Volume 36, Issue 2, Summer 2011, Pages 119-125
Medical Dosimetry

Implications of Artefacts Reduction in the Planning CT Originating from Implanted Fiducial Markers

https://doi.org/10.1016/j.meddos.2010.02.002Get rights and content

Abstract

The efficacy of metal artefact reduction (MAR) software to suppress artefacts in reconstructed computed tomography (CT) images originating from small metal objects, like tumor markers and surgical clips, was evaluated. In addition, possible implications of using digital reconstructed radiographs (DRRs), based on the MAR CT images, for setup verification were analyzed. A phantom and 15 patients with different tumor sites and implanted markers were imaged with a multislice CT scanner. The raw image data was reconstructed both with the clinically used filtered-backprojection (FBP) and with the MAR software. Using the MAR software, improvements in image quality were often observed in CT slices with markers or clips. Especially when several markers were located near to each other, fewer streak artefacts were observed than with the FBP algorithm. In addition, the shape and size of markers could be identified more accurately, reducing the contoured marker volumes by a factor of 2. For the phantom study, the CT numbers measured near to the markers corresponded more closely to the expected values. However, the MAR images were slightly more smoothed compared with the images reconstructed with FBP. For 8 prostate cancer patients in this study, the interobserver variation in 3D marker definition was similar (<0.4 mm) when using DRRs based on either FBP or MAR CT scans. Automatic marker matches also showed a similar success rate. However, differences in automatic match results up to 1 mm, caused by differences in the marker definition, were observed, which turned out to be (borderline) statistically significant (p = 0.06) for 2 patients. In conclusion, the MAR software might improve image quality by suppressing metal artefacts, probably allowing for a more reliable delineation of structures. When implanted markers or clips are used for setup verification, the accuracy may slightly be improved as well, which is relevant when using very tight clinical target volume (CTV) to planning target volume (PTV) margins for planning.

Introduction

In radiotherapy, computed tomography (CT) images are essential in 2 aspects of treatment planning: (1) 3D definition of the target volume and organs at risk (OAR) in relation to the body surface, and (2) provision of electron densities, which are derived from the Hounsfield units (HU) in the CT scan, to be used for tissue heterogeneity corrections in dose calculations.1 Often, high-density objects, such as metal prostheses, surgical clips, tumor markers, or dental fillings, generate streaklike artefacts in CT images. The main source for these artefacts is an inaccurate beam-hardening correction in filtered back-projection (FBP), the algorithm most commonly used for CT image reconstruction. FBP assumes that the X-ray beam is monoenergetic, yielding a total attenuation that is linear with the object thickness. But in reality the X-ray beam is polyenergetic. To reduce this problem, CT scanners generally have correction software that is optimized for soft tissue and bone, but this does not solve the problem for metal objects with a much higher attenuation coefficient. Because of the streak artefacts, accurate delineation of the target volume and sensitive structures may be difficult, and near the artefacts the obtained HU may not be representative of the real electron densities.

During patient treatment, metal markers or clips that are implanted in or near the target are used for online and offline position verification.2, 3 Although a tumour is generally not visible on megavoltage or kilovoltage transmission images acquired during treatment, the metal markers are often easily visible. An advantage of the use of the markers is that their position is more representative for the tumor position than the surrounding bony structures, allowing the use of smaller clinical target volume to planning target volume (CTV-PTV) margins during treatment planning. However, these small metal inserts also produce artefacts that are often visible in the CT images. When using the CT to derive digitally reconstructed radiographs (DRRs) as reference for position verification, the exact position and orientation of the markers might be obscured, especially when several of them are located closely to each other, which possibly results in inaccurate match results and a reduced success rate when an automatic match procedure is used.

To minimize metal artefacts in CT images, improved image reconstruction algorithms based on the raw projection data, i.e., the matrix of ray attenuations related to different angles acquired by the scanner, have been studied.1, 4, 5, 6, 7, 8 In iterative reconstruction methods, projection data associated with metal objects are ignored and the images are reconstructed based on the remaining data using a dedicated correction algorithm being applied in an iterative procedure.8, 9, 10 In projection interpolation–based methods, the projection through metal objects are replaced using an interpolation of neighbor projections.1, 6, 11, 12, 13 Recently, a metal artefact reduction (MAR) image reconstruction algorithm, which was described by Mahnken et al.,6 was implemented by Siemens in a research version of their CT software. For patients with large metal implants, it was demonstrated that this MAR software significantly improved CT image quality.1, 5, 14, 15

In this study we evaluate the efficacy of this MAR software to suppress artefacts from small metal objects, like tumor markers and surgical clips, both in reconstructed CT images and DRRs used for position verification. For a group of 8 prostate cancer patients, we evaluated the impact on the accuracy of marker definition and position verification when using DRRs derived from CT scans that were reconstructed without and with the MAR software, respectively.

Section snippets

MAR image reconstruction algorithm

The MAR image reconstruction algorithm, described in detail by Mahnken et al.6 uses a projection interpolation method. First, the metal objects are extracted from a conventionally reconstructed image using a threshold-based segmentation of pixels with a high HU value. These pixels are then projected forward to identify channels in the projection images that pass through the metal object. Each affected projection is then corrected by replacing metal shadows in the raw projection data using

Phantom study

Figure 1 shows a visual comparison of transversal CT images from the phantom study. The images in Fig. 1A and C (with the markers placed parallel and perpendicular to the CT scan direction, respectively) were reconstructed using the conventional FPB algorithm used in clinic. Because the algorithm does not suppress metal artefacts, bright streaks and a dark region are visible around the markers. Figure 1B and D shows the corresponding images obtained after using the MAR software. It is clear

Discussion and Conclusions

Artefacts in CT scans caused by metal implants, but also by small markers may be a serious problem for radiotherapy purposes. Because of the artefacts, accurate delineation of the target and sensitive organs may be impossible. To make sure that the target is adequately covered, larger CTV-PTV margins may be required to compensate for this problem, thereby often increasing the dose delivery to sensitive structures. In addition, the artefacts may hamper accurate position verification of patients

Acknowledgments

The authors would like to thank Siemens Medical Solutions for providing a research version of the MAR software and Marjolein van Os, Anneke Snoep, Elise Nicolai, and Leonie Schipper for their contribution to the study.

Cited by (12)

  • Post-mastectomy radiation therapy after breast reconstruction: Indications, timing and results

    2017, Breast
    Citation Excerpt :

    Internal high-density metallic ports through which saline is injected into tissue expanders may result in difficulties in imaging and perturbation in dose distribution around the port [24]. This problem may be partially alleviated through the implementing a metal artifact reduction image reconstruction algorithm [25]. According to a metaanalysis including 1105 patients, radiotherapy after prosthetic reconstruction is associated with over four times higher risk of complications compared to reconstruction without radiotherapy (OR 4.2; 95% CI, 2.4–7.2) [3].

  • The impact of orthopedic metal artifact reduction software on interreader variability when delineating areas of interest in the head and neck

    2015, Practical Radiation Oncology
    Citation Excerpt :

    Nonetheless, no objective results have been presented. Furthermore, 1 of the analyzed algorithms had a negative impact on structure contours, according to Kassim et al,16 and could increase the interreader variability. Our study results demonstrate a decrease in interreader variability when using O-MAR in the presence of a metal artifact.

  • Fiducial Placement for Guidance of Radiotherapy

    2020, Therapeutic Endoscopic Ultrasound
View all citing articles on Scopus
View full text