Skip to main content
Erschienen in: Journal of Clinical Monitoring and Computing 2/2016

01.04.2016 | Original Research

Endovascular coil detachment causing EMG artefact in BIS: a mechanistic exploration

verfasst von: Dhritiman Chakrabarti, Venkatapura J. Ramesh, Hima Pendharkar

Erschienen in: Journal of Clinical Monitoring and Computing | Ausgabe 2/2016

Einloggen, um Zugang zu erhalten

Abstract

Deployment of endovascular coils used in interventional neuroradiology commonly involves electrolytic detachment of the coil from the pusher catheter. This report describes a case of artefactual increase in electromyography (EMG) values of bispectral index (BIS) monitor during coil detachment. An explanation of this event is provided connecting mechanism of coil detachment and derivation of EMG values in a BIS monitor. While rising EMG values are thought to arise from frontalis contraction, they may as well be an unrecognized electrical artefact, especially in context of undistorted electroencephalography waveform.
Literatur
1.
Zurück zum Zitat Scheldrup RW, Schaller LB (inventors). Endovascular embolic device detachment detection method and apparatus. United States Patent US5669905 A. Target Therapeutics, Inc.; 1997. Scheldrup RW, Schaller LB (inventors). Endovascular embolic device detachment detection method and apparatus. United States Patent US5669905 A. Target Therapeutics, Inc.; 1997.
2.
Zurück zum Zitat Rampil IJ. A primer for EEG signal processing in anesthesia. Anesthesiology. 1998;89:980–1002.CrossRefPubMed Rampil IJ. A primer for EEG signal processing in anesthesia. Anesthesiology. 1998;89:980–1002.CrossRefPubMed
3.
Zurück zum Zitat Goncharova McFarland DJ, Vaughan TM, Wolpaw JR. EMG contamination of EEG: spectral and topographical characteristics. Clin Neurophysiol. 2003;114:1580–93.CrossRefPubMed Goncharova McFarland DJ, Vaughan TM, Wolpaw JR. EMG contamination of EEG: spectral and topographical characteristics. Clin Neurophysiol. 2003;114:1580–93.CrossRefPubMed
4.
Zurück zum Zitat Schnakers C, Ledoux D, Majerus S, Damas P, Damas F, Lambermont B. Diagnostic and prognostic use of bispectral index in coma, vegetative state and related disorders. Brain Inj. 2008;22:926–31.CrossRefPubMed Schnakers C, Ledoux D, Majerus S, Damas P, Damas F, Lambermont B. Diagnostic and prognostic use of bispectral index in coma, vegetative state and related disorders. Brain Inj. 2008;22:926–31.CrossRefPubMed
Metadaten
Titel
Endovascular coil detachment causing EMG artefact in BIS: a mechanistic exploration
verfasst von
Dhritiman Chakrabarti
Venkatapura J. Ramesh
Hima Pendharkar
Publikationsdatum
01.04.2016
Verlag
Springer Netherlands
Erschienen in
Journal of Clinical Monitoring and Computing / Ausgabe 2/2016
Print ISSN: 1387-1307
Elektronische ISSN: 1573-2614
DOI
https://doi.org/10.1007/s10877-015-9702-y

Weitere Artikel der Ausgabe 2/2016

Journal of Clinical Monitoring and Computing 2/2016 Zur Ausgabe

Update AINS

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.