Skip to main content
Erschienen in: Cardiovascular Ultrasound 1/2005

Open Access 01.12.2005 | Review

Relevance of tissue Doppler in the quantification of stress echocardiography for the detection of myocardial ischemia in clinical practice

verfasst von: Rosa Sicari

Erschienen in: Cardiovascular Ultrasound | Ausgabe 1/2005

download
DOWNLOAD
print
DRUCKEN
insite
SUCHEN

Abstract

In the present article we review the main published data on the application of Tissue Doppler Imaging (TDI) to stress echocardiography for the detection of myocardial ischemia. TDI has been applied to stress echocardiography in order to overcome the limitations of visual analysis for myocardial ischemia. The introduction of a new technology for clinical routine use should pass through the different phases of scientific assessment from feasibility studies to large multicenter studies, from efficacy to effectiveness studies. Nonetheless the pro-technology bias plays a major role in medicine and expensive and sophisticated techniques are accepted before their real usefulness and incremental value to the available ones is assessed. Apparently, TDI is not exempted by this approach : its applications are not substantiated by strong and sound results. Nonetheless, conventional stress echocardiography for myocardial ischemia detection is heavily criticized on the basis of its subjectivity. Stress echocardiography has a long lasting history and the evidence collected over 20 years positioned it as an established tool for the detection and prognostication of coronary artery disease. The quantitative assessment of myocardial ischemia remains a scientific challenge and a clinical goal but time has not come for these newer ultrasonographic techniques which should be restricted to research laboratories.
Pharmacologic stress echocardiography is an established cost-effective technique for the detection of coronary artery disease [1]. The widespread use in the clinical practice has become possible only after evidence collected through large scale multicenter studies that demonstrated its feasibility, safety, diagnostic and prognostic accuracy [48]. According to the guidelines of ACC/AHA – pharmacological stress echocardiography with either dobutamine or dipyridamole is a class I indication (of documented effectiveness and usefulness) for the diagnosis of coronary artery disease and for the prognostic stratification of patients with known coronary artery disease [2, 3]. Its major limitation is related to a high inter-observer variability and to operator-dependent expertise that might be overcome by an appropriate training and the use of strict reading criteria [911]. Nonetheless the hunt for an objective, operator-independent technique to be applied to the conventional black and white regional wall motion analysis remains a major goal in stress echocardiography. Tissue Doppler Imaging (TDI) provides a quantitative analysis of regional myocardial function through the analysis of myocardial velocities [12, 13]. Since velocity imaging is confounded by influence from velocities in other segments, the TDI – based modalities strain and strain rate imaging have been introduced to measure regional shortening fraction and shortening rate, respectively [14] Is the application of Tissue Doppler Imaging to stress echocardiography the technique that will solve it all? According to major journals the answer is yes: the diagnostic accuracy of stress echocardiography improves with TDI when analyzed in comparison with visual assessment of wall motion analysis for the detection of inducible ischemia. Inducible ischemia quantified in a number without the approximations of visual assessment. However the enthusiasm showed by some investigators is not substantiated by scientific results. In fact, a careful analysis of the data published so far raises more doubts than certainties.

What we talk about when we talk about TDI

The TDI modalities include myocardial velocity imaging, displacement imaging, strain rate imaging and strain imaging. TDI measures velocities by the Doppler shift of reflected ultrasound. Velocities are measured in the conventional imaging planes, from apical views as longitudinal velocities and from parasternal views as radial velocities. When we employ TDI, the velocities within a myocardial segment are the net result of motion caused by contractions in that segment, motion due to tethering to other segments, and overall motion of the heart. This tethering effect is the reason why longitudinal velocities increase progressively from the apex toward the base, when measured in an apical window. Therefore ischemia in the apical region reduces myocardial velocities not only in the apex, but also in the nonischemic basal segments [15]. In practical terms, the reduction of TDI velocities in basal segments is not synonymous of reduction of function in the same segments. The opposite effect, tethering of nonischemic segments might induce increase in velocity of adjacent ischemic segments. These limitations could be overcome by the employment of strain and strain rate. Strain rate reflects how fast regional myocardial shortening or lengthening occurs measured at two locations separated by a distance. This is the reason why some authors define strain rate as a spatial velocity gradient. Strain is calculated as the time integral of strain rate and is a dimensionless quantity. The limitations of TDI have been widely outlined [1618] and this is beyond the scope of the present review but they may be synthesized into two main problems: 1 – the amplitude of the estimated velocity is dependent on the angle at which the region is imaged; 2 – the overall cardiac motion, rotation, and contraction of adjacent segments will influence regional velocity estimates. Therefore, a more critical approach to this technology would have avoided the inconsistencies of scientific results when it was applied in the clinical arena.

TDI and Stress echocardiography for myocardial ischemia detection

Feasibility studies have been published demonstrating the applicability of TDI to stress echocardiography [1927] but only few studies addressed the issue of its diagnostic accuracy in a clinical context [2831].
Cain et al [28] applied myocardial Doppler velocity to dobutamine stress echocardiography in order to assess its diagnostic accuracy when compared to conventional visual assessment. They first identified the normality ranges of myocardial velocities in patients with normal coronary arteries or with a very low pretest probability of having coronary artery disease. Then they selected 114 patient with coronary artery disease assessed at coronary angiography and evaluated the diagnostic accuracy: see Table 1 (Additional file 1). Neither overall nor vascular territory accuracy was better for myocardial velocity when compared to visual wall motion scoring. The MYDISE Study [29] was the first multicenter study on the absolute value of TDI applied to dobutamine stress echocardiography. The study enrolled 289 patients separated in 3 groups: group 1 (n = 92) healthy volunteers or patients with normal coronary arteries, group 2 (n = 48) patients with known coronary artery disease and group 3 (n = 149) consecutive patients with known or suspected coronary artery disease. Exclusion criteria were: atrial fibrillation, previous myocardial infarction (Q waves on the electrocardiogram, or akinetic segments on the resting echocardiographic images), previous revascularization, unstable angina, complete bundle branch block, significant heart valve disease, contraindication to dobutamine or atropine). The diagnostic criteria were developed by comparing 92 normal subjects with 48 patients with coronary artery disease and applied in a prospective series of 149 patients referred to stress echo laboratory for the suspect of coronary artery disease. Velocity cut-off points were tested and discarded since they did nor perform well when compared to logistic regression models, using systolic velocities at peak stress in 7 myocardial segments and after adjusting for heart rate, age and gender [29].
The main concerns refer to strict stress echocardiographic issues:1.the lack of a comparison between conventional visual assessment of regional wall motion and TDI analysis. In absolute terms the diagnostic accuracy is not striking: see Table 1 (Additional file 1) and Fig 1. Even if we accept the hypothesis of a non-inferiority analysis of TDI versus dobutamine stress echocardiography we have to take into consideration some major limitations outlined by the authors: the optimal diagnostic accuracy was obtained by using peak systolic velocity after adjusting for maximal heart rate, age and gender: "ignoring these factors reduces both sensitivity and specificity" [29]. Moreover, authors applied a very complex regression model for diagnostic accuracy assessment. A recent meta-analysis on dobutamine stress echocardiography showed an overall sensitivity of 80% and a specificity of 87% [32] (see fig 2). 2 – The extent and severity of myocardial ischemia as defined by the number of ischemic segments and the pharmacologic load is never provided. The protocol was interrupted only in the presence of secondary criteria, but never for development of myocardial ischemia since the quantitative analysis was performed off-line. It has been demonstrated that diagnostic and prognostic accuracy of stress echocardiography increases when the response is stratified in the time and space domain, i.e. number of ischemic segments, severity of ischemia induced, the time of onset of ischemia and the pharmacologic dose. 3 – the apical segments have been excluded by the analysis since the systolic velocity is not reliable making the analysis possible only in 11 segments. Nonetheless, the apex and the apical segments are very often the site of stress echocardiographic positivity unless very proximal atherosclerotic lesions are present 4 – the time for performing analysis is never reported. We are informed that the comparison between systolic velocities at rest and at peak stress is disregarded since it is time consuming and increases the potential for observer variability without increasing diagnostic accuracy. 5 -, apparently, TDI cannot be applied to patients with wall motion abnormalities at rest. The exclusion of this group of patients makes this quantitative approach quite unfeasible for routine clinical application.
Voigt et al. [30] used a more realistic approach to the application of TDI to stress echocardiography. They first demonstrated in 44 patients with known or suspected coronary artery disease that strain rate quantitatively differentiates ischemic and nonischemic regional myocardial response to dobutamine stress echocardiography [30] and compared it with conventional visual assessment. The ratio of postsystolic shortening to maximal strain was the best quantitative parameter to identify dobutamine stress induced-ischemia. This quantitative analysis improved sensitivity from 81 (visual assessment) to 86% and specificity from 82% to 89%. The statistical significance is not provided in the manuscript. Then, in the same population of 44 [31], they compared the visual assessment of wall motion abnormalities with different parameters derived from TDI application such as peak systolic velocity, systolic displacement and strain rate imaging. They employed simultaneous perfusion scintigraphy as a gold standard of myocardial ischemia. The stress echocardiographic methodology employed was not a standard one for segmentation of the left ventricle (18 segments instead of 16 or 17), pharmacologic protocol (up to 2 mg atropine instead of 1 mg) and criteria for ischemia (worsening of wall motion only in 1 segment). Also in this case, the overall accuracy is not striking: Table 1 (Additional file 1). On the basis of these results TDI reduces significantly the diagnostic accuracy of dobutamine stress echocardiography whereas strain rate imaging equals the diagnostic accuracy but it does not improve it. Interestingly enough, the sensitivities and specificities of strain rate imaging are slightly different in the two papers even though the analysis was conducted in the very same set of patients. Moreover, since coronary angiography was performed in all patients, the diagnostic accuracy should have been calculated on this real gold standard instead of perfusion scintigraphy.
In a recent [33] article Marwick et al. questioned the hypothesis that false-negative results of dobutamine stress echocardiography reflect the underinterpretation of regional left ventricular function. On the opposite, the quantitative parameters such as strain rate and peak systolic strain rate were no different between false and true negative tests, suggesting that false-negative results are related to lack of ischemia in a functional sense. On the basis of this observation, quantitative markers are unlikely to increase the sensitivity of dobutamine stress echocardiography.

Conclusions

The quantitative interpretation of stress echocardiography is not superior to expert wall motion assessment. Open issues in the quantitative analysis remain at stake: which technique to be employed among systolic velocities, strain and strain rate, the assessment of normality criteria of myocardial velocities and how to interpret their values, the management of patients with regional and global left ventricular dysfunction, the analysis of the apical segments, the complexity of the analysis in a real clinical environment, the applicability to unselected populations, its unsuitability to exercise, the most widely used stressor in the clinical practice [34]. What is presented as a breakthrough technology should have already answered to these issues and when exported into the clinical arena should have provided an incremental value to the established and more easily accessible methods. It is at this point that we are lost in clinical translation: authoritative journals provide data that cannot be transferred into the daily life of a busy stress echocardiographic laboratory, although the general message is optimistic and tend to ignore flaws and limitations of the technique [35, 36]). The advantage/disadvantage balance of a new technology should clearly be stated. The potential advantages should always outweigh the disadvantages related to the higher costs and higher complexity of analysis. Perhaps, the shape of the quantitative technology to come has not been designed yet [37, 38]. TDI is one of the tools in our hands but apparently this is not its time. At least not on the basis of this evidence.
download
DOWNLOAD
print
DRUCKEN
Literatur
1.
Zurück zum Zitat Picano E: Stress echocardiography. From pathophysiological toy to diagnostic tool. Circulation. 1992, 85: 1604-1612.CrossRefPubMed Picano E: Stress echocardiography. From pathophysiological toy to diagnostic tool. Circulation. 1992, 85: 1604-1612.CrossRefPubMed
2.
Zurück zum Zitat Cheitlin MD, Armstrong WF, Aurigemma GP, Beller GA, Bierman FZ, Davis JL, Douglas PS, Faxon DP, Gillam LD, Kimball TR, Kussmaul WG, Pearlman AS, Philbrick JT, Rakowski H, Thys DM, Antman EM, Smith SC, Alpert JS, Gregoratos G, Anderson JL, Hiratzka LF, Hunt SA, Fuster V, Jacobs AK, Gibbons RJ, Russell RO, : ACC/AHA/ASE 2003 guideline update for the clinical application of echocardiography: summary article: a report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (ACC/AHA/ASE Committee to Update the 1997 Guidelines for the Clinical Application of Echocardiography). Circulation. 2003, 108: 1146-62.CrossRefPubMed Cheitlin MD, Armstrong WF, Aurigemma GP, Beller GA, Bierman FZ, Davis JL, Douglas PS, Faxon DP, Gillam LD, Kimball TR, Kussmaul WG, Pearlman AS, Philbrick JT, Rakowski H, Thys DM, Antman EM, Smith SC, Alpert JS, Gregoratos G, Anderson JL, Hiratzka LF, Hunt SA, Fuster V, Jacobs AK, Gibbons RJ, Russell RO, : ACC/AHA/ASE 2003 guideline update for the clinical application of echocardiography: summary article: a report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (ACC/AHA/ASE Committee to Update the 1997 Guidelines for the Clinical Application of Echocardiography). Circulation. 2003, 108: 1146-62.CrossRefPubMed
3.
Zurück zum Zitat Armstrong WF, Pellikka PA, Ryan T, Crouse L, Zoghbi WA: Stress echocardiography: recommendations for performance and interpretation of stress echocardiography. J Am Soc Echocardiogr. 1998, 11: 97-104.CrossRefPubMed Armstrong WF, Pellikka PA, Ryan T, Crouse L, Zoghbi WA: Stress echocardiography: recommendations for performance and interpretation of stress echocardiography. J Am Soc Echocardiogr. 1998, 11: 97-104.CrossRefPubMed
4.
Zurück zum Zitat Picano E, Marini C, Pirelli S, Maffei S, Bolognese L, Chiriatti G, Chiarella F, Orlandini A, Seveso G, : Safety of intravenous high-dose dipyridamole echocardiography. Am J Cardiol. 1992, 70: 252-256.CrossRefPubMed Picano E, Marini C, Pirelli S, Maffei S, Bolognese L, Chiriatti G, Chiarella F, Orlandini A, Seveso G, : Safety of intravenous high-dose dipyridamole echocardiography. Am J Cardiol. 1992, 70: 252-256.CrossRefPubMed
5.
Zurück zum Zitat Picano E, Mathias W, Pingitore A, Bigi R, : Safety and tolerability of dobutamine-atropine stress echocardiography: a prospective, large scale, multicenter trial. Lancet. 1994, 344: 1190-1192.CrossRefPubMed Picano E, Mathias W, Pingitore A, Bigi R, : Safety and tolerability of dobutamine-atropine stress echocardiography: a prospective, large scale, multicenter trial. Lancet. 1994, 344: 1190-1192.CrossRefPubMed
6.
Zurück zum Zitat Picano E, Landi P, Bolognese L, Chiaranda G, Chiarella F, Seveso G, Sclavo MG, Gandolfo N, Previtali M, : Prognostic value of dipyridamole-echocardiography early after uncomplicated myocardial infarction: a large scale multicenter trial. Am J Med. 1993, 95: 608-618.CrossRefPubMed Picano E, Landi P, Bolognese L, Chiaranda G, Chiarella F, Seveso G, Sclavo MG, Gandolfo N, Previtali M, : Prognostic value of dipyridamole-echocardiography early after uncomplicated myocardial infarction: a large scale multicenter trial. Am J Med. 1993, 95: 608-618.CrossRefPubMed
7.
Zurück zum Zitat Sicari R, Picano E, Landi P, Pingitore A, Bigi R, Coletta C, Heyman J, Casazza F, Previtali M, Mathias W, Dodi C, Minardi G, Lowenstein J, Garyfallidis X, Cortigiani L, Morales MA, : The prognostic value of dobutamine-atropine stress echocardiography early after acute myocardial infarction. J Am Coll Cardiol. 1997, 29: 254-60.CrossRefPubMed Sicari R, Picano E, Landi P, Pingitore A, Bigi R, Coletta C, Heyman J, Casazza F, Previtali M, Mathias W, Dodi C, Minardi G, Lowenstein J, Garyfallidis X, Cortigiani L, Morales MA, : The prognostic value of dobutamine-atropine stress echocardiography early after acute myocardial infarction. J Am Coll Cardiol. 1997, 29: 254-60.CrossRefPubMed
8.
Zurück zum Zitat , : Prognostic value of pharmacological stress echocardiography in patients with known or suspected coronary artery disease: a prospective, large scale, multicenter, head-to-head comparison between dipyridamole and dobutamine test. J Am Coll Cardiol. 1999, 6: 1769-1777. , : Prognostic value of pharmacological stress echocardiography in patients with known or suspected coronary artery disease: a prospective, large scale, multicenter, head-to-head comparison between dipyridamole and dobutamine test. J Am Coll Cardiol. 1999, 6: 1769-1777.
9.
Zurück zum Zitat Picano E, Lattanzi F, Orlandini A, Marini C, L'Abbate A: Stress echocardiography and the human factor: the importance of being expert. J Am Coll Cardiol. 1991, 17: 666-9.CrossRefPubMed Picano E, Lattanzi F, Orlandini A, Marini C, L'Abbate A: Stress echocardiography and the human factor: the importance of being expert. J Am Coll Cardiol. 1991, 17: 666-9.CrossRefPubMed
10.
Zurück zum Zitat Imran MB, Palinkas A, Pasanisi EM, De Nes M, Picano E: Optimal reading criteria in stress echocardiography. Am J Cardiol. 2002, 90: 444-5.CrossRefPubMed Imran MB, Palinkas A, Pasanisi EM, De Nes M, Picano E: Optimal reading criteria in stress echocardiography. Am J Cardiol. 2002, 90: 444-5.CrossRefPubMed
11.
Zurück zum Zitat Hoffmann R, Lethen H, Marwick T, Arnese M, Fioretti P, Pingitore A, Picano E, Buck T, Erbel R, Flachskampf FA, Hanrath P: Analysis of interinstitutional observer agreement in interpretation of dobutamine stress echocardiograms. J Am Coll Cardiol. 1996, 27: 330-6.CrossRefPubMed Hoffmann R, Lethen H, Marwick T, Arnese M, Fioretti P, Pingitore A, Picano E, Buck T, Erbel R, Flachskampf FA, Hanrath P: Analysis of interinstitutional observer agreement in interpretation of dobutamine stress echocardiograms. J Am Coll Cardiol. 1996, 27: 330-6.CrossRefPubMed
12.
Zurück zum Zitat McDicken WN, Sutherland GR, Moran CM, Gordon LN: Colour Doppler velocity imaging of the myocardium. Ultrasound Med Biol. 1992, 18: 651-654.CrossRefPubMed McDicken WN, Sutherland GR, Moran CM, Gordon LN: Colour Doppler velocity imaging of the myocardium. Ultrasound Med Biol. 1992, 18: 651-654.CrossRefPubMed
13.
Zurück zum Zitat Sutherland GR, Stewart MJ, Groundstroem KW, Moran CM, Fleming A, Guell-Peris FJ, Riemersma RA, Fenn LN, Fox KA, McDicken WN: Colour Doppler myocardial imaging: a new technique for assessment of myocardial function. J Am Soc Echocardiogr. 1994, 7: 441-458.CrossRefPubMed Sutherland GR, Stewart MJ, Groundstroem KW, Moran CM, Fleming A, Guell-Peris FJ, Riemersma RA, Fenn LN, Fox KA, McDicken WN: Colour Doppler myocardial imaging: a new technique for assessment of myocardial function. J Am Soc Echocardiogr. 1994, 7: 441-458.CrossRefPubMed
14.
Zurück zum Zitat Urheim S, Edvardsen T, Torp H, Angelsen B, Smiseth OA: Myocardial strain by Doppler echocardiography. Validation of a new method to quantify regional myocardial function. Circulation. 2000, 102: 1158-1164.CrossRefPubMed Urheim S, Edvardsen T, Torp H, Angelsen B, Smiseth OA: Myocardial strain by Doppler echocardiography. Validation of a new method to quantify regional myocardial function. Circulation. 2000, 102: 1158-1164.CrossRefPubMed
15.
Zurück zum Zitat Stoylen A, Skjaerpe T: Systolic long axis function of the left ventricle. Global and regional information. Scand Cardiovasc J. 2003, 37: 253-258.CrossRefPubMed Stoylen A, Skjaerpe T: Systolic long axis function of the left ventricle. Global and regional information. Scand Cardiovasc J. 2003, 37: 253-258.CrossRefPubMed
17.
Zurück zum Zitat Yip G, Abraham T, Belohlavek M, Khandheria BK: Clinical applications of strain rate imaging. J Am Soc Echocardiogr. 2003, 16: 1334-42.CrossRefPubMed Yip G, Abraham T, Belohlavek M, Khandheria BK: Clinical applications of strain rate imaging. J Am Soc Echocardiogr. 2003, 16: 1334-42.CrossRefPubMed
18.
Zurück zum Zitat Smiseth OA, Stoylen A, Ihlen H: Tissue Doppler Imaging for the diagnosis of coronary artery disease. Curr Opin Cardiol. 2004, 19: 421-9.CrossRefPubMed Smiseth OA, Stoylen A, Ihlen H: Tissue Doppler Imaging for the diagnosis of coronary artery disease. Curr Opin Cardiol. 2004, 19: 421-9.CrossRefPubMed
19.
Zurück zum Zitat Derumeaux G, Ovize M, Loufoua J, Andre-Fouet X, Minaire Y, Cribier A, Letac B: Doppler tissue imaging quantitates regional wall motion during myocardial ischemia and reperfusion. Circulation. 1998, 97: 1970-7.CrossRefPubMed Derumeaux G, Ovize M, Loufoua J, Andre-Fouet X, Minaire Y, Cribier A, Letac B: Doppler tissue imaging quantitates regional wall motion during myocardial ischemia and reperfusion. Circulation. 1998, 97: 1970-7.CrossRefPubMed
20.
Zurück zum Zitat Yamada E, Garcia M, Thomas JD, Marwick TH: Myocardial Doppler velocity imaging. A quantitative technique for interpretation of dobutamine echocardiography. Am J Cardiol. 1998, 82: 806-809.CrossRefPubMed Yamada E, Garcia M, Thomas JD, Marwick TH: Myocardial Doppler velocity imaging. A quantitative technique for interpretation of dobutamine echocardiography. Am J Cardiol. 1998, 82: 806-809.CrossRefPubMed
21.
Zurück zum Zitat Tsutsui H, Uematsu M, Shimizu H, Yamagishi M, Tanaka N, Matsuda H, Miyatake K: Comparative usefulness of myocardial velocity gradient in detecting ischemic myocardium by a dobutamine challenge. J Am Coll Cardiol. 1998, 3: 89-93. 10.1016/S0735-1097(97)00430-0.CrossRef Tsutsui H, Uematsu M, Shimizu H, Yamagishi M, Tanaka N, Matsuda H, Miyatake K: Comparative usefulness of myocardial velocity gradient in detecting ischemic myocardium by a dobutamine challenge. J Am Coll Cardiol. 1998, 3: 89-93. 10.1016/S0735-1097(97)00430-0.CrossRef
22.
Zurück zum Zitat Wilkenshoff UM, Sovany A, Wigstrom L, Olstad B, Lindstrom L, Engvall J, Janerot-Sjoberg B, Wranne B, Hatle L, Sutherland GR: Regional mean systolic myocardial velocity estimation by real-time color Doppler myocardial imaging: a new technique for quantifying regional systolic function. J Am Soc Echocardiogr. 1998, 11: 683-692.CrossRefPubMed Wilkenshoff UM, Sovany A, Wigstrom L, Olstad B, Lindstrom L, Engvall J, Janerot-Sjoberg B, Wranne B, Hatle L, Sutherland GR: Regional mean systolic myocardial velocity estimation by real-time color Doppler myocardial imaging: a new technique for quantifying regional systolic function. J Am Soc Echocardiogr. 1998, 11: 683-692.CrossRefPubMed
23.
Zurück zum Zitat Pasquet A, Armstrong G, Rimmerman C, Marwick TH: Correlation of myocardial Doppler velocity response to exercise with independent evidence of myocardial ischemia by dual-isotope single-photon emission computed tomography. Am J Cardiol. 2000, 85: 536-542.CrossRefPubMed Pasquet A, Armstrong G, Rimmerman C, Marwick TH: Correlation of myocardial Doppler velocity response to exercise with independent evidence of myocardial ischemia by dual-isotope single-photon emission computed tomography. Am J Cardiol. 2000, 85: 536-542.CrossRefPubMed
24.
Zurück zum Zitat von Bibra H, Tuchnitz A, Klein A, Schneider-Eicke J, Schomig A, Schwaiger M: Regional diastolic function by pulsed Doppler myocardial mapping for the detection of left ventricular ischemia during pharmacological stress testing. J Am Coll Cardiol. 2000, 36: 444-452.CrossRefPubMed von Bibra H, Tuchnitz A, Klein A, Schneider-Eicke J, Schomig A, Schwaiger M: Regional diastolic function by pulsed Doppler myocardial mapping for the detection of left ventricular ischemia during pharmacological stress testing. J Am Coll Cardiol. 2000, 36: 444-452.CrossRefPubMed
25.
Zurück zum Zitat Cain P, Short L, Baglin T, Case C, Bosch HG, Marwick TH: Development of a fully quantitative approach to the interpretation of stress echocardiography using radial and longitudinal myocardial velocities. J Am Soc Echocardiogr. 2002, 15: 752-767.CrossRef Cain P, Short L, Baglin T, Case C, Bosch HG, Marwick TH: Development of a fully quantitative approach to the interpretation of stress echocardiography using radial and longitudinal myocardial velocities. J Am Soc Echocardiogr. 2002, 15: 752-767.CrossRef
26.
Zurück zum Zitat Kowalski M, Herregods MC, Herbots L, Weidemann F, Simmons L, Strotmann J, Dommke C, D'hooge J, Claus P, Bijnens B, Hatle L, Sutherland GR: The feasibility of ultrasonic regional strain and strain rate imaging in quantifying dobutamine stress echocardiography. Eur J Echocardiogr. 2003, 4: 81-91.CrossRefPubMed Kowalski M, Herregods MC, Herbots L, Weidemann F, Simmons L, Strotmann J, Dommke C, D'hooge J, Claus P, Bijnens B, Hatle L, Sutherland GR: The feasibility of ultrasonic regional strain and strain rate imaging in quantifying dobutamine stress echocardiography. Eur J Echocardiogr. 2003, 4: 81-91.CrossRefPubMed
27.
Zurück zum Zitat Fraser AG, Payne N, Madler CF, Janerot-Sjoberg B, Lind B, Grocott-Mason RM, Ionescu AA, Florescu N, Wilkenshoff U, Lancellotti P, Wutte M, Brodin LA, : Feasibility and reproducibility of off-line tissue Doppler measurement of regional myocardial function during dobutamine stress echocardiography. Eur J Echocardiogr. 2003, 4: 43-53.CrossRefPubMed Fraser AG, Payne N, Madler CF, Janerot-Sjoberg B, Lind B, Grocott-Mason RM, Ionescu AA, Florescu N, Wilkenshoff U, Lancellotti P, Wutte M, Brodin LA, : Feasibility and reproducibility of off-line tissue Doppler measurement of regional myocardial function during dobutamine stress echocardiography. Eur J Echocardiogr. 2003, 4: 43-53.CrossRefPubMed
28.
Zurück zum Zitat Cain P, Baglin T, Case C, Spicer D, Short L, Marwick TH: Application of tissue Doppler to interpretation of dobutamine echocardiography and comparison with quantitative coronary angiography. Am J Cardiol. 2001, 87: 525-531.CrossRefPubMed Cain P, Baglin T, Case C, Spicer D, Short L, Marwick TH: Application of tissue Doppler to interpretation of dobutamine echocardiography and comparison with quantitative coronary angiography. Am J Cardiol. 2001, 87: 525-531.CrossRefPubMed
29.
Zurück zum Zitat Madler CF, Payne N, Wilkenshoff U, Cohen A, Derumeaux GA, Pierard LA, Engvall J, Brodin LA, Sutherland GR, Fraser AG, : Non-invasive diagnosis of coronary artery disease by quantitative stress echocardiography: optimal diagnostic models using off-line tissue Doppler in the MYDISE study. Eur Heart J. 2003, 24: 1584-94.CrossRefPubMed Madler CF, Payne N, Wilkenshoff U, Cohen A, Derumeaux GA, Pierard LA, Engvall J, Brodin LA, Sutherland GR, Fraser AG, : Non-invasive diagnosis of coronary artery disease by quantitative stress echocardiography: optimal diagnostic models using off-line tissue Doppler in the MYDISE study. Eur Heart J. 2003, 24: 1584-94.CrossRefPubMed
30.
Zurück zum Zitat Voigt JU, Exner B, Schmiedehausen K, Huchzermeyer C, Reulbach U, Nixdorff U, Platsch G, Kuwert T, Daniel WG, Flachskampf FA: Strain-rate imaging during dobutamine stress echocardiography provides objective evidence of inducible ischemia. Circulation. 2003, 107: 2120-6.CrossRefPubMed Voigt JU, Exner B, Schmiedehausen K, Huchzermeyer C, Reulbach U, Nixdorff U, Platsch G, Kuwert T, Daniel WG, Flachskampf FA: Strain-rate imaging during dobutamine stress echocardiography provides objective evidence of inducible ischemia. Circulation. 2003, 107: 2120-6.CrossRefPubMed
31.
Zurück zum Zitat Voigt JU, Nixdorff U, Bogdan R, Exner B, Schmiedehausen K, Platsch G, Kuwert T, Daniel WG, Flachskampf FA: Comparison of deformation imaging and velocity imaging for detecting regional inducible ischaemia during dobutamine stress echocardiography. Eur Heart J. 2004, 25: 1517-25.CrossRefPubMed Voigt JU, Nixdorff U, Bogdan R, Exner B, Schmiedehausen K, Platsch G, Kuwert T, Daniel WG, Flachskampf FA: Comparison of deformation imaging and velocity imaging for detecting regional inducible ischaemia during dobutamine stress echocardiography. Eur Heart J. 2004, 25: 1517-25.CrossRefPubMed
32.
Zurück zum Zitat Picano E, Bedetti G, Varga A, Cseh E: The comparable diagnostic accuracies of dobutamine-stress and dipyridamole-stress echocardiographies: a meta-analysis. Coron Artery Dis. 2000, 11: 151-9.CrossRefPubMed Picano E, Bedetti G, Varga A, Cseh E: The comparable diagnostic accuracies of dobutamine-stress and dipyridamole-stress echocardiographies: a meta-analysis. Coron Artery Dis. 2000, 11: 151-9.CrossRefPubMed
33.
Zurück zum Zitat Yuda S, Fang ZY, Leano R, Marwick TH: Is quantitative interpretation likely to increase sensitivity of dobutamine stress echocardiography? A study of false-negative results. J Am Soc Echocardiogr. 2004, 17: 448-53.CrossRefPubMed Yuda S, Fang ZY, Leano R, Marwick TH: Is quantitative interpretation likely to increase sensitivity of dobutamine stress echocardiography? A study of false-negative results. J Am Soc Echocardiogr. 2004, 17: 448-53.CrossRefPubMed
34.
Zurück zum Zitat Davidavicius G, Kowalski M, Williams RI, D'hooge J, Di Salvo G, Pierre-Justin G, Claus P, Rademakers F, Herregods MC, Fraser AG, Pierard LA, Bijnens B, Sutherland GR: Can regional strain and strain rate measurement be performed during both dobutamine and exercise echocardiography, and do regional deformation responses differ with different forms of stress testing?. J Am Soc Echocardiogr. 2003, 16: 299-308.CrossRefPubMed Davidavicius G, Kowalski M, Williams RI, D'hooge J, Di Salvo G, Pierre-Justin G, Claus P, Rademakers F, Herregods MC, Fraser AG, Pierard LA, Bijnens B, Sutherland GR: Can regional strain and strain rate measurement be performed during both dobutamine and exercise echocardiography, and do regional deformation responses differ with different forms of stress testing?. J Am Soc Echocardiogr. 2003, 16: 299-308.CrossRefPubMed
35.
Zurück zum Zitat Poldermans D, Bax JJ: Dobutamine echocardiography: a diagnostic tool comes of age. Eur Heart J. 2003, 24: 1541-2.CrossRefPubMed Poldermans D, Bax JJ: Dobutamine echocardiography: a diagnostic tool comes of age. Eur Heart J. 2003, 24: 1541-2.CrossRefPubMed
36.
Zurück zum Zitat Merli E, Sutherland G: Can we quantify ischaemia during dobutamine stress echocardiography in clinical practice?. Eur Heart J. 2004, 25: 1477-1479.CrossRefPubMed Merli E, Sutherland G: Can we quantify ischaemia during dobutamine stress echocardiography in clinical practice?. Eur Heart J. 2004, 25: 1477-1479.CrossRefPubMed
37.
Zurück zum Zitat Strotmann JM, Kvitting JPE, Wilkenshoff UM, Wranne B, Hatle L, Sutherland GR, : A new approach to assess regional myocardial function. J Am Soc Echocardiogr. 1999, 300-307. Strotmann JM, Kvitting JPE, Wilkenshoff UM, Wranne B, Hatle L, Sutherland GR, : A new approach to assess regional myocardial function. J Am Soc Echocardiogr. 1999, 300-307.
38.
Zurück zum Zitat Mele D, Pedini I, Alboni P, Levine RA: Anatomic M-mode: a new technique for quantitative assessment of left ventricular size and function. Am J Cardiol. 1998, 81: 82G-85G.CrossRefPubMed Mele D, Pedini I, Alboni P, Levine RA: Anatomic M-mode: a new technique for quantitative assessment of left ventricular size and function. Am J Cardiol. 1998, 81: 82G-85G.CrossRefPubMed
Metadaten
Titel
Relevance of tissue Doppler in the quantification of stress echocardiography for the detection of myocardial ischemia in clinical practice
verfasst von
Rosa Sicari
Publikationsdatum
01.12.2005
Verlag
BioMed Central
Erschienen in
Cardiovascular Ultrasound / Ausgabe 1/2005
Elektronische ISSN: 1476-7120
DOI
https://doi.org/10.1186/1476-7120-3-2

Weitere Artikel der Ausgabe 1/2005

Cardiovascular Ultrasound 1/2005 Zur Ausgabe

Leitlinien kompakt für die Innere Medizin

Mit medbee Pocketcards sicher entscheiden.

Seit 2022 gehört die medbee GmbH zum Springer Medizin Verlag

Update Innere Medizin

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.