Objective

Simultaneous acquisition of 99mTc and 123I was evaluated using a preclinical SPECT scanner with cadmium zinc telluride (CZT)-based detectors.

Methods

10-ml cylindrical syringes contained about 37 MBq 99mTc-tetrofosmin (99mTc-TF) or 37 MBq 123I-15-(p-iodophenyl)-3R,S-methyl pentadecanoic acid (123I-BMIPP) were used to assess the relationship between these SPECT radioactive counts and radioactivity. Two 10-ml syringes contained 100 or 300 MBq 99mTc-TF and 100 MBq 123I-BMIPP to assess the influence of 99mTc upscatter and 123I downscatter, respectively. A rat-sized cylindrical phantom also contained both 100 or 300 MBq 99mTc-TF and 100 MBq 123I-BMIPP. The two 10-ml syringes and phantom were scanned using a pinhole collimator for rats. Myocardial infarction model rats were examined using 300 MBq 99mTc-TF and 100 MBq 123I-BMIPP. Two 1-ml syringes contained 105 MBq 99mTc-labeled hexamethylpropyleneamine oxime (99mTc-HMPAO) and 35 MBq 123I-labeled N-ω-fluoropropyl-2β-carbomethoxy-3β-(4-iodophenyl) nortropane (123I-FP-CIT). The two 1-ml syringes were scanned using a pinhole collimator for mice. Normal mice were examined using 105 MBq 99mTc-HMPAO and 35 MBq 123I-FP-CIT.

Results

The relationship between SPECT radioactive counts and radioactivity was excellent. Downscatter contamination of 123I-BMIPP exhibited fewer radioactive counts for 300 MBq 99mTc-TF without scatter correction (SC) in 125–150 keV. There was no upscatter contamination of 99mTc-TF in 150–175 keV. In the rat-sized phantom, the radioactive count ratio decreased to 4.0 % for 300 MBq 99mTc-TF without SC in 125–150 keV. In the rats, myocardial images and radioactive counts of 99mTc-TF with the dual tracer were identical to those of the 99mTc-TF single injection. Downscatter contamination of 123I-FP-CIT was 4.2 % without SC in 125–150 keV. In the first injection of 99mTc-HMPAO and second injection of 123I-FP-CIT, brain images and radioactive counts of 99mTc-HMPAO with the dual tracer in normal mice also were the similar to those of the 99mTc-HMPAO single injection. In the first injection of 123I-FP-CIT and second injection of 99mTc-HMPAO, the brain images and radioactive counts with the dual tracer were not much different from those of the 123I-FP-CIT single injection.

Conclusions

Dual-tracer imaging of 99mTc- and 123I-labeled radiotracers is feasible in a preclinical SPECT scanner with CZT detector. When higher radioactivity of 99mTc-labeled radiotracers relative to 123I-labeled radiotracers is applied, correction methods are not necessarily required for the quantification of 99mTc- and 123I-labeled radiotracers when using a preclinical SPECT scanner with CZT detector.
Literatur
1.
Zurück zum Zitat Franc BL, Acton PD, Mari C, Hasegawa BH. Small-animal SPECT and SPECT/CT: important tools for preclinical investigation. J Nucl Med. 2008;49:1651–63.CrossRefPubMed Franc BL, Acton PD, Mari C, Hasegawa BH. Small-animal SPECT and SPECT/CT: important tools for preclinical investigation. J Nucl Med. 2008;49:1651–63.CrossRefPubMed
2.
Zurück zum Zitat Kim H, Furenlid LR, Crawford MJ, Wilson DW, Barber HB, Peterson TE, et al. SemiSPECT: a small-animal single-photon emission computed tomography (SPECT) imager based on eight cadmium zinc telluride (CZT) detector arrays. Med Phys. 2006;33:465–74.CrossRefPubMedPubMedCentral Kim H, Furenlid LR, Crawford MJ, Wilson DW, Barber HB, Peterson TE, et al. SemiSPECT: a small-animal single-photon emission computed tomography (SPECT) imager based on eight cadmium zinc telluride (CZT) detector arrays. Med Phys. 2006;33:465–74.CrossRefPubMedPubMedCentral
3.
Zurück zum Zitat Higaki Y, Kobayashi M, Uehara T, Hanaoka H, Arano Y, Kawai K. Appropriate collimators in a small animal SPECT scanner with CZT detector. Ann Nucl Med. 2013;27:271–8.CrossRefPubMed Higaki Y, Kobayashi M, Uehara T, Hanaoka H, Arano Y, Kawai K. Appropriate collimators in a small animal SPECT scanner with CZT detector. Ann Nucl Med. 2013;27:271–8.CrossRefPubMed
4.
Zurück zum Zitat Matsunari I, Miyazaki Y, Kobayashi M, Nishi K, Mizutani A, Kawai K, et al. Performance evaluation of the eXplore speCZT preclinical imaging system. Ann Nucl Med. 2014;28:484–97.CrossRefPubMed Matsunari I, Miyazaki Y, Kobayashi M, Nishi K, Mizutani A, Kawai K, et al. Performance evaluation of the eXplore speCZT preclinical imaging system. Ann Nucl Med. 2014;28:484–97.CrossRefPubMed
5.
Zurück zum Zitat Herzog BA, Buechel RA, Katz R, Brueckner M, Husmann L, Burger IA, et al. Nuclear myocardial perfusion imaging with a cadmium-zinc-telluride detector technique: optimized protocol for scan time reduction. J Nucl Med. 2010;51:46–51.CrossRefPubMed Herzog BA, Buechel RA, Katz R, Brueckner M, Husmann L, Burger IA, et al. Nuclear myocardial perfusion imaging with a cadmium-zinc-telluride detector technique: optimized protocol for scan time reduction. J Nucl Med. 2010;51:46–51.CrossRefPubMed
6.
Zurück zum Zitat Kacperski K, Erlandsson K, Ben-Haim S, Hutton BF. Iterative deconvolution of simultaneous 99mTc and 201Tl projection data measured on a CdZnTe-based cardiac SPECT scanner. Phys Med Biol. 2010;56:1397–414.CrossRef Kacperski K, Erlandsson K, Ben-Haim S, Hutton BF. Iterative deconvolution of simultaneous 99mTc and 201Tl projection data measured on a CdZnTe-based cardiac SPECT scanner. Phys Med Biol. 2010;56:1397–414.CrossRef
7.
Zurück zum Zitat El Fakhri G, Maksud P, Kijewski MF, Habert MO, Todd-Pokropek A, Aurengo A, et al. Scatter and cross-talk corrections in simultaneous Tc-99 m/I-123 brain SPECT using constrained factor analysis and artificial neural networks. IEEE Trans Nucl Sci. 2000;47:1573–80.CrossRef El Fakhri G, Maksud P, Kijewski MF, Habert MO, Todd-Pokropek A, Aurengo A, et al. Scatter and cross-talk corrections in simultaneous Tc-99 m/I-123 brain SPECT using constrained factor analysis and artificial neural networks. IEEE Trans Nucl Sci. 2000;47:1573–80.CrossRef
8.
Zurück zum Zitat El Fakhri G, Moore SC, Maksud P, Aurengo A, Kijewski MF. Absolute activity quantitation in simultaneous 123I/99mTc brain SPECT. J Nucl Med. 2001;42:300–8.PubMed El Fakhri G, Moore SC, Maksud P, Aurengo A, Kijewski MF. Absolute activity quantitation in simultaneous 123I/99mTc brain SPECT. J Nucl Med. 2001;42:300–8.PubMed
10.
Zurück zum Zitat Du Y, Tsui BM, Frey FC. Model-based crosstalk compensation for simultaneous 99mTc/123I dual-isotope brain SPECT imaging. Med Phys. 2007;34:3530–43.CrossRefPubMed Du Y, Tsui BM, Frey FC. Model-based crosstalk compensation for simultaneous 99mTc/123I dual-isotope brain SPECT imaging. Med Phys. 2007;34:3530–43.CrossRefPubMed
11.
Zurück zum Zitat Du Y, Frey EC. Quantitative evaluation of simultaneous reconstruction with model-based crosstalk compensation for 99mTc/123I dual-isotope simultaneous acquisition brain SPECT. Med Phys. 2009;36:2021–33.CrossRefPubMedPubMedCentral Du Y, Frey EC. Quantitative evaluation of simultaneous reconstruction with model-based crosstalk compensation for 99mTc/123I dual-isotope simultaneous acquisition brain SPECT. Med Phys. 2009;36:2021–33.CrossRefPubMedPubMedCentral
12.
Zurück zum Zitat Kao PF, Way SP, Yang AS. Simultaneous 99mTc and 123I dual-isotope brain striatal phantom single photon emission computed tomography: validation of 99mTc-TRODAT1 and 123I-IBZM simultaneous dopamine system brain imaging. Kaohsiung J Med Sci. 2009;25:601–7.CrossRefPubMed Kao PF, Way SP, Yang AS. Simultaneous 99mTc and 123I dual-isotope brain striatal phantom single photon emission computed tomography: validation of 99mTc-TRODAT1 and 123I-IBZM simultaneous dopamine system brain imaging. Kaohsiung J Med Sci. 2009;25:601–7.CrossRefPubMed
13.
Zurück zum Zitat Higley B, Smith FW, Smith T, Gemmell HG, Das Gupta P, Gvozdanovic DV, et al. Technetium-99m-1,2-bis [bis(2-ethoxyethyl) phosphino]ethane: human biodistribution, dosimetry and safety of a new myocardial perfusion imaging agent. J Nucl Med. 1993;34:30–8.PubMed Higley B, Smith FW, Smith T, Gemmell HG, Das Gupta P, Gvozdanovic DV, et al. Technetium-99m-1,2-bis [bis(2-ethoxyethyl) phosphino]ethane: human biodistribution, dosimetry and safety of a new myocardial perfusion imaging agent. J Nucl Med. 1993;34:30–8.PubMed
14.
Zurück zum Zitat Tateno M, Tamaki N, Yukihiro M, Kudoh T, Hattori N, Tadamura E, et al. Assessment of fatty acid uptake in ischemic heart disease without myocardial infarction. J Nucl Med. 1996;37:1981–5.PubMed Tateno M, Tamaki N, Yukihiro M, Kudoh T, Hattori N, Tadamura E, et al. Assessment of fatty acid uptake in ischemic heart disease without myocardial infarction. J Nucl Med. 1996;37:1981–5.PubMed
15.
Zurück zum Zitat Mochizuki T, Murase K, Higashino H, Miyagawa M, Sugawara Y, Kikuchi T, et al. Ischemic “memory image” in acute myocardial infarction of 123I-BMIPP after reperfusion therapy: a comparison with 99mTc-pyrophosphate and 201Tl dual-isotope SPECT. Ann Nucl Med. 2002;16:563–8.CrossRefPubMed Mochizuki T, Murase K, Higashino H, Miyagawa M, Sugawara Y, Kikuchi T, et al. Ischemic “memory image” in acute myocardial infarction of 123I-BMIPP after reperfusion therapy: a comparison with 99mTc-pyrophosphate and 201Tl dual-isotope SPECT. Ann Nucl Med. 2002;16:563–8.CrossRefPubMed
16.
Zurück zum Zitat Reutter BW, Huesman RH, Brennan KM, Boutchko R, Hanrahan SM, Gullberg GT. Longitudinal evaluation of fatty acid metabolism in normal and spontaneously hypertensive rat hearts with dynamic microSPECT imaging. Internal J Mol Imag. 2011;. doi:10.1155/2011/893129. Reutter BW, Huesman RH, Brennan KM, Boutchko R, Hanrahan SM, Gullberg GT. Longitudinal evaluation of fatty acid metabolism in normal and spontaneously hypertensive rat hearts with dynamic microSPECT imaging. Internal J Mol Imag. 2011;. doi:10.​1155/​2011/​893129.
17.
Zurück zum Zitat Walrand S, Jamar F, de Jong M, Pauwels S. Evaluation of novel whole-body high-resolution rodent SPECT (Linoview) based on direct acquisition of linogram projections. J Nucl Med. 2005;46:1872–80.PubMed Walrand S, Jamar F, de Jong M, Pauwels S. Evaluation of novel whole-body high-resolution rodent SPECT (Linoview) based on direct acquisition of linogram projections. J Nucl Med. 2005;46:1872–80.PubMed
18.
Zurück zum Zitat Zheng GL, Gagnon D. CdZnTe strip detector SPECT imaging with a slit collimator. Phys Med Biol. 2004;49:2257–71.CrossRefPubMed Zheng GL, Gagnon D. CdZnTe strip detector SPECT imaging with a slit collimator. Phys Med Biol. 2004;49:2257–71.CrossRefPubMed
19.
Zurück zum Zitat Strydhorst JH, Ruddy TD, Wells RG. Effects of CT-based attenuation correction of rat microSPECT images on relative myocardial perfusion and quantitative tracer uptake. Med Phys. 2015;42:1818–24.CrossRefPubMed Strydhorst JH, Ruddy TD, Wells RG. Effects of CT-based attenuation correction of rat microSPECT images on relative myocardial perfusion and quantitative tracer uptake. Med Phys. 2015;42:1818–24.CrossRefPubMed
20.
Zurück zum Zitat Ichihara T, Ogawa K, Motomura N, Kubo A, Hashimoto S. Compton scatter compensation using the triple-energy window method for single- and dual-isotope SPECT. J Nucl Med. 1993;34:2216–21.PubMed Ichihara T, Ogawa K, Motomura N, Kubo A, Hashimoto S. Compton scatter compensation using the triple-energy window method for single- and dual-isotope SPECT. J Nucl Med. 1993;34:2216–21.PubMed
21.
Zurück zum Zitat O’Reily MA, Muller A, Nynynen K. Ultrasound insertion loss of rat parietal bone appears to be proportional to animal mass at submegahertz frequencies. Ultrasound Med Biol. 2011;37:1930–7.CrossRef O’Reily MA, Muller A, Nynynen K. Ultrasound insertion loss of rat parietal bone appears to be proportional to animal mass at submegahertz frequencies. Ultrasound Med Biol. 2011;37:1930–7.CrossRef
22.
Zurück zum Zitat Sharp PF, Smith FW, Gemmell HG, Lyall D, Evans NT, Gvozdanovic D, Davidson J, Tyrrell DA, Pickett RD, Neirinckx RD. Technetium-99m HM-PAO stereoisomers as potential agents for imaging regional cerebral blood flow: human volunteer studies. J Nucl Med. 1986;27:171–7.PubMed Sharp PF, Smith FW, Gemmell HG, Lyall D, Evans NT, Gvozdanovic D, Davidson J, Tyrrell DA, Pickett RD, Neirinckx RD. Technetium-99m HM-PAO stereoisomers as potential agents for imaging regional cerebral blood flow: human volunteer studies. J Nucl Med. 1986;27:171–7.PubMed
23.
Zurück zum Zitat Miia Pitkonen, Eero Hippeläinen, Mari Raki, Jaan-Olle Andressoo, Arto Urtti, Pekka T Männistö, Sauli Savolainen, Mart Saarma, Kim Bergström. Advanced brain dopamine transporter imaging in mice using small-animal SPECT/CT. EJNMMI Res. 2012; 2: 55. Published online 2012 September 29. doi: 10.1186/2191- 219X-2-55. Miia Pitkonen, Eero Hippeläinen, Mari Raki, Jaan-Olle Andressoo, Arto Urtti, Pekka T Männistö, Sauli Savolainen, Mart Saarma, Kim Bergström. Advanced brain dopamine transporter imaging in mice using small-animal SPECT/CT. EJNMMI Res. 2012; 2: 55. Published online 2012 September 29. doi: 10.​1186/​2191- 219X-2-55.
Metadaten
Titel
Simultaneous acquisition of 99mTc- and 123I-labeled radiotracers using a preclinical SPECT scanner with CZT detectors
verfasst von
Masato Kobayashi
Ichiro Matsunari
Kodai Nishi
Asuka Mizutani
Yoshiharu Miyazaki
Kazuhiro Ogai
Jyunko Sugama
Kazuhiro Shiba
Keiichi Kawai
Seigo Kinuya
Publikationsdatum
08.01.2016
Verlag
Springer Japan
Erschienen in
Annals of Nuclear Medicine / Ausgabe 4/2016
Print ISSN: 0914-7187
Elektronische ISSN: 1864-6433
DOI
https://doi.org/10.1007/s12149-015-1055-6

Weitere Artikel der Ausgabe 4/2016

Annals of Nuclear Medicine 4/2016 Zur Ausgabe