Semi-invasive measurement of cardiac output based on pulse contour: a review and analysis

verfasst von: Thomas Schlöglhofer, BSc, Hermann Gilly, PhD, Heinrich Schima, PhD

Erschienen in: Canadian Journal of Anesthesia/Journal canadien d'anesthésie | Ausgabe 5/2014

Einloggen, um Zugang zu erhalten

Abstract

Purpose

The aim of this review was to provide a meta-analysis of all five of the most popular systems for arterial pulse contour analysis compared with pulmonary artery thermodilution, the established reference method for measuring cardiac output (CO). The five investigated systems are FloTrac/Vigileo®, PiCCO®, LiDCO/PulseCO®, PRAM/MostCare®, and Modelflow.

Source

In a comprehensive literature search through MEDLINE®, Web of Knowledge (v.5.11), and Google Scholar, we identified prospective studies and reviews that compared the pulse contour approach with the reference method (n = 316). Data extracted from the 93 selected studies included range and mean cardiac output, bias, percentage error, software versions, and study population. We performed a pooled weighted analysis of their precision in determining CO in various patient groups and clinical settings.

Principal findings

Results of the majority of studies indicate that the five investigated systems show acceptable accuracy during hemodynamically stable conditions. Forty-three studies provided adequate data for a pooled weighted analysis and resulted in a mean (SD) total pooled bias of −0.28 (1.25) L·min−1, percentage error of 40%, and a correlation coefficient of r = 0.71. In hemodynamically unstable patients (n = 8), we found a higher percentage error (45%) and bias of −0.54 (1.64) L·min−1.

Conclusion

During hemodynamic instability, CO measurement based on continuous arterial pulse contour analysis shows only limited agreement with intermittent bolus thermodilution. The calibrated systems seem to deliver more accurate measurements than the auto-calibrated or the non-calibrated systems. For reliable use of these semi-invasive systems, especially for critical therapeutic decisions during hemodynamic disorders, both a strategy for hemodynamic optimization and further technological improvements are necessary.
Anhänge
Nur mit Berechtigung zugänglich
Fußnoten
1
Hatib F, Roteliuk L, Pearce J (inventors). Pulse contour method and apparatus for continuous assessment of a cardiovascular parameter. International patent publication WO 2006/113337 A2, 2006 Oct. 26.
 
2
Roteliuk L (inventor). Arterial pressure-based automatic determination of a cardiovascular parameter. International patent publication WO2005/055825 A1, 2005 June 23.
 
3
LiDCO Ltd. User’s Manual LiDCO Rapid-Fluid management just got easier. http://​www.​lidcorapid.​co.​uk/​pdfs/​english-rapid-v1.​04-user-manual.​pdf (accessed February 2014).
 
4
Edwards. Lifesciences Inc. FloTrac System 3rd Generation Software. Available from URL: http://​www.​edwards.​com/​eu/​products/​mininvasive/​Pages/​flotrac3g.​aspx (accessed February 2014).
 
6
Edwards Lifesciences Inc. EV1000 Clinical Platform 510(k) Summary. Available from URL: http://​www.​accessdata.​fda.​gov/​cdrh_​docs/​pdf10/​K100709.​pdf (accessed February 2014).
 
7
Joeken S, Fahle M, Pfeiffer UJ (inventors). Devices for in-vivo determination of the compliance function and the systemic blood flow of a living being. US patent US 6315735 B1.
 
8
Band DM, Linton RA, O’Brien TK (inventors). Method and apparatus for the measurement of cardiac output. International patent publication WO 97/24982 A.
 
9
Band MS, Linton WM, Linton RA, O’Brien KT (inventors). Verfahren und Vorrichtung zum Messen der Herzleistung. DE patent 697 23 847 T2. 2004 Jun. 03.
 
10
Romano S (inventor). Method and apparatus for measuring cardiac output. US patent US 6758822 B2.
 
Literatur
1.
Zurück zum Zitat Koo KK, Sun JC, Zhou Q, et al. Pulmonary artery catheters: evolving rates and reasons for use. Crit Care Med 2011; 39: 1613-8.PubMed Koo KK, Sun JC, Zhou Q, et al. Pulmonary artery catheters: evolving rates and reasons for use. Crit Care Med 2011; 39: 1613-8.PubMed
2.
Zurück zum Zitat Sandham JD, Hull RD, Brant RF, et al. Canadian Critical Care Clinical Trials Group: a randomized, controlled trial of the use of pulmonary-artery catheters in high-risk surgical patients. N Engl J Med 2003; 348: 5-14.PubMed Sandham JD, Hull RD, Brant RF, et al. Canadian Critical Care Clinical Trials Group: a randomized, controlled trial of the use of pulmonary-artery catheters in high-risk surgical patients. N Engl J Med 2003; 348: 5-14.PubMed
3.
Zurück zum Zitat Evans DC, Doraiswamy VA, Prosciak MP, et al. Complications associated with pulmonary artery catheters: a comprehensive clinical review. Scand J Surg 2009; 98: 199-208.PubMed Evans DC, Doraiswamy VA, Prosciak MP, et al. Complications associated with pulmonary artery catheters: a comprehensive clinical review. Scand J Surg 2009; 98: 199-208.PubMed
4.
Zurück zum Zitat Connors AF, Speroff T, Dawson NV, et al. The effectiveness of right heart catheterization in the initial care of critically ill patients. JAMA 1996; 276: 889-97.PubMed Connors AF, Speroff T, Dawson NV, et al. The effectiveness of right heart catheterization in the initial care of critically ill patients. JAMA 1996; 276: 889-97.PubMed
5.
Zurück zum Zitat Schwann NM, Hillel Z, Hoeft A, et al. Lack of effectiveness of the pulmonary artery catheter in cardiac surgery. Anesth Analg 2011; 113: 994-1002.PubMed Schwann NM, Hillel Z, Hoeft A, et al. Lack of effectiveness of the pulmonary artery catheter in cardiac surgery. Anesth Analg 2011; 113: 994-1002.PubMed
6.
Zurück zum Zitat Rajaram SS, Desai NK, Kalra A, et al. Pulmonary artery catheters for adult patients in intensive care. Cochrane Database Syst Rev 2013; 2: CD003408. Rajaram SS, Desai NK, Kalra A, et al. Pulmonary artery catheters for adult patients in intensive care. Cochrane Database Syst Rev 2013; 2: CD003408.
8.
Zurück zum Zitat Benington S, Ferris P, Nirmalanc M. Emerging trends in minimally invasive haemodynamic monitoring and optimization of fluid therapy. Eur J Anaesth 2009; 26: 893-905. Benington S, Ferris P, Nirmalanc M. Emerging trends in minimally invasive haemodynamic monitoring and optimization of fluid therapy. Eur J Anaesth 2009; 26: 893-905.
9.
Zurück zum Zitat Campos MML, Sellés FA, de Vera GG, et al. Techniques available for hemodynamic monitoring. Advantages and limitations. Med Intensiva 2012; 36: 434-44. Campos MML, Sellés FA, de Vera GG, et al. Techniques available for hemodynamic monitoring. Advantages and limitations. Med Intensiva 2012; 36: 434-44.
10.
Zurück zum Zitat Drummond KE, Murphy M. Minimally invasive cardiac output monitors. Continuing Education in Anaesthesia, Crit Care Pain 2012; 12: 5-10. Drummond KE, Murphy M. Minimally invasive cardiac output monitors. Continuing Education in Anaesthesia, Crit Care Pain 2012; 12: 5-10.
11.
Zurück zum Zitat Litton E, Morgan M. The PiCCO Monitor: a review. Anaesth Intensive Care 2012; 40: 393-409.PubMed Litton E, Morgan M. The PiCCO Monitor: a review. Anaesth Intensive Care 2012; 40: 393-409.PubMed
12.
Zurück zum Zitat Frank O. Die Grundform des arteriellen Pulses. Erste Abhandlung. Mathematische Analyse. Zeitschrift für Biologie 1899; 37: 485-526. Frank O. Die Grundform des arteriellen Pulses. Erste Abhandlung. Mathematische Analyse. Zeitschrift für Biologie 1899; 37: 485-526.
13.
Zurück zum Zitat Erlanger J, Hooker DR. An experimental study of blood pressure and of pulse pressure in man. John Hopkins Hospital Records 1904; 12: 145-378. Erlanger J, Hooker DR. An experimental study of blood pressure and of pulse pressure in man. John Hopkins Hospital Records 1904; 12: 145-378.
14.
Zurück zum Zitat Rhodes A, Sunderland R. Arterial pulse power analysis: the LiDCO™ plus system. In: Pinsky MR, Payen D, editors. Functional Hemodynamic Monitoring-Update in Intensive Care and Emergency Medicine 42. Berlin, Heidelberg: Springer; 2005. p. 183-92. Rhodes A, Sunderland R. Arterial pulse power analysis: the LiDCO™ plus system. In: Pinsky MR, Payen D, editors. Functional Hemodynamic Monitoring-Update in Intensive Care and Emergency Medicine 42. Berlin, Heidelberg: Springer; 2005. p. 183-92.
15.
Zurück zum Zitat Maus TM, Lee DE. Arterial pressure-based cardiac output assessment. J Cardiothorac Vasc Anesth 2008; 22: 468-73.PubMed Maus TM, Lee DE. Arterial pressure-based cardiac output assessment. J Cardiothorac Vasc Anesth 2008; 22: 468-73.PubMed
16.
Zurück zum Zitat Romagnoli S, Romano SM, Bevilacqua S, et al. Dynamic response of liquid-filled catheter systems for measurement of blood pressure: precision of measurements and reliability of the pressure recording analytical method with different disposable systems. J Crit Care 2011; 26: 415-22.PubMed Romagnoli S, Romano SM, Bevilacqua S, et al. Dynamic response of liquid-filled catheter systems for measurement of blood pressure: precision of measurements and reliability of the pressure recording analytical method with different disposable systems. J Crit Care 2011; 26: 415-22.PubMed
17.
Zurück zum Zitat Thiele RH, Durieux ME. Arterial waveform analysis for the anesthesiologist: past, present, and future concepts. Anesth Analg 2011; 113: 766-76.PubMed Thiele RH, Durieux ME. Arterial waveform analysis for the anesthesiologist: past, present, and future concepts. Anesth Analg 2011; 113: 766-76.PubMed
18.
Zurück zum Zitat Alhashemi JA, Cecconi M, Hofer CK. Cardiac output monitoring: an integrative perspective. Crit Care 2011; 15: 214.PubMedCentralPubMed Alhashemi JA, Cecconi M, Hofer CK. Cardiac output monitoring: an integrative perspective. Crit Care 2011; 15: 214.PubMedCentralPubMed
19.
Zurück zum Zitat Morgan P, Al-Subaie N, Rhodes A. Minimally invasive cardiac output monitoring. Curr Opin Crit Care 2008; 14: 322-6.PubMed Morgan P, Al-Subaie N, Rhodes A. Minimally invasive cardiac output monitoring. Curr Opin Crit Care 2008; 14: 322-6.PubMed
20.
Zurück zum Zitat Funk DJ, Moretti EW, Gan TJ. Minimally invasive cardiac output monitoring in the perioperative setting. Anesth Analg 2009; 108: 887-97.PubMed Funk DJ, Moretti EW, Gan TJ. Minimally invasive cardiac output monitoring in the perioperative setting. Anesth Analg 2009; 108: 887-97.PubMed
21.
Zurück zum Zitat de Wilde RBP, Schreuder JJ, van den Berg PC, Jansen JR. An evaluation of cardiac output by five arterial pulse contour techniques during cardiac surgery. Anaesthesia 2007; 62: 760-8.PubMed de Wilde RBP, Schreuder JJ, van den Berg PC, Jansen JR. An evaluation of cardiac output by five arterial pulse contour techniques during cardiac surgery. Anaesthesia 2007; 62: 760-8.PubMed
22.
Zurück zum Zitat Mayer J, Boldt J, Poland R, et al. A continuous arterial pressure waveform-based cardiac output using the FloTrac/Vigileo: a review and meta-analysis. J Cardiothorac Vasc Anesth 2009; 23: 401-6.PubMed Mayer J, Boldt J, Poland R, et al. A continuous arterial pressure waveform-based cardiac output using the FloTrac/Vigileo: a review and meta-analysis. J Cardiothorac Vasc Anesth 2009; 23: 401-6.PubMed
23.
Zurück zum Zitat Peyton PJ, Chong SW. Minimally invasive measurement of cardiac output during surgery and critical care: a meta-analysis of accuracy and precision. Anesthesiology 2010; 113: 1220-35.PubMed Peyton PJ, Chong SW. Minimally invasive measurement of cardiac output during surgery and critical care: a meta-analysis of accuracy and precision. Anesthesiology 2010; 113: 1220-35.PubMed
24.
Zurück zum Zitat Camporota L, Beale R. Pitfalls in haemodynamic monitoring based on the arterial pressure waveform. Crit Care 2010; 14: 124.PubMedCentralPubMed Camporota L, Beale R. Pitfalls in haemodynamic monitoring based on the arterial pressure waveform. Crit Care 2010; 14: 124.PubMedCentralPubMed
25.
Zurück zum Zitat Marik PE. Noninvasive Cardiac Output Monitors: a State-of the-Art Review. J Cardiothorac Vasc Anesth 2013; 27: 121-34.PubMed Marik PE. Noninvasive Cardiac Output Monitors: a State-of the-Art Review. J Cardiothorac Vasc Anesth 2013; 27: 121-34.PubMed
26.
Zurück zum Zitat Montenij LJ, de Waal EE, Buhre WF. Arterial waveform analysis in anesthesia and critical care. Curr Opin Anaesthesiol 2011; 24: 651-6.PubMed Montenij LJ, de Waal EE, Buhre WF. Arterial waveform analysis in anesthesia and critical care. Curr Opin Anaesthesiol 2011; 24: 651-6.PubMed
27.
Zurück zum Zitat Stover JF, Stocker R, Lenherr R, et al. Noninvasive cardiac output and blood pressure monitoring cannot replace an invasive monitoring system in critically ill patients. BMC Anesthesiol 2009; 9: 6.PubMedCentralPubMed Stover JF, Stocker R, Lenherr R, et al. Noninvasive cardiac output and blood pressure monitoring cannot replace an invasive monitoring system in critically ill patients. BMC Anesthesiol 2009; 9: 6.PubMedCentralPubMed
28.
Zurück zum Zitat Bogert LW, Wesseling KH, Schraa O, et al. Pulse contour cardiac output derived from non-invasive arterial pressure in cardiovascular disease. Anaesthesia 2010; 65: 1119-25.PubMed Bogert LW, Wesseling KH, Schraa O, et al. Pulse contour cardiac output derived from non-invasive arterial pressure in cardiovascular disease. Anaesthesia 2010; 65: 1119-25.PubMed
29.
Zurück zum Zitat Hohn A, Defosse JM, Becker S, Steffen C, Wappler F, Sakka SG. Non-invasive continuous arterial pressure monitoring with Nexfin® does not sufficiently replace invasive measurements in critically ill patients. Br J Anaesth 2013; 111: 178-84.PubMed Hohn A, Defosse JM, Becker S, Steffen C, Wappler F, Sakka SG. Non-invasive continuous arterial pressure monitoring with Nexfin® does not sufficiently replace invasive measurements in critically ill patients. Br J Anaesth 2013; 111: 178-84.PubMed
30.
Zurück zum Zitat Critchley LA, Critchley JA. A meta-analysis of studies using bias and precision statistics to compare cardiac output measurement techniques. J Clin Monit Comput 1999; 15: 85-91.PubMed Critchley LA, Critchley JA. A meta-analysis of studies using bias and precision statistics to compare cardiac output measurement techniques. J Clin Monit Comput 1999; 15: 85-91.PubMed
31.
Zurück zum Zitat Hunter JE, Schmidt FL. Methods of Meta-analysis: correcting error and Bias in Research Findings. 2nd ed. Newbury Park, CA: Sage; 2004: 89. Hunter JE, Schmidt FL. Methods of Meta-analysis: correcting error and Bias in Research Findings. 2nd ed. Newbury Park, CA: Sage; 2004: 89.
32.
Zurück zum Zitat Gruenewald M, Renner J, Meybohm P, Hocker J, Scholz J, Bein B. Reliability of continuous cardiac output measurement during intra-abdominal hypertension relies on repeated calibrations: an experimental animal study. Critical Care 2008; 12: R132.PubMedCentralPubMed Gruenewald M, Renner J, Meybohm P, Hocker J, Scholz J, Bein B. Reliability of continuous cardiac output measurement during intra-abdominal hypertension relies on repeated calibrations: an experimental animal study. Critical Care 2008; 12: R132.PubMedCentralPubMed
33.
Zurück zum Zitat Gruenewald M, Meybohm P, Renner J, et al. Effect of norepinephrine dosage and calibration frequency on accuracy of pulse contour-derived cardiac output. Crit Care 2011; 15: R22.PubMedCentralPubMed Gruenewald M, Meybohm P, Renner J, et al. Effect of norepinephrine dosage and calibration frequency on accuracy of pulse contour-derived cardiac output. Crit Care 2011; 15: R22.PubMedCentralPubMed
34.
Zurück zum Zitat Rodig G, Prasser C, Keyl C, Liebold A, Hobbhahn J. Continuous cardiac output measurement: pulse contour analysis vs TD technique in cardiac surgical patients. Br J Anaesth 1999; 82: 525-30.PubMed Rodig G, Prasser C, Keyl C, Liebold A, Hobbhahn J. Continuous cardiac output measurement: pulse contour analysis vs TD technique in cardiac surgical patients. Br J Anaesth 1999; 82: 525-30.PubMed
35.
Zurück zum Zitat Godje O, Höke K, Goetz AE, et al. Reliability of a new algorithm for continuous cardiac output determination by pulse-contour analysis during hemodynamic instability. Crit Care Med 2002; 30: 52-8.PubMed Godje O, Höke K, Goetz AE, et al. Reliability of a new algorithm for continuous cardiac output determination by pulse-contour analysis during hemodynamic instability. Crit Care Med 2002; 30: 52-8.PubMed
36.
Zurück zum Zitat Sujatha P, Mehta Y, Dhar A, Sarkar D, Meharwal ZS, Trehan N. Comparison of cardiac output in OPCAB: bolus thermodilution technique versus pulse contour analysis. Ann Card Anaesth 2006; 9: 44-8.PubMed Sujatha P, Mehta Y, Dhar A, Sarkar D, Meharwal ZS, Trehan N. Comparison of cardiac output in OPCAB: bolus thermodilution technique versus pulse contour analysis. Ann Card Anaesth 2006; 9: 44-8.PubMed
37.
Zurück zum Zitat McGee WT, Horswell JL, Calderon J, et al. Validation of a continuous, arterial pressure-based cardiac output measurement: a multicenter, prospective clinical trial. Crit Care 2007; 11: R105.PubMedCentralPubMed McGee WT, Horswell JL, Calderon J, et al. Validation of a continuous, arterial pressure-based cardiac output measurement: a multicenter, prospective clinical trial. Crit Care 2007; 11: R105.PubMedCentralPubMed
38.
Zurück zum Zitat Zimmermann A, Kufner C, Hofbauer S, et al. The accuracy of the Vigileo/FloTrac continuous cardiac output monitor. J Cardiothorac Vasc Anesth 2008; 22: 388-93.PubMed Zimmermann A, Kufner C, Hofbauer S, et al. The accuracy of the Vigileo/FloTrac continuous cardiac output monitor. J Cardiothorac Vasc Anesth 2008; 22: 388-93.PubMed
39.
Zurück zum Zitat Sander M, Spies CD, Grubitzsch H, Foer A, Muller M, von Heymann C. Comparison of uncalibrated arterial waveform analysis in cardiac surgery patients with thermodilution cardiac output measurements. Crit Care 2006; 10: R164.PubMedCentralPubMed Sander M, Spies CD, Grubitzsch H, Foer A, Muller M, von Heymann C. Comparison of uncalibrated arterial waveform analysis in cardiac surgery patients with thermodilution cardiac output measurements. Crit Care 2006; 10: R164.PubMedCentralPubMed
40.
Zurück zum Zitat Opdam HI, Wan L, Bellomo R. A pilot assessment of the FloTrac™ cardiac output monitoring system. Intensive Care Med 2007; 33: 344-9.PubMed Opdam HI, Wan L, Bellomo R. A pilot assessment of the FloTrac™ cardiac output monitoring system. Intensive Care Med 2007; 33: 344-9.PubMed
41.
Zurück zum Zitat Prasser C, Bele S, Keyl C, et al. Evaluation of a new pressure-based cardiac output device requiring no external calibration. BMC Anesthesiol 2007; 7: 9.PubMedCentralPubMed Prasser C, Bele S, Keyl C, et al. Evaluation of a new pressure-based cardiac output device requiring no external calibration. BMC Anesthesiol 2007; 7: 9.PubMedCentralPubMed
42.
Zurück zum Zitat Cecconi M, Dawson D, Casaretti R, Grounds RM, Rhodes A. A prospective study of the accuracy and precision of continuous cardiac output monitoring devices as compared to intermittent thermodilution. Minerva Anestesiol 2010; 76: 1010-7.PubMed Cecconi M, Dawson D, Casaretti R, Grounds RM, Rhodes A. A prospective study of the accuracy and precision of continuous cardiac output monitoring devices as compared to intermittent thermodilution. Minerva Anestesiol 2010; 76: 1010-7.PubMed
43.
Zurück zum Zitat Manecke GR Jr, Auger WR. Cardiac output determination from the arterial pressure wave: clinical testing of a novel algorithm that does not require calibration. J Cardiothorac Vasc Anesth 2007; 21: 3-7.PubMed Manecke GR Jr, Auger WR. Cardiac output determination from the arterial pressure wave: clinical testing of a novel algorithm that does not require calibration. J Cardiothorac Vasc Anesth 2007; 21: 3-7.PubMed
44.
Zurück zum Zitat Breukers RM, Sepehrkhouv S, Spiegelenberg SR, Groeneveld J. Cardiac output measured by a new arterial pressure waveform analysis method without calibration compared with thermodilution after cardiac surgery. J Cardiothorac Vasc Anesth 2007; 21: 632-5.PubMed Breukers RM, Sepehrkhouv S, Spiegelenberg SR, Groeneveld J. Cardiac output measured by a new arterial pressure waveform analysis method without calibration compared with thermodilution after cardiac surgery. J Cardiothorac Vasc Anesth 2007; 21: 632-5.PubMed
45.
Zurück zum Zitat Senn A, Button D, Zollinger A, Hofer CK. Assessment of cardiac output changes using a modified FloTrac/Vigileo™ algorithm in cardiac surgery patients. Crit Care 2009; 13: R32.l. Senn A, Button D, Zollinger A, Hofer CK. Assessment of cardiac output changes using a modified FloTrac/Vigileo™ algorithm in cardiac surgery patients. Crit Care 2009; 13: R32.l.
46.
Zurück zum Zitat Chakravarthy M, Jayaprakash K, Kalligudd P, Prabhakumar D, Jawali V. Comparison of simultaneous estimation of cardiac output by four techniques in patients undergoing off-pump coronary artery bypass surgery—a prospective observational study. Ann Card Anaesth 2007; 10: 121-6.PubMed Chakravarthy M, Jayaprakash K, Kalligudd P, Prabhakumar D, Jawali V. Comparison of simultaneous estimation of cardiac output by four techniques in patients undergoing off-pump coronary artery bypass surgery—a prospective observational study. Ann Card Anaesth 2007; 10: 121-6.PubMed
47.
Zurück zum Zitat Button D, Weibel L, Reutehbuch O, Genoni M, Zollinger A, Hofer CK. Clinical evaluation of the FloTrac/Vigileo™ system and two established continuous cardiac output monitoring devices in patients undergoing cardiac surgery. Br J Anaesth 2007; 99: 329-36.PubMed Button D, Weibel L, Reutehbuch O, Genoni M, Zollinger A, Hofer CK. Clinical evaluation of the FloTrac/Vigileo™ system and two established continuous cardiac output monitoring devices in patients undergoing cardiac surgery. Br J Anaesth 2007; 99: 329-36.PubMed
48.
Zurück zum Zitat Mehta Y, Kumar R, Sawhney R, Bhise M, Singh A, Trehan N. Cardiac output monitoring: comparison of a new arterial pressure waveform analysis to the bolus thermodilution technique in patients undergoing off-pump coronary artery bypass surgery. J Cardiothorac Vasc Anesth 2008; 22: 394-9.PubMed Mehta Y, Kumar R, Sawhney R, Bhise M, Singh A, Trehan N. Cardiac output monitoring: comparison of a new arterial pressure waveform analysis to the bolus thermodilution technique in patients undergoing off-pump coronary artery bypass surgery. J Cardiothorac Vasc Anesth 2008; 22: 394-9.PubMed
49.
Zurück zum Zitat Prasser C, Trabold B, Schwab A, Keyl C, Ziegler S, Wiesenack C. Evaluation of an improved algorithm for arterial pressure-based cardiac output assessment without external calibration. Intensive Care Med 2007; 33: 2223-5.PubMed Prasser C, Trabold B, Schwab A, Keyl C, Ziegler S, Wiesenack C. Evaluation of an improved algorithm for arterial pressure-based cardiac output assessment without external calibration. Intensive Care Med 2007; 33: 2223-5.PubMed
50.
Zurück zum Zitat Della Rocca G, Costa MG, Chiarandini P, et al. Arterial pulse cardiac output agreement with thermodilution in patients in hyperdynamic conditions. J Cardiothorac Vasc Anesth 2008; 22: 681-7. Della Rocca G, Costa MG, Chiarandini P, et al. Arterial pulse cardiac output agreement with thermodilution in patients in hyperdynamic conditions. J Cardiothorac Vasc Anesth 2008; 22: 681-7.
51.
Zurück zum Zitat De Backer D, Marx G, Tan A, et al. Arterial pressure-based cardiac output monitoring: a multicenter validation of the third-generation software in septic patients. Intensive Care Med 2011; 37: 233-40.PubMedCentralPubMed De Backer D, Marx G, Tan A, et al. Arterial pressure-based cardiac output monitoring: a multicenter validation of the third-generation software in septic patients. Intensive Care Med 2011; 37: 233-40.PubMedCentralPubMed
52.
Zurück zum Zitat Zimmermann A, Steinwendner J, Hofbauer S, et al. The accuracy of the Vigileo/FloTrac system has been improved-follow-up after a software update: a blinded comparative study of 30 cardiosurgical patients. J Cardiothorac Vasc Anesth 2009; 23: 929-31.PubMed Zimmermann A, Steinwendner J, Hofbauer S, et al. The accuracy of the Vigileo/FloTrac system has been improved-follow-up after a software update: a blinded comparative study of 30 cardiosurgical patients. J Cardiothorac Vasc Anesth 2009; 23: 929-31.PubMed
53.
Zurück zum Zitat Cannesson M, Attof Y, Rosamel P, Joseph P, Bastien O, Lehot JJ. Comparison of FloTrac™ cardiac output monitoring system in patients undergoing coronary artery bypass grafting with pulmonary artery cardiac output measurements. Eur J Anaesthesiol 2007; 24: 832-9.PubMed Cannesson M, Attof Y, Rosamel P, Joseph P, Bastien O, Lehot JJ. Comparison of FloTrac™ cardiac output monitoring system in patients undergoing coronary artery bypass grafting with pulmonary artery cardiac output measurements. Eur J Anaesthesiol 2007; 24: 832-9.PubMed
54.
Zurück zum Zitat Staier K, Wiesenack C, Günkel L, Keyl C. Cardiac output determination by thermodilution and arterial pulse waveform analysis in patients undergoing aortic valve replacement. Can J Anesth 2008; 55: 22-8.PubMed Staier K, Wiesenack C, Günkel L, Keyl C. Cardiac output determination by thermodilution and arterial pulse waveform analysis in patients undergoing aortic valve replacement. Can J Anesth 2008; 55: 22-8.PubMed
55.
Zurück zum Zitat Hofer CK, Button D, Weibel L, Genoni M, Zollinger A. Uncalibrated radial and femoral arterial pressure waveform analysis for continuous cardiac output measurement: an evaluation in cardiac surgery patients. J Cardiothorac Vasc Anesth 2010; 24: 257-64.PubMed Hofer CK, Button D, Weibel L, Genoni M, Zollinger A. Uncalibrated radial and femoral arterial pressure waveform analysis for continuous cardiac output measurement: an evaluation in cardiac surgery patients. J Cardiothorac Vasc Anesth 2010; 24: 257-64.PubMed
56.
Zurück zum Zitat Vetrugno L, Costa MG, Spagnesi L, et al. Uncalibrated arterial pulse cardiac output measurements in patients with moderately abnormal left ventricular function. J Cardiothorac Vasc Anesth 2011; 25: 53-8.PubMed Vetrugno L, Costa MG, Spagnesi L, et al. Uncalibrated arterial pulse cardiac output measurements in patients with moderately abnormal left ventricular function. J Cardiothorac Vasc Anesth 2011; 25: 53-8.PubMed
57.
Zurück zum Zitat Maxeiner H, Wenzel S, Weigand MA, Mueller M. Cardiac output measured by FloTrac/Vigileo system: does the “plug and play” principle work? Applied Cardiopulmonary Pathophysiology 2010; 14: 131-8. Maxeiner H, Wenzel S, Weigand MA, Mueller M. Cardiac output measured by FloTrac/Vigileo system: does the “plug and play” principle work? Applied Cardiopulmonary Pathophysiology 2010; 14: 131-8.
58.
Zurück zum Zitat Biais M, Nouette-Gaulain K, Cottenceau V, et al. Cardiac output measurement in patients undergoing liver transplantation: pulmonary artery catheter versus uncalibrated arterial pressure waveform analysis. Anesth Analg 2008; 106: 1480-6.PubMed Biais M, Nouette-Gaulain K, Cottenceau V, et al. Cardiac output measurement in patients undergoing liver transplantation: pulmonary artery catheter versus uncalibrated arterial pressure waveform analysis. Anesth Analg 2008; 106: 1480-6.PubMed
59.
Zurück zum Zitat Slagt C, Beute J, Hoeksema M, Malagon I, Mulder JW, Groeneveld JA. Cardiac output derived from arterial pressure waveform analysis without calibration vs. thermodilution in septic shock: evolving accuracy of software versions. Eur J Anaesthesiol 2010; 27: 550-4.PubMed Slagt C, Beute J, Hoeksema M, Malagon I, Mulder JW, Groeneveld JA. Cardiac output derived from arterial pressure waveform analysis without calibration vs. thermodilution in septic shock: evolving accuracy of software versions. Eur J Anaesthesiol 2010; 27: 550-4.PubMed
60.
Zurück zum Zitat Schramm S, Albrecht E, Frascarolo P, Chassot PG, Spahn DR. Validity of an arterial pressure waveform analysis device: does the puncture site play a role in the agreement with intermittent pulmonary artery catheter TD measurements? J Cardiothorac Vasc Anesth 2010; 24: 250-6.PubMed Schramm S, Albrecht E, Frascarolo P, Chassot PG, Spahn DR. Validity of an arterial pressure waveform analysis device: does the puncture site play a role in the agreement with intermittent pulmonary artery catheter TD measurements? J Cardiothorac Vasc Anesth 2010; 24: 250-6.PubMed
61.
Zurück zum Zitat Saraceni E, Rossi S, Persona P, et al. Comparison of two methods for cardiac output measurement in critically ill patients. Br J Anaesth 2011; 106: 690-4.PubMed Saraceni E, Rossi S, Persona P, et al. Comparison of two methods for cardiac output measurement in critically ill patients. Br J Anaesth 2011; 106: 690-4.PubMed
62.
Zurück zum Zitat Biancofiore G, Critchley LA, Lee A, et al. Evaluation of an uncalibrated arterial pulse contour cardiac output monitoring system in cirrhotic patients undergoing liver surgery. Br J Anaesth 2009; 102: 47-54.PubMed Biancofiore G, Critchley LA, Lee A, et al. Evaluation of an uncalibrated arterial pulse contour cardiac output monitoring system in cirrhotic patients undergoing liver surgery. Br J Anaesth 2009; 102: 47-54.PubMed
63.
Zurück zum Zitat Krejci V, Vannucci A, Abbas A, Chapman W, Kangrga IM. Comparison of calibrated and uncalibrated arterial pressure-based cardiac output monitors during orthotopic liver transplantation. Liver Transpl 2010; 16: 773-82.PubMed Krejci V, Vannucci A, Abbas A, Chapman W, Kangrga IM. Comparison of calibrated and uncalibrated arterial pressure-based cardiac output monitors during orthotopic liver transplantation. Liver Transpl 2010; 16: 773-82.PubMed
64.
Zurück zum Zitat Junttila EK, Koskenkari JK, Ohtonen PP, Ala-Kokko TI. Uncalibrated arterial pressure waveform analysis for cardiac output monitoring is biased by low peripheral resistance in patients with intracranial haemorrhage. Br J Anaesth 2011; 107: 581-6.PubMed Junttila EK, Koskenkari JK, Ohtonen PP, Ala-Kokko TI. Uncalibrated arterial pressure waveform analysis for cardiac output monitoring is biased by low peripheral resistance in patients with intracranial haemorrhage. Br J Anaesth 2011; 107: 581-6.PubMed
65.
Zurück zum Zitat Tsai YF, Su BC, Lin CC, Liu FC, Lee WC, Yu HP. Cardiac Output derived from arterial pressure waveform analysis: validation of the third-generation software in patients undergoing orthotopic liver transplantation. Transplant Pro 2012; 44: 433-7. Tsai YF, Su BC, Lin CC, Liu FC, Lee WC, Yu HP. Cardiac Output derived from arterial pressure waveform analysis: validation of the third-generation software in patients undergoing orthotopic liver transplantation. Transplant Pro 2012; 44: 433-7.
66.
Zurück zum Zitat Biancofiore G, Critchley LA, Lee A, et al. Evaluation of a new software version of the FloTrac/Vigileo (Version 3.02) and a comparison with previous data in cirrhotic patients undergoing liver transplant surgery. Anesth Analg 2011; 113: 515-22.PubMed Biancofiore G, Critchley LA, Lee A, et al. Evaluation of a new software version of the FloTrac/Vigileo (Version 3.02) and a comparison with previous data in cirrhotic patients undergoing liver transplant surgery. Anesth Analg 2011; 113: 515-22.PubMed
67.
Zurück zum Zitat Slagt C, de Leeuw MA, Beute J, et al. Cardiac output measured by uncalibrated arterial pressure waveform analysis by recently released software version 3.02 versus thermodilution in septic shock. J Clin Monit Comput 2013; 27: 171-7.PubMed Slagt C, de Leeuw MA, Beute J, et al. Cardiac output measured by uncalibrated arterial pressure waveform analysis by recently released software version 3.02 versus thermodilution in septic shock. J Clin Monit Comput 2013; 27: 171-7.PubMed
68.
Zurück zum Zitat Vasdev S, Chauhan S, Choudhury M, Hote MP, Malik M, Kiran U. Arterial pressure waveform derived cardiac output FloTrac/Vigileo system (third generation software): comparison of two monitoring sites with the thermodilution cardiac output. J Clin Monit Comput 2012; 26: 115-20.PubMed Vasdev S, Chauhan S, Choudhury M, Hote MP, Malik M, Kiran U. Arterial pressure waveform derived cardiac output FloTrac/Vigileo system (third generation software): comparison of two monitoring sites with the thermodilution cardiac output. J Clin Monit Comput 2012; 26: 115-20.PubMed
69.
Zurück zum Zitat Akiyoshi K, Kandabashi T, Kaji J, et al. Accuracy of arterial pressure waveform analysis for cardiac output measurement in comparison with thermodilution methods in patients undergoing living donor liver transplantation. J Anesth 2011; 25: 178-83.PubMed Akiyoshi K, Kandabashi T, Kaji J, et al. Accuracy of arterial pressure waveform analysis for cardiac output measurement in comparison with thermodilution methods in patients undergoing living donor liver transplantation. J Anesth 2011; 25: 178-83.PubMed
70.
Zurück zum Zitat Eleftheriadis S, Galatoudis Z, Didilis V, et al. Variations in arterial blood pressure are associated with parallel changes in FlowTrac/Vigileo®-derived cardiac output measurements: a prospective comparison study. Crit Care 2009; 13: R179.PubMedCentralPubMed Eleftheriadis S, Galatoudis Z, Didilis V, et al. Variations in arterial blood pressure are associated with parallel changes in FlowTrac/Vigileo®-derived cardiac output measurements: a prospective comparison study. Crit Care 2009; 13: R179.PubMedCentralPubMed
71.
Zurück zum Zitat Hadian M, Kim HK, Severyn DA, Pinsky MR. Cross-comparison of cardiac output trending accuracy of LiDCO, PiCCO, FloTrac and pulmonary artery catheters. Crit Care 2010; 14: R212.PubMedCentralPubMed Hadian M, Kim HK, Severyn DA, Pinsky MR. Cross-comparison of cardiac output trending accuracy of LiDCO, PiCCO, FloTrac and pulmonary artery catheters. Crit Care 2010; 14: R212.PubMedCentralPubMed
72.
Zurück zum Zitat Buhre W, Weyland A, Kazmaier S, Hanekop, et al. Comparison of cardiac output assessed by pulse-contour analysis and TD in patients undergoing minimally invasive direct coronary artery bypass grafting. J Cardiothorac Vasc Anesth 1999; 13: 437-40. Buhre W, Weyland A, Kazmaier S, Hanekop, et al. Comparison of cardiac output assessed by pulse-contour analysis and TD in patients undergoing minimally invasive direct coronary artery bypass grafting. J Cardiothorac Vasc Anesth 1999; 13: 437-40.
73.
Zurück zum Zitat Zollner C, Haller M, Weis M, et al. Beat-to-beat measurement of cardiac output by intravascular pulse contour analysis: a prospective criterion standard study in patients after cardiac surgery. J Cardiothorac Vasc Anesthesia 2000; 14: 125-9. Zollner C, Haller M, Weis M, et al. Beat-to-beat measurement of cardiac output by intravascular pulse contour analysis: a prospective criterion standard study in patients after cardiac surgery. J Cardiothorac Vasc Anesthesia 2000; 14: 125-9.
74.
Zurück zum Zitat Hamzaoui O, Monnet X, Richard C, Osman D, Chemla D, Teboul JL. Effects of changes in vascular tone on the agreement between pulse contour and transpulmonary TD cardiac output measurements within an up to 6-hour calibration-free period. Crit Care Med 2008; 36: 434-40.PubMed Hamzaoui O, Monnet X, Richard C, Osman D, Chemla D, Teboul JL. Effects of changes in vascular tone on the agreement between pulse contour and transpulmonary TD cardiac output measurements within an up to 6-hour calibration-free period. Crit Care Med 2008; 36: 434-40.PubMed
75.
Zurück zum Zitat Bein B, Meybohm P, Cavus E, et al. The reliability of pulse contour-derived cardiac output during hemorrhage and after vasopressor administration. Anesth Analg 2007; 105: 107-13.PubMed Bein B, Meybohm P, Cavus E, et al. The reliability of pulse contour-derived cardiac output during hemorrhage and after vasopressor administration. Anesth Analg 2007; 105: 107-13.PubMed
76.
Zurück zum Zitat Staier K, Wilhelm M, Wiesenack Ch, et al. Pulmonary artery vs. transpulmonary thermodilution for the assessment of cardiac output in mitral regurgitation: a prospective observational study. Eur J Anaesthesiol 2012; 29: 431-7.PubMed Staier K, Wilhelm M, Wiesenack Ch, et al. Pulmonary artery vs. transpulmonary thermodilution for the assessment of cardiac output in mitral regurgitation: a prospective observational study. Eur J Anaesthesiol 2012; 29: 431-7.PubMed
77.
Zurück zum Zitat Tzenkov IG, Velasco DA, Perez Pena JM, Olmedilla Arnal L, Garutti Martınez I, Sanz Fernandez J. Cardiac output by femoral arterial thermodilution-calibrated pulse contour analysis during liver transplantation: comparison with pulmonary artery thermodilution. Transplant Proc 2003; 35: 1920-2. Tzenkov IG, Velasco DA, Perez Pena JM, Olmedilla Arnal L, Garutti Martınez I, Sanz Fernandez J. Cardiac output by femoral arterial thermodilution-calibrated pulse contour analysis during liver transplantation: comparison with pulmonary artery thermodilution. Transplant Proc 2003; 35: 1920-2.
78.
Zurück zum Zitat Wouters PF, Quaghebeur B, Sergeant P, Van Hemelrijck J, Vandermeersch E. Cardiac output monitoring using a brachial arterial catheter during off-pump coronary artery bypass grafting. J Cardiothorac Vasc Anesth 2005; 2: 160-4. Wouters PF, Quaghebeur B, Sergeant P, Van Hemelrijck J, Vandermeersch E. Cardiac output monitoring using a brachial arterial catheter during off-pump coronary artery bypass grafting. J Cardiothorac Vasc Anesth 2005; 2: 160-4.
79.
Zurück zum Zitat Rauch H, Müller M, Fleischer F, Bauer H, Martin E, Bottiger BW. Pulse contour analysis versus thermodilution in cardiac output surgery patients. Acta Anaesthesiol Scand 2002; 46: 424-9.PubMed Rauch H, Müller M, Fleischer F, Bauer H, Martin E, Bottiger BW. Pulse contour analysis versus thermodilution in cardiac output surgery patients. Acta Anaesthesiol Scand 2002; 46: 424-9.PubMed
80.
Zurück zum Zitat Mielck F, Buhre W, Hanekop G, Tirilomis T, Hilgers R, Sonntag H. Comparison of continuous cardiac output measurements in patients after cardiac surgery. J Cardiothorac Vasc Anesth 2003; 17: 211-6.PubMed Mielck F, Buhre W, Hanekop G, Tirilomis T, Hilgers R, Sonntag H. Comparison of continuous cardiac output measurements in patients after cardiac surgery. J Cardiothorac Vasc Anesth 2003; 17: 211-6.PubMed
81.
Zurück zum Zitat Halvorsen PS, Sokolov A, Cvancarova M, Hol PK, Lundblad R, Tonnessen TI. Continuous cardiac output during off-pump coronary artery bypass surgery: pulse contour analysis vs pulmonary artery thermodilution. Br J Anaesth 2007; 99: 484-92.PubMed Halvorsen PS, Sokolov A, Cvancarova M, Hol PK, Lundblad R, Tonnessen TI. Continuous cardiac output during off-pump coronary artery bypass surgery: pulse contour analysis vs pulmonary artery thermodilution. Br J Anaesth 2007; 99: 484-92.PubMed
82.
Zurück zum Zitat Sander M, von Heymann C, Foer A, et al. Pulse contour analysis after normothermic cardiopulmonary bypass in cardiac surgery patients. Crit Care 2005; 9: R729.PubMedCentralPubMed Sander M, von Heymann C, Foer A, et al. Pulse contour analysis after normothermic cardiopulmonary bypass in cardiac surgery patients. Crit Care 2005; 9: R729.PubMedCentralPubMed
83.
Zurück zum Zitat Ostergard M, Nielsen J, Rasmussen JP, Berthelsen PG. Cardiac output – pulse contour analysis vs. pulmonary artery thermodilution. Acta Anaesthesiol Scand 2006; 50: 1044-9. Ostergard M, Nielsen J, Rasmussen JP, Berthelsen PG. Cardiac output – pulse contour analysis vs. pulmonary artery thermodilution. Acta Anaesthesiol Scand 2006; 50: 1044-9.
84.
Zurück zum Zitat Della Rocca G, Costa MG, Coccia C, et al. Cardiac output monitoring: aortic transpulmonary thermodilution and pulse contour analysis agree with standard thermodilution methods in patients undergoing lung transplantation. Can J Anesth 2003; 50: 707-11. Della Rocca G, Costa MG, Coccia C, et al. Cardiac output monitoring: aortic transpulmonary thermodilution and pulse contour analysis agree with standard thermodilution methods in patients undergoing lung transplantation. Can J Anesth 2003; 50: 707-11.
85.
Zurück zum Zitat Yamashita K, Nishiyama T, Yokoyama T, Abe H, Manabe M. The effects of vasodilation on cardiac output measured by PiCCO. J Cardiothorac Vasc Anesth 2008; 22: 688-92.PubMed Yamashita K, Nishiyama T, Yokoyama T, Abe H, Manabe M. The effects of vasodilation on cardiac output measured by PiCCO. J Cardiothorac Vasc Anesth 2008; 22: 688-92.PubMed
86.
Zurück zum Zitat Della Rocca G, Costa MG, Pompei L, Coccia C, Pietropaoli P. Continuous and intermittent cardiac output measurement: pulmonary artery catheter versus aortic transpulmonary technique. Br J Anaesth 2002; 88: 350-6. Della Rocca G, Costa MG, Pompei L, Coccia C, Pietropaoli P. Continuous and intermittent cardiac output measurement: pulmonary artery catheter versus aortic transpulmonary technique. Br J Anaesth 2002; 88: 350-6.
87.
Zurück zum Zitat Irlbeck M, Forst H, Briegel J, Haller M, Peter K. Continuous measurement of cardiac output with pulse contour analysis. Anaesthesist 1995; 44: 493-500.PubMed Irlbeck M, Forst H, Briegel J, Haller M, Peter K. Continuous measurement of cardiac output with pulse contour analysis. Anaesthesist 1995; 44: 493-500.PubMed
88.
Zurück zum Zitat Felbinger TW, Reuter DA, Eltzschig HK, Moerstedt K, Goedje O, Goetz AE. Comparison of pulmonary arterial TD and arterial pulse contour analysis: evaluation of a new algorithm. J Clin Anesth 2002; 14: 296-301.PubMed Felbinger TW, Reuter DA, Eltzschig HK, Moerstedt K, Goedje O, Goetz AE. Comparison of pulmonary arterial TD and arterial pulse contour analysis: evaluation of a new algorithm. J Clin Anesth 2002; 14: 296-301.PubMed
89.
Zurück zum Zitat Goedje O, Hoeke K, Lichtwarck-Aschoff M, Faltchauser A, Lamm P, Reichart B. Continuous cardiac output by femoral arterial thermodilution calibrated pulse contour analysis: comparison with pulmonary arterial thermodilution. Crit Care Med 1999; 27: 2407-12.PubMed Goedje O, Hoeke K, Lichtwarck-Aschoff M, Faltchauser A, Lamm P, Reichart B. Continuous cardiac output by femoral arterial thermodilution calibrated pulse contour analysis: comparison with pulmonary arterial thermodilution. Crit Care Med 1999; 27: 2407-12.PubMed
90.
Zurück zum Zitat Linton R, Band D, O’Brien T, Jonas M, Leach R. Lithium dilution cardiac output measurement: a comparison with TD. Crit Care Med 1997; 25: 1796-800.PubMed Linton R, Band D, O’Brien T, Jonas M, Leach R. Lithium dilution cardiac output measurement: a comparison with TD. Crit Care Med 1997; 25: 1796-800.PubMed
91.
Zurück zum Zitat Garcia-Rodriquez C, Pittman J, Cassell C, et al. Lithium dilution cardiac output measurement: a clinical assessment of central venous and peripheral venous indicator injection. Crit Care Med 2002; 30: 2199-204. Garcia-Rodriquez C, Pittman J, Cassell C, et al. Lithium dilution cardiac output measurement: a clinical assessment of central venous and peripheral venous indicator injection. Crit Care Med 2002; 30: 2199-204.
92.
Zurück zum Zitat Hamilton TT, Huber LM, Jessen ME. PulseCO: a less-invasive method to monitor cardiac output from arterial pressure after cardiac surgery. Ann Thorac Surg 2002; 74: 1408-12. Hamilton TT, Huber LM, Jessen ME. PulseCO: a less-invasive method to monitor cardiac output from arterial pressure after cardiac surgery. Ann Thorac Surg 2002; 74: 1408-12.
93.
Zurück zum Zitat Mora B, Ince I, Birkenberg B, et al. Validation of cardiac output measurement with the LiDCOTM pulse contour system in patients with impaired left ventricular function after cardiac surgery. Anaesthesia 2011; 66: 675-81.PubMed Mora B, Ince I, Birkenberg B, et al. Validation of cardiac output measurement with the LiDCOTM pulse contour system in patients with impaired left ventricular function after cardiac surgery. Anaesthesia 2011; 66: 675-81.PubMed
94.
Zurück zum Zitat Costa MG, Della Rocca G, Chiarandini P, et al. Continuous and intermittent cardiac output measurement in hyperdynamic conditions: pulmonary artery catheter vs. lithium dilution technique. Intensive Care Med 2008; 34: 257-63. Costa MG, Della Rocca G, Chiarandini P, et al. Continuous and intermittent cardiac output measurement in hyperdynamic conditions: pulmonary artery catheter vs. lithium dilution technique. Intensive Care Med 2008; 34: 257-63.
95.
Zurück zum Zitat Dyer RA, Piercy JL, Reed AR, et al. Comparison between pulse waveform analysis and thermodilution cardiac output determination in patients with severe pre-eclampsia. Br J Anaesth 2011; 106: 77-81.PubMed Dyer RA, Piercy JL, Reed AR, et al. Comparison between pulse waveform analysis and thermodilution cardiac output determination in patients with severe pre-eclampsia. Br J Anaesth 2011; 106: 77-81.PubMed
96.
Zurück zum Zitat Missant C, Rex S, Wouters PF. Accuracy of cardiac output measurements with pulse contour analysis (PulseCOTM) and Doppler echocardiography during off-pump coronary artery bypass grafting. Eur J Anaesthesiol 2008; 25: 243-8.PubMed Missant C, Rex S, Wouters PF. Accuracy of cardiac output measurements with pulse contour analysis (PulseCOTM) and Doppler echocardiography during off-pump coronary artery bypass grafting. Eur J Anaesthesiol 2008; 25: 243-8.PubMed
97.
Zurück zum Zitat Yamashita K, Nishiyama T, Yokoyama T, Abe H, Manabe M. Cardiac output by PulseCO is not interchangeable with thermodilution in patients undergoing OPCAB. Can J Anesth 2005; 52: 530-4.PubMed Yamashita K, Nishiyama T, Yokoyama T, Abe H, Manabe M. Cardiac output by PulseCO is not interchangeable with thermodilution in patients undergoing OPCAB. Can J Anesth 2005; 52: 530-4.PubMed
98.
Zurück zum Zitat Nissen P, Van Lieshout JJ, Novovic S, Bundgaard-Nielsen M, Secher NH. Techniques of cardiac output measurement during liver transplantation: arterial pulse wave versus thermodilution. Liver Transpl 2009; 15: 287-91.PubMed Nissen P, Van Lieshout JJ, Novovic S, Bundgaard-Nielsen M, Secher NH. Techniques of cardiac output measurement during liver transplantation: arterial pulse wave versus thermodilution. Liver Transpl 2009; 15: 287-91.PubMed
99.
Zurück zum Zitat Jansen JRC, Schreuder JJ, Mulier JP, Smith NT, Settels JJ, Wesseling KH. A comparison of cardiac output derived from the arterial pressure wave against thermodilution in cardiac surgery patients. Br J Anaesth 2001; 87: 212-22.PubMed Jansen JRC, Schreuder JJ, Mulier JP, Smith NT, Settels JJ, Wesseling KH. A comparison of cardiac output derived from the arterial pressure wave against thermodilution in cardiac surgery patients. Br J Anaesth 2001; 87: 212-22.PubMed
100.
Zurück zum Zitat Wesseling KH, Jansen JRC, Settels JJ, Schreuder JJ. Computation of aortic flow from pressure in humans using a nonlinear, three-element model. J Appl Physiol 1993; 74: 2566-73.PubMed Wesseling KH, Jansen JRC, Settels JJ, Schreuder JJ. Computation of aortic flow from pressure in humans using a nonlinear, three-element model. J Appl Physiol 1993; 74: 2566-73.PubMed
101.
Zurück zum Zitat Jellema WT, Wesseling KH, Groeneveld AB, Stoutenbeek CP, Thijs LG, van Lieshout JJ. Continuous cardiac output in septic shock by simulating a model of the aortic input impedance. Anesthesiology 1999; 90: 1317-28.PubMed Jellema WT, Wesseling KH, Groeneveld AB, Stoutenbeek CP, Thijs LG, van Lieshout JJ. Continuous cardiac output in septic shock by simulating a model of the aortic input impedance. Anesthesiology 1999; 90: 1317-28.PubMed
102.
Zurück zum Zitat de Vaal HB, de Wilde RB, van den Berg PC, Schreuder JJ, Jansen JR. Less invasive determination of cardiac output from the arterial pressure by aortic diameter-calibrated pulse contour. Br J Anaesth 2005; 95: 326-31.PubMed de Vaal HB, de Wilde RB, van den Berg PC, Schreuder JJ, Jansen JR. Less invasive determination of cardiac output from the arterial pressure by aortic diameter-calibrated pulse contour. Br J Anaesth 2005; 95: 326-31.PubMed
103.
Zurück zum Zitat Hirschl MM, Kittler H, Woisetschlager C, et al. Simultaneous comparison of thoracic bioimpedance and arterial pulse waveform-derived cardiac output with thermodilution measurement. Crit Care Med 2000; 28: 1798-802.PubMed Hirschl MM, Kittler H, Woisetschlager C, et al. Simultaneous comparison of thoracic bioimpedance and arterial pulse waveform-derived cardiac output with thermodilution measurement. Crit Care Med 2000; 28: 1798-802.PubMed
104.
Zurück zum Zitat Romano SM, Pistolesi M. Assessment of cardiac output from systemic arterial pressure in humans. Crit Care Med 2002; 30: 1834-41.PubMed Romano SM, Pistolesi M. Assessment of cardiac output from systemic arterial pressure in humans. Crit Care Med 2002; 30: 1834-41.PubMed
105.
Zurück zum Zitat Romano SM, Conti AA, Giglioli C, et al. Blood flow assessment by arterial pressure wave without external calibration. Computers in Cardiology 2006; 33: 293-6. Romano SM, Conti AA, Giglioli C, et al. Blood flow assessment by arterial pressure wave without external calibration. Computers in Cardiology 2006; 33: 293-6.
106.
Zurück zum Zitat Giomarelli P, Biagioli B, Scolletta S. Cardiac output monitoring by pressure recording analytical method in cardiac surgery. Eur J Cardiothorac Surg 2004; 26: 515-20.PubMed Giomarelli P, Biagioli B, Scolletta S. Cardiac output monitoring by pressure recording analytical method in cardiac surgery. Eur J Cardiothorac Surg 2004; 26: 515-20.PubMed
107.
Zurück zum Zitat Romano SM, Scolletta S, Olivotto I, et al. Systemic arterial waveform analysis and assessment of blood flow during extracorporeal circulation. Perfusion 2006; 21: 109-16.PubMed Romano SM, Scolletta S, Olivotto I, et al. Systemic arterial waveform analysis and assessment of blood flow during extracorporeal circulation. Perfusion 2006; 21: 109-16.PubMed
108.
Zurück zum Zitat Zangrillo A, Maj G, Monaco F, et al. Cardiac index validation using the pressure recording analytic method in unstable patients. J Cardiothorac Vasc Anesth 2010; 24: 265-9.PubMed Zangrillo A, Maj G, Monaco F, et al. Cardiac index validation using the pressure recording analytic method in unstable patients. J Cardiothorac Vasc Anesth 2010; 24: 265-9.PubMed
109.
Zurück zum Zitat Scolletta S, Franchi F, Taccone FS, Donadello K, Biagioli B, Vincent JL. An uncalibrated pulse contour method to measure cardiac output during aortic counterpulsation. Anesth Analg 2011; 113: 1389-95.PubMed Scolletta S, Franchi F, Taccone FS, Donadello K, Biagioli B, Vincent JL. An uncalibrated pulse contour method to measure cardiac output during aortic counterpulsation. Anesth Analg 2011; 113: 1389-95.PubMed
110.
Zurück zum Zitat Franchi F, Silvestri R, Cubattoli L, et al. Comparison between an uncalibrated pulse contour method and thermodilution technique for cardiac output estimation in septic patients. Br J Anaesth 2011; 107: 202-8.PubMed Franchi F, Silvestri R, Cubattoli L, et al. Comparison between an uncalibrated pulse contour method and thermodilution technique for cardiac output estimation in septic patients. Br J Anaesth 2011; 107: 202-8.PubMed
111.
Zurück zum Zitat Paarmann H, Groesdonk HW, Sedemund-Adib B, Hanke T, Heinze H, Heringlake M. Lack of agreement between pulmonary arterial thermodilution cardiac output and the pressure recording analytical method in postoperative cardiac surgery patients. Br J Anaesth 2011; 106: 475-81.PubMed Paarmann H, Groesdonk HW, Sedemund-Adib B, Hanke T, Heinze H, Heringlake M. Lack of agreement between pulmonary arterial thermodilution cardiac output and the pressure recording analytical method in postoperative cardiac surgery patients. Br J Anaesth 2011; 106: 475-81.PubMed
112.
Zurück zum Zitat Maj G, Monaco F, Landoni G, et al. Cardiac index assessment by the pressure recording analytic method in unstable patients with atrial fibrillation. J Cardiothorac Anesth 2011; 25: 476-80. Maj G, Monaco F, Landoni G, et al. Cardiac index assessment by the pressure recording analytic method in unstable patients with atrial fibrillation. J Cardiothorac Anesth 2011; 25: 476-80.
113.
Zurück zum Zitat Belda FJ, Aguilar G, Teboul JL, et al. Complications related to less-invasive haemodynamic monitoring. Br J Anaesth 2011; 106: 482-6.PubMed Belda FJ, Aguilar G, Teboul JL, et al. Complications related to less-invasive haemodynamic monitoring. Br J Anaesth 2011; 106: 482-6.PubMed
114.
Zurück zum Zitat Reuter DA, Huang C, Edrich T, Shernan SK, Eltzschig HK. Cardiac output monitoring using indicator-dilution techniques: basics, limits, and perspectives. Anesth Analg 2010; 110: 799-811.PubMed Reuter DA, Huang C, Edrich T, Shernan SK, Eltzschig HK. Cardiac output monitoring using indicator-dilution techniques: basics, limits, and perspectives. Anesth Analg 2010; 110: 799-811.PubMed
115.
Zurück zum Zitat Richard C, Monnet X, Teboul JL. Pulmonary artery catheter monitoring in 2011. Current Opinion in Critical Care 2011; 17: 296-302.PubMed Richard C, Monnet X, Teboul JL. Pulmonary artery catheter monitoring in 2011. Current Opinion in Critical Care 2011; 17: 296-302.PubMed
116.
Zurück zum Zitat van Grondelle A, Ditchey RV, Groves BM, Wagner WW, Reeves JT. Thermodilution method overestimates low cardiac output in humans. Am J Physiol 1983; 245: H690-2.PubMed van Grondelle A, Ditchey RV, Groves BM, Wagner WW, Reeves JT. Thermodilution method overestimates low cardiac output in humans. Am J Physiol 1983; 245: H690-2.PubMed
117.
Zurück zum Zitat Nishikawa T, Dohi S. Errors in the measurement of cardiac output by thermodilution. Can J Anaesth 1993; 40: 142-53.PubMed Nishikawa T, Dohi S. Errors in the measurement of cardiac output by thermodilution. Can J Anaesth 1993; 40: 142-53.PubMed
118.
Zurück zum Zitat Jansen JR, Schreuder JJ, Bogaard JM, van Rooyen W, Versprille A. The thermodilution technique for measurement of cardiac output during artificial ventilation. J Appl Physiol 1981; 51: 584-91.PubMed Jansen JR, Schreuder JJ, Bogaard JM, van Rooyen W, Versprille A. The thermodilution technique for measurement of cardiac output during artificial ventilation. J Appl Physiol 1981; 51: 584-91.PubMed
119.
Zurück zum Zitat Bazaral MG, Petre J, Novoa R. Errors in TD cardiac output measurements caused by rapid pulmonary artery temperature decreases after cardiopulmonary bypass. Anesthesiology 1992; 77: 31-7.PubMed Bazaral MG, Petre J, Novoa R. Errors in TD cardiac output measurements caused by rapid pulmonary artery temperature decreases after cardiopulmonary bypass. Anesthesiology 1992; 77: 31-7.PubMed
120.
Zurück zum Zitat Jansen JR, Schreuder JJ, Punt KD, van den Berg PC, Alfieri O. Mean cardiac output by thermodilution with a single controlled injection. Crit Care Med 2001; 29: 1868-73.PubMed Jansen JR, Schreuder JJ, Punt KD, van den Berg PC, Alfieri O. Mean cardiac output by thermodilution with a single controlled injection. Crit Care Med 2001; 29: 1868-73.PubMed
121.
Zurück zum Zitat Nilsson LB, Nilsson JC, Skovgaard LT, Berthelsen PG. Thermodilution cardiac output — are three injections enough? Acta Anaesthesiol Scand 2004; 48: 1322-7.PubMed Nilsson LB, Nilsson JC, Skovgaard LT, Berthelsen PG. Thermodilution cardiac output — are three injections enough? Acta Anaesthesiol Scand 2004; 48: 1322-7.PubMed
122.
Zurück zum Zitat Yang X-X, Critchley LA, Joynt GM. Determination of the precision error of the pulmonary artery thermodilution catheter using an in vitro continuous flow test rig. Anesth Analg 2011; 112: 70-7.PubMed Yang X-X, Critchley LA, Joynt GM. Determination of the precision error of the pulmonary artery thermodilution catheter using an in vitro continuous flow test rig. Anesth Analg 2011; 112: 70-7.PubMed
123.
Zurück zum Zitat Marque S, Gros A, Chimot L, et al. Cardiac output monitoring in septic shock: evaluation of the third-generation Flotrac-Vigileo. J Clin Monit Comput 2013; 27: 273-9.PubMed Marque S, Gros A, Chimot L, et al. Cardiac output monitoring in septic shock: evaluation of the third-generation Flotrac-Vigileo. J Clin Monit Comput 2013; 27: 273-9.PubMed
124.
Zurück zum Zitat Cecconi M, Rhodes A, Poloniecki J, Della Rocca G, Grounds RM. Bench-to-bedside review: the importance of the precision of the reference technique in method comparison studies—with specific reference to the measurement of cardiac output. Crit Care 2009; 13: 201.PubMedCentralPubMed Cecconi M, Rhodes A, Poloniecki J, Della Rocca G, Grounds RM. Bench-to-bedside review: the importance of the precision of the reference technique in method comparison studies—with specific reference to the measurement of cardiac output. Crit Care 2009; 13: 201.PubMedCentralPubMed
125.
Zurück zum Zitat Hamilton C, Stamey J. Using Bland Altman to assess agreement between two medical devices! J Clin Monit Comput 2007; 21: 331-3.PubMed Hamilton C, Stamey J. Using Bland Altman to assess agreement between two medical devices! J Clin Monit Comput 2007; 21: 331-3.PubMed
126.
Zurück zum Zitat Hamm JB, Nguyen BV, Kiss G, et al. Assessment of a cardiac output device using arterial pulse waveform analysis, Vigileo, in cardiac surgery compared to pulmonary arterial TD. Anaesth Intensive Care 2010; 38: 295-301.PubMed Hamm JB, Nguyen BV, Kiss G, et al. Assessment of a cardiac output device using arterial pulse waveform analysis, Vigileo, in cardiac surgery compared to pulmonary arterial TD. Anaesth Intensive Care 2010; 38: 295-301.PubMed
127.
Zurück zum Zitat Jeong YB, Kim TH, Roh YJ, Choi IC, Suh JH. Comparison of uncalibrated arterial pressure waveform analysis with continuous thermodilution cardiac output measurements in patients undergoing elective off-pump coronary artery bypass surgery. J Cardiothorac Vasc Anesth 2010; 5: 767-71. Jeong YB, Kim TH, Roh YJ, Choi IC, Suh JH. Comparison of uncalibrated arterial pressure waveform analysis with continuous thermodilution cardiac output measurements in patients undergoing elective off-pump coronary artery bypass surgery. J Cardiothorac Vasc Anesth 2010; 5: 767-71.
128.
Zurück zum Zitat Haenggi M, Barthelmes D, Ulmer H, Takala J, Jakob SM. Comparison of non-calibrated pulse-contour analysis with continuous thermodilution for cardiac output assessment in patients with induced hypothermia after cardiac arrest. Resuscitation 2011; 82: 423-6.PubMed Haenggi M, Barthelmes D, Ulmer H, Takala J, Jakob SM. Comparison of non-calibrated pulse-contour analysis with continuous thermodilution for cardiac output assessment in patients with induced hypothermia after cardiac arrest. Resuscitation 2011; 82: 423-6.PubMed
129.
Zurück zum Zitat Jo YY, Song JW, Yoo YC, Park JY, Shim JK, Kwak YL. The uncalibrated pulse contour cardiac output during offpump coronary bypass surgery: performance in patients with a low cardiac output status and a reduced left ventricular function. Korean J Anesthesiol 2011; 60: 237-43.PubMedCentralPubMed Jo YY, Song JW, Yoo YC, Park JY, Shim JK, Kwak YL. The uncalibrated pulse contour cardiac output during offpump coronary bypass surgery: performance in patients with a low cardiac output status and a reduced left ventricular function. Korean J Anesthesiol 2011; 60: 237-43.PubMedCentralPubMed
130.
Zurück zum Zitat Su BC, Tsai YF, Chen CY, et al. Cardiac output derived from arterial pressure waveform analysis in patients undergoing liver transplantation: validity of a third-generation device. Transplant Proc 2012; 44: 424-8.PubMed Su BC, Tsai YF, Chen CY, et al. Cardiac output derived from arterial pressure waveform analysis in patients undergoing liver transplantation: validity of a third-generation device. Transplant Proc 2012; 44: 424-8.PubMed
131.
Zurück zum Zitat Monnet X, Anguel N, Jozwiak M, Richard C, Teboul JL. Third-generation FloTrac/Vigileo does not reliably track changes in cardiac output induced by norepinephrine in critically ill patients. Br J Anaesth 2012; 108: 615-22.PubMed Monnet X, Anguel N, Jozwiak M, Richard C, Teboul JL. Third-generation FloTrac/Vigileo does not reliably track changes in cardiac output induced by norepinephrine in critically ill patients. Br J Anaesth 2012; 108: 615-22.PubMed
132.
Zurück zum Zitat Metzelder S, Coburn M, Fries M, Reinges et al. Performance of cardiac output measurement derived from arterial pressure waveform analysis in patients requiring high-dose vasopressor therapy. Br J Anaesth 2011; 106: 776-84. Metzelder S, Coburn M, Fries M, Reinges et al. Performance of cardiac output measurement derived from arterial pressure waveform analysis in patients requiring high-dose vasopressor therapy. Br J Anaesth 2011; 106: 776-84.
133.
Zurück zum Zitat Monnet X, Anguel N, Naudin B, Jabot J, Richard C, Teboul JL. Arterial pressure-based cardiac output in septic patients: different accuracy of pulse contour and uncalibrated pressure waveform devices. Crit Care 2010; 14: R109.PubMedCentralPubMed Monnet X, Anguel N, Naudin B, Jabot J, Richard C, Teboul JL. Arterial pressure-based cardiac output in septic patients: different accuracy of pulse contour and uncalibrated pressure waveform devices. Crit Care 2010; 14: R109.PubMedCentralPubMed
134.
Zurück zum Zitat Chatti R, de Rudniki S, Marque S, et al. Comparison of two versions of the Vigileo-FloTrac system (1.03 and 1.07) for stroke volume estimation: a multicentre, blinded comparison with oesophageal Doppler measurements. Br J Anaesth 2009; 102: 463-9.PubMed Chatti R, de Rudniki S, Marque S, et al. Comparison of two versions of the Vigileo-FloTrac system (1.03 and 1.07) for stroke volume estimation: a multicentre, blinded comparison with oesophageal Doppler measurements. Br J Anaesth 2009; 102: 463-9.PubMed
135.
Zurück zum Zitat Broch O, Renner J, Hocker J, et al. Uncalibrated pulse power analysis fails to reliably measure cardiac output in patients undergoing coronary artery bypass surgery. Crit Care 2011; 15: R76.PubMedCentralPubMed Broch O, Renner J, Hocker J, et al. Uncalibrated pulse power analysis fails to reliably measure cardiac output in patients undergoing coronary artery bypass surgery. Crit Care 2011; 15: R76.PubMedCentralPubMed
136.
Zurück zum Zitat McCoy JV, Hollenberg SM, Dellinger RB, et al. Continuous cardiac index monitoring: a prospective observational study of agreement between a pulmonary artery catheter and a calibrated minimally invasive technique. Resuscitation 2009; 80: 893-7.PubMed McCoy JV, Hollenberg SM, Dellinger RB, et al. Continuous cardiac index monitoring: a prospective observational study of agreement between a pulmonary artery catheter and a calibrated minimally invasive technique. Resuscitation 2009; 80: 893-7.PubMed
137.
Zurück zum Zitat Reisner AT, Xu D, Ryan KL, Convertino VA, Rickards CA, Mukkamala R. Monitoring non-invasive cardiac output and stroke volume during experimental human hypovolaemia and resuscitation. Br J Anaesth 2011; 106: 23-30.PubMedCentralPubMed Reisner AT, Xu D, Ryan KL, Convertino VA, Rickards CA, Mukkamala R. Monitoring non-invasive cardiac output and stroke volume during experimental human hypovolaemia and resuscitation. Br J Anaesth 2011; 106: 23-30.PubMedCentralPubMed
138.
Zurück zum Zitat Singh S, Taylor MA. Con: the FloTrac device should not be used to follow cardiac output in cardiac surgical patients. J Cardiothorac Vasc Anesth 2010; 4: 709-11. Singh S, Taylor MA. Con: the FloTrac device should not be used to follow cardiac output in cardiac surgical patients. J Cardiothorac Vasc Anesth 2010; 4: 709-11.
139.
Zurück zum Zitat Ambrisko TD, Coppens P, Kabes R, Moens Y. Lithium dilution, pulse power analysis, and continuous thermodilution cardiac output measurements compared with bolus thermodilution in anaesthetized ponies. Br J Anaesth 2012; 109: 864-9.PubMed Ambrisko TD, Coppens P, Kabes R, Moens Y. Lithium dilution, pulse power analysis, and continuous thermodilution cardiac output measurements compared with bolus thermodilution in anaesthetized ponies. Br J Anaesth 2012; 109: 864-9.PubMed
140.
Zurück zum Zitat Ambrisko TD, Kabes R, Moens Y. Influence of drugs on the response characteristics of the LiDCO sensor: an in vitro study. Br J Anaesth 2013; 110: 305-10.PubMed Ambrisko TD, Kabes R, Moens Y. Influence of drugs on the response characteristics of the LiDCO sensor: an in vitro study. Br J Anaesth 2013; 110: 305-10.PubMed
141.
Zurück zum Zitat Critchley LA, Lee A, Ho A. A critical review of the ability of continuous cardiac output monitors to measure trends in cardiac output. Anesth Analg 2010; 111: 1180-92.PubMed Critchley LA, Lee A, Ho A. A critical review of the ability of continuous cardiac output monitors to measure trends in cardiac output. Anesth Analg 2010; 111: 1180-92.PubMed
142.
Zurück zum Zitat Suehiro K, Tanaka K, Funao T, et al. Systemic vascular resistance has an impact on the reliability of the Vigileo-FloTrac system in measuring cardiac output and tracking cardiac output changes. Br J Anaesth 2013; 111: 170-7.PubMed Suehiro K, Tanaka K, Funao T, et al. Systemic vascular resistance has an impact on the reliability of the Vigileo-FloTrac system in measuring cardiac output and tracking cardiac output changes. Br J Anaesth 2013; 111: 170-7.PubMed
143.
Zurück zum Zitat Chikhani M, Moppett IK. Minimally invasive cardiac output monitoring: what evidence do we need? Br J Anaesth 2011; 106: 451-3.PubMed Chikhani M, Moppett IK. Minimally invasive cardiac output monitoring: what evidence do we need? Br J Anaesth 2011; 106: 451-3.PubMed
144.
Zurück zum Zitat Geerts B, de Wilde R, Aarts L, Jansen J. Pulse contour analysis to assess hemodynamic response to passive leg raising. J Cardiothorac Vasc Anesth 2011; 25: 48-52.PubMed Geerts B, de Wilde R, Aarts L, Jansen J. Pulse contour analysis to assess hemodynamic response to passive leg raising. J Cardiothorac Vasc Anesth 2011; 25: 48-52.PubMed
145.
Zurück zum Zitat Squara P, Cecconi M, Rhodes M, Singer M, Chiche JD. Tracking changes in cardiac output: methodological considerations for the validation of monitoring devices. Intensive Care Med 2009; 35: 1801-8.PubMed Squara P, Cecconi M, Rhodes M, Singer M, Chiche JD. Tracking changes in cardiac output: methodological considerations for the validation of monitoring devices. Intensive Care Med 2009; 35: 1801-8.PubMed
146.
Zurück zum Zitat Critchley LA, Yang XX, Lee A. Assessment of trending ability of cardiac output monitors by polar plot methodology. J Cardiothorac Vasc Anesth 2011; 25: 536-46.PubMed Critchley LA, Yang XX, Lee A. Assessment of trending ability of cardiac output monitors by polar plot methodology. J Cardiothorac Vasc Anesth 2011; 25: 536-46.PubMed
147.
Zurück zum Zitat Critchley LA. Self-calibrating pulse contour cardiac output: do validation studies really show its clinical reliability? Crit Care 2009; 13: 123.PubMedCentralPubMed Critchley LA. Self-calibrating pulse contour cardiac output: do validation studies really show its clinical reliability? Crit Care 2009; 13: 123.PubMedCentralPubMed
148.
Zurück zum Zitat Maquire S, Rinehart J, Vakharia S, Cannesson M. Respiratory variation in pulse pressure and plethysmographic waveform: intraoperative applicability in a North American academic center. Anesth Analg 2011; 112: 94-6. Maquire S, Rinehart J, Vakharia S, Cannesson M. Respiratory variation in pulse pressure and plethysmographic waveform: intraoperative applicability in a North American academic center. Anesth Analg 2011; 112: 94-6.
149.
Zurück zum Zitat Hamilton MA, Cecconi M, Rhodes A. A systematic review and meta-analysis on the use of preemptive hemodynamic intervention to improve postoperative outcomes in moderate and high-risk surgical patients. Anesth Analg 2011; 112: 1392-402.PubMed Hamilton MA, Cecconi M, Rhodes A. A systematic review and meta-analysis on the use of preemptive hemodynamic intervention to improve postoperative outcomes in moderate and high-risk surgical patients. Anesth Analg 2011; 112: 1392-402.PubMed
150.
Zurück zum Zitat Scheeren TW, Wiesenack C, Gerlach H, Marx G. Goal-directed intraoperative fluid therapy guided by stroke volume and its variation in high-risk surgical patients: a prospective randomized multicentre. J Clin Monit Comput 2013; 27: 225-33.PubMed Scheeren TW, Wiesenack C, Gerlach H, Marx G. Goal-directed intraoperative fluid therapy guided by stroke volume and its variation in high-risk surgical patients: a prospective randomized multicentre. J Clin Monit Comput 2013; 27: 225-33.PubMed
151.
Zurück zum Zitat Pearse R, Dawson D, Fawcett J, Rhodes A, Grounds RM, Bennett ED. Early goal-directed therapy after major surgery reduces complications and duration of hospital stay. A randomised, controlled trial [ISRCTN38797445]. Crit Care 2005; 9: R687-93.PubMedCentralPubMed Pearse R, Dawson D, Fawcett J, Rhodes A, Grounds RM, Bennett ED. Early goal-directed therapy after major surgery reduces complications and duration of hospital stay. A randomised, controlled trial [ISRCTN38797445]. Crit Care 2005; 9: R687-93.PubMedCentralPubMed
152.
Zurück zum Zitat Mayer J, Boldt J, Mengistu AM, Rohm KD, Suttner S. Goal-directed intraoperative therapy based on autocalibrated arterial pressure waveform analysis reduces hospital stay in high-risk surgical patients: a randomized, controlled trial. Crit Care 2010; 14: R18.PubMedCentralPubMed Mayer J, Boldt J, Mengistu AM, Rohm KD, Suttner S. Goal-directed intraoperative therapy based on autocalibrated arterial pressure waveform analysis reduces hospital stay in high-risk surgical patients: a randomized, controlled trial. Crit Care 2010; 14: R18.PubMedCentralPubMed
153.
Zurück zum Zitat Manecke GR. Edwards FloTrac® sensor and Vigileo® monitor: easy, accurate, reliable cardiac output assessment using the arterial pulse wave. Expert Rev Med Devices 2005; 2: 523-7.PubMed Manecke GR. Edwards FloTrac® sensor and Vigileo® monitor: easy, accurate, reliable cardiac output assessment using the arterial pulse wave. Expert Rev Med Devices 2005; 2: 523-7.PubMed
154.
Zurück zum Zitat Boulain T, Achard JM, Teboul JL, Richard C, Perrotin D, Ginies G. Changes in BP induced by passive leg raising predict response to fluid loading in critically ill patients. Chest 2002; 121: 1245-52.PubMed Boulain T, Achard JM, Teboul JL, Richard C, Perrotin D, Ginies G. Changes in BP induced by passive leg raising predict response to fluid loading in critically ill patients. Chest 2002; 121: 1245-52.PubMed
155.
Zurück zum Zitat Cecconi M, Wilson J, Rhodes A. Pulse pressure analysis. In: Vincent JL, editor. I Yearbook of Intensive Care and Emergency Medicine. Berlin/Heidelberg/New York: Springer; 2006. p. 176-83. Cecconi M, Wilson J, Rhodes A. Pulse pressure analysis. In: Vincent JL, editor. I Yearbook of Intensive Care and Emergency Medicine. Berlin/Heidelberg/New York: Springer; 2006. p. 176-83.
156.
Zurück zum Zitat Langewouters GJ, Wesseling KH, Goedhard WJ. The pressure dependent dynamic elasticity of 35 thoracic and 16 abdominal human aortas in vitro described by a five component model. J Biomech 1985; 18: 613-20.PubMed Langewouters GJ, Wesseling KH, Goedhard WJ. The pressure dependent dynamic elasticity of 35 thoracic and 16 abdominal human aortas in vitro described by a five component model. J Biomech 1985; 18: 613-20.PubMed
157.
Zurück zum Zitat Wesseling KH, Smith NT, Nichols WW, Weber H, de Wit B, Beneken JE. Beat-to-beat cardiac output from the arterial pressure pulse contour. In: Feldman SA, Leigh JM, Spierdijk J, editors. Measurement in Anaesthesia. Leiden: Leiden University Press; 1974: 148-64. Wesseling KH, Smith NT, Nichols WW, Weber H, de Wit B, Beneken JE. Beat-to-beat cardiac output from the arterial pressure pulse contour. In: Feldman SA, Leigh JM, Spierdijk J, editors. Measurement in Anaesthesia. Leiden: Leiden University Press; 1974: 148-64.
158.
Zurück zum Zitat Wesseling KH, de Wit B, Smith NT. A simple device for the continuous measurement of cardiac output. Adv Cardiovasc Phys 1983; 5: 16-52. Wesseling KH, de Wit B, Smith NT. A simple device for the continuous measurement of cardiac output. Adv Cardiovasc Phys 1983; 5: 16-52.
159.
Zurück zum Zitat Jansen JR, van den Berg PC. Cardiac Output by thermodilution and arterial pulse contour techniques. In: Pinsky MR, Payen D, editors. Functional Hemodynamic Monitoring-Update in Intensive Care and Emergency Medicine 42. Berlin Heidelberg: Springer; 2005: 135-52. Jansen JR, van den Berg PC. Cardiac Output by thermodilution and arterial pulse contour techniques. In: Pinsky MR, Payen D, editors. Functional Hemodynamic Monitoring-Update in Intensive Care and Emergency Medicine 42. Berlin Heidelberg: Springer; 2005: 135-52.
160.
Zurück zum Zitat Jonas MM, Tanser SJ. Lithium dilution measurement of cardiac output and arterial pulse waveform analysis: an indicator dilution calibrated beat-by-beat system for continuous estimation of cardiac output. Curr Opin Crit Care 2002; 8: 257-61.PubMed Jonas MM, Tanser SJ. Lithium dilution measurement of cardiac output and arterial pulse waveform analysis: an indicator dilution calibrated beat-by-beat system for continuous estimation of cardiac output. Curr Opin Crit Care 2002; 8: 257-61.PubMed
161.
Zurück zum Zitat Jonas MM, Hett D, Morgan J. Real time, continuous monitoring of cardiac output and oxygen delivery. Int J Intensive Care 2002; 9: 1-6. Jonas MM, Hett D, Morgan J. Real time, continuous monitoring of cardiac output and oxygen delivery. Int J Intensive Care 2002; 9: 1-6.
162.
Zurück zum Zitat Langewouters GJ, Wesseling KH, Goedhard WJ. The static elastic properties of 45 human thoracic and 20 abdominal aortas in vitro and the parameters of a new model. Journal Biomech 1984; 17: 425-35. Langewouters GJ, Wesseling KH, Goedhard WJ. The static elastic properties of 45 human thoracic and 20 abdominal aortas in vitro and the parameters of a new model. Journal Biomech 1984; 17: 425-35.
163.
Zurück zum Zitat Bogert LWJ, Lieshout JJ. Non-invasive pulsatile arterial pressure and stroke volume changes from the human finger. Exp Physiol 2005; 90: 437-46.PubMed Bogert LWJ, Lieshout JJ. Non-invasive pulsatile arterial pressure and stroke volume changes from the human finger. Exp Physiol 2005; 90: 437-46.PubMed
164.
Zurück zum Zitat Messiah A. Change of representation and perturbation treatment of a part of the Hamiltonian. In: Messiah A (Ed). Quantum Mechanics. Amsterdam: North Holland Physics Publishing; 1985; 2: 722-60. Messiah A. Change of representation and perturbation treatment of a part of the Hamiltonian. In: Messiah A (Ed). Quantum Mechanics. Amsterdam: North Holland Physics Publishing; 1985; 2: 722-60.
165.
Zurück zum Zitat Scolletta S, Romano SM, Biagioli B, Capannini G, Giomarelli P. Pressure recording analytical method (PRAM) for measurement of cardiac output during various haemodynamic states. Br J Anaesth 2005; 95: 159-65.PubMed Scolletta S, Romano SM, Biagioli B, Capannini G, Giomarelli P. Pressure recording analytical method (PRAM) for measurement of cardiac output during various haemodynamic states. Br J Anaesth 2005; 95: 159-65.PubMed
Metadaten
Titel
Semi-invasive measurement of cardiac output based on pulse contour: a review and analysis
verfasst von
Thomas Schlöglhofer, BSc
Hermann Gilly, PhD
Heinrich Schima, PhD
Publikationsdatum
01.05.2014
Verlag
Springer US
Erschienen in
Canadian Journal of Anesthesia/Journal canadien d'anesthésie / Ausgabe 5/2014
Print ISSN: 0832-610X
Elektronische ISSN: 1496-8975
DOI
https://doi.org/10.1007/s12630-014-0135-8

Weitere Artikel der Ausgabe 5/2014

Canadian Journal of Anesthesia/Journal canadien d'anesthésie 5/2014 Zur Ausgabe

Update AINS

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.