Skip to main content
Erschienen in: Annals of Nuclear Medicine 8/2017

Open Access 08.07.2017 | Original Article

Cross calibration of 123I-meta-iodobenzylguanidine heart-to-mediastinum ratio with D-SPECT planogram and Anger camera

verfasst von: Kenichi Nakajima, Koichi Okuda, Kunihiko Yokoyama, Tatsuya Yoneyama, Shiro Tsuji, Hiroyuki Oda, Mitsuhiro Yoshita, Koji Kubota

Erschienen in: Annals of Nuclear Medicine | Ausgabe 8/2017

Abstract

Background

Cardiac 123I-meta-iodobenzylguanidine (MIBG) uptake is quantified using the heart-to-mediastinum ratio (HMR) with an Anger camera. The relationship between HMR determined using D-SPECT with a cadmium–zinc–telluride detector and an Anger camera is not fully understood. Therefore, the present study aimed to define this relationship using images derived from a phantom and from patients.

Methods

Cross-calibration phantom studies using an Anger camera with a low-energy high-resolution (LEHR) collimator and D-SPECT, and clinical 123I-MIBG studies proceeded in 40 consecutive patients (80 studies). In the phantom study, a conversion coefficient (CC) was defined based on phantom experiments and applied to the Anger camera and the D-SPECT detector. The HMR was calculated using anterior images with the Anger camera and anterior planograms with D-SPECT. First, the HMR from D-SPECT was cross-calibrated to the Anger camera, and then, the HMR from both cameras were converted to the medium-energy general-purpose collimator condition (CC 0.88; ME88 condition). The relationship between HMR and corrected and uncorrected methods was examined. A 123I-MIBG washout rate was calculated using both methods with and without background subtraction.

Results

Based on the phantom experiments, the CC of the Anger camera with an LEHR collimator and of D-SPECT using an anterior planogram was 0.55 and 0.63, respectively. The original HMR from the Anger camera and D-SPECT was 1.76 ± 0.42 and 1.86 ± 0.55, respectively (p < 0.0001). After D-SPECT HMR was converted to the Anger camera condition, the corrected D-SPECT HMR became comparable to the values under the Anger camera condition (1.75 ± 0.48, p = n. s.). When the HMR measured using the two cameras were converted under the ME88 condition, the average standardized HMR from the Anger camera and D-SPECT became comparable (2.21 ± 0.65 vs. 2.20 ± 0.75, p = n. s.). After standardization to the ME88 condition, a systematic difference in the linear regression lines disappeared, and the HMR from both the Anger (StdHMRAnger) and D-SPECT (StdHMRDSPECT) became comparable. Additional correction using a regression line further improved the relationship between both HMR [StdHMRDSPECT = 0.09 + 0.98 × StdHMRAnger (R 2 = 0.91)]. The washout rate closely correlated with and without background correction between both methods (R 2 = 0.83 and 0.65, respectively).

Conclusion

The phantom-based conversion method is applicable to D-SPECT and enables the common application of HMR irrespective of D-SPECT and the Anger camera.
Abkürzungen
CC
Conversion coefficient
CZT
Cadmium–zinc–telluride
HF
Heart failure
HMR
Heart-to-mediastinum ratio
LE
Low energy
ME
Medium energy
ME88
Medium-energy collimator condition with a conversion coefficient of 0.88
MIBG
Meta-iodobenzylguanidine
ROI
Region of interest

Introduction

Several multicenter studies and meta-analysis in Europe, the USA and Japan have indicated the value of sympathetic innervation imaging using 123I-meta-iodobenzylguanidine (MIBG) for patients with heart failure (HF) [15]. The Clinical Practice Guidelines of Nuclear Cardiology published by the Japanese Circulation Society included this procedure based on the considerable accumulation of clinical experience with 123I-MIBG in Japan [6, 7]. The European Association of Nuclear Medicine (EANM) Cardiovascular Committee and the European Council of Nuclear Cardiology have proposed MIBG protocols [8], and the American Society of Nuclear Cardiology (ASNC) imaging guidelines also summarize the application of 123I-MIBG and its methodology [9]. In addition to cardiology, 123I-MIBG has been used since the late 1990s with increasing frequency in patients with Parkinson’s disease and dementia with Lewy bodies, in whom cardiac 123I-MIBG uptake characteristically decreases due to neural degeneration [10, 11]. Thus, 123I-MIBG findings are considered as a biomarker of Lewy-body disease.
Although reproducibility of the heart-to-mediastinum ratio (HMR) is generally believed to be good [12], a major factor affecting HMR is differences among camera collimators at various hospitals [13]. For example, average normal values of late HMR are 2.5 with low-energy (LE) collimators and 3.0 for medium-energy (ME) collimators [14]. In fact, collimator designs are further divided into at least 6–7 collimator groups [15], and these differences are supposed to be mainly caused by different degrees of septal penetration and scatter in collimators, and the precise specifications of the size and length of holes and septal thickness are variable among vendors. We, therefore, developed a phantom-based correction method to cross-calibrate HMR among all Anger camera collimator systems [14]. Several phantom experiments have shown that even collimators of the same type, for example, low-energy high resolution (LEHR), have different specifications depending on the designs of vendors [17]. D-SPECT (Spectrum Dynamics, Israel; Biosensors Japan, Tokyo, Japan) has a cadmium–zinc–telluride (CZT) detector that enables high resolution and high sensitivity in myocardial perfusion imaging [17]. However, tomographic imaging is the standard output, and planar images commonly used with Anger cameras are not directly used. Differences between the Anger and D-SPECT cameras were investigated in the ADRECARD study, in which virtual anterior planograms were created with D-SPECT, and the HMR between the two methods correlated well [18].
The present study aimed to create a method of integrating HMR derived from D-SPECT planogram and Anger cameras using the same phantom-based conversion method to generate comparable quantitative parameters in 123I-MIBG study.

Methods

Phantom design and cross calibration of HMR

The structure of the calibration phantom is described elsewhere [14]. Briefly, the phantom was designed for planar imaging, and two reference HMR values can be obtained from one phantom using anterior and posterior sides (Hokuriku Yuuki, Co. Ltd., Kanazawa, Japan). Since the phantom has two compartments, one for 123I-MIBG and the other for water, the radionuclide concentration does not require adjustment, and HMR can be reproducibly calculated using dedicated software. The reference HMR values obtained from the anterior and posterior sides of the phantom were 2.6 and 3.5, respectively. The count decay in the acrylic and water compartment was calculated for thickness using an attenuation coefficient of 0.147/cm [19, 20].
A linear regression line that passes through a coordinate (1, 1) for the measured versus the reference HMR can be calculated, because two data points are obtained from the anterior and posterior sides. The slope of this regression line is defined as a conversion coefficient (CC) to the reference value, and it is unique for an institutional specific combination of scinticamera-collimator systems and acquisition conditions.

Planograms generated by D-SPECT

The standard D-SPECT output comprised tomographic reconstructed images. Therefore, a planogram equivalent to a planar anterior image was created based on all elementary two-dimensional images that shared the same angle onto one large field of view in a virtual plane as described [18]. A series of two-dimensional images equivalent to those of SPECT with the Anger camera were obtained for every angular position. The phantom structure was designed for planar images, which render three-dimensional reconstruction meaningless. Therefore, we used only anterior planograms and repeated the acquisition on the reverse side of the phantom.

Phantom experiments

Experiments with the calibration phantom proceeded for the Anger camera (Siemens Healthcare, Tokyo, Japan) with an LEHR collimator, and planar images were acquired from both sides of the phantom. Diluted 123I-MIBG (111 MBq) was poured into the phantom and images were acquired for 5 min each. A 15% energy window was set at 159 keV. Data were similarly acquired from both sides of the phantom for D-SPECT. The phantom was positioned horizontally on the backrest of an SPECT chair similar to how patients are positioned. A 3-cm acrylic plate filled with water was placed over the phantom when imaging was performed to simulate human body attenuation and scatter. A 15% asymmetric energy window was set at 159 keV (145–169 keV) for D-SPECT. Figure 1 shows typical phantom images.

Clinical study

We retrospectively selected 40 consecutive outpatients (average age, 73 ± 10 years; male, 50%) referred for 123I-MIBG assessment between July 2016 and February 2017. The indications for 123I-MIBG assessment were determined by internal medicine physicians, and their aims for the study were to diagnose Parkinson’s disease and dementia with Lewy bodies for neurology, and to evaluate the diagnosis and prognosis of HF. The diagnosis of neurological diseases was made in 24 patients. Among the patients with HF, the average left ventricular ejection fraction ranged 20–70%.
We used data acquired from all enrolled patients, since our aim was to develop an optimal methodology. The Ethics Committee at the Public Central Hospital of Matto Ishikawa approved this research protocol. Written informed consent from individual patients was waived, because the MIBG studies comprised part of regular clinical practice without additional imaging.
Early and late images were acquired at 15 min and 3 h after an intravenous injection of 111 MBq of 123I-MIBG (MyoMIBG; FUJIFILM RI Pharma, Tokyo, Japan) using an Anger camera. The anterior images were obtained for 5 min each with a 256 × 256 matrix. Just after Anger camera images, D-SPECT images were acquired for 10 min. An anterior-view equivalent planograms were generated, and the HMR was then calculated. Tomographic imaging was a standard clinical procedure used for evaluating segmental defect caused by ischemia and extensive decrease in Lewy-body disease, but it was not used in the present study.

Regions of interest and HMR

Phantom analysis We set regions of interest (ROI) on the heart and mediastinum of the phantom. The heart ROI was set as a circle on the heart, and the mediastinal ROI was set as a rectangle on the upper mediastinum with the Anger camera. The location was predefined for the phantom in all phantom experiments using the dedicated analytical software. Since the vertical image size was limited to 160 mm with D-SPECT, a similar ROI was manually set on the heart and mediastinum. The mediastinal ROI was set on the mid-mediastinum as high as possible, although lower than that in the Anger camera image.
Patient analysis The ROI was semi-automatically set as described for the clinical study (smartMIBG software, FUJIFILM RI Pharma, Japan) [21]. The operators selected a point at the center of the heart on the image, and then, a circular ROI was positioned on the heart. The subsequent processing was automatic, but can be modified manually if the location was inappropriate. A mediastinal ROI was determined as 30% of the height (center of the heart to the upper border of the mediastinum) and 10% of the body width. The circular cardiac ROI was similar to the setting in the study using the Anger camera. Since an upper mediastinum ROI could not be set for D-SPECT imaging, the highest mediastinal region of the lowest average count was selected. In a preliminary study, the inter-observer reproducibility of average mediastinal counts in the initial 40 data points was good, showing the first measurement = 1.01 × the second measurement −8 (r = 0.99, p < 0.0001).

Conversion of HMR between Anger camera and D-SPECT

To adjust HMR from D-SPECT to Anger camera conditions, the following equation was used:
Adjusted HMR = CC of Anger camera/CC of D-SPECT × (measured HMR − 1) + 1, and the effect of correction was examined.
In the next step, we used an MEGP collimator condition to standardize the HMR to provide better quantitative accuracy as stated in the European imaging proposal [8]. The average CC with MEGP was 0.88 [15], which is referred to herein as “standard ME88”. Based on the measured CC for the system, the standardized HMR to the ME88 condition was calculated as:
Standardized HMR = 0.88/CC of the institutional system × (measured HMR − 1) + 1.

Calculation of washout rate

Washout rate (WR) was calculated using the following formula for both the Anger camera and D-SPECT with early and late heart counts (H early, H late), mediastinal counts (M early, M late), and a decay correction factor (DCF):
$${\text{WR}} = ((H_{\text{early}} - M_{\text{early}} ) - (H_{\text{late}} - M_{\text{late}} )/{\text{DCF}})/(H_{\text{early}} - M_{\text{early}} ) \times 100,$$
$${\text{WR without background correction}} = (H_{\text{early}} - H_{\text{late}} /{\text{DCF}})/H_{\text{early}} \times 100,$$
where DCF = 0.5^(difference between early and late (h)/13).

Statistics

Data are shown as mean ± standard deviation (SD). Differences among groups were assessed using the one-way analysis of variance and Student’s t test. Paired values were analyzed using paired t tests with Bland–Altman plots and signed rank tests. Linear regression of the HMR between the two camera conditions was calculated using the least squares method. A variability of the average ROI count was also examined using coefficient of variation (CV, %). The statistics software was JMP version 12 (SAS Institute Inc., Cary, NC, USA), and we used Mathematica 11 (Wolfram Research Inc., Champaign, IL, USA) for some of the mathematical calculations.

Results

Phantom experiments and conversion coefficients

Figure 1 shows phantom images obtained with the Anger camera and D-SPECT planograms. Based on the two measurements, CC was calculated as 0.55 for the Anger camera with an LEHR collimator and 0.63 for the D-SPECT camera.

Cardiac and mediastinal counts

Cardiac and mediastinal counts per minute were compared between the Anger and D-SPECT cameras (Fig. 2). Linear correlation was good for both cardiac and mediastinal counts (R 2 = 0.95 for both), whereas acquired counts were higher with D-SPECT than the Anger camera. Mediastinal count variability was similar in two groups with lower (HMR < 1.6) and upper (HMR > 2.8) quartiles of HMR distribution. In the lower quartile group, mean count/pixel/min and CV (%) were 10.2 (22%) for Anger camera and 50.8 (23%) for D-SPECT. In the upper quartile group, they were 12.0 (24%) for Anger camera and 62.6 (25%) for D-SPECT.

Cross calibration of HMR ratio in the clinical study

Table 1 shows the HMR derived from the original Anger and D-SPECT images. Paired comparisons of the HMR derived from the original Anger and D-SPECT images showed that the latter was significantly higher (p < 0.0001), with a mean difference of 0.10. When HMR from D-SPECT was converted to the condition of the Anger camera with LEHR collimator, the difference between two systems disappeared (p = n. s.). Scatterplots and linear regression lines between HMR before and after correction showed that the conversion of HMR from D-SPECT to the Anger camera condition improved the systematic differences between the two camera systems (Fig. 3a, b).
Table 1
Original and standardized heart-to-mediastinum ratio
Camera-Collimator and CC
Original HMR—Anger camera
Original HMR—D-SPECT
D-SPECT adjusted to Anger LEHR condition
Anger LEHR standardized to ME88 condition
D-SPECT standardized to ME88 condition
Anger LEHR (CC = 0.55)
D-SPECT (CC = 0.63)
D-SPECT (CC = 0.63) to Anger LEHR (CC = 0.55) conditions
Anger LEHR (CC = 0.55) to Anger MEGP (CC = 0.88) conditions
D-SPECT (CC = 0.63) to Anger MEGP (CC = 0.88) conditions
Mean
1.76
1.86
1.75
2.21
2.20
SD
0.40
0.54
0.48
0.65
0.75
Minimum
1.03
0.93
0.94
1.05
0.90
Maximum
2.47
2.83
2.60
3.35
3.56
Analysis versus
Original HMR—Anger
Original HMR—Anger
Anger LEHR standardized to ME88
Mean difference
 
0.10
0.01
 
0.01
P
 
<0.0001
0.59
 
0.59
Wilcoxon signed rank
 
<0.0001
0.52
 
0.52
Correlation R
 
0.96
0.96
 
0.96
CC conversion coefficient, HMR heart-to-mediastinum ratio, LEHR low-energy high-resolution collimator, ME88 ME collimator with conversion coefficient of 0.88, SD standard deviation
Standardization to the ME88 condition similarly eliminated the difference between the Anger and D-SPECT findings. The average standardized HMR from the Anger camera (StdHMRAnger) and D-SPECT (StdHMRDSPECT) became comparable (2.21 ± 0.65 vs. 2.20 ± 0.75, p = n. s.) (Table 1). After standardization, a bivariate correlation plot showed good linearity: StdHMRDSPECT = −0.25 + 1.11 × StdHMRAnger (R 2 = 0.93; Fig. 3c).

Additional correction of HMR

Since standardized HMRDPSECT was slightly lower in the range of HMR < 1.3 and slightly higher in the range of HMR > 2.3 (Fig. 3c) compared with the standardized HMRAnger, further correction was attempted. As standardized HMR with Anger camera was calculated as StdHMRAnger = (StdHMRDSPECT + 0.30)/1.19 in the initial 40 data points, this regression equation was applied to the latter 40 data points for validation. Then, the bivariate correlation plot showed improved linearity: StdHMRDSPECT = 0.09 + 0.98 × StdHMRAnger (R 2 = 0.91; Fig. 4).

Washout rates

Washout rates from the D-SPECT and Anger cameras were compared with and without background correction (Fig. 5). Although they positively correlated (R 2 = 0.83 and 0.65, p < 0.0001, respectively), a few outliers persisted between the values derived from both cameras.

Discussion

The major purpose of this study was to create a conversion method between Anger and D-SPECT CZT cameras. Using CC values for D-SPECT image acquisition, we cross-calibrated HMR between Anger and D-SPECT camera systems and could also adjust the HMR to the ME88 condition. This cross calibration could enable the application of HMR to multicenter studies of patients with chronic HF and Lewy-body disease.

Need for standardization of MIBG parameters

Although HMR in a 123I-MIBG study is a simple parameter based on the average count ratio of the heart and mediastinum, a standardized approach is essential for diagnostic and prognostic evaluation [22]. Among various factors, the influence of the collimator on HMR calculations is too large to generate consistent results, particularly when the collimators are of low-(LE) and medium-(ME) energy. Several methods have been proposed, but we advocate using a calibration phantom that can be easily applied to any camera-collimator setting [15, 23]. The findings of 225 phantom experiments have shown that the key characteristics of collimators are not simply ME and LE, but can be more precisely defined, for example, as LEHR, LE general-purpose (LEGP), low–medium energy (LME), MEGP, and ME–low penetration (MELP) [15]. Whether or not a CC could be similarly determined for D-SPECT after the advent of the CZT camera has remained unknown. Even if CC could be measured, whether or not a D-SPECT HMR could be integrated with Anger camera conditions has not been determined.

Characteristics of HMR by D-SPECT

Anterior planar image-equivalent planograms generated by D-SPECT can serve as part of a quality control system for projection images, which was a convenient base for this study. Although we tried similar ROI settings, D-SPECT has some limitations. An upper mediastinal ROI cannot be set due to the vertical length of the view being 16 cm. We, therefore, tried to define the highest possible mediastinal region with the lowest average count, which corresponded to the mid position on the Anger camera image. If the large field of view CZT camera is available in future, the effect of small field of view on the accuracy of ROI setting could be validated. In addition, mediastinal count variability as examined by CV was not significantly different between the Anger camera and D-SPECT in patients with high and low HMR. However, an automatic ROI processing algorithm for D-SPECT could enhance reproducibility. The energy resolution of the CZT camera is better, which enabled better contrast in images derived from D-SPECT than from the Anger camera. The Compton scatter fraction might also differ between Anger and D-SPECT settings. Therefore, the CC determined in this study is a practical value with which to cross-calibrate the two camera conditions. The planogram is unique to the single D-SPECT system, whereas images from the Anger camera vary due to wide disparities among camera-collimator combinations.
Although the administration dose of 123I-MIBG was relatively low (111 MBq) compared with studies in the North America (370 MBq) and Europe (185 MBq), the image quality of planogram and SPECT was good by 10-min acquisition.

Comparison with ADRECARD study

The ADRECARD study was the first to compare HMR calibrated using D-SPECT and Anger cameras. A conversion equation for HMR (Corrected D-SPECT) = 0.5896 × HMR (D-SPECT) + 0.4649 was created based on a phantom experiment in that study [18]. Based on their original table and assuming that the LEHR of their camera collimator had a CC of 0.55, we tentatively calculated the standardized HMR, and found a good correlation even with our standardization method. However, HMR (D-SPECT) + 0.1 seemed to correlate more closely with the standardized HMR (Anger) [24]. Our HMR calculated with the Anger camera was higher than that in the ADRECARD study. Although agreement with our temporary calculation was generally good in the present study, the following factors should be considered. To calculate HMR, the square ROI over the heart applied in the ADRECARD study included a slight extra-cardiac area (lower heart count than ours), and a mediastinal rectangular region was placed in a lower position (higher mediastinal count than ours). Our heart ROI setting in the clinical study was circular and within the heart, and we identified the mediastinal region with the lowest count. As a result, the HMR calculated from the Anger camera using our algorithm was always higher. The location of the ROI could be a cause of variation in the clinical setting [25]. Although the ADRECARD study used 99mTc-tetrofosmin for localization of the heart, we did not use dual-nuclide acquisition to reduce the radiation burden and study cost. We obtained anterior planograms by localizing the heart by pre-test imaging for a short period. When the field of view is inappropriately located, measurements could be readily repeated using the high-sensitivity D-SPECT system.

Clinical implications

The difference in measured HMR between D-SPECT and Anger camera with an LEHR collimator was smaller compared with that between the Anger camera with LEHR and ME collimators. The CC with D-SPECT was between that of LEHR and LME collimators [15, 20, 23]. In multicenter prognostic studies, HMR values of 1.6–1.75 were thresholds for differentiating good and poor prognosis including cardiac death, serious arrhythmia, and progression of HF [1, 2, 4, 7]. When the linear regression line was observed, the impact of cross calibration was relatively small in the HMR range of <1.6. However, in the range of borderline to higher HMR, the discrepancy was increased between the two systems and appropriate correction methods should be used.
In a D-SPECT study, conversion of HMR to ME88 condition using conversion coefficient (0.63) works well around the HMR of 1.6–2.0 and can be used for clinical studies for differentiating good and poor prognosis. However, standardized HMR with D-SPECT showed slightly lower values in the HMR range of <1.3 and higher values in the HMR range of >2.3. This was probably due to lower mediastinal background and better contrast in D-SPECT study compared with the Anger camera condition. This systematic difference could be further corrected if we used regression line between standardized values. However, as the need for additional correction may depend on the individual D-SPECT system and acquisition conditions, further studies should be indicated in multiple centers, where D-SPECT is used.
If HMR could be consistently calculated in a wide range of HMR, possibility of using D-SPECT for a mortality risk model could be considered [26]. Since the uncorrected D-SPECT HMR was higher than that of the Anger camera with an LE collimator in the borderline to higher HMR range, a corrected HMR could avoid underestimating mortality risk, although further validation studies will be required.
Although the LEHR collimator is popular in the United States, many types of collimators other than LEHR such as LEGP, LME, MEGP, and MELP collimators are actually used in clinical practice. Therefore, CC can be applied to compare HMR from various conditions using the uniform acquisition conditions, such as ME88 condition and individual institutional LE collimator condition.

Washout rate

While the correlation between washout rates derived from the Anger camera and D-SPECT was also fair, reproducibility requires careful attention when background subtraction is applied to very low cardiac counts as seen in the outliers of washout rate plots between both cameras (Fig. 5). The need for background correction when patients have a low HMR should be further analyzed from both diagnostic and prognostic viewpoints [12].

Limitations

Only one D-SPECT and one Anger camera system were included in this study. Although D-SPECT planograms have no potential for variation, the adequacy of applying the present results to other hospitals should be further studied. We had already completed phantom experiments under 225 conditions at 84 institutions [15, 27] (>1000 conditions at present) in Japan by the end of 2016 and by that time studies under 210 conditions had also proceeded at 27 European institutions [20]. Although CC values are affected by specifications of the camera, collimators, detector crystals, and acquisition conditions, we postulate that the present findings could be applicable even for D-SPECT compared with other camera-collimator combinations. However, further studies are needed to validate this phantom methodology for universal applications. Three-dimensional SPECT quantitation was not used in the present study. Because whole heart quantitation has been achieved using MIBG imaging [28, 29], the potential variability of such three-dimensional methods including the need for an appropriate background, dependency of the results on software algorithm, and the relationship to conventional planar imaging should be further investigated. Whereas the current study confirmed that planar-equivalent HMR can be generated from D-SPECT images, whether this is the optimal use of the imaging capabilities of this D-SPECT system remains to be determined. Finally, in the clinical application, timing of the Anger and D-SPECT was not exactly the same. However, correlation of average counts was good between the Anger and D-SPECT cameras, and it has been shown that variation in acquisition time of 123I-MIBG between 2- and 4-h post-injection did not lead to a clinically significant change in the late H/M ratio [30].

Conclusion

The 123I-MIBG HMR can be similarly calculated with D-SPECT using a planogram as used in planar studies with an Anger camera. The HMR derived from D-SPECT can be calibrated to both LE collimator and ME collimator conditions using CC values based on institutional phantom experiments. A slight deviation of the regression line could be further improved using the regression line. The cross-calibration method supports diagnostic and prognostic uses of D-SPECT as used in Anger camera systems.

Acknowledgements

We appreciate the cooperation of technologists Shigeaki Hiko and Haruki Yamamoto. We also thank Dr. Tomofumi Yoshinaga, Department of Neurology, Public Hospital of Matto, Ishikawa, Hakusan, Japan for providing neurological data from the patients and Norma Foster for editorial assistance. This study was supported in part by Grants-in-Aid for Scientific Research in Japan (No. 15K09947; PI, Kenichi Nakajima).

Compliance with ethical standards

Conflict of interest

Kenichi Nakajima collaborates with FUJIFILM RI Pharma Co. Ltd. (Tokyo, Japan), supplier of 123I-MIBG in Japan, to develop software.
Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://​creativecommons.​org/​licenses/​by/​4.​0/​), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.

Unsere Produktempfehlungen

e.Med Interdisziplinär

Kombi-Abonnement

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

© Springer Medizin

Bis 11. April 2024 bestellen und im ersten Jahr 50 % sparen!

e.Med Radiologie

Kombi-Abonnement

Mit e.Med Radiologie erhalten Sie Zugang zu CME-Fortbildungen des Fachgebietes Radiologie, den Premium-Inhalten der radiologischen Fachzeitschriften, inklusive einer gedruckten Radiologie-Zeitschrift Ihrer Wahl.

© Springer Medizin

Bis 11. April 2024 bestellen und im ersten Jahr 50 % sparen!

Literatur
1.
Zurück zum Zitat Agostini D, Verberne HJ, Burchert W, Knuuti J, Povinec P, Sambuceti G, et al. I-123-mIBG myocardial imaging for assessment of risk for a major cardiac event in heart failure patients: insights from a retrospective European multicenter study. Eur J Nucl Med Mol Imaging. 2008;35:535–46.CrossRefPubMed Agostini D, Verberne HJ, Burchert W, Knuuti J, Povinec P, Sambuceti G, et al. I-123-mIBG myocardial imaging for assessment of risk for a major cardiac event in heart failure patients: insights from a retrospective European multicenter study. Eur J Nucl Med Mol Imaging. 2008;35:535–46.CrossRefPubMed
2.
Zurück zum Zitat Jacobson AF, Senior R, Cerqueira MD, Wong ND, Thomas GS, Lopez VA, et al. Myocardial iodine-123 meta-iodobenzylguanidine imaging and cardiac events in heart failure. Results of the prospective ADMIRE-HF (AdreView Myocardial Imaging for Risk Evaluation in Heart Failure) study. J Am Coll Cardiol. 2010;55:2212–21.CrossRefPubMed Jacobson AF, Senior R, Cerqueira MD, Wong ND, Thomas GS, Lopez VA, et al. Myocardial iodine-123 meta-iodobenzylguanidine imaging and cardiac events in heart failure. Results of the prospective ADMIRE-HF (AdreView Myocardial Imaging for Risk Evaluation in Heart Failure) study. J Am Coll Cardiol. 2010;55:2212–21.CrossRefPubMed
3.
Zurück zum Zitat Kuwabara Y, Tamaki N, Nakata T, Yamashina S, Yamazaki J. Determination of the survival rate in patients with congestive heart failure stratified by 123I-MIBG imaging: a meta-analysis from the studies performed in Japan. Ann Nucl Med. 2011;25:101–7.CrossRefPubMed Kuwabara Y, Tamaki N, Nakata T, Yamashina S, Yamazaki J. Determination of the survival rate in patients with congestive heart failure stratified by 123I-MIBG imaging: a meta-analysis from the studies performed in Japan. Ann Nucl Med. 2011;25:101–7.CrossRefPubMed
4.
Zurück zum Zitat Nakata T, Nakajima K, Yamashina S, Yamada T, Momose M, Kasama S, et al. A pooled analysis of multicenter cohort studies of 123I-mIBG imaging of sympathetic innervation for assessment of long-term prognosis in heart failure. JACC Cardiovasc Imaging. 2013;6:772–84.CrossRefPubMed Nakata T, Nakajima K, Yamashina S, Yamada T, Momose M, Kasama S, et al. A pooled analysis of multicenter cohort studies of 123I-mIBG imaging of sympathetic innervation for assessment of long-term prognosis in heart failure. JACC Cardiovasc Imaging. 2013;6:772–84.CrossRefPubMed
5.
Zurück zum Zitat Verschure DO, Veltman CE, Manrique A, Somsen GA, Koutelou M, Katsikis A, et al. For what endpoint does myocardial 123I-MIBG scintigraphy have the greatest prognostic value in patients with chronic heart failure? Results of a pooled individual patient data meta-analysis. Eur Heart J Cardiovasc Imaging. 2014;15:996–1003.CrossRefPubMed Verschure DO, Veltman CE, Manrique A, Somsen GA, Koutelou M, Katsikis A, et al. For what endpoint does myocardial 123I-MIBG scintigraphy have the greatest prognostic value in patients with chronic heart failure? Results of a pooled individual patient data meta-analysis. Eur Heart J Cardiovasc Imaging. 2014;15:996–1003.CrossRefPubMed
7.
Zurück zum Zitat Nakajima K, Nakata T. Cardiac 123I-MIBG Imaging for clinical decision making: 22-year experience in Japan. J Nucl Med. 2015;56(Suppl 4):11S–9S.CrossRefPubMed Nakajima K, Nakata T. Cardiac 123I-MIBG Imaging for clinical decision making: 22-year experience in Japan. J Nucl Med. 2015;56(Suppl 4):11S–9S.CrossRefPubMed
8.
Zurück zum Zitat Flotats A, Carrio I, Agostini D, Le Guludec D, Marcassa C, Schafers M, et al. Proposal for standardization of 123I-metaiodobenzylguanidine (MIBG) cardiac sympathetic imaging by the EANM Cardiovascular Committee and the European Council of Nuclear Cardiology. Eur J Nucl Med Mol Imaging. 2010;37:1802–12.CrossRefPubMed Flotats A, Carrio I, Agostini D, Le Guludec D, Marcassa C, Schafers M, et al. Proposal for standardization of 123I-metaiodobenzylguanidine (MIBG) cardiac sympathetic imaging by the EANM Cardiovascular Committee and the European Council of Nuclear Cardiology. Eur J Nucl Med Mol Imaging. 2010;37:1802–12.CrossRefPubMed
9.
Zurück zum Zitat Henzlova MJ, Duvall WL, Einstein AJ, Travin MI, Verberne HJ. ASNC imaging guidelines for SPECT nuclear cardiology procedures: stress, protocols, and tracers. J Nucl Cardiol. 2016;23:606–39.CrossRefPubMed Henzlova MJ, Duvall WL, Einstein AJ, Travin MI, Verberne HJ. ASNC imaging guidelines for SPECT nuclear cardiology procedures: stress, protocols, and tracers. J Nucl Cardiol. 2016;23:606–39.CrossRefPubMed
10.
Zurück zum Zitat King AE, Mintz J, Royall DR. Meta-analysis of 123I-MIBG cardiac scintigraphy for the diagnosis of Lewy body-related disorders. Mov Disord. 2011;26:1218–24.CrossRefPubMed King AE, Mintz J, Royall DR. Meta-analysis of 123I-MIBG cardiac scintigraphy for the diagnosis of Lewy body-related disorders. Mov Disord. 2011;26:1218–24.CrossRefPubMed
11.
Zurück zum Zitat Orimo S, Suzuki M, Inaba A, Mizusawa H. 123I-MIBG myocardial scintigraphy for differentiating Parkinson’s disease from other neurodegenerative parkinsonism: a systematic review and meta-analysis. Parkinsonism Relat Disord. 2012;18:494–500.CrossRefPubMed Orimo S, Suzuki M, Inaba A, Mizusawa H. 123I-MIBG myocardial scintigraphy for differentiating Parkinson’s disease from other neurodegenerative parkinsonism: a systematic review and meta-analysis. Parkinsonism Relat Disord. 2012;18:494–500.CrossRefPubMed
12.
Zurück zum Zitat Veltman CE, Boogers MJ, Meinardi JE, Al Younis I, Dibbets-Schneider P, Van der Wall EE, et al. Reproducibility of planar 123I-meta-iodobenzylguanidine (MIBG) myocardial scintigraphy in patients with heart failure. Eur J Nucl Med Mol Imaging. 2012;39:1599–608.CrossRefPubMedPubMedCentral Veltman CE, Boogers MJ, Meinardi JE, Al Younis I, Dibbets-Schneider P, Van der Wall EE, et al. Reproducibility of planar 123I-meta-iodobenzylguanidine (MIBG) myocardial scintigraphy in patients with heart failure. Eur J Nucl Med Mol Imaging. 2012;39:1599–608.CrossRefPubMedPubMedCentral
13.
Zurück zum Zitat Verberne HJ, Feenstra C, de Jong WM, Somsen GA, van Eck-Smit BL, Busemann Sokole E. Influence of collimator choice and simulated clinical conditions on 123I-MIBG heart/mediastinum ratios: a phantom study. Eur J Nucl Med Mol Imaging. 2005;32:1100–7.CrossRefPubMed Verberne HJ, Feenstra C, de Jong WM, Somsen GA, van Eck-Smit BL, Busemann Sokole E. Influence of collimator choice and simulated clinical conditions on 123I-MIBG heart/mediastinum ratios: a phantom study. Eur J Nucl Med Mol Imaging. 2005;32:1100–7.CrossRefPubMed
14.
Zurück zum Zitat Nakajima K, Okuda K, Matsuo S, Yoshita M, Taki J, Yamada M, et al. Standardization of metaiodobenzylguanidine heart to mediastinum ratio using a calibration phantom: effects of correction on normal databases and a multicentre study. Eur J Nucl Med Mol Imaging. 2012;39:113–9.CrossRefPubMed Nakajima K, Okuda K, Matsuo S, Yoshita M, Taki J, Yamada M, et al. Standardization of metaiodobenzylguanidine heart to mediastinum ratio using a calibration phantom: effects of correction on normal databases and a multicentre study. Eur J Nucl Med Mol Imaging. 2012;39:113–9.CrossRefPubMed
15.
Zurück zum Zitat Nakajima K, Okuda K, Yoshimura M, Matsuo S, Wakabayashi H, Imanishi Y, et al. Multicenter cross-calibration of I-123 metaiodobenzylguanidine heart-to-mediastinum ratios to overcome camera-collimator variations. J Nucl Cardiol. 2014;21:970–8.CrossRefPubMedPubMedCentral Nakajima K, Okuda K, Yoshimura M, Matsuo S, Wakabayashi H, Imanishi Y, et al. Multicenter cross-calibration of I-123 metaiodobenzylguanidine heart-to-mediastinum ratios to overcome camera-collimator variations. J Nucl Cardiol. 2014;21:970–8.CrossRefPubMedPubMedCentral
16.
Zurück zum Zitat Slomka PJ, Mehta PK, Germano G, Berman DS. Quantification of I-123-meta-iodobenzylguanidine heart-to-mediastinum ratios: not so simple after all. J Nucl Cardiol. 2014;21:979–83.CrossRefPubMed Slomka PJ, Mehta PK, Germano G, Berman DS. Quantification of I-123-meta-iodobenzylguanidine heart-to-mediastinum ratios: not so simple after all. J Nucl Cardiol. 2014;21:979–83.CrossRefPubMed
17.
Zurück zum Zitat Berman DS, Kang X, Tamarappoo B, Wolak A, Hayes SW, Nakazato R, et al. Stress thallium-201/rest technetium-99 m sequential dual isotope high-speed myocardial perfusion imaging. JACC Cardiovasc Imaging. 2009;2:273–82.CrossRefPubMed Berman DS, Kang X, Tamarappoo B, Wolak A, Hayes SW, Nakazato R, et al. Stress thallium-201/rest technetium-99 m sequential dual isotope high-speed myocardial perfusion imaging. JACC Cardiovasc Imaging. 2009;2:273–82.CrossRefPubMed
18.
Zurück zum Zitat Bellevre D, Manrique A, Legallois D, Bross S, Baavour R, Roth N, et al. First determination of the heart-to-mediastinum ratio using cardiac dual isotope (123I-MIBG/99mTc-tetrofosmin) CZT imaging in patients with heart failure: the ADRECARD study. Eur J Nucl Med Mol Imaging. 2015;42:1912–9.CrossRefPubMed Bellevre D, Manrique A, Legallois D, Bross S, Baavour R, Roth N, et al. First determination of the heart-to-mediastinum ratio using cardiac dual isotope (123I-MIBG/99mTc-tetrofosmin) CZT imaging in patients with heart failure: the ADRECARD study. Eur J Nucl Med Mol Imaging. 2015;42:1912–9.CrossRefPubMed
19.
Zurück zum Zitat Nakajima K, Matsubara K, Ishikawa T, Motomura N, Maeda R, Akhter N, et al. Correction of iodine-123-labeled meta-iodobenzylguanidine uptake with multi-window methods for standardization of the heart-to-mediastinum ratio. J Nucl Cardiol. 2007;14:843–51.CrossRefPubMed Nakajima K, Matsubara K, Ishikawa T, Motomura N, Maeda R, Akhter N, et al. Correction of iodine-123-labeled meta-iodobenzylguanidine uptake with multi-window methods for standardization of the heart-to-mediastinum ratio. J Nucl Cardiol. 2007;14:843–51.CrossRefPubMed
20.
Zurück zum Zitat Verschure DO, Poel E, Nakajima K, Okuda K, van Eck-Smit BL, Somsen GA, et al. A European myocardial 123I-mIBG cross-calibration phantom study. J Nucl Cardiol. 2017;. doi:10.1007/s12350-017-0782-6 (Epub ahead of print).PubMed Verschure DO, Poel E, Nakajima K, Okuda K, van Eck-Smit BL, Somsen GA, et al. A European myocardial 123I-mIBG cross-calibration phantom study. J Nucl Cardiol. 2017;. doi:10.​1007/​s12350-017-0782-6 (Epub ahead of print).PubMed
21.
Zurück zum Zitat Okuda K, Nakajima K, Hosoya T, Ishikawa T, Konishi T, Matsubara K, et al. Semi-automated algorithm for calculating heart-to-mediastinum ratio in cardiac Iodine-123 MIBG imaging. J Nucl Cardiol. 2011;18:82–9.CrossRefPubMed Okuda K, Nakajima K, Hosoya T, Ishikawa T, Konishi T, Matsubara K, et al. Semi-automated algorithm for calculating heart-to-mediastinum ratio in cardiac Iodine-123 MIBG imaging. J Nucl Cardiol. 2011;18:82–9.CrossRefPubMed
22.
Zurück zum Zitat Verberne HJ, Habraken JB, van Eck-Smit BL, Agostini D, Jacobson AF. Variations in 123I-metaiodobenzylguanidine (MIBG) late heart mediastinal ratios in chronic heart failure: a need for standardisation and validation. Eur J Nucl Med Mol Imaging. 2008;35:547–53.CrossRefPubMed Verberne HJ, Habraken JB, van Eck-Smit BL, Agostini D, Jacobson AF. Variations in 123I-metaiodobenzylguanidine (MIBG) late heart mediastinal ratios in chronic heart failure: a need for standardisation and validation. Eur J Nucl Med Mol Imaging. 2008;35:547–53.CrossRefPubMed
23.
Zurück zum Zitat Nakajima K, Verschure D, Okuda K, Verberne H. Standardization of 123I-meta-iodobenzylguanidine myocardial sympathetic activity imaging: phantom calibration and clinical applications. Clin Transl Imaging. 2017;5:255–63.CrossRefPubMedPubMedCentral Nakajima K, Verschure D, Okuda K, Verberne H. Standardization of 123I-meta-iodobenzylguanidine myocardial sympathetic activity imaging: phantom calibration and clinical applications. Clin Transl Imaging. 2017;5:255–63.CrossRefPubMedPubMedCentral
24.
Zurück zum Zitat Nakajima K, Okuda K, Matsuo S, Agostini D. The time has come to standardize 123I-MIBG heart-to-mediastinum ratios including planar and SPECT methods. Eur J Nucl Med Mol Imaging. 2016;43:386–8.CrossRefPubMed Nakajima K, Okuda K, Matsuo S, Agostini D. The time has come to standardize 123I-MIBG heart-to-mediastinum ratios including planar and SPECT methods. Eur J Nucl Med Mol Imaging. 2016;43:386–8.CrossRefPubMed
25.
Zurück zum Zitat Klene C, Jungen C, Okuda K, Kobayashi Y, Helberg A, Mester J, et al. Influence of ROI definition on the heart-to-mediastinum ratio in planar 123I-MIBG imaging. J Nucl Cardiol. 2016;. doi:10.1007/s12350-016-0708-8 (Epub ahead of print).PubMed Klene C, Jungen C, Okuda K, Kobayashi Y, Helberg A, Mester J, et al. Influence of ROI definition on the heart-to-mediastinum ratio in planar 123I-MIBG imaging. J Nucl Cardiol. 2016;. doi:10.​1007/​s12350-016-0708-8 (Epub ahead of print).PubMed
26.
Zurück zum Zitat Nakajima K, Nakata T, Matsuo S, Jacobson AF. Creation of mortality risk charts using 123I meta-iodobenzylguanidine heart-to-mediastinum ratio in patients with heart failure: 2- and 5-year risk models. Eur Heart J Cardiovasc Imaging. 2016;17:1138–45.CrossRefPubMed Nakajima K, Nakata T, Matsuo S, Jacobson AF. Creation of mortality risk charts using 123I meta-iodobenzylguanidine heart-to-mediastinum ratio in patients with heart failure: 2- and 5-year risk models. Eur Heart J Cardiovasc Imaging. 2016;17:1138–45.CrossRefPubMed
27.
Zurück zum Zitat Nakajima K, Scholte AJHA, Nakata T, Dimitriu-Leen AC, Chikamori T, Vitola JV, et al. Cardiac sympathetic nervous system imaging with 123I-meta-iodobenzylguanidine: perspectives from Japan and Europe. J Nucl Cardiol. 2017;24:952–60. doi:10.1007/s12350-017-0818-y.CrossRefPubMed Nakajima K, Scholte AJHA, Nakata T, Dimitriu-Leen AC, Chikamori T, Vitola JV, et al. Cardiac sympathetic nervous system imaging with 123I-meta-iodobenzylguanidine: perspectives from Japan and Europe. J Nucl Cardiol. 2017;24:952–60. doi:10.​1007/​s12350-017-0818-y.CrossRefPubMed
28.
Zurück zum Zitat Chen J, Folks RD, Verdes L, Manatunga DN, Jacobson AF, Garcia EV. Quantitative I-123 mIBG SPECT in differentiating abnormal and normal mIBG myocardial uptake. J Nucl Cardiol. 2012;19:92–9.CrossRefPubMed Chen J, Folks RD, Verdes L, Manatunga DN, Jacobson AF, Garcia EV. Quantitative I-123 mIBG SPECT in differentiating abnormal and normal mIBG myocardial uptake. J Nucl Cardiol. 2012;19:92–9.CrossRefPubMed
29.
Zurück zum Zitat Clements IP, Kelkar AA, Garcia EV, Butler J, Chen J, Folks R, et al. Prognostic significance of 123I-mIBG SPECT myocardial imaging in heart failure: differences between patients with ischaemic and non-ischaemic heart failure. Eur Heart J Cardiovasc Imaging. 2016;17:384–90.CrossRefPubMed Clements IP, Kelkar AA, Garcia EV, Butler J, Chen J, Folks R, et al. Prognostic significance of 123I-mIBG SPECT myocardial imaging in heart failure: differences between patients with ischaemic and non-ischaemic heart failure. Eur Heart J Cardiovasc Imaging. 2016;17:384–90.CrossRefPubMed
30.
Zurück zum Zitat Dimitriu-Leen AC, Gimelli A, Al Younis I, Veltman CE, Verberne HJ, Wolterbeek R, et al. The impact of acquisition time of planar cardiac 123I-MIBG imaging on the late heart to mediastinum ratio. Eur J Nucl Med Mol Imaging. 2016;43:326–32.CrossRefPubMed Dimitriu-Leen AC, Gimelli A, Al Younis I, Veltman CE, Verberne HJ, Wolterbeek R, et al. The impact of acquisition time of planar cardiac 123I-MIBG imaging on the late heart to mediastinum ratio. Eur J Nucl Med Mol Imaging. 2016;43:326–32.CrossRefPubMed
Metadaten
Titel
Cross calibration of 123I-meta-iodobenzylguanidine heart-to-mediastinum ratio with D-SPECT planogram and Anger camera
verfasst von
Kenichi Nakajima
Koichi Okuda
Kunihiko Yokoyama
Tatsuya Yoneyama
Shiro Tsuji
Hiroyuki Oda
Mitsuhiro Yoshita
Koji Kubota
Publikationsdatum
08.07.2017
Verlag
Springer Japan
Erschienen in
Annals of Nuclear Medicine / Ausgabe 8/2017
Print ISSN: 0914-7187
Elektronische ISSN: 1864-6433
DOI
https://doi.org/10.1007/s12149-017-1191-2

Weitere Artikel der Ausgabe 8/2017

Annals of Nuclear Medicine 8/2017 Zur Ausgabe