01.08.2011 | Symposium: UHMWPE for Arthroplasty: From Powder to Debris

History and Systematic Review of Wear and Osteolysis Outcomes for First-generation Highly Crosslinked Polyethylene

verfasst von: Steven M. Kurtz, PhD, Heather A. Gawel, MS, Jasmine D. Patel, PhD

Erschienen in: Clinical Orthopaedics and Related Research® | Ausgabe 8/2011

Einloggen, um Zugang zu erhalten

Abstract

Background

Highly crosslinked polyethylene (HXLPE) was introduced to reduce wear and osteolysis in total joint arthroplasty. While many studies report wear and osteolysis associated with HXLPE, analytical techniques, clinical study design and followup, HXLPE formulation and implant design characteristics, and patient populations differ substantially among investigations, complicating a unified perspective.

Questions/purposes

Literature on first-generation HXLPE was summarized. We systematically reviewed the radiographic wear data and incidence of osteolysis for HXLPE in hip and knee arthroplasty.

Methods

PubMed identified 391 studies; 28 met inclusion criteria for a weighted-averages analysis of two-dimensional femoral head penetration rates. To determine the incidence of osteolysis, we estimated a pooled odds ratio using a random-effects model.

Results

Weighted-averages analyses of femoral head penetration rates in HXLPE liners and conventional UHMWPE liners resulted, respectively, in a mean two-dimensional linear penetration rate of 0.042 mm/year based on 28 studies (n = 1503 hips) and 0.137 mm/year based on 18 studies (n = 695 hips). The pooled odds ratio for the risk of osteolysis in HXLPE versus conventional liners was 0.13 (95% confidence interval, 0.06–0.27) among studies with minimum 5-year followup. We identified two clinical studies of HXLPE in TKA, preventing systematic analysis of outcomes.

Conclusions

HXLPE liner studies consistently report lower femoral head penetration and an 87% lower risk of osteolysis. Reduction in femoral head penetration or osteolysis risk is not established for large-diameter (> 32 mm) metallic femoral heads or ceramic femoral heads of any size. Few studies document the clinical performance of HXLPE in knees.
Anhänge
Nur mit Berechtigung zugänglich
Literatur
1.
Zurück zum Zitat Ainsworth R, Farling G, Bardos D. An improved bearing material for joint replacement prostheses: carbon fiber-reinforced UHMW polyethylene. In: Transactions of the 3rd Annual Meeting of the Society for Biomaterials. Minneapolis, MN: Society for Biomaterials; 1977:119. Ainsworth R, Farling G, Bardos D. An improved bearing material for joint replacement prostheses: carbon fiber-reinforced UHMW polyethylene. In: Transactions of the 3rd Annual Meeting of the Society for Biomaterials. Minneapolis, MN: Society for Biomaterials; 1977:119.
2.
Zurück zum Zitat American Society for Testing and Materials. ASTM F2565-06. Standard Guide for Extensively Irradiation-Crosslinked Ultra-High Molecular Weight Polyethylene Fabricated Forms for Surgical Implant Applications. West Conshohocken, PA: American Society for Testing and Materials; 2006. American Society for Testing and Materials. ASTM F2565-06. Standard Guide for Extensively Irradiation-Crosslinked Ultra-High Molecular Weight Polyethylene Fabricated Forms for Surgical Implant Applications. West Conshohocken, PA: American Society for Testing and Materials; 2006.
3.
Zurück zum Zitat Amstutz HC, Campbell P, McKellop H, Schmalzreid TP, Gillespie WJ, Howie D, Jacobs J, Medley J, Merritt K. Metal on metal total hip replacement workshop consensus document. Clin Orthop Relat Res. 1996;329(Suppl):S297–S303.PubMed Amstutz HC, Campbell P, McKellop H, Schmalzreid TP, Gillespie WJ, Howie D, Jacobs J, Medley J, Merritt K. Metal on metal total hip replacement workshop consensus document. Clin Orthop Relat Res. 1996;329(Suppl):S297–S303.PubMed
4.
Zurück zum Zitat August AC, Aldam CH, Pynsent PB. The McKee-Farrar hip arthroplasty: a long-term study. J Bone Joint Surg Br. 1986;68:520–527.PubMed August AC, Aldam CH, Pynsent PB. The McKee-Farrar hip arthroplasty: a long-term study. J Bone Joint Surg Br. 1986;68:520–527.PubMed
5.
Zurück zum Zitat Ayers DC, Hays PL, Drew JM, Eskander MS, Osuch D, Bragdon CR. Two-year radiostereometric analysis evaluation of femoral head penetration in a challenging population of young total hip arthroplasty patients. J Arthroplasty. 2009;24:9–14.PubMed Ayers DC, Hays PL, Drew JM, Eskander MS, Osuch D, Bragdon CR. Two-year radiostereometric analysis evaluation of femoral head penetration in a challenging population of young total hip arthroplasty patients. J Arthroplasty. 2009;24:9–14.PubMed
6.
Zurück zum Zitat Baker DA, Bellare A, Pruitt L. The effects of degree of crosslinking on the fatigue crack initiation and propagation resistance of orthopedic-grade polyethylene. J Biomed Mater Res A. 2003;66:146–154.PubMed Baker DA, Bellare A, Pruitt L. The effects of degree of crosslinking on the fatigue crack initiation and propagation resistance of orthopedic-grade polyethylene. J Biomed Mater Res A. 2003;66:146–154.PubMed
7.
Zurück zum Zitat Baker DA, Hastings RS, Pruitt L. Study of fatigue resistance of chemical and radiation crosslinked medical grade ultrahigh molecular weight polyethylene. J Biomed Mater Res. 1999;46:573–581.PubMed Baker DA, Hastings RS, Pruitt L. Study of fatigue resistance of chemical and radiation crosslinked medical grade ultrahigh molecular weight polyethylene. J Biomed Mater Res. 1999;46:573–581.PubMed
8.
Zurück zum Zitat Bartel DL, Bicknell VL, Wright TM. The effect of conformity, thickness, and material on stresses in ultra-high molecular weight components for total joint replacement. J Bone Joint Surg Am. 1986;68:1041–1051.PubMed Bartel DL, Bicknell VL, Wright TM. The effect of conformity, thickness, and material on stresses in ultra-high molecular weight components for total joint replacement. J Bone Joint Surg Am. 1986;68:1041–1051.PubMed
9.
Zurück zum Zitat Bartel DL, Rawlinson JJ, Burstein AH, Ranawat CS, Flynn WF Jr. Stresses in polyethylene components of contemporary total knee replacements. Clin Orthop Relat Res. 1995;317:76–82.PubMed Bartel DL, Rawlinson JJ, Burstein AH, Ranawat CS, Flynn WF Jr. Stresses in polyethylene components of contemporary total knee replacements. Clin Orthop Relat Res. 1995;317:76–82.PubMed
10.
Zurück zum Zitat Beksac B, Salas A, Gonzalez Della Valle A, Salvati EA. Wear is reduced in THA performed with highly cross-linked polyethylene. Clin Orthop Relat Res. 2009;467:1765–1772.PubMed Beksac B, Salas A, Gonzalez Della Valle A, Salvati EA. Wear is reduced in THA performed with highly cross-linked polyethylene. Clin Orthop Relat Res. 2009;467:1765–1772.PubMed
11.
Zurück zum Zitat Bell CJ, Walker PS, Abeysundera MR, Simmons JM, King PM, Blunn GW. Effect of oxidation on delamination of ultrahigh-molecular-weight polyethylene tibial components. J Arthroplasty. 1998;13:280–290.PubMed Bell CJ, Walker PS, Abeysundera MR, Simmons JM, King PM, Blunn GW. Effect of oxidation on delamination of ultrahigh-molecular-weight polyethylene tibial components. J Arthroplasty. 1998;13:280–290.PubMed
12.
Zurück zum Zitat Bhambri SK, Gsell R, Kirkpatrick L, Swarts D, Blanchard CR, Crowninshield RD. The effect of aging on mechanical properties of melt-annealed highly crosslinked UHMWPE. In: Kurtz SM, Gsell R, Martell J, eds. ASTM STP 1445. Highly Crosslinked and Thermally Treated Ultra-High Molecular Weight Polyethylene For Joint Replacements. West Conshohocken, PA: American Society for Testing and Materials; 2004. Bhambri SK, Gsell R, Kirkpatrick L, Swarts D, Blanchard CR, Crowninshield RD. The effect of aging on mechanical properties of melt-annealed highly crosslinked UHMWPE. In: Kurtz SM, Gsell R, Martell J, eds. ASTM STP 1445. Highly Crosslinked and Thermally Treated Ultra-High Molecular Weight Polyethylene For Joint Replacements. West Conshohocken, PA: American Society for Testing and Materials; 2004.
13.
Zurück zum Zitat Bitsch RG, Loidolt T, Heisel C, Ball S, Schmalzried TP. Reduction of osteolysis with use of Marathon cross-linked polyethylene: a concise follow-up, at a minimum of five years, of a previous report. J Bone Joint Surg Am. 2008;90:1487–1491.PubMed Bitsch RG, Loidolt T, Heisel C, Ball S, Schmalzried TP. Reduction of osteolysis with use of Marathon cross-linked polyethylene: a concise follow-up, at a minimum of five years, of a previous report. J Bone Joint Surg Am. 2008;90:1487–1491.PubMed
14.
Zurück zum Zitat Bizot P, Nizard R, Lerouge S, Prudhommeaux F, Sedel L. Ceramic/ceramic total hip arthroplasty. J Orthop Sci. 2000;5:622–627.PubMed Bizot P, Nizard R, Lerouge S, Prudhommeaux F, Sedel L. Ceramic/ceramic total hip arthroplasty. J Orthop Sci. 2000;5:622–627.PubMed
15.
Zurück zum Zitat Blunn GW, Joshi AB, Minns RJ, Lidgren L, Lilley P, Ryd L, Engelbrecht E, Walker PS. Wear in retrieved condylar knee arthroplasties: a comparison of wear in different designs of 280 retrieved condylar knee prostheses. J Arthroplasty. 1997;12:281–290.PubMed Blunn GW, Joshi AB, Minns RJ, Lidgren L, Lilley P, Ryd L, Engelbrecht E, Walker PS. Wear in retrieved condylar knee arthroplasties: a comparison of wear in different designs of 280 retrieved condylar knee prostheses. J Arthroplasty. 1997;12:281–290.PubMed
16.
Zurück zum Zitat Bostrom MP, Bennett AP, Rimnac CM, Wright TM. The natural history of ultra high molecular weight polyethylene. Clin Orthop Relat Res. 1994;309:20–28.PubMed Bostrom MP, Bennett AP, Rimnac CM, Wright TM. The natural history of ultra high molecular weight polyethylene. Clin Orthop Relat Res. 1994;309:20–28.PubMed
17.
Zurück zum Zitat Boutin P, Christel P, Dorlot JM, Meunier A, de Roquancourt A, Blanquaert D, Herman S, Sedel L, Witvoet J. The use of dense alumina-alumina ceramic combination in total hip replacement. J Biomed Mater Res. 1988;22:1203–1232.PubMed Boutin P, Christel P, Dorlot JM, Meunier A, de Roquancourt A, Blanquaert D, Herman S, Sedel L, Witvoet J. The use of dense alumina-alumina ceramic combination in total hip replacement. J Biomed Mater Res. 1988;22:1203–1232.PubMed
18.
Zurück zum Zitat Bozic KJ, Kurtz S, Lau E, Ong K, Chiu V, Vail TP, Rubash HE, Berry DJ. The epidemiology of bearing surface usage in total hip arthroplasty in the United States. J Bone Joint Surg Am. 2009;91:1614–1620.PubMed Bozic KJ, Kurtz S, Lau E, Ong K, Chiu V, Vail TP, Rubash HE, Berry DJ. The epidemiology of bearing surface usage in total hip arthroplasty in the United States. J Bone Joint Surg Am. 2009;91:1614–1620.PubMed
19.
Zurück zum Zitat Bragdon CR, Barrett S, Martell JM, Greene ME, Malchau H, Harris WH. Steady-state penetration rates of electron beam-irradiated, highly cross-linked polyethylene at an average 45-month follow-up. J Arthroplasty. 2006;21:935–943.PubMed Bragdon CR, Barrett S, Martell JM, Greene ME, Malchau H, Harris WH. Steady-state penetration rates of electron beam-irradiated, highly cross-linked polyethylene at an average 45-month follow-up. J Arthroplasty. 2006;21:935–943.PubMed
20.
Zurück zum Zitat Bragdon CR, Greene ME, Freiberg AA, Harris WH, Malchau H. Radiostereometric analysis comparison of wear of highly cross-linked polyethylene against 36- vs 28-mm femoral heads. J Arthroplasty. 2007;22:125–129.PubMed Bragdon CR, Greene ME, Freiberg AA, Harris WH, Malchau H. Radiostereometric analysis comparison of wear of highly cross-linked polyethylene against 36- vs 28-mm femoral heads. J Arthroplasty. 2007;22:125–129.PubMed
21.
Zurück zum Zitat Bragdon CR, Kwon YM, Geller JA, Greene ME, Freiberg AA, Harris WH, Malchau H. Minimum 6-year followup of highly cross-linked polyethylene in THA. Clin Orthop Relat Res. 2007;465:122–127.PubMed Bragdon CR, Kwon YM, Geller JA, Greene ME, Freiberg AA, Harris WH, Malchau H. Minimum 6-year followup of highly cross-linked polyethylene in THA. Clin Orthop Relat Res. 2007;465:122–127.PubMed
22.
Zurück zum Zitat Calvert GT, Devane PA, Fielden J, Adams K, Horne JG. A double-blind, prospective, randomized controlled trial comparing highly cross-linked and conventional polyethylene in primary total hip arthroplasty. J Arthroplasty. 2009;24:505–510.PubMed Calvert GT, Devane PA, Fielden J, Adams K, Horne JG. A double-blind, prospective, randomized controlled trial comparing highly cross-linked and conventional polyethylene in primary total hip arthroplasty. J Arthroplasty. 2009;24:505–510.PubMed
23.
Zurück zum Zitat Champion AR, Li S, Saum K, Howard E, Simmons W. The effect of crystallinity on the physical properties of UHMWPE. Trans Orthop Res Soc. 1994;19:585. Champion AR, Li S, Saum K, Howard E, Simmons W. The effect of crystallinity on the physical properties of UHMWPE. Trans Orthop Res Soc. 1994;19:585.
24.
Zurück zum Zitat Chan FW, Bobyn JD, Medley JB, Krygier JJ, Yue S, Tanzer M. Engineering issues and wear performance of metal on metal hip implants. Clin Orthop Relat Res. 1996;333:96–107.PubMed Chan FW, Bobyn JD, Medley JB, Krygier JJ, Yue S, Tanzer M. Engineering issues and wear performance of metal on metal hip implants. Clin Orthop Relat Res. 1996;333:96–107.PubMed
25.
Zurück zum Zitat Charnley J. Arthroplasty of the hip: a new operation. Lancet 1961;I:1129–1132. Charnley J. Arthroplasty of the hip: a new operation. Lancet 1961;I:1129–1132.
26.
Zurück zum Zitat Charnley J. Tissue reaction to the polytetraflouroethyene. Lancet. 1963;II:1379. Charnley J. Tissue reaction to the polytetraflouroethyene. Lancet. 1963;II:1379.
27.
Zurück zum Zitat Charnley J. Low friction principle. In: Low Friction Arthroplasty of the Hip: Theory and Practice. Berlin, Germany: Springer-Verlag; 1979:3–16. Charnley J. Low friction principle. In: Low Friction Arthroplasty of the Hip: Theory and Practice. Berlin, Germany: Springer-Verlag; 1979:3–16.
28.
Zurück zum Zitat Charnley J. Personal communication, 1971: Sterilization of Charnley total hip sockets of high density polyethylene. In: Kurtz SM, ed. The UHMWPE Biomaterials Handbook: Ultra-High Molecular Weight Polyethylene in Total Joint Replacement and Medical Devices. 2nd ed. Burlington, MA: Academic Press; 2009:39–40. Charnley J. Personal communication, 1971: Sterilization of Charnley total hip sockets of high density polyethylene. In: Kurtz SM, ed. The UHMWPE Biomaterials Handbook: Ultra-High Molecular Weight Polyethylene in Total Joint Replacement and Medical Devices. 2nd ed. Burlington, MA: Academic Press; 2009:39–40.
29.
Zurück zum Zitat Charnley J, Halley DK. Rate of wear in total hip replacement. Clin Orthop Relat Res. 1975;112:170–179.PubMed Charnley J, Halley DK. Rate of wear in total hip replacement. Clin Orthop Relat Res. 1975;112:170–179.PubMed
30.
Zurück zum Zitat Collier JP, Sperling DK, Currier JH, Sutula LC, Saum KA, Mayor MB. Impact of gamma sterilization on clinical performance of polyethylene in the knee. J Arthroplasty. 1996;11:377–389.PubMed Collier JP, Sperling DK, Currier JH, Sutula LC, Saum KA, Mayor MB. Impact of gamma sterilization on clinical performance of polyethylene in the knee. J Arthroplasty. 1996;11:377–389.PubMed
31.
Zurück zum Zitat Collier JP, Sutula LC, Currier BH, Currier JH, Wooding RE, Williams IR, Farber KB, Mayor MB. Overview of polyethylene as a bearing material: comparison of sterilization methods. Clin Orthop Relat Res. 1996;333:76–86.PubMed Collier JP, Sutula LC, Currier BH, Currier JH, Wooding RE, Williams IR, Farber KB, Mayor MB. Overview of polyethylene as a bearing material: comparison of sterilization methods. Clin Orthop Relat Res. 1996;333:76–86.PubMed
32.
Zurück zum Zitat Collier MB, Engh CA Jr, Engh GA. Shelf age of the polyethylene tibial component and outcome of unicondylar knee arthroplasty. J Bone Joint Surg Am. 2004;86:763–769.PubMed Collier MB, Engh CA Jr, Engh GA. Shelf age of the polyethylene tibial component and outcome of unicondylar knee arthroplasty. J Bone Joint Surg Am. 2004;86:763–769.PubMed
33.
Zurück zum Zitat Connelly GM, Rimnac CM, Wright TM, Hertzberg RW, Manson JA. Fatigue crack propagation behavior of ultrahigh molecular weight polyethylene. J Orthop Res. 1984;2:119–125.PubMed Connelly GM, Rimnac CM, Wright TM, Hertzberg RW, Manson JA. Fatigue crack propagation behavior of ultrahigh molecular weight polyethylene. J Orthop Res. 1984;2:119–125.PubMed
34.
Zurück zum Zitat Costa L, Brach de Prever EM. UHMWPE for Arthroplasty. Turin, Italy: Edizioni Minerva Medica; 2000. Costa L, Brach de Prever EM. UHMWPE for Arthroplasty. Turin, Italy: Edizioni Minerva Medica; 2000.
35.
Zurück zum Zitat Currier BH, Currier JH, Mayor MB, Lyford KA, Collier JP, Van Citters DW. Evaluation of oxidation and fatigue damage of retrieved crossfire polyethylene acetabular cups. J Bone Joint Surg Am. 2007;89:2023–2029.PubMed Currier BH, Currier JH, Mayor MB, Lyford KA, Collier JP, Van Citters DW. Evaluation of oxidation and fatigue damage of retrieved crossfire polyethylene acetabular cups. J Bone Joint Surg Am. 2007;89:2023–2029.PubMed
36.
Zurück zum Zitat Currier BH, Currier JH, Mayor MB, Lyford KA, Van Citters DW, Collier JP. In vivo oxidation of gamma-barrier-sterilized ultra-high-molecular-weight polyethylene bearings. J Arthroplasty. 2007;22:721–731.PubMed Currier BH, Currier JH, Mayor MB, Lyford KA, Van Citters DW, Collier JP. In vivo oxidation of gamma-barrier-sterilized ultra-high-molecular-weight polyethylene bearings. J Arthroplasty. 2007;22:721–731.PubMed
37.
Zurück zum Zitat D’Antonio JA, Manley MT, Capello WN, Bierbaum BE, Ramakrishnan R, Naughton M, Sutton K. Five-year experience with Crossfire highly cross-linked polyethylene. Clin Orthop Relat Res. 2005;441:143–150.PubMed D’Antonio JA, Manley MT, Capello WN, Bierbaum BE, Ramakrishnan R, Naughton M, Sutton K. Five-year experience with Crossfire highly cross-linked polyethylene. Clin Orthop Relat Res. 2005;441:143–150.PubMed
38.
Zurück zum Zitat Digas G, Karrholm J, Thanner J, Herberts P. 5-year experience of highly cross-linked polyethylene in cemented and uncemented sockets: two randomized studies using radiostereometric analysis. Acta Orthop. 2007;78:746–754.PubMed Digas G, Karrholm J, Thanner J, Herberts P. 5-year experience of highly cross-linked polyethylene in cemented and uncemented sockets: two randomized studies using radiostereometric analysis. Acta Orthop. 2007;78:746–754.PubMed
39.
Zurück zum Zitat Digas G, Karrholm J, Thanner J, Malchau H, Herberts P. Highly cross-linked polyethylene in total hip arthroplasty: randomized evaluation of penetration rate in cemented and uncemented sockets using radiostereometric analysis. Clin Orthop Relat Res. 2004;429:6–16.PubMed Digas G, Karrholm J, Thanner J, Malchau H, Herberts P. Highly cross-linked polyethylene in total hip arthroplasty: randomized evaluation of penetration rate in cemented and uncemented sockets using radiostereometric analysis. Clin Orthop Relat Res. 2004;429:6–16.PubMed
40.
Zurück zum Zitat Dorr LD, Hilton KR, Wan Z, Markovich GD, Bloebaum R. Modern metal on metal articulation for total hip replacements. Clin Orthop Relat Res. 1996;333:108–117.PubMed Dorr LD, Hilton KR, Wan Z, Markovich GD, Bloebaum R. Modern metal on metal articulation for total hip replacements. Clin Orthop Relat Res. 1996;333:108–117.PubMed
41.
Zurück zum Zitat Dorr LD, Wan Z, Shahrdar C, Sirianni L, Boutary M, Yun A. Clinical performance of a Durasul highly cross-linked polyethylene acetabular liner for total hip arthroplasty at five years. J Bone Joint Surg Am. 2005;87:1816–1821.PubMed Dorr LD, Wan Z, Shahrdar C, Sirianni L, Boutary M, Yun A. Clinical performance of a Durasul highly cross-linked polyethylene acetabular liner for total hip arthroplasty at five years. J Bone Joint Surg Am. 2005;87:1816–1821.PubMed
42.
Zurück zum Zitat Duffy GP, Wannomae KK, Rowell SL, Muratoglu OK. Fracture of a cross-linked polyethylene liner due to impingement. J Arthroplasty. 2009;24:158.e15–19. Duffy GP, Wannomae KK, Rowell SL, Muratoglu OK. Fracture of a cross-linked polyethylene liner due to impingement. J Arthroplasty. 2009;24:158.e15–19.
43.
Zurück zum Zitat Dumbleton JH, D’Antonio JA, Manley MT, Capello WN, Wang A. The basis for a second-generation highly cross-linked UHMWPE. Clin Orthop Relat Res. 2006;453:265–271.PubMed Dumbleton JH, D’Antonio JA, Manley MT, Capello WN, Wang A. The basis for a second-generation highly cross-linked UHMWPE. Clin Orthop Relat Res. 2006;453:265–271.PubMed
44.
Zurück zum Zitat Engh CA Jr, Stepniewski AS, Ginn SD, Beykirch SE, Sychterz-Terefenko CJ, Hopper RH Jr, Engh CA. A randomized prospective evaluation of outcomes after total hip arthroplasty using cross-linked marathon and non-cross-linked Enduron polyethylene liners. J Arthroplasty. 2006;21:17–25.PubMed Engh CA Jr, Stepniewski AS, Ginn SD, Beykirch SE, Sychterz-Terefenko CJ, Hopper RH Jr, Engh CA. A randomized prospective evaluation of outcomes after total hip arthroplasty using cross-linked marathon and non-cross-linked Enduron polyethylene liners. J Arthroplasty. 2006;21:17–25.PubMed
45.
Zurück zum Zitat Fukui K, Kaneuji A, Sugimori T, Ichiseki T, Kitamura K, Matsumoto T. Wear comparison between a highly cross-linked polyethylene and conventional polyethylene against a zirconia femoral head: minimum 5-year follow-up. J Arthroplasty. 2011;26:45–49.PubMed Fukui K, Kaneuji A, Sugimori T, Ichiseki T, Kitamura K, Matsumoto T. Wear comparison between a highly cross-linked polyethylene and conventional polyethylene against a zirconia femoral head: minimum 5-year follow-up. J Arthroplasty. 2011;26:45–49.PubMed
46.
Zurück zum Zitat Garcia-Rey E, Garcia-Cimbrelo E, Cruz-Pardos A, Ortega-Chamarro J. New polyethylenes in total hip replacement: a prospective, comparative clinical study of two types of liner. J Bone Joint Surg Br. 2008;90:149–153.PubMed Garcia-Rey E, Garcia-Cimbrelo E, Cruz-Pardos A, Ortega-Chamarro J. New polyethylenes in total hip replacement: a prospective, comparative clinical study of two types of liner. J Bone Joint Surg Br. 2008;90:149–153.PubMed
47.
Zurück zum Zitat Garino JP. Modern ceramic-on-ceramic total hip systems in the United States: early results. Clin Orthop Relat Res. 2000;379:41–47.PubMed Garino JP. Modern ceramic-on-ceramic total hip systems in the United States: early results. Clin Orthop Relat Res. 2000;379:41–47.PubMed
48.
Zurück zum Zitat Garvin KL, Hartman CW, Mangla J, Murdoch N, Martell JM. Wear analysis in THA utilizing oxidized zirconium and crosslinked polyethylene. Clin Orthop Relat Res. 2009;467:141–145.PubMed Garvin KL, Hartman CW, Mangla J, Murdoch N, Martell JM. Wear analysis in THA utilizing oxidized zirconium and crosslinked polyethylene. Clin Orthop Relat Res. 2009;467:141–145.PubMed
49.
Zurück zum Zitat Geller JA, Malchau H, Bragdon C, Greene M, Harris WH, Freiberg AA. Large diameter femoral heads on highly cross-linked polyethylene: minimum 3-year results. Clin Orthop Relat Res. 2006;447:53–59.PubMed Geller JA, Malchau H, Bragdon C, Greene M, Harris WH, Freiberg AA. Large diameter femoral heads on highly cross-linked polyethylene: minimum 3-year results. Clin Orthop Relat Res. 2006;447:53–59.PubMed
50.
Zurück zum Zitat Gioe TJ, Sharma A, Tatman P, Mehle S. Do “premium” joint implants add value? Analysis of high cost joint implants in a community registry. Clin Orthop Relat Res. 2011;469:48–54.PubMed Gioe TJ, Sharma A, Tatman P, Mehle S. Do “premium” joint implants add value? Analysis of high cost joint implants in a community registry. Clin Orthop Relat Res. 2011;469:48–54.PubMed
51.
Zurück zum Zitat Glyn-Jones S, Isaac S, Hauptfleisch J, McLardy-Smith P, Murray DW, Gill HS. Does highly cross-linked polyethylene wear less than conventional polyethylene in total hip arthroplasty? A double-blind, randomized, and controlled trial using roentgen stereophotogrammetric analysis. J Arthroplasty. 2008;23:337–343.PubMed Glyn-Jones S, Isaac S, Hauptfleisch J, McLardy-Smith P, Murray DW, Gill HS. Does highly cross-linked polyethylene wear less than conventional polyethylene in total hip arthroplasty? A double-blind, randomized, and controlled trial using roentgen stereophotogrammetric analysis. J Arthroplasty. 2008;23:337–343.PubMed
52.
Zurück zum Zitat Glyn-Jones S, McLardy-Smith P, Gill HS, Murray DW. The creep and wear of highly cross-linked polyethylene: a three-year randomised, controlled trial using radiostereometric analysis. J Bone Joint Surg Br. 2008;90:556–561.PubMed Glyn-Jones S, McLardy-Smith P, Gill HS, Murray DW. The creep and wear of highly cross-linked polyethylene: a three-year randomised, controlled trial using radiostereometric analysis. J Bone Joint Surg Br. 2008;90:556–561.PubMed
53.
Zurück zum Zitat Good V, Widding K, Scott M, Jani S. The sensitivity of crosslinked UHMWPE to abrasive wear: hips vs. knees. In: Kurtz SM, Gsell R, Martell J, eds. ASTM STP 1445. Highly Crosslinked and Thermally Treated Ultra-High Molecular Weight Polyethylene For Joint Replacements. West Conshohocken, PA: American Society for Testing and Materials; 2004. Good V, Widding K, Scott M, Jani S. The sensitivity of crosslinked UHMWPE to abrasive wear: hips vs. knees. In: Kurtz SM, Gsell R, Martell J, eds. ASTM STP 1445. Highly Crosslinked and Thermally Treated Ultra-High Molecular Weight Polyethylene For Joint Replacements. West Conshohocken, PA: American Society for Testing and Materials; 2004.
54.
Zurück zum Zitat Greer K, King R, Chan FW. Effects of raw material, irradiation dose, and irradiation source on crosslinking of UHMWPE. In: Kurtz SM, Gsell R, Martell J, eds. ASTM STP 1445. Highly Crosslinked and Thermally Treated Ultra-High Molecular Weight Polyethylene For Joint Replacements. West Conshohocken, PA: American Society for Testing and Materials; 2004. Greer K, King R, Chan FW. Effects of raw material, irradiation dose, and irradiation source on crosslinking of UHMWPE. In: Kurtz SM, Gsell R, Martell J, eds. ASTM STP 1445. Highly Crosslinked and Thermally Treated Ultra-High Molecular Weight Polyethylene For Joint Replacements. West Conshohocken, PA: American Society for Testing and Materials; 2004.
55.
Zurück zum Zitat Griffith MJ, Seidenstein MK, Williams D, Charnley J. Socket wear in Charnley low friction arthroplasty of the hip. Clin Orthop Relat Res. 1978;137:37–47.PubMed Griffith MJ, Seidenstein MK, Williams D, Charnley J. Socket wear in Charnley low friction arthroplasty of the hip. Clin Orthop Relat Res. 1978;137:37–47.PubMed
56.
Zurück zum Zitat Grobbelaar CJ, Du Plessis TA, Marais F. The radiation improvment of polyethylene prostheses: a preliminary study. J Bone Joint Surg Br. 1978;60:370–374.PubMed Grobbelaar CJ, Du Plessis TA, Marais F. The radiation improvment of polyethylene prostheses: a preliminary study. J Bone Joint Surg Br. 1978;60:370–374.PubMed
57.
Zurück zum Zitat Halley D, Glassman A, Crowninshield RD. Recurrent dislocation after revision total hip replacement with a large prosthetic femoral head: a case report. J Bone Joint Surg Am. 2004;86:827–830.PubMed Halley D, Glassman A, Crowninshield RD. Recurrent dislocation after revision total hip replacement with a large prosthetic femoral head: a case report. J Bone Joint Surg Am. 2004;86:827–830.PubMed
58.
Zurück zum Zitat Hammerberg EM, Zhinian W, Manish D, Lawrence DD. Wear and range of motion of different femoral head sizes. J Arthroplasty. 2010;25:839–843.PubMed Hammerberg EM, Zhinian W, Manish D, Lawrence DD. Wear and range of motion of different femoral head sizes. J Arthroplasty. 2010;25:839–843.PubMed
59.
Zurück zum Zitat Harris WH. The problem is osteolysis. Clin Orthop Relat Res. 1995;311:46–53.PubMed Harris WH. The problem is osteolysis. Clin Orthop Relat Res. 1995;311:46–53.PubMed
60.
Zurück zum Zitat Heisel C, Silva M, dela Rosa MA, Schmalzried TP. Short-term in vivo wear of cross-linked polyethylene. J Bone Joint Surg Am. 2004;86:748–751.PubMed Heisel C, Silva M, dela Rosa MA, Schmalzried TP. Short-term in vivo wear of cross-linked polyethylene. J Bone Joint Surg Am. 2004;86:748–751.PubMed
61.
Zurück zum Zitat Heisel C, Silva M, Schmalzried TP. In vivo wear of bilateral total hip replacements: conventional versus crosslinked polyethylene. Arch Orthop Trauma Surg. 2005;125:555–557.PubMed Heisel C, Silva M, Schmalzried TP. In vivo wear of bilateral total hip replacements: conventional versus crosslinked polyethylene. Arch Orthop Trauma Surg. 2005;125:555–557.PubMed
62.
Zurück zum Zitat Hodrick JT, Severson EP, McAlister DS, Dahl B, Hofmann AA. Highly crosslinked polyethylene is safe for use in total knee arthroplasty. Clin Orthop Relat Res. 2008;466:2806–2812.PubMed Hodrick JT, Severson EP, McAlister DS, Dahl B, Hofmann AA. Highly crosslinked polyethylene is safe for use in total knee arthroplasty. Clin Orthop Relat Res. 2008;466:2806–2812.PubMed
63.
Zurück zum Zitat Hopper RH Jr, Young AM, Orishimo KF, Engh CA Jr. Effect of terminal sterilization with gas plasma or gamma radiation on wear of polyethylene liners. J Bone Joint Surg Am. 2003;85:464–468.PubMed Hopper RH Jr, Young AM, Orishimo KF, Engh CA Jr. Effect of terminal sterilization with gas plasma or gamma radiation on wear of polyethylene liners. J Bone Joint Surg Am. 2003;85:464–468.PubMed
64.
Zurück zum Zitat Hopper RH Jr, Young AM, Orishimo KF, McAuley JP. Correlation between early and late wear rates in total hip arthroplasty with application to the performance of marathon cross-linked polyethylene liners. J Arthroplasty. 2003;18:60–67.PubMed Hopper RH Jr, Young AM, Orishimo KF, McAuley JP. Correlation between early and late wear rates in total hip arthroplasty with application to the performance of marathon cross-linked polyethylene liners. J Arthroplasty. 2003;18:60–67.PubMed
65.
Zurück zum Zitat Ianuzzi A, Kurtz SM. Sterilization of joint replacement materials. In: Revell PA, ed. Joint Replacement Technology. Cambridge, UK: Woodhead Publishing Ltd; 2008:407–430. Ianuzzi A, Kurtz SM. Sterilization of joint replacement materials. In: Revell PA, ed. Joint Replacement Technology. Cambridge, UK: Woodhead Publishing Ltd; 2008:407–430.
66.
Zurück zum Zitat Ise K, Kawanabe K, Tamura J, Akiyama H, Goto K, Nakamura T. Clinical results of the wear performance of cross-linked polyethylene in total hip arthroplasty: prospective randomized trial. J Arthroplasty. 2009;24:1216–1220.PubMed Ise K, Kawanabe K, Tamura J, Akiyama H, Goto K, Nakamura T. Clinical results of the wear performance of cross-linked polyethylene in total hip arthroplasty: prospective randomized trial. J Arthroplasty. 2009;24:1216–1220.PubMed
67.
Zurück zum Zitat Jacobs JJ, Shanbhag A, Glant TT, Black J, Galante JO. Wear debris in total joint replacements. J Am Acad Orthop Surg. 1994;2:212–220.PubMed Jacobs JJ, Shanbhag A, Glant TT, Black J, Galante JO. Wear debris in total joint replacements. J Am Acad Orthop Surg. 1994;2:212–220.PubMed
68.
Zurück zum Zitat Jani S, Scott M. Wear debris generation in joint simulator testing of crosslinked UHMWPE. In: Kurtz SM, Gsell R, Martell J, eds. ASTM STP 1445. Highly Crosslinked and Thermally Treated Ultra-High Molecular Weight Polyethylene For Joint Replacements. West Conshohocken, PA: American Society for Testing and Materials; 2004. Jani S, Scott M. Wear debris generation in joint simulator testing of crosslinked UHMWPE. In: Kurtz SM, Gsell R, Martell J, eds. ASTM STP 1445. Highly Crosslinked and Thermally Treated Ultra-High Molecular Weight Polyethylene For Joint Replacements. West Conshohocken, PA: American Society for Testing and Materials; 2004.
69.
Zurück zum Zitat Kawate K, Ohmura T, Kawahara I, Tamai K, Ueha T, Takemura K. Differences in highly cross-linked polyethylene wear between zirconia and cobalt-chromium femoral heads in Japanese patients: a prospective, randomized study. J Arthroplasty. 2009;24:1221–1224.PubMed Kawate K, Ohmura T, Kawahara I, Tamai K, Ueha T, Takemura K. Differences in highly cross-linked polyethylene wear between zirconia and cobalt-chromium femoral heads in Japanese patients: a prospective, randomized study. J Arthroplasty. 2009;24:1221–1224.PubMed
70.
Zurück zum Zitat Kiely PD, Harty JA, McElwain JP. Hylamer wear rates and shelf life: a clinical correlation. Acta Orthop Belg. 2005;71:429–434.PubMed Kiely PD, Harty JA, McElwain JP. Hylamer wear rates and shelf life: a clinical correlation. Acta Orthop Belg. 2005;71:429–434.PubMed
71.
Zurück zum Zitat Kim YH, Kim JS, Choi YW, Kwon OR. Intermediate results of simultaneous alumina-on-alumina bearing and alumina-on-highly cross-linked polyethylene bearing total hip arthroplasties. J Arthroplasty. 2009;24:885–891.PubMed Kim YH, Kim JS, Choi YW, Kwon OR. Intermediate results of simultaneous alumina-on-alumina bearing and alumina-on-highly cross-linked polyethylene bearing total hip arthroplasties. J Arthroplasty. 2009;24:885–891.PubMed
72.
Zurück zum Zitat Kothari M, Bartel DL, Booker JF. Surface geometry of retrieved McKee-Farrar total hip replacements. Clin Orthop Relat Res. 1996;329(Suppl):S141–S147.PubMed Kothari M, Bartel DL, Booker JF. Surface geometry of retrieved McKee-Farrar total hip replacements. Clin Orthop Relat Res. 1996;329(Suppl):S141–S147.PubMed
73.
Zurück zum Zitat Krushell RJ, Fingeroth RH, Cushing MC. Early femoral head penetration of a highly crosslinked polyethylene liner vs a conventional polyethlene liner. J Arthroplasty. 2005;20:73–76.PubMed Krushell RJ, Fingeroth RH, Cushing MC. Early femoral head penetration of a highly crosslinked polyethylene liner vs a conventional polyethlene liner. J Arthroplasty. 2005;20:73–76.PubMed
74.
Zurück zum Zitat Kurtz SM. The UHMWPE Handbook: Ultra-High Molecular Weight Polyethylene in Total Joint Replacement. New York, NY: Academic Press; 2004. Kurtz SM. The UHMWPE Handbook: Ultra-High Molecular Weight Polyethylene in Total Joint Replacement. New York, NY: Academic Press; 2004.
75.
Zurück zum Zitat Kurtz SM. Packaging and sterilization of UHMWPE. In: Kurtz SM, ed. The UHMWPE Biomaterials Handbook: Ultra-High Molecular Weight Polyethylene in Total Joint Replacement and Medical Devices. 2nd ed. Burlington, MA: Academic Press; 2009:21–30. Kurtz SM. Packaging and sterilization of UHMWPE. In: Kurtz SM, ed. The UHMWPE Biomaterials Handbook: Ultra-High Molecular Weight Polyethylene in Total Joint Replacement and Medical Devices. 2nd ed. Burlington, MA: Academic Press; 2009:21–30.
76.
Zurück zum Zitat Kurtz SM, ed. The UHMWPE Biomaterials Handbook: Ultra-High Molecular Weight Polyethylene in Total Joint Replacement and Medical Devices. 2nd ed. Burlington, MA: Academic Press; 2009. Kurtz SM, ed. The UHMWPE Biomaterials Handbook: Ultra-High Molecular Weight Polyethylene in Total Joint Replacement and Medical Devices. 2nd ed. Burlington, MA: Academic Press; 2009.
77.
Zurück zum Zitat Kurtz SM, Gsell R, Martell J, eds. ASTM STP 1445. Highly Crosslinked and Thermally Treated Ultra-High Molecular Weight Polyethylene For Joint Replacements. West Conshohocken, PA: American Society for Testing and Materials; 2004. Kurtz SM, Gsell R, Martell J, eds. ASTM STP 1445. Highly Crosslinked and Thermally Treated Ultra-High Molecular Weight Polyethylene For Joint Replacements. West Conshohocken, PA: American Society for Testing and Materials; 2004.
78.
Zurück zum Zitat Kurtz SM, Hozack WJ, Purtill JJ, Marcolongo M, Kraay MJ, Goldberg VM, Sharkey PF, Parvizi J, Rimnac CM, Edidin AA. Significance of in vivo degradation for polyethylene in total hip arthroplasty. Clin Orthop Relat Res. 2006;453:47–57.PubMed Kurtz SM, Hozack WJ, Purtill JJ, Marcolongo M, Kraay MJ, Goldberg VM, Sharkey PF, Parvizi J, Rimnac CM, Edidin AA. Significance of in vivo degradation for polyethylene in total hip arthroplasty. Clin Orthop Relat Res. 2006;453:47–57.PubMed
79.
Zurück zum Zitat Kurtz SM, Manley M, Wang A, Taylor S, Dumbleton J. Comparison of the properties of annealed crosslinked (Crossfire) and conventional polyethylene as hip bearing materials. Bull Hosp Jt Dis. 2002-2003;61:17–26. Kurtz SM, Manley M, Wang A, Taylor S, Dumbleton J. Comparison of the properties of annealed crosslinked (Crossfire) and conventional polyethylene as hip bearing materials. Bull Hosp Jt Dis. 2002-2003;61:17–26.
80.
Zurück zum Zitat Kurtz SM, Mazzucco D, Rimnac CM, Schroeder D. Anisotropy and oxidative resistance of highly crosslinked UHMWPE after deformation processing by solid-state ram extrusion. Biomaterials. 2006;27:24–34.PubMed Kurtz SM, Mazzucco D, Rimnac CM, Schroeder D. Anisotropy and oxidative resistance of highly crosslinked UHMWPE after deformation processing by solid-state ram extrusion. Biomaterials. 2006;27:24–34.PubMed
81.
Zurück zum Zitat Kurtz SM, Muratoglu OK, Evans M, Edidin AA. Advances in the processing, sterilization, and crosslinking of ultra- high molecular weight polyethylene for total joint arthroplasty. Biomaterials. 1999;20:1659–1688.PubMed Kurtz SM, Muratoglu OK, Evans M, Edidin AA. Advances in the processing, sterilization, and crosslinking of ultra- high molecular weight polyethylene for total joint arthroplasty. Biomaterials. 1999;20:1659–1688.PubMed
82.
Zurück zum Zitat Kurtz SM, Rimnac CM, Hozack WJ, Turner J, Marcolongo M, Goldberg VM, Kraay MJ, Edidin AA. In vivo degradation of polyethylene liners after gamma sterilization in air. J Bone Joint Surg Am. 2005;87:815–823.PubMed Kurtz SM, Rimnac CM, Hozack WJ, Turner J, Marcolongo M, Goldberg VM, Kraay MJ, Edidin AA. In vivo degradation of polyethylene liners after gamma sterilization in air. J Bone Joint Surg Am. 2005;87:815–823.PubMed
83.
Zurück zum Zitat Lachiewicz PF, Heckman DS, Soileau ES, Mangla J, Martell JM. Femoral head size and wear of highly cross-linked polyethylene at 5 to 8 years. Clin Orthop Relat Res. 2009;467:3290–3296.PubMed Lachiewicz PF, Heckman DS, Soileau ES, Mangla J, Martell JM. Femoral head size and wear of highly cross-linked polyethylene at 5 to 8 years. Clin Orthop Relat Res. 2009;467:3290–3296.PubMed
84.
Zurück zum Zitat Laurent MP, Blanchard CR, Yao JQ, Johnson TS, Gilbertson LN, Swarts D, Crowninshield RD. The wear of highly crosslinked UHMWPE in the presence of abrasive particles: hip and knee simulator studies. In: Kurtz SM, Gsell R, Martell J, eds. ASTM STP 1445. Highly Crosslinked and Thermally Treated Ultra-High Molecular Weight Polyethylene For Joint Replacements. West Conshohocken, PA: American Society for Testing and Materials; 2004. Laurent MP, Blanchard CR, Yao JQ, Johnson TS, Gilbertson LN, Swarts D, Crowninshield RD. The wear of highly crosslinked UHMWPE in the presence of abrasive particles: hip and knee simulator studies. In: Kurtz SM, Gsell R, Martell J, eds. ASTM STP 1445. Highly Crosslinked and Thermally Treated Ultra-High Molecular Weight Polyethylene For Joint Replacements. West Conshohocken, PA: American Society for Testing and Materials; 2004.
85.
Zurück zum Zitat Leung SB, Egawa H, Stepniewski A, Beykirch S, Engh CA Jr, Engh CA Sr. Incidence and volume of pelvic osteolysis at early follow-up with highly cross-linked and noncross-linked polyethylene. J Arthroplasty. 2007;22:134–139.PubMed Leung SB, Egawa H, Stepniewski A, Beykirch S, Engh CA Jr, Engh CA Sr. Incidence and volume of pelvic osteolysis at early follow-up with highly cross-linked and noncross-linked polyethylene. J Arthroplasty. 2007;22:134–139.PubMed
86.
Zurück zum Zitat Li S, Burstein AH. Ultra-high molecular weight polyethylene: the material and its use in total joint implants. J Bone Joint Surg Am. 1994;76:1080–1090.PubMed Li S, Burstein AH. Ultra-high molecular weight polyethylene: the material and its use in total joint implants. J Bone Joint Surg Am. 1994;76:1080–1090.PubMed
87.
Zurück zum Zitat Li S, Chang JD, Barrena EG, Furman BD, Wright TM, Salvati E. Nonconsolidated polyethylene particles and oxidation in Charnley acetabular cups. Clin Orthop Relat Res. 1995;319:54–63.PubMed Li S, Chang JD, Barrena EG, Furman BD, Wright TM, Salvati E. Nonconsolidated polyethylene particles and oxidation in Charnley acetabular cups. Clin Orthop Relat Res. 1995;319:54–63.PubMed
88.
Zurück zum Zitat MacDonald D, Kurtz SM, Sakona A, Rimnac C. In vivo oxidation, oxidation potential, and mechanical properties of first generation highly crosslinked polyethylenes. Clin Orthop Relat Res. 2010 December 16 [Epub ahead of print]. MacDonald D, Kurtz SM, Sakona A, Rimnac C. In vivo oxidation, oxidation potential, and mechanical properties of first generation highly crosslinked polyethylenes. Clin Orthop Relat Res. 2010 December 16 [Epub ahead of print].
89.
Zurück zum Zitat Mall NA, Nunley RM, Zhu JJ, Maloney WJ, Barrack RL, Clohisy JC. The incidence of acetabular osteolysis in young patients with conventional versus highly crosslinked polyethylene. Clin Orthop Relat Res. 2011;469:372–381.PubMed Mall NA, Nunley RM, Zhu JJ, Maloney WJ, Barrack RL, Clohisy JC. The incidence of acetabular osteolysis in young patients with conventional versus highly crosslinked polyethylene. Clin Orthop Relat Res. 2011;469:372–381.PubMed
90.
Zurück zum Zitat Manning DW, Chiang PP, Martell JM, Galante JO, Harris WH. In vivo comparative wear study of traditional and highly cross-linked polyethylene in total hip arthroplasty. J Arthroplasty. 2005;20:880–886.PubMed Manning DW, Chiang PP, Martell JM, Galante JO, Harris WH. In vivo comparative wear study of traditional and highly cross-linked polyethylene in total hip arthroplasty. J Arthroplasty. 2005;20:880–886.PubMed
91.
Zurück zum Zitat Martell JM, Verner JJ, Incavo SJ. Clinical performance of a highly cross-linked polyethylene at two years in total hip arthroplasty: a randomized prospective trial. J Arthroplasty. 2003;18:55–59.PubMed Martell JM, Verner JJ, Incavo SJ. Clinical performance of a highly cross-linked polyethylene at two years in total hip arthroplasty: a randomized prospective trial. J Arthroplasty. 2003;18:55–59.PubMed
92.
Zurück zum Zitat McCalden RW, MacDonald SJ, Rorabeck CH, Bourne RB, Chess DG, Charron KD. Wear rate of highly cross-linked polyethylene in total hip arthroplasty: a randomized controlled trial. J Bone Joint Surg Am. 2009;91:773–782.PubMed McCalden RW, MacDonald SJ, Rorabeck CH, Bourne RB, Chess DG, Charron KD. Wear rate of highly cross-linked polyethylene in total hip arthroplasty: a randomized controlled trial. J Bone Joint Surg Am. 2009;91:773–782.PubMed
93.
Zurück zum Zitat McGovern TF, Ammeen DJ, Collier JP, Currier BH, Engh GA. Rapid polyethylene failure of unicondylar tibial components sterilized with gamma irradiation in air and implanted after a long shelf life. J Bone Joint Surg Am. 2002;84:901–906.PubMed McGovern TF, Ammeen DJ, Collier JP, Currier BH, Engh GA. Rapid polyethylene failure of unicondylar tibial components sterilized with gamma irradiation in air and implanted after a long shelf life. J Bone Joint Surg Am. 2002;84:901–906.PubMed
94.
Zurück zum Zitat McKee GK, Watson-Farrar J. Replacement of arthritic hips by the McKee-Farrar prosthesis. J Bone Joint Surg Br. 1966;48:245–259.PubMed McKee GK, Watson-Farrar J. Replacement of arthritic hips by the McKee-Farrar prosthesis. J Bone Joint Surg Br. 1966;48:245–259.PubMed
95.
Zurück zum Zitat McKellop H, Shen FW, Lu B, Campbell P, Salovey R. Development of an extremely wear-resistant ultra high molecular weight polyethylene for total hip replacements. J Orthop Res. 1999;17:157–167.PubMed McKellop H, Shen FW, Lu B, Campbell P, Salovey R. Development of an extremely wear-resistant ultra high molecular weight polyethylene for total hip replacements. J Orthop Res. 1999;17:157–167.PubMed
96.
Zurück zum Zitat McKellop HA, Shen FW, Campbell P, Ota T. Effect of molecular weight, calcium stearate, and sterilization methods on the wear of ultra-high molecular weight polyethylene acetabular cups in a hip simulator. J Orthop Res. 1999;17:329–339.PubMed McKellop HA, Shen FW, Campbell P, Ota T. Effect of molecular weight, calcium stearate, and sterilization methods on the wear of ultra-high molecular weight polyethylene acetabular cups in a hip simulator. J Orthop Res. 1999;17:329–339.PubMed
97.
Zurück zum Zitat McKnulty D, Swope S. The effect of crosslinking UHMWPE on in-vitro wear rates of fixed and mobile bearing knees. In: Kurtz SM, Gsell R, Martell J, eds. ASTM STP 1445. Highly Crosslinked and Thermally Treated Ultra-High Molecular Weight Polyethylene For Joint Replacements. West Conshohocken, PA: American Society for Testing and Materials; 2004. McKnulty D, Swope S. The effect of crosslinking UHMWPE on in-vitro wear rates of fixed and mobile bearing knees. In: Kurtz SM, Gsell R, Martell J, eds. ASTM STP 1445. Highly Crosslinked and Thermally Treated Ultra-High Molecular Weight Polyethylene For Joint Replacements. West Conshohocken, PA: American Society for Testing and Materials; 2004.
98.
Zurück zum Zitat Medel F, Kurtz SM, Klein G, Levine H, Sharkey P, Austin M, Kraay M, Rimnac CM. Clinical, surface damage and oxidative performance of Poly II tibial inserts after long-term implantation. J Long Term Eff Med Implants. 2008;18:151–165.PubMed Medel F, Kurtz SM, Klein G, Levine H, Sharkey P, Austin M, Kraay M, Rimnac CM. Clinical, surface damage and oxidative performance of Poly II tibial inserts after long-term implantation. J Long Term Eff Med Implants. 2008;18:151–165.PubMed
99.
Zurück zum Zitat Minoda Y, Aihara M, Sakawa A, Fukuoka S, Hayakawa K, Tomita M, Umeda N, Ohzono K. Comparison between highly cross-linked and conventional polyethylene in total knee arthroplasty. Knee. 2009;16:348–351.PubMed Minoda Y, Aihara M, Sakawa A, Fukuoka S, Hayakawa K, Tomita M, Umeda N, Ohzono K. Comparison between highly cross-linked and conventional polyethylene in total knee arthroplasty. Knee. 2009;16:348–351.PubMed
100.
Zurück zum Zitat Miyanishi K, Hara T, Kaminomachi S, Maekawa M, Iwamoto M, Torisu T. Short-term wear of Japanese highly cross-linked polyethylene in cementless THA. Arch Orthop Trauma Surg. 2008;128:995–1000.PubMed Miyanishi K, Hara T, Kaminomachi S, Maekawa M, Iwamoto M, Torisu T. Short-term wear of Japanese highly cross-linked polyethylene in cementless THA. Arch Orthop Trauma Surg. 2008;128:995–1000.PubMed
101.
Zurück zum Zitat Moore KD, Beck PR, Petersen DW, Cuckler JM, Lemons JE, Eberhardt AW. Early failure of a cross-linked polyethylene acetabular liner: a case report. J Bone Joint Surg Am. 2008;90:2499–2504.PubMed Moore KD, Beck PR, Petersen DW, Cuckler JM, Lemons JE, Eberhardt AW. Early failure of a cross-linked polyethylene acetabular liner: a case report. J Bone Joint Surg Am. 2008;90:2499–2504.PubMed
102.
Zurück zum Zitat Mu Z, Tian J, Wu T, Yang J, Pei F. A systematic review of radiological outcomes of highly cross-linked polyethylene versus conventional polyethylene in total hip arthroplasty. Int Orthop. 2009;33:599–604.PubMed Mu Z, Tian J, Wu T, Yang J, Pei F. A systematic review of radiological outcomes of highly cross-linked polyethylene versus conventional polyethylene in total hip arthroplasty. Int Orthop. 2009;33:599–604.PubMed
103.
Zurück zum Zitat Muratoglu OK, Bragdon CR, O’Connor DO, Jasty M, Harris WH. A novel method of cross-linking ultra-high-molecular-weight polyethylene to improve wear, reduce oxidation, and retain mechanical properties. Recipient of the 1999 HAP Paul Award. J Arthroplasty. 2001;16:149–160.PubMed Muratoglu OK, Bragdon CR, O’Connor DO, Jasty M, Harris WH. A novel method of cross-linking ultra-high-molecular-weight polyethylene to improve wear, reduce oxidation, and retain mechanical properties. Recipient of the 1999 HAP Paul Award. J Arthroplasty. 2001;16:149–160.PubMed
104.
Zurück zum Zitat Mutimer J, Devane PA, Adams K, Horne JG. Highly crosslinked polyethylene reduces wear in total hip arthroplasty at 5 years. Clin Orthop Relat Res. 2010;468:3228–3233.PubMed Mutimer J, Devane PA, Adams K, Horne JG. Highly crosslinked polyethylene reduces wear in total hip arthroplasty at 5 years. Clin Orthop Relat Res. 2010;468:3228–3233.PubMed
105.
Zurück zum Zitat Nair PD. Currently practised sterilization methods—some inadvertent consequences. J Biomater Appl. 1995;10:121–135.PubMed Nair PD. Currently practised sterilization methods—some inadvertent consequences. J Biomater Appl. 1995;10:121–135.PubMed
106.
Zurück zum Zitat Nakahara I, Nakamura N, Nishii T, Miki H, Sakai T, Sugano N. Minimum five-year follow-up wear measurement of longevity highly cross-linked polyethylene cup against cobalt-chromium or zirconia heads. J Arthroplasty. 2010;25:1182–1187.PubMed Nakahara I, Nakamura N, Nishii T, Miki H, Sakai T, Sugano N. Minimum five-year follow-up wear measurement of longevity highly cross-linked polyethylene cup against cobalt-chromium or zirconia heads. J Arthroplasty. 2010;25:1182–1187.PubMed
107.
Zurück zum Zitat Nizard RS, Sedel L, Christel P, Meunier A, Soudry M, Witvoet J. Ten-year survivorship of cemented ceramic-ceramic total hip prosthesis. Clin Orthop Relat Res. 1992;282:53–63.PubMed Nizard RS, Sedel L, Christel P, Meunier A, Soudry M, Witvoet J. Ten-year survivorship of cemented ceramic-ceramic total hip prosthesis. Clin Orthop Relat Res. 1992;282:53–63.PubMed
108.
Zurück zum Zitat Olyslaegers C, Defoort K, Simon JP, Vandenberghe L. Wear in conventional and highly cross-linked polyethylene cups: a 5-year follow-up study. J Arthroplasty. 2008;23:489–494.PubMed Olyslaegers C, Defoort K, Simon JP, Vandenberghe L. Wear in conventional and highly cross-linked polyethylene cups: a 5-year follow-up study. J Arthroplasty. 2008;23:489–494.PubMed
109.
Zurück zum Zitat Oonishi H, Clarke IC, Masuda S, Amino H. Study of retrieved acetabular sockets made from high-dose, cross-linked polyethylene. J Arthroplasty. 2001;16:129–133.PubMed Oonishi H, Clarke IC, Masuda S, Amino H. Study of retrieved acetabular sockets made from high-dose, cross-linked polyethylene. J Arthroplasty. 2001;16:129–133.PubMed
110.
Zurück zum Zitat Oonishi H, Kim SC, Takao Y, Kyomoto M, Iwamoto M, Ueno M. Wear of highly cross-linked polyethylene acetabular cup in Japan. J Arthroplasty. 2006;21:944–949.PubMed Oonishi H, Kim SC, Takao Y, Kyomoto M, Iwamoto M, Ueno M. Wear of highly cross-linked polyethylene acetabular cup in Japan. J Arthroplasty. 2006;21:944–949.PubMed
111.
Zurück zum Zitat Oonishi H, Takayama Y, Tsuji E. Improvement of polyethylene by irradiation in artificial joints. Rad Phys Chem. 1992;39:495–504. Oonishi H, Takayama Y, Tsuji E. Improvement of polyethylene by irradiation in artificial joints. Rad Phys Chem. 1992;39:495–504.
112.
Zurück zum Zitat Oral E, Wannomae KK, Hawkins N, Harris WH, Muratoglu OK. Alpha-tocopherol-doped irradiated UHMWPE for high fatigue resistance and low wear. Biomaterials. 2004;25:5515–5522.PubMed Oral E, Wannomae KK, Hawkins N, Harris WH, Muratoglu OK. Alpha-tocopherol-doped irradiated UHMWPE for high fatigue resistance and low wear. Biomaterials. 2004;25:5515–5522.PubMed
113.
Zurück zum Zitat Premnath V, Harris WH, Jasty M, Merrill EW. Gamma sterilization of UHMWPE articular implants: an analysis of the oxidation problem. Biomaterials. 1996;17:1741–1753.PubMed Premnath V, Harris WH, Jasty M, Merrill EW. Gamma sterilization of UHMWPE articular implants: an analysis of the oxidation problem. Biomaterials. 1996;17:1741–1753.PubMed
114.
Zurück zum Zitat Rajadhyaksha AD, Brotea C, Cheung Y, Kuhn C, Ramakrishnan R, Zelicof SB. Five-year comparative study of highly cross-linked (Crossfire) and traditional polyethylene. J Arthroplasty. 2009;24:161–167.PubMed Rajadhyaksha AD, Brotea C, Cheung Y, Kuhn C, Ramakrishnan R, Zelicof SB. Five-year comparative study of highly cross-linked (Crossfire) and traditional polyethylene. J Arthroplasty. 2009;24:161–167.PubMed
115.
Zurück zum Zitat Ramani K, Parasnis NC. Process-induced effects in compression molding of ultra-high molecular weight polyethylene (UHMWPE). In: Gsell RA, Stein HL, Ploskonka JJ, eds. Characterization and Properties of Ultra-High Molecular Weight Polyethylene. West Conshohocken, PA: American Society for Testing and Materials; 1998:5–23. Ramani K, Parasnis NC. Process-induced effects in compression molding of ultra-high molecular weight polyethylene (UHMWPE). In: Gsell RA, Stein HL, Ploskonka JJ, eds. Characterization and Properties of Ultra-High Molecular Weight Polyethylene. West Conshohocken, PA: American Society for Testing and Materials; 1998:5–23.
116.
Zurück zum Zitat Reeves EA, Barton DC, FitzPatrick DP, Fisher J. Comparison of gas plasma and gamma irradiation in air sterilization on the delamination wear of the ultra-high molecular weight polyethylene used in knee replacements. Proc Inst Mech Eng H. 2000;214:249–255.PubMed Reeves EA, Barton DC, FitzPatrick DP, Fisher J. Comparison of gas plasma and gamma irradiation in air sterilization on the delamination wear of the ultra-high molecular weight polyethylene used in knee replacements. Proc Inst Mech Eng H. 2000;214:249–255.PubMed
117.
Zurück zum Zitat Ries MD, Weaver K, Beals N. Safety and efficacy of ethylene oxide sterilized polyethylene in total knee arthroplasty. Clin Orthop Relat Res. 1996;331:159–163.PubMed Ries MD, Weaver K, Beals N. Safety and efficacy of ethylene oxide sterilized polyethylene in total knee arthroplasty. Clin Orthop Relat Res. 1996;331:159–163.PubMed
118.
Zurück zum Zitat Ries MD, Weaver K, Rose RM, Gunther J, Sauer W, Beals N. Fatigue strength of polyethylene after sterilization by gamma irradiation or ethylene oxide. Clin Orthop Relat Res. 1996;333:87–95.PubMed Ries MD, Weaver K, Rose RM, Gunther J, Sauer W, Beals N. Fatigue strength of polyethylene after sterilization by gamma irradiation or ethylene oxide. Clin Orthop Relat Res. 1996;333:87–95.PubMed
119.
Zurück zum Zitat Rimnac CM, Klein RW, Betts F, Wright TM. Post-irradiation aging of ultra-high molecular weight polyethylene. J Bone Joint Surg Am. 1994;76:1052–1056.PubMed Rimnac CM, Klein RW, Betts F, Wright TM. Post-irradiation aging of ultra-high molecular weight polyethylene. J Bone Joint Surg Am. 1994;76:1052–1056.PubMed
120.
Zurück zum Zitat Ring PA. Complete replacement arthroplasty of the hip by the ring prosthesis. J Bone Joint Surg Br. 1968;50:720–731.PubMed Ring PA. Complete replacement arthroplasty of the hip by the ring prosthesis. J Bone Joint Surg Br. 1968;50:720–731.PubMed
121.
Zurück zum Zitat Rohrl S, Nivbrant B, Mingguo L, Hewitt B. In vivo wear and migration of highly cross-linked polyethylene cups a radiostereometry analysis study. J Arthroplasty. 2005;20:409–413.PubMed Rohrl S, Nivbrant B, Mingguo L, Hewitt B. In vivo wear and migration of highly cross-linked polyethylene cups a radiostereometry analysis study. J Arthroplasty. 2005;20:409–413.PubMed
122.
Zurück zum Zitat Rohrl SM, Li MG, Nilsson KG, Nivbrant B. Very low wear of non-remelted highly cross-linked polyethylene cups: an RSA study lasting up to 6 years. Acta Orthop. 2007;78:739–745.PubMed Rohrl SM, Li MG, Nilsson KG, Nivbrant B. Very low wear of non-remelted highly cross-linked polyethylene cups: an RSA study lasting up to 6 years. Acta Orthop. 2007;78:739–745.PubMed
123.
Zurück zum Zitat Schmalzried TP, Jasty M, Harris WH. Periprosthetic bone loss in total hip arthroplasty: the concept of effective joint space. J Bone Joint Surg Am. 1992;74:849–863.PubMed Schmalzried TP, Jasty M, Harris WH. Periprosthetic bone loss in total hip arthroplasty: the concept of effective joint space. J Bone Joint Surg Am. 1992;74:849–863.PubMed
124.
Zurück zum Zitat Schmidt MB, Hamilton JV. The effects of calcium stearate on the properties of UHMWPE. Trans Orthop Res Soc. 1996;21:22. Schmidt MB, Hamilton JV. The effects of calcium stearate on the properties of UHMWPE. Trans Orthop Res Soc. 1996;21:22.
125.
Zurück zum Zitat Scott DL, Campbell PA, McClung CD, Schmalzried TP. Factors contributing to rapid wear and osteolysis in hips with modular acetabular bearings made of hylamer. J Arthroplasty. 2000;15:35–46.PubMed Scott DL, Campbell PA, McClung CD, Schmalzried TP. Factors contributing to rapid wear and osteolysis in hips with modular acetabular bearings made of hylamer. J Arthroplasty. 2000;15:35–46.PubMed
126.
Zurück zum Zitat Sedel L, Kerboull L, Christel P, Meunier A, Witvoet J. Alumina-on-alumina hip replacement: results and survivorship in young patients. J Bone Joint Surg Br. 1990;72:658–663.PubMed Sedel L, Kerboull L, Christel P, Meunier A, Witvoet J. Alumina-on-alumina hip replacement: results and survivorship in young patients. J Bone Joint Surg Br. 1990;72:658–663.PubMed
127.
Zurück zum Zitat Shia DS, Clohisy JC, Schinsky MF, Martell JM, Maloney WJ. THA with highly cross-linked polyethylene in patients 50 years or younger. Clin Orthop Relat Res. 2009;467:2059–2065.PubMed Shia DS, Clohisy JC, Schinsky MF, Martell JM, Maloney WJ. THA with highly cross-linked polyethylene in patients 50 years or younger. Clin Orthop Relat Res. 2009;467:2059–2065.PubMed
128.
Zurück zum Zitat Streicher RM. Ionizing irradiation for sterilization and modification of high molecular weight polyethylenes. Plast Rubber Proc Appl. 1988;10:221–229. Streicher RM. Ionizing irradiation for sterilization and modification of high molecular weight polyethylenes. Plast Rubber Proc Appl. 1988;10:221–229.
129.
Zurück zum Zitat Sutula LC, Collier JP, Saum KA, Currier BH, Currier JH, Sanford WM, Mayor MB, Wooding RE, Sperling DK, Williams IR, Kasprzack DJ, Suprenant VA. Impact of gamma sterilization on clinical performance of polyethylene in the hip. Clin Orthop Relat Res. 1995;319:28–40.PubMed Sutula LC, Collier JP, Saum KA, Currier BH, Currier JH, Sanford WM, Mayor MB, Wooding RE, Sperling DK, Williams IR, Kasprzack DJ, Suprenant VA. Impact of gamma sterilization on clinical performance of polyethylene in the hip. Clin Orthop Relat Res. 1995;319:28–40.PubMed
130.
Zurück zum Zitat Swarts D, Gsell R, King R, Devanathan D, Wallace S, Lin S, Rohr W. Aging of calcium stearate-free polyethylene. In: Transactions of the 5th World Biomaterials Congress. Mount Laurel, NJ: Society for Biomaterials; 1996:196. Swarts D, Gsell R, King R, Devanathan D, Wallace S, Lin S, Rohr W. Aging of calcium stearate-free polyethylene. In: Transactions of the 5th World Biomaterials Congress. Mount Laurel, NJ: Society for Biomaterials; 1996:196.
131.
Zurück zum Zitat Tower SS, Currier JH, Currier BH, Lyford KA, Van Citters DW, Mayor MB. Rim cracking of the cross-linked longevity polyethylene acetabular liner after total hip arthroplasty. J Bone Joint Surg Am. 2007;89:2212–2217.PubMed Tower SS, Currier JH, Currier BH, Lyford KA, Van Citters DW, Mayor MB. Rim cracking of the cross-linked longevity polyethylene acetabular liner after total hip arthroplasty. J Bone Joint Surg Am. 2007;89:2212–2217.PubMed
132.
Zurück zum Zitat Triclot P, Grosjean G, El Masri F, Courpied JP, Hamadouche M. A comparison of the penetration rate of two polyethylene acetabular liners of different levels of cross-linking: a prospective randomised trial. J Bone Joint Surg Br. 2007;89:1439–1445.PubMed Triclot P, Grosjean G, El Masri F, Courpied JP, Hamadouche M. A comparison of the penetration rate of two polyethylene acetabular liners of different levels of cross-linking: a prospective randomised trial. J Bone Joint Surg Br. 2007;89:1439–1445.PubMed
133.
Zurück zum Zitat Wang A, Essner A, Polineni VK, Stark C, Dumbleton JH. Lubrication and wear of ultra-high molecular weight polyethylene in total joint replacements. Tribol Int. 1998;31:17–33. Wang A, Essner A, Polineni VK, Stark C, Dumbleton JH. Lubrication and wear of ultra-high molecular weight polyethylene in total joint replacements. Tribol Int. 1998;31:17–33.
134.
Zurück zum Zitat Wang A, Manley M, Serekian P. Wear and structural fatigue simulation of crosslinked ultra-high molecular weight polyethylene for hip and knee bearing applications. In: Kurtz SM, Gsell R, Martell J, eds. ASTM STP 1445. Highly Crosslinked and Thermally Treated Ultra-High Molecular Weight Polyethylene For Joint Replacements. West Conshohocken, PA: American Society for Testing and Materials; 2004. Wang A, Manley M, Serekian P. Wear and structural fatigue simulation of crosslinked ultra-high molecular weight polyethylene for hip and knee bearing applications. In: Kurtz SM, Gsell R, Martell J, eds. ASTM STP 1445. Highly Crosslinked and Thermally Treated Ultra-High Molecular Weight Polyethylene For Joint Replacements. West Conshohocken, PA: American Society for Testing and Materials; 2004.
135.
Zurück zum Zitat Waugh W. John Charnley: The Man and the Hip. London, UK: Springer-Verlag; 1990. Waugh W. John Charnley: The Man and the Hip. London, UK: Springer-Verlag; 1990.
136.
Zurück zum Zitat White SE, Paxson RD, Tanner MG, Whiteside LA. Effects of sterilization on wear in total knee arthroplasty. Clin Orthop Relat Res. 1996;331:164–171.PubMed White SE, Paxson RD, Tanner MG, Whiteside LA. Effects of sterilization on wear in total knee arthroplasty. Clin Orthop Relat Res. 1996;331:164–171.PubMed
137.
Zurück zum Zitat Whittaker JP, Charron KD, McCalden RW, Macdonald SJ, Bourne RB. Comparison of steady state femoral head penetration rates between two highly cross-linked polyethylenes in total hip arthroplasty. J Arthroplasty. 2010;25:680–686.PubMed Whittaker JP, Charron KD, McCalden RW, Macdonald SJ, Bourne RB. Comparison of steady state femoral head penetration rates between two highly cross-linked polyethylenes in total hip arthroplasty. J Arthroplasty. 2010;25:680–686.PubMed
138.
Zurück zum Zitat Willert HG, Bertram H, Buchhorn GH. Osteolysis in alloarthroplasty of the hip: the role of ultra-high molecular weight polyethylene wear particles. Clin Orthop Relat Res. 1990;258:95–107.PubMed Willert HG, Bertram H, Buchhorn GH. Osteolysis in alloarthroplasty of the hip: the role of ultra-high molecular weight polyethylene wear particles. Clin Orthop Relat Res. 1990;258:95–107.PubMed
139.
Zurück zum Zitat Willert HG, Buchhorn GH, Eyerer P. Ultra-high Molecular Weight Polyethlene as Biomaterial in Orthopedic Surgery. Toronto, Canada: Hogrefe & Huber Publishers; 1991. Willert HG, Buchhorn GH, Eyerer P. Ultra-high Molecular Weight Polyethlene as Biomaterial in Orthopedic Surgery. Toronto, Canada: Hogrefe & Huber Publishers; 1991.
140.
Zurück zum Zitat Williams IR, Mayor MB, Collier JP. The impact of sterilization method on wear in knee arthroplasty. Clin Orthop Relat Res. 1998;356:170–180.PubMed Williams IR, Mayor MB, Collier JP. The impact of sterilization method on wear in knee arthroplasty. Clin Orthop Relat Res. 1998;356:170–180.PubMed
141.
Zurück zum Zitat Winter M, Griss P, Scheller G, Moser T. Ten- to 14-year results of a ceramic hip prosthesis. Clin Orthop Relat Res. 1992;282:73–80.PubMed Winter M, Griss P, Scheller G, Moser T. Ten- to 14-year results of a ceramic hip prosthesis. Clin Orthop Relat Res. 1992;282:73–80.PubMed
142.
Zurück zum Zitat Wright TM, Bartel DL. The problem of surface damage in polyethylene total knee components. Clin Orthop Relat Res. 1986;205:67–74.PubMed Wright TM, Bartel DL. The problem of surface damage in polyethylene total knee components. Clin Orthop Relat Res. 1986;205:67–74.PubMed
143.
Zurück zum Zitat Wright TM, Fukubayashi T, Burstein AH. The effect of carbon fiber reinforcement on contact area, contact pressure, and time-dependent deformation in polyethylene tibial components. J Biomed Mater Res. 1981;15:719–730.PubMed Wright TM, Fukubayashi T, Burstein AH. The effect of carbon fiber reinforcement on contact area, contact pressure, and time-dependent deformation in polyethylene tibial components. J Biomed Mater Res. 1981;15:719–730.PubMed
144.
Zurück zum Zitat Wroblewski BM, Siney PD, Dowson D, Collins SN. Prospective clinical and joint simulator studies of a new total hip arthroplasty using alumina ceramic heads and cross-linked polyethylene cups. J Bone Joint Surg Br. 1996;78:280–285.PubMed Wroblewski BM, Siney PD, Dowson D, Collins SN. Prospective clinical and joint simulator studies of a new total hip arthroplasty using alumina ceramic heads and cross-linked polyethylene cups. J Bone Joint Surg Br. 1996;78:280–285.PubMed
145.
Zurück zum Zitat Wrona M, Mayor MB, Collier JP, Jensen RE. The correlation between fusion defects and damage in tibial polyethylene bearings. Clin Orthop Relat Res. 1994;299:92–103.PubMed Wrona M, Mayor MB, Collier JP, Jensen RE. The correlation between fusion defects and damage in tibial polyethylene bearings. Clin Orthop Relat Res. 1994;299:92–103.PubMed
146.
Zurück zum Zitat Yao JQ, Lu MP, Johnson TS, Gilbertson LN, Swarts D, Blanchard CR, Crowninshield RD. Improved resistance to wear, delamination and posterior loading fatigue damage of electron beam irradiated, melt-annealed, highly crosslinked UHMWPE knee inserts. In: Kurtz SM, Gsell R, Martell J, eds. ASTM STP 1445. Highly Crosslinked and Thermally Treated Ultra-High Molecular Weight Polyethylene For Joint Replacements. West Conshohocken, PA: American Society for Testing and Materials; 2004. Yao JQ, Lu MP, Johnson TS, Gilbertson LN, Swarts D, Blanchard CR, Crowninshield RD. Improved resistance to wear, delamination and posterior loading fatigue damage of electron beam irradiated, melt-annealed, highly crosslinked UHMWPE knee inserts. In: Kurtz SM, Gsell R, Martell J, eds. ASTM STP 1445. Highly Crosslinked and Thermally Treated Ultra-High Molecular Weight Polyethylene For Joint Replacements. West Conshohocken, PA: American Society for Testing and Materials; 2004.
Metadaten
Titel
History and Systematic Review of Wear and Osteolysis Outcomes for First-generation Highly Crosslinked Polyethylene
verfasst von
Steven M. Kurtz, PhD
Heather A. Gawel, MS
Jasmine D. Patel, PhD
Publikationsdatum
01.08.2011
Verlag
Springer-Verlag
Erschienen in
Clinical Orthopaedics and Related Research® / Ausgabe 8/2011
Print ISSN: 0009-921X
Elektronische ISSN: 1528-1132
DOI
https://doi.org/10.1007/s11999-011-1872-4

Weitere Artikel der Ausgabe 8/2011

Clinical Orthopaedics and Related Research® 8/2011 Zur Ausgabe

Symposium: Bone Quality: From Bench to Bedside

Raman Assessment of Bone Quality

Symposium: Bone Quality: From Bench to Bedside

Infrared Assessment of Bone Quality: A Review

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.