Comparison of Different Cardiac MRI Sequences at 1.5 T/3.0 T with Respect to Signal-to-Noise and Contrast-to-Noise Ratios - Initial Experience[*]

Vergleich verschiedener kardialer MRT-Sequenzen bei 1.5 T/3.0 T bezüglich des Signal- und Kontrast-zu-Rausch-Verhältnisses - Erste Erfahrungen[*] M. Gutberlet1 , B. Spors1 , M. Grothoff1 , P. Freyhardt1 , K. Schwinge1 , M. Plotkin1 , H. Amthauer1 , R. Noeske2 , R. Felix1
  • 1Klinik für Strahlenheilkunde, Bereiche Diagnostische Radiologie und Nuklearmedizin, Charité, Universitätsmedizin Berlin, Campus Virchow-Klinikum, Germany
  • 2GE Medical Systems, Berlin, Germany
We would like to thank Anne Gale, ELS of the Deutsches Herzzentrum Berlin for editorial assistance
Further Information

Publication History

Publication Date:
02 June 2004 (online)

Zusammenfassung

Ziel: Es sollte die Bildqualität anhand des Signal- (SNR) und des Kontrast-zu-Rausch-Verhältnisses (CNR) verschiedener MR-Sequenzen für die kardiale Bildgebung bei 1.5 T und 3.0 T an Probanden verglichen werden. Material und Methoden: Es wurden 10 Probanden (5 Männer/5 Frauen) ohne kardiale Vorerkrankungen in einem mittleren Alter von 33 Jahren (± 8) an einem GE Signa 3.0 T und einem GE Signa 1.5 T TwinSpeed Excite (GE Medical Systems, Milwaukee, WI, USA) jeweils mit Hilfe einer 4-Elemente-„phased-array”-Oberflächenspule (gleiches Design) am selben Tag an beiden Scannern untersucht. Als Sequenzen für die Gewebscharakterisierung wurden EKG-getriggerte Fast-Spinechosequenzen (FSE) mit T1- (Double-IR), T1-STIR- (Triple-IR) und T2-Wichtung in transversaler Schichtorientierung verwendet. Für die kardiale Funktionsanalyse erfolgte die Akquisition von „steady state free precession” (SSFP) FIESTA-Sequenzen im 4-Kammerblick, 2-Kammerblick lange und kurze Achse. Der Flipwinkel wurde bei 3.0 T von 45° auf 30° reduziert, um das TR nicht aufgrund des höheren SAR verlängern zu müssen. Alle anderen Sequenzparameter wurden konstant gehalten. Ergebnisse: Die Messungen konnten bei allen 10 Probanden erfolgreich durchgeführt werden. Das mittlere SNR bei 3.0 T im Vergleich zu 1.5 T konnte sowohl bei der T2- (160 % SNR-Steigerung), der STIR-T1 (123 %) als auch der T1-gewichteten (91 %) FSE-Sequenz signifikant (p < 0,05) gesteigert werden, ebenso wie das CNR. Das mittlere SNR der SSFP-Sequenzen lag um mehr als das Doppelte (150 %) über dem bei 1.5 T, allerdings ohne signifikanten Effekt auf das CNR. Bezüglich der Bildqualität zeigte sich nur bei den SSFP-Sequenzen ein signifikanter Unterschied mit schlechterer Bildqualität bei 3.0 T. Bei den Artefakten waren signifikante Unterschiede bei den FSE-Sequenzen beim Auftreten von „shading” im Bereich des rechten Ventrikels (RV) bemerkbar. Nach Anwendung eines lokalen Shims bei 5/10 Patienten war auch bei den SSFP-Sequenzen kein signifikanter Unterschied bezüglich des Auftretens von Artefakten nachweisbar. Schlussfolgerungen: Mit Hilfe von 3.0 T konnte das SNR und CNR signifikant bei allen kardialen FSE-Sequenzen für die Gewebsdifferenzierung ohne wesentliche Veränderungen der Sequenzparameter bei 3.0 T gesteigert werden. Auch die modifizierte SSFP-Sequenz führte zu der erwarteten Steigerung des SNR bei 3.0 T, aber ohne Veränderung des CNR. Der subjektive Bildeindruck hingegen verbesserte sich nicht signifikant und wurde bei den SSFP-Sequenzen und im Bereich des RV sogar signifikant schlechter. Die Ergebnisse sind trotzdem ermutigend für den Einsatz der kardialen MRT bei 3.0 T zur Gewebscharakterisierung, insbesondere in Kombination mit den neuen Möglichkeiten der parallelen Bildgebung.

Abstract

Purpose: To compare image quality, signal-to-noise (SNR) and contrast-to-noise ratios (CNR) of different MRI sequences for cardiac imaging at 1.5 T and 3.0 T in volunteers. Material and Methods: 10 volunteers (5male, 5 female) with a mean age of 33 years (± 8) without any history of cardiac diseases were examined on a GE Signa 3.0 T and a GE Signa 1.5 T TwinSpeed Excite (GE Medical Systems, Milwaukee, WI, USA) scanner using a 4-element phased array surface coil (same design) on the same day. For tissue characterization ECG gated Fast Spinecho (FSE) T1- (Double IR), T1-STIR (Triple IR) and T2-weighted sequences in transverse orientation were used. For functional analysis a steady state free precession (SSFP - FIESTA) sequence was performed in the 4-chamber, 2-chamber long axis and short axis view. The flip angle used for the SSFP sequence at 3.0 T was reduced from 45° to 30° to keep short TR times while staying within the pre-defined SAR limitations. All other sequence parameters were kept constant. Results: All acquisitions could successfully be completed for the 10 volunteers. The mean SNR 3.0 T compared to 1.5 T was remarkably increased (p < 0.05) for the T2 - (160 % SNR increase), the STIR-T1- (123 %) and the T1- (91 %) weighted FSE. Similar results were found comparing CNR at 3.0 T and 1.5 T. The mean SNR achieved using the SSFP sequences was more than doubled by 3.0 T (150 %), but did not have any significant effect on the CNR. The image quality at 3.0 T did not appear to be improved, and was considered to be significantly worse when using SSFP sequences. Artefacts like shading in the area of the right ventricle (RV) were found to be more present at 3.0 T using FSE sequences. After a localized shim had been performed in 5/10 volunteers at the infero-lateral wall of the left ventricle (LV) with the SSFP sequences at 3.0 T no significant increase in artefacts could be detected. Conclusions: In all cardiac FSE sequences, SNR and CNR at 3.0 T were found to be increased compared to 1.5 T without any major changes of the sequence parameters. The adjusted SSFP sequences fulfilled the expected increase in SNR at 3.0 T but showed no increase in CNR. On the contrary, the overall image quality did not change or was even found to be significantly lower for the SSFP and the FSE sequences at the free wall of the RV. Nevertheless, the results are encouraging for the use of 3.0 T for cardiac tissue characterization and new applications with progressing use of parallel imaging.

1 Die Arbeit wurde von der DFG unter folgender Fördernummer unterstützt: Fe 206/8-1-723/01. Die Arbeit enthält Auszüge der Doktorarbeiten von Frau Kerstin Schwinge und Herrn Patrick Freyhardt.

References

  • 1 Wen H, Denison T J, Singerman R W. et al . The intrinsic signal-to-noise ratio in human cardiac imaging at 1.5, 3, and 4 T.  J Magn Reson. 1997;  125 65-71
  • 2 Noeske R, Seifert F, Rhein K H. et al . Human cardiac imaging at 3 T using phased array coils.  Magn Reson Med. 2000;  44 978-982
  • 3 Greenman R L, Shirosky J E, Mulkern R V. et al . Double inversion black-blood fast spin-echo imaging of the human heart: a comparison between 1.5T and 3.0T.  J Magn Reson Imaging. 2003;  17 648-655
  • 4 Hinton D P, Wald L L, Pitts J. et al . Comparison of cardiac MRI on 1.5 and 3.0 Tesla clinical whole body systems.  Invest Radiol. 2003;  38 436-442
  • 5 Kozerke S SM, Fischer S E. et al . Cardiac SSFP imaging using SENSE at 3.0 T.  Proc Intl Soc Magn Reson Med. 2002;  1736
  • 6 Ryf S KS, Spiegel M. et al . Myocardial tagging: comparing imaging at 3.0 T and 1.5T.  Proc Intl Soc Magn Reson Med. 2002;  1675
  • 7 Singerman R W, Denison T J, Wen H. et al . Simulation of B1 field distribution and intrinsic signal-to-noise in cardiac MRI as a function of static magnetic field.  J Magn Reson. 1997;  125 72-83
  • 8 Stuber M, Botnar R M, Fischer S E. et al . Preliminary report on in vivo coronary MRA at 3 Tesla in humans.  Magn Reson Med. 2002;  48 425-429
  • 9 IEC . Recommendation on MRI safety.  IEC 60 601-2-33/FDIS. 2002; 
  • 10 Kozerke S SM, Fischer S E. et al . Cardiac SSFP imaging using SENSE at 3.0 T.  Proc Intl Soc Magn Reson Med. 2002;  1736
  • 11 Hoffmann K T, Hosten N, Ehrenstein T. et al . The T2-weighted half-Fourier acquired single-shot turbo-spin-echo technic compared to the conventional T2-weighted turbo-spin-echo technic for cerebral magnetic resonance tomography. A sequence comparison.  Fortschr Röntgenstr. 2000;  172 521-526
  • 12 Miller S, Simonetti O P, Carr J. et al . MR Imaging of the heart with cine true fast imaging with steady-state precession: influence of spatial and temporal resolutions on left ventricular functional parameters.  Radiology. 2002;  223 263-269
  • 13 Jones R WWR. Signal intensity artifacts in clinical MR imaging.  Radiographics. 2000;  20 (3) 893-901
  • 14 Lunati E CP, Villa M. New adiabatic inversion pulses for magnetic resonance imaging.  Phys Med Biol. 1999;  44 (6) N99-105

1 Die Arbeit wurde von der DFG unter folgender Fördernummer unterstützt: Fe 206/8-1-723/01. Die Arbeit enthält Auszüge der Doktorarbeiten von Frau Kerstin Schwinge und Herrn Patrick Freyhardt.

PD Dr. med. Matthias Gutberlet

Klinik für Strahlenheilkunde, Charité, Campus Virchow Klinikum, Universitätsmedizin Berlin

Augustenburger Platz 1

13353 Berlin

Phone: +49/30/4 50 55-70 01

Fax: +49/30/4 50 55-79 01

Email: matthias.gutberlet@charite.de

    >