, Kozin ED, Laporte CM, O’Hara ML, Howell E, Dougherty D, Cohen JE, Southerland KW, Howard JL, Paulson EC, Acker MA, Morris RJ, Woo YJ (2009) Early planned institution of biventricular mechanical circulatory support results in improved outcomes compared with delayed conversion of a left ventricular assist device to a biventricular assist device. J Thorac Cardiovasc Surg 137(4):971–977PubMedCrossRef
83.
Zurück zum Zitat Frazier OH, Myers TJ, Gregoric I (2004) Biventricular assistance with the Jarvik FlowMaker: a case report. J Thorac Cardiovasc Surg 128(4):625–626PubMedCrossRef Frazier OH, Myers TJ, Gregoric I (2004) Biventricular assistance with the Jarvik FlowMaker: a case report. J Thorac Cardiovasc Surg 128(4):625–626PubMedCrossRef
84.
Zurück zum Zitat Stulak JM, Griffith KE, Nicklas JM, Pagani FD (2011) The use of the HeartWare HVAD for long-term right ventricular support after implantation of the HeartMate II device. J Thorac Cardiovasc Surg 142(3):e140–e142PubMedCrossRef Stulak JM, Griffith KE, Nicklas JM, Pagani FD (2011) The use of the HeartWare HVAD for long-term right ventricular support after implantation of the HeartMate II device. J Thorac Cardiovasc Surg 142(3):e140–e142PubMedCrossRef
85.
Zurück zum Zitat Strueber M, Meyer AL, Malehsa D, Haverich A (2010) Successful use of the HeartWare HVAD rotary blood pump for biventricular support. J Thorac Cardiovasc Surg 140(4):936–937PubMedCrossRef Strueber M, Meyer AL, Malehsa D, Haverich A (2010) Successful use of the HeartWare HVAD rotary blood pump for biventricular support. J Thorac Cardiovasc Surg 140(4):936–937PubMedCrossRef
86.
Zurück zum Zitat Saito S, Sakaguchi T, Miyagawa S, Yoshikawa Y, Yamauchi T, Ueno T, Kuratani T, Sawa Y (2011) Biventricular support using implantable continuous-flow ventricular assist devices. J Heart Lung Transpl 30(4):475–478CrossRef Saito S, Sakaguchi T, Miyagawa S, Yoshikawa Y, Yamauchi T, Ueno T, Kuratani T, Sawa Y (2011) Biventricular support using implantable continuous-flow ventricular assist devices. J Heart Lung Transpl 30(4):475–478CrossRef
87.
Zurück zum Zitat Stulak JM, Cowger J, Haft JW, Romano MA, Aaronson KD, Pagani FD (2011) 469 the natural history of tricuspid valve and right ventricular function following left ventricular assist device implantation. J Heart Lung Transpl 30(4, Supplement):S160 Stulak JM, Cowger J, Haft JW, Romano MA, Aaronson KD, Pagani FD (2011) 469 the natural history of tricuspid valve and right ventricular function following left ventricular assist device implantation. J Heart Lung Transpl 30(4, Supplement):S160
88.
Zurück zum Zitat Piacentino V III, Williams ML, Depp T, Garcia-Huerta K, Blue L, Lodge AJ, Mackensen GB, Swaminathan M, Rogers JG, Milano CA (2011) Impact of tricuspid valve regurgitation in patients treated with implantable left ventricular assist devices. Ann Thorac Surg 91(5):1342–1346; discussion 1346–1347 Piacentino V III, Williams ML, Depp T, Garcia-Huerta K, Blue L, Lodge AJ, Mackensen GB, Swaminathan M, Rogers JG, Milano CA (2011) Impact of tricuspid valve regurgitation in patients treated with implantable left ventricular assist devices. Ann Thorac Surg 91(5):1342–1346; discussion 1346–1347
89.
Zurück zum Zitat Saeed D, Kidambi T, Shalli S, Lapin B, Malaisrie SC, Lee R, Cotts WG, McGee EC Jr (2011) Tricuspid valve repair with left ventricular assist device implantation: is it warranted? J Heart Lung Transpl 30(5):530–535CrossRef Saeed D, Kidambi T, Shalli S, Lapin B, Malaisrie SC, Lee R, Cotts WG, McGee EC Jr (2011) Tricuspid valve repair with left ventricular assist device implantation: is it warranted? J Heart Lung Transpl 30(5):530–535CrossRef
90.
Zurück zum Zitat Piacentino V, Troupes C, Ganapathi A, Blue L, Mackensen GB, Swaminathan M, Felker GM, Stafford-Smith M, Lodge AJ, Rogers J, Milano C (2011) Clinical impact of concomitant tricuspid valve procedures during left ventricular assist device implantation. Paper presented at the Society of Thoracic Surgeons Annual Meeting, San Diego, CA Piacentino V, Troupes C, Ganapathi A, Blue L, Mackensen GB, Swaminathan M, Felker GM, Stafford-Smith M, Lodge AJ, Rogers J, Milano C (2011) Clinical impact of concomitant tricuspid valve procedures during left ventricular assist device implantation. Paper presented at the Society of Thoracic Surgeons Annual Meeting, San Diego, CA
91.
Zurück zum Zitat Park SJ, Liao KK, Segurola R, Madhu KP, Miller LW (2004) Management of aortic insufficiency in patients with left ventricular assist devices: a simple coaptation stitch method (Park’s stitch). J Heart Lung Transpl 127(1):264–266 Park SJ, Liao KK, Segurola R, Madhu KP, Miller LW (2004) Management of aortic insufficiency in patients with left ventricular assist devices: a simple coaptation stitch method (Park’s stitch). J Heart Lung Transpl 127(1):264–266
92.
Zurück zum Zitat Adamson RM, Dembitsky WP, Baradarian S, Chammas J, May-Newman K, Chillcott S, Stahovich M, McCalmont V, Ortiz K, Hoagland P, Jaski B (2011) Aortic valve closure associated with HeartMate left ventricular device support: technical considerations and long-term results. J Heart Lung Transpl 30(5):576–582CrossRef Adamson RM, Dembitsky WP, Baradarian S, Chammas J, May-Newman K, Chillcott S, Stahovich M, McCalmont V, Ortiz K, Hoagland P, Jaski B (2011) Aortic valve closure associated with HeartMate left ventricular device support: technical considerations and long-term results. J Heart Lung Transpl 30(5):576–582CrossRef
93.
Zurück zum Zitat Letsou GV, Connelly JH, Delgado RM III, Myers TJ, Gregoric ID, Smart FW, Frazier OH (2006) Is native aortic valve commissural fusion in patients with long-term left ventricular assist devices associated with clinically important aortic insufficiency? J Heart Lung Transpl 25(4):395–399CrossRef Letsou GV, Connelly JH, Delgado RM III, Myers TJ, Gregoric ID, Smart FW, Frazier OH (2006) Is native aortic valve commissural fusion in patients with long-term left ventricular assist devices associated with clinically important aortic insufficiency? J Heart Lung Transpl 25(4):395–399CrossRef
94.
Zurück zum Zitat Matthews JC, Aaronson KD, Jain R, Pagani FD, Kolias TJ (2009) 692: aortic insufficiency—trends over time in LVAD supported patients. J Heart Lung Transpl 28(2, Supplement):S306 Matthews JC, Aaronson KD, Jain R, Pagani FD, Kolias TJ (2009) 692: aortic insufficiency—trends over time in LVAD supported patients. J Heart Lung Transpl 28(2, Supplement):S306
95.
Zurück zum Zitat John R, Mantz K, Eckman P, Rose A, May-Newman K (2010) Aortic valve pathophysiology during left ventricular assist device support. J Heart Lung Transpl 29(12):1321–1329CrossRef John R, Mantz K, Eckman P, Rose A, May-Newman K (2010) Aortic valve pathophysiology during left ventricular assist device support. J Heart Lung Transpl 29(12):1321–1329CrossRef
96.
Zurück zum Zitat Pak S-W, Uriel N, Takayama H, Cappleman S, Song R, Colombo PC, Charles S, Mancini D, Gillam L, Naka Y, Jorde UP (2010) Prevalence of de novo aortic insufficiency during long-term support with left ventricular assist devices. J Heart Lung Transpl 29(10):1172–1176CrossRef Pak S-W, Uriel N, Takayama H, Cappleman S, Song R, Colombo PC, Charles S, Mancini D, Gillam L, Naka Y, Jorde UP (2010) Prevalence of de novo aortic insufficiency during long-term support with left ventricular assist devices. J Heart Lung Transpl 29(10):1172–1176CrossRef
97.
Zurück zum Zitat Cowger J, Pagani FD, Haft JW, Romano MA, Aaronson KD, Kolias TJ (2010) The development of aortic insufficiency in left ventricular assist device-supported patients/clinical perspective. Circ Heart Fail 3(6):668–674PubMedCrossRef Cowger J, Pagani FD, Haft JW, Romano MA, Aaronson KD, Kolias TJ (2010) The development of aortic insufficiency in left ventricular assist device-supported patients/clinical perspective. Circ Heart Fail 3(6):668–674PubMedCrossRef
98.
Zurück zum Zitat Mudd JO, Cuda JD, Halushka M, Soderlund KA, Conte JV, Russell SD (2008) Fusion of aortic valve commissures in patients supported by a continuous axial flow left ventricular assist device. J Heart Lung Transpl 27(12):1269–1274CrossRef Mudd JO, Cuda JD, Halushka M, Soderlund KA, Conte JV, Russell SD (2008) Fusion of aortic valve commissures in patients supported by a continuous axial flow left ventricular assist device. J Heart Lung Transpl 27(12):1269–1274CrossRef
99.
Zurück zum Zitat Toda K, Fujita T, Domae K, Shimahara Y, Kobayashi J, Nakatani T (2011) Late aortic insufficiency related to poor prognosis during left ventricular assist device support. Ann Thorac Surg 92(3):929–934PubMedCrossRef Toda K, Fujita T, Domae K, Shimahara Y, Kobayashi J, Nakatani T (2011) Late aortic insufficiency related to poor prognosis during left ventricular assist device support. Ann Thorac Surg 92(3):929–934PubMedCrossRef
100.
Zurück zum Zitat Bedi M, Kormos R, Winowich S, McNamara DM, Mathier MA, Murali S (2007) Ventricular arrhythmias during left ventricular assist device support. Am J Cardiol 99(8):1151–1153PubMedCrossRef Bedi M, Kormos R, Winowich S, McNamara DM, Mathier MA, Murali S (2007) Ventricular arrhythmias during left ventricular assist device support. Am J Cardiol 99(8):1151–1153PubMedCrossRef
101.
Zurück zum Zitat Andersen M, Videbaek R, Boesgaard S, Sander K, Hartsen PB, Gustafsson F (2009) Incidence of ventricular arrhythmias in patients on long-term support with a continuous-flow assist device (HeartMate II). J Heart Lung Transpl 28(7):733–735CrossRef Andersen M, Videbaek R, Boesgaard S, Sander K, Hartsen PB, Gustafsson F (2009) Incidence of ventricular arrhythmias in patients on long-term support with a continuous-flow assist device (HeartMate II). J Heart Lung Transpl 28(7):733–735CrossRef
102.
Zurück zum Zitat Milano CA, Simeone AA, Blue LJ, Rogers JG (2011) Presentation and management of left ventricular assist device inflow cannula malposition. J Heart Lung Transpl 30(7):838–840CrossRef Milano CA, Simeone AA, Blue LJ, Rogers JG (2011) Presentation and management of left ventricular assist device inflow cannula malposition. J Heart Lung Transpl 30(7):838–840CrossRef
103.
Zurück zum Zitat Ambardekar AV, Allen LA, Lindenfeld J, Lowery CM, Cannon AP, Cleveland JC Jr, Brieke A, Sauer WH (2010) Implantable cardioverter-defibrillator shocks in patients with a left ventricular assist device. J Heart Lung Transpl 29(7):771–776CrossRef Ambardekar AV, Allen LA, Lindenfeld J, Lowery CM, Cannon AP, Cleveland JC Jr, Brieke A, Sauer WH (2010) Implantable cardioverter-defibrillator shocks in patients with a left ventricular assist device. J Heart Lung Transpl 29(7):771–776CrossRef
104.
Zurück zum Zitat Miller LW (2011) Left ventricular assist devices are underutilized/response to miller. Circulation 123(14):1552–1558PubMedCrossRef Miller LW (2011) Left ventricular assist devices are underutilized/response to miller. Circulation 123(14):1552–1558PubMedCrossRef
105.
Zurück zum Zitat Russo MJ, Gelijns A, Aaronson K, Miller L, Hong K, Oz M, Ascheim D, Pagani F, Naka Y, Rose E, Moskowitz A (2010) Changing cost-effectiveness of left ventricular assist devices as destination therapy. J Am Coll Cardiol 55(10, Supplement 1):A18.E169–A118.E169 Russo MJ, Gelijns A, Aaronson K, Miller L, Hong K, Oz M, Ascheim D, Pagani F, Naka Y, Rose E, Moskowitz A (2010) Changing cost-effectiveness of left ventricular assist devices as destination therapy. J Am Coll Cardiol 55(10, Supplement 1):A18.E169–A118.E169
106.
Zurück zum Zitat Slaughter MS, Russo M, Bostic R, Rogers J (2010) Abstract 12375: cost effectiveness of continuous-flow left ventricular assist device as destination therapy in advanced heart failure. Circulation 122(21 Supplement):A12375 Slaughter MS, Russo M, Bostic R, Rogers J (2010) Abstract 12375: cost effectiveness of continuous-flow left ventricular assist device as destination therapy in advanced heart failure. Circulation 122(21 Supplement):A12375
Metadaten
Titel
Mechanical circulatory support: devices, outcomes and complications
verfasst von
Carmelo A. Milano
Alan A. Simeone
Publikationsdatum
01.01.2013
Verlag
Springer US
Erschienen in
Heart Failure Reviews / Ausgabe 1/2013
Print ISSN: 1382-4147
Elektronische ISSN: 1573-7322
DOI
https://doi.org/10.1007/s10741-012-9303-5

Weitere Artikel der Ausgabe 1/2013

Heart Failure Reviews 1/2013 Zur Ausgabe

Update Kardiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.