Adverse events following implantable cardioverter defibrillator implantation: a systematic review

verfasst von: Rebecca Persson, Amy Earley, Ann C. Garlitski, Ethan M. Balk, Katrin Uhlig

Erschienen in: Journal of Interventional Cardiac Electrophysiology | Ausgabe 2/2014

Einloggen, um Zugang zu erhalten

Abstract

Purpose

Implantable cardioverter defibrillators (ICDs) prevent sudden cardiac death, but patients need to be counseled about potential harms. We summarized the evidence on adverse events from ICDs with a focus on ICD use for primary prevention of sudden cardiac death.

Methods

We searched MEDLINE and Cochrane Central Register of Controlled Trials from 2002 through 2012 for reports of adverse events from ICDs implanted for primary prevention or mixed indications (primary and secondary prevention). Studies had to have ≥500 patients and specify patient numerators and denominators.

Results

Data from 35 independent cohorts reported in 53 articles were included. Reports from one registry provided high quality evidence on adverse events during hospitalization for ICD implantation. Adverse events ranged from 2.8 to 3.6 %. Serious adverse events ranged from 1.2 to 1.4 %. The most frequent serious adverse events were pneumothorax (0.4–0.5 %) and cardiac arrest (0.3 %). The quality of the evidence for long-term adverse events was low. Frequency of adverse events post-hospitalization was variable, as was follow-up: device-related complications <0.1–6.4 % (2–49 months), lead-related complications <0.1–3.9 % (1.5–40 months), infection 0.2–3.7 % (1.5–49 months), and thrombosis 0.2–2.9 % (1.5–49 months). Evidence for inappropriate shock was of moderate quality with 3–21 % of patients experiencing at least one inappropriate shock during 1 to 5 years of follow-up.

Limitations

The limitation of the evidence reviewed in this study is low quality evidence for adverse events post-hospitalization. Evidence is predominantly from mixed primary and secondary prevention populations.

Conclusions

In-hospital adverse events after ICD implantation are infrequent. The estimates for long-term adverse events are uncertain. Up to one-fifth of patients receive inappropriate shocks.
Anhänge
Nur mit Berechtigung zugänglich
Literatur
1.
Zurück zum Zitat Earley, A., Persson, R., Garlitski, A., Balk, E., & Uhlig, K. (2014). Effectiveness of implantable cardioverter defibrillators for primary prevention of sudden cardiac death in subgroups: a systematic review. Annals of Internal Medicine, 160, 111–121.PubMedCrossRef Earley, A., Persson, R., Garlitski, A., Balk, E., & Uhlig, K. (2014). Effectiveness of implantable cardioverter defibrillators for primary prevention of sudden cardiac death in subgroups: a systematic review. Annals of Internal Medicine, 160, 111–121.PubMedCrossRef
2.
Zurück zum Zitat Bernard, M. L., Shotwell, M., Nietert, P. J., & Gold, M. R. (2012). Meta-analysis of bleeding complications associated with cardiac rhythm device implantation. Circulation. Arrhythmia and Electrophysiology, 5, 468–474.PubMedCentralPubMedCrossRef Bernard, M. L., Shotwell, M., Nietert, P. J., & Gold, M. R. (2012). Meta-analysis of bleeding complications associated with cardiac rhythm device implantation. Circulation. Arrhythmia and Electrophysiology, 5, 468–474.PubMedCentralPubMedCrossRef
3.
Zurück zum Zitat Cheng, A., Wang, Y., Curtis, J. P., & Varosy, P. D. (2010). Acute lead dislodgements and in-hospital mortality in patients enrolled in the national cardiovascular data registry implantable cardioverter defibrillator registry. Journal of the American College of Cardiology, 56, 1651–1656.PubMedCrossRef Cheng, A., Wang, Y., Curtis, J. P., & Varosy, P. D. (2010). Acute lead dislodgements and in-hospital mortality in patients enrolled in the national cardiovascular data registry implantable cardioverter defibrillator registry. Journal of the American College of Cardiology, 56, 1651–1656.PubMedCrossRef
4.
Zurück zum Zitat Polin, G. M., Zado, E., Nayak, H., Cooper, J. M., Russo, A. M., Dixit, S., et al. (2006). Proper management of pericardial tamponade as a late complication of implantable cardiac device placement. American Journal of Cardiology, 98, 223–225.PubMedCrossRef Polin, G. M., Zado, E., Nayak, H., Cooper, J. M., Russo, A. M., Dixit, S., et al. (2006). Proper management of pericardial tamponade as a late complication of implantable cardiac device placement. American Journal of Cardiology, 98, 223–225.PubMedCrossRef
5.
Zurück zum Zitat Sohail, M. R., Henrikson, C. A., Braid-Forbes, M. J., Forbes, K. F., & Lerner, D. J. (2011). Mortality and cost associated with cardiovascular implantable electronic device infections. Archives of Internal Medicine, 171, 1821–1828.PubMedCrossRef Sohail, M. R., Henrikson, C. A., Braid-Forbes, M. J., Forbes, K. F., & Lerner, D. J. (2011). Mortality and cost associated with cardiovascular implantable electronic device infections. Archives of Internal Medicine, 171, 1821–1828.PubMedCrossRef
6.
Zurück zum Zitat van Rees, J. B., de Bie, M. K., Thijssen, J., Borleffs, C. J., Schalij, M. J., & van Erven, L. (2011). Implantation-related complications of implantable cardioverter-defibrillators and cardiac resynchronization therapy devices: a systematic review of randomized clinical trials. Journal of the American College of Cardiology, 58, 995–1000.PubMedCrossRef van Rees, J. B., de Bie, M. K., Thijssen, J., Borleffs, C. J., Schalij, M. J., & van Erven, L. (2011). Implantation-related complications of implantable cardioverter-defibrillators and cardiac resynchronization therapy devices: a systematic review of randomized clinical trials. Journal of the American College of Cardiology, 58, 995–1000.PubMedCrossRef
7.
Zurück zum Zitat Borne, R. T., Varosy, P. D., & Masoudi, F. A. (2013). Implantable cardioverter-defibrillator shocks: epidemiology, outcomes, and therapeutic approaches. JAMA Internal Medicine, 173, 859–865.PubMedCrossRef Borne, R. T., Varosy, P. D., & Masoudi, F. A. (2013). Implantable cardioverter-defibrillator shocks: epidemiology, outcomes, and therapeutic approaches. JAMA Internal Medicine, 173, 859–865.PubMedCrossRef
8.
Zurück zum Zitat Bostwick, J. M., & Sola, C. L. (2007). An updated review of implantable cardioverter/defibrillators, induced anxiety, and quality of life. The Psychiatric Clinics of North America, 30, 677–688.PubMedCrossRef Bostwick, J. M., & Sola, C. L. (2007). An updated review of implantable cardioverter/defibrillators, induced anxiety, and quality of life. The Psychiatric Clinics of North America, 30, 677–688.PubMedCrossRef
9.
Zurück zum Zitat Germano, J. J., Reynolds, M., Essebag, V., & Josephson, M. E. (2006). Frequency and causes of implantable cardioverter-defibrillator therapies: is device therapy proarrhythmic? American Journal of Cardiology, 97, 1255–1261.PubMedCrossRef Germano, J. J., Reynolds, M., Essebag, V., & Josephson, M. E. (2006). Frequency and causes of implantable cardioverter-defibrillator therapies: is device therapy proarrhythmic? American Journal of Cardiology, 97, 1255–1261.PubMedCrossRef
10.
Zurück zum Zitat Hauptman, P. J., Chibnall, J. T., Guild, C., & Armbrecht, E. S. (2013). Patient perceptions, physician communication, and the implantable cardioverter-defibrillator. JAMA Internal Medicine, 173, 571–577.PubMedCrossRef Hauptman, P. J., Chibnall, J. T., Guild, C., & Armbrecht, E. S. (2013). Patient perceptions, physician communication, and the implantable cardioverter-defibrillator. JAMA Internal Medicine, 173, 571–577.PubMedCrossRef
11.
Zurück zum Zitat Matlock, D. D., Nowels, C. T., Masoudi, F. A., Sauer, W. H., Bekelman, D. B., Main, D. S., et al. (2011). Patient and cardiologist perceptions on decision making for implantable cardioverter-defibrillators: a qualitative study. Pacing and Clinical Electrophysiology, 34, 1634–1644.PubMedCrossRef Matlock, D. D., Nowels, C. T., Masoudi, F. A., Sauer, W. H., Bekelman, D. B., Main, D. S., et al. (2011). Patient and cardiologist perceptions on decision making for implantable cardioverter-defibrillators: a qualitative study. Pacing and Clinical Electrophysiology, 34, 1634–1644.PubMedCrossRef
12.
Zurück zum Zitat Uhlig, K., Balk, E. M., Earley, A., Persson, R., Garlitski, A. C., Chen, M., et al. (2013). Assessment on implantable defibrillators and the evidence for primary prevention of sudden cardiac death. Evidence Report/Technology Assessment. (Prepared by the Tufts Evidence-based Practice Center under Contract No. 290-2007-10055-I.) Rockville: Agency for Healthcare Research and Quality. www.effectivehealthcare.gov/reports/final.cfm. Uhlig, K., Balk, E. M., Earley, A., Persson, R., Garlitski, A. C., Chen, M., et al. (2013). Assessment on implantable defibrillators and the evidence for primary prevention of sudden cardiac death. Evidence Report/Technology Assessment. (Prepared by the Tufts Evidence-based Practice Center under Contract No. 290-2007-10055-I.) Rockville: Agency for Healthcare Research and Quality. www.​effectivehealthc​are.​gov/​reports/​final.​cfm.
13.
Zurück zum Zitat Hammill, S. C., Kremers, M. S., Kadish, A. H., Stevenson, L. W., Heidenreich, P. A., Lindsay, B. D., et al. (2009). Review of the ICD Registry’s third year, expansion to include lead data and pediatric ICD procedures, and role for measuring performance. Heart Rhythm, 6, 1397–1401.PubMedCrossRef Hammill, S. C., Kremers, M. S., Kadish, A. H., Stevenson, L. W., Heidenreich, P. A., Lindsay, B. D., et al. (2009). Review of the ICD Registry’s third year, expansion to include lead data and pediatric ICD procedures, and role for measuring performance. Heart Rhythm, 6, 1397–1401.PubMedCrossRef
14.
Zurück zum Zitat Wallace, B. C., Trikalinos, T. A., Lau, J., Brodley, C., & Schmid, C. H. (2010). Semi-automated screening of biomedical citations for systematic reviews. BMC Bioinformatics, 11, 55.PubMedCentralPubMedCrossRef Wallace, B. C., Trikalinos, T. A., Lau, J., Brodley, C., & Schmid, C. H. (2010). Semi-automated screening of biomedical citations for systematic reviews. BMC Bioinformatics, 11, 55.PubMedCentralPubMedCrossRef
15.
Zurück zum Zitat Chou, R., Aronson, N., Atkins, D., Ismaila, A. S., Santaguida, P., Smith, D. H., et al. (2010). AHRQ series paper 4: assessing harms when comparing medical interventions: AHRQ and the effective health-care program. Journal of Clinical Epidemiology, 63, 502–512.PubMedCrossRef Chou, R., Aronson, N., Atkins, D., Ismaila, A. S., Santaguida, P., Smith, D. H., et al. (2010). AHRQ series paper 4: assessing harms when comparing medical interventions: AHRQ and the effective health-care program. Journal of Clinical Epidemiology, 63, 502–512.PubMedCrossRef
17.
Zurück zum Zitat Owens, D. K., Lohr, K. N., Atkins, D., Treadwell, J. R., Reston, J. T., Bass, E. B., et al. (2010). AHRQ series paper 5: grading the strength of a body of evidence when comparing medical interventions—agency for healthcare research and quality and the effective health-care program. Journal of Clinical Epidemiology, 63, 513–523.PubMedCrossRef Owens, D. K., Lohr, K. N., Atkins, D., Treadwell, J. R., Reston, J. T., Bass, E. B., et al. (2010). AHRQ series paper 5: grading the strength of a body of evidence when comparing medical interventions—agency for healthcare research and quality and the effective health-care program. Journal of Clinical Epidemiology, 63, 513–523.PubMedCrossRef
18.
Zurück zum Zitat Aggarwal, A., Wang, Y., Rumsfeld, J. S., Curtis, J. P., & Heidenreich, P. A. (2009). Clinical characteristics and in-hospital outcome of patients with end-stage renal disease on dialysis referred for implantable cardioverter-defibrillator implantation. Heart Rhythm, 6, 1565–1571.PubMedCrossRef Aggarwal, A., Wang, Y., Rumsfeld, J. S., Curtis, J. P., & Heidenreich, P. A. (2009). Clinical characteristics and in-hospital outcome of patients with end-stage renal disease on dialysis referred for implantable cardioverter-defibrillator implantation. Heart Rhythm, 6, 1565–1571.PubMedCrossRef
19.
Zurück zum Zitat Curtis, J. P., Luebbert, J. J., Wang, Y., Rathore, S. S., Chen, J., Heidenreich, P. A., et al. (2009). Association of physician certification and outcomes among patients receiving an implantable cardioverter-defibrillator. JAMA, 301, 1661–1670.PubMedCentralPubMedCrossRef Curtis, J. P., Luebbert, J. J., Wang, Y., Rathore, S. S., Chen, J., Heidenreich, P. A., et al. (2009). Association of physician certification and outcomes among patients receiving an implantable cardioverter-defibrillator. JAMA, 301, 1661–1670.PubMedCentralPubMedCrossRef
20.
Zurück zum Zitat Dewland, T. A., Pellegrini, C. N., Wang, Y., Marcus, G. M., Keung, E., & Varosy, P. D. (2011). Dual-chamber implantable cardioverter-defibrillator selection is associated with increased complication rates and mortality among patients enrolled in the NCDR implantable cardioverter-defibrillator registry. Journal of the American College of Cardiology, 58, 1007–1013.PubMedCrossRef Dewland, T. A., Pellegrini, C. N., Wang, Y., Marcus, G. M., Keung, E., & Varosy, P. D. (2011). Dual-chamber implantable cardioverter-defibrillator selection is associated with increased complication rates and mortality among patients enrolled in the NCDR implantable cardioverter-defibrillator registry. Journal of the American College of Cardiology, 58, 1007–1013.PubMedCrossRef
21.
Zurück zum Zitat Freeman, J. V., Wang, Y., Curtis, J. P., Heidenreich, P. A., & Hlatky, M. A. (2010). The relation between hospital procedure volume and complications of cardioverter-defibrillator implantation from the implantable cardioverter-defibrillator registry. Journal of the American College of Cardiology, 56, 1133–1139.PubMedCrossRef Freeman, J. V., Wang, Y., Curtis, J. P., Heidenreich, P. A., & Hlatky, M. A. (2010). The relation between hospital procedure volume and complications of cardioverter-defibrillator implantation from the implantable cardioverter-defibrillator registry. Journal of the American College of Cardiology, 56, 1133–1139.PubMedCrossRef
22.
Zurück zum Zitat Freeman, J. V., Wang, Y., Curtis, J. P., Heidenreich, P. A., & Hlatky, M. A. (2012). Physician procedure volume and complications of cardioverter-defibrillator implantation. Circulation, 125, 57–64.PubMedCrossRef Freeman, J. V., Wang, Y., Curtis, J. P., Heidenreich, P. A., & Hlatky, M. A. (2012). Physician procedure volume and complications of cardioverter-defibrillator implantation. Circulation, 125, 57–64.PubMedCrossRef
23.
Zurück zum Zitat Haines, D. E., Wang, Y., & Curtis, J. (2011). Implantable cardioverter-defibrillator registry risk score models for acute procedural complications or death after implantable cardioverter-defibrillator implantation. Circulation, 123, 2069–2076.PubMedCrossRef Haines, D. E., Wang, Y., & Curtis, J. (2011). Implantable cardioverter-defibrillator registry risk score models for acute procedural complications or death after implantable cardioverter-defibrillator implantation. Circulation, 123, 2069–2076.PubMedCrossRef
24.
Zurück zum Zitat Peterson, P. N., Daugherty, S. L., Wang, Y., Vidaillet, H. J., Heidenreich, P. A., Curtis, J. P., et al. (2009). Gender differences in procedure-related adverse events in patients receiving implantable cardioverter-defibrillator therapy. Circulation, 119, 1078–1084.PubMedCrossRef Peterson, P. N., Daugherty, S. L., Wang, Y., Vidaillet, H. J., Heidenreich, P. A., Curtis, J. P., et al. (2009). Gender differences in procedure-related adverse events in patients receiving implantable cardioverter-defibrillator therapy. Circulation, 119, 1078–1084.PubMedCrossRef
25.
Zurück zum Zitat Tsai, V., Goldstein, M. K., Hsia, H. H., Wang, Y., Curtis, J., & Heidenreich, P. A. (2011). Age differences in primary prevention implantable cardioverter-defibrillator use in U.S. individuals. Journal of American Geriatrics Society, 59, 1589–1595.CrossRef Tsai, V., Goldstein, M. K., Hsia, H. H., Wang, Y., Curtis, J., & Heidenreich, P. A. (2011). Age differences in primary prevention implantable cardioverter-defibrillator use in U.S. individuals. Journal of American Geriatrics Society, 59, 1589–1595.CrossRef
26.
Zurück zum Zitat Tsai, V., Goldstein, M. K., Hsia, H. H., Wang, Y., Curtis, J., & Heidenreich, P. A. (2011). Influence of age on perioperative complications among patients undergoing implantable cardioverter-defibrillators for primary prevention in the United States. Circulation. Cardiovascular Quality and Outcomes, 4, 549–556.PubMedCrossRef Tsai, V., Goldstein, M. K., Hsia, H. H., Wang, Y., Curtis, J., & Heidenreich, P. A. (2011). Influence of age on perioperative complications among patients undergoing implantable cardioverter-defibrillators for primary prevention in the United States. Circulation. Cardiovascular Quality and Outcomes, 4, 549–556.PubMedCrossRef
27.
Zurück zum Zitat Sengupta, J., Kendig, A. C., Goormastic, M., Hwang, E. S., Ching, E. A., Chung, R., et al. (2012). Implantable cardioverter-defibrillator FDA safety advisories: impact on patient mortality and morbidity. Heart Rhythm, 9, 1619–1626.PubMedCentralPubMedCrossRef Sengupta, J., Kendig, A. C., Goormastic, M., Hwang, E. S., Ching, E. A., Chung, R., et al. (2012). Implantable cardioverter-defibrillator FDA safety advisories: impact on patient mortality and morbidity. Heart Rhythm, 9, 1619–1626.PubMedCentralPubMedCrossRef
28.
Zurück zum Zitat Lyman, S., Sedrakyan, A., Do, H., Razzano, R., & Mushlin, A. I. (2011). Infrequent physician use of implantable cardioverter-defibrillators risks patient safety. Heart, 97, 1655–1660.PubMedCrossRef Lyman, S., Sedrakyan, A., Do, H., Razzano, R., & Mushlin, A. I. (2011). Infrequent physician use of implantable cardioverter-defibrillators risks patient safety. Heart, 97, 1655–1660.PubMedCrossRef
29.
Zurück zum Zitat Thijssen, J., Borleffs, C. J., van Rees, J. B., Man, S., de Bie, M. K., Venlet, J., et al. (2012). Implantable cardioverter-defibrillator longevity under clinical circumstances: an analysis according to device type, generation, and manufacturer. Heart Rhythm, 9, 513–519.PubMedCrossRef Thijssen, J., Borleffs, C. J., van Rees, J. B., Man, S., de Bie, M. K., Venlet, J., et al. (2012). Implantable cardioverter-defibrillator longevity under clinical circumstances: an analysis according to device type, generation, and manufacturer. Heart Rhythm, 9, 513–519.PubMedCrossRef
30.
Zurück zum Zitat Tsai, F. S., Aronow, W. S., Devabhaktuni, S., Desai, H., Kruger, A., Lai, H. M., et al. (2010). Prevalence of complications during implantation and during 38-month follow-up of 1060 consecutive patients with implantable cardioverter-defibrillators. American Journal of Therapy, 17, e8–e10.CrossRef Tsai, F. S., Aronow, W. S., Devabhaktuni, S., Desai, H., Kruger, A., Lai, H. M., et al. (2010). Prevalence of complications during implantation and during 38-month follow-up of 1060 consecutive patients with implantable cardioverter-defibrillators. American Journal of Therapy, 17, e8–e10.CrossRef
31.
Zurück zum Zitat Varma, N., Michalski, J., Epstein, A. E., & Schweikert, R. (2010). Automatic remote monitoring of implantable cardioverter-defibrillator lead and generator performance: the Lumos-T Safely RedUceS RouTine Office Device Follow-Up (TRUST) trial. Circulation. Arrhythmia and Electrophysiology, 3, 428–436.PubMedCrossRef Varma, N., Michalski, J., Epstein, A. E., & Schweikert, R. (2010). Automatic remote monitoring of implantable cardioverter-defibrillator lead and generator performance: the Lumos-T Safely RedUceS RouTine Office Device Follow-Up (TRUST) trial. Circulation. Arrhythmia and Electrophysiology, 3, 428–436.PubMedCrossRef
32.
Zurück zum Zitat Landolina, M., Gasparini, M., Lunati, M., Iacopino, S., Boriani, G., Bonanno, C., et al. (2011). Long-term complications related to biventricular defibrillator implantation: rate of surgical revisions and impact on survival: insights from the Italian Clinical Service Database. Circulation, 123, 2526–2535.PubMedCrossRef Landolina, M., Gasparini, M., Lunati, M., Iacopino, S., Boriani, G., Bonanno, C., et al. (2011). Long-term complications related to biventricular defibrillator implantation: rate of surgical revisions and impact on survival: insights from the Italian Clinical Service Database. Circulation, 123, 2526–2535.PubMedCrossRef
33.
Zurück zum Zitat Gradaus, R., Block, M., Brachmann, J., Breithardt, G., Huber, H. G., Jung, W., et al. (2003). Mortality, morbidity, and complications in 3344 patients with implantable cardioverter defibrillators: results fron the German ICD Registry EURID. Pacing and Clinical Electrophysiology, 26, t-8.CrossRef Gradaus, R., Block, M., Brachmann, J., Breithardt, G., Huber, H. G., Jung, W., et al. (2003). Mortality, morbidity, and complications in 3344 patients with implantable cardioverter defibrillators: results fron the German ICD Registry EURID. Pacing and Clinical Electrophysiology, 26, t-8.CrossRef
34.
Zurück zum Zitat Porterfield, J. G., Porterfield, L. M., Kuck, K. H., Corbisiero, R., Greenberg, S. M., Hindricks, G., et al. (2010). Clinical performance of the St. Jude Medical Riata defibrillation lead in a large patient population. Journal of Cardiovascular Electrophysiology, 21, 551–556.PubMedCrossRef Porterfield, J. G., Porterfield, L. M., Kuck, K. H., Corbisiero, R., Greenberg, S. M., Hindricks, G., et al. (2010). Clinical performance of the St. Jude Medical Riata defibrillation lead in a large patient population. Journal of Cardiovascular Electrophysiology, 21, 551–556.PubMedCrossRef
35.
Zurück zum Zitat Dichtl, W., Wolber, T., Paoli, U., Brüllmann, S., Stühlinger, M., Berger, T., et al. (2011). Appropriate therapy but not inappropriate shocks predict survival in implantable cardioverter defibrillator patients. Clinical Cardiology, 34, 433–436.PubMedCrossRef Dichtl, W., Wolber, T., Paoli, U., Brüllmann, S., Stühlinger, M., Berger, T., et al. (2011). Appropriate therapy but not inappropriate shocks predict survival in implantable cardioverter defibrillator patients. Clinical Cardiology, 34, 433–436.PubMedCrossRef
36.
Zurück zum Zitat Bode, F., Himmel, F., Reppel, M., Mortensen, K., Schunkert, H., & Wiegand, U. K. (2012). Should all dysfunctional high-voltage leads be extracted? Results of a single-centre long-term registry. Europace, 14, 1764–1770.PubMedCrossRef Bode, F., Himmel, F., Reppel, M., Mortensen, K., Schunkert, H., & Wiegand, U. K. (2012). Should all dysfunctional high-voltage leads be extracted? Results of a single-centre long-term registry. Europace, 14, 1764–1770.PubMedCrossRef
37.
Zurück zum Zitat Moss, A. J., Zareba, W., Hall, W. J., Klein, H., Wilber, D. J., Cannom, D. S., et al. (2002). Prophylactic implantation of a defibrillator in patients with myocardial infarction and reduced ejection fraction. New England Journal of Medicine, 346, 877–883.PubMedCrossRef Moss, A. J., Zareba, W., Hall, W. J., Klein, H., Wilber, D. J., Cannom, D. S., et al. (2002). Prophylactic implantation of a defibrillator in patients with myocardial infarction and reduced ejection fraction. New England Journal of Medicine, 346, 877–883.PubMedCrossRef
38.
Zurück zum Zitat Borleffs, C. J., van Erven, L., van Bommel, R. J., van der Velde, E. T., van der Wall, E. E., Bax, J. J., et al. (2009). Risk of failure of transvenous implantable cardioverter-defibrillator leads. Circulation. Arrhythmia and Electrophysiology, 2, 411–416.PubMedCrossRef Borleffs, C. J., van Erven, L., van Bommel, R. J., van der Velde, E. T., van der Wall, E. E., Bax, J. J., et al. (2009). Risk of failure of transvenous implantable cardioverter-defibrillator leads. Circulation. Arrhythmia and Electrophysiology, 2, 411–416.PubMedCrossRef
39.
Zurück zum Zitat Hauser, R. G., Mugglin, A. S., Friedman, P. A., Kramer, D. B., Kallinen, L., McGriff, D., et al. (2012). Early detection of an underperforming implantable cardiovascular device using an automated safety surveillance tool. Circulation. Cardiovascular Quality and Outcomes, 5, 189–196.PubMedCrossRef Hauser, R. G., Mugglin, A. S., Friedman, P. A., Kramer, D. B., Kallinen, L., McGriff, D., et al. (2012). Early detection of an underperforming implantable cardiovascular device using an automated safety surveillance tool. Circulation. Cardiovascular Quality and Outcomes, 5, 189–196.PubMedCrossRef
40.
Zurück zum Zitat Morrison, T. B., Friedman, P. A., Kallinen, L. M., Hodge, D. O., Crusan, D., Kumar, K., et al. (2011). Impact of implanted recalled sprint Fidelis lead on patient mortality. Journal of the American College of Cardiology, 58, 278–283.PubMedCrossRef Morrison, T. B., Friedman, P. A., Kallinen, L. M., Hodge, D. O., Crusan, D., Kumar, K., et al. (2011). Impact of implanted recalled sprint Fidelis lead on patient mortality. Journal of the American College of Cardiology, 58, 278–283.PubMedCrossRef
41.
Zurück zum Zitat Parkash, R., Thibault, B., Sterns, L., Sapp, J., Krahn, A., Talajic, M., et al. (2012). Sprint Fidelis lead fractures in patients with cardiac resynchronization therapy devices: insight from the Resynchronization/Defibrillation for Ambulatory Heart Failure (RAFT) study. Circulation, 126, 2928–2934.PubMedCrossRef Parkash, R., Thibault, B., Sterns, L., Sapp, J., Krahn, A., Talajic, M., et al. (2012). Sprint Fidelis lead fractures in patients with cardiac resynchronization therapy devices: insight from the Resynchronization/Defibrillation for Ambulatory Heart Failure (RAFT) study. Circulation, 126, 2928–2934.PubMedCrossRef
42.
Zurück zum Zitat van Rees, J. B., van Welsenes, G. H., Borleffs, C. J., Thijssen, J., van der Velde, E. T., van der Wall, E. E., et al. (2012). Update on small-diameter implantable cardioverter-defibrillator leads performance. Pacing and Clinical Electrophysiology, 35, 652–658.PubMedCrossRef van Rees, J. B., van Welsenes, G. H., Borleffs, C. J., Thijssen, J., van der Velde, E. T., van der Wall, E. E., et al. (2012). Update on small-diameter implantable cardioverter-defibrillator leads performance. Pacing and Clinical Electrophysiology, 35, 652–658.PubMedCrossRef
43.
Zurück zum Zitat MacFadden, D. R., Crystal, E., Krahn, A. D., Mangat, I., Healey, J. S., Dorian, P., et al. (2012). Sex differences in implantable cardioverter-defibrillator outcomes: findings from a prospective defibrillator database. Annals of Internal Medicine, 156, 195–203.PubMedCrossRef MacFadden, D. R., Crystal, E., Krahn, A. D., Mangat, I., Healey, J. S., Dorian, P., et al. (2012). Sex differences in implantable cardioverter-defibrillator outcomes: findings from a prospective defibrillator database. Annals of Internal Medicine, 156, 195–203.PubMedCrossRef
44.
Zurück zum Zitat Charytan, D. M., Patrick, A. R., Liu, J., Setoguchi, S., Herzog, C. A., Brookhart, M. A., et al. (2011). Trends in the use and outcomes of implantable cardioverter-defibrillators in patients undergoing dialysis in the United States. American Journal of Kidney Diseases, 58, 409–417.PubMedCrossRef Charytan, D. M., Patrick, A. R., Liu, J., Setoguchi, S., Herzog, C. A., Brookhart, M. A., et al. (2011). Trends in the use and outcomes of implantable cardioverter-defibrillators in patients undergoing dialysis in the United States. American Journal of Kidney Diseases, 58, 409–417.PubMedCrossRef
45.
Zurück zum Zitat Brüllmann, S., Dichtl, W., Paoli, U., Haegeli, L., Schmied, C., Steffel, J., et al. (2012). Comparison of benefit and mortality of implantable cardioverter-defibrillator therapy in patients aged >=75 years versus those <75 years. American Journal of Cardiology, 109, 712–717.PubMedCrossRef Brüllmann, S., Dichtl, W., Paoli, U., Haegeli, L., Schmied, C., Steffel, J., et al. (2012). Comparison of benefit and mortality of implantable cardioverter-defibrillator therapy in patients aged >=75 years versus those <75 years. American Journal of Cardiology, 109, 712–717.PubMedCrossRef
46.
Zurück zum Zitat Birnie, D. H., Parkash, R., Exner, D. V., Essebag, V., Healey, J. S., Verma, A., et al. (2012). Clinical predictors of Fidelis lead failure: report from the Canadian Heart Rhythm Society Device Committee. Circulation, 125, 1217–1225.PubMedCrossRef Birnie, D. H., Parkash, R., Exner, D. V., Essebag, V., Healey, J. S., Verma, A., et al. (2012). Clinical predictors of Fidelis lead failure: report from the Canadian Heart Rhythm Society Device Committee. Circulation, 125, 1217–1225.PubMedCrossRef
47.
Zurück zum Zitat Cheung, J. W., Tobin-Hess, A., Patel, A., Slotwiner, D. J., & Goldner, B. G. (2012). Trends in Fidelis lead survival: transition from an exponential to linear pattern of lead failure over time. Circulation. Arrhythmia and Electrophysiology, 5, 906–912.PubMedCrossRef Cheung, J. W., Tobin-Hess, A., Patel, A., Slotwiner, D. J., & Goldner, B. G. (2012). Trends in Fidelis lead survival: transition from an exponential to linear pattern of lead failure over time. Circulation. Arrhythmia and Electrophysiology, 5, 906–912.PubMedCrossRef
48.
Zurück zum Zitat Tzogias, L., Bellavia, D., Sharma, S., Donohue, T. J., & Schoenfeld, M. H. (2012). Natural history of the Sprint Fidelis lead: survival analysis from a large single-center study. Journal of Interventional Cardiac Electrophysiology, 34, 37–44.PubMedCrossRef Tzogias, L., Bellavia, D., Sharma, S., Donohue, T. J., & Schoenfeld, M. H. (2012). Natural history of the Sprint Fidelis lead: survival analysis from a large single-center study. Journal of Interventional Cardiac Electrophysiology, 34, 37–44.PubMedCrossRef
49.
Zurück zum Zitat Brumberg, G. E., Kaseer, B., Shah, H., Saba, S., & Jain, S. (2012). Biventricular defibrillator patients have higher complication rates after revision of recalled leads. Pacing and Clinical Electrophysiology, 35, 665–671.PubMedCrossRef Brumberg, G. E., Kaseer, B., Shah, H., Saba, S., & Jain, S. (2012). Biventricular defibrillator patients have higher complication rates after revision of recalled leads. Pacing and Clinical Electrophysiology, 35, 665–671.PubMedCrossRef
50.
Zurück zum Zitat de Bie, M. K., van Rees, J. B., Thijssen, J., Borleffs, C. J., Trines, S. A., Cannegieter, S. C., et al. (2012). Cardiac device infections are associated with a significant mortality risk. Heart Rhythm, 9, 494–498.PubMedCrossRef de Bie, M. K., van Rees, J. B., Thijssen, J., Borleffs, C. J., Trines, S. A., Cannegieter, S. C., et al. (2012). Cardiac device infections are associated with a significant mortality risk. Heart Rhythm, 9, 494–498.PubMedCrossRef
51.
Zurück zum Zitat Remmelts, H. H., Meine, M., Loh, P., Hauer, R., Doevendans, P. A., van Herwerden, L. A., et al. (2009). Infection after ICD implantation: operating room versus cardiac catheterisation laboratory. Netherlands Heart Journal, 17, 95–100.PubMedCentralPubMedCrossRef Remmelts, H. H., Meine, M., Loh, P., Hauer, R., Doevendans, P. A., van Herwerden, L. A., et al. (2009). Infection after ICD implantation: operating room versus cardiac catheterisation laboratory. Netherlands Heart Journal, 17, 95–100.PubMedCentralPubMedCrossRef
52.
Zurück zum Zitat Lee, D. S., Krahn, A. D., Healey, J. S., Birnie, D., Crystal, E., Dorian, P., et al. (2010). Evaluation of early complications related to de novo cardioverter defibrillator implantation insights from the Ontario ICD database. Journal of the American College of Cardiology, 55, 774–782.PubMedCrossRef Lee, D. S., Krahn, A. D., Healey, J. S., Birnie, D., Crystal, E., Dorian, P., et al. (2010). Evaluation of early complications related to de novo cardioverter defibrillator implantation insights from the Ontario ICD database. Journal of the American College of Cardiology, 55, 774–782.PubMedCrossRef
53.
Zurück zum Zitat Freudenberger, R. S., Hellkamp, A. S., Halperin, J. L., Poole, J., Anderson, J., Johnson, G., et al. (2007). Risk of thromboembolism in heart failure: an analysis from the Sudden Cardiac Death in Heart Failure Trial (SCD-HeFT). Circulation, 115, 2637–2641.PubMedCrossRef Freudenberger, R. S., Hellkamp, A. S., Halperin, J. L., Poole, J., Anderson, J., Johnson, G., et al. (2007). Risk of thromboembolism in heart failure: an analysis from the Sudden Cardiac Death in Heart Failure Trial (SCD-HeFT). Circulation, 115, 2637–2641.PubMedCrossRef
54.
Zurück zum Zitat Bardy, G. H., Lee, K. L., Mark, D. B., Poole, J. E., Packer, D. L., Boineau, R., et al. (2005). Amiodarone or an implantable cardioverter-defibrillator for congestive heart failure. New England Journal of Medicine, 352, 225–237.PubMedCrossRef Bardy, G. H., Lee, K. L., Mark, D. B., Poole, J. E., Packer, D. L., Boineau, R., et al. (2005). Amiodarone or an implantable cardioverter-defibrillator for congestive heart failure. New England Journal of Medicine, 352, 225–237.PubMedCrossRef
55.
Zurück zum Zitat Desai, H., Aronow, W. S., Ahn, C., Gandhi, K., Hussain, S., Lai, H. M., et al. (2010). Risk factors for appropriate cardioverter-defibrillator shocks, inappropriate cardioverter-defibrillator shocks, and time to mortality in 549 patients with heart failure. American Journal of Cardiology, 105, 1336–1338.PubMedCrossRef Desai, H., Aronow, W. S., Ahn, C., Gandhi, K., Hussain, S., Lai, H. M., et al. (2010). Risk factors for appropriate cardioverter-defibrillator shocks, inappropriate cardioverter-defibrillator shocks, and time to mortality in 549 patients with heart failure. American Journal of Cardiology, 105, 1336–1338.PubMedCrossRef
56.
Zurück zum Zitat Gold, M. R., Ahmad, S., Browne, K., Berg, K. C., Thackeray, L., & Berger, R. D. (2012). Prospective comparison of discrimination algorithms to prevent inappropriate ICD therapy: primary results of the Rhythm ID Going Head to Head Trial. Heart Rhythm, 9, 370–377.PubMedCrossRef Gold, M. R., Ahmad, S., Browne, K., Berg, K. C., Thackeray, L., & Berger, R. D. (2012). Prospective comparison of discrimination algorithms to prevent inappropriate ICD therapy: primary results of the Rhythm ID Going Head to Head Trial. Heart Rhythm, 9, 370–377.PubMedCrossRef
57.
Zurück zum Zitat Kleemann, T., Hochadel, M., Strauss, M., Skarlos, A., Seidl, K., & Zahn, R. (2012). Comparison between atrial fibrillation-triggered implantable cardioverter-defibrillator (ICD) shocks and inappropriate shocks caused by lead failure: different impact on prognosis in clinical practice. Journal of Cardiovascular Electrophysiology, 23, 735–740.PubMedCrossRef Kleemann, T., Hochadel, M., Strauss, M., Skarlos, A., Seidl, K., & Zahn, R. (2012). Comparison between atrial fibrillation-triggered implantable cardioverter-defibrillator (ICD) shocks and inappropriate shocks caused by lead failure: different impact on prognosis in clinical practice. Journal of Cardiovascular Electrophysiology, 23, 735–740.PubMedCrossRef
58.
Zurück zum Zitat Larsen, P. D., De, S. P., Harding, S. A., Woodcock, E., & Lever, N. A. (2010). Use of implantable cardioverter defibrillators in the New Zealand context from 2000 to 2007. New Zealand Medical Journal, 123, 76–85.PubMed Larsen, P. D., De, S. P., Harding, S. A., Woodcock, E., & Lever, N. A. (2010). Use of implantable cardioverter defibrillators in the New Zealand context from 2000 to 2007. New Zealand Medical Journal, 123, 76–85.PubMed
59.
Zurück zum Zitat Moss, A. J., Schuger, C., Beck, C. A., Brown, M. W., Cannom, D. S., Daubert, J. P., et al. (2012). Reduction in inappropriate therapy and mortality through ICD programming. New England Journal of Medicine, 367, 2275–2283.PubMedCrossRef Moss, A. J., Schuger, C., Beck, C. A., Brown, M. W., Cannom, D. S., Daubert, J. P., et al. (2012). Reduction in inappropriate therapy and mortality through ICD programming. New England Journal of Medicine, 367, 2275–2283.PubMedCrossRef
60.
Zurück zum Zitat Sandesara, C. M., Sullivan, R. M., Russo, A. M., Li, W., Kendig, A., Day, J. D., et al. (2011). Older persons with diabetes receive fewer inappropriate ICD shocks: results from the INTRINSIC RV trial. Journal of Cardiovascular Translational Research, 4, 27–34.PubMedCrossRef Sandesara, C. M., Sullivan, R. M., Russo, A. M., Li, W., Kendig, A., Day, J. D., et al. (2011). Older persons with diabetes receive fewer inappropriate ICD shocks: results from the INTRINSIC RV trial. Journal of Cardiovascular Translational Research, 4, 27–34.PubMedCrossRef
61.
Zurück zum Zitat Saxon, L. A., Hayes, D. L., Gilliam, F. R., Heidenreich, P. A., Day, J., Seth, M., et al. (2010). Long-term outcome after ICD and CRT implantation and influence of remote device follow-up: the ALTITUDE survival study. Circulation, 122, 2359–2367.PubMedCrossRef Saxon, L. A., Hayes, D. L., Gilliam, F. R., Heidenreich, P. A., Day, J., Seth, M., et al. (2010). Long-term outcome after ICD and CRT implantation and influence of remote device follow-up: the ALTITUDE survival study. Circulation, 122, 2359–2367.PubMedCrossRef
62.
Zurück zum Zitat Schaer, B., Sticherling, C., Szili-Torok, T., Osswald, S., Jordaens, L., & Theuns, D. A. (2011). Impact of left ventricular ejection fraction on occurrence of ventricular events in defibrillator patients with coronary artery disease. Europace, 13, 1562–1567.PubMedCrossRef Schaer, B., Sticherling, C., Szili-Torok, T., Osswald, S., Jordaens, L., & Theuns, D. A. (2011). Impact of left ventricular ejection fraction on occurrence of ventricular events in defibrillator patients with coronary artery disease. Europace, 13, 1562–1567.PubMedCrossRef
63.
Zurück zum Zitat Sweeney, M. O., Sherfesee, L., DeGroot, P. J., Wathen, M. S., & Wilkoff, B. L. (2010). Differences in effects of electrical therapy type for ventricular arrhythmias on mortality in implantable cardioverter-defibrillator patients. Heart Rhythm, 7, 353–360.PubMedCrossRef Sweeney, M. O., Sherfesee, L., DeGroot, P. J., Wathen, M. S., & Wilkoff, B. L. (2010). Differences in effects of electrical therapy type for ventricular arrhythmias on mortality in implantable cardioverter-defibrillator patients. Heart Rhythm, 7, 353–360.PubMedCrossRef
64.
Zurück zum Zitat van Welsenes, G. H., van Rees, J. B., Borleffs, C. J., Cannegieter, S. C., Bax, J. J., van Erven, L., et al. (2011). Long-term follow-up of primary and secondary prevention implantable cardioverter defibrillator patients. Europace, 13, 389–394.PubMedCrossRef van Welsenes, G. H., van Rees, J. B., Borleffs, C. J., Cannegieter, S. C., Bax, J. J., van Erven, L., et al. (2011). Long-term follow-up of primary and secondary prevention implantable cardioverter defibrillator patients. Europace, 13, 389–394.PubMedCrossRef
65.
Zurück zum Zitat Martin, D. T., McNitt, S., Nesto, R. W., Rutter, M. K., & Moss, A. J. (2011). Cardiac resynchronization therapy reduces the risk of cardiac events in patients with diabetes enrolled in the multicenter automatic defibrillator implantation trial with cardiac resynchronization therapy (MADIT-CRT). Circulation. Heart Failure, 4, 332–338.PubMedCrossRef Martin, D. T., McNitt, S., Nesto, R. W., Rutter, M. K., & Moss, A. J. (2011). Cardiac resynchronization therapy reduces the risk of cardiac events in patients with diabetes enrolled in the multicenter automatic defibrillator implantation trial with cardiac resynchronization therapy (MADIT-CRT). Circulation. Heart Failure, 4, 332–338.PubMedCrossRef
66.
Zurück zum Zitat Sweeney, M. O., Sakaguchi, S., Simons, G., Machado, C., Connett, J. E., Yang, F., et al. (2012). Response to the Center for Medicare & Medicaid Services coverage with evidence development request for primary prevention implantable cardioverter-defibrillators: data from the OMNI study. Heart Rhythm, 9, 1058–1066.PubMedCrossRef Sweeney, M. O., Sakaguchi, S., Simons, G., Machado, C., Connett, J. E., Yang, F., et al. (2012). Response to the Center for Medicare & Medicaid Services coverage with evidence development request for primary prevention implantable cardioverter-defibrillators: data from the OMNI study. Heart Rhythm, 9, 1058–1066.PubMedCrossRef
67.
Zurück zum Zitat van Rees, J. B., Borleffs, C. J., Thijssen, J., de Bie, M. K., van Erven, L., Cannegieter, S. C., et al. (2012). Prophylactic implantable cardioverter-defibrillator treatment in the elderly: therapy, adverse events, and survival gain. Europace, 14, 66–73.PubMedCrossRef van Rees, J. B., Borleffs, C. J., Thijssen, J., de Bie, M. K., van Erven, L., Cannegieter, S. C., et al. (2012). Prophylactic implantable cardioverter-defibrillator treatment in the elderly: therapy, adverse events, and survival gain. Europace, 14, 66–73.PubMedCrossRef
68.
Zurück zum Zitat Berenbom, L. D., Weiford, B. C., Vacek, J. L., Emert, M. P., Hall, W. J., Andrews, M. L., et al. (2005). Differences in outcomes between patients treated with single- versus dual-chamber implantable cardioverter defibrillators: a substudy of the Multicenter Automatic Defibrillator Implantation Trial II. Annals of Noninvasive Electrocardiology, 10, 429–435.PubMedCrossRef Berenbom, L. D., Weiford, B. C., Vacek, J. L., Emert, M. P., Hall, W. J., Andrews, M. L., et al. (2005). Differences in outcomes between patients treated with single- versus dual-chamber implantable cardioverter defibrillators: a substudy of the Multicenter Automatic Defibrillator Implantation Trial II. Annals of Noninvasive Electrocardiology, 10, 429–435.PubMedCrossRef
69.
Zurück zum Zitat Sweeney, M. O., Wathen, M. S., Volosin, K., Abdalla, I., DeGroot, P. J., Otterness, M. F., et al. (2005). Appropriate and inappropriate ventricular therapies, quality of life, and mortality among primary and secondary prevention implantable cardioverter defibrillator patients: results from the Pacing Fast VT REduces Shock ThErapies (PainFREE Rx II) trial. Circulation, 111, 2898–2905.PubMedCrossRef Sweeney, M. O., Wathen, M. S., Volosin, K., Abdalla, I., DeGroot, P. J., Otterness, M. F., et al. (2005). Appropriate and inappropriate ventricular therapies, quality of life, and mortality among primary and secondary prevention implantable cardioverter defibrillator patients: results from the Pacing Fast VT REduces Shock ThErapies (PainFREE Rx II) trial. Circulation, 111, 2898–2905.PubMedCrossRef
70.
Zurück zum Zitat van Rees, J. B., Borleffs, C. J., de Bie, M. K., Stijnen, T., van Erven, L., Bax, J. J., et al. (2011). Inappropriate implantable cardioverter-defibrillator shocks: incidence, predictors, and impact on mortality. Journal of the American College of Cardiology, 57, 556–562.PubMedCrossRef van Rees, J. B., Borleffs, C. J., de Bie, M. K., Stijnen, T., van Erven, L., Bax, J. J., et al. (2011). Inappropriate implantable cardioverter-defibrillator shocks: incidence, predictors, and impact on mortality. Journal of the American College of Cardiology, 57, 556–562.PubMedCrossRef
71.
Zurück zum Zitat Wei, S., Loyo-Berríos, N. I., Haigney, M. C., Cheng, H., Pinnow, E. E., Mitchell, K. R., et al. (2011). Elevated B-type natriuretic peptide is associated with increased in-hospital mortality or cardiac arrest in patients undergoing implantable cardioverter-defibrillator implantation. Circulation. Cardiovascular Quality and Outcomes, 4, 346–354.PubMedCrossRef Wei, S., Loyo-Berríos, N. I., Haigney, M. C., Cheng, H., Pinnow, E. E., Mitchell, K. R., et al. (2011). Elevated B-type natriuretic peptide is associated with increased in-hospital mortality or cardiac arrest in patients undergoing implantable cardioverter-defibrillator implantation. Circulation. Cardiovascular Quality and Outcomes, 4, 346–354.PubMedCrossRef
72.
Zurück zum Zitat Peterson, P. N., Varosy, P. D., Heidenreich, P. A., Wang, Y., Dewland, T. A., Curtis, J. P., et al. (2013). Association of single- vs dual-chamber ICDs with mortality, readmissions, and complications among patients receiving an ICD for primary prevention. JAMA, 309, 2025–2034.PubMedCentralPubMedCrossRef Peterson, P. N., Varosy, P. D., Heidenreich, P. A., Wang, Y., Dewland, T. A., Curtis, J. P., et al. (2013). Association of single- vs dual-chamber ICDs with mortality, readmissions, and complications among patients receiving an ICD for primary prevention. JAMA, 309, 2025–2034.PubMedCentralPubMedCrossRef
73.
Zurück zum Zitat Resnic, F. S., & Normand, S. L. (2012). Postmarketing surveillance of medical devices—filling in the gaps. New England Journal of Medicine, 366, 875–877.PubMedCrossRef Resnic, F. S., & Normand, S. L. (2012). Postmarketing surveillance of medical devices—filling in the gaps. New England Journal of Medicine, 366, 875–877.PubMedCrossRef
74.
Zurück zum Zitat Hauser, R. G. (2012). Here we go again—failure of postmarketing device surveillance. New England Journal of Medicine, 366, 873–875.PubMedCrossRef Hauser, R. G. (2012). Here we go again—failure of postmarketing device surveillance. New England Journal of Medicine, 366, 873–875.PubMedCrossRef
75.
Zurück zum Zitat Kremers, M. S., Hammill, S. C., Berul, C. I., Koutras, C., Curtis, J. S., Wang, Y., et al. (2013). The National ICD Registry Report: version 2.1 including leads and pediatrics for years 2010 and 2011. Heart Rhythm, 10, e59–e65.PubMedCrossRef Kremers, M. S., Hammill, S. C., Berul, C. I., Koutras, C., Curtis, J. S., Wang, Y., et al. (2013). The National ICD Registry Report: version 2.1 including leads and pediatrics for years 2010 and 2011. Heart Rhythm, 10, e59–e65.PubMedCrossRef
76.
Zurück zum Zitat Tan, V. H., Wilton, S. B., Kuriachan, V., Sumner, G. L., & Exner, D. V. (2014). Impact of programming strategies aimed at reducing nonessential implantable cardioverter defibrillator therapies on mortality: a systematic review and meta-analysis. Circulation. Arrhythmia and Electrophysiology, 7(1), 164–170.PubMedCrossRef Tan, V. H., Wilton, S. B., Kuriachan, V., Sumner, G. L., & Exner, D. V. (2014). Impact of programming strategies aimed at reducing nonessential implantable cardioverter defibrillator therapies on mortality: a systematic review and meta-analysis. Circulation. Arrhythmia and Electrophysiology, 7(1), 164–170.PubMedCrossRef
Metadaten
Titel
Adverse events following implantable cardioverter defibrillator implantation: a systematic review
verfasst von
Rebecca Persson
Amy Earley
Ann C. Garlitski
Ethan M. Balk
Katrin Uhlig
Publikationsdatum
01.08.2014
Verlag
Springer US
Erschienen in
Journal of Interventional Cardiac Electrophysiology / Ausgabe 2/2014
Print ISSN: 1383-875X
Elektronische ISSN: 1572-8595
DOI
https://doi.org/10.1007/s10840-014-9913-z

Weitere Artikel der Ausgabe 2/2014

Journal of Interventional Cardiac Electrophysiology 2/2014 Zur Ausgabe

Update Kardiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.