Highly Crosslinked Polyethylene Does Not Reduce Aseptic Loosening in Cemented THA 10-year Findings of a Randomized Study

verfasst von: Per-Erik Johanson, MD, Georgios Digas, MD, PhD, Peter Herberts, MD, PhD, Jonas Thanner, MD, PhD, Johan Kärrholm, MD, PhD

Erschienen in: Clinical Orthopaedics and Related Research® | Ausgabe 11/2012

Einloggen, um Zugang zu erhalten

Abstract

Background

Polyethylene (PE) wear particles are believed to cause aseptic loosening and thereby impair function in hip arthroplasty. Highly crosslinked polyethylene (XLPE) has low short- and medium-term wear rates. However, the long-term wear characteristics are unknown and it is unclear whether reduced wear particle burden improves function and survival of cemented hip arthroplasty.

Questions/purposes

We asked whether XLPE wear rates remain low up to 10 years and whether this leads to improved implant fixation, periprosthetic bone quality, and clinical function compared to conventional PE.

Methods

We randomized 60 patients (61 hips) to receive either PE or XLPE cemented cups combined with a cemented stem. At 10 years postoperatively, 51 patients (52 hips) were evaluated for polyethylene wear and component migration estimation by radiostereometry, for radiolucent lines, bone densitometry, and Harris hip and pain scores. Revisions were recorded.

Results

XLPE cups had a lower mean three-dimensional wear rate between 2 and 10 years compared to conventional PE hips: 0.005 mm/year versus 0.056 mm/year. We found no differences in cup migration, bone mineral density, radiolucencies, functional scores, and revision rate. There was a trend toward improved stem fixation in the XLPE group. The overall stem failure rate was comparably high, without influencing wear rate in XLPE hips.

Conclusions

XLPE displayed a low wear rate up to 10 years when used in cemented THA, but we found no clear benefits in any other parameters. Further research is needed to determine whether cemented THA designs with XLPE are less prone to stem loosening.

Level of Evidence

Level I, therapeutic study. See the Instructions for Authors for a complete description of levels of evidence.
Anhänge
Nur mit Berechtigung zugänglich
Literatur
1.
Zurück zum Zitat Baxter RM, Ianuzzi A, Freeman TA, Kurtz SM, Steinbeck MJ. Distinct immunohistomorphologic changes in periprosthetic hip tissues from historical and highly crosslinked UHMWPE implant retrievals. J Biomed Mater Res A. 2010;95:68–78.PubMed Baxter RM, Ianuzzi A, Freeman TA, Kurtz SM, Steinbeck MJ. Distinct immunohistomorphologic changes in periprosthetic hip tissues from historical and highly crosslinked UHMWPE implant retrievals. J Biomed Mater Res A. 2010;95:68–78.PubMed
2.
Zurück zum Zitat Bitsch RG, Loidolt T, Heisel C, Ball S, Schmalzried TP. Reduction of osteolysis with use of Marathon cross-linked polyethylene: a concise follow-up, at a minimum of five years, of a previous report. J Bone Joint Surg Am. 2008;90:1487–1491.PubMedCrossRef Bitsch RG, Loidolt T, Heisel C, Ball S, Schmalzried TP. Reduction of osteolysis with use of Marathon cross-linked polyethylene: a concise follow-up, at a minimum of five years, of a previous report. J Bone Joint Surg Am. 2008;90:1487–1491.PubMedCrossRef
3.
Zurück zum Zitat Boss JH, Misselevich I, Behar J, Mendes DG. Histologic analysis of the periprosthetic tissues of long-term surviving cemented total hip arthroplasties. J Long Term Eff Med Implants. 1996;6:73–90.PubMed Boss JH, Misselevich I, Behar J, Mendes DG. Histologic analysis of the periprosthetic tissues of long-term surviving cemented total hip arthroplasties. J Long Term Eff Med Implants. 1996;6:73–90.PubMed
4.
Zurück zum Zitat Callaghan JJ, Cuckler JM, Huddleston JI, Galante JO. How have alternative bearings (such as metal-on-metal, highly cross-linked polyethylene, and ceramic-on-ceramic) affected the prevention and treatment of osteolysis? J Am Acad Orthop Surg. 2008;16(suppl 1):S33–S38.PubMed Callaghan JJ, Cuckler JM, Huddleston JI, Galante JO. How have alternative bearings (such as metal-on-metal, highly cross-linked polyethylene, and ceramic-on-ceramic) affected the prevention and treatment of osteolysis? J Am Acad Orthop Surg. 2008;16(suppl 1):S33–S38.PubMed
5.
Zurück zum Zitat Capello WN, D’Antonio JA, Ramakrishnan R, Naughton M. Continued improved wear with an annealed highly cross-linked polyethylene. Clin Orthop Relat Res. 2011;469:825–830.PubMedCrossRef Capello WN, D’Antonio JA, Ramakrishnan R, Naughton M. Continued improved wear with an annealed highly cross-linked polyethylene. Clin Orthop Relat Res. 2011;469:825–830.PubMedCrossRef
6.
Zurück zum Zitat Catelas I, Wimmer MA, Utzschneider S. Polyethylene and metal wear particles: characteristics and biological effects. Semin Immunopathol. 2011;33:257–271.PubMedCrossRef Catelas I, Wimmer MA, Utzschneider S. Polyethylene and metal wear particles: characteristics and biological effects. Semin Immunopathol. 2011;33:257–271.PubMedCrossRef
7.
Zurück zum Zitat DeLee JG, Charnley J. Radiological demarcation of cemented sockets in total hip replacement. Clin Orthop Relat Res. 1976;121:20–32.PubMed DeLee JG, Charnley J. Radiological demarcation of cemented sockets in total hip replacement. Clin Orthop Relat Res. 1976;121:20–32.PubMed
8.
Zurück zum Zitat Digas G, Karrholm J, Olofsson K. Five-year DEXA study of 88 hips with cemented femoral stem: influence of design variations on early migration of a cemented stem in THA. Int Orthop. 2009;33:1495–1500.PubMedCrossRef Digas G, Karrholm J, Olofsson K. Five-year DEXA study of 88 hips with cemented femoral stem: influence of design variations on early migration of a cemented stem in THA. Int Orthop. 2009;33:1495–1500.PubMedCrossRef
9.
Zurück zum Zitat Digas G, Karrholm J, Thanner J. Different loss of BMD using uncemented press-fit and whole polyethylene cups fixed with cement: repeated DXA studies in 96 hips randomized to 3 types of fixation. Acta Orthop. 2006;77:218–226.PubMedCrossRef Digas G, Karrholm J, Thanner J. Different loss of BMD using uncemented press-fit and whole polyethylene cups fixed with cement: repeated DXA studies in 96 hips randomized to 3 types of fixation. Acta Orthop. 2006;77:218–226.PubMedCrossRef
10.
Zurück zum Zitat Digas G, Karrholm J, Thanner J, Herberts P. 5-year experience of highly cross-linked polyethylene in cemented and uncemented sockets: two randomized studies using radiostereometric analysis. Acta Orthop. 2007;78:746–754.PubMedCrossRef Digas G, Karrholm J, Thanner J, Herberts P. 5-year experience of highly cross-linked polyethylene in cemented and uncemented sockets: two randomized studies using radiostereometric analysis. Acta Orthop. 2007;78:746–754.PubMedCrossRef
11.
Zurück zum Zitat Digas G, Karrholm J, Thanner J, Malchau H, Herberts P. Highly cross-linked polyethylene in cemented THA: randomized study of 61 hips. Clin Orthop Relat Res. 2003;417:126–138.PubMed Digas G, Karrholm J, Thanner J, Malchau H, Herberts P. Highly cross-linked polyethylene in cemented THA: randomized study of 61 hips. Clin Orthop Relat Res. 2003;417:126–138.PubMed
12.
Zurück zum Zitat Digas G, Karrholm J, Thanner J, Malchau H, Herberts P. The Otto Aufranc Award. Highly cross-linked polyethylene in total hip arthroplasty: randomized evaluation of penetration rate in cemented and uncemented sockets using radiostereometric analysis. Clin Orthop Relat Res. 2004;429:6–16.PubMedCrossRef Digas G, Karrholm J, Thanner J, Malchau H, Herberts P. The Otto Aufranc Award. Highly cross-linked polyethylene in total hip arthroplasty: randomized evaluation of penetration rate in cemented and uncemented sockets using radiostereometric analysis. Clin Orthop Relat Res. 2004;429:6–16.PubMedCrossRef
13.
Zurück zum Zitat Dorr LD, Wan Z. Ten years of experience with porous acetabular components for revision surgery. Clin Orthop Relat Res. 1995;319:191–200.PubMed Dorr LD, Wan Z. Ten years of experience with porous acetabular components for revision surgery. Clin Orthop Relat Res. 1995;319:191–200.PubMed
14.
Zurück zum Zitat Dumbleton JH, Manley MT, Edidin AA. A literature review of the association between wear rate and osteolysis in total hip arthroplasty. J Arthroplasty. 2002;17:649–661.PubMedCrossRef Dumbleton JH, Manley MT, Edidin AA. A literature review of the association between wear rate and osteolysis in total hip arthroplasty. J Arthroplasty. 2002;17:649–661.PubMedCrossRef
15.
Zurück zum Zitat Espehaug B, Furnes O, Engesaeter LB, Havelin LI. 18 years of results with cemented primary hip prostheses in the Norwegian Arthroplasty Register: concerns about some newer implants. Acta Orthop. 2009;80:402–412.PubMedCrossRef Espehaug B, Furnes O, Engesaeter LB, Havelin LI. 18 years of results with cemented primary hip prostheses in the Norwegian Arthroplasty Register: concerns about some newer implants. Acta Orthop. 2009;80:402–412.PubMedCrossRef
16.
Zurück zum Zitat Fukui K, Kaneuji A, Sugimori T, Ichiseki T, Kitamura K, Matsumoto T. Wear comparison between a highly cross-linked polyethylene and conventional polyethylene against a zirconia femoral head: minimum 5-year follow-up. J Arthroplasty. 2011;26:45–49.PubMedCrossRef Fukui K, Kaneuji A, Sugimori T, Ichiseki T, Kitamura K, Matsumoto T. Wear comparison between a highly cross-linked polyethylene and conventional polyethylene against a zirconia femoral head: minimum 5-year follow-up. J Arthroplasty. 2011;26:45–49.PubMedCrossRef
17.
Zurück zum Zitat Garcia-Rey E, Garcia-Cimbrelo E, Cruz-Pardos A, Ortega-Chamarro J. New polyethylenes in total hip replacement: a prospective, comparative clinical study of two types of liner. J Bone Joint Surg Br. 2008;90:149–153.PubMedCrossRef Garcia-Rey E, Garcia-Cimbrelo E, Cruz-Pardos A, Ortega-Chamarro J. New polyethylenes in total hip replacement: a prospective, comparative clinical study of two types of liner. J Bone Joint Surg Br. 2008;90:149–153.PubMedCrossRef
18.
Zurück zum Zitat Glyn-Jones S, McLardy-Smith P, Gill HS, Murray DW. The creep and wear of highly cross-linked polyethylene: a three-year randomised, controlled trial using radiostereometric analysis. J Bone Joint Surg Br. 2008;90:556–561.PubMedCrossRef Glyn-Jones S, McLardy-Smith P, Gill HS, Murray DW. The creep and wear of highly cross-linked polyethylene: a three-year randomised, controlled trial using radiostereometric analysis. J Bone Joint Surg Br. 2008;90:556–561.PubMedCrossRef
19.
Zurück zum Zitat Grose A, Gonzalez Della Valle A, Bullough P, Lyman S, Tomek I, Pellicci P. High failure rate of a modern, proximally roughened, cemented stem for total hip arthroplasty. Int Orthop. 2006;30:243–247.PubMedCrossRef Grose A, Gonzalez Della Valle A, Bullough P, Lyman S, Tomek I, Pellicci P. High failure rate of a modern, proximally roughened, cemented stem for total hip arthroplasty. Int Orthop. 2006;30:243–247.PubMedCrossRef
20.
Zurück zum Zitat Gruen TA, McNeice GM, Amstutz HC. “Modes of failure” of cemented stem-type femoral components: a radiographic analysis of loosening. Clin Orthop Relat Res. 1979;141:17–27.PubMed Gruen TA, McNeice GM, Amstutz HC. “Modes of failure” of cemented stem-type femoral components: a radiographic analysis of loosening. Clin Orthop Relat Res. 1979;141:17–27.PubMed
21.
Zurück zum Zitat Harris WH. Traumatic arthritis of the hip after dislocation and acetabular fractures: treatment by mold arthroplasty. An end-result study using a new method of result evaluation. J Bone Joint Surg Am. 1969;51:737–755.PubMed Harris WH. Traumatic arthritis of the hip after dislocation and acetabular fractures: treatment by mold arthroplasty. An end-result study using a new method of result evaluation. J Bone Joint Surg Am. 1969;51:737–755.PubMed
22.
Zurück zum Zitat Herberts P, Malchau H. Long-term registration has improved the quality of hip replacement: a review of the Swedish THR Register comparing 160,000 cases. Acta Orthop Scand. 2000;71:111–121.PubMedCrossRef Herberts P, Malchau H. Long-term registration has improved the quality of hip replacement: a review of the Swedish THR Register comparing 160,000 cases. Acta Orthop Scand. 2000;71:111–121.PubMedCrossRef
23.
Zurück zum Zitat Illgen RL 2nd, Bauer LM, Hotujec BT, Kolpin SE, Bakhtiar A, Forsythe TM. Highly crosslinked vs conventional polyethylene particles: relative in vivo inflammatory response. J Arthroplasty. 2009;24:117–124.PubMedCrossRef Illgen RL 2nd, Bauer LM, Hotujec BT, Kolpin SE, Bakhtiar A, Forsythe TM. Highly crosslinked vs conventional polyethylene particles: relative in vivo inflammatory response. J Arthroplasty. 2009;24:117–124.PubMedCrossRef
24.
Zurück zum Zitat Illgen RL 2nd, Forsythe TM, Pike JW, Laurent MP, Blanchard CR. Highly crosslinked vs conventional polyethylene particles—an in vitro comparison of biologic activities. J Arthroplasty. 2008;23:721–731.PubMedCrossRef Illgen RL 2nd, Forsythe TM, Pike JW, Laurent MP, Blanchard CR. Highly crosslinked vs conventional polyethylene particles—an in vitro comparison of biologic activities. J Arthroplasty. 2008;23:721–731.PubMedCrossRef
25.
Zurück zum Zitat Ingham E, Fisher J. The role of macrophages in osteolysis of total joint replacement. Biomaterials. 2005;26:1271–1286.PubMedCrossRef Ingham E, Fisher J. The role of macrophages in osteolysis of total joint replacement. Biomaterials. 2005;26:1271–1286.PubMedCrossRef
26.
Zurück zum Zitat Ingham E, Fisher J, Rubash HE, Sinha RK, Shanbhag AS, Kim SY. Biological reactions to wear debris in total joint replacement: pathogenesis of bone loss after total hip arthroplasty. Proc Inst Mech Eng H. 2000;214:21–37.PubMed Ingham E, Fisher J, Rubash HE, Sinha RK, Shanbhag AS, Kim SY. Biological reactions to wear debris in total joint replacement: pathogenesis of bone loss after total hip arthroplasty. Proc Inst Mech Eng H. 2000;214:21–37.PubMed
27.
Zurück zum Zitat Ishiguro N, Kojima T, Ito T, Saga S, Anma H, Kurokouchi K, Iwahori Y, Iwase T, Iwata H. Macrophage activation and migration in interface tissue around loosening total hip arthroplasty components. J Biomed Mater Res. 1997;35:399–406.PubMedCrossRef Ishiguro N, Kojima T, Ito T, Saga S, Anma H, Kurokouchi K, Iwahori Y, Iwase T, Iwata H. Macrophage activation and migration in interface tissue around loosening total hip arthroplasty components. J Biomed Mater Res. 1997;35:399–406.PubMedCrossRef
28.
Zurück zum Zitat Jasty M, Maloney WJ, Bragdon CR, O’Connor DO, Haire T, Harris WH. The initiation of failure in cemented femoral components of hip arthroplasties. J Bone Joint Surg Br. 1991;73:551–558.PubMed Jasty M, Maloney WJ, Bragdon CR, O’Connor DO, Haire T, Harris WH. The initiation of failure in cemented femoral components of hip arthroplasties. J Bone Joint Surg Br. 1991;73:551–558.PubMed
29.
Zurück zum Zitat Kärrholm J, Razaznejad R. Fixation and bone remodeling around a low stiffness stem in revision surgery. Clin Orthop Relat Res. 2008;466:380–388.PubMedCrossRef Kärrholm J, Razaznejad R. Fixation and bone remodeling around a low stiffness stem in revision surgery. Clin Orthop Relat Res. 2008;466:380–388.PubMedCrossRef
30.
Zurück zum Zitat Kurtz SM, Gawel HA, Patel JD. History and systematic review of wear and osteolysis outcomes for first-generation highly crosslinked polyethylene. Clin Orthop Relat Res. 2011;469:2262–2277.PubMedCrossRef Kurtz SM, Gawel HA, Patel JD. History and systematic review of wear and osteolysis outcomes for first-generation highly crosslinked polyethylene. Clin Orthop Relat Res. 2011;469:2262–2277.PubMedCrossRef
31.
Zurück zum Zitat Leung SB, Egawa H, Stepniewski A, Beykirch S, Engh CA Jr, Engh CA Sr. Incidence and volume of pelvic osteolysis at early follow-up with highly cross-linked and noncross-linked polyethylene. J Arthroplasty. 2007;22:134–139.PubMedCrossRef Leung SB, Egawa H, Stepniewski A, Beykirch S, Engh CA Jr, Engh CA Sr. Incidence and volume of pelvic osteolysis at early follow-up with highly cross-linked and noncross-linked polyethylene. J Arthroplasty. 2007;22:134–139.PubMedCrossRef
32.
Zurück zum Zitat Livermore J, Ilstrup D, Morrey B. Effect of femoral head size on wear of the polyethylene acetabular component. J Bone Joint Surg Am. 1990;72:518–528.PubMed Livermore J, Ilstrup D, Morrey B. Effect of femoral head size on wear of the polyethylene acetabular component. J Bone Joint Surg Am. 1990;72:518–528.PubMed
33.
Zurück zum Zitat Martell JM, Berdia S. Determination of polyethylene wear in total hip replacements with use of digital radiographs. J Bone Joint Surg Am. 1997;79:1635–1641.PubMed Martell JM, Berdia S. Determination of polyethylene wear in total hip replacements with use of digital radiographs. J Bone Joint Surg Am. 1997;79:1635–1641.PubMed
34.
Zurück zum Zitat McCalden RW, MacDonald SJ, Rorabeck CH, Bourne RB, Chess DG, Charron KD. Wear rate of highly cross-linked polyethylene in total hip arthroplasty: a randomized controlled trial. J Bone Joint Surg Am. 2009;91:773–782.PubMedCrossRef McCalden RW, MacDonald SJ, Rorabeck CH, Bourne RB, Chess DG, Charron KD. Wear rate of highly cross-linked polyethylene in total hip arthroplasty: a randomized controlled trial. J Bone Joint Surg Am. 2009;91:773–782.PubMedCrossRef
35.
Zurück zum Zitat McKellop H, Shen FW, Lu B, Campbell P, Salovey R. Development of an extremely wear-resistant ultra high molecular weight polyethylene for total hip replacements. J Orthop Res. 1999;17:157–167.PubMedCrossRef McKellop H, Shen FW, Lu B, Campbell P, Salovey R. Development of an extremely wear-resistant ultra high molecular weight polyethylene for total hip replacements. J Orthop Res. 1999;17:157–167.PubMedCrossRef
36.
Zurück zum Zitat Mjoberg B. Theories of wear and loosening in hip prostheses: wear-induced loosening vs loosening-induced wear—a review. Acta Orthop Scand. 1994;65:361–371.PubMedCrossRef Mjoberg B. Theories of wear and loosening in hip prostheses: wear-induced loosening vs loosening-induced wear—a review. Acta Orthop Scand. 1994;65:361–371.PubMedCrossRef
37.
Zurück zum Zitat Murray DW, Rushton N. Macrophages stimulate bone resorption when they phagocytose particles. J Bone Joint Surg Br. 1990;72:988–992.PubMed Murray DW, Rushton N. Macrophages stimulate bone resorption when they phagocytose particles. J Bone Joint Surg Br. 1990;72:988–992.PubMed
38.
Zurück zum Zitat Mutimer J, Devane PA, Adams K, Horne JG. Highly crosslinked polyethylene reduces wear in total hip arthroplasty at 5 years. Clin Orthop Relat Res. 2010;468:3228–3233.PubMedCrossRef Mutimer J, Devane PA, Adams K, Horne JG. Highly crosslinked polyethylene reduces wear in total hip arthroplasty at 5 years. Clin Orthop Relat Res. 2010;468:3228–3233.PubMedCrossRef
39.
Zurück zum Zitat Olyslaegers C, Defoort K, Simon JP, Vandenberghe L. Wear in conventional and highly cross-linked polyethylene cups: a 5-year follow-up study. J Arthroplasty. 2008;23:489–494.PubMedCrossRef Olyslaegers C, Defoort K, Simon JP, Vandenberghe L. Wear in conventional and highly cross-linked polyethylene cups: a 5-year follow-up study. J Arthroplasty. 2008;23:489–494.PubMedCrossRef
40.
Zurück zum Zitat Oparaugo PC, Clarke IC, Malchau H, Herberts P. Correlation of wear debris-induced osteolysis and revision with volumetric wear-rates of polyethylene: a survey of 8 reports in the literature. Acta Orthop Scand. 2001;72:22–28.PubMedCrossRef Oparaugo PC, Clarke IC, Malchau H, Herberts P. Correlation of wear debris-induced osteolysis and revision with volumetric wear-rates of polyethylene: a survey of 8 reports in the literature. Acta Orthop Scand. 2001;72:22–28.PubMedCrossRef
41.
Zurück zum Zitat Periasamy K, Watson WS, Mohammed A, Murray H, Walker B, Patil S, Meek RM. A randomised study of peri-prosthetic bone density after cemented versus trabecular fixation of a polyethylene acetabular component. J Bone Joint Surg Br. 2011;93:1033–1044.PubMed Periasamy K, Watson WS, Mohammed A, Murray H, Walker B, Patil S, Meek RM. A randomised study of peri-prosthetic bone density after cemented versus trabecular fixation of a polyethylene acetabular component. J Bone Joint Surg Br. 2011;93:1033–1044.PubMed
42.
Zurück zum Zitat Purdue PE, Koulouvaris P, Nestor BJ, Sculco TP. The central role of wear debris in periprosthetic osteolysis. HSS J. 2006;2:102–113.PubMedCrossRef Purdue PE, Koulouvaris P, Nestor BJ, Sculco TP. The central role of wear debris in periprosthetic osteolysis. HSS J. 2006;2:102–113.PubMedCrossRef
43.
Zurück zum Zitat Rajadhyaksha AD, Brotea C, Cheung Y, Kuhn C, Ramakrishnan R, Zelicof SB. Five-year comparative study of highly cross-linked (Crossfire) and traditional polyethylene. J Arthroplasty. 2009;24:161–167.PubMedCrossRef Rajadhyaksha AD, Brotea C, Cheung Y, Kuhn C, Ramakrishnan R, Zelicof SB. Five-year comparative study of highly cross-linked (Crossfire) and traditional polyethylene. J Arthroplasty. 2009;24:161–167.PubMedCrossRef
44.
Zurück zum Zitat Ranstam J, Ryd L, Onsten I. Erratum: Accurate accuracy assessment: review of basic principles (Acta Orthopaedica Scandinavica (1999) 70 4 (319-321)). Acta Orthop Scand. 2000;71:106–108.PubMedCrossRef Ranstam J, Ryd L, Onsten I. Erratum: Accurate accuracy assessment: review of basic principles (Acta Orthopaedica Scandinavica (1999) 70 4 (319-321)). Acta Orthop Scand. 2000;71:106–108.PubMedCrossRef
45.
Zurück zum Zitat Ries MD, Pruitt L. Effect of cross-linking on the microstructure and mechanical properties of ultra-high molecular weight polyethylene. Clin Orthop Relat Res. 2005;440:149–156.PubMedCrossRef Ries MD, Pruitt L. Effect of cross-linking on the microstructure and mechanical properties of ultra-high molecular weight polyethylene. Clin Orthop Relat Res. 2005;440:149–156.PubMedCrossRef
46.
Zurück zum Zitat Rohrl SM, Li MG, Nilsson KG, Nivbrant B. Very low wear of non-remelted highly cross-linked polyethylene cups: an RSA study lasting up to 6 years. Acta Orthop. 2007;78:739–745.PubMedCrossRef Rohrl SM, Li MG, Nilsson KG, Nivbrant B. Very low wear of non-remelted highly cross-linked polyethylene cups: an RSA study lasting up to 6 years. Acta Orthop. 2007;78:739–745.PubMedCrossRef
47.
Zurück zum Zitat Rohrl SM, Nivbrant B, Nilsson KG. No adverse effects of submelt-annealed highly crosslinked polyethylene in cemented cups. Acta Orthop. 2012;83:148–152.PubMedCrossRef Rohrl SM, Nivbrant B, Nilsson KG. No adverse effects of submelt-annealed highly crosslinked polyethylene in cemented cups. Acta Orthop. 2012;83:148–152.PubMedCrossRef
48.
Zurück zum Zitat Schmalzried TP, Dorey FJ, McKellop H. The multifactorial nature of polyethylene wear in vivo. J Bone Joint Surg Am. 1998;80:1234–1243.PubMed Schmalzried TP, Dorey FJ, McKellop H. The multifactorial nature of polyethylene wear in vivo. J Bone Joint Surg Am. 1998;80:1234–1243.PubMed
49.
Zurück zum Zitat Schmalzried TP, Kwong LM, Jasty M, Sedlacek RC, Haire TC, O’Connor DO, Bragdon CR, Kabo JM, Malcolm AJ, Harris WH. The mechanism of loosening of cemented acetabular components in total hip arthroplasty: analysis of specimens retrieved at autopsy. Clin Orthop Relat Res. 1992;274:60–78.PubMed Schmalzried TP, Kwong LM, Jasty M, Sedlacek RC, Haire TC, O’Connor DO, Bragdon CR, Kabo JM, Malcolm AJ, Harris WH. The mechanism of loosening of cemented acetabular components in total hip arthroplasty: analysis of specimens retrieved at autopsy. Clin Orthop Relat Res. 1992;274:60–78.PubMed
50.
51.
Zurück zum Zitat Sundfeldt M, Carlsson LV, Johansson CB, Thomsen P, Gretzer C. Aseptic loosening, not only a question of wear: a review of different theories. Acta Orthop. 2006;77:177–197.PubMedCrossRef Sundfeldt M, Carlsson LV, Johansson CB, Thomsen P, Gretzer C. Aseptic loosening, not only a question of wear: a review of different theories. Acta Orthop. 2006;77:177–197.PubMedCrossRef
52.
Zurück zum Zitat Thomas GE, Simpson DJ, Mehmood S, Taylor A, McLardy-Smith P, Gill HS, Murray DW, Glyn-Jones S. The seven-year wear of highly cross-linked polyethylene in total hip arthroplasty: a double-blind, randomized controlled trial using radiostereometric analysis. J Bone Joint Surg Am. 2011;93:716–722.PubMedCrossRef Thomas GE, Simpson DJ, Mehmood S, Taylor A, McLardy-Smith P, Gill HS, Murray DW, Glyn-Jones S. The seven-year wear of highly cross-linked polyethylene in total hip arthroplasty: a double-blind, randomized controlled trial using radiostereometric analysis. J Bone Joint Surg Am. 2011;93:716–722.PubMedCrossRef
53.
Zurück zum Zitat Triclot P, Grosjean G, El Masri F, Courpied JP, Hamadouche M. A comparison of the penetration rate of two polyethylene acetabular liners of different levels of cross-linking: a prospective randomised trial. J Bone Joint Surg Br. 2007;89:1439–1445.PubMedCrossRef Triclot P, Grosjean G, El Masri F, Courpied JP, Hamadouche M. A comparison of the penetration rate of two polyethylene acetabular liners of different levels of cross-linking: a prospective randomised trial. J Bone Joint Surg Br. 2007;89:1439–1445.PubMedCrossRef
54.
Zurück zum Zitat Valstar ER, Gill R, Ryd L, Flivik G, Börlin N, Kärrholm J. Guidelines for standardization of radiostereometry (RSA) of implants. Acta Orthop. 2005;76:563–572.PubMedCrossRef Valstar ER, Gill R, Ryd L, Flivik G, Börlin N, Kärrholm J. Guidelines for standardization of radiostereometry (RSA) of implants. Acta Orthop. 2005;76:563–572.PubMedCrossRef
55.
Zurück zum Zitat Venesmaa PK, Kroger HP, Jurvelin JS, Miettinen HJ, Suomalainen OT, Alhava EM, Yli-Kyyny T, Tamminen I, Syri J, Venesmaa P, Kroger H. Periprosthetic bone loss after cemented total hip arthroplasty: a prospective 5-year dual energy radiographic absorptiometry study of 15 patients. Acta Orthop Scand. 2003;74:31–36.PubMedCrossRef Venesmaa PK, Kroger HP, Jurvelin JS, Miettinen HJ, Suomalainen OT, Alhava EM, Yli-Kyyny T, Tamminen I, Syri J, Venesmaa P, Kroger H. Periprosthetic bone loss after cemented total hip arthroplasty: a prospective 5-year dual energy radiographic absorptiometry study of 15 patients. Acta Orthop Scand. 2003;74:31–36.PubMedCrossRef
56.
Zurück zum Zitat Zhang HY, Blunt L, Jiang XQ, Brown L, Barrans S, Zhao Y. Femoral stem wear in cemented total hip replacement. Proc Inst Mech Eng H. 2008;222:583–592.PubMedCrossRef Zhang HY, Blunt L, Jiang XQ, Brown L, Barrans S, Zhao Y. Femoral stem wear in cemented total hip replacement. Proc Inst Mech Eng H. 2008;222:583–592.PubMedCrossRef
Metadaten
Titel
Highly Crosslinked Polyethylene Does Not Reduce Aseptic Loosening in Cemented THA 10-year Findings of a Randomized Study
verfasst von
Per-Erik Johanson, MD
Georgios Digas, MD, PhD
Peter Herberts, MD, PhD
Jonas Thanner, MD, PhD
Johan Kärrholm, MD, PhD
Publikationsdatum
01.11.2012
Verlag
Springer-Verlag
Erschienen in
Clinical Orthopaedics and Related Research® / Ausgabe 11/2012
Print ISSN: 0009-921X
Elektronische ISSN: 1528-1132
DOI
https://doi.org/10.1007/s11999-012-2400-x

Weitere Artikel der Ausgabe 11/2012

Clinical Orthopaedics and Related Research® 11/2012 Zur Ausgabe

Symposium: Papers Presented at the 2011 Meeting of the International Hip Society

Socket Position Determines Hip Resurfacing 10-Year Survivorship

Symposium: Papers Presented at the 2011 Meeting of the International Hip Society

Intraarticular Abnormalities in Residual Perthes and Perthes-like Hip Deformities

Symposium: Papers Presented at the 2011 Meeting of the International Hip Society

Wear-Corrosion Synergism in a CoCrMo Hip Bearing Alloy Is Influenced by Proteins

Symposium: Papers Presented at the 2011 Meeting of the International Hip Society

Head Material Influences Survival of a Cemented Total Hip Prosthesis in the Norwegian Arthroplasty Register

Symposium: Papers Presented at the 2011 Meeting of the International Hip Society

What Factors Predict Failure 4 to 12 Years After Periacetabular Osteotomy?

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.