Validity of linear measurements of the jaws using ultralow-dose MDCT and the iterative techniques of ASIR and MBIR

verfasst von: Asma’a A. Al-Ekrish, Reema Al-Shawaf, Peter Schullian, Ra’ed Al-Sadhan, Romed Hörmann, Gerlig Widmann

Erschienen in: International Journal of Computer Assisted Radiology and Surgery | Ausgabe 10/2016

Einloggen, um Zugang zu erhalten

Abstract

Purpose

To assess the comparability of linear measurements of dental implant sites recorded from multidetector computed tomography (MDCT) images obtained using standard-dose filtered backprojection (FBP) technique with those from various ultralow doses combined with FBP, adaptive statistical iterative reconstruction (ASIR), and model-based iterative reconstruction (MBIR) techniques. The results of the study may contribute to MDCT dose optimization for dental implant site imaging.

Methods

MDCT scans of two cadavers were acquired using a standard reference protocol and four ultralow-dose test protocols (TP). The volume CT dose index of the different dose protocols ranged from a maximum of 30.48–36.71 mGy to a minimum of 0.44–0.53 mGy. All scans were reconstructed using FBP, ASIR-50, ASIR-100, and MBIR, and either a bone or standard reconstruction kernel. Linear measurements were recorded from standardized images of the jaws by two examiners. Intra- and inter-examiner reliability of the measurements were analyzed using Cronbach’s alpha and inter-item correlation. Agreement between the measurements obtained with the reference-dose/FBP protocol and each of the test protocols was determined with Bland–Altman plots and linear regression. Statistical significance was set at a P-value of 0.05.

Results

No systematic variation was found between the linear measurements obtained with the reference protocol and the other imaging protocols. The only exceptions were TP3/ASIR-50 (bone kernel) and TP4/ASIR-100 (bone and standard kernels). The mean measurement differences between these three protocols and the reference protocol were within ±0.1 mm, with the 95 % confidence interval limits being within the range of ±1.15 mm.

Conclusions

A nearly 97.5 % reduction in dose did not significantly affect the height and width measurements of edentulous jaws regardless of the reconstruction algorithm used.
Anhänge
Nur mit Berechtigung zugänglich
Literatur
1.
Zurück zum Zitat United Nations Scientific Committee on the Effects of Atomic Radiation Sources and Effects of Ionizing Radiation (2010) In: Official Records of the General Assembly, Sixty-third Session, Supplement No. 46., New York, (2010) 2008. United Nations United Nations Scientific Committee on the Effects of Atomic Radiation Sources and Effects of Ionizing Radiation (2010) In: Official Records of the General Assembly, Sixty-third Session, Supplement No. 46., New York, (2010) 2008. United Nations
2.
Zurück zum Zitat Loubele M, Bogaerts R, Van Dijck E, Pauwels R, Vanheusden S, Suetens P, Marchal G, Sanderink G, Jacobs R (2009) Comparison between effective radiation dose of CBCT and MSCT scanners for dentomaxillofacial applications. Eur J Radiol 71(3):461–468CrossRefPubMed Loubele M, Bogaerts R, Van Dijck E, Pauwels R, Vanheusden S, Suetens P, Marchal G, Sanderink G, Jacobs R (2009) Comparison between effective radiation dose of CBCT and MSCT scanners for dentomaxillofacial applications. Eur J Radiol 71(3):461–468CrossRefPubMed
4.
Zurück zum Zitat Suomalainen A, Kiljunen T, Kaser Y, Peltola J, Kortesniemi M (2009) Dosimetry and image quality of four dental cone beam computed tomography scanners compared with multislice computed tomography scanners. Dentomaxillofac Radiol 38(6):367–378. doi:10.1259/dmfr/15779208 CrossRefPubMed Suomalainen A, Kiljunen T, Kaser Y, Peltola J, Kortesniemi M (2009) Dosimetry and image quality of four dental cone beam computed tomography scanners compared with multislice computed tomography scanners. Dentomaxillofac Radiol 38(6):367–378. doi:10.​1259/​dmfr/​15779208 CrossRefPubMed
6.
Zurück zum Zitat Tyndall DA, Price JB, Tetradis S, Ganz SD, Hildebolt C, Scarfe WC (2012) Position statement of the American Academy of Oral and Maxillofacial Radiology on selection criteria for the use of radiology in dental implantology with emphasis on cone beam computed tomography. Oral Surg Oral Med Oral Pathol Oral Radiol 113(6):817–826CrossRefPubMed Tyndall DA, Price JB, Tetradis S, Ganz SD, Hildebolt C, Scarfe WC (2012) Position statement of the American Academy of Oral and Maxillofacial Radiology on selection criteria for the use of radiology in dental implantology with emphasis on cone beam computed tomography. Oral Surg Oral Med Oral Pathol Oral Radiol 113(6):817–826CrossRefPubMed
7.
Zurück zum Zitat Al-Ekrish A, Ekram M (2011) A comparative study of the accuracy and reliability of multidetector computed tomography and cone beam computed tomography in the assessment of dental implant site dimensions. Dentomaxillofac Radiol 40(2):67–75CrossRefPubMedPubMedCentral Al-Ekrish A, Ekram M (2011) A comparative study of the accuracy and reliability of multidetector computed tomography and cone beam computed tomography in the assessment of dental implant site dimensions. Dentomaxillofac Radiol 40(2):67–75CrossRefPubMedPubMedCentral
8.
Zurück zum Zitat Suomalainen A, Vehmas T, Kortesniemi M, Robinson S, Peltola J (2008) Accuracy of linear measurements using dental cone beam and conventional multislice computed tomography. Dentomaxillofac Radiol 37(1):10–17. doi:10.1259/dmfr/14140281 CrossRefPubMed Suomalainen A, Vehmas T, Kortesniemi M, Robinson S, Peltola J (2008) Accuracy of linear measurements using dental cone beam and conventional multislice computed tomography. Dentomaxillofac Radiol 37(1):10–17. doi:10.​1259/​dmfr/​14140281 CrossRefPubMed
9.
Zurück zum Zitat Rustemeyer P, Streubühr U, Suttmoeller J (2004) Low-dose dental computed tomography: significant dose reduction without loss of image quality. Acta Radiol 45:847–853CrossRefPubMed Rustemeyer P, Streubühr U, Suttmoeller J (2004) Low-dose dental computed tomography: significant dose reduction without loss of image quality. Acta Radiol 45:847–853CrossRefPubMed
10.
Zurück zum Zitat Loubele M, Jacobs R, Maes F, Schutyser F, Debaveye D, Bogaerts R, Coudyzer W, Vandermeulen D, Van Cleynenbreugel J, Marchal G (2005) Radiation dose vs. image quality for low-dose CT protocols of the head for maxillofacial surgery and oral implant planning. Radiat Protect Dosim 117(1–3):211–216 Loubele M, Jacobs R, Maes F, Schutyser F, Debaveye D, Bogaerts R, Coudyzer W, Vandermeulen D, Van Cleynenbreugel J, Marchal G (2005) Radiation dose vs. image quality for low-dose CT protocols of the head for maxillofacial surgery and oral implant planning. Radiat Protect Dosim 117(1–3):211–216
11.
Zurück zum Zitat Rustemeyer P, Streubühr U, Suttmoeller J (2004) Low-dose dental computed tomography: significant dose reduction without loss of image quality. Acta Radiol 45(8):847–853CrossRefPubMed Rustemeyer P, Streubühr U, Suttmoeller J (2004) Low-dose dental computed tomography: significant dose reduction without loss of image quality. Acta Radiol 45(8):847–853CrossRefPubMed
12.
Zurück zum Zitat Silva AC, Lawder HJ, Hara A, Kujak J, Pavlicek W (2010) Innovations in CT dose reduction strategy: application of the adaptive statistical iterative reconstruction algorithm. Am J Roentgenol 194(1):191–199. doi:10.2214/ajr.09.2953 CrossRef Silva AC, Lawder HJ, Hara A, Kujak J, Pavlicek W (2010) Innovations in CT dose reduction strategy: application of the adaptive statistical iterative reconstruction algorithm. Am J Roentgenol 194(1):191–199. doi:10.​2214/​ajr.​09.​2953 CrossRef
13.
Zurück zum Zitat Widmann G, Dalla Torre D, Hoermann R, Schullian P, Gassner EM, Bale R, Puelacher W (2015) Ultralow-dose computed tomography imaging for surgery of midfacial and orbital fractures using ASIR and MBIR. Int J Oral Maxillofac Surg 44(4):441–446. doi:10.1016/j.ijom.2015.01.011 CrossRefPubMed Widmann G, Dalla Torre D, Hoermann R, Schullian P, Gassner EM, Bale R, Puelacher W (2015) Ultralow-dose computed tomography imaging for surgery of midfacial and orbital fractures using ASIR and MBIR. Int J Oral Maxillofac Surg 44(4):441–446. doi:10.​1016/​j.​ijom.​2015.​01.​011 CrossRefPubMed
14.
Zurück zum Zitat McCollough CH, Leng S, Sunnegardh J, Vrieze TJ, Yu L, Lane J, Raupach R, Stierstorfer K, Flohr T (2013) Spatial resolution improvement and dose reduction potential for inner ear CT imaging using a z-axis deconvolution technique. Med Phys 40(6):061904. doi:10.1118/1.4802730 CrossRefPubMed McCollough CH, Leng S, Sunnegardh J, Vrieze TJ, Yu L, Lane J, Raupach R, Stierstorfer K, Flohr T (2013) Spatial resolution improvement and dose reduction potential for inner ear CT imaging using a z-axis deconvolution technique. Med Phys 40(6):061904. doi:10.​1118/​1.​4802730 CrossRefPubMed
15.
Zurück zum Zitat Niu YT, Mehta D, Zhang ZR, Zhang YX, Liu YF, Kang TL, Xian JF, Wang ZC (2012) Radiation dose reduction in temporal bone CT with iterative reconstruction technique. AJNR Am J Neuroradiol 33(6):1020–1026. doi:10.3174/ajnr.A2941 CrossRefPubMed Niu YT, Mehta D, Zhang ZR, Zhang YX, Liu YF, Kang TL, Xian JF, Wang ZC (2012) Radiation dose reduction in temporal bone CT with iterative reconstruction technique. AJNR Am J Neuroradiol 33(6):1020–1026. doi:10.​3174/​ajnr.​A2941 CrossRefPubMed
16.
Zurück zum Zitat Schulz B, Beeres M, Bodelle B, Bauer R, Al-Butmeh F, Thalhammer A, Vogl TJ, Kerl JM (2013) Performance of iterative image reconstruction in CT of the paranasal sinuses: a phantom study. AJNR Am J Neuroradiol 34(5):1072–1076. doi:10.3174/ajnr.A3339 CrossRefPubMed Schulz B, Beeres M, Bodelle B, Bauer R, Al-Butmeh F, Thalhammer A, Vogl TJ, Kerl JM (2013) Performance of iterative image reconstruction in CT of the paranasal sinuses: a phantom study. AJNR Am J Neuroradiol 34(5):1072–1076. doi:10.​3174/​ajnr.​A3339 CrossRefPubMed
17.
Zurück zum Zitat Widmann G, Schullian P, Gassner E-M, Hoermann R, Bale R, Puelacher W (2015) Ultralow-dose CT of the craniofacial bone for navigated surgery using adaptive statistical iterative reconstruction and model-based iterative reconstruction: 2D and 3D image quality. Am J Roentgenol 204(3):563–569CrossRef Widmann G, Schullian P, Gassner E-M, Hoermann R, Bale R, Puelacher W (2015) Ultralow-dose CT of the craniofacial bone for navigated surgery using adaptive statistical iterative reconstruction and model-based iterative reconstruction: 2D and 3D image quality. Am J Roentgenol 204(3):563–569CrossRef
18.
Zurück zum Zitat McCollough CH, Chen GH, Kalender W, Leng S, Samei E, Taguchi K, Wang G, Yu L, Pettigrew RI (2012) Achieving routine submillisievert CT scanning: report from the summit on management of radiation dose in CT. Radiology 264(2):567–580CrossRefPubMedPubMedCentral McCollough CH, Chen GH, Kalender W, Leng S, Samei E, Taguchi K, Wang G, Yu L, Pettigrew RI (2012) Achieving routine submillisievert CT scanning: report from the summit on management of radiation dose in CT. Radiology 264(2):567–580CrossRefPubMedPubMedCentral
19.
Zurück zum Zitat McHanwell S, Brenner E, Chirculescu A, Drukker J, van Mameren H, Mazzotti G, Pais D, Paulsen F, Plaisant O, Caillaud M, Laforêt E, Riederer B, Sanudo J, Bueno-Lopez J, Donate F, Sprumont P, Teofilovski-Parapid G, Moxham B (2008) The legal and ethical framework governing body donation in Europe—a review of current practice and recommendations for good practice. Eur J Anat 12(1):1–24 McHanwell S, Brenner E, Chirculescu A, Drukker J, van Mameren H, Mazzotti G, Pais D, Paulsen F, Plaisant O, Caillaud M, Laforêt E, Riederer B, Sanudo J, Bueno-Lopez J, Donate F, Sprumont P, Teofilovski-Parapid G, Moxham B (2008) The legal and ethical framework governing body donation in Europe—a review of current practice and recommendations for good practice. Eur J Anat 12(1):1–24
20.
Zurück zum Zitat Riederer B, Bolt S, Brenner E, Bueno-Lopez J, Cirulescu A, Davies D, De Caro R, Gerrits P, McHanwell S, Pais D, Paulsen F, Plaisant O, Sendemir E, Stabile I, Moxham B (2012) The legal and ethical framework governing body donation in Europe—1st update on current practice. Eur J Anat 16(1):1–21 Riederer B, Bolt S, Brenner E, Bueno-Lopez J, Cirulescu A, Davies D, De Caro R, Gerrits P, McHanwell S, Pais D, Paulsen F, Plaisant O, Sendemir E, Stabile I, Moxham B (2012) The legal and ethical framework governing body donation in Europe—1st update on current practice. Eur J Anat 16(1):1–21
21.
Zurück zum Zitat Platzer W, Putz R, Poisel S (1978) Ein neues Konservierungs- und Aufbewahrungssystem für anatomisches Material. Acta Anat (Basel) 102(1):60–67CrossRef Platzer W, Putz R, Poisel S (1978) Ein neues Konservierungs- und Aufbewahrungssystem für anatomisches Material. Acta Anat (Basel) 102(1):60–67CrossRef
22.
Zurück zum Zitat ICRP (2007) The 2007 Recommendations of the International Commission on Radiological Protection. ICRP Publication 103. Ann ICRP 37:2–4 ICRP (2007) The 2007 Recommendations of the International Commission on Radiological Protection. ICRP Publication 103. Ann ICRP 37:2–4
23.
Zurück zum Zitat Bland JM, Altman D (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 327(8476):307–310CrossRef Bland JM, Altman D (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 327(8476):307–310CrossRef
24.
Zurück zum Zitat Watanabe H, Abdul MM, Kurabayashi T, Aoki H (2010) Mandible size and morphology determined with CT on a premise of dental implant operation. Surg Radiol Anat 32(4):343–349CrossRefPubMed Watanabe H, Abdul MM, Kurabayashi T, Aoki H (2010) Mandible size and morphology determined with CT on a premise of dental implant operation. Surg Radiol Anat 32(4):343–349CrossRefPubMed
25.
Zurück zum Zitat Widmann G, Bischel A, Stratis A, Kakar A, Bosmans H, Jacobs R, Gassner E-M, Puelacher W, Pauwels R (2016) Ultralow dose dentomaxillofacial CT imaging and iterative reconstruction techniques: variability of Hounsfield Units and contrast-to-noise ratio. Br J Radiol 89(1060):20151055. doi:10.1259/bjr.20151055 Widmann G, Bischel A, Stratis A, Kakar A, Bosmans H, Jacobs R, Gassner E-M, Puelacher W, Pauwels R (2016) Ultralow dose dentomaxillofacial CT imaging and iterative reconstruction techniques: variability of Hounsfield Units and contrast-to-noise ratio. Br J Radiol 89(1060):20151055. doi:10.​1259/​bjr.​20151055
26.
Zurück zum Zitat Hara AK, Paden RG, Silva AC, Kujak JL, Lawder HJ, Pavlicek W (2009) Iterative reconstruction technique for reducing body radiation dose at CT: feasibility study. Am J Roentgenol 193(3):764–771CrossRef Hara AK, Paden RG, Silva AC, Kujak JL, Lawder HJ, Pavlicek W (2009) Iterative reconstruction technique for reducing body radiation dose at CT: feasibility study. Am J Roentgenol 193(3):764–771CrossRef
27.
Zurück zum Zitat Widmann G, Fasser M, Schullian P, Zangerl A, Puelacher W, Kral F, Riechelmann H, Jaschke W, Bale R (2012) Substantial dose reduction in modern multi-slice spiral computed tomography (MSCT)-guided craniofacial and skull base surgery. RoFo. Fortschritte auf dem Gebiete der Rontgenstrahlen und der Nuklearmedizin 184(2):136–142CrossRefPubMed Widmann G, Fasser M, Schullian P, Zangerl A, Puelacher W, Kral F, Riechelmann H, Jaschke W, Bale R (2012) Substantial dose reduction in modern multi-slice spiral computed tomography (MSCT)-guided craniofacial and skull base surgery. RoFo. Fortschritte auf dem Gebiete der Rontgenstrahlen und der Nuklearmedizin 184(2):136–142CrossRefPubMed
29.
Zurück zum Zitat GE Medical Systems (2010) Advantage Dentascan operator manual 2182400-100. vol revision 4. General Electric Company GE Medical Systems (2010) Advantage Dentascan operator manual 2182400-100. vol revision 4. General Electric Company
31.
Zurück zum Zitat Bornstein MM, Scarfe WC, Vaughn VM, Jacobs R (2014) Cone beam computed tomography in implant dentistry: a systematic review focusing on guidelines, indications, and radiation dose risks. Int J Oral Maxillofac Implants 29(Suppl):55–77CrossRefPubMed Bornstein MM, Scarfe WC, Vaughn VM, Jacobs R (2014) Cone beam computed tomography in implant dentistry: a systematic review focusing on guidelines, indications, and radiation dose risks. Int J Oral Maxillofac Implants 29(Suppl):55–77CrossRefPubMed
32.
Zurück zum Zitat Ludlow J, Timothy R, Walker C, Hunter R, Benavides E, Samuelson D, Scheske M (2015) Effective dose of dental CBCT—a meta analysis of published data and additional data for nine CBCT units. Dento Maxillofac Radiol 44(1):20140197CrossRef Ludlow J, Timothy R, Walker C, Hunter R, Benavides E, Samuelson D, Scheske M (2015) Effective dose of dental CBCT—a meta analysis of published data and additional data for nine CBCT units. Dento Maxillofac Radiol 44(1):20140197CrossRef
34.
Zurück zum Zitat Kyriakou Y, Kolditz D, Langner O, Krause J, Kalender W (2011) Digital volume tomography (DVT) and multislice spiral CT (MSCT): an objective examination of dose and image quality. Rofo 183(2):144–153. doi:10.1055/s-0029-1245709 CrossRefPubMed Kyriakou Y, Kolditz D, Langner O, Krause J, Kalender W (2011) Digital volume tomography (DVT) and multislice spiral CT (MSCT): an objective examination of dose and image quality. Rofo 183(2):144–153. doi:10.​1055/​s-0029-1245709 CrossRefPubMed
35.
Zurück zum Zitat Vercruyssen M, Fortin T, Widmann G, Jacobs R, Quirynen M (2014) Different techniques of static/dynamic guided implant surgery: modalities and indications. Periodontol 2000 66(1):214–227CrossRefPubMed Vercruyssen M, Fortin T, Widmann G, Jacobs R, Quirynen M (2014) Different techniques of static/dynamic guided implant surgery: modalities and indications. Periodontol 2000 66(1):214–227CrossRefPubMed
Metadaten
Titel
Validity of linear measurements of the jaws using ultralow-dose MDCT and the iterative techniques of ASIR and MBIR
verfasst von
Asma’a A. Al-Ekrish
Reema Al-Shawaf
Peter Schullian
Ra’ed Al-Sadhan
Romed Hörmann
Gerlig Widmann
Publikationsdatum
01.10.2016
Verlag
Springer Berlin Heidelberg
Erschienen in
International Journal of Computer Assisted Radiology and Surgery / Ausgabe 10/2016
Print ISSN: 1861-6410
Elektronische ISSN: 1861-6429
DOI
https://doi.org/10.1007/s11548-016-1419-y

Weitere Artikel der Ausgabe 10/2016

International Journal of Computer Assisted Radiology and Surgery 10/2016 Zur Ausgabe

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.