Reproducibility of in-vivo diffusion tensor cardiovascular magnetic resonance in hypertrophic cardiomyopathy
verfasst von:
Laura-Ann McGill, Tevfik F Ismail, Sonia Nielles-Vallespin, Pedro Ferreira, Andrew D Scott, Michael Roughton, Philip J Kilner, S Yen Ho, Karen P McCarthy, Peter D Gatehouse, Ranil de Silva, Peter Speier, Thorsten Feiweier, Choukkri Mekkaoui, David E Sosnovik, Sanjay K Prasad, David N Firmin, Dudley J Pennell
Myocardial disarray is an important histological feature of hypertrophic cardiomyopathy (HCM) which has been studied post-mortem, but its in-vivo prevalence and extent is unknown. Cardiac Diffusion Tensor Imaging (cDTI) provides information on mean intravoxel myocyte orientation and potentially myocardial disarray. Recent technical advances have improved in-vivo cDTI, and the aim of this study was to assess the interstudy reproducibility of quantitative in-vivo cDTI in patients with HCM.
Methods and results
A stimulated-echo single-shot-EPI sequence with zonal excitation and parallel imaging was implemented. Ten patients with HCM were each scanned on 2 different days. For each scan 3 short axis mid-ventricular slices were acquired with cDTI at end systole. Fractional anisotropy (FA), mean diffusivity (MD), and helix angle (HA) maps were created using a cDTI post-processing platform developed in-house. The mean ± SD global FA was 0.613 ± 0.044, MD was 0.750 ± 0.154 × 10-3 mm2/s and HA was epicardium −34.3 ± 7.6°, mesocardium 3.5 ± 6.9° and endocardium 38.9 ± 8.1°. Comparison of initial and repeat studies showed global interstudy reproducibility for FA (SD = ± 0.045, Coefficient of Variation (CoV) = 7.2%), MD (SD = ± 0.135 × 10-3 mm2/s, CoV = 18.6%) and HA (epicardium SD = ± 4.8°; mesocardium SD = ± 3.4°; endocardium SD = ± 2.9°). Reproducibility of FA was superior to MD (p = 0.003). Global MD was significantly higher in the septum than the reference lateral wall (0.784 ± 0.188 vs 0.750 ± 0.154 x10-3 mm2/s, p < 0.001). Septal HA was significantly lower than the reference lateral wall in all 3 transmural layers (from −8.3° to −10.4°, all p < 0.001).
Conclusions
To the best of our knowledge, this is the first study to assess the interstudy reproducibility of DTI in the human HCM heart in-vivo and the largest cDTI study in HCM to date. Our results show good reproducibility of FA, MD and HA which indicates that current technology yields robust in-vivo measurements that have potential clinical value. The interpretation of regional differences in the septum requires further investigation.
Prof Dudley Pennell is a consultant to Siemens and a director of Cardiovascular Imaging Solutions. Royal Brompton Hospital has research collaboration agreements with Siemens and Medical Solutions. The other authors declare that they have no competing interests.
Authors’ contributions
LAM, TFI, SNV, PF, ADS, SKP, DNF recruited patients, acquired patient data and performed image analysis and data handling. MR performed all study statistics. SNV, PF, PS, TF, CM, DES, PDG, ADS, DNF contributed to image acquisition and processing. DJP, DNF, PJK, SYH, KPM, RDS conceived the study. DJP, DNF were responsible for the final manuscript. All authors read and approved the final manuscript.
Hypertrophic cardiomyopathy (HCM) is the most common inherited cardiomyopathy and affects 1 in 500 of the general population[1, 2]. The majority of patients experience a benign clinical course, however, a small but significant minority experience major adverse sequelae, including sudden cardiac death and heart failure mortality[2, 3]. Histologically, HCM manifests as fibrosis, microvascular abnormalities, myocyte hypertrophy and myocyte disarray. Disarray may play a key role the in the genesis of the substrate responsible for malignant ventricular arrhythmias and myocardial contractile dysfunction,[4] but to date, insights into the extent and spatial distribution of myocyte disarray have been derived largely from small post-mortem series examining hearts obtained after sudden cardiac death[5‐8] These studies have considerable selection bias and as a result, the in-vivo prevalence and functional consequences of disarray in HCM remain unknown. Furthermore, the diagnostic utility of disarray in HCM is unclear as the phenomenon has also been described in a variety of congenital and acquired myocardial diseases[9, 10]. These include other conditions associated with left ventricular hypertrophy, such as aortic stenosis and hypertensive heart disease,[11] the latter frequently posing a challenge in the differential diagnosis of HCM[12].
Cardiovascular magnetic resonance (CMR) is assuming an increasingly important role in the diagnosis and evaluation of HCM[13, 14]. It allows the accurate assessment of hypertrophy;[15] the detection and quantification of fibrosis through late gadolinium enhancement (LGE) imaging and T1-relaxometry;[16, 17] and the identification of microvascular dysfunction through first-pass perfusion imaging[18]. The assessment of disarray however has proved more challenging. The emerging CMR-based technique of cardiovascular diffusion tensor imaging (cDTI) may afford this opportunity[19] and could have diagnostic utility in the differential diagnosis of left ventricular hypertrophy. This technology has been successfully applied to assess nerve fibre architecture in the central nervous system. The orientation of nerve tracts is inferred by assessing the diffusivity of water molecules which preferentially occurs parallel to the nerve fibres[20]. However, the in-vivo application of such techniques to the myocardial architecture has been limited by the challenge of resolving the molecular diffusion of water molecules in the presence of bulk motion of the heart due to cardiac contraction and respiratory motion[21, 22]. Studies in animal models have suggested that this technique may give useful insights into disarray in HCM,[23, 24] but to date there is only one previous in-vivo human pilot study of 5 HCM patients[25]. We previously described the development of a novel pulse sequence for cDTI which draws on technical improvements and 3 T imaging in normal subjects[22]. In the current study, we applied this novel technique to quantify the reproducibility of cDTI in HCM for Fractional Anisotropy (FA), Mean Diffusivity (MD), and Helix Angle (HA) maps to determine its robustness for potential clinical use.
Anzeige
Methods
Study population
We prospectively enrolled 10 patients with HCM referred for CMR at Royal Brompton Hospital. HCM was diagnosed in accordance with standard clinical guidelines[12]. Patients were excluded if they had atrial fibrillation, significant comorbidity, or contraindications to CMR and gadolinium-based contrast agents. The study was approved by the National Research Ethics Service and was conducted in accordance with the principles set out in the declaration of Helsinki, with written informed consent obtained from all patients.
Image acquisition
Images were acquired using a clinical 3.0 T scanner (Magnetom Skyra, Siemens AG Healthcare Sector, Erlangen, Germany) with an anterior 18 element matrix coil and 8–12 elements of a matrix spine coil. The 10 patients were each scanned on two separate days (subsequently referred to as initial and repeat study) in order to assess interstudy reproducibility.
After the localization steps to determine the short axis of the left ventricle (LV), a retro-gated cine sequence with a temporal resolution of 40 ms, acquired in a mid-ventricular short axis plane, was used to find the timing and duration of the subject-specific end-systolic pause. Localized first and second-order shimming and frequency adjustment were performed with an adjustment box positioned to cover the extent of the left ventricle in the imaged slices. cDTI was performed in the end-systolic pause in the mid left ventricle using a diffusion-weighted stimulated echo acquisition mode (DW-STEAM) single-shot echo-planar imaging (EPI) sequence gated to every other cardiac cycle[22]. Zonal-excitation was used to reduce the field of view in the phase encode direction and therefore, the echo train length[26]. The following sequence parameters were used: b = 0 s/mm2 plus 6 diffusion encoding directions, fat saturation, TR = 2 RR intervals = 2000 ms (assuming a heart rate of 60 beats per minute), TE = 23 ms, BW = 2442 Hz/pixel, GRAPPA parallel imaging acceleration factor of 2,[27] field of view = 360 × 123–157 mm2, slice thickness 8 mm, EPI echo train length = 22–28 readouts depending on field of view, EPI readout duration = 11–14 ms, in-plane spatial resolution = 2.7 × 2.7 mm2 interpolated to 1.35 × 1.35 mm2, 3 slices, 4 mm slice gap. Typically 10 image repetitions were obtained for each slice and each diffusion direction before averaging to improve signal to noise ratio (SNR). When motion artefacts were present due to inadequate breath holding and ectopy, up to 13 repetitions were acquired and the order that the diffusion directions were acquired in was rotated to avoid losing data from the same direction at the end of each breath hold. For diffusion encoding, the maximum available on axis gradient strength of 45 mT/m was used with a trapezoidal gradient pulse duration of 10 ms, leading to a diffusion sensitivity of b = 350 s/mm2. Breathold duration was for 18 cardiac cycles, which was typically 18 s. The typical duration of the patient scan was 45 minutes.
In order to provide functional and morphological data, breath-hold CMR was performed. Retrospectively gated balanced steady-state free-precession cine sequences were acquired in three long-axis planes, followed by a contiguous stack of short axis slices from the atrioventricular ring to the apex[28].
Anzeige
Diffusion tensor analysis
A cDTI post-processing software tool was custom-built in house using MATLAB (Mathworks, MA, USA). It included an interface to reject images corrupted by artefacts based on a visual analysis and an automatic cross-correlation algorithm to perform a rigid co-registration of the remaining images before averaging (translation, no rotation)[29]. A rank 2 diffusion tensor was generated for each voxel using the signal intensity data from the six diffusion-weighted images and the reference image where b = 0 as described in Kingsley 2006[30]. The eigensystem (eigenvalues and eigenvectors) was then calculated for each tensor. Due to noise, high local anisotropy, misregistration, and or artefacts, a small percentage of the measured eigenvalues are negative, violating the assumption of a positive definite diffusion tensor. These were set to the mean of the corresponding non-negative eigenvalues in neighbouring voxels. Two quantitative diffusion parameter maps were calculated from the eigensystems: Fractional Anisotropy (FA) and Mean Diffusivity (MD)[31]. FA is an index of the degree of deviation of the observed diffusion from isotropic, and MD is the first moment of the diffusion tensor, which is the average diffusivity.
The principal eigenvector of the diffusion tensor was taken to represent mean intravoxel myocyte orientation. The Helix Angle (HA) was then calculated as the angle between the short axis plane and the projection of the myocyte vector onto a plane tangential to the local endocardium[32]. Left-handed epicardial helix orientations were assigned a negative angle and right handed endocardial helix orientations a positive helix angle, as originally described by Streeter,[33, 34] with zero angle being aligned with the local circumferential direction.
For quantitative analysis, the myocardium was divided into 4 segments for each slice and the same segments in each slice were grouped into the conventional LV walls (anterior, septal, inferior, lateral)[35]. Care was taken to exclude papillary muscle regions. HA analysis further segmented the myocardium transmurally into endocardial, mesocardial and epicardial layers. This was performed by dividing the local myocardium into three equal thickness layers. All cDTI analysis was performed by a single observer blinded to clinical data. 3D visualisation of the tensor with superquadric glyphs[36] was implemented using Python and Paraview (Kitware, NM, USA) software. The post-processing and analysis time was approximately 3 hours.
Image analysis
Ventricular volumes, function, mass, and ejection fraction for all patients were measured for the LV using a semi-automated threshold-based technique (CMRtools, Cardiovascular Imaging Solutions, London). All volume and mass measurements were indexed to body surface area calculated using the Du Bois method[37, 38]. Late gadolinium enhancement (LGE) imaging was analysed to assess fibrosis, and recorded as present/not present with its location, and in addition the extent of LGE was quantified using CMR42 software with a full-width half-maximum threshold and expressed as percent of LV mass.
Statistical analysis
MD and FA values were analysed globally (all segments, all slices), by slice (3 slices), and by LV wall (common wall segments averaged from all 3 slices). HA was also analysed globally, by slice and by wall but also in 3 layers (epicardium, mesocardium, endocardium) because of the known transmural variation in HA. Values from the initial study were compared between LV walls, slices and layers (for HA) using a hierarchical (patients - slices - LV walls - layers) mixed effects model. The lateral wall and mid slice were chosen as reference regions for comparison, because the lateral wall is less commonly affected by hypertrophy in HCM, and the mid-wall is furthest from any anomalous measurement relating to the apex or mitral valve. All values were found to be normally distributed using the Kolmogorov-Smirnov test, and are therefore shown as mean ± standard deviation (SD). Statistical significance was set to p <0.01 to account for the number of statistical tests performed. Interstudy reproducibility was assessed using the SD of the difference between the scans, and the coefficient of variation (CoV) where reasonable (SD of the difference between the scans divided by the measurement mean). Bland-Altman analysis was also performed. Differences in the CoV were assessed using a variance ratio test.
Results
Study population
The baseline demographic, clinical and CMR characteristics of the study population are summarised in Table1.
Table 1
Baseline patient characteristics [mean ± SD, or number of patients (%)]
HCM patients (n = 10)
Age [years]
57 ± 9
Male
9 (90%)
Body Surface Area [m2]
1.99 ± 0.2
Body mass index [kg/m2]
27.5 ± 4.1
LV morphology &risk factors
Asymmetrical septal hypertrophy
7 (70%)
Apical hypertrophy
3 (30%)
Number of risk factors for sudden death: median (range)
1 (0-1)
Presence and location of late gadolinium enhancement
9 (90%)
Septum
47%
Anterior
20%
Inferior
20%
Lateral
7%
Extent of late gadolinium enhancement: [% of LV mass]
4.5% ± 5.6%
CMR dimensions and function
Indexed LV EDV [mL/m2]
70 ± 12
Indexed LV ESV [mL/m2]
16 ± 4
LV Ejection Fraction [%]
77 ± 5
LV Mass Index [g/m2]
63 ± 12
Maximum end-diastolic Wall Thickness [mm]
22 ± 5
Diffusion tensor imaging
DTI was successfully acquired in all cases. Figure1 demonstrates a DTI dataset showing typical images before averaging. b0 images (references images without diffusion weighting) after motion correction and averaging, and derived FA, MD and HA maps at three slice locations, together with FA and MD plots for each LV myocardial segment, both for the initial and repeat scans, are shown in Figures2 and3. While Figure2 displays a reproducible data set without obvious artifacts, Figure3 displays the least reproducible data set of this study. In Figure2, the MD and FA maps appear homogeneous, however the transmural progression of HA can be clearly observed and looks as expected in a healthy volunteer. In Figure3, the HA maps look perturbed.
×
×
×
Fractional anisotropy
The mean global FA in the 10 patients in the initial study was 0.613 ± 0.044 (Table2). When analysed by slice, FA was greatest in the mid-slice (0.623 ± 0.043) and smaller, but not significantly so in the basal and apical slices (using the mid slice as reference). When analysed by LV wall, FA was greatest in the anterior wall at 0.615 ± 0.039 and least in the septal wall at 0.610 ± 0.059. Using the lateral wall as a reference, there was no statistically significant difference between walls.
Table 2
Fractional anisotropy (FA): results from initial study
FA
N
Mean
SD
Difference
95% Confidence interval
p value
Global
10
0.613
0.044
Slice
Mid
10
0.623
0.043
Reference
Apical
10
0.608
0.05
-0.015
-0.04
0.01
0.25
Basal
10
0.608
0.046
-0.015
-0.04
0.01
0.25
LV wall
Lateral
10
0.614
0.041
Reference
Anterior
10
0.615
0.039
0.002
-0.021
0.024
0.90
Inferior
10
0.611
0.04
-0.003
-0.026
0.019
0.79
Septal
10
0.61
0.059
-0.003
-0.026
0.019
0.77
Anzeige
Mean diffusivity
The mean global MD was 0.750 ± 0.154 × 10-3 mm2/s (Table3). MD was greatest in the apical slice (0.785 ± 0.152 × 10-3 mm2/s) and least in the basal slice (0.726 ± 0.145 × 10-3 mm2/s), but there were no statistically significant differences between slices. When analysed by LV wall, MD was significantly greater in the septal wall than the reference lateral wall (0.784 ± 0.188 × 10-3 mm2/s vs 0.714 ± 0.155 × 10-3 mm2/s; p < 0.001).
Table 3
Mean diffusivity (MD): Results from initial study [×10-3mm2/s]
MD
N
Mean
SD
Difference
95% Confidence interval
p value
Global
10
0.75
0.154
Slice
Mid
10
0.742
0.178
Reference
Apical
10
0.785
0.152
0.042
-0.017
0.101
0.16
Basal
10
0.726
0.145
-0.016
-0.075
0.043
0.60
LV wall
Lateral
10
0.714
0.155
Reference
Anterior
10
0.758
0.145
0.044
0.004
0.083
0.029
Inferior
10
0.742
0.137
0.028
-0.011
0.067
0.16
Septal
10
0.784
0.188
0.07
0.031
0.109
<0.001
Helix angle
Global LV helical angles (Table4) showed transmural progression from left handed in the epicardium (−34.3 ± 7.6°), near perpendicular to the imaging plane in the mesocardium (3.5 ± 6.9°), to right handed in the endocardium (38.9 ± 8.1°). When analysed by slice, the helical angles were not statistically different between slices. When analysed by LV wall, the endocardial HA was statistically more right handed in the inferior and septal walls compared to the lateral wall (33.9 ± 8.4°, 34.1 ± 5.2° vs 42.6 ± 5.5°; both p < 0.001). In the mesocardial layer the inferior and septal walls HA were significantly more perpendicular to the plane than in the lateral wall (−0.2 ± 5.7°, -0.7 ± 6.9° vs 7.6 ± 4.5°; both p < 0.001). In the epicardial layer, the anterior and septal HA was significantly more left handed than lateral wall (−36.9 ± 6.9°, -39.1 ± 8.2° vs −28.7 ± 6.0° p = 0.004 & <0.001).
Table 4
Helix angle (HA): Results from initial study [degrees]
HA
N
Mean
SD
Difference
95% Confidence interval
p value
Global
Endo
10
38.9
8.1
Meso
10
3.5
6.9
Epi
10
−34.3
7.6
Slice
Endo layer
Mid
10
40
4.6
Reference
Apical
10
41
7.2
1
−2.6
4.7
0.58
Basal
10
36.3
6
−3.7
−7.4
−0.1
0.044
Meso layer
Mid
10
3
3.2
Reference
Apical
10
4.3
5.3
1.3
−1.5
4
0.37
Basal
10
3.5
4.3
0.5
−2.2
3.2
0.72
Epi layer
Mid
10
−34.4
3.2
Reference
Apical
10
−33.8
6.5
0.6
−3
4.2
0.73
Basal
10
−34.8
4.1
−0.4
−4
3.2
0.82
LV wall
Endo layer
Lateral
10
42.6
5.5
Reference
Anterior
10
45.1
6.6
2.5
−1.6
6.7
0.24
Inferior
10
33.9
8.4
−8.7
−12.8
−4.5
<0.001
Septal
10
34.1
5.2
−8.5
−12.6
−4.3
<0.001
Meso layer
Lateral
10
7.6
4.5
Reference
Anterior
10
7.3
6.1
−0.4
−4.6
3.9
0.87
Inferior
10
−0.2
5.7
−7.8
−12.1
−3.5
<0.001
Septal
10
−0.7
6.9
−8.3
−12.6
−4
<0.001
Epi layer
Lateral
10
−28.7
6
Reference
Anterior
10
−36.9
6.9
−8.3
−13.8
−2.7
0.004
Inferior
10
−32.4
5.4
−3.7
−9.3
1.8
0.19
Septal
10
−39.1
8.2
−10.4
−16
−4.9
<0.001
Reproducibility
Initial and repeat studies were performed 167 ± 21 days apart. There were no clinical events or changes in treatment in the intervening period. FA, MD and HA maps appeared similar between initial and repeat studies in the majority of cases (Figure2). Figure4 shows two pairs of HA maps (initial vs repeat study) showing good reproducibility (top), and an example which appears less reproducible (bottom). Small regions with high levels of noise (arrows) can be seen in the HA maps, thought to be a consequence of motion, either cardiac or respiratory or both.
×
FA reproducibility data, comparing the initial and repeat studies are provided in Table5 and the associated Bland-Altman plots are shown in Figure5. FA data analysed globally, per slice and by LV wall showed good reproducibility with minimal bias and low variance. The global SD between the initial and repeat studies was ± 0.045 and the CoV was 7.2%.
Table 5
Fractional anisotropy (FA) reproducibility
FA
N
Mean value
Mean difference
SD of difference
95% limits of agreement
Coefficient of variation
Global
10
0.617
−0.008
0.045
−0.096
0.079
7.2%
Slice
Apical
10
0.614
−0.013
0.054
−0.119
0.093
8.8%
Mid
10
0.622
0.001
0.051
−0.099
0.102
8.2%
Basal
10
0.614
−0.013
0.043
−0.098
0.072
7.1%
LV wall
Anterior
10
0.620
−0.009
0.042
−0.091
0.073
6.8%
Lateral
10
0.608
0.012
0.059
−0.104
0.128
9.7%
Inferior
10
0.620
−0.018
0.053
−0.123
0.087
8.6%
Septal
10
0.619
−0.017
0.053
−0.121
0.086
8.5%
×
Anzeige
MD reproducibility data, comparing the initial and repeat studies, are provided in Table6 and Figure5. MD data showed good reproducibility at each level of analysis. The global SD between the initial and repeat studies was ± 0.135 × 10-3 mm2/s and the coefficient of variation was 18.6%. The interstudy reproducibility of MD was significantly less good than for FA (p = 0.003).
Table 6
Mean diffusivity (MD) reproducibility (×10-3mm2/s)
MD
N
Mean value
Mean difference
SD of difference
95% limits of agreement
Coefficient of variation
Global
10
0.728
0.045
0.135
−0.220
0.310
18.6%
Slice
Apical
10
0.764
0.041
0.176
−0.304
0.385
23.0%
Mid
10
0.720
0.045
0.160
−0.268
0.358
22.2%
Basal
10
0.704
0.045
0.115
−0.181
0.271
16.4%
LV wall
Anterior
10
0.727
0.061
0.112
−0.158
0.281
15.4%
Lateral
10
0.713
0.003
0.164
−0.317
0.324
22.9%
Inferior
10
0.716
0.053
0.140
−0.221
0.327
19.5%
Septal
10
0.751
0.067
0.232
−0.388
0.521
30.9%
HA reproducibility data, comparing the initial and repeat studies are provided in detail in Table7. The associated plots are shown in Figures5 and6. The global mean SD for HA between the initial and repeat studies was ± 4.8°, ± 3.4° and ± 2.9° in the endocardial, mesocardial and epicardial layers. Analysis of HA by slice and LV wall was similarly reproducible. Three-dimensional analysis of the DTI data is shown in Figure7 as superquadric glyphs. The colour and orientation/shape of the glyphs are determined by the helix angle and the eigensystem respectively.
Table 7
Helix angle (HA) reproducibility (degrees)
HA
N
Mean value
Mean difference
SD of difference
95% limits of agreement
Global
Endo
10
38.4
1.2
4.8
−8.3
10.6
Meso
10
3.9
−0.6
3.4
−7.3
6.0
Epi
10
−33.1
−2.7
2.9
−8.4
3.0
Apical endo
10
39.5
3.0
10.6
−17.8
23.9
Apical meso
10
4.3
0.0
6.1
−12.0
12.0
Apical epi
10
−32.0
−3.6
6.4
−16.2
8.9
Mid endo
10
39.1
1.9
6.2
−10.3
14.1
Mid meso
10
3.4
−0.7
4.7
−9.9
8.4
Mid epi
10
−33.0
−2.9
6.8
−16.2
10.4
Basal endo
10
36.6
−0.8
4.4
−9.3
7.8
Basal meso
10
4.1
−1.1
3.2
−7.3
5.2
Basal epi
10
−33.9
−1.8
4.7
−11.0
7.4
LV wall
Anterior endo
10
43.2
3.7
8.6
−13.2
20.5
Anterior meso
10
6.0
2.6
5.5
−8.1
13.3
Anterior epi
10
−36.5
−0.9
7.3
−15.2
13.5
Lateral endo
10
42.6
0.0
4.9
−9.7
9.7
Lateral meso
10
8.7
−2.2
6.9
−15.8
11.4
Lateral epi
10
−27.7
−2.0
7.8
−17.4
13.3
Inferior endo
10
34.2
−0.6
6.7
−13.6
12.5
Inferior meso
10
0.7
−1.7
7.2
−15.9
12.4
Inferior epi
10
−30.8
−3.2
9.4
−21.7
15.3
Septal endo
10
33.3
1.6
8.0
−14.1
17.4
Septal meso
10
0.2
−1.8
6.8
−15.1
11.6
Septal epi
10
−36.7
−4.9
7.3
−19.3
9.5
The CoV was not calculated for HA, see text for details.
×
×
Discussion
In-vivo cDTI is challenging due to cardiac motion and the intrinsically low SNR of the technique. However, our previous study showed that cDTI using the DW-STEAM sequence at 3T can be performed in-vivo in healthy volunteers with good reproducibility[22]. We now show that this technique can be performed apparently successfully in patients with HCM and have assessed its interstudy reproducibility and the variation of quantitative measures around the myocardium.
FA had a global value of 0.613 ± 0.044 and analysis by slice and wall yielded similar results with no statistical differences. The interstudy reproducibility of global FA was very good at 7.2%, with values ranging from 6.8% to 9.7% for slices and walls. This very robust reproducibility suggests that the values have clinical validity. MD had a global value of 0.750 ± 0.154 × 10-3 mm2/s and was significantly higher in the septum (0.784 ± 0.188 × 10-3 mm2/s, p < 0.001). The interstudy reproducibility of global MD was reasonable at 18.6%, with values ranging from 15.4% to 30.9% for slices and walls. The interstudy reproducibility of MD was significantly less good than for FA (p = 0.003).
Anzeige
One possible interpretation of the finding of higher MD in the septum might be that it resulted from the commonly found septal hypertrophy in HCM in association with an enlarged interstitial space from disarray or fibrosis. However, considerable caution must be observed in the interpretation of this result because of known and unknown technical issues. One issue might be that the SNR across the LV is not homogeneous due to the decreasing sensitivity of the receiver surface coils with depth (the signal being higher the closer to the coil elements). It is well documented in diffusion tensor imaging that FA and MD values are biased by SNR due to the Rician noise distribution and due to the fact that only magnitude images are used to derive the diffusion tensor parameters[39]. Therefore, higher SNR may lead to higher values of MD. The evaluation of the SNR in these images is beyond the scope of this paper, but will be the subject of future work. Another possible confounding issue is sensitivity to the difference in myocardial position between the 2 successive cardiac cycles used for imaging, which might affect different regions of the LV in a different manner and create artefactual MD differences across the LV. A further potential issue is mis-registration of the regions of interest used to segment the LV, with influence of right ventricular papillary muscles affecting the septal measurements. Finally, there will also be unidentified technical issues which could have a bearing on the results. The relative contribution of each of these issues is currently unknown, and in-depth analysis of larger numbers of patients and healthy volunteers will be necessary to determine whether the increased MD is an actual clinical finding or caused by limitations of the present cDTI technique.
Analysis of HA was divided into 3 transmural layers because of prior findings. The global value for endocardial HA was 38.9° ± 8.1°, and SD of the difference between the measurement for the 2 scans was 4.8°. For the mesocardium, the global value was 3.5° ± 6.9° with an interstudy SD of 3.4°. For the epicardium, the global value was −34.3° ± 7.6° with an interstudy SD of 2.9°. These findings indicate very good reproducibility between scans of HA in all 3 layers and suggest clinical validity. CoVs were not calculated because their magnitude is a function of the coordinate angles chosen, and in the current system were incalculable in the mesocardial layer as the mean helical angle approximates to zero as myocyte orientation rotates from a left-handed to right-handed helix creating unstable values by division. Significant differences in HA were found between the reference lateral wall and the septum in all 3 transmural layers (ranging from 8.5° to −10.4°, all p < 0.001). Significant differences also affected the inferior wall in the endo and mesocardium (p < 0.001), and the anterior wall in the epicardium (p = 0.004). One possible interpretation of the finding of different HA values in the septum is disturbance from myocardial disarray. However, once again this interpretation needs to be regarded with considerable caution. In this study the myocardium was segmented transmurally into 3 equal thickness layers. This does not necessarily represent the real left-handed, circular and right-handed HA structure of the myocardium and therefore averaging within these 3 layers is bound to happen. Although this should not largely affect the reproducibility, it could influence the LV wall comparison due to the varying myocardial thickness per wall segment. The segmentation might be subject-dependent due to the fact that the papillary muscles and RV wall are visually excluded and might differ between initial and repeat scans, affecting the values due to mis-registration of the region of interest, and also across patients. The design, implementation, optimisation and comparison of segmentation algorithms to extract robust and meaningful quantitative HA data will be a very important task for the translation of cardiac DTI into routine, and is the subject of ongoing research. On the other hand, HA distribution maps appeared healthy in the majority of cases. We hypothesise that this might be due to the fact that the patient cohort scanned for this study was composed of stable HCM patients. There is a possibility that this is the reason why the HA maps appear normal, and that more severe forms of HCM might correlate with disturbed HA distribution maps.
There is relatively little previous in-vivo work on the absolute values of FA and MD. In healthy human subjects we reported an MD value of 0.8 ± 0.2 × 10-3 mm2/s using a similar breath hold DW-STEAM protocol,[22] and Reese et al.[40] reported an MD of 0.9 ± 0.3 × 10-3 mm2/s, which are both within one standard deviation of the global value of 0.750 ± 0.154 × 10-3 mm2/s found in the current study of HCM patients. Tseng et al. performed in-vivo DTI in 5 patients with HCM and 5 volunteers[25]. They measured FA at 0.78 in the free wall and 0.56 in the septum in HCM patients (p = 0.03 for the difference) and 0.78 and 0.72 respectively in normal subjects (p = 0.15 for the difference). In our current study, we measured slightly lower FA at 0.614 ± 0.041 × 10-3 mm2/s in the lateral wall and 0.610 ± 0.059 × 10-3 mm2/s in the septum, with no significant difference between our values (p = 0.77). Further studies will be needed to compare measures, standardise and optimise the methods and further validate results. It should be noted that the quantitative results in this study are obtained from acquisitions at end-systole. We might expect differences between in-vivo and ex-vivo data, and also between fresh and fixed ex-vivo data, and finally also between data acquired at different cardiac phases.
The general limitations of this study are the following: The modest in-plane resolution of 2.7 mm might lead to volume averaging of the blood pool in endocardial layers or the RV papillary muscles in epicardial layers, which could bias the FA, MD and HA values in these areas; Data acquisition was performed at end-systole, which is considered a phase when the diffusion acquisition is not affected by strain,[40] and although the systolic pause in HCM is longer than in healthy volunteers, motion artifacts may still have occurred related to the optimal positioning of the acquisition window; The time between the 2 scans in the study might have resulted in alteration in the myocardium due to disease progression, but this is unlikely given the slow rate of progression and the lack of any clinical events in the intervening period; This is a preliminary study on a small sample size of only 10 HCM patients, and further work with a larger number of patients is required, which will include quantitative regional comparisons with analysis of LGE; The application of the current technique to patients with limited breathold capability or frequent ectopy remains to be determined, but would be more challenging.
Conclusions
In conclusion, we have implemented a robust quantitative in-vivo cDTI technique that has the potential to enhance our understanding of in-vivo structure-function relationships in the HCM heart. We have shown that cDTI of the human HCM heart in-vivo can be performed using a recently reported technique, and we have assessed the reproducibility of the results of this technique in HCM patients.
Authors’ information
David N Firmin and Dudley J Pennell are joint senior authors.
Laura-Ann McGill, Tevfik F Ismail, Sonia Nielles-Vallespin and Pedro Ferreira are joint first authors.
Acknowledgements
The authors are grateful to the staff of the CMR Unit and the Cardiovascular Biomedical Research Unit, Royal Brompton Hospital for their support with this work.
Sources of Funding
This work was supported by the National Institute of Health Research Cardiovascular Biomedical Research Unit at the Royal Brompton Hospital and Imperial College, London. Dr Ismail is funded by the British Heart Foundation.
Disclosures
Prof Dudley Pennell is a consultant to Siemens and a director of Cardiovascular Imaging Solutions. The Royal Brompton Hospital has research collaboration agreements with Siemens AG Medical Solutions.
Open Access
This article is published under license to BioMed Central Ltd. This is an Open Access article is distributed under the terms of the Creative Commons Attribution License (
https://creativecommons.org/licenses/by/2.0
), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
Competing interests
Prof Dudley Pennell is a consultant to Siemens and a director of Cardiovascular Imaging Solutions. Royal Brompton Hospital has research collaboration agreements with Siemens and Medical Solutions. The other authors declare that they have no competing interests.
Authors’ contributions
LAM, TFI, SNV, PF, ADS, SKP, DNF recruited patients, acquired patient data and performed image analysis and data handling. MR performed all study statistics. SNV, PF, PS, TF, CM, DES, PDG, ADS, DNF contributed to image acquisition and processing. DJP, DNF, PJK, SYH, KPM, RDS conceived the study. DJP, DNF were responsible for the final manuscript. All authors read and approved the final manuscript.
Reproducibility of in-vivo diffusion tensor cardiovascular magnetic resonance in hypertrophic cardiomyopathy