Skip to main content
Erschienen in: Journal of Cardiovascular Magnetic Resonance 1/2012

Open Access 01.12.2012 | Research

T1 mapping of the myocardium: Intra-individual assessment of the effect of field strength, cardiac cycle and variation by myocardial region

verfasst von: Nadine Kawel, Marcelo Nacif, Anna Zavodni, Jacquin Jones, Songtao Liu, Christopher T Sibley, David A Bluemke

Erschienen in: Journal of Cardiovascular Magnetic Resonance | Ausgabe 1/2012

Abstract

Background

Myocardial T1 relaxation time (T1 time) and extracellular volume fraction (ECV) are altered in the presence of myocardial fibrosis. The purpose of this study was to evaluate acquisition factors that may result in variation of measured T1 time and ECV including magnetic field strength, cardiac phase and myocardial region.

Methods

31 study subjects were enrolled and underwent one cardiovascular MR exam at 1.5 T and two exams at 3 T, each on separate days. A Modified Look-Locker Inversion Recovery (MOLLI) sequence was acquired before and 5, 10, 12, 20, 25 and 30 min after administration of 0.15 mmol/kg gadopentetate dimeglumine (Gd-DTPA; Magnevist) at 1.5 T (exam 1). For exam 2, MOLLI sequences were acquired at 3 T both during diastole and systole, before and after administration of Gd-DTPA (0.15 mmol/kg Magnevist).Exam 3 was identical to exam 2 except gadobenate dimeglumine was administered (Gd-BOPTA; 0.1 mmol/kg Multihance). T1 times were measured in myocardium and blood. ECV was calculated by (ΔR1myocardium/ΔR1blood)*(1-hematocrit).

Results

Before gadolinium, T1 times of myocardium and blood were significantly greater at 3 T versus 1.5 T (28% and 31% greater, respectively, p < 0.001); after gadolinium, 3 T values remained greater than those at 1.5 T (14% and 12% greater for myocardium and blood at 3 T with Gd-DTPA, respectively, p < 0.0001 and 18% and 15% greater at 3 T with Gd-BOPTA, respectively, p < 0.0001). However, ECV did not vary significantly with field strength when using the same contrast agent at equimolar dose (p = 0.2). Myocardial T1 time was 1% shorter at systole compared to diastole pre-contrast and 2% shorter at diastole compared to systole post-contrast (p < 0.01). ECV values were greater during diastole compared to systole on average by 0.01 (p < 0.01 to p < 0.0001). ECV was significantly higher for the septum compared to the non-septal myocardium for all three exams (p < 0.0001-0.01) with mean absolute differences of 0.01, 0.004, and 0.07, respectively, for exams 1, 2 and 3.

Conclusion

ECV is similar at field strengths of 1.5 T and 3 T. Due to minor variations in T1 time and ECV during the cardiac cycle and in different myocardial regions, T1 measurements should be obtained at the same cardiac phase and myocardial region in order to obtain consistent results.
Hinweise

Electronic supplementary material

The online version of this article (doi:10.​1186/​1532-429X-14-27) contains supplementary material, which is available to authorized users.

Competing interests

The authors declare they have no competing interests.

Authors’ contribution

NK: study design, data acquisition, data analysis, data interpretation, manuscript drafting; MN: study desing, data analysis, manuscript revision; AZ: data acquisition, data analysis, data interpretation, manuscript revision; JJ: data acquisition, manuscript revision; SL: study design, data acquisition; CS: study design, manuscript revision; DB principal investigator, study design, data interpretation, manuscript revision. All authors read and approved the final manuscript.

Background

Focal ischemic and non-ischemic scar of the myocardium can be reliably detected with late gadolinium enhancement (LGE) magnetic resonance (MR) [1]. Using LGE, the threshold limit of detection of focally damaged myocardium was reported to be 5% per left ventricular slice [2]. A wide variety of non-ischemic cardiomyopathies are associated with diffuse myocardial fibrosis that is characterized by an increase in collagen fibers [3, 4]. In these conditions, the LGE technique is unlikely to identify subtle alterations in diffuse fibrosis.
An alternative approach to characterize myocardial tissue composition is direct measurement of myocardial T1 relaxation time (T1 time). Alterations in T1 time have been related to variation in myocardial fibrosis: an increase in extracellular volume (associated with higher collagen content) is associated with a greater T1 time on non-contrast MR imaging. After gadolinium administration, T1 times are lower in association with greater collagen content [5]. The modified Look-Locker Inversion Recovery (MOLLI) is a convenient method to measure T1 times during a single breath-hold [6]. The extracellular volume fraction (ECV) characterizes the distribution of a gadolinium contrast agent in the myocardium relative to blood pool. ECV can be approximated from the relaxation rates (R1) of myocardium and blood pool before versus after gadolinium administration, corrected for the blood hematocrit [79].
Several technical parameters have the potential to alter T1 time and ECV values. T1 mapping has been performed at both 1.5 T [5, 6, 10, 11] and 3 T [8, 12], but intra-individual comparison in a larger sample size has not been performed yet. Myocardial blood volume varies between systole and diastole but the sensitivity of T1 mapping to these variations is also unknown. In addition, T1 time and ECV values may be measured globally or regionally; regional measurement may be affected by local alteration in field homogeneity or differences in motion related artefacts. Thus the purpose of this paper was to evaluate the effect of these parameters (field strength, region by region comparison and cardiac cycle) in order to establish the magnitude of the effect size in relationship to T1 time and ECV.

Methods

Study population

31 study participants (20 women; mean age ± SD, 28 ± 6 years; age range, 18–40 years) signed informed consent as part of an ongoing institutional review board approved study. Each participant was asked to undergo three MR scans with different field strength and contrast media, respectively. At each of the 3 separate MR examinations, clinical history and physical examination were assessed to be stable, and laboratory values were repeated. Since “normal aging” is associated with increasing diffuse fibrosis by histology, the upper limit of age to participate in the study was set to 40 years in order to establish a homogenous group of healthy volunteers and reduce the potential for age related differences [13]. All study participants were free of clinically evident illness or elevated cardiovascular risk factors. The ECG obtained prior to the first MR exam did not show an abnormality and the physical exam performed by a physician did not reveal a pathologic finding in any of the participants.

MR acquisition

Each study subject was scanned three times: once using a 1.5 T MR scanner (Avanto, Siemens Medical Solutions, Erlangen, Germany) (exam 1) and twice using a 3 T MR scanner (Verio, Siemens Medical Solutions, Erlangen, Germany) (exam 2 and 3) using a 32-channel cardiac coil for all exams. Although the MR scans are labeled as exam 1, 2 and 3, the order of the exams was influenced by scanner availability and varied in several subjects.
The MOLLI sequence was acquired in mid-ventricular short axis views pre contrast and 5, 10, 12, 20, 25 and 30 minutes after intravenous contrast administration at the 1.5 T MR scanner (exam 1). The exams at the 3 T scanner were performed at mid- to end-diastole and systole (exam 2 and 3). In order to have more time points for comparison and a longer time span after contrast administration, additional acquisitions of MOLLI were performed at 15, 35, 40 and 45 min. In preliminary tests ideal delay time for image acquisition at systole was evaluated in additional volunteers. A delay time of 150 ms (early systole) after the R-wave resulted consistently in the best image quality and was therefore used for the current study.
For the scans at 1.5 T (exam 1 ) and one of the scans performed at 3 T (exam 2), gadopentetate dimeglumine (Gd-DTPA; Magnevist, Bayer Healthcare) at a dose of 0.15 mmol/kg was administered as a bolus of 2 ml/s followed by a 30 ml saline chaser bolus at the same flow rate using a power injector. In exam 3, imaging was also performed at 3 T and a bolus of gadobenate dimeglumine (Gd-BOPTA; Multihance, Bracco Dignostics) was administered at a dose of 0.1 mmol/kg at the same flow rate. A lower dose was chosen in order to compensate for the higher relaxivity and to be consistent with our institution’s clinical routine [1416]. Scan parameters were as follows: TE/TR 1.14/2.7 ms (1.5 T), 1.03/2.4 ms (3 T); flip angle 35°; bandwidth 977 Hz/Px (1.5 T), 1002 Hz/Px (3 T); minimum TI 125–150 ms; TI increment 80 ms; slice thickness 8 mm; iPAT factor (GRAPPA) 2.
At the time of the first MR study, a complete cardiac MR examination including cine steady state free precession (SSFP) images of the entire heart in short and long axis as well as late gadolinium enhancement (LGE) sequences were obtained in order to exclude occult myocardial disease.

Image analysis

T1 maps were generated using MRmap [17]. Manual motion correction was performed when necessary. T1 maps were transferred to QMass V.7.2 (Medis Medical Imaging Systems, Netherlands). Left ventricular endocardial and epicardial contours were drawn manually while segments were defined automatically by the software after marking the border between segment 7 and 8. T1 times at the mid-cavity short axis slice of the left ventricle (American heart association [AHA] segments 7–12 [18]) were determined for each segment. An overall T1 time for each slice was also calculated, excluding images with severely impaired image quality; score 4 (see below). T1 time of the blood pool was measured by manually drawing a region of interest in the blood pool of the left ventricular cavity taking care to avoid the papillary muscles. Since a heart rate dependency for long T1 times has been demonstrated, heart rate correction was performed for the pre contrast T1 maps acquired at the 3 T scanner using custom made IDL-based software [19]. For 10 randomly chosen subjects measurements were performed by a second independent reader to assess inter-observer variability. The ECV was calculated as ΔR1myocardium/ΔR1blood * (1-hematocrit), where ΔR1myocardium = 1/T1myocardium pre contrast - 1/T1myocardium post contrast and ΔR1blood = 1/T1blood pre contrast - 1/T1blood post contrast[7]. For images acquired at systole and diastole each myocardial segment was assigned a quality score ranging from 1 to 4 (1 = perfect image quality without artifacts, 2 = slightly impaired image quality, e.g. by slight blurring, 3 = severely impaired image quality, but not involving the whole segment, 4 = severely impaired image quality involving the whole segment). An overall image score for the entire mid-cavity slice was calculated by the sum of the 6 segments.
For comparison of regional variation of T1 times the mean value of AHA segments 8 and 9 were defined as the septum and the mean value of AHA segments 7, 10, 11 and 12 were defined as non-septal myocardium. The septal and non-septal myocardium were chosen for comparison related to results of preliminary analysis that revealed a difference of T1 times in these two regions.
Left ventricular ejection fraction was obtained using Cardiac Image Modeller (CIM) (Auckland, New Zealand) [20].

Statistical analysis

Statistical analysis was performed using PASW (version 19) and SAS (version 9.2) statistical software. A p-value <0.05 was considered to be statistically significant.
The continuous variables T1 time (ms) and ECV are expressed as mean ± SD. Myocardial T1 times obtained before contrast administration were compared using a paired t-test. A linear mixed-model analysis was performed with the log transformed T1 times for blood and myocardium to compare T1 time among groups over time after contrast administration. Linear mixed-model analysis with the ECV values to identify group related differences was also performed. Both sets of models evaluated group, time and group-by-time interaction by study subjects as random effects.
Quality scores for images acquired at systole and diastole were compared using the Wilcoxon signed rank test.
To compare inter-observer agreement, analysis according to Bland-Altman was performed and Pearson’s correlation coefficient (poor agreement = 0; slight = 0.01-0.20; fair = 0.21-0.40; moderate = 0.41-0.60; good = 0.61-0.80, and excellent = 0.81-1.00 agreement) and the intraclass correlation coefficient (ICC) using a two-way random model (ICC < 0.40 = poor; ICC ≥ 0.40 to 0.75 = fair to good; ICC > 0.75 = excellent agreement) were calculated.

Results

Of 31 study participants, 28 completed the scan at 1.5 T with Gd-DTPA (exam 1), 29 completed the scan at 3 T with Gd-DTPA (exam 2) and 24 completed the scan at 3 T with Gd-BOPTA (exam 3). 23 volunteers completed all three exams. Average delay between one exam and the following exam was 58 ± 31 days. Of the 8 volunteers who did not complete all scans, 6 volunteers refused to complete the three scans for various reasons, in one subject the scan had to be terminated after contrast injection (Gd-BOPTA) related to nausea and vomiting and one volunteer was pregnant.
Of 1176 segments evaluated on images acquired at 1.5 T, 27 (2.3%) had to be excluded due to severe artifacts (quality score 4). Similar percentages of segments were excluded of the other exams: 39/1914 (2%) (3 T Gd-DTPA, diastole), 27/1914 (1.6%) (3 T Gd-DTPA, systole), 15/1566 (1.0%) (3 T Gd-BOPTA, diastole) and 5/1368 (0.4%) (3 T Gd-BOPTA, systole). In 2 volunteers scanned at 3 T with Gd-BOPTA (exam 3), 5 maps could not be obtained due to a scanner malfunction and in one other volunteer images were acquired at diastole only.
Global and regional left ventricular function was normal in all volunteers (mean ejection fraction ± SD: 60.4 ± 4.5%). Late gadolinium enhancement (LGE) MR images did not show myocardial scar in any of the participants. Table 1 shows further characteristics of the study participants.
Table 1
Characteristics of study participants (n = 31)
Characteristics
1.5 T Gd-DTPA
(exam 1)
3 T Gd-DTPA
(exam 2)
3 T Gd-BOPTA
(exam 3)
Subjects, n
28
29
24
Age, years
27.6 ± 5.8
28.3 ± 6.1
28.1 ± 5.9
Men, n (%)
10 (36)
9 (31)
8 (33)
Weight in kg, mean ± SD
72 ± 13
72 ± 14
69 ± 13
Hematocrit in %, mean ± SD
41 ± 6
40 ± 4
40 ± 4
Creatinine in mg/dl, mean ± SD
0.7 ± 0.2
0.8 ± 0.2
0.8 ± 0.2
Gadolinium in ml, mean ± SD
21.5 ± 4
21.5 ± 4
13.9 ± 3
Of note, differences of T1 times are shown as absolute difference in milliseconds and/or relative difference in percent. ECV is reported as a ratio. Differences in ECV are mentioned as absolute differences of the ratio and/or the relative difference in percent.

Intra-individual evaluation of ECV and T1 time at 1.5 T versus 3 T

Comparison of 1.5 T and 3 T was performed for volunteers who completed all three exams (n = 23). As expected, before contrast administration, T1 time of myocardium was 28% higher and T1 time of blood was 31% higher for exams acquired at 3 T compared to 1.5 T (p < 0.0001).
For the measurements between 5 and 30 minutes after contrast administration, mean myocardial T1 time was on average 67 ms (14%) higher for exam 2 (3 T, Gd-DTPA) compared to exam 1 (1.5 T, Gd-DTPA) and it was 88 ms (18%) higher for exam 3 (3 T, Gd-BOPTA) compared to exam 1 (1.5 T, Gd-DTPA) (p < 0.0001, both comparisons). Results for T1 time of blood were similar: T1 time of blood was on average 43 ms (12%) higher for exam 2 (3 T, Gd-DTPA) compared to exam 1 (1.5 T, Gd-DTPA) and it was 52 ms (15%) higher for exam 3 compared to exam 1 (1.5 T, Gd-DTPA) (p < 0.0001, both comparisons) (Figure 1).
There was no significant difference for ECV between 1.5 T and 3 T using Gd-DTPA (p = 0.2) while there was a significant difference for the comparison between 1.5 T using Gd-DTPA (exam 1) and 3 T using Gd-BOPTA (exam 3) (p = 0.01). ECV obtained at 5, 10 and 12 minutes was on average 0.006 (2.0%) lower for exam 3 (3 T, Gd-BOPTA) compared to exam 1 (1.5 T, Gd-DTPA) while mean ECV obtained at 20, 25 and 30 minutes was 0.005 (2.1%) higher for exam 3 (3 T, Gd-BOPTA) compared to exam 1 (1.5 T, Gd-DTPA) (Figure 1).
Intra-individual variation in ECV was rather large. Including all 24 subjects who completed all three exams, the mean absolute difference between ECV values of all time points was between 0.09 and 0.1.

Effect of cardiac phase

In exam 2 (3 T, Gd-DTPA) and exam 3 (3 T, Gd-BOPTA) MOLLI was acquired at mid- to end-diastole and early systole. In these subjects, we assessed the variation in T1 time and ECV during diastole compared to systole.
Prior to contrast administration myocardial T1 time was significantly lower at systole compared to diastole for both exams at 3 T (p < 0.01), consistent with lower myocardial blood volume at systole. However, the mean difference averaged over all time points was only 15 ms (about 1% change in the mean T1 time) for exam 2 and 5 ms (about 0.4%) for exam 3 (Figure 2).
After contrast administration, mean myocardial T1 time was significantly lower at diastole compared to systole (p < 0.01) but mean differences in myocardial T1 time between cardiac phases were small: 12 ms (2%) for Gd-DTPA and 13 ms (2%) for Gd-BOPTA. As expected, blood pool showed no variation with cardiac cycle (p = 0.3 [Gd-DTPA]; p = 0.2 [Gd-BOPTA]). Consequently ECV varied as a function of cardiac cycle; although mean group differences for ECV over repeated measurements were statistically significant (p < 0.0001 [Gd-DTPA]; p < 0.01 [Gd-BOPTA]), the mean ECV was on average only 0.01 higher in diastole compared to systole (Figure 2).
Image quality scores per segment are displayed in Table 2. For the exam with Gd-DTPA, the overall image quality score for the entire slice of 280 T1 maps of images acquired at diastole was 12.4 ± 2.9 and 11.7 ± 3.3 for images acquired in systole (p < 0.0001). For 230 T1 maps of images acquired with Gd-BOPTA the mean image quality was 11.4 ± 2.3 for diastole and 10.1 ± 2.7 for systole (p < 0.0001) (lower score indicated better image quality).
Table 2
Image quality score
 
Exam 2 (3 T Gd-DTPA)
Exam 3 (3 T Gd-BOPTA)
diastole
systole
diastole
systole
Seg 7 (anterior)
2.2 ± 0.7
2.2 ± 0.6
2.0 ± 0.6
2.0 ± 0.5
Seg 8 (anteroseptal)
1.7 ± 0.6
1.6 ± 0.6
1.7 ± 0.5
1.5 ± 0.6
Seg 9 (inferoseptal)
1.6 ± 0.6
1.5 ± 0.6
1.4 ± 0.6
1.3 ± 0.5
Seg 10 (inferior)
2.4 ± 0.7
2.1 ± 0.8
2.1 ± 0.6
1.8 ± 0.6
Seg 11 (inferolateral)
2.1 ± 0.7
2.2 ± 0.8
2.2 ± 0.6
1.8 ± 0.7
Seg 12 (interolateral)
2.2 ± 0.7
2.1 ± 0.8
2.0 ± 0.5
1.7 ± 0.7
Lower image quality score indicates better image quality; Seg = segment.
Regionally, the septal segments 8 and 9 had the best image quality score over all scans and time points. In the majority of scans segment 12 (anterolateral) had the third best quality score. Overall, segment 7 (anterior) and 11 (inferolateral) had the worst quality scores (Table 2).

Regional variation in T1 time and ECV: Septum versus non-septal myocardium

Before contrast administration T1 times were statistically significantly longer for the septum compared to the non-septal myocardium at 3 T (p < 0.01 [exam 2] and p < 0.0001 [exam 3], respectively) but not at 1.5 T (p = 0.09). However, T1 time at 3 T was on average only 19 ms (1.5%) (exam 2) and 24 ms (1.9%) (exam 3), respectively longer for the septal myocardium.
Mean T1 time after contrast administration differed statistically significantly between the two groups (septum versus non-septal myocardium) for the exams performed at 1.5 T (p = 0.02) but not for the exams at 3 T (p = 0.4 [Gd-DTPA]; p = 0.6 [Gd-BOPTA]). Although differences in T1 time after contrast administration between the septum and the non-septal myocardium were statistically significant at 1.5 T, the mean T1 time of the non-septal myocardium was on average only 9 ms (or about 2%) longer compared to the septal myocardium. Differences tended to be larger for the acquisitions in the first 20 minutes and decreased after 20 minutes (Figure 3).
ECV was also significantly higher for the septal compared to the non-septal myocardium for all three exams: 1.5 T Gd-DTPA (p < 0.001), 3 T Gd-DTPA (p < 0.01), and 3 T Gd-BOPTA (p < 0.0001). Mean absolute differences were: 0.01, 0.004, and 0.07, respectively (Figure 3).

Inter-observer agreement

Inter-observer agreement was excellent for measurements of T1 time for myocardium and blood at both field strengths, and for measurements obtained at 3 T in systole and diastole, with ICC ranging from 0.984 to 1.0 (Table 3).
Table 3
Inter-observer agreement
 
Pearson’s correlation coefficient (r)
p
Bias
95% Limits of agreement
ICC
1.5 T Gd-DTPA
     
T1 myocardium
1.0
<0.0001
2.3
−15.7 to 20.2
0.999
T1 blood
1.0
<0.0001
0.7
−18.8 to 20.2
1.0
3 T Gd-DTPA diastole
     
T1 myocardium
1.0
<0.0001
−0.5
−8.5 to 7.4
1.0
T1 blood
1.0
<0.0001
−0.8
−36.7 to 35.1
0.999
3 T Gd-DTPA systole
     
T1 myocardium
1.0
<0.0001
5.5
−24.0 to 35.0
0.998
T1 blood
1.0
<0.0001
−1.3
−16.4 to 13.7
1.0
3 T Gd-BOPTA diastole
     
T1 myocardium
1.0
<0.0001
−7.8
−43.2 to 27.6
0.997
T1 blood
1.0
<0.0001
−7.2
−56.3 to 41.9
0.998
3 T Gd-BOPTA systole
     
T1 myocardium
0.97
<0.0001
−6.4
−80.4 to 67.7
0.984
T1 blood
0.99
<0.0001
−10.9
−135.2 to 113.5
0.993
ICC = intraclass correlation coefficient.

Discussion

In the current study we demonstrated that ECV does not vary significantly by field strength (1.5 T versus 3 T) when using the same contrast agent at equimolar dose. Myocardial T1 times and ECV vary significantly between systole and diastole and ECV values vary with myocardial region. However, absolute differences were small and might not be clinically relevant or detectable on an individual patient basis.

Intra-individual evaluation of ECV and T1 time at 1.5 T versus 3 T

As expected and according to previous publications, T1 times for blood and myocardium varied significantly with field strength [21, 22]. Sharma et al. also described a significant difference for T1 time of myocardium but not for blood [8]. The authors explain the difference by a lesser field dependence due to greater free water content and shorter molecular correlation times in blood. This discrepancy compared to the current study and previous publications might be attributed to a relatively low number of study subjects in the study by Sharma et al. (n = 10) receiving two different contrast doses and a different pulse sequence. Sharma et al. also found a statistically significant difference for the partition coefficient for values obtained at 1.5 T and 3 T. The partition coefficient is similar to ECV used in this study, but is not corrected for hematocrit. In the current study ECV did not vary with field strength when the same contrast agent was used at equimolar dose.
The variation of ECV between Gd-DTPA and Gd-BOPTA can be explained by the higher relaxivity of Gd-BOPTA related to a weak protein binding capacity. In plasma the contrast agent mainly binds to human serum albumin. Since albumin is present in blood, the distribution between blood and myocardium is expected to be different between the two contrast agents whereas pre-contrast values are the same. This should affect the ratio of the change in relaxation rate of myocardium and blood and could explain the difference in ECV values [23].
Of note is the relatively large inter-individual variation in ECV values. The mean absolute inter-individual variation was 0.09 to 0.1. This will hamper establishment of cut-off values between normal and disease and needs to be further evaluated in future studies.

Effect of cardiac phase

Differences in T1 time between systole and diastole are mainly related to changes in the myocardial blood volume. In the rabbit heart vascular space was measured as 0.15 ± 0.04 ml/g [24]. In the left ventricular free wall of rat hearts Judd et al. demonstrated a decrease in intramyocardial blood volume by 42% from diastole to systole. Blood volume changed from 8.6 ± 1.3 ml/100 g to 5.0 ± 0.7 ml/100 g [25]. Changes from diastole to systole in capillary volumes are larger compared to pre and post capillary microvessels (decrease of 32% versus 48%) and changes in the septum have been shown to be larger compared to the lateral wall (decrease of 25% versus 18.2%) [26, 27]. Differences in reported numbers are related to different measurement techniques and study subjects as well as in vivo versus ex vivo studies.
In a recent publication Wansapura et al. evaluated cyclic changes of myocardial blood volume by T1 mapping in seven volunteers without contrast administration using a dual flip angle technique and a fast gradient echo sequence [28]. They describe a decrease in T1 time from diastole to systole of 70% in the septum and 43% in the lateral wall. With a decrease in T1 time from diastole to systole of about 1% prior to contrast administration and an increase from diastole to systole of about 2% after contrast administration, differences in T1 time of the current study are substantially smaller compared to measurements reported by Wansapura et al. Although myocardial blood volume changes significantly from diastole to systole, the blood volume accounts for a small fraction of the entire myocardial volume only and decreases by 3.6 ml/100 g from diastole to systole, assuming that conditions of the human heart are similar to the animal heart [25]. The small change of T1 time from diastole to systole in the current study concurs with the change in myocardial blood volume as a relatively small fraction of the entire myocardial volume.
Since the myocardium is thicker during systole and images yielded a better image quality in the current study, it may be advantageous to measure T1 times at systole. In the current study, absolute variation of myocardial T1 time and ECV values over the cardiac cycle was significant. However, absolute changes were small (<2%) and might not be clinically relevant.
During a breath hold of about 15 seconds, variation of heart rate might be substantial and changes of up to 20 beats per minute may occur. This variation in heart rate mainly affects duration of diastole and therefore might cause motion related artifacts. In the current study in contrary to the general opinion and despite the rather long acquisition window of about 150 ms of the MOLLI sequence, images acquired at early systole resulted in a constantly good image quality (Figure 4).

Regional variation in T1 time and ECV measurement: Septum versus non-septal myocardium

In animal experiments, a high regional variability of the myocardial partition coefficient has been demonstrated [29]. Since the septum shows less motion compared to the remaining myocardium, it has been postulated that T1 times might vary with myocardial region due to less motion related measurement errors. The remaining myocardium is also more prone to susceptibility artifacts, particular at 3 T. Further, regional differences in myocardial perfusion might cause variation in T1 time [30]. Messroghli et al. compared T1 times of the base, mid-cavity and apical level of the left ventricle before contrast administration and did not find regional differences [6]. Statistical analysis for differences between the septum and the non-septal myocardium was not performed. In the current study, ECV values varied between regions for both 1.5 T and 3 T with absolute mean differences ranging from 0.004 to 0.07. Therefore, for comparison measurements should be obtained in the same region or cover the entire myocardium of a slice, respectively.

Strength and limitations

Acquisitions of the MOLLI sequence at systole were performed after the diastolic acquisition. However, the time delay between both acquisitions was in the order of 20 seconds only and T1 times of blood remained stable during this period. Thus this short delay between sequences, although systematic, is unlikely to be responsible for T1 and ECV differences between systole and diastole.
A strength of the study is the higher number of subjects compared to recent publications and the fact that the same subject was scanned three times in order to obtain intra-individual comparison. Since T1 times and ECV values show a relatively high variation between healthy subjects, a study based on intra-individual comparison assures that the detected differences are not related to inter-individual variation.

Conclusions

T1 mapping is a valuable MR technique to measure myocardial T1 time and calculate the ECV. When using the same contrast at equimolar dose, ECV does not vary with field strength. MOLLI permits measurement of small changes in myocardial blood volume over the cardiac cycle. Further, minor but statistically significant differences in ECV between myocardial regions were detected. But absolute differences were small and might not be relevant depending on the clinical or research application. However, to avoid addition of small differences, repetitive measurements should be obtained in the same cardiac phase and including the same myocardial region.

Acknowledgement

This study was supported by the NIH intramural research program.
Open Access This article is published under license to BioMed Central Ltd. This is an Open Access article is distributed under the terms of the Creative Commons Attribution License ( https://​creativecommons.​org/​licenses/​by/​2.​0 ), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Competing interests

The authors declare they have no competing interests.

Authors’ contribution

NK: study design, data acquisition, data analysis, data interpretation, manuscript drafting; MN: study desing, data analysis, manuscript revision; AZ: data acquisition, data analysis, data interpretation, manuscript revision; JJ: data acquisition, manuscript revision; SL: study design, data acquisition; CS: study design, manuscript revision; DB principal investigator, study design, data interpretation, manuscript revision. All authors read and approved the final manuscript.
Anhänge

Authors’ original submitted files for images

Literatur
1.
Zurück zum Zitat Cummings KW, Bhalla S, Javidan-Nejad C, Bierhals AJ, Gutierrez FR, Woodard PK: A pattern-based approach to assessment of delayed enhancement in nonischemic cardiomyopathy at MR imaging. Radiographics. 2009, 29: 89-103. 10.1148/rg.291085052.CrossRefPubMed Cummings KW, Bhalla S, Javidan-Nejad C, Bierhals AJ, Gutierrez FR, Woodard PK: A pattern-based approach to assessment of delayed enhancement in nonischemic cardiomyopathy at MR imaging. Radiographics. 2009, 29: 89-103. 10.1148/rg.291085052.CrossRefPubMed
2.
Zurück zum Zitat Nassenstein K, Breuckmann F, Bucher C, Kaiser G, Konorza T, Schafer L, Konietzka I, de Greiff A, Heusch G, Erbel R, Barkhausen J: How much myocardial damage is necessary to enable detection of focal late gadolinium enhancement at cardiac MR imaging?. Radiology. 2008, 249: 829-835. 10.1148/radiol.2493080457.CrossRefPubMed Nassenstein K, Breuckmann F, Bucher C, Kaiser G, Konorza T, Schafer L, Konietzka I, de Greiff A, Heusch G, Erbel R, Barkhausen J: How much myocardial damage is necessary to enable detection of focal late gadolinium enhancement at cardiac MR imaging?. Radiology. 2008, 249: 829-835. 10.1148/radiol.2493080457.CrossRefPubMed
3.
Zurück zum Zitat Jellis C, Martin J, Narula J, Marwick TH: Assessment of nonischemic myocardial fibrosis. J Am Coll Cardiol. 2010, 56: 89-97. 10.1016/j.jacc.2010.02.047.CrossRefPubMed Jellis C, Martin J, Narula J, Marwick TH: Assessment of nonischemic myocardial fibrosis. J Am Coll Cardiol. 2010, 56: 89-97. 10.1016/j.jacc.2010.02.047.CrossRefPubMed
4.
Zurück zum Zitat Mewton N, Liu CY, Croisille P, Bluemke D, Lima JA: Assessment of myocardial fibrosis with cardiovascular magnetic resonance. J Am Coll Cardiol. 2011, 57: 891-903. 10.1016/j.jacc.2010.11.013.CrossRefPubMed Mewton N, Liu CY, Croisille P, Bluemke D, Lima JA: Assessment of myocardial fibrosis with cardiovascular magnetic resonance. J Am Coll Cardiol. 2011, 57: 891-903. 10.1016/j.jacc.2010.11.013.CrossRefPubMed
5.
Zurück zum Zitat Iles L, Pfluger H, Phrommintikul A, Cherayath J, Aksit P, Gupta SN, Kaye DM, Taylor AJ: Evaluation of diffuse myocardial fibrosis in heart failure with cardiac magnetic resonance contrast-enhanced T1 mapping. J Am Coll Cardiol. 2008, 52: 1574-1580. 10.1016/j.jacc.2008.06.049.CrossRefPubMed Iles L, Pfluger H, Phrommintikul A, Cherayath J, Aksit P, Gupta SN, Kaye DM, Taylor AJ: Evaluation of diffuse myocardial fibrosis in heart failure with cardiac magnetic resonance contrast-enhanced T1 mapping. J Am Coll Cardiol. 2008, 52: 1574-1580. 10.1016/j.jacc.2008.06.049.CrossRefPubMed
6.
Zurück zum Zitat Messroghli DR, Plein S, Higgins DM, Walters K, Jones TR, Ridgway JP, Sivananthan MU: Human myocardium: single-breath-hold MR T1 mapping with high spatial resolution–reproducibility study. Radiology. 2006, 238: 1004-1012. 10.1148/radiol.2382041903.CrossRefPubMed Messroghli DR, Plein S, Higgins DM, Walters K, Jones TR, Ridgway JP, Sivananthan MU: Human myocardium: single-breath-hold MR T1 mapping with high spatial resolution–reproducibility study. Radiology. 2006, 238: 1004-1012. 10.1148/radiol.2382041903.CrossRefPubMed
7.
Zurück zum Zitat Arheden H, Saeed M, Higgins CB, Gao DW, Ursell PC, Bremerich J, Wyttenbach R, Dae MW, Wendland MF: Reperfused rat myocardium subjected to various durations of ischemia: estimation of the distribution volume of contrast material with echo-planar MR imaging. Radiology. 2000, 215: 520-528.CrossRefPubMed Arheden H, Saeed M, Higgins CB, Gao DW, Ursell PC, Bremerich J, Wyttenbach R, Dae MW, Wendland MF: Reperfused rat myocardium subjected to various durations of ischemia: estimation of the distribution volume of contrast material with echo-planar MR imaging. Radiology. 2000, 215: 520-528.CrossRefPubMed
8.
Zurück zum Zitat Sharma P, Socolow J, Patel S, Pettigrew RI, Oshinski JN: Effect of Gd-DTPA-BMA on blood and myocardial T1 at 1.5T and 3T in humans. J Magn Reson Imaging. 2006, 23: 323-330. 10.1002/jmri.20504.CrossRefPubMed Sharma P, Socolow J, Patel S, Pettigrew RI, Oshinski JN: Effect of Gd-DTPA-BMA on blood and myocardial T1 at 1.5T and 3T in humans. J Magn Reson Imaging. 2006, 23: 323-330. 10.1002/jmri.20504.CrossRefPubMed
9.
Zurück zum Zitat Messroghli DR, Nordmeyer S, Dietrich T, Dirsch O, Kaschina E, Savvatis K, Oh-Ici D, Klein C, Berger F, Kuehne T: Assessment of diffuse myocardial fibrosis in rats using small animal look-locker inversion recovery T1 mapping. Circ Cardiovasc Imaging. 2011, 4 (6): 636-640. 10.1161/CIRCIMAGING.111.966796.CrossRefPubMed Messroghli DR, Nordmeyer S, Dietrich T, Dirsch O, Kaschina E, Savvatis K, Oh-Ici D, Klein C, Berger F, Kuehne T: Assessment of diffuse myocardial fibrosis in rats using small animal look-locker inversion recovery T1 mapping. Circ Cardiovasc Imaging. 2011, 4 (6): 636-640. 10.1161/CIRCIMAGING.111.966796.CrossRefPubMed
10.
Zurück zum Zitat Flacke SJ, Fischer SE, Lorenz CH: Measurement of the gadopentetate dimeglumine partition coefficient in human myocardium in vivo: normal distribution and elevation in acute and chronic infarction. Radiology. 2001, 218: 703-710.CrossRefPubMed Flacke SJ, Fischer SE, Lorenz CH: Measurement of the gadopentetate dimeglumine partition coefficient in human myocardium in vivo: normal distribution and elevation in acute and chronic infarction. Radiology. 2001, 218: 703-710.CrossRefPubMed
11.
Zurück zum Zitat Schelbert EB, Testa SM, Meier CG, Ceyrolles WJ, Levenson JE, Blair AJ, Kellman P, Jones BL, Ludwig DR, Schwartzman D: Myocardial extravascular extracellular volume fraction measurement by gadolinium cardiovascular magnetic resonance in humans: slow infusion versus bolus. J Cardiovasc Magn Reson. 2011, 13: 16-10.1186/1532-429X-13-16.PubMedCentralCrossRefPubMed Schelbert EB, Testa SM, Meier CG, Ceyrolles WJ, Levenson JE, Blair AJ, Kellman P, Jones BL, Ludwig DR, Schwartzman D: Myocardial extravascular extracellular volume fraction measurement by gadolinium cardiovascular magnetic resonance in humans: slow infusion versus bolus. J Cardiovasc Magn Reson. 2011, 13: 16-10.1186/1532-429X-13-16.PubMedCentralCrossRefPubMed
12.
Zurück zum Zitat Broberg CS, Chugh SS, Conklin C, Sahn DJ, Jerosch-Herold M: Quantification of diffuse myocardial fibrosis and its association with myocardial dysfunction in congenital heart disease. Circ Cardiovasc Imaging. 2010, 3: 727-734. 10.1161/CIRCIMAGING.108.842096.PubMedCentralCrossRefPubMed Broberg CS, Chugh SS, Conklin C, Sahn DJ, Jerosch-Herold M: Quantification of diffuse myocardial fibrosis and its association with myocardial dysfunction in congenital heart disease. Circ Cardiovasc Imaging. 2010, 3: 727-734. 10.1161/CIRCIMAGING.108.842096.PubMedCentralCrossRefPubMed
13.
Zurück zum Zitat Pugh KG, Wei JY: Clinical implications of physiological changes in the aging heart. Drugs Aging. 2001, 18: 263-276. 10.2165/00002512-200118040-00004.CrossRefPubMed Pugh KG, Wei JY: Clinical implications of physiological changes in the aging heart. Drugs Aging. 2001, 18: 263-276. 10.2165/00002512-200118040-00004.CrossRefPubMed
14.
Zurück zum Zitat Pintaske J, Martirosian P, Graf H, Erb G, Lodemann KP, Claussen CD, Schick F: Relaxivity of Gadopentetate Dimeglumine (Magnevist), Gadobutrol (Gadovist), and Gadobenate Dimeglumine (MultiHance) in human blood plasma at 0.2, 1.5, and 3 Tesla. Invest Radiol. 2006, 41: 213-221. 10.1097/01.rli.0000197668.44926.f7.CrossRefPubMed Pintaske J, Martirosian P, Graf H, Erb G, Lodemann KP, Claussen CD, Schick F: Relaxivity of Gadopentetate Dimeglumine (Magnevist), Gadobutrol (Gadovist), and Gadobenate Dimeglumine (MultiHance) in human blood plasma at 0.2, 1.5, and 3 Tesla. Invest Radiol. 2006, 41: 213-221. 10.1097/01.rli.0000197668.44926.f7.CrossRefPubMed
15.
Zurück zum Zitat Rohrer M, Bauer H, Mintorovitch J, Requardt M, Weinmann HJ: Comparison of magnetic properties of MRI contrast media solutions at different magnetic field strengths. Invest Radiol. 2005, 40: 715-724. 10.1097/01.rli.0000184756.66360.d3.CrossRefPubMed Rohrer M, Bauer H, Mintorovitch J, Requardt M, Weinmann HJ: Comparison of magnetic properties of MRI contrast media solutions at different magnetic field strengths. Invest Radiol. 2005, 40: 715-724. 10.1097/01.rli.0000184756.66360.d3.CrossRefPubMed
16.
Zurück zum Zitat van der Molen AJ, Bellin MF: Extracellular gadolinium-based contrast media: differences in diagnostic efficacy. Eur J Radiol. 2008, 66: 168-174. 10.1016/j.ejrad.2008.02.010.CrossRefPubMed van der Molen AJ, Bellin MF: Extracellular gadolinium-based contrast media: differences in diagnostic efficacy. Eur J Radiol. 2008, 66: 168-174. 10.1016/j.ejrad.2008.02.010.CrossRefPubMed
17.
Zurück zum Zitat Messroghli DR, Rudolph A, Abdel-Aty H, Wassmuth R, Kuhne T, Dietz R, Schulz-Menger J: An open-source software tool for the generation of relaxation time maps in magnetic resonance imaging. BMC Med Imaging. 2010, 10: 16-10.1186/1471-2342-10-16.PubMedCentralCrossRefPubMed Messroghli DR, Rudolph A, Abdel-Aty H, Wassmuth R, Kuhne T, Dietz R, Schulz-Menger J: An open-source software tool for the generation of relaxation time maps in magnetic resonance imaging. BMC Med Imaging. 2010, 10: 16-10.1186/1471-2342-10-16.PubMedCentralCrossRefPubMed
18.
Zurück zum Zitat Cerqueira MD, Weissman NJ, Dilsizian V, Jacobs AK, Kaul S, Laskey WK, Pennell DJ, Rumberger JA, Ryan T, Verani MS: Standardized myocardial segmentation and nomenclature for tomographic imaging of the heart: a statement for healthcare professionals from the Cardiac Imaging Committee of the Council on Clinical Cardiology of the American Heart Association. Circulation. 2002, 105: 539-542. 10.1161/hc0402.102975.CrossRefPubMed Cerqueira MD, Weissman NJ, Dilsizian V, Jacobs AK, Kaul S, Laskey WK, Pennell DJ, Rumberger JA, Ryan T, Verani MS: Standardized myocardial segmentation and nomenclature for tomographic imaging of the heart: a statement for healthcare professionals from the Cardiac Imaging Committee of the Council on Clinical Cardiology of the American Heart Association. Circulation. 2002, 105: 539-542. 10.1161/hc0402.102975.CrossRefPubMed
19.
Zurück zum Zitat Lee JJ, Liu S, Nacif MS, Ugander M, Kawel N, Sibley CT, Kellman P, Arai A, Bluemke DA: Myocardial T1 and Extracellular Volume Fraction Mapping at 3 Tesla. J Cardiovasc Magn Reson: Offic J Soc Cardiovasc Magn Reson. 2011, 13: 75-CrossRef Lee JJ, Liu S, Nacif MS, Ugander M, Kawel N, Sibley CT, Kellman P, Arai A, Bluemke DA: Myocardial T1 and Extracellular Volume Fraction Mapping at 3 Tesla. J Cardiovasc Magn Reson: Offic J Soc Cardiovasc Magn Reson. 2011, 13: 75-CrossRef
20.
Zurück zum Zitat Young AA, Cowan BR, Thrupp SF, Hedley WJ, Dell'Italia LJ: Left ventricular mass and volume: fast calculation with guide-point modeling on MR images. Radiology. 2000, 216: 597-602.CrossRefPubMed Young AA, Cowan BR, Thrupp SF, Hedley WJ, Dell'Italia LJ: Left ventricular mass and volume: fast calculation with guide-point modeling on MR images. Radiology. 2000, 216: 597-602.CrossRefPubMed
21.
Zurück zum Zitat Henriksen O, de Certaines JD, Spisni A, Cortsen M, Muller RN, Ring PB: In vivo field dependence of proton relaxation times in human brain, liver and skeletal muscle: a multicenter study. Magn Reson Imaging. 1993, 11: 851-856. 10.1016/0730-725X(93)90202-O.CrossRefPubMed Henriksen O, de Certaines JD, Spisni A, Cortsen M, Muller RN, Ring PB: In vivo field dependence of proton relaxation times in human brain, liver and skeletal muscle: a multicenter study. Magn Reson Imaging. 1993, 11: 851-856. 10.1016/0730-725X(93)90202-O.CrossRefPubMed
22.
Zurück zum Zitat Stanisz GJ, Odrobina EE, Pun J, Escaravage M, Graham SJ, Bronskill MJ, Henkelman RM: T1, T2 relaxation and magnetization transfer in tissue at 3 T. Magn Reson Med. 2005, 54: 507-512. 10.1002/mrm.20605.CrossRefPubMed Stanisz GJ, Odrobina EE, Pun J, Escaravage M, Graham SJ, Bronskill MJ, Henkelman RM: T1, T2 relaxation and magnetization transfer in tissue at 3 T. Magn Reson Med. 2005, 54: 507-512. 10.1002/mrm.20605.CrossRefPubMed
23.
Zurück zum Zitat Spinazzi A, Lorusso V, Pirovano G, Taroni P, Kirchin M, Davies A: Multihance clinical pharmacology: biodistribution and MR enhancement of the liver. Acad Radiol. 1998, 5 (Suppl 1): S86-S89. discussion S93-84CrossRefPubMed Spinazzi A, Lorusso V, Pirovano G, Taroni P, Kirchin M, Davies A: Multihance clinical pharmacology: biodistribution and MR enhancement of the liver. Acad Radiol. 1998, 5 (Suppl 1): S86-S89. discussion S93-84CrossRefPubMed
24.
Zurück zum Zitat Gonzalez F, Bassingthwaighte JB: Heterogeneities in regional volumes of distribution and flows in rabbit heart. Am J Physiol. 1990, 258: H1012-H1024.PubMedCentralPubMed Gonzalez F, Bassingthwaighte JB: Heterogeneities in regional volumes of distribution and flows in rabbit heart. Am J Physiol. 1990, 258: H1012-H1024.PubMedCentralPubMed
25.
Zurück zum Zitat Judd RM, Levy BI: Effects of barium-induced cardiac contraction on large- and small-vessel intramyocardial blood volume. Circ Res. 1991, 68: 217-225. 10.1161/01.RES.68.1.217.CrossRefPubMed Judd RM, Levy BI: Effects of barium-induced cardiac contraction on large- and small-vessel intramyocardial blood volume. Circ Res. 1991, 68: 217-225. 10.1161/01.RES.68.1.217.CrossRefPubMed
26.
Zurück zum Zitat Toyota E, Fujimoto K, Ogasawara Y, Kajita T, Shigeto F, Matsumoto T, Goto M, Kajiya F: Dynamic changes in three-dimensional architecture and vascular volume of transmural coronary microvasculature between diastolic- and systolic-arrested rat hearts. Circulation. 2002, 105: 621-626. 10.1161/hc0502.102964.CrossRefPubMed Toyota E, Fujimoto K, Ogasawara Y, Kajita T, Shigeto F, Matsumoto T, Goto M, Kajiya F: Dynamic changes in three-dimensional architecture and vascular volume of transmural coronary microvasculature between diastolic- and systolic-arrested rat hearts. Circulation. 2002, 105: 621-626. 10.1161/hc0502.102964.CrossRefPubMed
27.
Zurück zum Zitat Wu EX, Tang H, Wong KK, Wang J: Mapping cyclic change of regional myocardial blood volume using steady-state susceptibility effect of iron oxide nanoparticles. J Magn Reson Imaging. 2004, 19: 50-58. 10.1002/jmri.10426.CrossRefPubMed Wu EX, Tang H, Wong KK, Wang J: Mapping cyclic change of regional myocardial blood volume using steady-state susceptibility effect of iron oxide nanoparticles. J Magn Reson Imaging. 2004, 19: 50-58. 10.1002/jmri.10426.CrossRefPubMed
28.
Zurück zum Zitat Wansapura J, Gottliebson W, Crotty E, Fleck R: Cyclic variation of T1 in the myocardium at 3 T. Magn Reson Imaging. 2006, 24: 889-893. 10.1016/j.mri.2006.04.016.CrossRefPubMed Wansapura J, Gottliebson W, Crotty E, Fleck R: Cyclic variation of T1 in the myocardium at 3 T. Magn Reson Imaging. 2006, 24: 889-893. 10.1016/j.mri.2006.04.016.CrossRefPubMed
29.
Zurück zum Zitat Harris PA, Lorenz CH, Holburn GE, Overholser KA: Regional measurement of the Gd-DTPA tissue partition coefficient in canine myocardium. Magn Reson Med. 1997, 38: 541-545. 10.1002/mrm.1910380406.CrossRefPubMed Harris PA, Lorenz CH, Holburn GE, Overholser KA: Regional measurement of the Gd-DTPA tissue partition coefficient in canine myocardium. Magn Reson Med. 1997, 38: 541-545. 10.1002/mrm.1910380406.CrossRefPubMed
30.
Zurück zum Zitat Muehling OM, Jerosch-Herold M, Panse P, Zenovich A, Wilson BV, Wilson RF, Wilke N: Regional heterogeneity of myocardial perfusion in healthy human myocardium: assessment with magnetic resonance perfusion imaging. J Cardiovasc Magn Reson. 2004, 6: 499-507. 10.1081/JCMR-120030570.CrossRefPubMed Muehling OM, Jerosch-Herold M, Panse P, Zenovich A, Wilson BV, Wilson RF, Wilke N: Regional heterogeneity of myocardial perfusion in healthy human myocardium: assessment with magnetic resonance perfusion imaging. J Cardiovasc Magn Reson. 2004, 6: 499-507. 10.1081/JCMR-120030570.CrossRefPubMed
Metadaten
Titel
T1 mapping of the myocardium: Intra-individual assessment of the effect of field strength, cardiac cycle and variation by myocardial region
verfasst von
Nadine Kawel
Marcelo Nacif
Anna Zavodni
Jacquin Jones
Songtao Liu
Christopher T Sibley
David A Bluemke
Publikationsdatum
01.12.2012
Verlag
BioMed Central
Erschienen in
Journal of Cardiovascular Magnetic Resonance / Ausgabe 1/2012
Elektronische ISSN: 1532-429X
DOI
https://doi.org/10.1186/1532-429X-14-27

Weitere Artikel der Ausgabe 1/2012

Journal of Cardiovascular Magnetic Resonance 1/2012 Zur Ausgabe

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.