Volume 16, Issue 7, July 2016, Pages 835-841
The Spine Journal

Clinical Study
Differences in bone mineral density of fixation points between lumbar cortical and traditional pedicle screws

https://doi.org/10.1016/j.spinee.2015.11.034Get rights and content

Abstract

Background Context

The use of a novel lumbar pedicle cortical bone trajectory (CBT) screw has recently gained popularity, allowing for a minimally invasive approach and potentially improved screw purchase. However, to date, no studies have identified the ideal patient population to utilize this technology.

Purpose

This study reports the bone mineral density (BMD) using Hounsfield units (HUs) along a CBT screw pathway. Patients with a greater difference in density of bone in the lumbar vertebrae between the fixation points of the CBT and traditional pedicle screw may be optimal candidates to realize the advantages of this technique.

Study Design/Setting

A cross-sectional observational anatomic study was carried out.

Patient Sample

The sample comprised 180 randomly selected patients with lumbar computed tomography imaging from L1 to L5 spinal levels.

Outcome Measures

This study used computed tomography image-derived HUs as a metric for BMD.

Methods

A total of 180 patients without previous lumbar surgery with computed tomography imaging of the lumbar spine met the inclusion criteria. Patients were chosen randomly from an institutional database based on age (evenly distributed by decade of life) and gender. Hounsfield units were measured at the expected end fixation point for both a cortical (superior/posterior portion of the vertebral body) and traditional pedicle trajectory (mid-vertebral body).

Results

Hounsfield unit values measured at the end fixation point for the CBT screw were significantly greater than that of the traditional pedicle screw in all age groups. The relative difference in HU values significantly increased with each decade of age (p<.001) and caudal lumbar level (p<.001). In the osteoporotic group, as determined by well-established HU values, there was a significantly greater difference in the BMD of the CBT fixation point compared with the traditional trajectory (p=.048–<.001).

Conclusions

Bone mineral density as measured by HU values for the fixation point of the CBT screw is significantly greater than that of the traditional pedicle screw. This difference is even more pronounced when comparing osteoporotic and elderly patients to the general population. The data in this study suggest that the potential advantages from the CBT screw such as screw purchase may increase linearly with age and in osteoporotic patients.

Introduction

Traditional lumbar pedicle screws are widely used in the treatment of lumbar spinal pathology. Despite its success in achieving stability in these patients, complications such as screw loosening, migration, and pullout lead to suboptimal outcomes and the need for reoperation [1], [2]. In 2009, Santoni et al. introduced data demonstrating that a cortical bone trajectory (CBT) screw in the pedicle enhanced pullout strength by 30% when compared with the traditional pedicle screw [3]. Similarly, Baluch et al. reported that cortical screws had superior resistance to craniocaudal toggling and required force for displacement [4]. This has led to clinical adoption of this technique in certain clinical situations. The technique of insertion for the cortical screw allows for a smaller incision and less exposure in comparison to the traditional lumbar pedicle screw where the transverse processes need to be visualized. Because its insertion point is just medial to the lateral aspect of the pars interarticularis and uses an 8° lateral and 25° cranial screw angulation, operative time and, potentially, recover time are reduced [5]. Although limited in sample size, early clinical studies have reported good outcomes with the use of CBT for a variety of pathologies [6], [7], [8].

Bone mineral density (BMD) has been demonstrated to positively correlate with increased pullout strength [3]. Since the end point of the traditional pedicle screw often lies within the cancellous bone of the mid-vertebral body, which preferentially diminishes with age [9], [10], there is an increased risk of toggle that may eventually loosen the screw purchase. On the other hand, the CBT screw fixation point lies within the cortical bone of the pedicle and vertebral body, which theoretically reduces this risk. Furthermore, this area of bone maintains its density more effectively than that of the mid-vertebral body during the aging process [11]. This technique also allows for a shorter screw while achieving mechanical purchase equivalent to or better than that of the traditional pedicle screw, most notably in osteopenic or osteoporotic patients in cadaveric studies [12].

Dual X-ray absorptiometry (DXA) is often used to determine BMD in patients. However, recently, the use of Hounsfield units (HU) measured on computed tomography (CT) has been demonstrated to be a reliable proxy for BMD [13], [14], [15]. Schreiber et al. noted the distinct advantages that CT imaging provides over traditional DXA to the spine surgeon, including the ability to determine density in specific regions of interest (ROIs) including the vertebral body trabeculae [16], which has been shown to be most predictive of fracture risk [17]. Further, DXA can be unreliable in cases of severe degeneration, scoliosis, or after instrumented surgery [16], which commonly afflicts the patient population who requires spine surgery. Consequently, HUs may provide an equivalent if not improved measure for relative BMD in this patient population.

Several studies investigating the utility of the CBT technique have yielded a favorable outlook regarding its incorporation into surgical practice [4], [18]. However, it is yet unclear what the ideal patient population is to utilize this novel technology. Patients who have a greater difference in the BMD along the trajectory of the CBT versus traditional pedicle screw may best realize the advantages of the CBT screws.

Section snippets

Methods

After appropriate Institutional Review Board approval, an institutional medical database was queried for patients with current procedural terminology (CPT) codes 72131, 72132, and 72133 (CT lumbar spine with, without, or with and without contrast) over a time period of one decade (2004–2014). Exclusion criteria included patients less than 30 or greater than 90 years of age, recent trauma to the lumbar spine, incomplete imaging of the lumbar spine, or prior instrumentation of the lumbar

Results

Calculations of HU values for 180 patients in this database demonstrated a significantly greater BMD for the fixation point of the CBT than that of the traditional pedicle screw at each lumbar level (p<.001 for each level) (Table 1). Relative differences in BMD significantly increased with each decade of age at each lumbar spinal level (p=.0003–<.0001) (Table 2). Further analysis using bivariate correlation revealed age to be positively correlated with an increaseddifference in BMD between the

Discussion

The CBT pedicle screw is becoming increasingly popular as an alternative, more minimally invasive approach for lumbar fusion procedures [9]. Coupled with a diminished necessity for posterolateral dissection and a screw insertion that avoids the adjacent spinal cord, nerve roots, and vascular structure, a number of advantages exist with this novel technique [4]. Wray et al. illustrated that an increased bone quality was present in CBT screw trajectory compared with traditional pedicle screw

Conclusions

An examination of the lumbar vertebrae in 180 patients has demonstrated a higher BMD for the anchoring of the CBT screw, as measured by HU, when compared with that of the traditional pedicle screw. This difference was greater in caudal lumbar levels as well as osteoporotic and elderly patients compared with the general population. The data in this study suggest that the advantages from the CBT screw may be best realized in this patient population.

References (28)

  • J. Lee et al.

    The insertional torque of a pedicle screw has a positive correlation with bone mineral density in posterior lumbar pedicle screw fixation

    J Bone Joint Surg Br

    (2012)
  • S. Soshi et al.

    An experimental study on transpedicular screw fixation in relation to osteoporosis of the lumbar spine

    Spine

    (1991)
  • SuB.W. et al.

    An anatomical study of the mid-lateral pars relative to the pedicle footprint in the lower lumbar spine

    Spine

    (2009)
  • S. Wray et al.

    Pedicle screw placement in the lumbar spine: effect of trajectory and screw design on acute biomechanical purchase

    J Neurosurg Spine

    (2015)
  • Cited by (0)

    FDA device/drug status: Not applicable.

    Author disclosures: HTM: Nothing to disclose. SMM: Nothing to disclose. SZH: Nothing to disclose. TJJ: Nothing to disclose. AAP: Royalties: Amedica (C), Ulrich (C), Biomet (C); Private Investments: Amedica (<1%), Cytonics (<1%), Nocimed (<1%), Vital5 (<1%); Consulting: Biomet (B), Stryker (B), DePuy (B); Board of Directors: Cervical Spine Research Society (None), outside the submitted work. WKH: Consulting: Medtronic, Inc (B), Stryker (E), Bacterin (B), Graftys (A), Lifenet (A), AONA (Nonfinancial), Bioventus (B), Globus (B), Spinesmith (Nonfinancial), outside the submitted work.