Evaluation of cardiac output variations with the peripheral pulse pressure to mean arterial pressure ratio

verfasst von: Audrey Tantot, Anais Caillard, Arthur Le Gall, Joaquim Mateo, Sandrine Millasseau, Alexandre Mebazaa, Etienne Gayat, Fabrice Vallée

Erschienen in: Journal of Clinical Monitoring and Computing | Ausgabe 4/2019

Einloggen, um Zugang zu erhalten

Abstract

Cardiac output (CO) optimisation during surgery reduces post-operative morbidity. Various methods based on pulse pressure analysis have been developed to overcome difficulties to measure accurate CO variations in standard anaesthetic settings. Several of these methods include, among other parameters, the ratio of pulse pressure to mean arterial pressure (PP/MAP). The aim of this study was to evaluate whether the ratio of radial pulse pressure to mean arterial pressure (ΔPPrad/MAP) could track CO variations (ΔCO) induced by various therapeutic interventions such as fluid infusions and vasopressors boluses [phenylephrine (PE), norepinephrine (NA) or ephedrine (EP)] in the operating room. Trans-oesophageal Doppler signal and pressure waveforms were recorded in patients undergoing neurosurgery. CO and PPrad/MAP were recorded before and after fluid challenges, PE, NA and EP bolus infusions as medically required during their anaesthesia. One hundred and three patients (mean age: 52 ± 12 years old, 38 men) have been included with a total of 636 sets of measurement. During fluids challenges (n = 188), a positive correlation was found between ΔPPrad/MAP and ΔCO (r = 0.22, p = 0.003). After PE (n = 256) and NA (n = 121) boluses, ΔPPrad/MAP positively tracked ΔCO (r = 0.53 and 0.41 respectively, p < 0.001). By contrast, there was no relation between ΔPPrad/MAP and ΔCO after EP boluses (r = 0.10, p = 0.39). ΔPPrad/MAP tracked ΔCO variations during PE and NA vasopressor challenges. However, after positive fluid challenge or EP boluses, ΔPPrad/MAP was not as performant to track ΔCO which could make the use of this ratio difficult in current clinical practice.
Literatur
1.
Zurück zum Zitat Hamilton MA, Cecconi M, Rhodes A. A systematic review and meta-analysis on the use of preemptive hemodynamic intervention to improve postoperative outcomes in moderate and high-risk surgical patients. Anesth Analg. 2011;112:1392–402.CrossRefPubMed Hamilton MA, Cecconi M, Rhodes A. A systematic review and meta-analysis on the use of preemptive hemodynamic intervention to improve postoperative outcomes in moderate and high-risk surgical patients. Anesth Analg. 2011;112:1392–402.CrossRefPubMed
2.
Zurück zum Zitat Cecconi M, Rhodes A, Poloniecki J, Della RG, Grounds RM. Bench-to-bedside review: the importance of the precision of the reference technique in method comparison studies–with specific reference to the measurement of cardiac output. Crit Care. 2009;13:201.CrossRefPubMedPubMedCentral Cecconi M, Rhodes A, Poloniecki J, Della RG, Grounds RM. Bench-to-bedside review: the importance of the precision of the reference technique in method comparison studies–with specific reference to the measurement of cardiac output. Crit Care. 2009;13:201.CrossRefPubMedPubMedCentral
3.
Zurück zum Zitat Pearse RM, Harrison DA, MacDonald N, Gillies MA, Blunt M, Ackland G, et al. Effect of a perioperative, cardiac output-guided hemodynamic therapy algorithm on outcomes following major gastrointestinal surgery: a randomized clinical trial and systematic review. JAMA 2014;311:2181.CrossRefPubMed Pearse RM, Harrison DA, MacDonald N, Gillies MA, Blunt M, Ackland G, et al. Effect of a perioperative, cardiac output-guided hemodynamic therapy algorithm on outcomes following major gastrointestinal surgery: a randomized clinical trial and systematic review. JAMA 2014;311:2181.CrossRefPubMed
4.
Zurück zum Zitat Westerhof N, Elzinga G. Normalized input impedance and arterial decay time over heart period are independent of animal size. Am J Physiol. 1991;261:R126–33.PubMed Westerhof N, Elzinga G. Normalized input impedance and arterial decay time over heart period are independent of animal size. Am J Physiol. 1991;261:R126–33.PubMed
5.
7.
Zurück zum Zitat Metzelder S, Coburn M, Fries M, Reinges M, Reich S, Rossaint R, et al. Performance of cardiac output measurement derived from arterial pressure waveform analysis in patients requiring high-dose vasopressor therapy. Br J Anaesth. 2011;106:776–84.CrossRefPubMed Metzelder S, Coburn M, Fries M, Reinges M, Reich S, Rossaint R, et al. Performance of cardiac output measurement derived from arterial pressure waveform analysis in patients requiring high-dose vasopressor therapy. Br J Anaesth. 2011;106:776–84.CrossRefPubMed
8.
Zurück zum Zitat Gruenewald M, Meybohm P, Renner J, Broch O, Caliebe A, Weiler N, et al. Effect of norepinephrine dosage and calibration frequency on accuracy of pulse contour-derived cardiac output. Crit Care. 2011;15:R22.CrossRefPubMedPubMedCentral Gruenewald M, Meybohm P, Renner J, Broch O, Caliebe A, Weiler N, et al. Effect of norepinephrine dosage and calibration frequency on accuracy of pulse contour-derived cardiac output. Crit Care. 2011;15:R22.CrossRefPubMedPubMedCentral
9.
Zurück zum Zitat Bein B, Meybohm P, Cavus E, Renner J, Tonner PH, Steinfath M, et al. The reliability of pulse contour-derived cardiac output during hemorrhage and after vasopressor administration. Anesth Analg. 2007;105:107–13.CrossRefPubMed Bein B, Meybohm P, Cavus E, Renner J, Tonner PH, Steinfath M, et al. The reliability of pulse contour-derived cardiac output during hemorrhage and after vasopressor administration. Anesth Analg. 2007;105:107–13.CrossRefPubMed
10.
Zurück zum Zitat Monge García MI, Romero MG, Cano AG, Rhodes A, Grounds RM, Cecconi M. Impact of arterial load on the agreement between pulse pressure analysis and esophageal Doppler. Crit Care. 2013;17:R113.CrossRefPubMedPubMedCentral Monge García MI, Romero MG, Cano AG, Rhodes A, Grounds RM, Cecconi M. Impact of arterial load on the agreement between pulse pressure analysis and esophageal Doppler. Crit Care. 2013;17:R113.CrossRefPubMedPubMedCentral
11.
Zurück zum Zitat Sun J. Cardiac output estimation using arterial blood pressure waveforms. Cambridge: Massachusetts Institute of Technology; 2006. Sun J. Cardiac output estimation using arterial blood pressure waveforms. Cambridge: Massachusetts Institute of Technology; 2006.
12.
Zurück zum Zitat Caillard A, Gayat E, Tantot A, Dubreuil G, M’Bakulu E, Madadaki C, et al. Comparison of cardiac output measured by oesophageal Doppler ultrasonography or pulse pressure contour wave analysis. Br J Anaesth. 2015;114:893–900.CrossRefPubMed Caillard A, Gayat E, Tantot A, Dubreuil G, M’Bakulu E, Madadaki C, et al. Comparison of cardiac output measured by oesophageal Doppler ultrasonography or pulse pressure contour wave analysis. Br J Anaesth. 2015;114:893–900.CrossRefPubMed
13.
Zurück zum Zitat Critchley LA, Lee A, Ho AM. A critical review of the ability of continuous cardiac output monitors to measure trends in cardiac output. Anesth Analg. 2010;111:1180–92.CrossRefPubMed Critchley LA, Lee A, Ho AM. A critical review of the ability of continuous cardiac output monitors to measure trends in cardiac output. Anesth Analg. 2010;111:1180–92.CrossRefPubMed
14.
Zurück zum Zitat Meng L, Tran NP, Alexander BS, Laning K, Chen G, Kain ZN, et al. The impact of phenylephrine, ephedrine, and increased preload on third-generation Vigileo-FloTrac and esophageal doppler cardiac output measurements. Anesth Analg. 2011;113:751–7.CrossRefPubMed Meng L, Tran NP, Alexander BS, Laning K, Chen G, Kain ZN, et al. The impact of phenylephrine, ephedrine, and increased preload on third-generation Vigileo-FloTrac and esophageal doppler cardiac output measurements. Anesth Analg. 2011;113:751–7.CrossRefPubMed
15.
Zurück zum Zitat Adji A, O’Rourke MF. Brachial artery tonometry and the Popeye phenomenon: explanation of anomalies in generating central from upper limb pressure waveforms. J Hypertens. 2012;30:1540–51.CrossRefPubMed Adji A, O’Rourke MF. Brachial artery tonometry and the Popeye phenomenon: explanation of anomalies in generating central from upper limb pressure waveforms. J Hypertens. 2012;30:1540–51.CrossRefPubMed
16.
Zurück zum Zitat O’Rourke MF, Adji A. An updated clinical primer on large artery mechanics: implications of pulse waveform analysis and arterial tonometry. Curr Opin Cardiol. 2005;20:275–81.CrossRefPubMed O’Rourke MF, Adji A. An updated clinical primer on large artery mechanics: implications of pulse waveform analysis and arterial tonometry. Curr Opin Cardiol. 2005;20:275–81.CrossRefPubMed
17.
Zurück zum Zitat Learoyd BM, Taylor MG. Alterations with age in the viscoelastic properties of human arterial walls. Circ Res. 1966;18:278–92.CrossRefPubMed Learoyd BM, Taylor MG. Alterations with age in the viscoelastic properties of human arterial walls. Circ Res. 1966;18:278–92.CrossRefPubMed
18.
Zurück zum Zitat Monnet X, Chemla D, Osman D, Anguel N, Richard C, Pinsky MR, Teboul JL. Measuring aortic diameter improves accuracy of oesophageal Doppler in assessing fluid responsiveness. Crit Care Med. 2007;35(2):477–82.CrossRefPubMed Monnet X, Chemla D, Osman D, Anguel N, Richard C, Pinsky MR, Teboul JL. Measuring aortic diameter improves accuracy of oesophageal Doppler in assessing fluid responsiveness. Crit Care Med. 2007;35(2):477–82.CrossRefPubMed
Metadaten
Titel
Evaluation of cardiac output variations with the peripheral pulse pressure to mean arterial pressure ratio
verfasst von
Audrey Tantot
Anais Caillard
Arthur Le Gall
Joaquim Mateo
Sandrine Millasseau
Alexandre Mebazaa
Etienne Gayat
Fabrice Vallée
Publikationsdatum
25.10.2018
Verlag
Springer Netherlands
Erschienen in
Journal of Clinical Monitoring and Computing / Ausgabe 4/2019
Print ISSN: 1387-1307
Elektronische ISSN: 1573-2614
DOI
https://doi.org/10.1007/s10877-018-0210-8

Weitere Artikel der Ausgabe 4/2019

Journal of Clinical Monitoring and Computing 4/2019 Zur Ausgabe

Update AINS

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.