Entwicklung eines 3-dimensionalen Messverfahrens zur Bestimmung des tibialen Gefälles im Spiral-CT

Development of a 3-Dimensional Method to Determine the Tibial Slope with Multislice-CTM.  A.  Kessler1 , A.  Burkart1 , V.  Martinek1 , A.  Beer2 , A.  B.  Imhoff1
  • 1Abteilung und Poliklinik für Sportorthopädie der Technischen Universität München, Leiter: Prof. A. B. Imhoff
  • 2Institut für Röntgendiagnostik, TU München, Klinikum rechts der Isar, Leiter: Prof. E. J. Rummeny
Further Information

Publication History

Publication Date:
15 April 2003 (online)

Zusammenfassung

Studienziel: Die Vermessung und ggf. operative Korrektur des Slopes hat in der Knieendoprothetik starke Bedeutung und findet jetzt auch in der Kreuzbandchirurgie zunehmendes Interesse. Die valide Messung in vivo mittels konventionellem Röntgen ist dabei jedoch nur mäßig reproduzierbar. Ziel unserer Studie war es, den Messfehler in der konventionellen Röntgentechnik zu quantifizieren und mit unserem neu entwickelten Verfahren (MPR-CT) zu vergleichen. Methode: Bei 6 humanen Tibiapräparaten wurde der Slope mittels konventionellem Röntgen (bei 0°, 2.5°, 5°, 10°, 15°, 20°, 30° IRO) und Spiral CT (3D-MPR-Rekonstruktion) bestimmt und miteinander verglichen. Ergebnisse: Die Streubreite der Messung lag im konventionellen Röntgen selbst bei optimaler Einstellung im seitlichen Strahlengang (0° Rotation) bei > 5°. Mit zunehmender Rotation steigt der Messfehler bis auf 14° an. Mit dem neuen Verfahren (MPR-CT) konnte ein Messfehler von < 3° erzielt werden. Schlussfolgerung: Untersuchungen in der konventionellen Röntgentechnik zeigen eine hohe Streubreite v. a. bei Rotationsfehlstellung in Bezug auf den seitlichen Strahlengang. Die Spiral-CT hat eine hohe Genauigkeit und eliminiert methodisch den Rotationsfehler. Die Spiral-CT ist eine präzise Messmethode für die In-vivo-Bestimmung des Slopes mit einem Messfehler von < 3°. Für detaillierte Fragestellungen ist die konventionelle Röntgentechnik mit einem Messfehler von > 5° ungeeignet.

Abstract

Aim: The measurement or, respectively, the correction of the tibial slope is an important subject in the field of knee arthroplasty and in the procedure of cruciate ligament reconstruction. However, a valid value of the tibial slope cannot be obtained from the conventional plain X-rays in a reproducible way. The aim of this study was to evaluate the proximal tibial slope with a new CT measurement system and to compare the values with the X-ray method. Method: Using 6 cadaver tibiae, the antero-posterior slope of the proximal tibia was measured at 0°, 2.5°, 5°, 10°, 15°, 20°, 30° internal rotation using the plain X-ray and spiral CT (3D-MPR reconstruction). Both methods were compared. Results: The variation of the measurement of the tibial slope with the X-ray method was > 5° even if the optimal 0° rotation in the lateral view was applied. Moreover, the measurement error increased to 14° while increasing the rotation of the tibia. Using the new CT system, the error was less than 3° in all measurements. Conclusion: The measurement of the tibial slope in the conventional X-ray technique showed a high variation of the values depending on the rotation of the tibia in the lateral view. In contrast, the measurements with the new CT system represented a precise method with a small variation of the tibial slope values. For this reason detailed questions regarding the precise anatomy of the proximal tibia cannot be answered precisely with plain X-rays.

Literatur

  • 1 Bai B. et al . Effect of posterior cut angle on tibial component loading.  J Arthroplasty. 2000;  15 (7) 916-920
  • 2 Migaud H. et al . Influence of the tibial slope on tibial translation and mobility of non-constrained total knee prosthesis.  Rev Chir Orthop Reparatrice Appar Mot. 1996;  82 (1) 7-13
  • 3 Singerman R. et al . Decreased posterior tibial slope increases strain in the posterior cruciate ligament following total knee arthroplasty.  J Arthroplasty. 1996;  11 (1) 99-103
  • 4 Waelchli B, Romero J. Dislocation of the polyethylene inlay due to anterior tibial slope in revision total knee arthroplasty.  Knee Surg Sports Traumatol Arthrosc. 2001;  9 (5) 296-298
  • 5 Caylor K B. et al . Intra- and interobserver measurement variability of tibial plateau slope from lateral radiographs in dogs.  J Am Anim Hosp Assoc. 2001;  37 (3) 263-268
  • 6 Whiteside L A, Amador D D. The effect of posterior tibial slope on knee stability after Ortholoc total knee arthroplasty.  J Arthroplasty. 1988;  3 (Suppl) S51-S57
  • 7 Meister K. et al . Caudal slope of the tibia and its relationship to noncontact injuries to the ACL.  Am J Knee Surg. 1998;  11 (4) 217-219
  • 8 Imhoff A B. Stellenwert der Osteotomie bei der posterioren/posteriolateralen Knieinstabilität. in 17. Kongress der Arbeitsgemeinschaft für Arthroskopie. 5. - 7.10.2000. Basel
  • 9 Brazier J. et al . Evaluation of methods for radiographic measurement of the tibial slope. A study of 83 healthy knees.  Rev Chir Orthop Reparatrice Appar Mot. 1996;  82 (3) 195-200
  • 10 Genin P, Weill G, Julliard R. The tibial slope. Proposal for a measurement method.  J Radiol. 1993;  74 (1) 27-33
  • 11 Julliard R. et al . The median functional slope of the tibia. Principle. Technique of measurement. Value. Interest.  Rev Chir Orthop Reparatrice Appar Mot. 1993;  79 (8) 625-634
  • 12 Imhoff A B. HTO und VKB/HKB - Stellenwert des tibialen Slope. in ASG Fortbildungskurse der DGOOC, Abstraktband. 3. - 6.10.2001. Berlin

Prof. Dr. med. A. B. Imhoff

Abteilung und Poliklinik für Sportorthopädie der Technischen Universität München

Connollystr. 32

80809 München

Phone: 089-2892 4462

Fax: 089-2892 4484

Email: info@markus-kessler.de

    >