Original ArticleCross Calibration of Hologic QDR2000 and GE Lunar Prodigy for Forearm Bone Mineral Density Measurements
Introduction
The International Society for Clinical Densitometry (ISCD) recommends the use of forearm dual energy X-ray absorptiometry (DXA) scans when hip and/or spine cannot be measured or interpreted for technical reasons (e.g., the presence of osteophytes or implants), when the subject has hyperparathyroidism, where there is a reduction in the cortical thickness of bone, and in very obese patients who are over the weight limit for DXA table 1, 2. The ISCD recommends the use of the 33% radius (or one-third radius) region of interest (ROI) on the nondominant forearm for the diagnosis of osteoporosis, because it is the ROI that is at almost the same site when measured on instruments from different manufacturers (3). The site also predicts fracture as well as the more distal ROIs (3), which are less consistently placed by different manufacturers. Forearm bone mineral density (BMD) is useful in studies of bone growth (4), in investigating the effects of exercise (5) and in clinical trials, where the differential effect of new therapies on cortical and trabecular bone is of interest 6, 7.
There are very few studies that have considered cross calibration of forearm DXA scans 8, 9, 10, 11, 12. Most have dealt with comparisons between instruments from the same manufacturers and only 3 have compared equipment from different manufacturers 8, 11, 12. The purpose of this study was to carry out an in vivo cross calibration for forearm BMD between an Hologic QDR2000 (Hologic Inc., Bedford MA, USA) and Lunar Prodigy (GE Healthcare, Madison WI, USA) during equipment upgrade.
Section snippets
Methods
Nineteen subjects attending for routine bone densitometry were recruited to the study. Ethical approval for the study was obtained from the Nottingham Research Ethics Committee 2. Forearm BMD of the nondominant forearm was measured on a Hologic QDR2000 (QDR2000) in array mode and a Lunar Prodigy (Prodigy). BMD results were derived for the total forearm ROI and the 33% radius, ultradistal radius, 33% ulna, and the ultradistal ulna (Fig. 1). During the period of the study, daily quality control
Results
Nineteen subjects (17 females and 2 males, mean age 57 yr, range: 42–79 yr) were recruited. Figure 1 shows the positioning and definition of the forearm ROIs for the same subject on the 2 DXA instruments. Table 1 summarizes the BMD data and the mean difference and limits of agreement from the Bland & Altman plots. Figure 2 shows a typical Bland & Altman plot for the 33% radius BMD.
Figure 3 shows the in vivo cross calibrations for radius BMD. Both the 33% and ultradistal regions are shown. There
Discussion
As expected, there are systematic differences in BMD between the QDR2000 and the Prodigy (Table 1). The BMD measured on the Prodigy is 15% higher in the total forearm in common with the difference in BMD between Hologic and Lunar instruments at other sites (16). The difference in BMD is correlated with the BMD (Fig. 2) so is 20–25% higher in the cortical regions of the 33% ulna and 33% radius and smaller in the ultradistal ROIs, where the BMD is lower. The fact that a single cross-calibration
Conclusions
There are systematic differences in forearm BMD between scanners. At low BMD, the Prodigy appears to underestimate forearm BMD. The standard error of estimate about the in vivo cross calibration is such that it cannot be used to transfer individual patients between instruments, but could be applied to clinical trial data. Spine phantoms do not provide an adequate cross calibration for forearm BMD.
Acknowledgments
We would like to acknowledge the grant funding to support this work from Bio-Imaging Technologies Inc and NPS Pharmaceuticals. We also gratefully acknowledge the support and review of this work from Dr Colin Miller from Bio-Imaging and Dr Stuart Jackson.
References (21)
- et al.
Bone measurements in asymptomatic primary hyperparathyroidism
Bone
(1998) - et al.
Calcium supplementation and bone mineral density in females from childhood to young adulthood: a randomized controlled trial
Am J Clin Nutr
(2005) - et al.
Cortical and trabecular bone at the forearm show different adaptation patterns in response to tennis playing
J Clin Densitom
(2004) - et al.
Bone density changes in Paget's disease 2 years after iv pamidronate: profound, sustained increases in pagetic bone with severity-related loss in forearm nonpagetic cortical bone
Bone
(2003) - et al.
Cross calibration of QDR-2000 and QDR-1000 dual-energy X-ray densitometers for bone mineral and soft-tissue measurements
Bone
(1995) - et al.
Pencil-beam vs fan-beam dual-energy X-ray absorptiometry comparisons across four systems: body composition and bone mineral
J Clin Densitom
(2004) - et al.
Impact of using the ultradistal radius region of interest on diagnostic classification
J Clin Densitom
(2004) - et al.
Cross calibration of DXA as part of an equipment replacement program
J Clin Densitom
(2006) - et al.
European semi-anthropomorphic phantom for the cross-calibration of peripheral bone densitometers: assessment of precision accuracy and stability
Bone Miner
(1994) - et al.
Which central dual X-ray absorptiometry skeletal sites and regions of interest should be used to determine the diagnosis of osteoporosis?
J Clin Densitom
(2002)
Cited by (7)
Cross-Calibration Study of The Stratos And Hologic QDR 4500A Dual-Energy X-Ray Absorptiometers to Assess Bone Mineral Density And Body Composition
2023, Journal of Clinical DensitometryQuantitative Comparison of 2 Dual-Energy X-ray Absorptiometry Systems in Assessing Body Composition and Bone Mineral Measurements
2016, Journal of Clinical DensitometryCitation Excerpt :There have been efforts to make cross-system comparisons between the products of a single manufacturer (9,10), between different DXA systems (7,11–13), and between DXA and other imaging modalities (14–16). BMD was cross-compared between Lunar Prodigy (GE Healthcare, Madison, WI) and QDR 2000 (Hologic Inc., Bedford, MA) (7). Bone mineral content (BMC) and fat mass were cross-compared between Lunar Expert (GE Healthcare, Madison, WI), Lunar Prodigy (GE Healthcare, Madison, WI), and Hologic Delphi W (Hologic Inc., Bedford MA) (11).
Forearm Bone Mineral Density Measurement With Different Scanning Positions: A Study in Right-Handed Chinese Using Dual-Energy X-Ray Absorptiometry
2012, Journal of Clinical DensitometryCitation Excerpt :The CV is a popular measure of precision; it is valuable when comparing the precision of BMD measurement at different skeletal sites or between different techniques. The precision of forearm BMD measurement has been the focus of various studies (16,17,19–22). In a study by Boyanov (19), forearm BMD was measured using single X-ray absorptiometry (DTX-100 device), and the %CV at the distal radius + ulna was 1.65, whereas the %CV at the UD radius was 2.01.
Canadian association of radiologists technical standards for bone mineral densitometry reporting
2011, Canadian Association of Radiologists JournalOpportunistic hand radiographs to screen for low forearm bone mineral density: a prospective and retrospective cohort study
2024, BMC Musculoskeletal Disorders