Defibrillation safety in emergency helicopter transport

https://doi.org/10.1016/S0196-0644(89)80318-XGet rights and content

Rotary aircraft play a growing role in the transport of critically ill patients who may require emergency treatment, including defibrillation, during transport. The close quarters and proximity of vital electronic equipment have generated concern among personnel carrying out defibrillation in the air. We address the chief safety issues in helicopter defibrillation by providing measurements of the transient leakage current resulting from contact with a paddle and tested in-flight electronic interference and survey the defibrillation experience of helicopter programs. Our data show that airborne defibrillation is safe. A maximum of 1.5 mA of transient leakage current was measured from a standard battery-powered defibrillator, well within the accepted safety standard of 50 mA. In flight, there was no interference with the avionics or medical equipment, and adequate clearance was available for personnel. Of the helicopter programs surveyed, 69 (87%) had defibrillated in flight without incident. We conclude that defibrillation can be performed in the helicopter without hesitation whether on the ground or in the air, provided standard defibrillation precautions are observed.