Introduction

Clear corneal incision phacoemulsification is the preferred practice for cataract removal in the present scenario.1 Over the past few decades, there is a trend toward reduction of incision size.2, 3 Smaller incisions offer faster rehabilitation, lesser astigmatism, lesser inflammation, and fewer chances of wound leak and post-operative endophthalmitis.4, 5, 6, 7

Autosert (Alcon Laboratories, Inc., Fort Worth, TX, USA) is a new motorized intraocular lens (IOL) injector system that has been launched recently in India. There are safety reports available with the use of Autosert for IOL implantation.8 It is easy to use and was found to cause less incision enlargement during IOL implantation by Allen et al8 We compared this device with the manual Monarch injector, provided by the same manufacturer, with respect to IOL safety, incision size, and wound integrity using Anterior Segment Optical Coherence Tomography (ASOCT; Visante; Carl Zeiss Meditec, Inc., Dublin, CA, USA).

Materials and methods

This was a prospective, randomized study conducted at our tertiary care centre. Patients were randomized using a list of computer-generated random numbers at the time of deciding the surgical date. Consecutive patients who had no ocular pathology other than cataract and planned for standard phacoemulsification surgery were recruited in the study. All preoperative investigations for cataract surgery were done by a single, experienced observer.

Informed consent was taken and patients received standard 2.2 mm phacoemulsification by a single surgeon and on a single machine. They were randomized to receive IOL implantation using manual Monarch injector (group I) or motorized Autosert injector (group II) with a D cartridge for hydrophobic acrylic aspheric intraocular lens (Acrysof SN60WF IOL, Alcon Laboratories, Inc.). Incision width was measured before and immediately after IOL implantation in both the groups using incision gauge (Appasamy incision Gauge model AA2102, Appasamy Associates, Arumbakkam, Chennai, India) (Figure 1). Patients in whom the incision size enlarged by more than 2.2 mm after irrigation–aspiration and before IOL was implanted were excluded from the study. Presence of any IOL nicks or scratches if observed during IOL insertion were noted and documented. ASOCT was done on first post-operative day and at 1 month by a randomization blind examiner.

Figure 1
figure 1

Incision gauges used for measuring incision size.

Surgical technique

All surgeries were performed by a single experienced surgeon (SK) in the same operating room for phacoemulsification. Following an aseptic technique, a clear corneal incision was made temporally using a 2.2-mm metal trapezoidal keratome (ClearCut slit knife, single bevel; Alcon Laboratories, Inc.) under topical anesthesia. A side port incision was created using a microvitreoretinal blade (Alcon Laboratories, Inc.). Cohesive viscoelastic was injected to fill the anterior chamber and circular continuous capsulorrhexis was performed using capsulotomy needle. Cortical cleaving hydrodissection was done and nucleus was emulsified using horizontal chop technique. Phacoemulsification in all cases was done using torsional phaco with the Infiniti Vision System (Alcon Laboratories, Inc.) with a 0.9-mm Kelman miniflare tip (Alcon Laboratories, Inc.) and 0.9 mm Ultrasleeve (Alcon Laboratories, Inc.). Irrigation–aspiration of cortical material was done using Intrepid silicone-sleeved co-axial system (Alcon Laboratories, Inc.). The incision width was measured after filling the anterior chamber and the bag with a cohesive viscoelastic. Patients with incision >2.2 mm before IOL implantation were excluded from the study. Hydrophobic acrylic aspheric intraocular lens (Acrysof SN60WF IOL) implantation was done using a D cartridge with a Monarch manual injector in one group and Autosert motorized injector in the other group. Proper IOL loading was done by the same experienced surgeon (SK) with the haptics folded over the optic. Wound-assisted IOL insertion was done in both groups. Autosert was used with an initial velocity of 2.5 mm/s followed by a pause of 0.5 s and then a final velocity of 2.5 mm/s. Final positioning of the IOL in bag was done using the plunger end. The incision width was again measured immediately after IOL implantation and recorded. Finally the viscoelastic was removed using irrigation–aspiration. The IOL status was checked for the presence of any nicks or scratches and final position was noted. Incisions were hydrated and checked for leakage after which antibiotics and steroids were instilled.

ASOCT was done for all the patients for main 2.2 mm incision using ‘Enhanced high resolution corneal protocol’ on post-operative day 1 and day 30 by a randomization blind investigator. Presence and extent of wound gape was noted for both the groups. Extent of gape or non-apposition of the incision was observed on ASOCT. Extent of descemet membrane detachment and posterior wound retraction, if any, were also assessed. Posterior wound retraction was defined as a step-off or recession of the limbal edge of the posterior wound surface.9

Statistical analysis was done using SPSS software version 15.0 (SPSS, Inc., Chicago, IL, USA) and P<0.05 was considered significant.

Results

Sixty-two patients were included in the study, 32 in the group with manual injector (group I) and 30 in the group with Autosert motorized injector (group II). The difference between the mean age and mean IOL power in either group was comparable (Table 1). On assessing IOL safety, small nicks were seen in three (9.37%) cases in IOL periphery in group I but none were seen in group II (Figure 2). Only those patients who had incision size 2.2 mm after irrigation–aspiration were considered. In group I, final incision after IOL insertion increased to 2.3 mm in 27 eyes and in the remaining 5 eyes it increased to 2.4 mm, while in Autosert group, 28 of the incisions remained at 2.2 mm after insertion and only 2 increased to 2.3 mm. (P<0.0001; Fisher’s Exact test; Table 2).

Table 1 Comparison of age and intraocular lens power between manual injector group and motorized injector group
Figure 2
figure 2

Arrow showing IOL nick in periphery in a patient in which IOL implantation was done using manual injector.

Table 2 Comparison of incision size, IOL nicks, and wound characteristics on ASOCT between manual injector group and motorized injector group

Localized descemet membrane detachment was seen in 14 (43.75%) eyes in manual group and in 7 (23.33%) eyes in group II (P=0.09) using ASOCT on post-operative day 1 (Figures 3a and b; Table 2). However, on follow up at 1 month, none of the patients had descemet detachment. No anterior wound gape was seen in either of the groups. Posterior wound gape was seen in 16 (50%) patients in manual group and 6 (20%) in the motorized group at day 1 (P=0.018); however, none of the patients had evidence of gape at 1 month follow-up (Figures 3d and e). None of the patients had posterior wound gape, which extended in size by more than one-fourth of the total incision length. Posterior wound retraction was seen in only one patient in the manual group at 1 month (Figure 3c; Table 2).

Figure 3
figure 3

(a) ASOCT image showing descemet membrane detachment in a patient in the manual injector group on day 1. (b) ASOCT image showing descemet membrane detachment in a patient in the motorized injector group on day 1. (c) ASOCT image showing posterior wound retraction in a patient in the manual injector group at 1 month. (d) ASOCT image showing posterior gape in a patient in the manual injector group on day 1. (e) ASOCT image showing posterior gape in a patient in the motorized injector group on day 1. Note that the incision architecture is arcuate thus imparting greater wound strength.

Discussion

There is an increasing trend toward smaller incision size in phacoemulsification along with greater preference for square-edged intraocular lenses to decrease the incidence of posterior capsular opacification. This requires lenses to be inserted through smaller incisions and smaller bore cartridges. For IOL implantation through incisions ≤2.2 mm, injector systems are safer as compared with forceps. However an ideal injector system still eludes the medical fraternity. With available reports on Autosert,8 we tried to find the IOL safety, changes in incision size after the IOL insertion, and wound integrity on ASOCT comparing manual and motorized injector systems.

Phacoemulsification with IOL implantation may lead to an increase in original incision size occurring at various steps.10, 11 Maximum incision stretch has been shown to occur during IOL implantation.11 Steinert and Deacon12 reported that an irreversible increase in incision size follows phacoemulsification surgery due to initial stretch-related incision tearing. Various factors have been seen to affect the amount of incision enlargement, including dioptric power of IOL,13 type of IOL,14 type of keratome,15 and speed of IOL insertion.16 Contrarily, Kohnen et al found that IOL power, material, and shape of cartridge do not affect incision size.17 According to them, width of injector is the only factor which affects incision enlargement significantly. However, most of these studies have been done on incisions of size >2.5 mm.12, 13, 17 In our study, we compared incision enlargement with 2.2 mm incision and found the incision to enlarge in 100% of the cases, when using D cartridge with manual injector. Our results approximate that of Osher18 who recorded 100% enlargement with 2.2-mm incision size. However they used a C cartridge and hence, the enlargement was not due to IOL implantation alone. Similarly, Espiritu and Bernardo11 used 2.2 mm incision with D cartridge using Monarch III manual injector and found enlargement in 93.5% of the eyes. The mean enlargement in incision size after cortical aspiration to after IOL implantation in their study was 0.046 mm, which is a little less than ours (0.12 mm). We also compared the difference in final incision sizes with D cartridge and Autosert and found lesser incision enlargement with Autosert, which was statistically significant; results similar to those reported by Allen et al.8 Forcing implantation of IOL through incision of smaller than adequate size may lead to corneal damage manifesting as descemet’s tears, irregularities, distortions, and improper wound healing. The ideal incision size for use of D cartridge has been observed by Espiritu and Bernardo11 to be 2.3 mm with wound-assisted implantation of IOL. Our study also favors this observation as in 84.4% of patients the incision enlarged to 2.3 mm with the same technique. In addition, we found no enlargement of incision in cases with Autosert-assisted implantation in 93.33% of the eyes. Thus 2.2 mm is an adequate-sized incision for IOL implantation using Autosert, when used in conjunction with D cartridge.

We have also found IOL nicks in the periphery in group I, when using D cartridge. This has not been reported previously in literature. As both hands are engaged when using manual injector and the eye usually rolls nasally during insertion through temporal incision under topical anesthesia, this causes more force to be applied on the wound as well as during the final plunge of the IOL. This can sometimes cause over-riding of the plunger onto the folded IOL inside the cartridge resulting in IOL nicks and associated wound distortion.

ASOCT is a useful investigative modality for determination of wound morphology and evolution over time when compared with clinical appreciation on slit lamp biomicroscopy. Though ASOCT-assisted in vivo evaluation of wound architecture in patients undergoing phacoemulsification has been done previously,19, 20, 21, 22, 23 we used ASOCT to compare manual vs the new motorized injector and to assess its safety, which has not been reported previously. No differences could be elicited between the two groups with respect to descemet’s membrane detachment, anterior wound gape, and posterior gape at 1 month. None of our eyes developed epithelial gape. This may be because a two-step clear corneal incision is arcuate in configuration (Figures 3a–e), rather than linear, and hence is more stable.19 Lyles et al20 similarly reported preservation of epithelial integrity on first post-operative day in their series. The presence of transient descemet membrane detachment and endothelial gape was elicited on first post-operative day in both groups. Torres et al21 reported similar wound distortion using a 3.2 mm incision with a manual injector. On serial ASOCT examination at 1 month, all 22 eyes showed resolution of posterior wound gape, a trend similar to that seen by Chee et al23 and Dupont-Monod et al.22 They also reported that the posterior gape did not exceed by >25% of total incision length in their series, an observation similar to ours. Posterior wound gape has been shown to correlate to greater contrast enhancement at wound site, especially in the first 24 h, thus predisposing these eyes for endophthalmitis.24 Thus, Autosert may provide safer wound architecture in the immediate post-operative period, a time most crucial for development of endophthalmitis, in comparison to manual IOL injection.

Though a small sample size and the use of incision gauges with minimum calibration up to 0.1 mm resolution comprises our limitations, an in vivo comparison of wound architecture between the two groups, which has not been done previously, brings forth novel observations. We conclude that use of D cartridge with Autosert motorized injector is safe as far as wound morphology and IOL insertion is concerned; it causes lesser incision enlargement, fewer IOL nicks, and is easier to use.