Erschienen in:
01.04.2016 | Original Research
Endovascular coil detachment causing EMG artefact in BIS: a mechanistic exploration
verfasst von:
Dhritiman Chakrabarti, Venkatapura J. Ramesh, Hima Pendharkar
Erschienen in:
Journal of Clinical Monitoring and Computing
|
Ausgabe 2/2016
Einloggen, um Zugang zu erhalten
Abstract
Deployment of endovascular coils used in interventional neuroradiology commonly involves electrolytic detachment of the coil from the pusher catheter. This report describes a case of artefactual increase in electromyography (EMG) values of bispectral index (BIS) monitor during coil detachment. An explanation of this event is provided connecting mechanism of coil detachment and derivation of EMG values in a BIS monitor. While rising EMG values are thought to arise from frontalis contraction, they may as well be an unrecognized electrical artefact, especially in context of undistorted electroencephalography waveform.