01.06.2010

A protocol for patients with cardiovascular implantable devices undergoing magnetic resonance imaging (MRI): should defibrillation threshold testing be performed post-(MRI)

verfasst von: Peter Thomas Burke, Hamid Ghanbari, Patrick B. Alexander, Michael K. Shaw, Marcos Daccarett, Christian Machado

Erschienen in: Journal of Interventional Cardiac Electrophysiology | Ausgabe 1/2010

Einloggen, um Zugang zu erhalten

Abstract

Purpose

Magnetic resonance imaging (MRI) in patients with Cardiovascular Implantable Electronic Devices (CIED) has not been approved by the Food and Drug Administration. Recent data suggests MRI as a relative rather than absolute contraindication in CIED patients. Recently, the American Heart Association has recommended defibrillation threshold testing (DFTT) in implantable cardioverter defibrillator (ICD) patients undergoing MRI. We evaluated the feasibility and safety of a protocol for MRI in CIED patients, incorporating the new recommendations on DFTT.

Methods

Consecutive patients with CIED undergoing MRI were included. The protocol consisted of continuous monitoring during imaging, device interrogation pre- and post-MRI, reprogramming of the pacemaker to an asynchronous mode in pacemaker-dependent (PMD) patients and a non-tracking/sensing mode for non-PMD patients. All tachyarrhythmia therapies were disabled. Devices were interrogated for lead impedance, battery life, pacing, and sensing thresholds. All patients with ICD underwent DFTT/defibrillator safety margin testing (DSMT) post-MRI.

Results

A total of 92 MRI’s at 1.5 Tesla were performed in 38 patients. A total of 13 PMD patients, ten ICD patients, four cardiac resynchronization therapy with defibrillator (CRT-D) patients, and 11 non-PMD patients were scanned from four major manufacturers. No device circuitry damage, programming alterations, inappropriate shocks, failure to pace, or changes in sensing, pacing, or defibrillator thresholds were found on single or multiple MRI sessions.

Conclusions

Our protocol for MRI in CIED patients appears safe, feasible, and reproducible. This is irrespective of the type of CIED, pacemaker dependancy or multiple 24-h scanning sessions. Our protocol addresses early detection of potential complications and establishes a response system for potential device-related complications. Our observation suggests that routine DFTT/DSMT post-MRI may not be necessary.
Literatur
1.
Zurück zum Zitat Luechinger, R., Zeijlemaker, V. A., et al. (2005). In vivo heating of pacemaker leads during magnetic resonance imaging. European Heart Journal, 26, 376–383.CrossRefPubMed Luechinger, R., Zeijlemaker, V. A., et al. (2005). In vivo heating of pacemaker leads during magnetic resonance imaging. European Heart Journal, 26, 376–383.CrossRefPubMed
2.
Zurück zum Zitat Pinski, S. L., Trohman, R. G., et al. (2002). Interference in implantable cardiac devices, part II (review). Pacing and Clinical Electrophysiology, 25(10), 1496–1509.CrossRefPubMed Pinski, S. L., Trohman, R. G., et al. (2002). Interference in implantable cardiac devices, part II (review). Pacing and Clinical Electrophysiology, 25(10), 1496–1509.CrossRefPubMed
3.
Zurück zum Zitat Gimbel, R., Bailey, S., et al. (2005). Strategies for the safe magnetic resonance imaging of pacemaker-dependant patients. Pacing and Clinical Electrophysiology, 28, 1041–1046.CrossRefPubMed Gimbel, R., Bailey, S., et al. (2005). Strategies for the safe magnetic resonance imaging of pacemaker-dependant patients. Pacing and Clinical Electrophysiology, 28, 1041–1046.CrossRefPubMed
4.
Zurück zum Zitat Martin, E. T., Coman, J. A., et al. (2004). Magnetic resonance imaging and cardiac pacemaker safety at 1.5 Tesla. Journal of the American College of Cardiology, 43, 1315–1324.CrossRefPubMed Martin, E. T., Coman, J. A., et al. (2004). Magnetic resonance imaging and cardiac pacemaker safety at 1.5 Tesla. Journal of the American College of Cardiology, 43, 1315–1324.CrossRefPubMed
5.
Zurück zum Zitat Roguin, A., Zviman, M., et al. (2004). Modern pacemaker and implantable cardioverter/defibrillator systems can be magnetic imaging safe. in vitro and in vivo assesment of safety and function at 1.5 T. Circulation, 110, 475–482.CrossRefPubMed Roguin, A., Zviman, M., et al. (2004). Modern pacemaker and implantable cardioverter/defibrillator systems can be magnetic imaging safe. in vitro and in vivo assesment of safety and function at 1.5 T. Circulation, 110, 475–482.CrossRefPubMed
6.
Zurück zum Zitat Derek, V. E. (2005). Is antitachycardia pacing a safe and efficacious alternative to shocks for fast ventricular tachyarrhythmia treatment? Nature Clinical Practice Cardiovascular Medicine, 2, 68–69.CrossRef Derek, V. E. (2005). Is antitachycardia pacing a safe and efficacious alternative to shocks for fast ventricular tachyarrhythmia treatment? Nature Clinical Practice Cardiovascular Medicine, 2, 68–69.CrossRef
7.
Zurück zum Zitat Martin, E. T. (2005). Can cardiac pacemakers and magnetic resonance imaging systems co-exist? European Heart Journal, 26, 325–327.CrossRefPubMed Martin, E. T. (2005). Can cardiac pacemakers and magnetic resonance imaging systems co-exist? European Heart Journal, 26, 325–327.CrossRefPubMed
8.
Zurück zum Zitat Shellock, F., Fieno, D., et al. (2006). Cardiac pacemeaker: in vitro assesment at 1.5 T. American Heart Journal, 151, 436–443.CrossRefPubMed Shellock, F., Fieno, D., et al. (2006). Cardiac pacemeaker: in vitro assesment at 1.5 T. American Heart Journal, 151, 436–443.CrossRefPubMed
9.
Zurück zum Zitat Gimbel, J. R. (2008). Magnetic resonance imaging of implantable cardiac rhythm Devicesat 3.0 Tesla. Pacing and Clinical Electrophysiology, 31, 795–801.CrossRefPubMed Gimbel, J. R. (2008). Magnetic resonance imaging of implantable cardiac rhythm Devicesat 3.0 Tesla. Pacing and Clinical Electrophysiology, 31, 795–801.CrossRefPubMed
10.
Zurück zum Zitat Sierra, M., & Machado, C. (2008). Magnetic resonance imaging in patients with implantable cardiac devices. Review in Cardiovascular Medicine, 9(4), 232–238. Sierra, M., & Machado, C. (2008). Magnetic resonance imaging in patients with implantable cardiac devices. Review in Cardiovascular Medicine, 9(4), 232–238.
11.
Zurück zum Zitat Levine, G., Gomes, A., et al. (2007). Safety of magnetic resonance imaging in patients with cardiovascular devices. Circulation, 116, 2878–2891.CrossRefPubMed Levine, G., Gomes, A., et al. (2007). Safety of magnetic resonance imaging in patients with cardiovascular devices. Circulation, 116, 2878–2891.CrossRefPubMed
12.
Zurück zum Zitat Gimbel, J. R., Kanal, E., et al. (2005). Outcomes of Magnetic Resonance Imaging (MRI) in selected patients with Implantable Cardioverter Defibrillators (ICD). Pacing and Clinical Electrophysiology, 28, 270–273.CrossRefPubMed Gimbel, J. R., Kanal, E., et al. (2005). Outcomes of Magnetic Resonance Imaging (MRI) in selected patients with Implantable Cardioverter Defibrillators (ICD). Pacing and Clinical Electrophysiology, 28, 270–273.CrossRefPubMed
13.
Zurück zum Zitat Roguin, A., Schwitter, J., Vahlhaus, C., Lombardi, M., Priori, S., & Sommer, T. (2008). ESC position paper—magnetic resonance imaging in individuals with pacemakers or implantable cardioverter defibrillator systems. Europace, 10, 336–346.CrossRefPubMed Roguin, A., Schwitter, J., Vahlhaus, C., Lombardi, M., Priori, S., & Sommer, T. (2008). ESC position paper—magnetic resonance imaging in individuals with pacemakers or implantable cardioverter defibrillator systems. Europace, 10, 336–346.CrossRefPubMed
14.
Zurück zum Zitat Gold, M., Breiter, D., & Hahn, S. (2003). Safety of a single successful conversion of ventricular fibrillation before the implantation of cardioverter-defibrillators. Pacing and Clinical Electrophysiology, 26, 483–486.CrossRefPubMed Gold, M., Breiter, D., & Hahn, S. (2003). Safety of a single successful conversion of ventricular fibrillation before the implantation of cardioverter-defibrillators. Pacing and Clinical Electrophysiology, 26, 483–486.CrossRefPubMed
15.
Zurück zum Zitat Higgins, S., Mann, D., Calkins, H., et al. (2005). One conversion of ventricular fibrillation is adequate for implantable cardioverter-debibrillator. An analysis from the low energy safety study (LESS). Heart Rhythm, 2, 117–122.CrossRefPubMed Higgins, S., Mann, D., Calkins, H., et al. (2005). One conversion of ventricular fibrillation is adequate for implantable cardioverter-debibrillator. An analysis from the low energy safety study (LESS). Heart Rhythm, 2, 117–122.CrossRefPubMed
16.
Zurück zum Zitat Gimbel, J. R., & Kanal, E. (2004). Can patients with implantable pacemakers safely undergo magnetic resonance imaging? Journal of the American College of Cardiology, 43, 1325–1327.CrossRefPubMed Gimbel, J. R., & Kanal, E. (2004). Can patients with implantable pacemakers safely undergo magnetic resonance imaging? Journal of the American College of Cardiology, 43, 1325–1327.CrossRefPubMed
17.
Zurück zum Zitat Rozner, M. A., Burton, A. W., & Kumar, A. (2005). Pacemaker complication during magnetic resonance imaging. Journal of the American College of Cardiology, 45, 161–162.CrossRefPubMed Rozner, M. A., Burton, A. W., & Kumar, A. (2005). Pacemaker complication during magnetic resonance imaging. Journal of the American College of Cardiology, 45, 161–162.CrossRefPubMed
18.
Zurück zum Zitat Shellock, F. G., O'Neil, M., et al. (1999). Cardiac pacemeakers and implantable cardioverter defibrillators are unaffected by operation of extremity MR imaging system. American Journal of Roentgenology, 172, 165–170.PubMed Shellock, F. G., O'Neil, M., et al. (1999). Cardiac pacemeakers and implantable cardioverter defibrillators are unaffected by operation of extremity MR imaging system. American Journal of Roentgenology, 172, 165–170.PubMed
19.
Zurück zum Zitat Naehle, C. P., Zeijlemaker, V., Thomas, D., Meyer, C., Strach, K., Fimmers, R., et al. (2009). Evaluation of cumulative effects of MR imaging on pacemaker systems at 1.5 Tesla. Pacing and Clinical Electrophysiology, 32(12), 1526–1535.CrossRefPubMed Naehle, C. P., Zeijlemaker, V., Thomas, D., Meyer, C., Strach, K., Fimmers, R., et al. (2009). Evaluation of cumulative effects of MR imaging on pacemaker systems at 1.5 Tesla. Pacing and Clinical Electrophysiology, 32(12), 1526–1535.CrossRefPubMed
20.
Zurück zum Zitat Naehle, C. P., Meyer, C., Thomas, D., Remerie, S., Krautmacher, C., Litt, H., et al. (2008). Safety of brain 3-T MR imaging with transmit-receive head coil in patients with cardiac pacemakers: pilot prospective study with 51 examinations. Radiology, 249(3), 991–1001.CrossRefPubMed Naehle, C. P., Meyer, C., Thomas, D., Remerie, S., Krautmacher, C., Litt, H., et al. (2008). Safety of brain 3-T MR imaging with transmit-receive head coil in patients with cardiac pacemakers: pilot prospective study with 51 examinations. Radiology, 249(3), 991–1001.CrossRefPubMed
21.
Zurück zum Zitat Naehle, C. P., Strach, K., Thomas, D., Meyer, C., Linhart, M., Bitaraf, S., et al. (2009). Magnetic resonance imaging at 1.5 Tesla in patients with implantable cardioverters/defibrillators. Journal of the American College of Cardiology, 54(6), 549–555.CrossRefPubMed Naehle, C. P., Strach, K., Thomas, D., Meyer, C., Linhart, M., Bitaraf, S., et al. (2009). Magnetic resonance imaging at 1.5 Tesla in patients with implantable cardioverters/defibrillators. Journal of the American College of Cardiology, 54(6), 549–555.CrossRefPubMed
22.
Zurück zum Zitat Hayes, D. L., Holmes, D. R., Jr., et al. (1987). Effect of 1.5 Tesla nuclear magnetic resonance imaging scanner on implanted permanent pacemakers. Journal of the American College of Cardiology, 10(4), 782–786.PubMedCrossRef Hayes, D. L., Holmes, D. R., Jr., et al. (1987). Effect of 1.5 Tesla nuclear magnetic resonance imaging scanner on implanted permanent pacemakers. Journal of the American College of Cardiology, 10(4), 782–786.PubMedCrossRef
23.
Zurück zum Zitat Holmes, D. R., Jr., Hayes, D. L., et al. (1986). The effects of magnetic resonance imaging on implantable pulse generators. Pacing and Clinical Electrophysiology, 9(3), 360–370.CrossRefPubMed Holmes, D. R., Jr., Hayes, D. L., et al. (1986). The effects of magnetic resonance imaging on implantable pulse generators. Pacing and Clinical Electrophysiology, 9(3), 360–370.CrossRefPubMed
24.
Zurück zum Zitat Luechinger, R., Duru, F., et al. (2001). Force and torque effects of a 1.5-Tesla MRI scanner on cardiac pacemakers and ICDs. Pacing and Clinical Electrophysiology, 24(2), 199–205.CrossRefPubMed Luechinger, R., Duru, F., et al. (2001). Force and torque effects of a 1.5-Tesla MRI scanner on cardiac pacemakers and ICDs. Pacing and Clinical Electrophysiology, 24(2), 199–205.CrossRefPubMed
25.
Zurück zum Zitat Gimbel, J. R. (2008). The AHA scientific statement of MRI in patients with devices: neat, but incomplete. Unwise and unsupported. Pacing and Clinical Electrophysiology, 31(6), 649–651.CrossRefPubMed Gimbel, J. R. (2008). The AHA scientific statement of MRI in patients with devices: neat, but incomplete. Unwise and unsupported. Pacing and Clinical Electrophysiology, 31(6), 649–651.CrossRefPubMed
26.
Zurück zum Zitat Coman, J. A., Martin, E. T., et al. (2004). Implantable cardiac defibrillator interactions with magnetic resonance imaging at 1.5 Tesla. Journal of the American College of Cardiology, 43, 138A.CrossRef Coman, J. A., Martin, E. T., et al. (2004). Implantable cardiac defibrillator interactions with magnetic resonance imaging at 1.5 Tesla. Journal of the American College of Cardiology, 43, 138A.CrossRef
27.
Zurück zum Zitat Sandler, D. A., Coman, J. A., et al. (2006). Magnetic resonance imaging of patients with pacemakers and implantable cardioverter-defibrillators in a 1.5-Tesla magnet: A five-year experience. Heart Rhythm Society, 27th Annual Scientific Sessions, Boston, MA: May 17–20, 2006 Sandler, D. A., Coman, J. A., et al. (2006). Magnetic resonance imaging of patients with pacemakers and implantable cardioverter-defibrillators in a 1.5-Tesla magnet: A five-year experience. Heart Rhythm Society, 27th Annual Scientific Sessions, Boston, MA: May 17–20, 2006
28.
Zurück zum Zitat Fiek, M., Remp, T., Reithmann, C., & Steinbeck, G. (2004). Complete loss of ICD programmability after magnetic resonance imaging. Pacing and Clinical Electrophysiology, 27, 1002–1004.CrossRefPubMed Fiek, M., Remp, T., Reithmann, C., & Steinbeck, G. (2004). Complete loss of ICD programmability after magnetic resonance imaging. Pacing and Clinical Electrophysiology, 27, 1002–1004.CrossRefPubMed
29.
Zurück zum Zitat Glikson, M., Luria, D., et al. (2000). Are routine arrhythmia inductions necessary in patients with pectoral implantable cardioverter defibrillators? Journal of Cardiovascular Electrophysiology, 2, 127–135.CrossRef Glikson, M., Luria, D., et al. (2000). Are routine arrhythmia inductions necessary in patients with pectoral implantable cardioverter defibrillators? Journal of Cardiovascular Electrophysiology, 2, 127–135.CrossRef
30.
Zurück zum Zitat Nazarian, S., Roguin, A., Zviman, M. M., Lardo, A. C., Dickfeld, T. L., Calkins, H., et al. (2006). Clinical utility and safety of a protocol for noncardiac and cardiac magnetic resonance imaging of patients with permanent pacemakers and implantable-cardioverter defibrillators at 1.5 Tesla. Circulation, 114, 1277–1284.CrossRefPubMed Nazarian, S., Roguin, A., Zviman, M. M., Lardo, A. C., Dickfeld, T. L., Calkins, H., et al. (2006). Clinical utility and safety of a protocol for noncardiac and cardiac magnetic resonance imaging of patients with permanent pacemakers and implantable-cardioverter defibrillators at 1.5 Tesla. Circulation, 114, 1277–1284.CrossRefPubMed
31.
Zurück zum Zitat Birnie, D., Tung, S., Simpson, C., Crystal, E., Exner, D., Ayala Paredes, F. A., et al. (2008). Complications associated with defibrillation threshold testing: the Canadian experience. Heart Rhythm, 5(3), 387–390.CrossRefPubMed Birnie, D., Tung, S., Simpson, C., Crystal, E., Exner, D., Ayala Paredes, F. A., et al. (2008). Complications associated with defibrillation threshold testing: the Canadian experience. Heart Rhythm, 5(3), 387–390.CrossRefPubMed
Metadaten
Titel
A protocol for patients with cardiovascular implantable devices undergoing magnetic resonance imaging (MRI): should defibrillation threshold testing be performed post-(MRI)
verfasst von
Peter Thomas Burke
Hamid Ghanbari
Patrick B. Alexander
Michael K. Shaw
Marcos Daccarett
Christian Machado
Publikationsdatum
01.06.2010
Verlag
Springer US
Erschienen in
Journal of Interventional Cardiac Electrophysiology / Ausgabe 1/2010
Print ISSN: 1383-875X
Elektronische ISSN: 1572-8595
DOI
https://doi.org/10.1007/s10840-009-9463-y

Weitere Artikel der Ausgabe 1/2010

Journal of Interventional Cardiac Electrophysiology 1/2010 Zur Ausgabe

Update Kardiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.