A novel breast ultrasound system for providing coronal images: System development and feasibility study

https://doi.org/10.1016/j.ultras.2014.09.009Get rights and content

Highlights

  • A new 3D ultrasound imaging system was developed to generate breast coronal images to further provide more diagnosis information in breast examinations.

  • This system was based on a clinical free-hand 2D probe to facilitate routine examination.

  • For coronal imaging purpose, the omitted volume formation procedure could reduce resampling approximation.

  • Good accuracy and repeatability of this system were demonstrated in phantom experiments.

  • The potential of this new system for clinical applications was demonstrated by in vivo experiments.

Abstract

Breast ultrasound images along coronal plane contain important diagnosis information. However, conventional clinical 2D ultrasound cannot provide such images. In order to solve this problem, we developed a novel ultrasound system aimed at providing breast coronal images. In this system, a spatial sensor was fixed on an ultrasound probe to obtain the image spatial data. A narrow-band rendering method was used to form coronal images based on B-mode images and their corresponding spatial data. Software was developed for data acquisition, processing, rendering and visualization. In phantom experiments, 20 inclusions with different size (5–20 mm) were measured using this new system. The results obtained by the new method well correlated with those measured by a micrometer (y = 1.0147x, R2 = 0.9927). The phantom tests also showed that this system had excellent intra- and inter-operator repeatability (ICC > 0.995). Three subjects with breast lesions were scanned in vivo using this new system and a commercially available three-dimensional (3D) probe. The average scanning times for the two systems were 64 s and 74 s, respectively. The results revealed that this new method required shorter scanning time. The tumor sizes measured on the coronal plane provided by the new method were smaller by 5.6–11.9% in comparison with the results of the 3D probe. The phantom tests and preliminary subject tests indicated the feasibility of this system for clinical applications by providing additional information for clinical breast ultrasound diagnosis.

Introduction

Breast cancer is the most common cancer in women worldwide. In the Global Health Estimates of World Health Organization (WHO), it was estimated that 508,482 women died of breast cancer in 2011 in the world [1]. In America, it was reported that 226,870 women were diagnosed with breast cancer and 39,510 of them died of breast cancer in 2012 [2]. According to the report of Breast Cancer UK, breast cancer accounted for 31% of cancers diagnosed in women [3]. Up to now, there has not been an effective method to prevent breast cancer and early detection has remained the cornerstone for breast cancer control [4]. Among all breast cancer detection methods, ultrasound plays an important role in breast cancer deaths decline for its advantages of radiation-free, real-time and suitable for dense breast [5], [6]. Ultrasound has long been recognized as a valuable tool to distinguish between cysts and solid masses. With the rapid development of ultrasound techniques and greatly increased images quality, breast ultrasound can now not only be used for characterizing cysts, but also differentiating benign from malignant lesions. In a breast abnormalities (259 carcinomas, 1820 benign) examination, ultrasound could help to avoid unnecessary biopsy with benign diagnosis results in 71 suspicious cases at palpation or mammography [7]. Therefore, routine ultrasound examination can help to reduce unnecessary biopsies.

In clinical breast ultrasound examination, 2D ultrasound probe is routinely used which can only provide transverse and longitudinal images but no coronal images. However, information on this plane has been proved to be beneficial for clinical diagnosis [8], [9], [10], [11], [12], [13], [14]. Rotten et al. analyzed images of normal breast tissue and breast lesions and found four diagnosis features on coronal plane [8]. Among these features, one was defined as compressive pattern which was thought to be associated with benign lesions. In this pattern, the continuous hyperechoic bands of tissue peripheral to the masses appeared to be distinct from the central part. Another feature was called converging pattern which was a typical characteristic of malignant lesions. For this pattern, a stellate distortion consisting of alternating hypoechoic and hyperechoic lines converged towards to the hypoechoic central masses [8]. In the study of Chen et al., the two features were described as hyperechoic rim and retraction phenomenon. They were used independently to differentiate breast lesions and good accuracy (95.9% for hyperechoic rim; 96.8% for retraction phenomenon) and specificity (92.8% for hyperechoic rim; 100% for retraction phenomenon) were reported [9]. In fact, the feature of retraction phenomenon on coronal plane, also called spiculation, was reported by other researchers with high specificity (98.4% in [10]; 94.6% in [11]) in tumor diagnosis. In the report of Meyberg-Solomayer et al., the coronal plane could provide tumor classification information when the infiltrative zone was not visible (17 of 39 cases) or unclear (6 of 8 cases) in 2D ultrasound imaging [12]. This result demonstrated that the image on this plane could offer a better assessment when the infiltrative zone surrounding the lesion was unclear or not visible on conventional 2D images, which could help to reduce biopsies. Images along coronal plane were also beneficial for tumor extent measurement [13] and ultrasound-guided vacuum-assisted core-needle biopsy [14]. Based on images on coronal images, various computer-aided diagnosis (CAD) methods were presented to help to automatically detect tumor candidate [15], mark spiculated masses [16], [17], and classify tumor stages [18], [19].

Breast coronal images can be provided by the technique of three-dimensional (3D) ultrasound imaging. Many researchers have been studying on this technique. One approach of 3D breast ultrasound imaging was to scan the breast using the conventional 2D probe, which was driven by a mechanical motor [20], [21], [22]. A typical representation of this approach was the method proposed by Kotsianos-Hermle et al. [20]. In this method, breast was compressed by two paddles and a probe was driven mechanically on the top of the paddle to acquire the breast images. Another approach was to scan the patient with a specially designed probe when the patient was in supine position, such as the commercial products Automated Breast Volume Scanner (ABVS) of Siemens and the Automated Whole-Breast Ultrasound (AWBU) of Sonocine [23], [24], [25]. Another approach for 3D ultrasound imaging is the 2D array ultrasound probe which uses electronic pyramidal scanning [26]. They have been used successfully for real-time 3D imaging of the heart, where high volume frame rate (40 volumes per second) is required [27], but are seldom used for breast imaging. A high volume frame is not as necessary for breast imaging, unless breast tissue motion needs to be tracked in three dimensions in real time, as in 3D elastography [28]. Therefore, the first two approaches of 3D ultrasound imaging can provide breast coronal images. However, in these methods, the driven motor or specially designed probes were required for scanning. These equipments were bulky and large, which were inconvenient for clinical scanning to offer regular motions. In addition, the moving manner of the probe in these systems was predefined so the operator could not move the probe to the desired position freely. Some regions such as axillary region and tissue against the chest wall were not accessible by using these systems. Therefore, there are still many works to be done before 3D ultrasound imaging technique can be widely used in clinical breast examination. 2D ultrasound imaging remains the dominant scanning mode for clinical breast ultrasound diagnosis.

Accordingly, this study was aimed to develop a breast ultrasound system for providing coronal images based on the clinical 2D ultrasound scanner. In the following sections, this system is described in details. The system accuracy and reliability tests based on phantoms are presented. Preliminary clinical tests were also performed to demonstrate the system feasibility.

Section snippets

System overview

A corresponding freehand 3D ultrasound annotation system was previously developed and successfully used for annotating breast ultrasound images [29]. Fig. 1 shows the diagram of this system. It consisted of three main components: an ultrasound machine (EUB-8500, Hitachi, Tokyo, Japan) with a linear 2D probe (EUP-L65/6–14 MHz, Hitachi, focused probe, 6–14 MHz), an electromagnetic spatial sensing device (med-SAFE, Ascension Technology, Burlington, VT, USA) and a computer with Intel Core i5 3.35 GHz

Validation results

Fig. 6 is a typical coronal image of the phantom. The dark part indicated the inclusion. The dimensional measurement was also shown on the image. The measurement results were compared with the actual sizes of inclusions to demonstrate the system accuracy. Fig. 7 illustrates the correlation about the phantom diameters measured by the 3D ultrasound rendering image and the micrometer. The results demonstrated a very good linear correlation (y = 1.0147x, R2 = 0.9927) between the results obtained by the

Conclusions and discussion

Based on a clinical ultrasound scanner and a spatial sensor, a new system for providing breast coronal images was successfully developed in this paper. The tests on phantoms and three patients in vivo were conducted to preliminarily demonstrate the system feasibility. Unlike the existed 3D breast ultrasound scanner which requires a specially designed 3D probe [20], [21], [22], [23], [24], [25], [26], this method simply adds a spatial sensor on the routinely used 1D array linear probe. The

Acknowledgments

This work was partially supported by the Hong Kong Innovative Technology Fund for HK-Guangdong Collaboration (GHP/047/09) and Hong Kong Polytechnic University (G-YL74).

References (61)

  • Q.H. Huang et al.

    A new scanning approach for limb extremities using a water bag in freehand 3-D ultrasound

    Ultrasound Med. Biol.

    (2005)
  • R.W. Prager et al.

    Rapid calibration of 3-D freehand ultrasound

    Ultrasound Med. Biol.

    (1998)
  • A.K. Thittai et al.

    Small breast lesion classification performance using the normalized axial-shear strain area feature

    Ultrasound Med. Biol.

    (2013)
  • R. Prager et al.

    Stradx: real-time acquisition and visualization of freehand three-dimensional ultrasound

    Med. Image Anal.

    (1999)
  • G. Treece et al.

    Fast surface and volume estimation from non-parallel cross-sections, for freehand 3-D ultrasound

    Med. Image Anal.

    (1999)
  • A. Gee et al.

    Narrow-band volume rendering for freehand 3D ultrasound

    Comput Graph.

    (2002)
  • O.V. Solberg et al.

    Freehand 3D ultrasound reconstruction algorithms – a review

    Ultrasound Med. Biol.

    (2007)
  • Q.H. Huang et al.

    Development of a portable 3D ultrasound imaging system for musculoskeletal tissues

    Ultrasonics

    (2005)
  • Q.H. Huang et al.

    An adaptive squared-distance-weighted interpolation for volume reconstruction in 3D freehand ultrasound

    Ultrasonics

    (2006)
  • T.M. Mak et al.

    Liver fibrosis assessment using transient elastography guided with real-time B-mode ultrasound imaging: a feasibility study

    Ultrasound Med. Biol.

    (2013)
  • C. Leung et al.

    A real-time intrasubject elastic registration algorithm for dynamic 2-D ultrasound images

    Ultrasound Med. Biol.

    (2009)
  • G.M. Treece et al.

    Correction of probe pressure artifacts in freehand 3D ultrasound

    Med. Image Anal.

    (2002)
  • ...
  • R. Siegel, D. Naishadham, A. Jemal. Cancer Statistics, 2012. pp....
  • ...
  • American Cancer Society

    Breast Cancer Facts & Figures, 2011–2012

    (2011)
  • G.F. Xiao, 3-D free-hand ultrasound imaging and image analysis of the breast, PhD Dissertation. Department of...
  • S. Ciatto et al.

    The contribution of ultrasonography to the differential-diagnosis of breast-cancer

    Neoplasma

    (1994)
  • D. Rotten et al.

    Analysis of normal breast tissue and of solid breast masses using three-dimensional ultrasound mammography

    Ultrasound Obstet. Gynecol.

    (1999)
  • Z.L. Wang et al.

    Comparison of automated breast volume scanning to hand-held ultrasound and mammography

    Radiol. Med.

    (2012)
  • Cited by (11)

    • Automated Breast Ultrasound Interpretation Times: A Reader Performance Study

      2018, Academic Radiology
      Citation Excerpt :

      Although ultrasound can help address the challenges of cancer detection in dense breasts, traditional handheld ultrasound has several important limitations, including operator dependency, variable scan quality and reproducibility, and long acquisition times, which raise concerns for broad-scale implementation as an adjunct breast cancer screening tool (25–30). Automated breast ultrasound (ABUS), which is a US Food and Drug Administration–approved adjunct-screening tool specifically for women with dense breast tissue (31), has the potential to overcome several limitations of handheld screening ultrasound because it acquires three-dimensional (3D) volumes viewable in three orthogonal planes, reducing operator variability and subjectivity (32,33). However, as alternative screening methodologies such as ABUS are gaining popularity, questions have been raised regarding time required for radiologist interpretation of the ABUS examination.

    • 1000-Case Reader Study of Radiologists' Performance in Interpretation of Automated Breast Volume Scanner Images with a Computer-Aided Detection System

      2018, Ultrasound in Medicine and Biology
      Citation Excerpt :

      The images acquired in the transverse plane are reconstructed in three dimensions to allow interpretation in the coronal plane, which has proved valuable in diagnosis and differential diagnosis in breast cancer. The ABVS has been found to have benefits in standardization of image acquisition and quality, high reproducibility for lesion characterization and ease of understanding for clinicians (Jiang et al. 2015; Tozaki and Fukuma 2012). However, the interpretation of ABVS images remains associated with certain degree of missed diagnoses and false-negative interpretations (Dromain et al. 2013) influenced by artifacts, viewing area and visual fatigue and distraction of the radiologist.

    • Freehand three-dimensional ultrasound system for assessment of scoliosis

      2015, Journal of Orthopaedic Translation
      Citation Excerpt :

      The feasibility of using landmarks in cross-sectional B-mode ultrasound images acquired with continuous scanning approach, including laminae and transverse process (TP) as the landmarks, to assess spine curvature has been demonstrated in in vitro experiments [21–23]. A number of 3-D ultrasound systems have been advanced to overcome the limitations of 2-D viewing of 3-D anatomy [22–31]. Purnama et al [29] showed that imaging of the human spine using freehand 3-D ultrasound was feasible, but no measurement of spine curvature angle was conducted.

    View all citing articles on Scopus
    View full text