Preterm infants suffering from respiratory distress syndrome have a reduced end-expiratory lung volume (EELV) and lung compliance, leading to impaired ventilation and oxygenation (1). A substantial proportion of these preterm infants needs to be intubated and mechanically ventilated (2). Despite the beneficial effect on lung function and gas exchange, mechanical ventilation also has a downside as it can lead to secondary lung injury and subsequent bronchopulmonary dysplasia (3,4). For this reason, preterm infants are extubated as soon as possible and transferred back to noninvasive respiratory support, predominantly nasal continuous positive airway pressure (nCPAP). To make these transitions successful, infants should be able to maintain an optimal EELV and normal ventilation during nCPAP. Placing infants in prone position might facilitate this transition to nCPAP as this posture seems to increase EELV and improve respiratory mechanics (5,6,7,8).

Despite the fact that EELV might play an important role in successful transition from invasive to noninvasive respiratory support, the changes in lung volume following extubation have been poorly investigated. Only two studies have assessed changes in functional residual capacity before and after extubation, but these changes were not measured continuously and the infants were extubated to a head box and not nCPAP (9,10). Moreover, the role of positioning was not addressed in these studies.

The fact that lung volume changes have been poorly studied in preterm infants is mainly caused by the lack of easily applicable, inexpensive, methods to measure lung volume continuously at bedside, which do not alter respiratory mechanics or increase dead space in this vulnerable population. Electrical impedance tomography (EIT) is a relatively new, noninvasive technique, which measures changes in lung electrical impedance in a cross-sectional slice of the chest that correlate well with changes in aeration (11). Furthermore, EIT also provides information on regional changes in volume and ventilation (12,13,14). Recent studies have shown that EIT is feasible in the preterm population and that the lung volume changes measures in the examined slice are representative for the whole lung (15,16,17).

We have therefore conducted an observational study using EIT to gain more insight into the (regional) changes in EELV and ventilation following transition from invasive to non-invasive respiratory support in premature infants with respiratory distress syndrome. Furthermore, we have assessed the additional effect of prone position shortly after extubation on these parameters.

Results

We included a convenience sample of 20 infants with a mean gestational age of 28.7 ± 1.7 wk and birth weight of 1,125 ± 388 g. Patient characteristics are shown in Table 1 . Two patients did not tolerate high-frequency oscillatory ventilation (HFOV) without oscillations and in two patients the electrodes detached when turning them from supine to prone positioning, leaving only 18 measurements at these time points. The continuous distending pressure (CDP) before extubation, after extubation, and after switching to prone position was, respectively, 7.9 ± 0.5, 6.0 ± 0.2, and 6.0 ± 0.2 cmH2O. The FiO2 remained stable at these three time points ( Table 2 ).

Table 1 Patient characteristics
Table 2 Ventilatory parameters

Compared to ET-CPAP, EIT showed a significant increase in EELV (median 3.8 AU/kg, interquartile range: −0.8–27.4 AU/kg, P < 0.05) after extubation to nCPAP in supine position. The increase in EELV was located mostly in the dorsal lung regions ( Figure 1 ). In all patients, the EELV stabilized within the 10-min measurement period. Both respiratory rate and VT increased after extubation, but only the latter reached statistical significance. In response to these changes, TcPCO2 showed a modest, nonsignificant reduction and SpO2 a modest but significant increase ( Table 2 ). Ventilation distribution did not change after extubation ( Table 2 ; Figure 2 ).

Figure 1
figure 1

Changes in end-expiratory lung volume, global (solid line, closed circles), ventral (dashed line, open diamonds), and dorsal (dotted line, open triangles). Data are presented as median with interquartile range.

PowerPoint slide

Figure 2
figure 2

Percentage of tidal impedance change normalized to body weight in 32 anterior-to-posterior chest slices at different study phases; endotracheal continuous positive airway pressure (CPAP) (closed squares) in supine position, nasal CPAP in supine position (open circles) and nCPAP in prone position (open triangles). The slices were denoted by setting the most anterior part to 0% and the most posterior part to 100%. Data are presented as mean values.

PowerPoint slide

After changing to prone position, EELV significantly increased compared to nCPAP in the supine position (median: 36.0, interquartile range: 23.7–63.3 AU/kg, P < 0.01) ( Figure 1 ). Again, the increase in EELV favored the dorsal lung regions, although there was also an increase observed in the ventral regions ( Figure 1 ). The positional change resulted in a significant decrease in respiratory rate compared to supine position with no change in VT and TcPCO2 ( Table 2 ). SpO2 remained stable after switching to prone position. Ventilation significantly increased in the ventral regions as indicated by significant changes in area under the curve and geometrical center of ventilation ( Table 2 ).

Discussion

This study shows that preterm infants are able to maintain their EELV when transitioned from endotracheal to nasal CPAP. Placing these infants in prone position results in an increase in EELV and a ventral shift in tidal ventilation distribution.

To our knowledge, this is the first study that has continuously monitored changes in EELV and ventilation during transition from invasive to noninvasive respiratory support. To avoid any interference of mechanical ventilation with these parameters, we stopped the oscillations during HFOV, making extubation a true transition from endotracheal to nasal CPAP. According to our local protocol, infants were extubated to a nasal CDP 2 cmH2O lower than during HFOV, because it has been suggested that the alveolar CDP delivered by the SensorMedics 3100 ventilator is 1–2 cmH2O lower than the pressure recorded at the hub of the endotracheal tube when delivered with an inspiratory time of 33%, as is custom in our unit (18,19,20). Despite this lower CDP EELV remained stable after extubation. This finding is probably best explained by the fact that the infant regains control of the upper airways and laryngeal regulation after removal of the endotracheal tube. Previous studies have shown that the upper airways play an important role in maintaining functional residual capacity in infants (21,22). The fact that tidal volumes also increased after extubation may also have contributed to the stability of the EELV.

EIT also allowed us to assess the regional changes in EELV following extubation. Although the changes in EELV were relatively small, it was interesting to observe that the dorsal (dependent) but not the ventral (nondependent) lung regions showed a modest increase in EELV. We can only speculate on possible explanations for this finding. First, spontaneous (negative pressure) breathing with control of the upper airways may have a different effect on EELV in the dependent lung regions compared to breathing through an endotracheal tube supported by positive airway pressures. Second, the absence of endotracheally delivered positive airway pressure following extubation may have increased venous return to the heart. This in turn may have limited an increase in lung aeration and thus lung impedance in the ventral part of the chest, as blood has a five times lower impedance than lung tissue at expiration (23).

Many infants on noninvasive support are placed in the prone position as previous studies have demonstrated that this reduces thoraco-abdominal asynchrony, improves oxygenation and increases functional residual capacity (5,7,24). Our study also shows an increase in EELV when switching from supine to prone position but adds important information on its regional distributions. It shows that the increase in EELV is most prominent in the dorsal regions (25), which may imply a gravitational effect resulting in an increased aeration in these now nondependent regions. Although this cannot be substantiated by the present study, this effect on aeration in the dorsal regions may be accompanied by a reduction in blood and fluid content.

In addition to changes in EELV, we have also assessed the effect of extubation and body positioning on oxygenation and ventilation. The changes in EELV following extubation and prone positioning only had a small impact on oxygenation, suggesting that the changes in EELV were mainly caused by distension and not recruitment. The low FiO2 also seems to support this assumption. Removal of the endotracheal tube resulted in a significant increase in VT, which is probably best explained by the reduction in airway resistance (26,27). This increase in VT resulted in only a small, nonsignificant, reduction in TcPCO2. This indicates that airway dead space may have also increased after extubation due to the slight increase in EELV. Placing the infant in prone position did not change tidal volume but did result in a significant reduction in respiratory rate compared to supine position. As TcPCO2 remained stable this suggests more efficient breathing, probably caused by a reduction in thoraco-abdominal asynchrony (7).

In the present study, we have also determined the distribution of tidal ventilation. The paradigm that ventilation distribution in infants follows a reversed pattern compared to adults with a preferable distribution to the nondependent (uppermost) lung has recently been challenged (28,29,30,31). It has been suggested that this pattern only applies to sedated and mechanically ventilated infants. In spontaneously breathing infants, ventilation is distributed much more evenly throughout the lung (28,29). Our study showed that application of CDP via the endotracheal or nasal route did not impact ventilation distribution in these spontaneously breathing preterm infants. Comparable to other studies (28,30), ventilation slightly favored the dorsal lung regions. However, and in contrast to a previous study (30), placing the infants in prone position resulted in a clear shift in ventilation distribution to the ventral, dependent lung regions. It is unclear why our findings are different from these previous studies. It might be that the lung behaves different to positional change in the first hours after extubation to nCPAP. A previous report in a term born infant seems to support this hypothesis as ventilation distribution clearly shifted to the ventral region after placing this infant in prone position 1 h after extubation (15). Whether this shift in ventilation distribution to the ventral lung regions is consistent over time cannot be answered by this study.

Our study has several limitations that need to be addressed. First, the positioning of the infants during and after extubation was not randomized. Extubation in our unit is routinely performed in supine positioning as this provides the best access for airway management and placement of nCPAP. For this reason, the nursing staff was not comfortable extubating infants in prone position. In theory, the increase in EELV following prone positioning may have been an ongoing improvement after extubation, not related to the positional change. The fact that the EELV stabilized within the 10-min measurement period after extubation and the increase in EELV is consistent with previous studies makes this very unlikely. Second, this study only addressed the immediate changes in EELV and ventilation, using 10-min measurement periods. Whether these effects are maintained over a longer period of time needs further study. Third, this study only included patients with respiratory distress syndrome, extubated in the first 72 h after birth. The results might differ when dealing with other causes of respiratory failure or older infants. Finally, this study does not offer insight into absolute lung volume changes as it is not yet possible to calibrate the impedance changes during HFOV. However, this does not compromise the results of this study as we were interested in relative and not absolute changes in EELV and ventilation.

In conclusion, this study shows that preterm infants are able to maintain their EELV when transitioned from invasive to noninvasive respiratory support. Furthermore, it reveals that prone positioning increases EELV and provides more efficient ventilation, making this the preferred position following extubation in preterm infants.

Methods

The study was performed in the neonatal intensive care unit of the Emma Children’s Hospital, Academic Medical Center (Amsterdam, The Netherlands), where preterm infants (< 37 wk) with a suspected diagnosis of respiratory distress syndrome and failing nCPAP are treated with open lung HFOV. All patients are ventilated in the supine position and are not sedated or paralyzed. Infants were included in the study if they were eligible for extubation according to our local protocol, i.e., a CDP ≤ 8 cmH2O and a fraction of inspired oxygen (FiO2) ≤ 0.30. Exclusion criteria were congenital anomalies of the chest and/or abdomen and a fragile skin condition. Written informed consent was obtained from both parents and the study was approved by the Institutional Review Board of the Academic Medical Center, Amsterdam, The Netherlands.

Patient Data Collection

Patient characteristics including gestational age, birth weight, postnatal age, and Apgar scores were collected. Together with the EIT recordings, FiO2, heart rate, transcutaneous oxygen saturation (SpO2), and transcutaneous partial pressure of carbon dioxide (TcPCO2) were registered throughout the experiment.

Study Protocol

The study consisted of two phases during which impedance changes were continuously measured with EIT. During the first phase, within 10 min before extubation, patients were switched from HFOV (SensorMedics 3100, CareFusion, Yorba Linda, CA) to endotracheal CPAP (ET-CPAP) for 2 min prior to extubation by stopping the oscillations and allowing for spontaneous breathing. Next, the patients were extubated to nCPAP, delivered by the Infant Flow SiPAP system (CareFusion) with a CDP of 6 cmH2O. Following extubation, the impedance changes were registered for 10 min in supine position.

During the second phase, initiated 10 min after extubation, the infants were transferred from supine to prone position. Lung volume and ventilation changes were again measured for 10 min using EIT.

EIT Data Collection

Before extubation, 16 hand-trimmed electrocardiography electrodes (Blue Sensor, BRS-50-K, Ambu, Linthicum Heights, MD) were placed equidistantly on the thorax circumference, just above the nipple line and connected to the Goettingen Goe-MF II EIT system (CareFusion). Repetitive electrical currents (5 mA, 100 kHz) were injected cyclically (scan rate: 44 Hz) through adjacent electrode pairs. Voltage changes were registered between all other passive adjacent electrode pairs. A back-projection algorithm generated a series of 32 × 32 pixel matrices of relative impedance changes (ΔZ) by comparing these voltage changes to a reference period. Impedance changes were registered using the Veit software (CareFusion). Data analysis was performed off-line with AUSPEX version 1.6 (VUmc, Amsterdam, The Netherlands) after all data collection was completed.

EIT Data Analysis

As EIT cannot measure absolute changes in lung volumes, the change in EELV (ΔEELV) was calculated by selecting a stable 30-s period at the start of the study protocol, which was used as reference for the subsequent recordings. Next, the relative changes in EELV were calculated for each step, by averaging the relative ΔZ at the troughs of the spontaneous breaths in a stable 30-s recording period at the end of each step to allow for lung volume to reach equilibrium. ΔEELV was then normalized for body weight.

EIT-derived tidal volume (VT) changes were calculated for each examination phase by referencing the selected 30-s period to the average ΔZ in that period. The EIT waveforms of relative ΔZ values were then band-pass filtered in the band of spontaneous breathing rate (10/min below the actual breathing frequency and 10/min above its second harmonic). The peaks and troughs of the signal were identified and averaged over the selected period. The average amplitude was normalized for body weight.

Regional ventilation distribution of the spontaneous breaths was assessed by means of functional EIT (fEIT) images. fEIT images were generated by calculating the SD of the impedance waveforms in each individual pixel within the 32 × 32 pixel matrix. A ventilation profile was derived from each fEIT image showing the distribution of ventilation in 32 anteroposterior slices, as previously described (32). Subsequently, the areas under the curve (AUC) for the ventral or anterior (slice 1-16, AUCant) and dorsal or posterior (slice 17-32) regions were calculated. The AUCant describes the proportion of the total tidal impedance changes in the anterior lung regions. An AUCant less than 50% indicates a posterior oriented and more than 50% an anterior oriented distribution of tidal ventilation. Finally, the geometrical center of ventilation was determined as an additional measure describing the ventilation distribution (32,33). The geometrical center of ventilation represents the level in the anterior–posterior axis (0–100%) where the AUC of tidal impedance changes is identical in the lung regions above and below this level. A geometrical center of ventilation more than 50% indicates a posterior oriented and less than 50% an anterior oriented distribution of tidal ventilation.

Statistical Analysis

Statistical analysis was performed using GraphPad Prism 5.0 (GraphPad Software, San Diego, CA) and SPSS version 20.0 (SPSS, Chicago, IL). Data were expressed as mean with SD or median with interquartile range, as appropriate. Comparative analysis was performed using the Wilcoxon signed rank test for skewed data and the paired student t-test for normally distributed data.

Disclosure

This work was financially supported by Chiesi Pharmaceutical, Parma, Italy. The EIT device was kindly provided by CareFusion, Yorba Linda, CA.