Erschienen in: Insights into Imaging 5/2013

Open Access 01.10.2013 | Original Article

Comparison of image quality between filtered back-projection and the adaptive statistical and novel model-based iterative reconstruction techniques in abdominal CT for renal calculi

verfasst von: Varut Vardhanabhuti, Sumaira Ilyas, Catherine Gutteridge, Simon J. Freeman, Carl A. Roobottom

Erschienen in: Insights into Imaging | Ausgabe 5/2013

Abstract

Objectives

To compare image quality on computed tomographic (CT) images acquired with filtered back-projection (FBP), adaptive statistical iterative reconstruction (ASIR) and model-based iterative reconstruction (MBIR) techniques in CT kidney/ureter/bladder (KUB) examination.

Methods

Eighteen patients underwent standard protocol CT KUB at our institution. The same raw data were reconstructed using FBP, ASIR and MBIR. Objective [mean image noise, contrast-to-noise ratio (CNR) for kidney and mean attenuation values of subcutaneous fat] and subjective image parameters (image noise, image contrast, overall visibility of kidneys/ureters/bladder, visibility of small structures, and overall diagnostic confidence) were assessed using a scoring system from 1 (best) to 5 (worst).

Results

Objective image measurements revealed significantly less image noise and higher CNR and the same fat attenuation values for the MBIR technique (P < 0.05). MBIR scored best in all the subjective image parameters (P < 0.001) with averages ranging between 2.05–2.73 for MBIR, 2.95–3.10 for ASIR and 3.08–3.31 for FBP. No significant difference was observed between FBP and ASIR (P > 0.05), while there was a significant difference between ASIR vs. MBIR (P < 0.05). The mean effective dose was 3 mSv.

Conclusion

MBIR shows superior reduction in noise and improved image quality (both objective and subjective analysis) compared with ASIR and FBP CT KUB examinations.

Main Messages

There are many reconstruction options in CT.
Novel model-based iterative reconstruction (MBIR) showed the least noise and optimal image quality.
For CT of the kidneys/ureters/bladder, MBIR should be utilised, if available.
Further studies to reduce the dose while maintaining image quality should be pursued.

Introduction

The CT KUB is now regarded as the imaging investigation of choice for most patients with suspected renal stone disease because of its unrivalled stone detection capacity, speed and non-dependence on intravenous contrast medium administration [1, 2]. There are well-established practices of using low-dose CT in the detection of renal stone disease resulting in inherently ‘noiser’ images than in conventional CT examinations but without compromising diagnostic confidence for this clinical entity [3, 4]. In this article, we focus on the emergence of new iterative reconstruction techniques that have developed over the past few years. Traditional filtered back-projection (FBP) has given way to novel iterative reconstruction algorithms, and its use has increasingly been employed. For example, General Electric (GE, Milwaukee, WI, USA) has introduced adaptive statistical iterative reconstruction (ASIR), which uses a blend of filtered back-projection images with iteratively reconstructed images. Centres that have this technology use a varying degree of ASIR with most adopting a value between a 20 and 40 % ASIR blend. More recently, GE has brought out model-based iterative reconstruction (MBIR) and obtained FDA approval for this technique in September 2011. The specifics of each reconstruction algorithm will be discussed in detail later. Phantom studies have shown promising results using MBIR in noise reduction [5]. Some clinical studies have also shown promising results using ASIR [68], but to date few assessments have been made of the use of MBIR in clinical practice [911]. There is no recent literature on the use of MBIR in CT KUB. The purpose of this study was to perform a comparison of the image quality of CT KUB examinations acquired with three different reconstruction algorithms—FBP, ASIR and MBIR.

Materials and methods

The study was approved by our institutional review board. Due to the nature of the study, no consent or ethical approval was required. Eighteen consecutive patients (in a 4-week period) who underwent standard protocol CT KUB as an outpatient episode for investigation of suspected or known urolithiasis at our institution were chosen. The mean age was 42 years (range, 20–67; SD, 14.8) with a mean weight of 74.8 kg (range, 62–128; SD 17.5). Retrospectively, the same raw data were reconstructed using FBP, ASIR and MBIR. This resulted in 54 image data sets in total. The DLP for each examination was recorded and the effective dose calculated using a conversion factor of 0.015 mSv/(mGy × cm) [12].

CT technique

CT was performed with a commercial CT system (Discovery CT750 HD; GE Healthcare). We use automatic tube current modulation using noise index (NI) parameters for prescribing an acceptable image noise in clinical practice so that the amount of noise in an image would remain constant despite differing patient sizes [13]. At our institution we use an NI of 50 for CT KUB with acquisition performed at 1.25-mm slice thickness. Other scanning parameters are as follows: tube voltage, 120 kVp; pitch, 0.984:1; table speed, 39.37 mm per gantry rotation; helical acquisition mode; detector configuration, 64*0.625 mm; gantry rotation time, 0.5 s; reconstructed section thickness, 0.625, 2.5 mm and 5 mm; reconstructed section interval, 0.625 mm, 2.5 mm and 5 mm; standard reconstruction kernel.

Reconstruction algorithms

The differences among the three reconstruction techniques are related to the assumptions that each method makes in producing the final image from the raw data. FBP assumes that the focal spot on the x-ray tube is a point source, with a perfect pencil beam shape, and a point at the patient’s body and at the detector is assumed to be a pixel (two- rather than three-dimensional). ASIR uses FBP as the building block for image reconstruction assuming the beam to be a perfect point source. It aims to improve image quality by focusing on noise reduction. In our institution we use a blending of 30 % of ASIR with FBP. Whilst ASIR still relies on FBP data sets, model-based iterative reconstruction builds a forward projection using dedicated system optics, taking into account every x-ray projection (in its true three-dimensional domain), and produces an image based on the raw data. Multiple iterations are performed to correct the residual error between the forward projection and acquired image. These algorithms also incorporate statistical noise information in the reconstruction process. The combination of system optic modelling and statistical modelling helps in noise reduction and results in truer image characteristics compared to FBP and ASIR. In addition, MBIR also accounts for noise from photon flux as well as system noise (e.g. electronic noise) from the CT system itself.

Quantitative image analysis

Objective image analysis was performed using circular regions of interest (ROIs) drawn over several areas (size of between 1 and 3 cm2). For each patient, ROIs were drawn over five contiguous images for each anatomical area. Image noise was taken from standard deviation values derived over three areas of subcutaneous fat (anterior abdominal wall, left buttock and right buttock). Mean attenuation values were taken as an average of the mean Hounsfield numbers over these same areas of subcutaneous fat. ROIs were also drawn over the upper poles of both kidneys. These values were used to calculate the contrast-to-noise ratio (CNR) for kidneys across three different reconstruction algorithms using the following equation:
CNR = ROI o ROI sf / SD n
ROIo is the mean attenuation for the organ of interest, ROIsf is the mean attenuation for the subcutaneous fat, and SDn is the mean image noise.

Qualitative image analysis

Subjective analysis was performed by anonymising 54 data sets and displaying this in randomised order to two radiologists (with 7 and 15 years of consultant urogenital radiology experience). Each radiologist had access to axial and multi-planar reformats. Parameters assessed were: A: image noise (1: minimal, 2: less than average, 3: average, 4: above average, 5: unacceptable), B: image contrast (1: excellent, 2: above average, 3: acceptable, 4: suboptimal, 5: very poor), C: overall visibility of kidneys/ureters/bladder, D: visibility of small structures, e.g. small lymph nodes and adrenal glands (1: excellent visualisation, 2: above average visibility, 3: acceptable visibility, 4: suboptimal visibility, 5: unacceptable) and E: overall diagnostic confidence (1: completely confident, 2: probably confident, 3: confident only for limited clinical entity, 4: poor confident, 5: non-diagnostic examination). The scoring criteria were based on the European Guidelines for Quality Criteria for CT [14].

Statistical analysis

Quantitative data such as objective image noise and mean attenuation were analysed by comparing standard deviations, the 95 % confidence interval and statistical differences analysed using repeated analysis of variance (ANOVA) with post-test correction using the Bonferroni method. The interobserver variation between the two radiologists for each of the assessed subjective image quality parameters was estimated by using weighted kappa statistics. The Friedman test (with Dunn post-test) was used to test for equality of median scores among all subjective parameters.

Results

Quantitative

Table 1 shows the average quantitative values for the image noise, mean attenuation values and contrast-to-noise ratio of three different reconstruction algorithms. Table 2 shows repeated ANOVA with Bonferroni multiple comparison tests of mean differences in image noise. This supports the hypothesis that objective image noise shows a significant reduction with the new MBIR technique (P < 0.0001). No significant differences were seen when comparing mean attenuation values for each method (P > 0.05 with non-overlapping 95 % CI); see Fig. 1. MBIR shows superior CNR for the kidneys in comparison with subcutaneous fat (P < 0.05), but this effect is likely to be predominantly due to the marked reduction in image noise seen with MBIR (see Fig. 2).
Table 1
Average quantitative values for image noise, mean attenuation values and contrast-to-noise ratio of three different reconstruction algorithms
 
FBP
ASIR
MBIR
Mean noise
 Mean
47.65
39.36
15.74
 SD
15.84
13.40
3.74
Mean attenuation values
 Mean
−106.74
−106.75
−106.58
 SD
9.68
9.70
9.66
Mean contrast-to-noise ratio
 
LT kidney
RT kidney
LT kidney
RT kidney
LT kidney
RT kidney
 Mean
3.18
3.19
3.86
3.88
9.13
9.09
 SD
0.97
0.97
1.20
1.20
2.02
1.82
Table 2
Bonferroni’s multiple comparison test of mean differences in image noise for three different reconstruction algorithms
 
Mean Diff.
95 % CI
 
FBP vs. ASIR
8.285
1.252 to 15.32
p < 0.0001
FBP vs. MBIR
31.91
24.88 to 38.94
p < 0.0001
ASIR vs. MBIR
23.63
16.59 to 30.66
p < 0.0001

Qualitative

Mean qualitative scores between two raters are summarised in Table 3. MBIR scored best in all the subjective assessments of image parameters (image noise, image contrast, visibility of kidneys/ureters/bladder, visibility of small structures and overall diagnostic confidence) followed by ASIR then FBP (P < 0.001). This is graphically illustrated in Fig. 3 (image noise), Fig. 4 (visibility of kidneys/ureters/bladder) and Fig. 5 (overall diagnostic confidence). Table 4 shows a summary of the Friedman test with Dunn’s post-test for multiple comparisons for qualitative image noise. This illustrates that the scoring between FBP and ASIR shows no difference (P > 0.05), but significance exists between ASIR vs. MBIR and FBP vs. MBIR (P < 0.05). The interobserver variation (weighted kappa and standard errors) between the two radiologists were fair to moderate as follows: image noise [0.549 (SE:0.091)], image contrast [0.261 (SE: 0.096)], visibility of kidneys/ureters/bladder [0.231 (0.098), visibility of small structures [0.394 (0.105) and overall diagnostic confidence [0.434 (0.094).
Table 3
Mean qualitative scores between two raters
 
FBP
ASIR
MBIR
Image noise
 Mean
3.23
2.95
2.05
 SD
0.53
0.46
0.22
 Lower 95 % CI
2.98
2.74
1.95
 Upper 95 % CI
3.47
3.16
2.16
Image contrast
 Mean
3.31
2.81
2.25
 SD
0.42
0.55
0.31
 Lower 95 % CI
3.09
2.53
2.10
 Upper 95 % CI
3.52
3.08
2.40
Visibility of kidneys/ureters/bladder
 Mean
3.08
3.05
2.65
 SD
0.24
0.28
0.33
 Lower 95 % CI
2.96
2.92
2.50
 Upper 95 % CI
3.19
3.18
2.80
Visibility of small structures
 Mean
3.13
3.10
2.73
 SD
0.43
0.42
0.34
 Lower 95 % CI
2.93
2.91
2.56
 Upper 95 % CI
3.32
3.30
2.89
Overall diagnostic confidence
 Mean
3.08
3.08
2.58
 SD
0.44
0.44
0.29
 Lower 95 % CI
2.87
2.87
2.44
 Upper 95 % CI
3.28
3.28
2.71
Table 4
Friedman test with Dunn’s multiple comparison test
Friedman test
P value
<0.0001
Friedman statistic
36.21
Dunn’s multiple comparison test
Difference in rank sum
Significant? P < 0.05?
ASIR vs. FBP
−9.5
No
ASIR vs. MBIR
24.5
Yes
FBP vs. MBIR
34
Yes

Radiation dose

Images were acquired at a mean effective dose of 3 mSv (SD 2 mSv; mean DLP 202 mGy, SD 145.7); see Table 5.
Table 5
Radiation dose
 
DLP
Effective dose (mSv)
Average
195.03
2.93
SD
138.91
2.08
Min.
93.57
1.40
Max.
692.49
10.39
Some side-by-side examples are shown in Figs. 6, 7, 8 and 9 for different-sized stones.

Discussion

Computed tomography has largely replaced excretory urography as the investigation of choice for assessment of renal colic and ureteric stones, with high sensitivity and specificity of 94 %–100 % and 97 % respectively [1517]. In recent years, iterative reconstruction techniques have been introduced, predominantly by reducing noise to improve image quality. ASIR, one of the most widely studied iterative reconstruction techniques, is associated not only with improved image quality but also with significant dose reduction [58]. Better image quality with ASIR as compared to FBP was also observed in our series of patients, which is in keeping with previous study [18]. The novel iterative technique MBIR was associated with greater noise reduction and improved image quality compared to both ASIR and FBP. This was observed on both objective and subjective analysis in our study. Our results are in keeping with recent findings of some dose studies in other examinations such as spine [19, 20], posterior fossa angiography [21], abdomen [22] and ex vivo heart [23]. All these studies show that there is improved image quality, in particular reduced objective and subjective image noise compared with traditional FBP and ASIR.
With improved image quality, the next logical step would be to reduce the radiation dose whilst maintaining diagnostic image quality. There is emerging evidence that MBIR has significant dose reduction potential in chest [2426], abdominal [11, 27] and paediatric cardiac CT [28]. Of note, Pickhardt et al. [27] have shown that MBIR has significant potential when aggressive dose reduction strategies are utilised. This was a preliminary trial with pooled data from low-dose contrast-enhanced abdominal CT, CT colonography and unenhanced CT KUB. The small sample size and diversity of examinations included in this study did not allow firm conclusions to be drawn in respect to lesion detectability or diagnostic confidence in aggressive dose reduction protocols. Due to the retrospective nature of our study, we did not study the low-dose utility of MBIR. However, extrapolating from results of previous studies, application of MBIR can potentially be used to perform diagnostic examinations with further reduction in dose. In the context of CT KUB, this can be of particular value as many patients with urolithiasis are young and will require multiple CT examinations during their lives. However, further research is required to ascertain the exact levels of dose reduction achievable with MBIR. Further studies using lower dose scans with assessment of diagnostic accuracies should be performed to gain maximal benefit of the noise reduction achieved with MBIR.
The major limitation of MBIR is the time required to obtain multiple iterations. In our study 35–40 min was required for image reconstruction with a model-based iterative reconstruction algorithm. This limits the use of this technique in emergency situations, which was not the case with ASIR [28]. It is however still possible to perform an initial reconstruction with FBP or ASIR (to detect large or easily detectable lesions or life-threatening conditions, for example) followed by a second reconstruction with MBIR (where a detailed and thorough examination can be performed to formalise a final report).
In summary, MBIR is superior to both FBP and ASIR in terms of image noise and quality in both objective and subjective analysis at the same radiation dose, and if available should be utilised. Using this new iterative reconstruction algorithm, it may be possible to acquire images with diagnostic quality similar to FBP or ASIR at a reduced dose, but further studies are required to substantiate this claim.

Conflicts of interest

The authors declare no conflicts of interest. No funding was received for this work. Written patient consent was waived by the Institutional Review Board.
Open Access This article is distributed under the terms of the Creative Commons Attribution License which permits any use, distribution, and reproduction in any medium, provided the original author(s) and the source are credited.
Literatur
1.
Zurück zum Zitat Gottlieb RH, La TC, Erturk EN et al (2002) CT in detecting urinary tract calculi: influence on patient imaging and clinical outcomes. Radiology 225:441–449PubMedCrossRef Gottlieb RH, La TC, Erturk EN et al (2002) CT in detecting urinary tract calculi: influence on patient imaging and clinical outcomes. Radiology 225:441–449PubMedCrossRef
2.
Zurück zum Zitat Smith RC, Verga M, McCarthy S, Rosenfield AT (1996) Diagnosis of acute flank pain: value of unenhanced helical CT. AJR Am J Roentgenol 166:97–101PubMedCrossRef Smith RC, Verga M, McCarthy S, Rosenfield AT (1996) Diagnosis of acute flank pain: value of unenhanced helical CT. AJR Am J Roentgenol 166:97–101PubMedCrossRef
3.
Zurück zum Zitat Ciaschini MW, Remer EM, Baker ME et al (2009) Urinary calculi: radiation dose reduction of 50% and 75% at CT–effect on sensitivity 1. Radiology 251:105–111PubMedCrossRef Ciaschini MW, Remer EM, Baker ME et al (2009) Urinary calculi: radiation dose reduction of 50% and 75% at CT–effect on sensitivity 1. Radiology 251:105–111PubMedCrossRef
4.
Zurück zum Zitat Jin DH, Lamberton GR, Broome DR et al (2010) Effect of reduced radiation CT protocols on the detection of renal calculi. Radiology 255:100–107PubMedCrossRef Jin DH, Lamberton GR, Broome DR et al (2010) Effect of reduced radiation CT protocols on the detection of renal calculi. Radiology 255:100–107PubMedCrossRef
5.
Zurück zum Zitat Vardhanabhuti V, Olubaniyi B, Loader R et al (2012) Image quality assessment in torso phantom comparing effects of varying automatic current modulation with filtered back projection, adaptive statistical, and model-based iterative reconstruction techniques in CT. J Med Imaging Radiat Sci 43:228–238CrossRef Vardhanabhuti V, Olubaniyi B, Loader R et al (2012) Image quality assessment in torso phantom comparing effects of varying automatic current modulation with filtered back projection, adaptive statistical, and model-based iterative reconstruction techniques in CT. J Med Imaging Radiat Sci 43:228–238CrossRef
6.
Zurück zum Zitat Cornfeld D, Israel G, Detroy E et al (2011) Impact of adaptive statistical iterative reconstruction (ASIR) on radiation dose and image quality in aortic dissection studies: a qualitative and quantitative analysis. Am J Roentgenol 196:W336–W340CrossRef Cornfeld D, Israel G, Detroy E et al (2011) Impact of adaptive statistical iterative reconstruction (ASIR) on radiation dose and image quality in aortic dissection studies: a qualitative and quantitative analysis. Am J Roentgenol 196:W336–W340CrossRef
7.
Zurück zum Zitat Singh S, Kalra MK, Gilman MD et al (2011) Adaptive statistical iterative reconstruction technique for radiation dose reduction in chest CT: a pilot study. Radiology 259:565–573PubMedCrossRef Singh S, Kalra MK, Gilman MD et al (2011) Adaptive statistical iterative reconstruction technique for radiation dose reduction in chest CT: a pilot study. Radiology 259:565–573PubMedCrossRef
8.
Zurück zum Zitat Singh S, Kalra MK, Hsieh J et al (2010) Abdominal CT: comparison of adaptive statistical iterative and filtered back projection reconstruction techniques. Radiology 257:373–383PubMedCrossRef Singh S, Kalra MK, Hsieh J et al (2010) Abdominal CT: comparison of adaptive statistical iterative and filtered back projection reconstruction techniques. Radiology 257:373–383PubMedCrossRef
9.
Zurück zum Zitat Yamada Y, Jinzaki M, Tanami Y et al (2012) Model-based iterative reconstruction technique for ultralow-dose computed tomography of the lung: a pilot study. Invest Radiol 47:482–489PubMedCrossRef Yamada Y, Jinzaki M, Tanami Y et al (2012) Model-based iterative reconstruction technique for ultralow-dose computed tomography of the lung: a pilot study. Invest Radiol 47:482–489PubMedCrossRef
10.
Zurück zum Zitat Katsura M, Matsuda I, Akahane M et al (2012) Model-based iterative reconstruction technique for radiation dose reduction in chest CT: comparison with the adaptive statistical iterative reconstruction technique. Eur Radiol 22:1613–1623PubMedCrossRef Katsura M, Matsuda I, Akahane M et al (2012) Model-based iterative reconstruction technique for radiation dose reduction in chest CT: comparison with the adaptive statistical iterative reconstruction technique. Eur Radiol 22:1613–1623PubMedCrossRef
11.
Zurück zum Zitat Husarik DB, Marin D, Samei E et al (2012) Radiation dose reduction in abdominal computed tomography during the late hepatic arterial phase using a model-based iterative reconstruction algorithm: how low can we go? Invest Radiol 47:468–474PubMedCrossRef Husarik DB, Marin D, Samei E et al (2012) Radiation dose reduction in abdominal computed tomography during the late hepatic arterial phase using a model-based iterative reconstruction algorithm: how low can we go? Invest Radiol 47:468–474PubMedCrossRef
12.
Zurück zum Zitat Deak PD, Smal Y, Kalender WA (2010) Multisection CT protocols: sex- and age-specific conversion factors used to determine effective dose from dose length product. Radiology 257:158–166PubMedCrossRef Deak PD, Smal Y, Kalender WA (2010) Multisection CT protocols: sex- and age-specific conversion factors used to determine effective dose from dose length product. Radiology 257:158–166PubMedCrossRef
13.
Zurück zum Zitat Li J, Udayasankar UK, Tang X, Toth TL, Small WC, Kalra MK (2011) Patient size compensated automatic tube current modulation in multi-detector row CT of the abdomen and pelvis. Acad Radiol 18:205–211PubMedCrossRef Li J, Udayasankar UK, Tang X, Toth TL, Small WC, Kalra MK (2011) Patient size compensated automatic tube current modulation in multi-detector row CT of the abdomen and pelvis. Acad Radiol 18:205–211PubMedCrossRef
14.
Zurück zum Zitat Leonardi M, Bongartz G, Geleijns DJ, Golding SJ, Jessen K, Jurik AG et al (1999) European guidelines on quality criteria for computed tomography. Brussels, Belgium: European Commission, 2000:EUR 16262 Leonardi M, Bongartz G, Geleijns DJ, Golding SJ, Jessen K, Jurik AG et al (1999) European guidelines on quality criteria for computed tomography. Brussels, Belgium: European Commission, 2000:EUR 16262
15.
Zurück zum Zitat Yilmaz S, Sindel T, Arslan G et al (1998) Renal colic: comparison of spiral CT, US and IVU in the detection of ureteral calculi. Eur Radiol 8:212–217PubMedCrossRef Yilmaz S, Sindel T, Arslan G et al (1998) Renal colic: comparison of spiral CT, US and IVU in the detection of ureteral calculi. Eur Radiol 8:212–217PubMedCrossRef
16.
Zurück zum Zitat Katz DS, Hines J, Rausch DR et al (1999) Unenhanced helical CT for suspected renal colic. AJR Am J Roentgenol 173:425–430PubMedCrossRef Katz DS, Hines J, Rausch DR et al (1999) Unenhanced helical CT for suspected renal colic. AJR Am J Roentgenol 173:425–430PubMedCrossRef
17.
Zurück zum Zitat Sudah M, Vanninen RL, Partanen K et al (2002) Patients with acute flank pain: comparison of MR urography with unenhanced helical CT. Radiology 223:98–105PubMedCrossRef Sudah M, Vanninen RL, Partanen K et al (2002) Patients with acute flank pain: comparison of MR urography with unenhanced helical CT. Radiology 223:98–105PubMedCrossRef
18.
Zurück zum Zitat Kulkarni NM, Uppot RN, Eisner BH, Sahani DV (2012) Radiation dose reduction at multidetector CT with adaptive statistical iterative reconstruction for evaluation of urolithiasis: how low can we go? Radiology 265(1):158–166PubMedCrossRef Kulkarni NM, Uppot RN, Eisner BH, Sahani DV (2012) Radiation dose reduction at multidetector CT with adaptive statistical iterative reconstruction for evaluation of urolithiasis: how low can we go? Radiology 265(1):158–166PubMedCrossRef
19.
Zurück zum Zitat Machida H, Tanaka I, Fukui R, Kita K, Shen Y, Ueno E, Suzuki S (2013) Improved delineation of the anterior spinal artery with model-based iterative reconstruction in CT angiography: a clinical pilot study. AJR Am J Roentgenol 200(2):442–446PubMedCrossRef Machida H, Tanaka I, Fukui R, Kita K, Shen Y, Ueno E, Suzuki S (2013) Improved delineation of the anterior spinal artery with model-based iterative reconstruction in CT angiography: a clinical pilot study. AJR Am J Roentgenol 200(2):442–446PubMedCrossRef
20.
Zurück zum Zitat Katsura M, Sato J, Akahane M, Matsuda I, Ishida M, Yasaka K, Kunimatsu A, Ohtomo K (2013) Comparison of pure and hybrid iterative reconstruction techniques with conventional filtered back projection: image quality assessment in the cervicothoracic region. Eur J Radiol 82(2):356–360PubMedCrossRef Katsura M, Sato J, Akahane M, Matsuda I, Ishida M, Yasaka K, Kunimatsu A, Ohtomo K (2013) Comparison of pure and hybrid iterative reconstruction techniques with conventional filtered back projection: image quality assessment in the cervicothoracic region. Eur J Radiol 82(2):356–360PubMedCrossRef
21.
Zurück zum Zitat Machida H, Takeuchi H, Tanaka I, Fukui R, Shen Y, Ueno E, Suzuki S, Lin XZ (2013) Improved delineation of arteries in the posterior fossa of the brain by model-based iterative reconstruction in volume-rendered 3D CT angiography. AJNR Am J Neuroradiol 34(5):971–5 Machida H, Takeuchi H, Tanaka I, Fukui R, Shen Y, Ueno E, Suzuki S, Lin XZ (2013) Improved delineation of arteries in the posterior fossa of the brain by model-based iterative reconstruction in volume-rendered 3D CT angiography. AJNR Am J Neuroradiol 34(5):971–5
22.
Zurück zum Zitat Deák Z, Grimm JM, Treitl M, Geyer LL, Linsenmaier U, Körner M, Reiser MF, Wirth S (2013) Filtered back projection, adaptive statistical iterative reconstruction, and a model-based iterative reconstruction in abdominal CT: an experimental clinical study. Radiology 266(1):197–206PubMedCrossRef Deák Z, Grimm JM, Treitl M, Geyer LL, Linsenmaier U, Körner M, Reiser MF, Wirth S (2013) Filtered back projection, adaptive statistical iterative reconstruction, and a model-based iterative reconstruction in abdominal CT: an experimental clinical study. Radiology 266(1):197–206PubMedCrossRef
23.
Zurück zum Zitat Scheffel H, Stolzmann P, Schlett CL, Engel LC, Major GP, Károlyi M, Do S, Maurovich-Horvat P, Hoffmann U (2012) Coronary artery plaques: cardiac CT with model-based and adaptive-statistical iterative reconstruction technique. Eur J Radiol 81(3):e363–e369PubMedCrossRef Scheffel H, Stolzmann P, Schlett CL, Engel LC, Major GP, Károlyi M, Do S, Maurovich-Horvat P, Hoffmann U (2012) Coronary artery plaques: cardiac CT with model-based and adaptive-statistical iterative reconstruction technique. Eur J Radiol 81(3):e363–e369PubMedCrossRef
24.
Zurück zum Zitat Katsura M, Matsuda I, Akahane M, Yasaka K, Hanaoka S, Akai H, Sato J, Kunimatsu A, Ohtomo K (2013) Model-based iterative reconstruction technique for ultralow-dose chest CT: comparison of pulmonary nodule detectability with the adaptive statistical iterative reconstruction technique. Invest Radiol 48(4):206–212PubMed Katsura M, Matsuda I, Akahane M, Yasaka K, Hanaoka S, Akai H, Sato J, Kunimatsu A, Ohtomo K (2013) Model-based iterative reconstruction technique for ultralow-dose chest CT: comparison of pulmonary nodule detectability with the adaptive statistical iterative reconstruction technique. Invest Radiol 48(4):206–212PubMed
25.
Zurück zum Zitat Neroladaki A, Botsikas D, Boudabbous S, Becker CD, Montet X (2013) Computed tomography of the chest with model-based iterative reconstruction using a radiation exposure similar to chest X-ray examination: preliminary observations. Eur Radiol 23(2):360–366PubMedCrossRef Neroladaki A, Botsikas D, Boudabbous S, Becker CD, Montet X (2013) Computed tomography of the chest with model-based iterative reconstruction using a radiation exposure similar to chest X-ray examination: preliminary observations. Eur Radiol 23(2):360–366PubMedCrossRef
26.
Zurück zum Zitat Vardhanabhuti V, Loader RJ, Mitchell GR, Riordan RD, Roobottom CA (2013) Image quality assessment of standard- and low-dose chest CT using filtered back projection, adaptive statistical iterative reconstruction, and novel model-based iterative reconstruction algorithms. AJR Am J Roentgenol 200:10.2214CrossRef Vardhanabhuti V, Loader RJ, Mitchell GR, Riordan RD, Roobottom CA (2013) Image quality assessment of standard- and low-dose chest CT using filtered back projection, adaptive statistical iterative reconstruction, and novel model-based iterative reconstruction algorithms. AJR Am J Roentgenol 200:10.2214CrossRef
27.
Zurück zum Zitat Pickhardt PJ, Lubner MG, Kim DH, Tang J, Ruma JA, del Rio AM, Chen GH (2012) Abdominal CT with model-based iterative reconstruction (MBIR): initial results of a prospective trial comparing ultralow-dose with standard-dose imaging. AJR Am J Roentgenol 199(6):1266–1274PubMedPubMedCentralCrossRef Pickhardt PJ, Lubner MG, Kim DH, Tang J, Ruma JA, del Rio AM, Chen GH (2012) Abdominal CT with model-based iterative reconstruction (MBIR): initial results of a prospective trial comparing ultralow-dose with standard-dose imaging. AJR Am J Roentgenol 199(6):1266–1274PubMedPubMedCentralCrossRef
28.
Zurück zum Zitat Tricarico F, Hlavacek AM, Schoepf UJ, Ebersberger U, Nance JW Jr, Vliegenthart R, Cho YJ, Spears JR, Secchi F, Savino G, Marano R, Schoenberg SO, Bonomo L, Apfaltrer P (2013) Cardiovascular CT angiography in neonates and children: image quality and potential for radiation dose reduction with iterative image reconstruction techniques. Eur Radiol 23(5):1306–1315PubMedCrossRef Tricarico F, Hlavacek AM, Schoepf UJ, Ebersberger U, Nance JW Jr, Vliegenthart R, Cho YJ, Spears JR, Secchi F, Savino G, Marano R, Schoenberg SO, Bonomo L, Apfaltrer P (2013) Cardiovascular CT angiography in neonates and children: image quality and potential for radiation dose reduction with iterative image reconstruction techniques. Eur Radiol 23(5):1306–1315PubMedCrossRef
Metadaten
Titel
Comparison of image quality between filtered back-projection and the adaptive statistical and novel model-based iterative reconstruction techniques in abdominal CT for renal calculi
verfasst von
Varut Vardhanabhuti
Sumaira Ilyas
Catherine Gutteridge
Simon J. Freeman
Carl A. Roobottom
Publikationsdatum
01.10.2013
Verlag
Springer Berlin Heidelberg
Erschienen in
Insights into Imaging / Ausgabe 5/2013
Elektronische ISSN: 1869-4101
DOI
https://doi.org/10.1007/s13244-013-0273-5

Weitere Artikel der Ausgabe 5/2013

Insights into Imaging 5/2013 Zur Ausgabe

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.