Skip to main content
Erschienen in: Journal of Interventional Cardiac Electrophysiology 2/2016

21.01.2016

Analysis of catheter contact force during atrial fibrillation ablation using the robotic navigation system: results from a randomized study

verfasst von: Antonio Dello Russo, Gaetano Fassini, Sergio Conti, Michela Casella, Antonio Di Monaco, Eleonora Russo, Stefania Riva, Massimo Moltrasio, Fabrizio Tundo, Giuseppe De Martino, G. Joseph Gallinghouse, Luigi Di Biase, Andrea Natale, Claudio Tondo

Erschienen in: Journal of Interventional Cardiac Electrophysiology | Ausgabe 2/2016

Einloggen, um Zugang zu erhalten

Abstract

Purpose

Contact with cardiac tissue is an important determinant of lesion efficacy during atrial fibrillation (AF) ablation. The Sensei X™ robotic navigation system (RNS) (Hansen Medical, Mountain View, CA, USA) has been validated for contact force (CF) sensing expressed in grams (g). The Thermocool® SmartTouch™ catheter enables the measurement of catheter tip CF and direction inside the heart. We aimed to investigate the catheter CF with and without RNS during pulmonary vein isolation (PVI) procedures.

Methods

Eighty patients with symptomatic AF (56 males, age 63 ± 18) were enrolled in this study. Fifty-seven patients had paroxysmal AF and 23 early persistent AF. All procedures were performed with the Thermocool® SmartTouch™ ablation catheter. Forty patients were randomized to perform PVI with the Sensei X™ RNS (group 1), while in the other 40 patients (group 2), PVI was performed without the RNS.

Results

AF ablation was performed successfully in all patients without complications, while contact force was kept in the established 10–40 g range. A significantly higher CF was documented on the PVs in group 1 compared to group 2. The 1-year freedom from AF recurrence was higher in group 1 compared to group 2 (90 vs. 65 %, p = 0.04). Moreover, a significant reduction of fluoroscopy time was noted in the RNS group (13 ± 10 vs. 20 ± 10 min, respectively, p = 0.05).

Conclusions

The Sensei X™ RNS permits a significantly higher CF during transcatheter AF ablation with a low rate of AF recurrence at clinical follow-up.
Literatur
1.
Zurück zum Zitat Camm, A. J., Lip, G. Y., De Caterina, R., Savelieva, I., Atar, D., Hohnloser, S. H., et al. (2012). ESC Committee for Practice Guidelines (CPG). 2012 focused update of the ESC Guidelines for the management of atrial fibrillation: an update of the 2010 ESC Guidelines for the management of atrial fibrillation. Developed with the special contribution of the European Heart Rhythm Association. European Heart Journal, 33, 2719–2747.CrossRefPubMed Camm, A. J., Lip, G. Y., De Caterina, R., Savelieva, I., Atar, D., Hohnloser, S. H., et al. (2012). ESC Committee for Practice Guidelines (CPG). 2012 focused update of the ESC Guidelines for the management of atrial fibrillation: an update of the 2010 ESC Guidelines for the management of atrial fibrillation. Developed with the special contribution of the European Heart Rhythm Association. European Heart Journal, 33, 2719–2747.CrossRefPubMed
2.
Zurück zum Zitat Calkins, H., Kuck, K. H., Cappato, R., Brugada, J., Camm, A. J., Chen, S. A., et al. (2012). 2012 HRS/EHRA/ECAS expert consensus statement on catheter and surgical ablation of atrial fibrillation: recommendations for patient selection, procedural techniques, patient management and follow-up, definitions, endpoints, and research trial design. Heart Rhythm, 9, 632–696.CrossRefPubMed Calkins, H., Kuck, K. H., Cappato, R., Brugada, J., Camm, A. J., Chen, S. A., et al. (2012). 2012 HRS/EHRA/ECAS expert consensus statement on catheter and surgical ablation of atrial fibrillation: recommendations for patient selection, procedural techniques, patient management and follow-up, definitions, endpoints, and research trial design. Heart Rhythm, 9, 632–696.CrossRefPubMed
3.
Zurück zum Zitat Di Biase, L., Paoletti Perini, A., Mohanty, P., Goldenberg, A. S., Grifoni, G., Santangeli, P., et al. (2014). Visual, Tactile and Contact Force Feedback: which one is more important for catheter ablation? Results from an in vitro experimental study. Heart Rhythm, 11(3), 506–513.CrossRefPubMed Di Biase, L., Paoletti Perini, A., Mohanty, P., Goldenberg, A. S., Grifoni, G., Santangeli, P., et al. (2014). Visual, Tactile and Contact Force Feedback: which one is more important for catheter ablation? Results from an in vitro experimental study. Heart Rhythm, 11(3), 506–513.CrossRefPubMed
4.
Zurück zum Zitat Kumar, S., Chan, M., Lee, J., Wong, M. C., Yudi, M., Morton, J. B., et al. (2014). Catheter-tissue contact force determines atrial electrogram characteristics before and lesion efficacy after antral pulmonary vein isolation in humans. Journal of Cardiovascular Electrophysiology, 25(2), 122–129.CrossRefPubMed Kumar, S., Chan, M., Lee, J., Wong, M. C., Yudi, M., Morton, J. B., et al. (2014). Catheter-tissue contact force determines atrial electrogram characteristics before and lesion efficacy after antral pulmonary vein isolation in humans. Journal of Cardiovascular Electrophysiology, 25(2), 122–129.CrossRefPubMed
5.
Zurück zum Zitat Olson, M. D., Phreaner, N., Schuller, J. L., Nguyen, D. T., Katz, D. F., Aleong, R. G., et al. (2013). Effect of catheter movement and contact during application of radiofrequency energy on ablation lesion characteristics. Journal of Interventional Cardiac Electrophysiology, 38(2), 123–129.CrossRefPubMed Olson, M. D., Phreaner, N., Schuller, J. L., Nguyen, D. T., Katz, D. F., Aleong, R. G., et al. (2013). Effect of catheter movement and contact during application of radiofrequency energy on ablation lesion characteristics. Journal of Interventional Cardiac Electrophysiology, 38(2), 123–129.CrossRefPubMed
6.
Zurück zum Zitat Haldar, S., Jarman, J. W., Panikker, S., Jones, D. G., Salukhe, T., Gupta, D., et al. (2013). Contact force sensing technology identifies sites of inadequate contact and reduces acute pulmonary vein reconnection: a prospective case control study. International Journal of Cardiology, 168, 1160–1166.CrossRefPubMed Haldar, S., Jarman, J. W., Panikker, S., Jones, D. G., Salukhe, T., Gupta, D., et al. (2013). Contact force sensing technology identifies sites of inadequate contact and reduces acute pulmonary vein reconnection: a prospective case control study. International Journal of Cardiology, 168, 1160–1166.CrossRefPubMed
7.
Zurück zum Zitat Martinek, M., Lemes, C., Sigmund, E., Derndorfer, M., Aichinger, J., Winter, S., et al. (2012). Clinical impact of an open-irrigated radiofrequency catheter with direct force measurement on atrial fibrillation ablation. Pacing and Clinical Electrophysiology, 35, 1312–1318.CrossRefPubMed Martinek, M., Lemes, C., Sigmund, E., Derndorfer, M., Aichinger, J., Winter, S., et al. (2012). Clinical impact of an open-irrigated radiofrequency catheter with direct force measurement on atrial fibrillation ablation. Pacing and Clinical Electrophysiology, 35, 1312–1318.CrossRefPubMed
8.
Zurück zum Zitat Okumura, Y., Johnson, S. B., Bunch, T. J., Henz, B. D., O’Brien, C. J., & Packer, D. L. (2008). A systematical analysis of in vivo contact forces on virtual catheter tip/tissue surface contact during cardiac mapping and intervention. Journal of Cardiovascular Electrophysiology, 19, 632–640.CrossRefPubMed Okumura, Y., Johnson, S. B., Bunch, T. J., Henz, B. D., O’Brien, C. J., & Packer, D. L. (2008). A systematical analysis of in vivo contact forces on virtual catheter tip/tissue surface contact during cardiac mapping and intervention. Journal of Cardiovascular Electrophysiology, 19, 632–640.CrossRefPubMed
9.
Zurück zum Zitat Bai, R., DI Biase, L., Valderrabano, M., Lorgat, F., Micochova, H., Tilz, R., et al. (2012). Worldwide experience with the robotic navigation system in catheter ablation of atrial fibrillation: methodology, efficacy and safety. Journal of Cardiovascular Electrophysiology, 23, 820–826.CrossRefPubMed Bai, R., DI Biase, L., Valderrabano, M., Lorgat, F., Micochova, H., Tilz, R., et al. (2012). Worldwide experience with the robotic navigation system in catheter ablation of atrial fibrillation: methodology, efficacy and safety. Journal of Cardiovascular Electrophysiology, 23, 820–826.CrossRefPubMed
10.
Zurück zum Zitat Duncan, E. R., Finlay, M., Page, S. P., Hunter, R., Goromonzi, F., Richmond, L., et al. (2012). Improved electrogram attenuation during ablation of paroxysmal atrial fibrillation with the Hansen robotic system. Pacing and Clinical Electrophysiology, 35(6), 730–738.CrossRefPubMed Duncan, E. R., Finlay, M., Page, S. P., Hunter, R., Goromonzi, F., Richmond, L., et al. (2012). Improved electrogram attenuation during ablation of paroxysmal atrial fibrillation with the Hansen robotic system. Pacing and Clinical Electrophysiology, 35(6), 730–738.CrossRefPubMed
11.
Zurück zum Zitat Lorgat, F., Pudney, E., van Deventer, H., & Chitsaz, S. (2012). Robotically controlled ablation for atrial fibrillation: the first real-world experience in Africa with the Hansen robotic system. Cardiovascular Journal of Africa, 23(5), 274–280.CrossRefPubMedPubMedCentral Lorgat, F., Pudney, E., van Deventer, H., & Chitsaz, S. (2012). Robotically controlled ablation for atrial fibrillation: the first real-world experience in Africa with the Hansen robotic system. Cardiovascular Journal of Africa, 23(5), 274–280.CrossRefPubMedPubMedCentral
12.
Zurück zum Zitat Willems, S., Steven, D., Servatius, H., Hoffmann, B. A., Drewitz, I., Müllerleile, K., et al. (2010). Persistence of pulmonary vein isolation after robotic remote-navigated ablation for atrial fibrillation and its relation to clinical outcome. Journal of Cardiovascular Electrophysiology, 21, 1079–1084.CrossRefPubMed Willems, S., Steven, D., Servatius, H., Hoffmann, B. A., Drewitz, I., Müllerleile, K., et al. (2010). Persistence of pulmonary vein isolation after robotic remote-navigated ablation for atrial fibrillation and its relation to clinical outcome. Journal of Cardiovascular Electrophysiology, 21, 1079–1084.CrossRefPubMed
13.
Zurück zum Zitat Di Biase, L., Wang, Y., Horton, R., Gallinghouse, G. J., Mohanty, P., Sanchez, J., et al. (2009). Ablation of atrial fibrillation utilizing robotic catheter navigation in comparison to manual navigation and ablation: single-center experience. Journal of Cardiovascular Electrophysiology, 20, 1328–1335.CrossRefPubMed Di Biase, L., Wang, Y., Horton, R., Gallinghouse, G. J., Mohanty, P., Sanchez, J., et al. (2009). Ablation of atrial fibrillation utilizing robotic catheter navigation in comparison to manual navigation and ablation: single-center experience. Journal of Cardiovascular Electrophysiology, 20, 1328–1335.CrossRefPubMed
14.
Zurück zum Zitat Wazni, O. M., Barrett, C., Martin, D. O., Shaheen, N., Tarakji, K., Baranowski, B., et al. (2009). Experience with the hansen robotic system for atrial fibrillation ablation-lessons learned and techniques modified: Hansen in the real world. Journal of Cardiovascular Electrophysiology, 20(11), 1193–1196.CrossRefPubMed Wazni, O. M., Barrett, C., Martin, D. O., Shaheen, N., Tarakji, K., Baranowski, B., et al. (2009). Experience with the hansen robotic system for atrial fibrillation ablation-lessons learned and techniques modified: Hansen in the real world. Journal of Cardiovascular Electrophysiology, 20(11), 1193–1196.CrossRefPubMed
15.
Zurück zum Zitat Di Biase, L., Natale, A., Barrett, C., Tan, C., Elayi, C. S., Ching, C. K., et al. (2009). Relationship between catheter forces, lesion characteristics, “popping”, and char formation: experience with Robotic Navigation System. Journal of Cardiovascular Electrophysiology, 20, 436–440.CrossRefPubMed Di Biase, L., Natale, A., Barrett, C., Tan, C., Elayi, C. S., Ching, C. K., et al. (2009). Relationship between catheter forces, lesion characteristics, “popping”, and char formation: experience with Robotic Navigation System. Journal of Cardiovascular Electrophysiology, 20, 436–440.CrossRefPubMed
16.
Zurück zum Zitat Society, B. H. R. (2014). Standard for Interventional Electrophysiology Study and Catheter Ablation in Adults. Society, B. H. R. (2014). Standard for Interventional Electrophysiology Study and Catheter Ablation in Adults.
17.
Zurück zum Zitat Kuck, K. H., Reddy, V. Y., Schmidt, B., Natale, A., Neuzil, P., Saoudi, N., et al. (2012). A novel radiofrequency ablation catheter using contact force sensing: Toccata study. Heart Rhythm, 9(1), 18–23.CrossRefPubMed Kuck, K. H., Reddy, V. Y., Schmidt, B., Natale, A., Neuzil, P., Saoudi, N., et al. (2012). A novel radiofrequency ablation catheter using contact force sensing: Toccata study. Heart Rhythm, 9(1), 18–23.CrossRefPubMed
18.
Zurück zum Zitat Reddy, V. Y., Shah, D. C., Kautzenr, J., Schmidt, B., Saoudi, N., Herrera, C., et al. (2012). The relationship between contact force and clinical outcome during radiofrequency catheter ablation of atrial fibrillation in the TOCCATA study. Heart Rhythm, 9(11), 1789–1795.CrossRefPubMed Reddy, V. Y., Shah, D. C., Kautzenr, J., Schmidt, B., Saoudi, N., Herrera, C., et al. (2012). The relationship between contact force and clinical outcome during radiofrequency catheter ablation of atrial fibrillation in the TOCCATA study. Heart Rhythm, 9(11), 1789–1795.CrossRefPubMed
19.
Zurück zum Zitat Mizuno, H., Vergara, P., Maccabelli, G., Trevisi, N., Eng, S. C., Brombin, C., et al. (2013). Contact force monitoring for cardiac mapping in patients with ventricular tachycardia. Journal of Cardiovascular Electrophysiology, 24, 519–524.CrossRefPubMed Mizuno, H., Vergara, P., Maccabelli, G., Trevisi, N., Eng, S. C., Brombin, C., et al. (2013). Contact force monitoring for cardiac mapping in patients with ventricular tachycardia. Journal of Cardiovascular Electrophysiology, 24, 519–524.CrossRefPubMed
20.
Zurück zum Zitat Dello Russo, A., Fassini, G., Casella, M., Bologna, F., Al-Nono, O., Colombo, D., et al. (2014). Simultaneous assessment of contact pressure and local electrical coupling index using robotic navigation. Journal of Interventional Cardiac Electrophysiology, 40, 23–31.CrossRefPubMed Dello Russo, A., Fassini, G., Casella, M., Bologna, F., Al-Nono, O., Colombo, D., et al. (2014). Simultaneous assessment of contact pressure and local electrical coupling index using robotic navigation. Journal of Interventional Cardiac Electrophysiology, 40, 23–31.CrossRefPubMed
21.
Zurück zum Zitat Nakagawa, H., Kautzner, J., Natale, A., Peichl, P., Cihak, R., Wichterle, D., et al. (2013). Locations of high contact force during left atrial mapping in atrial fibrillation patients: electrogram amplitude and impedance are poor predictors of electrode-tissue contact force for ablation of atrial fibrillation. Circulation. Arrhythmia and Electrophysiology, 6(4), 746–753.CrossRefPubMed Nakagawa, H., Kautzner, J., Natale, A., Peichl, P., Cihak, R., Wichterle, D., et al. (2013). Locations of high contact force during left atrial mapping in atrial fibrillation patients: electrogram amplitude and impedance are poor predictors of electrode-tissue contact force for ablation of atrial fibrillation. Circulation. Arrhythmia and Electrophysiology, 6(4), 746–753.CrossRefPubMed
22.
Zurück zum Zitat Steven, D., Servatius, H., Rostock, T., Hoffman, B., Drewitz, I., Mullerleile, K., et al. (2010). Reduced fluoroscopy during atrial fibrillation ablation: benefits of robotic guided navigation. Journal of Cardiovascular Electrophysiology, 21, 6–12.CrossRefPubMed Steven, D., Servatius, H., Rostock, T., Hoffman, B., Drewitz, I., Mullerleile, K., et al. (2010). Reduced fluoroscopy during atrial fibrillation ablation: benefits of robotic guided navigation. Journal of Cardiovascular Electrophysiology, 21, 6–12.CrossRefPubMed
23.
Zurück zum Zitat Ullah, W., McLean, A., Hunter, R. J., Baker, V., Richmond, L., Cantor, E. J., et al. (2014). Randomized trial comparing robotic to manual ablation for atrial fibrillation. Heart Rhythm, 11, 1862–1869.CrossRefPubMed Ullah, W., McLean, A., Hunter, R. J., Baker, V., Richmond, L., Cantor, E. J., et al. (2014). Randomized trial comparing robotic to manual ablation for atrial fibrillation. Heart Rhythm, 11, 1862–1869.CrossRefPubMed
24.
Zurück zum Zitat Ullah, W., Hunter, R. J., Haldar, S., McLean, A., Dhinoja, M., Sporton, S., et al. (2014). Comparison of robotic and manual persistent AF ablation using catheter contact force sensing: an international multicenter registry study. PACE, 37, 1427–1435.CrossRefPubMed Ullah, W., Hunter, R. J., Haldar, S., McLean, A., Dhinoja, M., Sporton, S., et al. (2014). Comparison of robotic and manual persistent AF ablation using catheter contact force sensing: an international multicenter registry study. PACE, 37, 1427–1435.CrossRefPubMed
25.
Zurück zum Zitat Casella, M., Pelargonio, G., Dello Russo, A., Riva, S., Bartoletti, S., Santangeli, P., et al. (2011). "Near-zero" fluoroscopic exposure in supraventricular arrhythmia ablation using the EnSite NavX™ mapping system: personal experience and review of the literature. Journal of Interventional Cardiac Electrophysiology, 31, 109–118.CrossRefPubMed Casella, M., Pelargonio, G., Dello Russo, A., Riva, S., Bartoletti, S., Santangeli, P., et al. (2011). "Near-zero" fluoroscopic exposure in supraventricular arrhythmia ablation using the EnSite NavX™ mapping system: personal experience and review of the literature. Journal of Interventional Cardiac Electrophysiology, 31, 109–118.CrossRefPubMed
26.
Zurück zum Zitat Ait-Ali, L., Andreassi, M. G., Foffa, I., Spadoni, I., Vano, E., & Picano, E. (2010). Cumulative patient effective dose and acute radiation-induced chromosomal DNA damage in children with congenital heart disease. Heart, 96, 269–274.CrossRefPubMed Ait-Ali, L., Andreassi, M. G., Foffa, I., Spadoni, I., Vano, E., & Picano, E. (2010). Cumulative patient effective dose and acute radiation-induced chromosomal DNA damage in children with congenital heart disease. Heart, 96, 269–274.CrossRefPubMed
Metadaten
Titel
Analysis of catheter contact force during atrial fibrillation ablation using the robotic navigation system: results from a randomized study
verfasst von
Antonio Dello Russo
Gaetano Fassini
Sergio Conti
Michela Casella
Antonio Di Monaco
Eleonora Russo
Stefania Riva
Massimo Moltrasio
Fabrizio Tundo
Giuseppe De Martino
G. Joseph Gallinghouse
Luigi Di Biase
Andrea Natale
Claudio Tondo
Publikationsdatum
21.01.2016
Verlag
Springer US
Erschienen in
Journal of Interventional Cardiac Electrophysiology / Ausgabe 2/2016
Print ISSN: 1383-875X
Elektronische ISSN: 1572-8595
DOI
https://doi.org/10.1007/s10840-016-0102-0

Weitere Artikel der Ausgabe 2/2016

Journal of Interventional Cardiac Electrophysiology 2/2016 Zur Ausgabe

Update Kardiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.