05.08.2019 | ORIGINAL ARTICLE

Head-to-head comparison of a CZT-based all-purpose SPECT camera and a dedicated CZT cardiac device for myocardial perfusion and functional analysis

verfasst von: Alessia Gimelli, MD, Riccardo Liga, MD, Michelangelo Bertasi, RT, Annette Kusch, RT, Paolo Marzullo, MD

Erschienen in: Journal of Nuclear Cardiology | Ausgabe 4/2021

Einloggen, um Zugang zu erhalten

Abstract

Purpose

To compare the outputs of a novel all-purpose SPECT camera equipped with CZT detectors (Discovery NM/CT 670) with the state-of-the-art represented by a dedicated CZT (Alcyone, Discovery 530c) cardiac camera in patients submitted to myocardial perfusion imaging (MPI).

Methods

We included 19 patients that underwent sequential low-dose 99mTc-tetrofosmin (148-185 MBq during stress and 296-370 MBq at rest) MPI with Alcyone and Discovery 670 cameras. Quantitative (% tracer’s uptake) and semi-quantitative analyses of perfusion data were performed for each scan. Moreover, major left ventricular (LV) functional and structural parameters were derived from each camera and compared.

Results

The two cameras showed excellent correlation for segmental myocardial % uptake at stress (R = 0.90; P < 0.001) and at rest (R = 0.88; P < 0.001) with narrow Bland-Altman limits of agreement. The level of diagnostic agreement of Discovery 670 and Alcyone cameras regarding perfusion analysis was excellent (Cohen’s κ 0.85). Similarly, the two cameras showed excellent correlation in the evaluation of LV ejection fraction (R = 0.95), peak filling rate (R = 0.97), and mass (R = 0.98).

Conclusions

Our preliminary results suggest that MPI with an all-purpose Discovery 670 CZT-SPECT camera is feasible, comparing well with the current state-of-the-art technology.
Anhänge
Nur mit Berechtigung zugänglich
Literatur
1.
Zurück zum Zitat Douglas PS, Hoffmann U, Patel MR, Mark DB, Al-Khalidi HR, Cavanaugh B, et al. Outcomes of anatomical versus functional testing for coronary artery disease. N Engl J Med 2015;372:1291-300.CrossRef Douglas PS, Hoffmann U, Patel MR, Mark DB, Al-Khalidi HR, Cavanaugh B, et al. Outcomes of anatomical versus functional testing for coronary artery disease. N Engl J Med 2015;372:1291-300.CrossRef
2.
Zurück zum Zitat Abbott BG, Case JA, Dorbala S, Einstein AJ, Galt JR, Pagnanelli R, et al. Contemporary cardiac SPECT imaging-innovations and best practices: An information statement from the American Society of Nuclear Cardiology. Circ Cardiovasc Imaging 2018;11:e000020.CrossRef Abbott BG, Case JA, Dorbala S, Einstein AJ, Galt JR, Pagnanelli R, et al. Contemporary cardiac SPECT imaging-innovations and best practices: An information statement from the American Society of Nuclear Cardiology. Circ Cardiovasc Imaging 2018;11:e000020.CrossRef
3.
Zurück zum Zitat Herzog BA, Buechel RR, Katz R, Brueckner M, Husmann L, Burger IA, et al. Nuclear myocardial perfusion imaging with a cadmium–zinc–telluride detector technique: Optimized protocol for scan time reduction. J Nucl Med 2010;51:46-51.CrossRef Herzog BA, Buechel RR, Katz R, Brueckner M, Husmann L, Burger IA, et al. Nuclear myocardial perfusion imaging with a cadmium–zinc–telluride detector technique: Optimized protocol for scan time reduction. J Nucl Med 2010;51:46-51.CrossRef
4.
Zurück zum Zitat Nkoulou R, Pazhenkottil AP, Kuest SM, Ghadri JR, Wolfrum M, Husmann L, et al. Semiconductor detectors allow low-dose-low-dose 1-day SPECT myocardial perfusion imaging. J Nucl Med 2011;52:1204-9.CrossRef Nkoulou R, Pazhenkottil AP, Kuest SM, Ghadri JR, Wolfrum M, Husmann L, et al. Semiconductor detectors allow low-dose-low-dose 1-day SPECT myocardial perfusion imaging. J Nucl Med 2011;52:1204-9.CrossRef
5.
Zurück zum Zitat Gimelli A, Bottai M, Giorgetti A, Genovesi D, Kusch A, Ripoli A, et al. Comparison between ultrafast and standard single-photon emission CT in patients with coronary artery disease: A pilot study. Circ Cardiovasc Imaging 2011;4:51-8.CrossRef Gimelli A, Bottai M, Giorgetti A, Genovesi D, Kusch A, Ripoli A, et al. Comparison between ultrafast and standard single-photon emission CT in patients with coronary artery disease: A pilot study. Circ Cardiovasc Imaging 2011;4:51-8.CrossRef
6.
Zurück zum Zitat Gimelli A, Liga R, Duce V, Kusch A, Clemente A, Marzullo P. Accuracy of myocardial perfusion imaging in detecting multivessel coronary artery disease: A cardiac CZT study. J Nucl Cardiol 2017;24:687-95.CrossRef Gimelli A, Liga R, Duce V, Kusch A, Clemente A, Marzullo P. Accuracy of myocardial perfusion imaging in detecting multivessel coronary artery disease: A cardiac CZT study. J Nucl Cardiol 2017;24:687-95.CrossRef
7.
Zurück zum Zitat Fiechter M, Gebhard C, Fuchs TA, Ghadri JR, Stehli J, Kazakauskaite E, et al. Cadmium-zinc-telluride myocardial perfusion imaging in obese patients. J Nucl Med 2012;53:1401-6.CrossRef Fiechter M, Gebhard C, Fuchs TA, Ghadri JR, Stehli J, Kazakauskaite E, et al. Cadmium-zinc-telluride myocardial perfusion imaging in obese patients. J Nucl Med 2012;53:1401-6.CrossRef
8.
Zurück zum Zitat Keidar Z, Raysberg I, Lugassi R, Frenkel A, Israel O. Novel cadmium zinc telluride based detector general purpose gamma camera: Initial evaluation and comparison with a standard camera. J Nucl Med. 2016;57:259. Keidar Z, Raysberg I, Lugassi R, Frenkel A, Israel O. Novel cadmium zinc telluride based detector general purpose gamma camera: Initial evaluation and comparison with a standard camera. J Nucl Med. 2016;57:259.
9.
Zurück zum Zitat Gregoire B, Pina-Jomir G, Bani-Sadr A, Moreau-Triby C, Janier M, Scheiber C. Four-minute bone SPECT using large-field cadmium–zinc–telluride camera. Clin Nucl Med 2018;43:389-95.CrossRef Gregoire B, Pina-Jomir G, Bani-Sadr A, Moreau-Triby C, Janier M, Scheiber C. Four-minute bone SPECT using large-field cadmium–zinc–telluride camera. Clin Nucl Med 2018;43:389-95.CrossRef
10.
11.
Zurück zum Zitat Gimelli A, Liga R, Clemente A, Pasanisi EM, Favilli B, Marzullo P. Appropriate choice of stress modality in patients undergoing myocardial perfusion scintigraphy with a cardiac camera equipped with solid-state detectors: the role of diabetes mellitus. Eur Heart J Cardiovasc Imaging 2018;19:1268-75.CrossRef Gimelli A, Liga R, Clemente A, Pasanisi EM, Favilli B, Marzullo P. Appropriate choice of stress modality in patients undergoing myocardial perfusion scintigraphy with a cardiac camera equipped with solid-state detectors: the role of diabetes mellitus. Eur Heart J Cardiovasc Imaging 2018;19:1268-75.CrossRef
12.
Zurück zum Zitat Gimelli A, Liga R, Pasanisi EM, Casagranda M, Marzullo P. Myocardial ischemia in the absence of obstructive coronary lesion: The role of post-stress diastolic dysfunction in detecting early coronary atherosclerosis. J Nucl Cardiol 2017;24:1542-50.CrossRef Gimelli A, Liga R, Pasanisi EM, Casagranda M, Marzullo P. Myocardial ischemia in the absence of obstructive coronary lesion: The role of post-stress diastolic dysfunction in detecting early coronary atherosclerosis. J Nucl Cardiol 2017;24:1542-50.CrossRef
13.
Zurück zum Zitat Kapur A, Latus KA, Davies G, Dhawan RT, Eastick S, Jarritt PH, et al. A comparison of three radionuclide myocardial perfusion tracers in clinical practice: the ROBUST study. Eur J Nucl Med Mol Imaging 2002;29:1608-16.CrossRef Kapur A, Latus KA, Davies G, Dhawan RT, Eastick S, Jarritt PH, et al. A comparison of three radionuclide myocardial perfusion tracers in clinical practice: the ROBUST study. Eur J Nucl Med Mol Imaging 2002;29:1608-16.CrossRef
14.
Zurück zum Zitat Gimelli A, Liga R, Startari U, Giorgetti A, Pieraccini L, Marzullo P. Evaluation of ischaemia in patients with atrial fibrillation: impact of stress protocol on myocardial perfusion imaging accuracy. Eur Heart J Cardiovasc Imaging 2015;16:781-7.CrossRef Gimelli A, Liga R, Startari U, Giorgetti A, Pieraccini L, Marzullo P. Evaluation of ischaemia in patients with atrial fibrillation: impact of stress protocol on myocardial perfusion imaging accuracy. Eur Heart J Cardiovasc Imaging 2015;16:781-7.CrossRef
16.
Zurück zum Zitat Agostini D, Marie PY, Ben-Haim S, Rouzet F, Songy B, Giordano A, et al. Performance of cardiac cadmium-zinc-telluride gamma camera imaging in coronary artery disease: a review from the cardiovascular committee of the European Association of Nuclear Medicine (EANM). Eur J Nucl Med Mol Imaging 2016;43:2423-32.CrossRef Agostini D, Marie PY, Ben-Haim S, Rouzet F, Songy B, Giordano A, et al. Performance of cardiac cadmium-zinc-telluride gamma camera imaging in coronary artery disease: a review from the cardiovascular committee of the European Association of Nuclear Medicine (EANM). Eur J Nucl Med Mol Imaging 2016;43:2423-32.CrossRef
17.
Zurück zum Zitat Hindorf C, Oddstig J, Hedeer F, Hansson MJ, Jögi J, Engblom H. Importance of correct patient positioning in myocardial perfusion SPECT when using a CZT camera. J Nucl Cardiol 2014;21:695-702.CrossRef Hindorf C, Oddstig J, Hedeer F, Hansson MJ, Jögi J, Engblom H. Importance of correct patient positioning in myocardial perfusion SPECT when using a CZT camera. J Nucl Cardiol 2014;21:695-702.CrossRef
Metadaten
Titel
Head-to-head comparison of a CZT-based all-purpose SPECT camera and a dedicated CZT cardiac device for myocardial perfusion and functional analysis
verfasst von
Alessia Gimelli, MD
Riccardo Liga, MD
Michelangelo Bertasi, RT
Annette Kusch, RT
Paolo Marzullo, MD
Publikationsdatum
05.08.2019
Verlag
Springer International Publishing
Erschienen in
Journal of Nuclear Cardiology / Ausgabe 4/2021
Print ISSN: 1071-3581
Elektronische ISSN: 1532-6551
DOI
https://doi.org/10.1007/s12350-019-01835-4

Weitere Artikel der Ausgabe 4/2021

Journal of Nuclear Cardiology 4/2021 Zur Ausgabe

Update Kardiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.