Agreement between diurnal variations of intraocular pressure by Tono-Pen and Goldmann applanation tonometer in patients on topical anti-glaucoma medication

verfasst von: Shikha Gupta, Gautam Sinha, Reetika Sharma, Bhagabat Nayak, Bharat Patil, Bibhuti Kashyap, Abdul Shameer, Tanuj Dada

Erschienen in: International Ophthalmology | Ausgabe 1/2016

Einloggen, um Zugang zu erhalten

Abstract

To estimate agreement in diurnal variations of intraocular pressure (IOP) by Tono-Pen (TP) and Goldmann applanation tonometer (GAT) in glaucoma patients on topical anti-glaucoma medication(s). IOP was measured at every 3 h from 7 a.m. to 10 a.m. in 50 eyes of glaucoma patients on topical medication(s). Diurnal fluctuation of IOP by each method was calculated as maximum–minimum IOP in a day. Central corneal thickness (CCT) was measured by ultrasonic pachymeter. There was good correlation between TP and GAT at all times during a day, minimum, and maximum IOPs during a day (Correlation coefficient, 0.706 at 7 a.m., 0.624 at 10 a.m., 0.682 at 1 p.m., 0.814 at 4 p.m., 0.652 at 7 p.m., 0.572 at 10 p.m., 0.668 minimum IOP, 0.689 maximum IOP). Mean IOPs by TP were always higher than GAT at all times during a day. Bland–Altman plots suggested a close relationship between the two sets of readings, and that this relationship was consistent at different times in a day, in maximum IOPs, minimum IOPs and also in fluctuation of IOPs. Linear regression analysis between the differences of diurnal fluctuation (diurnal fluctuation by GAT-diurnal fluctuation by TP) and CCT showed strong association (R 2 = 0.857, p < 0.001). The mean change in difference of diurnal fluctuation (GAT–TP) for a 10-micron increase in CCT was 0.69 mmHg. TP can be considered a reliable alternative to GAT in glaucoma patients for knowing the diurnal control of IOP; however these two methods should not be used interchangeably. Difference of diurnal fluctuation between two methods is dependent on CCT.
Literatur
1.
Zurück zum Zitat Comparison of glaucomatous progression between untreated patients with normal-tension glaucoma and patients with therapeutically reduced intraocular pressures. Collaborative Normal-Tension Glaucoma Study Group. Am J Ophthalmol 1998 Oct; 126(4):487-97 Comparison of glaucomatous progression between untreated patients with normal-tension glaucoma and patients with therapeutically reduced intraocular pressures. Collaborative Normal-Tension Glaucoma Study Group. Am J Ophthalmol 1998 Oct; 126(4):487-97
2.
Zurück zum Zitat Kass MA, Heuer DK, Higginbotham EJ, Johnson CA, Keltner JL, Miller JP, Parrish RK 2nd, Wilson MR, Gordon MO (2002) The Ocular hypertension treatment study: a randomized trial determines that topical ocular hypotensive medication delays or prevents the onset of primary open-angle glaucoma. Arch Ophthalmol 120(6):701–713PubMedCrossRef Kass MA, Heuer DK, Higginbotham EJ, Johnson CA, Keltner JL, Miller JP, Parrish RK 2nd, Wilson MR, Gordon MO (2002) The Ocular hypertension treatment study: a randomized trial determines that topical ocular hypotensive medication delays or prevents the onset of primary open-angle glaucoma. Arch Ophthalmol 120(6):701–713PubMedCrossRef
3.
Zurück zum Zitat Miglior S, Torri V, Zeyen T (2007) Intercurrent factors associated with the development of open angle glaucoma in the European glaucoma prevention study. Am J Ophthalmol 144(2):266–275PubMedCrossRef Miglior S, Torri V, Zeyen T (2007) Intercurrent factors associated with the development of open angle glaucoma in the European glaucoma prevention study. Am J Ophthalmol 144(2):266–275PubMedCrossRef
4.
Zurück zum Zitat Brubaker RF (1996) Delayed functional loss in glaucoma. LII Edward Jackson memorial lecture. Am J Ophthalmol 121:473–483PubMedCrossRef Brubaker RF (1996) Delayed functional loss in glaucoma. LII Edward Jackson memorial lecture. Am J Ophthalmol 121:473–483PubMedCrossRef
5.
Zurück zum Zitat Drance SM (1963) Diurnal variation of intraocular pressure in treated glaucoma. Arch Ophthalmol 70:302–311PubMedCrossRef Drance SM (1963) Diurnal variation of intraocular pressure in treated glaucoma. Arch Ophthalmol 70:302–311PubMedCrossRef
6.
7.
Zurück zum Zitat Whitacre MM, Stein R (1993) Sources of error with use of Goldmann-type tonometers. Surv Ophthalmol 38:1–30PubMedCrossRef Whitacre MM, Stein R (1993) Sources of error with use of Goldmann-type tonometers. Surv Ophthalmol 38:1–30PubMedCrossRef
8.
Zurück zum Zitat Mackay RS, Marg E (1959) Fast, automatic, electronic tonometers based on an exact theory. Acta Ophthalmol (Copenh) 13:495–507 Mackay RS, Marg E (1959) Fast, automatic, electronic tonometers based on an exact theory. Acta Ophthalmol (Copenh) 13:495–507
9.
Zurück zum Zitat Iester M, Mermoud A, Achache F, Roy S (2001) New Tonopen XL: comparison with the Goldmann tonometer. Eye (Lond) 13(Pt 1):52–58CrossRef Iester M, Mermoud A, Achache F, Roy S (2001) New Tonopen XL: comparison with the Goldmann tonometer. Eye (Lond) 13(Pt 1):52–58CrossRef
10.
Zurück zum Zitat Bhartiya S, Bali SJ, Sharma R, Chaturvedi N, Dada T (2011) Comparative evaluation of TonoPen AVIA, Goldmann applanation tonometry and non-contact tonometry. Int Ophthalmol 13(4):297–302CrossRef Bhartiya S, Bali SJ, Sharma R, Chaturvedi N, Dada T (2011) Comparative evaluation of TonoPen AVIA, Goldmann applanation tonometry and non-contact tonometry. Int Ophthalmol 13(4):297–302CrossRef
11.
Zurück zum Zitat Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1:307–310PubMedCrossRef Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1:307–310PubMedCrossRef
12.
Zurück zum Zitat Asrani S, Zeimer R, Wilensky J et al (2000) Large diurnal fluctuations in intraocular pressure are an independent risk factor in patients with glaucoma. J Glaucoma 9(2):134–142PubMedCrossRef Asrani S, Zeimer R, Wilensky J et al (2000) Large diurnal fluctuations in intraocular pressure are an independent risk factor in patients with glaucoma. J Glaucoma 9(2):134–142PubMedCrossRef
13.
Zurück zum Zitat Nouri-Mahdavi K, Hoffman D, Coleman AL et al (2004) Advanced glaucoma intervention study. Predictive factors for glaucomatous visual field progression in the advanced glaucoma intervention study. Ophthalmology 111:1627–1635PubMedCrossRef Nouri-Mahdavi K, Hoffman D, Coleman AL et al (2004) Advanced glaucoma intervention study. Predictive factors for glaucomatous visual field progression in the advanced glaucoma intervention study. Ophthalmology 111:1627–1635PubMedCrossRef
14.
Zurück zum Zitat Yoshita T, Kobayashi A, Takahashi M et al (2004) Reliability of intraocular pressure by Tono-Pen XL over amniotic membrane patch in human. J Glaucoma 13(5):413–416PubMedCrossRef Yoshita T, Kobayashi A, Takahashi M et al (2004) Reliability of intraocular pressure by Tono-Pen XL over amniotic membrane patch in human. J Glaucoma 13(5):413–416PubMedCrossRef
16.
Zurück zum Zitat Horowitz GS, Byles J, Lee J et al (2004) Comparison of the Tono-Pen and Goldmann tonometer for measuring intraocular pressure in patients with glaucoma. Clin Exp Ophthalmol 32(6):584–589CrossRef Horowitz GS, Byles J, Lee J et al (2004) Comparison of the Tono-Pen and Goldmann tonometer for measuring intraocular pressure in patients with glaucoma. Clin Exp Ophthalmol 32(6):584–589CrossRef
17.
Zurück zum Zitat Bandyopadhyay M, Raychaudhuri A, Lahiri SK, Schwartz EC, Myatt M, Johnson GJ (2002) Comparison of Goldmann applanation tonometry with the Tonopen for measuring intraocular pressure in a population-based glaucoma survey in rural West Bengal. Ophthalmic Epidemiol 9(3):215–224PubMedCrossRef Bandyopadhyay M, Raychaudhuri A, Lahiri SK, Schwartz EC, Myatt M, Johnson GJ (2002) Comparison of Goldmann applanation tonometry with the Tonopen for measuring intraocular pressure in a population-based glaucoma survey in rural West Bengal. Ophthalmic Epidemiol 9(3):215–224PubMedCrossRef
18.
Zurück zum Zitat Viestenz A, Lausen B, Jünemann AM, Mardin CY (2002) Comparison of precision of the TonoPenXL with the Goldmann and Draeger applanation tonometer in a sitting and recumbent position of the patients—a clinical study on 251 eyes. Klin Monbl Augenheilkd 219(11):785–790PubMedCrossRef Viestenz A, Lausen B, Jünemann AM, Mardin CY (2002) Comparison of precision of the TonoPenXL with the Goldmann and Draeger applanation tonometer in a sitting and recumbent position of the patients—a clinical study on 251 eyes. Klin Monbl Augenheilkd 219(11):785–790PubMedCrossRef
19.
Zurück zum Zitat Rao VJ, Gnanaraj L, Mitchell KW, Figueiredo FC (2001) Clinical comparison of ocular blood flow tonometer, Tonopen, and Goldmann applanation tonometer for measuring intraocular pressure in postkeratoplasty eyes. Cornea 20(8):834–838PubMedCrossRef Rao VJ, Gnanaraj L, Mitchell KW, Figueiredo FC (2001) Clinical comparison of ocular blood flow tonometer, Tonopen, and Goldmann applanation tonometer for measuring intraocular pressure in postkeratoplasty eyes. Cornea 20(8):834–838PubMedCrossRef
20.
Zurück zum Zitat Li J, Herndon LW, Asrani SG, Stinnett S, Allingham RR (2004) Clinical comparison of the proview eye pressure monitor with the Goldmann applanation tonometer and the Tonopen. Arch Ophthalmol 122(8):1117–1121PubMedCrossRef Li J, Herndon LW, Asrani SG, Stinnett S, Allingham RR (2004) Clinical comparison of the proview eye pressure monitor with the Goldmann applanation tonometer and the Tonopen. Arch Ophthalmol 122(8):1117–1121PubMedCrossRef
21.
Zurück zum Zitat Bafa M, Lambrinakis I, Dayan M, Birch M (2001) Clinical comparison of the measurement of the IOP with the ocular blood flow tonometer, the Tonopen XL and the Goldmann applanation tonometer. Acta Ophthalmol Scand 79(1):15–18PubMedCrossRef Bafa M, Lambrinakis I, Dayan M, Birch M (2001) Clinical comparison of the measurement of the IOP with the ocular blood flow tonometer, the Tonopen XL and the Goldmann applanation tonometer. Acta Ophthalmol Scand 79(1):15–18PubMedCrossRef
22.
Zurück zum Zitat Mollan SP, Wolffsohn JS, Nessim M et al (2008) Accuracy of Goldmann, ocular response analyser, Pascal and TonoPen XL tonometry in keratoconic and normal eyes. Br J Ophtalmol 92(12):1661–1665CrossRef Mollan SP, Wolffsohn JS, Nessim M et al (2008) Accuracy of Goldmann, ocular response analyser, Pascal and TonoPen XL tonometry in keratoconic and normal eyes. Br J Ophtalmol 92(12):1661–1665CrossRef
23.
Zurück zum Zitat Salvetat ML, Zeppieri M, Tosoni C, Brusini P (2007) Comparisons between Pascal dynamic contour tonometry, the TonoPen, and Goldmann applanation tonometry in patients with glaucoma. Acta Ophthalmol Scand 85(3):272–279PubMedCrossRef Salvetat ML, Zeppieri M, Tosoni C, Brusini P (2007) Comparisons between Pascal dynamic contour tonometry, the TonoPen, and Goldmann applanation tonometry in patients with glaucoma. Acta Ophthalmol Scand 85(3):272–279PubMedCrossRef
24.
Zurück zum Zitat Hsu S-Y, Sheu M-M, Hsu A-H, Wu K-Y, Yeh J-I, Tien J-N, Tsai R-K (2009) Comparisons of intraocular pressure measurements: Goldmann applanation tonometry, noncontact tonometry, Tono-Pen tonometry, and dynamic contour tonometry. Eye (Lond) 23(7):1582–1588CrossRef Hsu S-Y, Sheu M-M, Hsu A-H, Wu K-Y, Yeh J-I, Tien J-N, Tsai R-K (2009) Comparisons of intraocular pressure measurements: Goldmann applanation tonometry, noncontact tonometry, Tono-Pen tonometry, and dynamic contour tonometry. Eye (Lond) 23(7):1582–1588CrossRef
25.
Zurück zum Zitat Mosaed S, Chamberlain WD, Liu JH, Medeiros FA (2008) Weinreb RN association of central corneal thickness and 24-hour intraocular pressure fluctuation. J Glaucoma 17(2):85–88PubMedCrossRef Mosaed S, Chamberlain WD, Liu JH, Medeiros FA (2008) Weinreb RN association of central corneal thickness and 24-hour intraocular pressure fluctuation. J Glaucoma 17(2):85–88PubMedCrossRef
26.
Zurück zum Zitat Fogagnolo P, Capizzi F, Orzalesi N, Figus M, Ferreras A, Rossetti LJ (2010) Can mean central corneal thickness and its 24-hour fluctuation influence fluctuation of intraocular pressure? J Glaucoma 19:418–423PubMedCrossRef Fogagnolo P, Capizzi F, Orzalesi N, Figus M, Ferreras A, Rossetti LJ (2010) Can mean central corneal thickness and its 24-hour fluctuation influence fluctuation of intraocular pressure? J Glaucoma 19:418–423PubMedCrossRef
Metadaten
Titel
Agreement between diurnal variations of intraocular pressure by Tono-Pen and Goldmann applanation tonometer in patients on topical anti-glaucoma medication
verfasst von
Shikha Gupta
Gautam Sinha
Reetika Sharma
Bhagabat Nayak
Bharat Patil
Bibhuti Kashyap
Abdul Shameer
Tanuj Dada
Publikationsdatum
01.02.2016
Verlag
Springer Netherlands
Erschienen in
International Ophthalmology / Ausgabe 1/2016
Print ISSN: 0165-5701
Elektronische ISSN: 1573-2630
DOI
https://doi.org/10.1007/s10792-015-0067-5

Weitere Artikel der Ausgabe 1/2016

International Ophthalmology 1/2016 Zur Ausgabe

Neu im Fachgebiet Augenheilkunde

Open Access 22.02.2024 | Peeling | Originalien

Selektive Vitalfarbstoffe in der Makulachirurgie

Erhöhen sie die Wahrscheinlichkeit der intraoperativen ILM-Identifizierung auch bei einem erfahrenen Operateur?

Update Augenheilkunde

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.