Original contribution
In vitro and in vivo ablation of porcine renal tissues using high-intensity focused ultrasound

https://doi.org/10.1016/S0301-5629(03)00981-5Get rights and content

Abstract

The aim of this paper is to present issues regarding the thermal ablation of porcine renal tissues in vitro and in vivo using high-intensity focused ultrasound (HIFU). Production of lesions in the cortex in vitro is consistent, whereas lesions in the medulla are created whenever there are no air spaces in the medulla. Typically, the lesion length at 2000 W/cm2 and 5-s pulse duration is around 20 mm and the corresponding width around 3 mm. Lesioning of a large volume was achieved by moving the transducer in a grid formation. Lesioning through a fat layer is possible provided that there are no air spaces between the fat and kidney interface. It was found that, above 3200 W/cm2 with 5-s pulse duration at 4 MHz, cavitation activity occurred in most of the lesions created. (E-mail: [email protected])

Introduction

High-intensity focused ultrasound (HIFU) is a noninvasive procedure for heating tumours without affecting the healthy tissue surrounding the tumour. Therefore, application of HIFU is being investigated as an alternative to standard surgical techniques. Although the idea of using HIFU was proposed in the middle of this century by Lynn et al. (1942), its maximum potential for clinical use has been established only recently, due to the developments of sophisticated systems (for example, Chapelon et al., 1992b, Hynynen et al., 1993a and Birhle et al. 1994). HIFU has been proven over nearly 60 years to be an effective and efficient method for ablating soft tissue, but its commercial success is still under trial.

HIFU was explored in almost every tissue that is accessible by ultrasound. The following represent some examples of some applications explored: eye (Lizzi et al. 1984), prostate Sanghvi et al., 1991, Chapelon et al., 1999, liver (ter Haar et al. 1989), brain Fry et al., 1954, Lele, 1962, Vykhodtseva et al., 1994 and kidney Linke et al., 1973, Chapelon et al., 1992a, Hynynen et al., 1995.

Recently, the technology of HIFU systems has improved because the ultrasonic therapy can be guided by imaging. Ultrasonic imaging is the simplest and most inexpensive method; however, it has poor contrast between soft tissues. On the other hand, magnetic resonance imaging (MRI) offers superior contrast, but it is more expensive. Several trials have been conducted in the area of ultrasonic imaging (for example, Seip and Ebbini, 1995, Maass-Moreno et al., 1996 and in the area of MRI (for example, Cline et al., 1992, Hynynen et al., 1993a, Hynynen et al., 1995, Hynynen et al., 1993b, thus enhancing the potential of HIFU.

The main goal of HIFU is to maintain a temperature between 50 to 100°C for a few s (typically < 10 s), in order to cause tissue necrosis. Typically, focal peak intensity between 1000 to 10,000 W/cm2 is used with pulse duration between 1 to 10 s and a frequency of 1 to 5 MHz.

This paper reports additional experience in the ablation of kidney tissue. For applications such as the treatment of benign prostatic hyperplasia (BPH), it is sufficient to destroy as much tissue as possible for the purpose of tissue debulking. For applications in oncology, destruction of all viable tumor cells is required and, therefore, protocols in this area must be very accurate and reliable. In this work, methods suggested by Malcolm and ter Haar (1996) were used to produce complete, reliable and consistent ablation of renal tissues (creation of a contiguous array of touching lesions of thermal origin, avoiding boiling and cavitation, and use of cooling to avoid merging of lesions in front of the focus).

Experiments were performed using a generic HIFU system that includes a signal generator, a radiofrequency (RF) amplifier, a 3-D robotic system, a transducer and a personal computer (PC) that controls the entire system. The transducer of 4-cm diameter was spherically focused, operating at 4 MHz. Pulse duration of 5 s was used in all the experiments to minimise effects of blood perfusion.

Results of HIFU ablation of porcine kidney in vitro and in vivo are presented. Single lesions in the cortex and medulla with or without a fat layer were created. The ultimate goal was to create large lesions, which was accomplished by moving the transducer in patterned schemes. The experimental results are compared with the results of a simulation model that includes the effect of rapid increase of attenuation during the transition of a tissue from healthy to necrotic. Thus, the kidney attenuations of porcine cortex, medulla, muscle and fat were measured as a function of thermal dose.

The attenuation was measured using a system that includes two low-intensity transducers, a signal generator, a data-acquisition card and a PC. Attenuation includes absorption, scattering and reflection but, when minimising scattering and reflection, attenuation will reflect mostly losses due to absorption. Scattering and reflection was minimised by using a homogeneous sample with a minimum number of interfaces or air ducts, and by removing acoustically troublesome gas bubbles using degassing techniques. Using these precautions in the study by Damianou et al. (1997), it was found that the value of absorption was very close to the value of attenuation, indicating that the contribution of reflection and scattering was minimised. Absorption is an acoustical property of tissue with wide variation from tissue to tissue and represents the rate at which energy in tissue is converted to heat. For short pulses (< 5 s), it is one of the primary factors contributing to the temperature elevation in tissue because the effect of blood flow is minimised.

Section snippets

Ultrasonic system

Figure 1 shows the block diagram of the system, with photographs of the actual instruments. The system consists of a signal generator (HP 33120A Hewlett Packard, now Agilent technologies, Englewood, CO), a RF amplifier (LA 100-CE, Kalmus, Bothell, WA), a 3-D positioning system (MD-2, Arrick Robotics, Hurst, TX) and a 10-cm spherically shaped bowl transducer made from piezoelectric ceramic PZT4 (Etalon, Lebanon, IN). The transducer operates at 4 MHz, has a focal length of 10 cm and diameter of

Results

Figure 7 shows the attenuation of porcine kidney (cortex and medulla), muscle and fat as a function of thermal dose. Note that the attenuation increases rapidly during necrosis and, eventually, stabilises after total necrosis has been produced.

To eliminate the effect of blood perfusion, the pulse duration of 5 s was used (Billard et al. 1990) as the standard pulse duration to be used throughout the in vitro and in vivo experiments. Pulse duration of 1 s would be desired, because treatment time

Discussion

The trend of attenuation with dose shows a similar trend to the results published by Damianou et al. (1997). The normal kidney tissue included both medulla and renal cortex, for which we found no significant difference in terms of ultrasonic attenuation. The research on muscle and fat is justified because these two tissues usually surround the kidney. The trend of attenuation with dose in all tissues was similar. There was a rapid increase of attenuation with thermal dose and, when necrosis

Acknowledgements

This work was supported by the Research Promotion Foundation (RPF) of Cyprus (contract 25/99).

References (29)

  • J.Y Chapelon et al.

    A

    Effects of high-energy focused ultrasound on kidney tissue in the rat and the dog. J Eur Urol

    (1992)
  • J.Y Chapelon et al.

    A

    In vivo effects of high-intensity ultrasound on prostatic adenocarcinoma Dunning R3327. Cancer Res

    (1992)
  • H.E Cline et al.

    MR-guided focused ultrasound surgery

    J Comput Assist Tomogr

    (1992)
  • C Damianou et al.

    The effect of various physical parameters on the size and shape of necrosed tissue volume during ultrasound surgery

    J Acoust Soc Am

    (1994)
  • Cited by (36)

    • Heart ablation using a planar rectangular high intensity ultrasound transducer and MRI guidance

      2012, Ultrasonics
      Citation Excerpt :

      Therapeutic ultrasound was explored as a therapeutic modality in almost every accessible tissue. Several studies have examined the histological changes related to therapeutic ultrasound ablation in various tissues [20,19,21,22,18,23–27]. Within the last decade, therapeutic ultrasound has been used in various in vitro models [28–32].

    • In vitro atrial septal ablation using high-intensity focused ultrasound

      2012, Journal of the American Society of Echocardiography
    • High-Intensity Focused Ultrasound (HIFU) Therapy Applications

      2009, Ultrasound Clinics
      Citation Excerpt :

      Although they concluded that HIFU therapy for breast cancer was safe and feasible, they also pointed out that more clinical trials are necessary to validate this noninvasive method. A number of animal studies have been under way to investigate the efficiency of HIFU treatment for renal tumors.106–113 Susani and colleagues114 studied the effect of HIFU ablation on the morphologic change of renal tissue in a phase I trial.

    • The feasibility of heating on tumor periphery by using high intensity focused ultrasound thermal surgery

      2008, International Communications in Heat and Mass Transfer
      Citation Excerpt :

      The main goal of HIFU is to maintain a temperature between 50 and 100 °C for a few seconds (typically less than 10 s), in order to cause tissue necrosis. Typically, focal peak intensity between 1000 and 10,000 W/cm2 is used with pulse duration between 1 and 10 s and a frequency of 1 to 5 MHz [6,7]. However, according to the vascular characteristics of tumor is that the vascular supply and blood flow in tumors are markedly different from those in normal tissues [8–10].

    • In Vitro Mitral Chordal Cutting by High Intensity Focused Ultrasound

      2008, Ultrasound in Medicine and Biology
      Citation Excerpt :

      Ultrasonic energy can be applied in a target volume to induce molecular agitation, absorptive heating and ultimately thermal coagulative tissue necrosis. Several studies have confirmed the histologic changes related to HIFU ablations in the liver, kidney, prostate, breast, brain and cardiac tissues (Chen et al. 1999; Damianou 2003; Engel et al. 2006; Fujikura et al. 2006; Lee et al. 2000; Otsuka et al. 2007; Strickberger et al. 1999; Susani et al. 1993; Vykhodtseva et al. 1994; Wu et al. 2003). We previously reported that HIFU can perforate cardiac valves by its thermal ablative effect and might prove useful for an ultrasound-based valvuloplasty (Otsuka et al. 2005).

    View all citing articles on Scopus
    View full text