Erschienen in: BMC Infectious Diseases 1/2009

Open Access 01.12.2009 | Research article

Sensitivity of direct versus concentrated sputum smear microscopy in HIV-infected patients suspected of having pulmonary tuberculosis

verfasst von: Adithya Cattamanchi, David W Dowdy, J Lucian Davis, William Worodria, Samuel Yoo, Moses Joloba, John Matovu, Philip C Hopewell, Laurence Huang

Erschienen in: BMC Infectious Diseases | Ausgabe 1/2009

Abstract

Background

Sputum concentration increases the sensitivity of smear microscopy for the diagnosis of tuberculosis (TB), but few studies have investigated this method in human immunodeficiency virus (HIV)-infected individuals.

Methods

We performed a prospective, blinded evaluation of direct and concentrated Ziehl-Neelsen smear microscopy on a single early-morning sputum sample in HIV-infected patients with > 2 weeks of cough hospitalized in Kampala, Uganda. Direct and concentrated smear results were compared with results of Lowenstein-Jensen culture.

Results

Of 279 participants, 170 (61%) had culture-confirmed TB. The sensitivity of direct and concentrated smear microscopy was not significantly different (51% vs. 52%, difference 1%, 95% confidence interval (CI): [-7%, 10%], p = 0.88). However, when results of both direct and concentrated smears were considered together, sensitivity was significantly increased compared with either method alone (64%, 95% CI: [56%, 72%], p < 0.01 for both comparisons) and was similar to that of direct smear results from consecutive (spot and early-morning) specimens (64% vs. 63%, difference 1%, 95% CI: [-6%, 8%], p = 0.85). Among 109 patients with negative cultures, one had a positive direct smear and 12 had positive concentrated smears (specificity 99% vs. 89%, difference 10%, 95% CI: [2%, 18%], p = 0.003). Of these 13 patients, 5 (38%) had improved on TB therapy after two months.

Conclusion

Sputum concentration did not increase the sensitivity of light microscopy for TB diagnosis in this HIV-infected population. Given the resource requirements for sputum concentration, additional studies using maximal blinding, high-quality direct microscopy, and a rigorous gold standard should be conducted before universally recommending this technique.
Hinweise

Electronic supplementary material

The online version of this article (doi:10.​1186/​1471-2334-9-53) contains supplementary material, which is available to authorized users.
Adithya Cattamanchi, David W Dowdy contributed equally to this work.

Competing interests

The authors declare that they have no competing interests.

Authors' contributions

AC and JLD participated in study design, data collection, statistical analysis, and drafting of the manuscript. DWD performed the primary statistical analysis and drafted the initial manuscript. WW, SY, MJ, and JM participated in study design, data collection, and drafting of the manuscript. PCH and LH participated in study design and drafting of the manuscript. All authors approved the final manuscript.
Abkürzungen
TB
tuberculosis
HIV
human immunodeficiency virus
CI
confidence interval
BAL
bronchoalveolar lavage
NTRL
National Tuberculosis and Leprosy Programme Reference Laboratory
NALC
N-acetyl-L-cysteine
NaOH
sodium hydroxide
AFB
acid-fast bacilli
WHO
World Health Organization
IUATLD
International Union Against Tuberculosis and Lung Disease
HPF
high-powered field.

Background

Direct sputum smear microscopy is the cornerstone of tuberculosis (TB) diagnosis worldwide [1]. Direct smear microscopy is rapid, inexpensive [24], highly specific [57], and capable of identifying the most infectious cases of TB [79], but its sensitivity is limited, particularly in those with human immunodeficiency virus (HIV) co-infection [1013]. Processing of sputum with subsequent concentration by centrifugation or sedimentation may increase the sensitivity of smear microscopy [14], and some investigators have recommended sputum processing and concentration as a global standard [15]. However, others have called for more evidence before implementing such a policy change, citing concerns such as increased cost for materials and training, higher biosafety requirements, and difficulty standardizing techniques across sites [16]. Of equally great concern, sputum concentration carries a risk of decreased sensitivity (e.g., through destruction of bacilli during concentration) [17, 18] and specificity (e.g., through contamination during additional transfer steps) [19, 20]. In addition, the difficulty in blinding readers as to whether a specimen is concentrated presents an inherent risk of bias in studies of sputum concentration. Few studies have employed the clear design and reporting requirements required to minimize this inherent risk [21] or evaluated sputum concentration in populations with high HIV prevalence [2226].
We therefore performed a prospective, blinded evaluation of direct and concentrated smear microscopy – performed simultaneously on a single early-morning sputum specimen – in a population of hospitalized, HIV-infected patients with cough for 2 or more weeks in Kampala, Uganda.

Methods

Study Population

Consecutive HIV-infected patients admitted to the medical wards of Mulago Hospital (Kampala, Uganda) between September 2007 and April 2008 for respiratory illness with cough of at least 2 weeks' duration were eligible for the study. We included patients who provided informed consent and an early-morning sputum specimen for TB diagnosis. We excluded patients who were receiving anti-TB treatment or had clinical evidence of congestive heart failure. The study protocol was approved by the institutional review boards at Makerere University, Mulago Hospital, the Uganda National Council for Sciences and Technology, and the University of California, San Francisco.

Patient Evaluation

All patients were tested for HIV infection with a sequential testing algorithm incorporating three rapid enzyme immunoassay kits. For TB diagnosis, patients provided a randomly-timed sputum sample for direct smear microscopy at the time of enrollment. In addition, patients provided an early-morning sputum sample on the morning following admission; this sample was sent for both direct and concentrated smear microscopy (as described below). Patients without any positive smear examinations were offered bronchoscopy with bronchoalveolar lavage (BAL) if the procedure was deemed safe and appropriate by the chest medicine consultant. All sputum and BAL specimens were sent for mycobacterial culture. Patients with suspected TB (determined by the treating ward physician) began treatment with isoniazid, rifampin, ethambutol, and pyrazinamide. Patients were evaluated during an outpatient visit or by telephone interview between two and four months after hospital discharge to assess for clinical/radiographic improvement.

Laboratory Methods

Sputum and BAL samples were analyzed at the Uganda National Tuberculosis and Leprosy Programme Reference Laboratory (NTRL). Both direct and concentrated smears were prepared from the same specimen. Direct smears were prepared and stained using the hot Ziehl-Neelsen method (1% carbol-fuchsin dye) [27]. Specimens were then decontaminated with a 1% N-acetyl-L-cysteine (NALC)-2% sodium hydroxide (NaOH)-2% sodium citrate solution and concentrated by centrifugation at 3000 × g for 10 minutes [28]. The concentrated specimen was then used to prepare mycobacterial cultures by inoculation onto two separate Lowenstein-Jensen slants and, for the early-morning specimen only, a concentrated smear stained using the hot Ziehl-Neelsen method [29]. Concentrated smears were labeled with random identification numbers immediately after preparation so that readers could not determine if direct and concentrated smears were from the same patient.
NTRL staff, who were also blinded to all clinical information, read all smears within 48 hours of preparation using a standard light microscope (magnification 1000×). They reported the presence or absence of acid-fast bacilli (AFB) using the WHO/IUATLD scale, with a positive result corresponding to ≥ 1 AFB per 100 high-power fields (HPFs) [29]. They also read cultures weekly, designating as negative any slant with no growth after eight weeks of incubation. Positive cultures (defined as ≥ 1 colony forming units) were confirmed with Ziehl-Neelsen staining. Study staff promptly communicated all laboratory results to ward physicians responsible for treatment decisions. Since 2005, the Uganda NTRL has participated in a semi-annual external quality assurance program for smear microscopy administered by the World Health Organization and has passed all quality assurance assessments.

Outcome Definitions

The primary outcome for our analyses was culture-positive TB, defined as a positive Lowenstein-Jensen (LJ) culture result from the randomly-timed sputum specimen, early morning sputum specimen, or, when available, BAL specimen. We performed two secondary analyses using different "gold standard" definitions of TB. First, we restricted the definition of TB to include only patients with a positive culture on the same specimen from which smears were prepared. Second, we broadened the definition of TB to also include patients who improved clinically on empiric TB therapy, as documented by a study medical officer and a chest consultant (W.W. or S.Y.) between two and four months after hospital discharge.

Sample Size

We aimed to collect concentrated sputum specimens from 329 patients, in order to provide 90% power to detect a difference between 50% and 60% sensitivity for direct and concentrated sputum smear, respectively, assuming a 2-sided alpha of 0.05, phi (correlation coefficient) of 0.5, and a projected 20% dropout rate due to contamination or failure to perform culture (i.e., final sample size of 262 patients). Sample size calculations were performed using PS: Power and Sample Size Calculation, version 2.1.31 [30].

Statistical Analysis

Analyses were performed using STATA 9.0 (Stata Corp., College Station, TX). Sensitivity and specificity were calculated in reference to the outcomes defined above, and compared between diagnostic strategies using McNemar's test. Bivariate comparisons were made using Fisher's exact test for dichotomous variables and the Wilcoxon rank-sum test for continuous variables. Concordance was measured using the kappa statistic. All p-values were two-sided, with statistical significance defined as p < 0.05.

Results

Study Population

Of 388 eligible patients, 39 (10%) were unable to provide an early-morning sputum specimen (unable or unwilling to spontaneously expectorate), 20 (5%) did not have a concentrated smear performed, 48 (12%) had a contaminated sputum culture, and 2 (1%) did not have culture performed despite the availability of concentrated smear, giving a final sample size of 279 HIV-infected TB suspects (Figure 1). The majority of exclusions (other than for contamination) occurred during a two-week period when the NTRL lacked sufficient staffing to process research samples in addition to their routine clinical work. Patients excluded from the study did not differ from those included with regard to baseline characteristics (gender, age, CD4 count, antiretroviral use) or the proportion of TB cases confirmed by positive direct or concentrated smear results (p > 0.15 for all comparisons). However, excluded patients were more likely to die in the hospital (28% vs. 8%, p < 0.001).
Of the 279 patients included in the study, 143 (51%) were positive for TB on culture of the early-morning sputum specimen used for comparison of direct and concentrated smear results. An additional 27 (10%) were positive on culture of another specimen. Of the remaining 109 patients, 103 (98%) had two or more negative cultures and six had only a single negative culture. Exclusion of these latter six patients did not materially affect results (all had negative direct and concentrated smears). Patients with at least one positive TB culture had significantly lower CD4+ T-lymphocyte counts than patients with negative TB cultures (median 33 vs. 101, p < 0.0001), but these two groups did not differ significantly by gender, age, education, antiretroviral use, or by mortality at hospital discharge or at two months (Table 1). Regarding sputum quality, of the 279 specimens, 33 (12%) were described as salivary, 191 (68%) mucoid, 41 (15%) purulent, and 14 (5%) bloody. Exclusion of salivary specimens or restriction to mucoid specimens reduced the sample size but did not materially affect results.
Table 1
Characteristics of Study Population, by Tuberculosis Culture Status
Characteristic
Total
(n = 279)
Culture-positive
(n = 170)
Culture-negative
(n = 109)
p-value
Female gender, n (%)
166 (60)
98 (58)
68 (62)
0.46
Age, median (IQR)
34 (27–40)
33 (27–40)
34 (28–40)
0.88
Education beyond primary level, n (%)
97 (35)
61 (36)
36 (33)
0.70
CD4 count, median (IQR) (n = 274)
49 (15–169)
33 (12–115)
101 (31–219)
< 0.0001
Antiretroviral use on admission, n (%)
42 (15)
24 (14)
18 (17)
0.37
Mortality, n (%)
    
   In hospital (n = 279)
23 (8)
13 (8)
10 (9)
0.66
   Two-month (n = 241)
65 (27)
43 (30)
22 (23)
0.30
Clinical presentation, n (%)
    
   Fever/chills/sweats
262 (94)
158 (93)
104 (95)
0.45
   Weight loss
265 (95)
167 (98)
98 (90)
0.003
   Hemoptysis
68 (28)
31 (21)
37 (40)
0.001
IQR, interquartile range

Sensitivity and Specificity of Direct and Concentrated Sputum Smear Microscopy

As shown in Table 2, the sensitivity of direct and concentrated smear microscopy was similar when using any positive culture result as the gold standard (51% vs. 52%, difference 1%, 95% confidence interval: [-7%, 10%], p = 0.88). However, concordance between the two methods was only fair, both among all specimens (80% overall agreement; unweighted kappa 0.56, 95% CI: [0.44, 0.68]) and among culture-positive specimens (75% overall agreement; unweighted kappa 0.51, 95% CI: [0.35, 0.66]). As a result, the performance of both smear techniques on the same specimen yielded an absolute increase of 12–13% in sensitivity over either smear alone (p < 0.01 for increase in sensitivity over direct or concentrated smear alone). In addition, combining the results of direct and concentrated smears of a single specimen had a similar sensitivity to performing direct smear microscopy on two consecutive (randomly-timed and early-morning) sputum specimens (64% vs. 63%, difference 1%, 95% confidence interval: [-6%, 8%], p = 0.85). Among the 109 patients with negative cultures, one had a positive direct smear (specificity 99%, 95% CI: [94%, 100%]), and 12 had positive concentrated smears (specificity 89%, 95% CI: [81%, 95%], p = 0.003 for difference).
Table 2
Sensitivity and Specificity of Concentrated and Direct Smear, By Reference Outcome
Concentrated Smear Result
Direct Smear Result
Total
Gold Standard: Sputum Culture
A. Sensitivity
   
 
Positive
Negative
 
Positive
67 (39%)
22 (13%)
89 (52%)
Negative
20 (12%)
61 (36%)
81 (48%)
Total
87 (51%)
83 (49%)
170 (100%)
Sensitivity of Direct Smear = 0.51 (0.43–0.59)
Sensitivity of Concentrated Smear = 0.52 (0.44–0.61)
Sensitivity of Direct + Concentrated Smear = 0.64 (0.56–0.72) a, b
B. Specificity
   
 
Positive
Negative
 
Positive
0 (0%)
12 (11%)
12 (11%)
Negative
1 (1%)
96 (88%)
97 (89%)
Total
1 (1%)
108 (99%)
109 (100%)
Specificity of Direct Smear = 0.99 (0.94–1.0) a
Specificity of Concentrated Smear = 0.89 (0.81–0.95) b
Specificity of Direct + Concentrated Smear = 0.88 (0.80–0.94) b
Gold Standard: Sputum Culture + Clinical Criteria
A. Sensitivity
   
 
Positive
Negative
 
Positive
67 (34%)
27 (14%)
94 (48%)
Negative
20 (10%)
82 (42%)
102 (52%)
Total
87 (44%)
109 (56%)
196 (100%)
Sensitivity of Direct Smear = 0.44 (0.36–0.52)
Sensitivity of Concentrated Smear = 0.48 (0.40–0.56)
Sensitivity of Direct + Concentrated Smear = 0.58 (0.50–0.66) a, b
B. Specificity
   
 
Positive
Negative
 
Positive
0 (0%)
7 (8%)
7 (8%)
Negative
1 (1%)
75 (90%)
76 (92%)
Total
1 (1%)
82 (99%)
83 (100%)
Specificity of Direct Smear = 0.99 (0.93–1.0)
Specificity of Concentrated Smear = 0.92 (0.83–0.97)
Specificity of Direct + Concentrated Smear = 0.90 (0.81–0.96) b
a p < 0.05 for comparison with concentrated smear alone
b p < 0.05 for comparison with direct smear alone
We performed two secondary analyses using different definitions for the gold standard. First, when we restricted the gold standard definition to a positive culture result of the same specimen from which smears were prepared, the sensitivities of direct and concentrated smear remained similar (59% vs. 58%, p = 1.0), and the specificity of direct smear remained higher than that of concentrated smear (97% vs. 87%, p = 0.003). Second, when we expanded the gold standard to include patients with negative cultures but clinical response to TB therapy, the sensitivity of the two methods remained similar (44% vs. 48%, p = 0.38). However, the difference in specificity was no longer statistically significant (92% vs. 99%, p = 0.07) because 5 (42%) of the 12 patients with positive concentrated smears but negative cultures had documented clinical improvement on TB therapy at 2-month follow-up (Table 3).
Table 3
Data on Patients with Positive Smears and Negative Cultures
Patient ID
Admission CD4 Count
Smear Resulta
Final Diagnosis
Treated for TB After Discharge?
Status 2 Months After Discharge
  
Direct
Concentrated
   
1
31
 
1+
TB
Yes
Improved
2
193
 
1+
TB
Yes
Improved
3
196
 
1+
TB
Yes
Improved
4
398
 
Scanty
TB
Yes
Improved
5
687
 
3+
TB
Yes
Improved
6
110
 
1+
Aspergillus
Yes
No Improvement
7
3
 
1+
PCP
Died in hospital
Dead
8
5
 
Scanty
Bacterial Pneumonia
No
Improved
9 b
12
Scanty
 
Unknown
Died in hospital
Dead
10
31
 
1+
Unknown
Died in hospital
Dead
11
59
 
1+
Unknown
Died in hospital
Dead
12
81
 
Scanty
Bacterial Pneumonia
No
Improved
13
399
 
1+
Unknown
No
No Improvement
TB, tuberculosis; PCP, Pneumocystis jirovecii pneumonia
a Scanty = 1–9 acid-fast bacilli (AFB) per 100 high-power fields (HPF), 1+ = 1–9 AFB per 10 HPF, 3+ = >10 AFB per HPF.
b Direct smear alone was positive; for other patients, the concentrated smear alone was positive.

Density of Acid-Fast Bacilli

Table 4 compares the density of AFB on specimens from patients with culture-confirmed TB. Of 170 confirmed TB patients, 67 (39%) were positive on both direct and concentrated smear, 61 (36%) were negative on both, 20 (12%) were positive only on concentrated smear, and 22 (13%) were positive only on direct smear. Among 20 specimens that were positive only after concentration, 8 (40%) had 10 or more AFB per 100 HPF. By contrast, among 22 specimens that were positive only on direct smear, 19 (86%) had 10 or more AFB per 100 HPF.
Table 4
Density of Acid-Fast Bacilli (per 100 High-Power Fields) in Patients with Culture-Positive Tuberculosis
  
Direct Smear Result
Total
Concentrated Smear Result
 
0
1–9
≥ 10
 
 
0
61
3
19
83
 
1–9
12
2
2
16
 
≥ 10
8
7
56
71
Total
 
81
12
77
170

Discussion

Sputum concentration did not increase the sensitivity of light microscopy for TB diagnosis in this prospective, blinded evaluation of 279 hospitalized, HIV-infected adults in Kampala, Uganda. Moreover, sputum concentration decreased specificity, though this difference was attenuated when clinical response to TB therapy was considered in the gold standard. The performance of concentrated smear microscopy requires further evaluation before recommending universal implementation of this technique.
Our finding that sputum concentration prior to Ziehl-Neelsen staining did not significantly increase sensitivity for TB over direct smear microscopy differs from the results of five prior studies that found higher sensitivity in HIV-endemic populations after sputum concentration [3135]. This discrepancy may reflect differences in study site (e.g., research versus field setting), patient population (e.g., specially-selected outpatients versus consecutive hospitalized patients), gold standard (e.g., multiple cultures and follow-up data not routinely used versus used), or study methodology (e.g., sodium hypochlorite and sedimentation versus NALC-NaOH and centrifugation). In addition, although many studies employed some form of blinding, it is difficult to blind readers to processing method (direct vs. concentrated) due to the different appearance of unprocessed versus processed slides. This imperfect blinding is known to influence the results of sputum smear microscopy [36] and would be expected to inflate measured sensitivity for concentrated smear. Although we cannot exclude random chance or weaknesses in our study design as the explanation for our negative results, a number of additional considerations speak to the veracity of our findings. First, our reported 52% sensitivity of concentrated smear is similar to that in previous studies of HIV-infected individuals (i.e., 50% [37], 52% [38], and 54% [39]), suggesting that the discrepancy in findings may reflect higher sensitivity of direct microscopy in the present study, rather than higher sensitivity of concentrated microscopy in other studies. In addition, our study incorporated several features that strengthen its internal validity – consecutive enrollment of a relevant population, clear description of included and excluded patients, blinded reading of index and reference tests by trained non-research staff, and a gold standard which incorporated multiple culture results and follow-up assessment [40, 41]. Thus, while our findings differ from those of previous studies, the present study provides sound evidence that sputum concentration does not universally increase smear sensitivity among HIV-infected TB suspects, and it suggests that provision of high-quality direct microscopy may diminish the benefit of sputum concentration.
In this study, direct and concentrated sputum smear had surprisingly low concordance among patients with culture-confirmed TB (kappa = 0.51), even though both techniques were performed on the same sputum specimen. Although direct and concentrated smear microscopy had similar sensitivities, they detected different patients: 39% (42/109) of all smear-positive patients with culture-confirmed TB were only positive by a single method (Table 2). Although most prior studies [42, 43] have found moderate-to-high concordance between direct and concentrated smear, one earlier study [44] – also performed using high-speed centrifugation in an African reference lab – found results similar to those presented here (sensitivity of 43% for direct smear, 44% for concentrated smear, and 55% for direct plus concentrated smear). Collection of a single sputum specimen would reduce patient burden in terms of number of visits to health care centers, and could also decrease the delay between clinical presentation and initiation of treatment. Given the similar sensitivity of direct and concentrated smear microscopy also observed in our study, future studies should investigate whether performing multiple direct smears on a single specimen can substantially increase sensitivity for TB, avoiding the need for multiple patient visits.
As with any evaluation of smear microscopy, this study has certain limitations. First, our study was designed to measure a clinically relevant difference between direct and concentrated microscopy and was not powered to demonstrate equivalence. However, our 95% confidence intervals exclude more than a 10% absolute difference in sensitivity between the two methods. Second, this study was conducted in a national reference laboratory using NALC-NaOH on early-morning specimens collected from a population of HIV-infected, hospitalized patients. Thus, our results may not fully generalize to other settings (e.g., peripheral laboratories, laboratories using alternative sputum processing methods, non-HIV populations, and healthier outpatient populations). Estimates of diagnostic performance are known to vary between ambulatory and hospital settings. However, the choice of study population is less likely to impact a comparison between two diagnostic techniques. In addition, given the rigorous training required of microscopists at the Uganda NTRL, it is unlikely that laboratory inexperience explains the results of the present study. Finally, in order to better replicate actual test conditions, internal quality assurance was not performed during the study period. Though we would not expect reliability to differentially affect direct versus concentrated sputum smear results, we were unable to quantify inter-reader and intra-reader agreement.

Conclusion

In conclusion, we failed to find a difference in sensitivity between direct and concentrated sputum smear microscopy performed in a national reference laboratory serving an HIV-infected hospitalized adult population. Before widely recommending sputum concentration, additional field evaluations that demonstrate benefit when incorporating strict blinding, high quality direct smear microscopy, and a clear gold standard are needed. Such studies should also investigate whether simpler modifications (e.g., parallel performance of multiple direct smears on a single sputum specimen) can similarly increase sensitivity and cost-effectiveness. Ultimately, modifications in smear microscopy may increase the yield of TB diagnosis only marginally, a possibility which emphasizes the need for development and testing of novel rapid diagnostic technologies.

Acknowledgements

The authors would like to thank Elisha Hatanga, Peter Awongo, Fred Kangave, William Ssenfuma, Nicholas Ezati, and Raymond Assimwe at the Uganda NTRL for performing smear microscopy and mycobacterial culture for this study. The authors also would like to thank Saskia den Boon, Florence Nankya, Rachel Kyeyune, John Kiidha, and Patrick Byanyima for enrolling and caring for the patients included in this study.
Financial Support:
This work was supported by grant numbers K24 HL087713 (LH), F32 HL088990 (JLD), and K23 AI080147 (JLD) from the National Institutes of Health, as well as an unrestricted grant from the Foundation for Innovative New Diagnostics.
This article is published under license to BioMed Central Ltd. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://​creativecommons.​org/​licenses/​by/​2.​0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Competing interests

The authors declare that they have no competing interests.

Authors' contributions

AC and JLD participated in study design, data collection, statistical analysis, and drafting of the manuscript. DWD performed the primary statistical analysis and drafted the initial manuscript. WW, SY, MJ, and JM participated in study design, data collection, and drafting of the manuscript. PCH and LH participated in study design and drafting of the manuscript. All authors approved the final manuscript.
Anhänge

Authors’ original submitted files for images

Below are the links to the authors’ original submitted files for images.
Literatur
1.
Zurück zum Zitat World Health Organization: Global tuberculosis control 2008: surveillance, planning, financing. 2008, Geneva: WHO World Health Organization: Global tuberculosis control 2008: surveillance, planning, financing. 2008, Geneva: WHO
2.
Zurück zum Zitat Albert H: Economic analysis of the diagnosis of smear-negative pulmonary tuberculosis in South Africa: incorporation of a new rapid test, FASTPlaqueTB, into the diagnostic algorithm. Int J Tuberc Lung Dis. 2004, 8: 240-247.PubMed Albert H: Economic analysis of the diagnosis of smear-negative pulmonary tuberculosis in South Africa: incorporation of a new rapid test, FASTPlaqueTB, into the diagnostic algorithm. Int J Tuberc Lung Dis. 2004, 8: 240-247.PubMed
3.
Zurück zum Zitat Suarez PG, Floyd K, Portocarrero J, Alarcon E, Rapiti E, Ramos G, et al: Feasibility and cost-effectiveness of standardised second-line drug treatment for chronic tuberculosis patients: a national cohort study in Peru. Lancet. 2002, 359: 1980-1989. 10.1016/S0140-6736(02)08830-X.CrossRefPubMed Suarez PG, Floyd K, Portocarrero J, Alarcon E, Rapiti E, Ramos G, et al: Feasibility and cost-effectiveness of standardised second-line drug treatment for chronic tuberculosis patients: a national cohort study in Peru. Lancet. 2002, 359: 1980-1989. 10.1016/S0140-6736(02)08830-X.CrossRefPubMed
4.
Zurück zum Zitat van Cleeff M, Kivihya-Ndugga L, Githui W, Ng'ang'a L, Kibuga D, Odhiambo J, et al: Cost-effectiveness of polymerase chain reaction versus Ziehl-Neelsen smear microscopy for diagnosis of tuberculosis in Kenya. Int J Tuberc Lung Dis. 2005, 9: 877-883.PubMed van Cleeff M, Kivihya-Ndugga L, Githui W, Ng'ang'a L, Kibuga D, Odhiambo J, et al: Cost-effectiveness of polymerase chain reaction versus Ziehl-Neelsen smear microscopy for diagnosis of tuberculosis in Kenya. Int J Tuberc Lung Dis. 2005, 9: 877-883.PubMed
5.
Zurück zum Zitat Keeler E, Perkins MD, Small P, Hanson C, Reed S, Cunningham J, et al: Reducing the global burden of tuberculosis: the contribution of improved diagnostics. Nature. 2006, 444 (Suppl 1): 49-57. 10.1038/nature05446.CrossRefPubMed Keeler E, Perkins MD, Small P, Hanson C, Reed S, Cunningham J, et al: Reducing the global burden of tuberculosis: the contribution of improved diagnostics. Nature. 2006, 444 (Suppl 1): 49-57. 10.1038/nature05446.CrossRefPubMed
6.
Zurück zum Zitat Levy H, Feldman C, Sacho H, van der MH, Kallenbach J, Koornhof H: A reevaluation of sputum microscopy and culture in the diagnosis of pulmonary tuberculosis. Chest. 1989, 95: 1193-1197. 10.1378/chest.95.6.1193.CrossRefPubMed Levy H, Feldman C, Sacho H, van der MH, Kallenbach J, Koornhof H: A reevaluation of sputum microscopy and culture in the diagnosis of pulmonary tuberculosis. Chest. 1989, 95: 1193-1197. 10.1378/chest.95.6.1193.CrossRefPubMed
7.
Zurück zum Zitat Toman's tuberculosis. Case detection, treatment, and monitoring. 2004, Geneva: World Health Organization, 2 Toman's tuberculosis. Case detection, treatment, and monitoring. 2004, Geneva: World Health Organization, 2
8.
Zurück zum Zitat Behr MA, Warren SA, Salamon H, Hopewell PC, Ponce dL, Daley CL, et al: Transmission of Mycobacterium tuberculosis from patients smear-negative for acid-fast bacilli. Lancet. 1999, 353: 444-449. 10.1016/S0140-6736(98)03406-0.CrossRefPubMed Behr MA, Warren SA, Salamon H, Hopewell PC, Ponce dL, Daley CL, et al: Transmission of Mycobacterium tuberculosis from patients smear-negative for acid-fast bacilli. Lancet. 1999, 353: 444-449. 10.1016/S0140-6736(98)03406-0.CrossRefPubMed
9.
Zurück zum Zitat Narain R, Rao MS, Chandrasekhar P, Pyarelal : Microscopy positive and microscopy negative cases of pulmonary tuberculosis. Am Rev Respir Dis. 1971, 103: 761-773.PubMed Narain R, Rao MS, Chandrasekhar P, Pyarelal : Microscopy positive and microscopy negative cases of pulmonary tuberculosis. Am Rev Respir Dis. 1971, 103: 761-773.PubMed
10.
Zurück zum Zitat Getahun H, Harrington M, O'Brien R, Nunn P: Diagnosis of smear-negative pulmonary tuberculosis in people with HIV infection or AIDS in resource-constrained settings: informing urgent policy changes. Lancet. 2007, 369: 2042-2049. 10.1016/S0140-6736(07)60284-0.CrossRefPubMed Getahun H, Harrington M, O'Brien R, Nunn P: Diagnosis of smear-negative pulmonary tuberculosis in people with HIV infection or AIDS in resource-constrained settings: informing urgent policy changes. Lancet. 2007, 369: 2042-2049. 10.1016/S0140-6736(07)60284-0.CrossRefPubMed
11.
Zurück zum Zitat Corbett EL, Watt CJ, Walker N, Maher D, Williams BG, Raviglione MC, et al: The growing burden of tuberculosis: global trends and interactions with the HIV epidemic. Arch Intern Med. 2003, 163: 1009-1021. 10.1001/archinte.163.9.1009.CrossRefPubMed Corbett EL, Watt CJ, Walker N, Maher D, Williams BG, Raviglione MC, et al: The growing burden of tuberculosis: global trends and interactions with the HIV epidemic. Arch Intern Med. 2003, 163: 1009-1021. 10.1001/archinte.163.9.1009.CrossRefPubMed
12.
Zurück zum Zitat Elliott AM, Halwiindi B, Hayes RJ, Luo N, Tembo G, Machiels L, et al: The impact of human immunodeficiency virus on presentation and diagnosis of tuberculosis in a cohort study in Zambia. J Trop Med Hyg. 1993, 96: 1-11.PubMed Elliott AM, Halwiindi B, Hayes RJ, Luo N, Tembo G, Machiels L, et al: The impact of human immunodeficiency virus on presentation and diagnosis of tuberculosis in a cohort study in Zambia. J Trop Med Hyg. 1993, 96: 1-11.PubMed
13.
Zurück zum Zitat Lawson L, Yassin MA, Ramsay A, Emenyonu NE, Squire SB, Cuevas LE: Comparison of scanty AFB smears against culture in an area with high HIV prevalence. Int J Tuberc Lung Dis. 2005, 9: 933-935.PubMed Lawson L, Yassin MA, Ramsay A, Emenyonu NE, Squire SB, Cuevas LE: Comparison of scanty AFB smears against culture in an area with high HIV prevalence. Int J Tuberc Lung Dis. 2005, 9: 933-935.PubMed
14.
Zurück zum Zitat Steingart KR, Ng V, Henry M, Hopewell PC, Ramsay A, Cunningham J, et al: Sputum processing methods to improve the sensitivity of smear microscopy for tuberculosis: a systematic review. Lancet Infect Dis. 2006, 6: 664-674. 10.1016/S1473-3099(06)70602-8.CrossRefPubMed Steingart KR, Ng V, Henry M, Hopewell PC, Ramsay A, Cunningham J, et al: Sputum processing methods to improve the sensitivity of smear microscopy for tuberculosis: a systematic review. Lancet Infect Dis. 2006, 6: 664-674. 10.1016/S1473-3099(06)70602-8.CrossRefPubMed
15.
Zurück zum Zitat Angeby KA, Hoffner SE, Diwan VK: Should the 'bleach microscopy method' be recommended for improved case detection of tuberculosis? Literature review and key person analysis. Int J Tuberc Lung Dis. 2004, 8: 806-815.PubMed Angeby KA, Hoffner SE, Diwan VK: Should the 'bleach microscopy method' be recommended for improved case detection of tuberculosis? Literature review and key person analysis. Int J Tuberc Lung Dis. 2004, 8: 806-815.PubMed
16.
Zurück zum Zitat Ramsay A, Squire SB, Siddiqi K, Cunningham J, Perkins MD: The bleach microscopy method and case detection for tuberculosis control. Int J Tuberc Lung Dis. 2006, 10: 256-258.PubMed Ramsay A, Squire SB, Siddiqi K, Cunningham J, Perkins MD: The bleach microscopy method and case detection for tuberculosis control. Int J Tuberc Lung Dis. 2006, 10: 256-258.PubMed
17.
Zurück zum Zitat Van Deun A, Maug AK, Cooreman E, Hossain MA, Chambuganj N, Rema V, et al: Bleach sedimentation method for increased sensitivity of sputum smear microscopy: does it work?. Int J Tuberc Lung Dis. 2000, 4: 371-376.PubMed Van Deun A, Maug AK, Cooreman E, Hossain MA, Chambuganj N, Rema V, et al: Bleach sedimentation method for increased sensitivity of sputum smear microscopy: does it work?. Int J Tuberc Lung Dis. 2000, 4: 371-376.PubMed
18.
Zurück zum Zitat Wilkinson D, Sturm AW: Diagnosing tuberculosis in a resource-poor setting: the value of sputum concentration. Trans R Soc Trop Med Hyg. 1997, 91: 420-421. 10.1016/S0035-9203(97)90263-7.CrossRefPubMed Wilkinson D, Sturm AW: Diagnosing tuberculosis in a resource-poor setting: the value of sputum concentration. Trans R Soc Trop Med Hyg. 1997, 91: 420-421. 10.1016/S0035-9203(97)90263-7.CrossRefPubMed
19.
Zurück zum Zitat Perera J, Arachchi DM: The optimum relative centrifugal force and centrifugation time for improved sensitivity of smear and culture for detection of Mycobacterium tuberculosis from sputum. Trans R Soc Trop Med Hyg. 1999, 93: 405-409. 10.1016/S0035-9203(99)90135-9.CrossRefPubMed Perera J, Arachchi DM: The optimum relative centrifugal force and centrifugation time for improved sensitivity of smear and culture for detection of Mycobacterium tuberculosis from sputum. Trans R Soc Trop Med Hyg. 1999, 93: 405-409. 10.1016/S0035-9203(99)90135-9.CrossRefPubMed
20.
Zurück zum Zitat Angeby KA, Hoffner SE, Diwan VK: Should the 'bleach microscopy method' be recommended for improved case detection of tuberculosis? Literature review and key person analysis. Int J Tuberc Lung Dis. 2004, 8: 806-815.PubMed Angeby KA, Hoffner SE, Diwan VK: Should the 'bleach microscopy method' be recommended for improved case detection of tuberculosis? Literature review and key person analysis. Int J Tuberc Lung Dis. 2004, 8: 806-815.PubMed
21.
Zurück zum Zitat Pai M, O'Brien R: Tuberculosis diagnostics trials: do they lack methodological rigor?. Expert Rev Mol Diagn. 2006, 6: 509-514. 10.1586/14737159.6.4.509.CrossRefPubMed Pai M, O'Brien R: Tuberculosis diagnostics trials: do they lack methodological rigor?. Expert Rev Mol Diagn. 2006, 6: 509-514. 10.1586/14737159.6.4.509.CrossRefPubMed
22.
Zurück zum Zitat Eyangoh SI, Torrea G, Tejiokem MC, Kamdem Y, Piam FF, Noeske J, et al: HIV-related incremental yield of bleach sputum concentration and fluorescence technique for the microscopic detection of tuberculosis. Eur J Clin Microbiol Infect Dis. 2008, 27: 849-855. 10.1007/s10096-008-0516-4.CrossRefPubMed Eyangoh SI, Torrea G, Tejiokem MC, Kamdem Y, Piam FF, Noeske J, et al: HIV-related incremental yield of bleach sputum concentration and fluorescence technique for the microscopic detection of tuberculosis. Eur J Clin Microbiol Infect Dis. 2008, 27: 849-855. 10.1007/s10096-008-0516-4.CrossRefPubMed
23.
Zurück zum Zitat Bonnet M, Ramsay A, Githui W, Gagnidze L, Varaine F, Guerin PJ: Bleach sedimentation: an opportunity to optimize smear microscopy for tuberculosis diagnosis in settings of high prevalence of HIV. Clin Infect Dis. 2008, 46: 1710-1716. 10.1086/587891.CrossRefPubMed Bonnet M, Ramsay A, Githui W, Gagnidze L, Varaine F, Guerin PJ: Bleach sedimentation: an opportunity to optimize smear microscopy for tuberculosis diagnosis in settings of high prevalence of HIV. Clin Infect Dis. 2008, 46: 1710-1716. 10.1086/587891.CrossRefPubMed
24.
Zurück zum Zitat Lawson L, Yassin MA, Ramsay A, Emenyonu EN, Thacher TD, Davies PD, et al: Short-term bleach digestion of sputum in the diagnosis of pulmonary tuberculosis in patients co-infected with HIV. Tuberculosis (Edinb). 2007, 87: 368-372. 10.1016/j.tube.2007.02.002.CrossRef Lawson L, Yassin MA, Ramsay A, Emenyonu EN, Thacher TD, Davies PD, et al: Short-term bleach digestion of sputum in the diagnosis of pulmonary tuberculosis in patients co-infected with HIV. Tuberculosis (Edinb). 2007, 87: 368-372. 10.1016/j.tube.2007.02.002.CrossRef
25.
Zurück zum Zitat Githui WA, Matu SW, Muthami LN, Juma E: Improved diagnosis of Ziehl-Neelsen smear negative tuberculosis using sodium hypochlorite sedimentation method. East Afr Med J. 2007, 84: 455-459.PubMed Githui WA, Matu SW, Muthami LN, Juma E: Improved diagnosis of Ziehl-Neelsen smear negative tuberculosis using sodium hypochlorite sedimentation method. East Afr Med J. 2007, 84: 455-459.PubMed
26.
Zurück zum Zitat Bruchfeld J, Aderaye G, Palme IB, Bjorvatn B, Kallenius G, Lindquist L: Sputum concentration improves diagnosis of tuberculosis in a setting with a high prevalence of HIV. Trans R Soc Trop Med Hyg. 2000, 94: 677-680. 10.1016/S0035-9203(00)90230-X.CrossRefPubMed Bruchfeld J, Aderaye G, Palme IB, Bjorvatn B, Kallenius G, Lindquist L: Sputum concentration improves diagnosis of tuberculosis in a setting with a high prevalence of HIV. Trans R Soc Trop Med Hyg. 2000, 94: 677-680. 10.1016/S0035-9203(00)90230-X.CrossRefPubMed
27.
Zurück zum Zitat Rieder HL, Van Deun A, Kam KM, Kim SJ, Chonde TM, Trebucq A, et al: Priorities for Tuberculosis Bacteriology Services in Low Income Countries. 2007, Paris: International Union Against Tuberculosis and Lung Disease Rieder HL, Van Deun A, Kam KM, Kim SJ, Chonde TM, Trebucq A, et al: Priorities for Tuberculosis Bacteriology Services in Low Income Countries. 2007, Paris: International Union Against Tuberculosis and Lung Disease
28.
Zurück zum Zitat Kent P, Kubica G: Public Health Mycobacteriology – A Guide for the Level III Laboratory. 1985, Atlanta: Centers for Disease Control Kent P, Kubica G: Public Health Mycobacteriology – A Guide for the Level III Laboratory. 1985, Atlanta: Centers for Disease Control
29.
Zurück zum Zitat World Health Organization: Laboratory services in tuberculosis control. WHO/tuberculosis/98.258. 1998, Geneva: WHO World Health Organization: Laboratory services in tuberculosis control. WHO/tuberculosis/98.258. 1998, Geneva: WHO
30.
Zurück zum Zitat Dupont WD, Plummer WD: Power and sample size calculations. A review and computer program. Control Clin Trials. 1990, 11: 116-128. 10.1016/0197-2456(90)90005-M.CrossRefPubMed Dupont WD, Plummer WD: Power and sample size calculations. A review and computer program. Control Clin Trials. 1990, 11: 116-128. 10.1016/0197-2456(90)90005-M.CrossRefPubMed
31.
Zurück zum Zitat Eyangoh SI, Torrea G, Tejiokem MC, Kamdem Y, Piam FF, Noeske J, et al: HIV-related incremental yield of bleach sputum concentration and fluorescence technique for the microscopic detection of tuberculosis. Eur J Clin Microbiol Infect Dis. 2008, 27 (9): 849-55. 10.1007/s10096-008-0516-4. Epub 2008 May 8.CrossRefPubMed Eyangoh SI, Torrea G, Tejiokem MC, Kamdem Y, Piam FF, Noeske J, et al: HIV-related incremental yield of bleach sputum concentration and fluorescence technique for the microscopic detection of tuberculosis. Eur J Clin Microbiol Infect Dis. 2008, 27 (9): 849-55. 10.1007/s10096-008-0516-4. Epub 2008 May 8.CrossRefPubMed
32.
Zurück zum Zitat Bonnet M, Ramsay A, Githui W, Gagnidze L, Varaine F, Guerin PJ: Bleach sedimentation: an opportunity to optimize smear microscopy for tuberculosis diagnosis in settings of high prevalence of HIV. Clin Infect Dis. 2008, 46: 1710-1716. 10.1086/587891.CrossRefPubMed Bonnet M, Ramsay A, Githui W, Gagnidze L, Varaine F, Guerin PJ: Bleach sedimentation: an opportunity to optimize smear microscopy for tuberculosis diagnosis in settings of high prevalence of HIV. Clin Infect Dis. 2008, 46: 1710-1716. 10.1086/587891.CrossRefPubMed
33.
Zurück zum Zitat Lawson L, Yassin MA, Ramsay A, Emenyonu EN, Thacher TD, Davies PD, et al: Short-term bleach digestion of sputum in the diagnosis of pulmonary tuberculosis in patients co-infected with HIV. Tuberculosis (Edinb). 2007, 87: 368-372. 10.1016/j.tube.2007.02.002.CrossRef Lawson L, Yassin MA, Ramsay A, Emenyonu EN, Thacher TD, Davies PD, et al: Short-term bleach digestion of sputum in the diagnosis of pulmonary tuberculosis in patients co-infected with HIV. Tuberculosis (Edinb). 2007, 87: 368-372. 10.1016/j.tube.2007.02.002.CrossRef
34.
Zurück zum Zitat Githui WA, Matu SW, Muthami LN, Juma E: Improved diagnosis of Ziehl-Neelsen smear negative tuberculosis using sodium hypochlorite sedimentation method. East Afr Med J. 2007, 84: 455-459.PubMed Githui WA, Matu SW, Muthami LN, Juma E: Improved diagnosis of Ziehl-Neelsen smear negative tuberculosis using sodium hypochlorite sedimentation method. East Afr Med J. 2007, 84: 455-459.PubMed
35.
Zurück zum Zitat Bruchfeld J, Aderaye G, Palme IB, Bjorvatn B, Kallenius G, Lindquist L: Sputum concentration improves diagnosis of tuberculosis in a setting with a high prevalence of HIV. Trans R Soc Trop Med Hyg. 2000, 94: 677-680. 10.1016/S0035-9203(00)90230-X.CrossRefPubMed Bruchfeld J, Aderaye G, Palme IB, Bjorvatn B, Kallenius G, Lindquist L: Sputum concentration improves diagnosis of tuberculosis in a setting with a high prevalence of HIV. Trans R Soc Trop Med Hyg. 2000, 94: 677-680. 10.1016/S0035-9203(00)90230-X.CrossRefPubMed
36.
Zurück zum Zitat Pai M, O'Brien R: Tuberculosis diagnostics trials: do they lack methodological rigor?. Expert Rev Mol Diagn. 2006, 6: 509-514. 10.1586/14737159.6.4.509.CrossRefPubMed Pai M, O'Brien R: Tuberculosis diagnostics trials: do they lack methodological rigor?. Expert Rev Mol Diagn. 2006, 6: 509-514. 10.1586/14737159.6.4.509.CrossRefPubMed
37.
Zurück zum Zitat Bruchfeld J, Aderaye G, Palme IB, Bjorvatn B, Kallenius G, Lindquist L: Sputum concentration improves diagnosis of tuberculosis in a setting with a high prevalence of HIV. Trans R Soc Trop Med Hyg. 2000, 94: 677-680. 10.1016/S0035-9203(00)90230-X.CrossRefPubMed Bruchfeld J, Aderaye G, Palme IB, Bjorvatn B, Kallenius G, Lindquist L: Sputum concentration improves diagnosis of tuberculosis in a setting with a high prevalence of HIV. Trans R Soc Trop Med Hyg. 2000, 94: 677-680. 10.1016/S0035-9203(00)90230-X.CrossRefPubMed
38.
Zurück zum Zitat Lawson L, Yassin MA, Ramsay A, Emenyonu EN, Thacher TD, Davies PD, et al: Short-term bleach digestion of sputum in the diagnosis of pulmonary tuberculosis in patients co-infected with HIV. Tuberculosis (Edinb). 2007, 87: 368-372. 10.1016/j.tube.2007.02.002.CrossRef Lawson L, Yassin MA, Ramsay A, Emenyonu EN, Thacher TD, Davies PD, et al: Short-term bleach digestion of sputum in the diagnosis of pulmonary tuberculosis in patients co-infected with HIV. Tuberculosis (Edinb). 2007, 87: 368-372. 10.1016/j.tube.2007.02.002.CrossRef
39.
Zurück zum Zitat Eyangoh SI, Torrea G, Tejiokem MC, Kamdem Y, Piam FF, Noeske J, et al: HIV-related incremental yield of bleach sputum concentration and fluorescence technique for the microscopic detection of tuberculosis. Eur J Clin Microbiol Infect Dis. 2008, 27 (9): 849-55. 10.1007/s10096-008-0516-4. Epub 2008 May 8.CrossRefPubMed Eyangoh SI, Torrea G, Tejiokem MC, Kamdem Y, Piam FF, Noeske J, et al: HIV-related incremental yield of bleach sputum concentration and fluorescence technique for the microscopic detection of tuberculosis. Eur J Clin Microbiol Infect Dis. 2008, 27 (9): 849-55. 10.1007/s10096-008-0516-4. Epub 2008 May 8.CrossRefPubMed
40.
Zurück zum Zitat Bossuyt PM, Reitsma JB, Bruns DE, Gatsonis CA, Glasziou PP, Irwig LM, et al: Towards complete and accurate reporting of studies of diagnostic accuracy: The STARD Initiative. Ann Intern Med. 2003, 138: 40-44.CrossRefPubMed Bossuyt PM, Reitsma JB, Bruns DE, Gatsonis CA, Glasziou PP, Irwig LM, et al: Towards complete and accurate reporting of studies of diagnostic accuracy: The STARD Initiative. Ann Intern Med. 2003, 138: 40-44.CrossRefPubMed
41.
Zurück zum Zitat Pai M, O'Brien R: Tuberculosis diagnostics trials: do they lack methodological rigor?. Expert Rev Mol Diagn. 2006, 6: 509-514. 10.1586/14737159.6.4.509.CrossRefPubMed Pai M, O'Brien R: Tuberculosis diagnostics trials: do they lack methodological rigor?. Expert Rev Mol Diagn. 2006, 6: 509-514. 10.1586/14737159.6.4.509.CrossRefPubMed
42.
Zurück zum Zitat Steingart KR, Ng V, Henry M, Hopewell PC, Ramsay A, Cunningham J, et al: Sputum processing methods to improve the sensitivity of smear microscopy for tuberculosis: a systematic review. Lancet Infect Dis. 2006, 6: 664-674. 10.1016/S1473-3099(06)70602-8.CrossRefPubMed Steingart KR, Ng V, Henry M, Hopewell PC, Ramsay A, Cunningham J, et al: Sputum processing methods to improve the sensitivity of smear microscopy for tuberculosis: a systematic review. Lancet Infect Dis. 2006, 6: 664-674. 10.1016/S1473-3099(06)70602-8.CrossRefPubMed
43.
Zurück zum Zitat Bonnet M, Ramsay A, Githui W, Gagnidze L, Varaine F, Guerin PJ: Bleach sedimentation: an opportunity to optimize smear microscopy for tuberculosis diagnosis in settings of high prevalence of HIV. Clin Infect Dis. 2008, 46: 1710-1716. 10.1086/587891.CrossRefPubMed Bonnet M, Ramsay A, Githui W, Gagnidze L, Varaine F, Guerin PJ: Bleach sedimentation: an opportunity to optimize smear microscopy for tuberculosis diagnosis in settings of high prevalence of HIV. Clin Infect Dis. 2008, 46: 1710-1716. 10.1086/587891.CrossRefPubMed
44.
Zurück zum Zitat Wilkinson D, Sturm AW: Diagnosing tuberculosis in a resource-poor setting: the value of sputum concentration. Trans R Soc Trop Med Hyg. 1997, 91: 420-421. 10.1016/S0035-9203(97)90263-7.CrossRefPubMed Wilkinson D, Sturm AW: Diagnosing tuberculosis in a resource-poor setting: the value of sputum concentration. Trans R Soc Trop Med Hyg. 1997, 91: 420-421. 10.1016/S0035-9203(97)90263-7.CrossRefPubMed
Metadaten
Titel
Sensitivity of direct versus concentrated sputum smear microscopy in HIV-infected patients suspected of having pulmonary tuberculosis
verfasst von
Adithya Cattamanchi
David W Dowdy
J Lucian Davis
William Worodria
Samuel Yoo
Moses Joloba
John Matovu
Philip C Hopewell
Laurence Huang
Publikationsdatum
01.12.2009
Verlag
BioMed Central
Erschienen in
BMC Infectious Diseases / Ausgabe 1/2009
Elektronische ISSN: 1471-2334
DOI
https://doi.org/10.1186/1471-2334-9-53

Weitere Artikel der Ausgabe 1/2009

BMC Infectious Diseases 1/2009 Zur Ausgabe

Leitlinien kompakt für die Innere Medizin

Mit medbee Pocketcards sicher entscheiden.

Seit 2022 gehört die medbee GmbH zum Springer Medizin Verlag

Update Innere Medizin

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.