27.02.2019 | Original Article
Left ventricular ejection fraction determined with the simulation of a very low-dose CZT-SPECT protocol and an additional count-calibration on planar radionuclide angiographic data
verfasst von:
Hubert Tissot, MD, Véronique Roch, MSc, Olivier Morel, MD, Nicolas Veran, PharmD, Mathieu Perrin, MD, Marine Claudin, MD, Antoine Verger, MD, PhD, Gilles Karcher, MD, PhD, Pierre-Yves Marie, MD, PhD, Laetitia Imbert, PhD
Erschienen in:
Journal of Nuclear Cardiology
|
Ausgabe 5/2019
Einloggen, um Zugang zu erhalten
Abstract
Purpose
To determine whether the left ventricular ejection fractions (EFs), measured on a high-sensitivity CZT single photon emission computed tomography (SPECT)-camera with a 70% reduction in recording times and a prevention of EF overestimation through an additional count-calibration, are concordant with reference EF from planar radionuclide angiography (2D-RNA).
Methods
An additional 10-minute CZT-SPECT recording was performed in patients referred to 2D-RNA for cardiomyopathy (n = 23) or chemotherapy monitoring (n = 50) with an in vivo red blood cell labeling with 850 MBq \( {}^{{99{\text{m}}}}{\text{TcO}}_{4}^{ - } \). The EF, obtained from CZT-SPECT with 100% (SPECT100) or 30% (SPECT30) projection times and with a SPECT-count calibration on the 2D-RNA counts of corresponding cavity volumes, were compared to EF from 2D-RNA.
Results
Strong and equivalent relationships were documented between the EF from 2D-RNA and the calibrated EF from SPECT100 (y = 0.89x + 6.62; R2 = 0.87) and SPECT30 (y = 0.87x + 8.40; R2 = 0.85), and the mean EF from SPECT100 (54% ± 15%) and SPECT30 (53% ± 16%) were close to that from 2D-RNA (55% ± 15%). However, upward shifts in these mean values were documented in the absence of count calibration for both SPECT100 (60% ± 18%) and SPECT30 (60% ± 18%).
Conclusion
Left ventricular EF may be determined on a high-sensitivity CZT-camera, a 70% reduction in injected activities, and an additional count-calibration for further enhancing the concordance with 2D-RNA values.