Elsevier

Dental Materials

Volume 21, Issue 12, December 2005, Pages 1119-1128
Dental Materials

Minimization of the inevitable residual monomer in denture base acrylic

https://doi.org/10.1016/j.dental.2005.03.003Get rights and content

Summary

Objectives

Residual monomer ([MMA]R) in denture base acrylic continues to be of concern. The response surface of concentration vs. time and temperature for the equilibration of methyl methacrylate (MMA) and its polymer (PMMA) allows a prediction of the time to the minimum at any temperature for a closed system. It was the purpose here to determine whether this prediction applies to normal denture base processing, and whether optimum conditions could be identified.

Methods

Denture bases were processed following normal laboratory procedures, including pre-cure for 3 h at 70 °C for all tests. Commercial powder and liquid were used at either 95 or 100 °C, or a plain PMMA powder and the same liquid at 95 °C, for times ranging from 5 to 192 h. Residual MMA was determined by gas chromatography.

Results

[MMA]R decreased steadily from ∼0.25% to as low as ∼0.07% with increasing time at temperature, but did not approach equilibrium. The rate of diffusive loss of MMA appears to exceed the rate of depolymerization.

Significance

Residual monomer is inevitable for all PMMA-based products no matter what the curing conditions are. However, extended time at high temperature can allow low values to be attained, and the time allowed can compensate for processing temperatures somewhat lower than the ordinarily recommended 100 °C. It is suggested that overnight processing at 95 °C should be adopted to minimize [MMA]R and save energy. This result is of importance for work at high altitude.

Introduction

It has long been known that there is residual monomer left in denture base acrylic after processing, with many reports on the biological and mechanical consequences since the use of poly(methyl methacrylate) (PMMA) became widespread in about the 1950s. This monomer, methyl methacrylate (MMA), diffuses out of the denture and thus into adjacent oral tissues; this may result in irritant and allergic reactions, including a ‘burning mouth’ sensation (BMS). Numerous studies have dealt with such leaching [1], [2], [3], [4], [5], [6], [7], [8], [9], [10], [11], [12] and the biological effects of MMA on the acrylic denture-wearing patient [13], [14], [15], [16], [17], [18], [19], [20], [21], [22]. Broadly, it was found that the amount leached was proportional to the amount of residual MMA (MMAR) present, so more was leached from cold-cured resins. Most MMAR was lost in the first few days of immersion but became almost constant after 2 weeks. This diffusive loss of MMAR was also found to increase with immersion temperature.

Skin-patch testing for ingredients of acrylic denture materials such as PMMA, MMA, benzoyl peroxide and hydroquinone has shown mild to strong reactions toward MMA, which was therefore concluded to be an allergen [13], [14], [15], [16], [17], [18], [19], [20], [21], [22]. In addition, the mechanical properties of the polymer are affected by MMAR concentration, [MMA]R: tensile strength, modulus of elasticity and surface hardness have been found to be lower with greater [MMA]R [23], [24], [25], [26], [27], [28], [29], [30], [31], [32], [33], [34]. The explanation is the plasticizing effect of the MMA. Therefore, it is desirable on both accounts to reduce the value of [MMA]R in the processed denture to the point where both biological and mechanical effects are negligible.

There have been many studies of the effect of curing conditions on the value of [MMA]R in denture acrylic (Table 1). Thus, the effects of temperature, time, initiator concentration, curing environment, water bath vs. oven, pressure and mixing ratio (polymer: monomer) have been investigated. In summary, it has been reported that higher curing temperature and longer curing time lower the value of [MMA]R [33], [35], [36], [37], [38], [39], [40], [41]. Curing under pressure was said to allow shorter processing time with [MMA]R comparable with that for the more usual long cure without the formation of porosity [42]. It has been claimed that increasing the proportion of monomer in the mixture decreased [MMA]R on the basis that there was greater temperature rise during polymerization such that more monomer reacted [27]. This ignores the increased polymerization shrinkage that would result. Of the many curing cycles studied, 70 °C for 7 h+‘terminal boil’ was concluded to give the lowest value of [MMA]R [27], [38], [39], [40].

However, what is curious about this large literature is that there is nowhere a mention of the essential equilibrium between monomer and polymer in such a free-radical polymerization system. It seems to have been implicit that (a) MMMR is merely a consequence of inadequate processing that could be fixed if only the correct conditions could be identified, and (b) that a target of zero for [MMA]R was realistic, although never known to be attained. Furthermore, there is nowhere in the dental literature on this topic or in text books, as far as is known, any mention of the ordinary chemical considerations of the effects of temperature on the rate of reaction, or indeed of the expected progress of the reaction with time. Given that free-radical polymerization has been well-enough understood for a long time, this is all the more extraordinary.

A systematic investigation of [MMA]R against time and temperature was recently reported [43] in which the following equation was deduced as broadly representing the response surface for the reaction in a sealed system:log[M]=log[[M0]2erfc[z+bc]+d×10ze]wherez=alog(t)1000T

Here, [M0] is the initial concentration of MMA, t is the time in h and T the temperature in K; logarithms are to base 10. The values of the fitted constants and their standard errors are given in Table 2. The locus of the fitted surface minimum with respect to either variable is shown in Fig. 1. Two points may be made immediately: the time to attain the minimum at 70 °C is in excess of 300 h; and that at 100 °C around 24 h would be required. The contrast with normal processing conditions is marked.

The questions then arise as to whether (a) this result (i.e. Eq. (1)) was applicable to the effectively open system of normal denture base processing (i.e. the mould is porous), (b) whether the minimum is accessible in a reasonable time with available equipment and other normal operating circumstances, (c) what the minimum might be in such a context. Accordingly, it was the purpose of this work to study the effect of processing time and temperature on the value of [MMA]R in PMMA denture bases in the light of the predictions made from Eq. (1).

Section snippets

Materials and methods

A commercial dental acrylic product for denture bases, PMMA-based polymer powder and MMA-based monomer liquid (Pro Base Hot, Ivoclar, Schaan, Liechtenstein) was used for denture base preparation. A plain PMMA powder (i.e. with no benzoyl peroxide or other additives) (Struers, Copenhagen, Denmark) was also used as a control, but with the commercial monomer. The process followed the usual procedure employed in the Prince Philip Dental Hospital, Hong Kong [44], [45] except that teeth were omitted.

Results

The results are shown in Fig. 2: log[MMA]R declines almost linearly against log(time). There is little distinction to be made between the three sets of results.

Discussion

It is noteworthy that there is no evidence that a minimum value for [MMA]R was approached under these normal processing conditions, even at 192 h. It is therefore concluded to be essentially inaccessible in practical terms. In a closed system, [MMA] is simply the outcome of the thermodynamics and kinetics. However, in an open system, the situation is complicated by MMA loss to the surroundings (i.e. the mould). Given the clarity of the minimum documented for the sealed system, which had no

Conclusion

The pursuit of zero residual monomer has been futile, and much labor would have been avoided if the fundamental chemistry of this free-radical polymerization system had been recognized. The closed-system results do not apply to normal denture base processing in that diffusive loss of MMA dominates in comparison with depolymerization. This is seen to be an advantage in that extended processing time can lead to lower values than are ordinarily attained. In addition, a more conservative processing

Acknowledgements

We are grateful to Mrs Hui Oi Ling of the Prince Philip Dental Hospital, Hong Kong, for her help in the preparation of the acrylic denture bases. Grateful thanks are also due to the provision of GC facilities by the Department of Chemistry, The University of Hong Kong, and for the technical assistance of Mr Ho Kam Wing of that department. This work was done in partial fulfillment of the requirements for the degree of PhD by CYKL at, and supported by a studentship from The University of Hong

References (56)

  • J.L. Giunta et al.

    Allergic contact stomatitis caused by acrylic resin

    J Prosthet Dent

    (1979)
  • H.J. Caul

    Relationship between residual monomer and some physical properties of self cured resins

    J Am Dent Assoc

    (1956)
  • A. Harrison et al.

    Do self-curing acrylic resin repairs gain strength with age?

    J Dent

    (1977)
  • J. Arab et al.

    The effect of an elevated level of residual monomer on the whitening of a denture base and its physical properties

    J Dent

    (1989)
  • P. Honorez et al.

    The effect of three processing cycles on some physical and chemical properties of a heat-cured acrylic resin

    J Prosthet Dent

    (1989)
  • A. Dogan et al.

    The effect of preparation conditions of acrylic denture base materials on the level of residual monomer, mechanical properties and water absorption

    J Dent

    (1995)
  • A. Harrison et al.

    Effect of the curing cycle on residual monomer levels of acrylic resin denture base polymers

    J Dent

    (1992)
  • V. Jerolimov et al.

    Rapid curing of acrylic denture-base materials

    Dent Mater

    (1989)
  • J.S. Shim et al.

    Residual monomer concentrations in denture base acrylic resin after an additional, soft-liner, heat-cure cycle

    Dent Mater

    (1999)
  • G. Szabo et al.

    The loss of residual monomer from denture base polymers coated with an ultraviolet light-activated polymer

    Dent Mater

    (1987)
  • I.M. Harman et al.

    Effects of time and temperature on polymerization of a methacrylate resin denture base

    J Am Dent Assoc

    (1949)
  • W.F.E. Yau et al.

    Pressure and temperature changes in heat-cured acrylic resin during processing

    Dent Mater

    (2002)
  • P. Pfeiffer et al.

    Residual methyl methacrylate monomer, water sorption, and water solubility of hypoallergenic denture base materials

    J Prosthet Dent

    (2004)
  • D.C. Smith et al.

    Residual monomer methyl methacrylate in the denture base and its relation to denture sore mouth

    Br Dent J

    (1955)
  • D.C. Smith et al.

    The acrylic denture base—some effects of residual monomer and peroxide

    Br Dent J

    (1959)
  • B. Axelsson et al.

    The leaching and biological effect of the residual monomer of methyl methacrylate

    Odont Revy

    (1962)
  • S. Baker et al.

    The release of residual monomeric methyl methacrylate from acrylic appliances in the human mouth: an assay for monomer in saliva

    J Dent Res

    (1988)
  • U. Kedjarune et al.

    Release of methyl methacrylate from heat-cured and autopolymerized resins: cytotoxicity testing related to residual monomer

    Aust Dent J

    (1999)
  • Cited by (66)

    • Maxillary resection prosthesis fabricated from urethane dimethacrylate for a patient with polymethyl methacrylate allergy: A clinical report

      2023, Journal of Prosthetic Dentistry
      Citation Excerpt :

      Inadequate heat polymerization cycles or the use of certain autopolymerizing acrylates has been associated with the presence of residual acrylic monomer.10 Residual MMA left in dentures and other dental prostheses can diffuse into adjacent tissues and can result in allergic and irritating reactions.11 In this patient’s situation, placing the PMMA on her arm signified a true allergic reaction to PMMA, as opposed to an allergy to the residual MMA as is more typically seen.

    • Analysis of the residual monomer content in milled and 3D-printed removable CAD-CAM complete dentures: an in vitro study

      2022, Journal of Dentistry
      Citation Excerpt :

      In contrast, a study by Steinmassl et al. showed no statistical difference in the amount of residual monomer release with four different CAD-CAM milled RCDs when compared to the conventional heat-processed dentures [25]. The amount of monomer conversion during the conventional heat process and CAD-CAM milled complete denture were believed to be correlated to the application of high pressure and extended processing time for polymerization [17,19]. However, with 3D-printed denture base materials, the degree of residual monomer is unknown, as the photopolymerization fabrication mechanisms are different from those of heat-processed and milled dentures

    • Determination of Polyetheretherketone (PEEK) mechanical properties as a denture material

      2019, Saudi Dental Journal
      Citation Excerpt :

      Some research has proposed the use or development of alternative materials (Ucar et al., 2012). Although PMMA is a non-metallic denture base material, there is still the possibility of toxic reactions or irritations for the wearer (Alanko et al., 1996; Kanerva et al., 1997; Lung and Darvell, 2005) and for dental technicians involved in handling and fabrication (Kanerva et al., 1993; Kiec-Swierczynska, 1996). There has been interested in discovering materials that of very similar to PMMA but have better properties.

    • Thermoset polymethacrylate-based materials for dental applications

      2019, Materials for Biomedical Engineering: Thermoset and Thermoplastic Polymers
    View all citing articles on Scopus
    View full text