Telemetric Intracranial Pressure Recording via a Shunt System Integrated Sensor: A Safety and Feasibility Study

Philipp Ertl
1   Department of Neurosurgery, Medical School Hannover, Hannover, Germany
,
Elvis J. Hermann
1   Department of Neurosurgery, Medical School Hannover, Hannover, Germany
,
Hans E. Heissler
1   Department of Neurosurgery, Medical School Hannover, Hannover, Germany
,
Joachim K. Krauss
1   Department of Neurosurgery, Medical School Hannover, Hannover, Germany
› Author Affiliations
Further Information

Publication History

29 June 2016

06 April 2017

Publication Date:
06 June 2017 (online)

Abstract

Background The dynamics of intracranial pressure (ICP) after shunt surgery in patients with normal pressure hydrocephalus (NPH) are poorly known. Temporarily implanted parenchymal pressure probes are reliable for accurate ICP monitoring; however, a disadvantage of this method is that the ICP probe has to be explanted after a period of time, requiring additional surgery. We present two patients with NPH with an integrated ICP measuring device in the shunt system that allows for long-term postoperative telemetric monitoring of ICP.

Methods Two patients (one man, 66 years old, and one woman, 78 years old) with normal pressure hydrocephalus underwent shunt surgery using the Aesculap-Miethke Sensor Reservoir (Potsdam) as an ICP measuring device integrated in the shunt system. On the first postoperative day and during follow-up examination 3 months later, several measurements of ICP were performed using a handheld device to read the Sensor Reservoir data. Postural changes such as lying down, sitting, and standing with different head postures were assessed according to an experimental protocol in a randomized order during which the ICP was measured.

Results There were clear ICP responses due to postural changes, in line with physiologically expected values. Because the highest sampling rate of the Aesculap-Miethke Sensor Reservoir is 1 Hz for continuous measurement of ICP, however, the collected data have to be considered an approximation for actual ICP dynamics.

Conclusion The Aesculap-Miethke Sensor Reservoir is an easy-to-use tool to measure ICP changes reliably in patients with a shunt system. The sampling rate and handling of data acquisition may eventually be developed further.

 
  • References

  • 1 Lilja A, Andresen M, Hadi A, Christoffersen D, Juhler M. Clinical experience with telemetric intracranial pressure monitoring in a Danish neurosurgical center. Clin Neurol Neurosurg 2014; 120: 36-40
  • 2 Welschehold S, Schmalhausen E, Dodier P. , et al. First clinical results with a new telemetric intracranial pressure-monitoring system. Neurosurgery 2012; 70 (1, Suppl Operative): 44-49 ; discussion 49
  • 3 Sprung C, Schlosser H-G, Lemcke J. , et al. The adjustable proGAV shunt: a prospective safety and reliability multicenter study. Neurosurgery 2010; 66 (03) 465-474
  • 4 Droste DW, Krauss JK. Intracranial pressure B-waves precede corresponding arterial blood pressure oscillations in patients with suspected normal pressure hydrocephalus. Neurol Res 1999; 21 (07) 627-630
  • 5 Droste DW, Krauss JK. Oscillations of cerebrospinal fluid pressure in nonhydrocephalic persons. Neurol Res 1997; 19 (02) 135-138
  • 6 Droste DW, Krauss JK, Berger W, Schuler E, Brown MM. Rhythmic oscillations with a wavelength of 0.5–2 min in transcranial Doppler recordings. Acta Neurol Scand 1994; 90 (02) 99-104
  • 7 Andresen M, Hadi A, Petersen LG, Juhler M. Effect of postural changes on ICP in healthy and ill subjects. Acta Neurochir (Wien) 2015; 157 (01) 109-113
  • 8 Krauss JK, Droste DW. Predictability of intracranial pressure oscillations in patients with suspected normal pressure hydrocephalus by transcranial Doppler ultrasound. Neurol Res 1994; 16 (05) 398-402
  • 9 Gaab MR, Heissler HE, Ehrhardt K. Physical characteristics of various methods for measuring ICP. In: Hoff J, Betz AL. , eds. Intracranial Pressure VII. Berlin, Germany: Springer; 1989: 16-21
  • 10 Gaab MR, Heissler HE. ICP monitoring. Crit Rev Biomed Eng 1984; 11 (03) 189-250
  • 11 Freimann FB, Schulz M, Haberl H, Thomale UW. Feasibility of telemetric ICP-guided valve adjustments for complex shunt therapy. Childs Nerv Syst 2014; 30 (04) 689-697
  • 12 Maeske M, Mayer S, Blanc SM, Schulz C, Kunz U, Mauer UM. Telemetric intracranial pressure measurement: a graphical approach to data analysis. J Neurol Surg A Cent Eur Neurosurg 2016; 77 (03) 258-263