Routine evaluation of left ventricular function using CZT-SPECT, with low injected activities and limited recording times
verfasst von:
Marine Claudin, MD, Laetitia Imbert, PhD, Wassila Djaballah, MD, Nicolas Veran, PharmD, Sylvain Poussier, PhD, Véronique Roch, CRA, Mathieu Perrin, MD, Antoine Verger, MD, Henri Boutley, Tech, Gilles Karcher, MD, PhD, Pierre-Yves Marie, MD, PhD
Erschienen in:
Journal of Nuclear Cardiology
|
Ausgabe 1/2018
Einloggen, um Zugang zu erhalten
Abstract
Purpose
Gamma-cameras, with Cadmium-Zinc-Telluride (CZT) detectors, allow to perform myocardial perfusion imaging (MPI) with limited injected activities and recorded times. This study aimed at determining whether the routine assessment of left ventricular (LV) function with such limited counts protocols compares well with reference values from cardiac MRI.
Methods
The study included patients who have undergone cardiac MRI and an MPI routinely planned on a CZT camera with a low-dose protocol (120 MBq of Sestamibi for stress and 360 MBq at rest for 75 kg body weight), while targeting the recording of only 500 myocardial kcounts in order to limit the recording times (<10 minutes for stress, <4 minutes for rest). SPECT images were reconstructed with a method maintaining rather high spatial (8 mm) and temporal (16 frames/cycle) resolutions.
Results
Seventy-six patients were included, and mean effective dose was 3.5 ± 1.7 mSv for the total MPI protocol. Correlations between CZT-SPECT and MRI were good to excellent for ejection fraction (r
2 = 0.77), end-diastolic (r
2 = 0.88) and end-systolic (r
2 = 0.93) volumes, and the analysis of segmental contractility correlated well between the two techniques (kappa score = 0.72 ± 0.02).
Conclusion
LV function, assessed on a CZT camera with low injected activities and limited recording times, correlates well with the reference assessment from cardiac MRI.