Skip to main content
Erschienen in: Neurocritical Care 2/2011

01.04.2011 | Practical Pearl

Ventilator Autocycling and Delayed Recognition of Brain Death

verfasst von: William T. McGee, Patrick Mailloux

Erschienen in: Neurocritical Care | Ausgabe 2/2011

Einloggen, um Zugang zu erhalten

Abstract

Background

Improvements in technology play an important role in caring for critically ill patients. One example is the advance in ventilator design to facilitate triggering of mechanical breaths. Minimal changes in circuit flow unrelated to respiratory effort can trigger a ventilator breath and may mislead caregivers in recognizing brain death.

Methods

We observed patients with devastating brain injuries in a mixed medical/surgical intensive care unit (ICU) with a high clinical suspicion for brain death including the absence of cranial nerve function with apparent spontaneous breathing during patient-triggered modes of mechanical ventilation. Further clinical observation for spontaneous respirations was assessed upon removal of ventilatory support.

Results

Nine patients with brain injury due to multiple etiologies were identified and demonstrated no spontaneous respirations when formally assessed for apnea. Length of time between brain death and its recognition could not be determined.

Conclusion

When brain-dead patients who are suitable organ donors are mistakenly identified as having cerebral activity, the diagnosis of brain death is delayed. This delay impacts resource utilization, impedes recovery and function of organs for donation, and adversely affects donor families, potential recipients of organs, and patient donors who may have testing and treatment that cannot be beneficial. Patients with catastrophic brain injury and absent cranial nerve function should undergo immediate formal apnea testing.
Literatur
2.
Zurück zum Zitat Ethics Committee. Recommendations for non-heart beating organ donation: a position paper by the Ethics Committee, American College of Critical Care Medicine, Society of Critical Care Medicine. Crit Care Med. 2001;29:1826–31.CrossRef Ethics Committee. Recommendations for non-heart beating organ donation: a position paper by the Ethics Committee, American College of Critical Care Medicine, Society of Critical Care Medicine. Crit Care Med. 2001;29:1826–31.CrossRef
3.
Zurück zum Zitat Frontera JA, Kalb T. How I manage the adult potential organ donor: donation after neurological death. Neurocrit Care. 2010;12(1):111–6.PubMedCrossRef Frontera JA, Kalb T. How I manage the adult potential organ donor: donation after neurological death. Neurocrit Care. 2010;12(1):111–6.PubMedCrossRef
4.
Zurück zum Zitat Imanaka H, Nishimura M, Takeuchi M, et al. Autotriggering caused by cardiogenic oscillation during flow-triggered mechanical ventilation. Crit Care Med. 2000;28:402–7.PubMedCrossRef Imanaka H, Nishimura M, Takeuchi M, et al. Autotriggering caused by cardiogenic oscillation during flow-triggered mechanical ventilation. Crit Care Med. 2000;28:402–7.PubMedCrossRef
5.
Zurück zum Zitat Willatts SM, Drummond G. Brainstem death and ventilator trigger settings. Anaesthesia. 2000;55:676–7.PubMedCrossRef Willatts SM, Drummond G. Brainstem death and ventilator trigger settings. Anaesthesia. 2000;55:676–7.PubMedCrossRef
6.
Zurück zum Zitat Wijdicks EFM, Manno EM, Holets SR. Ventilator self-cycling may falsely suggest patient effort during brain death determination. Neurology. 2005;65:774.PubMedCrossRef Wijdicks EFM, Manno EM, Holets SR. Ventilator self-cycling may falsely suggest patient effort during brain death determination. Neurology. 2005;65:774.PubMedCrossRef
7.
Zurück zum Zitat Arbour R. Cardiogenic oscillation and ventilator autotriggering in brain-dead patients: a case series. Am J Crit Care. 2009;18(5):488–95.CrossRef Arbour R. Cardiogenic oscillation and ventilator autotriggering in brain-dead patients: a case series. Am J Crit Care. 2009;18(5):488–95.CrossRef
8.
Zurück zum Zitat Harboe S. Autocycling and increase in intrinsic positive end-expiratory pressure during mechanical ventilation. Acta Anaesthesiol Scand. 2001;45:1295–7.PubMedCrossRef Harboe S. Autocycling and increase in intrinsic positive end-expiratory pressure during mechanical ventilation. Acta Anaesthesiol Scand. 2001;45:1295–7.PubMedCrossRef
9.
Zurück zum Zitat Shoham AB, Patel B, Arabia FA, et al. Mechanical ventilation and the total artificial heart: optimal ventilator trigger to avoid post-operative autocycling—a case series and literature review. J Cardiothorac Surg. 2010;5:39.PubMedCrossRef Shoham AB, Patel B, Arabia FA, et al. Mechanical ventilation and the total artificial heart: optimal ventilator trigger to avoid post-operative autocycling—a case series and literature review. J Cardiothorac Surg. 2010;5:39.PubMedCrossRef
10.
Zurück zum Zitat Schwab RJ, Schnader J. Ventilator autocycling due to an endotracheal tube cuff leak. Chest. 1991;100:1172–3.PubMedCrossRef Schwab RJ, Schnader J. Ventilator autocycling due to an endotracheal tube cuff leak. Chest. 1991;100:1172–3.PubMedCrossRef
11.
Zurück zum Zitat Baumann MH, Sahn SA. Medical management and therapy of bronchopleural fistulas in the mechanically ventilated patient. Chest. 1990;97:721–8.PubMedCrossRef Baumann MH, Sahn SA. Medical management and therapy of bronchopleural fistulas in the mechanically ventilated patient. Chest. 1990;97:721–8.PubMedCrossRef
12.
Zurück zum Zitat Pierson DJ. Persistent bronchopleural air leak during mechanical ventilation: a review. Respir Care. 1982;27:408–16. Pierson DJ. Persistent bronchopleural air leak during mechanical ventilation: a review. Respir Care. 1982;27:408–16.
13.
Zurück zum Zitat Tilles RB, Don HF. Complications of high pleural suction in bronchopleural fistulas. Anesthesiology. 1975;43:486–7.PubMedCrossRef Tilles RB, Don HF. Complications of high pleural suction in bronchopleural fistulas. Anesthesiology. 1975;43:486–7.PubMedCrossRef
14.
Zurück zum Zitat Zimmerman JE, Coglan DL, Mills M. Management of bronchopleural fistula complicating therapy with positive and expiratory pressure (PEEP). Chest. 1973;64:526–9.PubMedCrossRef Zimmerman JE, Coglan DL, Mills M. Management of bronchopleural fistula complicating therapy with positive and expiratory pressure (PEEP). Chest. 1973;64:526–9.PubMedCrossRef
15.
Zurück zum Zitat Aslanian P, El Atrous S, Isabey D, et al. Effects of flow triggering on breathing effort during partial ventilatory support. Am J Respir Crit Care Med. 1998;157:135–43.PubMed Aslanian P, El Atrous S, Isabey D, et al. Effects of flow triggering on breathing effort during partial ventilatory support. Am J Respir Crit Care Med. 1998;157:135–43.PubMed
16.
Zurück zum Zitat West JB, Hugh-Jones P. Pulsatile gas flow in bronchi caused by the heart beat. J Appl Phys. 1961;16(4):697–702. West JB, Hugh-Jones P. Pulsatile gas flow in bronchi caused by the heart beat. J Appl Phys. 1961;16(4):697–702.
17.
Zurück zum Zitat Verheije JL, Rady MY, McGregor J. Recovery of transplantable organs after cardiac or circulatory death: transforming the paradigm for the ethics of organ donation. Philos Ethics Humanit Med. 2007;2(May):8.CrossRef Verheije JL, Rady MY, McGregor J. Recovery of transplantable organs after cardiac or circulatory death: transforming the paradigm for the ethics of organ donation. Philos Ethics Humanit Med. 2007;2(May):8.CrossRef
18.
Zurück zum Zitat Lewis J, Peltier J, Nelson H, et al. Development of the University of Wisconsin donation after cardiac death evaluation tool. Prog Transplant. 2003;13(4):265–73.PubMed Lewis J, Peltier J, Nelson H, et al. Development of the University of Wisconsin donation after cardiac death evaluation tool. Prog Transplant. 2003;13(4):265–73.PubMed
19.
Zurück zum Zitat Bernat JL, D’Alessandro AM, Port FK, et al. Report of a national conference on donation after cardiac death. Am J Transplant. 2006;6:281–91.PubMedCrossRef Bernat JL, D’Alessandro AM, Port FK, et al. Report of a national conference on donation after cardiac death. Am J Transplant. 2006;6:281–91.PubMedCrossRef
20.
Zurück zum Zitat Wood KE, Becker BN, McCartney JG, et al. Care of the potential organ donor. N Engl J Med. 2004;351:2730–9.PubMedCrossRef Wood KE, Becker BN, McCartney JG, et al. Care of the potential organ donor. N Engl J Med. 2004;351:2730–9.PubMedCrossRef
21.
Zurück zum Zitat Cantin B, Kwok BW, Chan MC, et al. The impact of brain death on survival after heart transplantation: time is of the essence. Transplantation. 2003;76:1275–9.PubMedCrossRef Cantin B, Kwok BW, Chan MC, et al. The impact of brain death on survival after heart transplantation: time is of the essence. Transplantation. 2003;76:1275–9.PubMedCrossRef
22.
Zurück zum Zitat McGee WT, Teres D. Scoring systems in the ICU: outcome predictors for responsible resource utilization. In: Orlowski JP, editor. Ethics in critical care. Frederick: University Publishing Group; 1999. p. 153–85. McGee WT, Teres D. Scoring systems in the ICU: outcome predictors for responsible resource utilization. In: Orlowski JP, editor. Ethics in critical care. Frederick: University Publishing Group; 1999. p. 153–85.
23.
Zurück zum Zitat Wijdicks E, Varelas P, Gronseth G, Greer D. Evidence-based guideline update: determining brain death in adults. Neurology. 2010;74:1911–8.PubMedCrossRef Wijdicks E, Varelas P, Gronseth G, Greer D. Evidence-based guideline update: determining brain death in adults. Neurology. 2010;74:1911–8.PubMedCrossRef
Metadaten
Titel
Ventilator Autocycling and Delayed Recognition of Brain Death
verfasst von
William T. McGee
Patrick Mailloux
Publikationsdatum
01.04.2011
Verlag
Humana Press Inc
Erschienen in
Neurocritical Care / Ausgabe 2/2011
Print ISSN: 1541-6933
Elektronische ISSN: 1556-0961
DOI
https://doi.org/10.1007/s12028-010-9491-6

Weitere Artikel der Ausgabe 2/2011

Neurocritical Care 2/2011 Zur Ausgabe

Leitlinien kompakt für die Neurologie

Mit medbee Pocketcards sicher entscheiden.

Seit 2022 gehört die medbee GmbH zum Springer Medizin Verlag

Update Neurologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.