Skip to main content
Erschienen in: Clinical Orthopaedics and Related Research® 7/2015

01.07.2015 | Clinical Research

Revision of Metal-on-metal Hip Prostheses Results in Marked Reduction of Blood Cobalt and Chromium Ion Concentrations

verfasst von: Olli Lainiala, BM, Aleksi Reito, MD, Petra Elo, MD, PhD, Jorma Pajamäki, MD, PhD, Timo Puolakka, MD, PhD, Antti Eskelinen, MD, PhD

Erschienen in: Clinical Orthopaedics and Related Research® | Ausgabe 7/2015

Einloggen, um Zugang zu erhalten

Abstract

Background

High revision rates attributable to adverse reactions to metal debris have been reported for total hip arthroplasties (THAs) with metal-on-metal implants and hip resurfacings. The effect of revision on blood metal ion levels is described only in small series, the clinical results of revisions have been contradictory, and concerns regarding component loosening have been presented.

Questions/purposes

We asked: (1) Did revision surgery result in a reduction to normal for whole blood cobalt (Co) and chromium (Cr) levels (2) What changes to the Oxford Hip Score were observed after revision of these hips with metal-on-metal implants? (3) Were there radiologic signs of component loosening observed on 1-year followup radiographs?

Methods

Between September 2010 and April 2013, 154 patients (166 hips) who had THAs with implantation of the Articular Surface Replacement (ASR™) system and 44 patients (49 hips) who had hip resurfacings of the ASR™ implant underwent revision surgery for adverse reactions to metal debris at our institution, after recall of these components in August 2010. General indications for revision of these implants included a symptomatic hip and/or a predominantly solid pseudotumor seen on cross-sectional imaging. Since recall, patients were systematically followed after revision with Oxford Hip Score questionnaires, blood Co and Cr measurements (analyzed from whole blood with dynamic reaction-cell inductively coupled plasma-mass spectrometry), and plain radiographs at 2 and 12 months after revision surgery, and thereafter at 2-year intervals. Preoperative and 1-year postoperative blood Co and Cr values were available for 93% (185 of 198 patients), Oxford Hip Score for 76% (151 of 198 patients), and plain radiographs for all patients.

Results

Whole-blood levels of Co decreased below the 7 ppb cut-off value in all patients with revision of unilateral THA or resurfacing, however, blood Cr levels remained elevated in four of 90 patients (4%) in the unilateral THA group and four of 34 patients (12%) in the unilateral resurfacing group. All had ultrahigh (> 40 ppb) preoperative Cr levels. Cr levels remained elevated in six of the patients at the 3-year followup. The median Oxford Hip Score improved from preoperative to 1-year postoperative in the unilateral THA group (38 [4–48] to 40 [9–48], p = 0.049) and in the unilateral hip resurfacing group (37.5 [9–48] to 44 [13–48], p = 0.011). No improvement was seen in patients who had bilateral THAs (37 [14–48] to 41 [9–48], p = 0.196). Only minor radiographic abnormalities were seen, with no suspicion of component loosening.

Conclusions

Metal-on-metal THAs and resurfacings have raised concerns and an emerging rate of revisions has been seen for many different metal-on-metal hip prostheses worldwide. Revision surgery seems to be effective for removal of the systemic metal ion burden, even though blood Cr remained elevated in a few patients for more than 3 years after removal of the metal-on-metal implant. In patients with bilateral metal-on-metal hip replacements the remaining metal-on-metal implant still supplies the body with Co and Cr ions after a unilateral revision, and therefore followup should be continued. Adverse reactions to metal debris do not seem to compromise implant ingrowth after revision surgery. However, as some of our patients still had a poor functional outcome at 12 months after revision surgery, additional research is warranted to determine the optimal time for patients to undergo revision surgery for suspected adverse reactions to metal debris.

Level of Evidence

Level IV, therapeutic study.
Literatur
3.
Zurück zum Zitat Ball ST, Le Duff MJ, Amstutz HC. Early results of conversion of a failed femoral component in hip resurfacing arthroplasty. J Bone Joint Surg Am. 2007;89:735–741.PubMedCrossRef Ball ST, Le Duff MJ, Amstutz HC. Early results of conversion of a failed femoral component in hip resurfacing arthroplasty. J Bone Joint Surg Am. 2007;89:735–741.PubMedCrossRef
4.
Zurück zum Zitat Ball ST, Severns D, Linn M, Meyer RS, Swenson FC. What happens to serum metal ion levels after a metal-on-metal bearing is removed? J Arthroplasty. 2013;28(8 suppl):53–55.PubMedCrossRef Ball ST, Severns D, Linn M, Meyer RS, Swenson FC. What happens to serum metal ion levels after a metal-on-metal bearing is removed? J Arthroplasty. 2013;28(8 suppl):53–55.PubMedCrossRef
5.
Zurück zum Zitat Brent J, Devlin JJ. Dilemmas about the toxicological consequences of metal-on-metal hip prostheses: what we do and do not know, and what we should do? Clin Toxicol (Phila). 2013;51:195–198.PubMedCrossRef Brent J, Devlin JJ. Dilemmas about the toxicological consequences of metal-on-metal hip prostheses: what we do and do not know, and what we should do? Clin Toxicol (Phila). 2013;51:195–198.PubMedCrossRef
6.
Zurück zum Zitat Campbell P, Ebramzadeh E, Nelson S, Takamura K, De Smet K, Amstutz HC. Histological features of pseudotumor-like tissues from metal-on-metal hips. Clin Orthop Relat Res. 2010;468:2321–2327.PubMedCentralPubMedCrossRef Campbell P, Ebramzadeh E, Nelson S, Takamura K, De Smet K, Amstutz HC. Histological features of pseudotumor-like tissues from metal-on-metal hips. Clin Orthop Relat Res. 2010;468:2321–2327.PubMedCentralPubMedCrossRef
7.
Zurück zum Zitat De Haan R, Campbell PA, Su EP, De Smet KA. Revision of metal-on-metal resurfacing arthroplasty of the hip: the influence of malpositioning of the components. J Bone Joint Surg Br. 2008;90:1158–1163.PubMedCrossRef De Haan R, Campbell PA, Su EP, De Smet KA. Revision of metal-on-metal resurfacing arthroplasty of the hip: the influence of malpositioning of the components. J Bone Joint Surg Br. 2008;90:1158–1163.PubMedCrossRef
8.
Zurück zum Zitat De Smet KA, Van Der Straeten C, Van Orsouw M, Doubi R, Backers K, Grammatopoulos G. Revisions of metal-on-metal hip resurfacing: lessons learned and improved outcome. Orthop Clin North Am. 2011;42:259–269, ix. De Smet KA, Van Der Straeten C, Van Orsouw M, Doubi R, Backers K, Grammatopoulos G. Revisions of metal-on-metal hip resurfacing: lessons learned and improved outcome. Orthop Clin North Am. 2011;42:259–269, ix.
9.
Zurück zum Zitat de Steiger RN, Miller LN, Prosser GH, Graves SE, Davidson DC, Stanford TE. Poor outcome of revised resurfacing hip arthroplasty. Acta Orthop. 2010;81:72–76.PubMedCentralPubMedCrossRef de Steiger RN, Miller LN, Prosser GH, Graves SE, Davidson DC, Stanford TE. Poor outcome of revised resurfacing hip arthroplasty. Acta Orthop. 2010;81:72–76.PubMedCentralPubMedCrossRef
10.
Zurück zum Zitat DeLee JG, Charnley J. Radiological demarcation of cemented sockets in total hip replacement. Clin Orthop Relat Res. 1976;121:20–32.PubMed DeLee JG, Charnley J. Radiological demarcation of cemented sockets in total hip replacement. Clin Orthop Relat Res. 1976;121:20–32.PubMed
12.
Zurück zum Zitat Durrani SK, Noble PC, Sampson B, Panetta T, Liddle AD, Sabah SA, Chan NK, Skinner JA, Hart AJ. Changes in blood ion levels after removal of metal-on-metal hip replacements: 16 patients followed for 0–12 months. Acta Orthop. 2014;85:259–265.PubMedCentralPubMedCrossRef Durrani SK, Noble PC, Sampson B, Panetta T, Liddle AD, Sabah SA, Chan NK, Skinner JA, Hart AJ. Changes in blood ion levels after removal of metal-on-metal hip replacements: 16 patients followed for 0–12 months. Acta Orthop. 2014;85:259–265.PubMedCentralPubMedCrossRef
13.
Zurück zum Zitat Ebreo D, Khan A, El-Meligy M, Armstrong C, Peter V. Metal ion levels decrease after revision for metallosis arising from large-diameter metal-on-metal hip arthroplasty. Acta Orthop Belg. 2011;77:777–781.PubMed Ebreo D, Khan A, El-Meligy M, Armstrong C, Peter V. Metal ion levels decrease after revision for metallosis arising from large-diameter metal-on-metal hip arthroplasty. Acta Orthop Belg. 2011;77:777–781.PubMed
14.
Zurück zum Zitat Eswaramoorthy VK, Biant LC, Field RE. Clinical and radiological outcome of stemmed hip replacement after revision from metal-on-metal resurfacing. J Bone Joint Surg Br. 2009;91:1454–1458.PubMedCrossRef Eswaramoorthy VK, Biant LC, Field RE. Clinical and radiological outcome of stemmed hip replacement after revision from metal-on-metal resurfacing. J Bone Joint Surg Br. 2009;91:1454–1458.PubMedCrossRef
15.
Zurück zum Zitat Fabi D, Levine B, Paprosky W, Della Valle C, Sporer S, Klein G, Levine H, Hartzband M. Metal-on-metal total hip arthroplasty: causes and high incidence of early failure. Orthopedics. 2012;35:e1009–e1016.PubMedCrossRef Fabi D, Levine B, Paprosky W, Della Valle C, Sporer S, Klein G, Levine H, Hartzband M. Metal-on-metal total hip arthroplasty: causes and high incidence of early failure. Orthopedics. 2012;35:e1009–e1016.PubMedCrossRef
17.
Zurück zum Zitat Gilbert RE, Cheung G, Carrothers AD, Meyer C, Richardson JB. Functional results of isolated femoral revision of hip resurfacing arthroplasty. J Bone Joint Surg Am. 2010;92:1600–1604.PubMedCrossRef Gilbert RE, Cheung G, Carrothers AD, Meyer C, Richardson JB. Functional results of isolated femoral revision of hip resurfacing arthroplasty. J Bone Joint Surg Am. 2010;92:1600–1604.PubMedCrossRef
18.
Zurück zum Zitat Grammatopolous G, Pandit H, Kwon YM, Gundle R, McLardy-Smith P, Beard DJ, Murray DW, Gill HS. Hip resurfacings revised for inflammatory pseudotumour have a poor outcome. J Bone Joint Surg Br. 2009;91:1019–1024.PubMedCrossRef Grammatopolous G, Pandit H, Kwon YM, Gundle R, McLardy-Smith P, Beard DJ, Murray DW, Gill HS. Hip resurfacings revised for inflammatory pseudotumour have a poor outcome. J Bone Joint Surg Br. 2009;91:1019–1024.PubMedCrossRef
19.
Zurück zum Zitat Gross TP, Liu F. Outcomes after revision of metal-on-metal hip resurfacing arthroplasty. J Arthroplasty. 2014;29(9 suppl):219–223.PubMedCrossRef Gross TP, Liu F. Outcomes after revision of metal-on-metal hip resurfacing arthroplasty. J Arthroplasty. 2014;29(9 suppl):219–223.PubMedCrossRef
20.
Zurück zum Zitat Gruen TA, McNeice GM, Amstutz HC. “Modes of failure” of cemented stem-type femoral components: a radiographic analysis of loosening. Clin Orthop Relat Res. 1979;141:17–27.PubMed Gruen TA, McNeice GM, Amstutz HC. “Modes of failure” of cemented stem-type femoral components: a radiographic analysis of loosening. Clin Orthop Relat Res. 1979;141:17–27.PubMed
21.
Zurück zum Zitat Hart AJ, Quinn PD, Sampson B, Sandison A, Atkinson KD, Skinner JA, Powell JJ, Mosselmans JF. The chemical form of metallic debris in tissues surrounding metal-on-metal hips with unexplained failure. Acta Biomater. 2010;6:4439–4446.PubMedCrossRef Hart AJ, Quinn PD, Sampson B, Sandison A, Atkinson KD, Skinner JA, Powell JJ, Mosselmans JF. The chemical form of metallic debris in tissues surrounding metal-on-metal hips with unexplained failure. Acta Biomater. 2010;6:4439–4446.PubMedCrossRef
22.
Zurück zum Zitat Langton DJ, Jameson SS, Joyce TJ, Hallab NJ, Natu S, Nargol AV. Early failure of metal-on-metal bearings in hip resurfacing and large-diameter total hip replacement: a consequence of excess wear. J Bone Joint Surg Br. 2010;92:38–46.PubMedCrossRef Langton DJ, Jameson SS, Joyce TJ, Hallab NJ, Natu S, Nargol AV. Early failure of metal-on-metal bearings in hip resurfacing and large-diameter total hip replacement: a consequence of excess wear. J Bone Joint Surg Br. 2010;92:38–46.PubMedCrossRef
23.
Zurück zum Zitat Liddle AD, Satchithananda K, Henckel J, Sabah SA, Vipulendran KV, Lewis A, Skinner JA, Mitchell AW, Hart AJ. Revision of metal-on-metal hip arthroplasty in a tertiary center: a prospective study of 39 hips with between 1 and 4 years of follow-up. Acta Orthop. 2013;84:237–245.PubMedCentralPubMedCrossRef Liddle AD, Satchithananda K, Henckel J, Sabah SA, Vipulendran KV, Lewis A, Skinner JA, Mitchell AW, Hart AJ. Revision of metal-on-metal hip arthroplasty in a tertiary center: a prospective study of 39 hips with between 1 and 4 years of follow-up. Acta Orthop. 2013;84:237–245.PubMedCentralPubMedCrossRef
25.
Zurück zum Zitat Munro JT, Masri BA, Duncan CP, Garbuz DS. High complication rate after revision of large-head metal-on-metal total hip arthroplasty. Clin Orthop Relat Res. 2014;472;523–528.PubMedCentralPubMedCrossRef Munro JT, Masri BA, Duncan CP, Garbuz DS. High complication rate after revision of large-head metal-on-metal total hip arthroplasty. Clin Orthop Relat Res. 2014;472;523–528.PubMedCentralPubMedCrossRef
26.
Zurück zum Zitat Murray DW, Fitzpatrick R, Rogers K, Pandit H, Beard DJ, Carr AJ, Dawson J. The use of the Oxford hip and knee scores. J Bone Joint Surg Br. 2007;89:1010–1014.PubMedCrossRef Murray DW, Fitzpatrick R, Rogers K, Pandit H, Beard DJ, Carr AJ, Dawson J. The use of the Oxford hip and knee scores. J Bone Joint Surg Br. 2007;89:1010–1014.PubMedCrossRef
28.
Zurück zum Zitat Pritchett JW. One-component revision of failed hip resurfacing from adverse reaction to metal wear debris. J Arthroplasty. 2014;29:219–224.PubMedCrossRef Pritchett JW. One-component revision of failed hip resurfacing from adverse reaction to metal wear debris. J Arthroplasty. 2014;29:219–224.PubMedCrossRef
29.
Zurück zum Zitat Reito A, Puolakka T, Elo P, Pajamaki J, Eskelinen A. High prevalence of adverse reactions to metal debris in small-headed ASR™ hips. Clin Orthop Relat Res. 2013;471:2954–2961.PubMedCentralPubMedCrossRef Reito A, Puolakka T, Elo P, Pajamaki J, Eskelinen A. High prevalence of adverse reactions to metal debris in small-headed ASR™ hips. Clin Orthop Relat Res. 2013;471:2954–2961.PubMedCentralPubMedCrossRef
30.
Zurück zum Zitat Stryker LS, Odum SM, Fehring TK, Springer BD. Revisions of monoblock metal-on-metal THAs have high early complication rates. Clin Orthop Relat Res. 2014 Jul 16. [Epub ahead of print] Stryker LS, Odum SM, Fehring TK, Springer BD. Revisions of monoblock metal-on-metal THAs have high early complication rates. Clin Orthop Relat Res. 2014 Jul 16. [Epub ahead of print]
31.
Zurück zum Zitat Su EP, Su SL. Surface replacement conversion: results depend upon reason for revision. Bone Joint J. 2013;95(11 suppl A):88–91. Su EP, Su SL. Surface replacement conversion: results depend upon reason for revision. Bone Joint J. 2013;95(11 suppl A):88–91.
32.
Zurück zum Zitat Urban RM, Tomlinson MJ, Hall DJ, Jacobs JJ. Accumulation in liver and spleen of metal particles generated at nonbearing surfaces in hip arthroplasty. J Arthroplasty. 2004;19:94–101.PubMedCrossRef Urban RM, Tomlinson MJ, Hall DJ, Jacobs JJ. Accumulation in liver and spleen of metal particles generated at nonbearing surfaces in hip arthroplasty. J Arthroplasty. 2004;19:94–101.PubMedCrossRef
33.
Zurück zum Zitat Wyles CC, Van Demark RE 3rd, Sierra RJ, Trousdale RT. High rate of infection after aseptic revision of failed metal-on-metal total hip arthroplasty. Clin Orthop Relat Res. 2014;472:509–516. Wyles CC, Van Demark RE 3rd, Sierra RJ, Trousdale RT. High rate of infection after aseptic revision of failed metal-on-metal total hip arthroplasty. Clin Orthop Relat Res. 2014;472:509–516.
Metadaten
Titel
Revision of Metal-on-metal Hip Prostheses Results in Marked Reduction of Blood Cobalt and Chromium Ion Concentrations
verfasst von
Olli Lainiala, BM
Aleksi Reito, MD
Petra Elo, MD, PhD
Jorma Pajamäki, MD, PhD
Timo Puolakka, MD, PhD
Antti Eskelinen, MD, PhD
Publikationsdatum
01.07.2015
Verlag
Springer US
Erschienen in
Clinical Orthopaedics and Related Research® / Ausgabe 7/2015
Print ISSN: 0009-921X
Elektronische ISSN: 1528-1132
DOI
https://doi.org/10.1007/s11999-015-4156-6

Weitere Artikel der Ausgabe 7/2015

Clinical Orthopaedics and Related Research® 7/2015 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.