Elsevier

Radiotherapy and Oncology

Volume 109, Issue 3, December 2013, Pages 367-369
Radiotherapy and Oncology

Contact radiotheraphy
Monte Carlo dosimetry for the Papillon P50 contact radiotherapy and IORT device

https://doi.org/10.1016/j.radonc.2013.10.032Get rights and content

Abstract

Contact radiotherapy uses small field sizes and very short SSDs to deliver low-energy X-rays very close to the position of a tumour. We present a Monte Carlo simulation of the Papillon contact radiotherapy machine, and in particular the backscatter factor, to investigate whether the factors given in current Codes of practice for kV dosimetry remain valid under these conditions.

Section snippets

Materials and methods

Two systems of dosimetry are in common use for devices with an accelerating potential in the tens of kilovolts range and are described in detail in [2], [3], [8]. Both methods are based around an air kerma calibration and determine the dose at the surface of a water phantom.

The chamber factor [2], [3] has been determined experimentally [9], [10]. We performed similar measurements in air and then at the surface of a water phantom according to the IPEMB Code of practice [2], [3]. Setting the two

Results

The water-to-air mass energy-absorption coefficient ratio is relatively insensitive to the photon spectrum, varying on the level of tenths of a percent over the tens of kV range. The value found using the Penelope physics models implemented in Geant 4 was 1.022 with a statistical uncertainty of approximately 0.05%.

The calculated backscatter factor is shown in Fig. 1 for two of the applicators, the 22 mm and the 25 mm applicators. As the backscatter factor applies to an abstract point at the

Discussion

The mass energy absorption coefficient ratios were 0.2–0.3% smaller than those quoted in [2], [3]. Clinically a 0.3% difference is not significant. The backscatter factor calculated by the Monte Carlo was greater by approximately 1% than would be found by using the HVL, FSD and field-size based tables in [2], [3]. Assuming the locally measured HVL of 0.64 mm Al, then the backscatter factor according to [4], [7] is approximately 1.069, compared to our 1.082 for the 25 mm applicator and 1.068

Conclusions

We find that the dosimetry system used for other kV treatments is applicable to the Papillon P50. The mass energy-absorption coefficient ratio is consistent with previous values although this value is not very sensitive to beam quality or set-up conditions. The backscatter factor has been found to be 1.076 + 0.001 for a 22 mm applicator, at 2.9 mm FSD. Similarly the chamber factor was found to be consistent with the 1.06 currently recommended, albeit with large experimental errors. A full Monte

References (14)

  • S. Agostinelli et al.

    Geant4—a simulation toolkit

    NIM A

    (2003)
  • O. Croce et al.

    Contact Radiotherapy using a 50 kV X-ray system: evaluation of relative dose distribution with the Monte Carlo code PENELOPE and comparison with measurements

    Radiat Phys Chem

    (2012)
  • J.-P. Gerard et al.

    Renaissance of contact X-ray therapy for treating rectal cancer

    Expert Rev Med Devices

    (2011)
  • S.C. Klevenhagen et al.

    The IPEMB code of practice for the determination of absorbed dose for X-rays below 300 kV generating potential (0.035 mm Al-4 mm Cu HVL; 10–300 kV generating potential

    Phys Med Biol

    (1996)
  • R.J. Aukett et al.

    Addendum to the IPEMB code of practice for the determination of absorbed dose for X-rays below 300 kV generating potential (0.035 mm Al-4 mm Cu HVL)

    Phys Med Biol

    (2005)
  • B. Grosswendt

    Dependence of the photon backscatter factor for water on source-to-phantom distance and irradiation field size

    Phys Med Biol

    (1990)
  • S.C. Klevenhagen et al.

    Memorandum from the Institute of Physical Sciences in Medicine: back-scatter and F-factors for low- and medium-energy X-ray beams in radiotherapy

    Br J Radiol

    (1991)
There are more references available in the full text version of this article.

Cited by (6)

View full text