J Neurol Surg A Cent Eur Neurosurg 2013; 74(05): 321-324
DOI: 10.1055/s-0032-1315790
Technical Note
Georg Thieme Verlag KG Stuttgart · New York

Frameless Navigated Biopsy with the BrainLAB® VarioGuide System: A Technical Note

Rüdiger Buchalla
1   DIAKO Flensburg, Neurochirurgische Klinik, Flensburg, Germany
,
Silke Hopf-Jensen
2   DIAKO Flensburg, Radiologisches Institut, Flensburg, Germany
,
Ortwin Rubarth
1   DIAKO Flensburg, Neurochirurgische Klinik, Flensburg, Germany
,
Wolfgang Börm
1   DIAKO Flensburg, Neurochirurgische Klinik, Flensburg, Germany
› Author Affiliations
Further Information

Publication History

09 March 2012

24 March 2012

Publication Date:
07 July 2012 (online)

Abstract

Objective Implementation of a new neuronavigation system in the department included the opportunity to change the brain biopsy procedures from a frame-based system to a frameless system—the BrainLAB® VarioGuide. We report our first 20 consecutive procedures within 2 years, analyzing the effectiveness of this new tool.

Patients, Materials, and Methods 15 men and 5 women with various intracerebral lesions were biopsied. Their average age was 62 years (range, 41–83 years). Preoperative planning was performed using computed tomography (CT) or magnetic resonance imaging (MRI) data provided by a radiological picture archiving and communication system (PACS) sent to the planning station. Planning was based on either MRI or CT in 14 and 4 cases, respectively, and a combination of both in 2 cases.

Results The lesions had a mean maximal diameter of 42.97 mm (range, 5.8–109.8 mm). 19 lesions were supratentorial, 1 of those intraventricular, and 1 lesion was infratentorial, within the cerebellar peduncle and brainstem. During surgery, the samples were sent to neuropathological examination, and the procedure was completed after confirmation that pathological tissue was obtained. In 18 of 20 cases, definitive pathological tissue was harvested. Histopathology confirmed 8 glioblastomas, 1 oligoastrocytoma, 4 anaplastic astrocytomas, 1 astrocytoma, and 2 primary central nervous system lymphomas, as well as 1 encephalitis and 1 necrotic tissue of unknown origin. In 3 cases, including the case of necrotic tissue, there was no final diagnosis possible. Referencing the BrainLAB® navigation system and software-wizard guided sterile adjustment of the VarioGuide arm over the entry point took 5 to 10 minutes. The mean operation time was 49 minutes (range, 24–69 minutes) including the time until first neuropathological examination was completed. No adverse events occurred during or after the operation.

 
  • References

  • 1 Barnett GH, Miller DW, Weisenberger J. Frameless stereotaxy with scalp-applied fiducial markers for brain biopsy procedures: experience in 218 cases. J Neurosurg 1999; 91 (4) 569-576
  • 2 Gralla J, Nimsky C, Buchfelder M, Fahlbusch R, Ganslandt O. Frameless stereotactic brain biopsy procedures using the Stealth Station: indications, accuracy and results. Zentralbl Neurochir 2003; 64 (4) 166-170
  • 3 Ringel F, Ingerl D, Ott S, Meyer B. VarioGuide: a new frameless image-guided stereotactic system—accuracy study and clinical assessment. Neurosurgery 2009; 64 (5, Suppl 2) 365-371 , discussion 371–373
  • 4 Paleologos TS, Dorward NL, Wadley JP, Thomas DG. Clinical validation of true frameless stereotactic biopsy: analysis of the first 125 consecutive cases. Neurosurgery 2001; 49 (4) 830-835 , discussion 835–837
  • 5 Woodworth GF, McGirt MJ, Samdani A, Garonzik I, Olivi A, Weingart JD. Frameless image-guided stereotactic brain biopsy procedure: diagnostic yield, surgical morbidity, and comparison with the frame-based technique. J Neurosurg 2006; 104 (2) 233-237
  • 6 Air EL, Leach JL, Warnick RE, McPherson CM. Comparing the risks of frameless stereotactic biopsy in eloquent and noneloquent regions of the brain: a retrospective review of 284 cases. J Neurosurg 2009; 111 (4) 820-824
  • 7 Dammers R, Haitsma IK, Schouten JW, Kros JM, Avezaat CJ, Vincent AJ. Safety and efficacy of frameless and frame-based intracranial biopsy techniques. Acta Neurochir (Wien) 2008; 150 (1) 23-29
  • 8 Widmann G, Eisner W, Kovacs P , et al. Accuracy and clinical use of a novel aiming device for frameless stereotactic brain biopsy. Minim Invasive Neurosurg 2008; 51 (6) 361-369
  • 9 Shooman D, Belli A, Grundy PL. Image-guided frameless stereotactic biopsy without intraoperative neuropathological examination. J Neurosurg 2010; 113 (2) 170-178