Equipment review: An appraisal of the LiDCO™plus method of measuring cardiac output

verfasst von: Rupert M Pearse, Kashif Ikram, John Barry

Erschienen in: Critical Care | Ausgabe 3/2004

Einloggen, um Zugang zu erhalten

Abstract

The LiDCO™plus system is a minimally/non-invasive technique of continuous cardiac output measurement. In common with all cardiac output monitors this technology has both strengths and weaknesses. This review discusses the technological basis of the device and its clinical application.
Literatur
1.
2.
Zurück zum Zitat Technology Assessment Task Force of the Society of Critical Care Medicine: A model for technology assessment applied to pulse oximetry. Crit Care Med 1993, 21: 615-624.CrossRef Technology Assessment Task Force of the Society of Critical Care Medicine: A model for technology assessment applied to pulse oximetry. Crit Care Med 1993, 21: 615-624.CrossRef
3.
Zurück zum Zitat Hamilton TT, Huber LM, Jessen ME: PulseCO: a less-invasive technique to monitor cardiac output from arterial pressure after cardiac surgery. Ann Thorac Surg 2002, 74: S1408-S1412. 10.1016/S0003-4975(02)04059-6CrossRefPubMed Hamilton TT, Huber LM, Jessen ME: PulseCO: a less-invasive technique to monitor cardiac output from arterial pressure after cardiac surgery. Ann Thorac Surg 2002, 74: S1408-S1412. 10.1016/S0003-4975(02)04059-6CrossRefPubMed
4.
Zurück zum Zitat Heller LB, Fisher M, Pfanzelter N, Jayakar D, Jeevanandam V, Aronson S: Continuous intraoperative cardiac output determination with arterial pulse wave analysis (PulseCO™) is valid and precise. Anesth Analg 2002, 93: SCA1-SCA112. Heller LB, Fisher M, Pfanzelter N, Jayakar D, Jeevanandam V, Aronson S: Continuous intraoperative cardiac output determination with arterial pulse wave analysis (PulseCO™) is valid and precise. Anesth Analg 2002, 93: SCA1-SCA112.
5.
Zurück zum Zitat Mason DJ, O'Grady M, Woods JP, McDonell W: Assessment of lithium dilution cardiac output as a technique for measurement of cardiac output in dogs. Am J Vet Res 2001, 62: 1255-1261.CrossRefPubMed Mason DJ, O'Grady M, Woods JP, McDonell W: Assessment of lithium dilution cardiac output as a technique for measurement of cardiac output in dogs. Am J Vet Res 2001, 62: 1255-1261.CrossRefPubMed
6.
Zurück zum Zitat Linton RA, Jonas MM, Tibby SM, Murdoch IA, O'Brien TK, Linton NW, Band DM: Cardiac output measured by lithium dilution and transpulmonary thermodilution in patients in a paediatric intensive care unit. Intensive Care Med 2000, 26: 1507-1511. 10.1007/s001340051347CrossRefPubMed Linton RA, Jonas MM, Tibby SM, Murdoch IA, O'Brien TK, Linton NW, Band DM: Cardiac output measured by lithium dilution and transpulmonary thermodilution in patients in a paediatric intensive care unit. Intensive Care Med 2000, 26: 1507-1511. 10.1007/s001340051347CrossRefPubMed
7.
Zurück zum Zitat Linton RA, Young LE, Marlin DJ, Blissett KJ, Brearley JC, Jonas MM, O'Brien TK, Linton NW, Band DM, Jones RS: Cardiac output measured by lithium dilution, thermodilution and transesophageal Doppler echocardiography in anesthetized horses. Am J Vet Res 2000, 61: 731-737.CrossRefPubMed Linton RA, Young LE, Marlin DJ, Blissett KJ, Brearley JC, Jonas MM, O'Brien TK, Linton NW, Band DM, Jones RS: Cardiac output measured by lithium dilution, thermodilution and transesophageal Doppler echocardiography in anesthetized horses. Am J Vet Res 2000, 61: 731-737.CrossRefPubMed
8.
Zurück zum Zitat Linton R, Band D, O'Brien T, Jonas MM, Leach R: Lithium dilution cardiac output measurement: a comparison with thermodilution. Crit Care Med 1997, 25: 1796-1800. 10.1097/00003246-199711000-00015CrossRefPubMed Linton R, Band D, O'Brien T, Jonas MM, Leach R: Lithium dilution cardiac output measurement: a comparison with thermodilution. Crit Care Med 1997, 25: 1796-1800. 10.1097/00003246-199711000-00015CrossRefPubMed
9.
Zurück zum Zitat Kurita T, Morita K, Kato S, Kikura M, Horie M, Ikeda K: Comparison of the accuracy of the lithium dilution technique with the thermodilution technique for measurement of cardiac output. Br J Anaesth 1997, 79: 770-775.CrossRefPubMed Kurita T, Morita K, Kato S, Kikura M, Horie M, Ikeda K: Comparison of the accuracy of the lithium dilution technique with the thermodilution technique for measurement of cardiac output. Br J Anaesth 1997, 79: 770-775.CrossRefPubMed
10.
Zurück zum Zitat Hatfield C, McDonell W, Lemke D, Black W: Pharmacokinetics and toxic effects of lithium chloride after intravenous administration in conscious horses. Am J Vet Res 2001, 62: 1387-1392.CrossRefPubMed Hatfield C, McDonell W, Lemke D, Black W: Pharmacokinetics and toxic effects of lithium chloride after intravenous administration in conscious horses. Am J Vet Res 2001, 62: 1387-1392.CrossRefPubMed
11.
Zurück zum Zitat Jonas MM, Kelly FE, Linton RAF, Band DM, O'Brien TK, Linton NWF: A comparison of lithium dilution cardiac output measurements made using central and atecubital venous injection of lithium chloride. J Clin Monit Comput 1999, 15: 525-528. 10.1023/A:1009914714769CrossRefPubMed Jonas MM, Kelly FE, Linton RAF, Band DM, O'Brien TK, Linton NWF: A comparison of lithium dilution cardiac output measurements made using central and atecubital venous injection of lithium chloride. J Clin Monit Comput 1999, 15: 525-528. 10.1023/A:1009914714769CrossRefPubMed
12.
Zurück zum Zitat Leibowitz AB: Do pulmonary artery catheters improve patient outcome? No. Crit Care Clin 1996, 12: 559-568.CrossRefPubMed Leibowitz AB: Do pulmonary artery catheters improve patient outcome? No. Crit Care Clin 1996, 12: 559-568.CrossRefPubMed
13.
Zurück zum Zitat Connors AF Jr: Right heart catheterisation: is it effective? New Horiz 1997, 5: 195-200.PubMed Connors AF Jr: Right heart catheterisation: is it effective? New Horiz 1997, 5: 195-200.PubMed
14.
Zurück zum Zitat Polanczyk CA, Rohde LE, Goldman L, Cook EF, Thomas EJ, Marcantonio ER, Mangione CM, Lee TH: Right heart catheterisation and cardiac complications in patients undergoing noncardiac surgery: an observational study. JAMA 2001, 286: 309-314. 10.1001/jama.286.3.309CrossRefPubMed Polanczyk CA, Rohde LE, Goldman L, Cook EF, Thomas EJ, Marcantonio ER, Mangione CM, Lee TH: Right heart catheterisation and cardiac complications in patients undergoing noncardiac surgery: an observational study. JAMA 2001, 286: 309-314. 10.1001/jama.286.3.309CrossRefPubMed
15.
Zurück zum Zitat Connors AF Jr, Speroff T, Dawson NV, Thomas C, Harrell FE Jr, Wagner D, Desbiens N, Goldman L, Wu AW, Califf RM, Fulkerson WJ Jr, Vidaillet H, Broste S, Bellamy P, Lynn J, Knaus WA: The effectiveness of right heart catheterisation in the initial care of the critically ill patients. JAMA 1996, 276: 889-977. 10.1001/jama.276.11.889CrossRefPubMed Connors AF Jr, Speroff T, Dawson NV, Thomas C, Harrell FE Jr, Wagner D, Desbiens N, Goldman L, Wu AW, Califf RM, Fulkerson WJ Jr, Vidaillet H, Broste S, Bellamy P, Lynn J, Knaus WA: The effectiveness of right heart catheterisation in the initial care of the critically ill patients. JAMA 1996, 276: 889-977. 10.1001/jama.276.11.889CrossRefPubMed
16.
Zurück zum Zitat Manecke GR Jr, Brown JC, Landau AA, Kapelanski DP, St Laurent CM, Auger WR: An unusual case of pulmonary artery catheter malfunction. Anesth Analg 2002, 95: 302-304.PubMed Manecke GR Jr, Brown JC, Landau AA, Kapelanski DP, St Laurent CM, Auger WR: An unusual case of pulmonary artery catheter malfunction. Anesth Analg 2002, 95: 302-304.PubMed
17.
Zurück zum Zitat Keus , et al.: The use of invasive (pulmonary artery) monitoring in combination with multiple co-morbidity: indispensable or hazardous? Int J Intensive Care 2002, 9: 86-92. Keus , et al.: The use of invasive (pulmonary artery) monitoring in combination with multiple co-morbidity: indispensable or hazardous? Int J Intensive Care 2002, 9: 86-92.
18.
Zurück zum Zitat Farber DL, Rose DM, Bassell GM, Eugene J: Hemopytysts and pneumothorax after removal of a persistently wedged pulmonary artery catheter. Crit Care Med 1981, 9: 494-495.CrossRefPubMed Farber DL, Rose DM, Bassell GM, Eugene J: Hemopytysts and pneumothorax after removal of a persistently wedged pulmonary artery catheter. Crit Care Med 1981, 9: 494-495.CrossRefPubMed
19.
Zurück zum Zitat Ehrie M, Morgan A, Moore F, Connor N: Endocarditis with the indwelling balloon tipped pulmonary artery catheter in burn patients. J Trauma 1978, 18: 664-666.CrossRefPubMed Ehrie M, Morgan A, Moore F, Connor N: Endocarditis with the indwelling balloon tipped pulmonary artery catheter in burn patients. J Trauma 1978, 18: 664-666.CrossRefPubMed
20.
Zurück zum Zitat Sasaki TM, Panke TW, Dorethy JF, Lindberg RB, Pruitt BA: The relationship of central venous and pulmonary artery catheter position to acute right-sided endocarditis in severe thermal injury. J Trauma 1979, 19: 740-743.CrossRefPubMed Sasaki TM, Panke TW, Dorethy JF, Lindberg RB, Pruitt BA: The relationship of central venous and pulmonary artery catheter position to acute right-sided endocarditis in severe thermal injury. J Trauma 1979, 19: 740-743.CrossRefPubMed
21.
Zurück zum Zitat McDaniel DD, Stone JG, Faltas AN, Khambatta HJ, Thys DM, Antunes AM, Bregman D: Catheter induced pulmonary artery hemorrhage. Diagnosis and management in cardiac operations. J Thoracic Cardiovasc Surg 1981, 82: 1-4. McDaniel DD, Stone JG, Faltas AN, Khambatta HJ, Thys DM, Antunes AM, Bregman D: Catheter induced pulmonary artery hemorrhage. Diagnosis and management in cardiac operations. J Thoracic Cardiovasc Surg 1981, 82: 1-4.
22.
Zurück zum Zitat Barash PG, Nardi D, Hammond G, Walker-Smith G, Capuano D, Laks H, Kopriva CJ, Baue AE, Geha AS: Catheter induced pulmonary artery perforation. Mechanisms, management, and modifications. J Thoracic Cardiovasc Surg 1981, 82: 5-12. Barash PG, Nardi D, Hammond G, Walker-Smith G, Capuano D, Laks H, Kopriva CJ, Baue AE, Geha AS: Catheter induced pulmonary artery perforation. Mechanisms, management, and modifications. J Thoracic Cardiovasc Surg 1981, 82: 5-12.
23.
Zurück zum Zitat Paulson DM, Scott SM, Sethi GK: Pulmonary hemorrhage associated with balloon flotation catheters: a case report of a case and review of the literature. J Thoracic Cardiovasc Surg 1980, 80: 453-458. Paulson DM, Scott SM, Sethi GK: Pulmonary hemorrhage associated with balloon flotation catheters: a case report of a case and review of the literature. J Thoracic Cardiovasc Surg 1980, 80: 453-458.
24.
Zurück zum Zitat Melter R, Klint PP, Simoons M: Hemoptysis after flushing Swan-Ganz catheters in the wedge position [letter]. N Engl J Med 1981, 304: 1170-1171.CrossRef Melter R, Klint PP, Simoons M: Hemoptysis after flushing Swan-Ganz catheters in the wedge position [letter]. N Engl J Med 1981, 304: 1170-1171.CrossRef
25.
Zurück zum Zitat McLoud TC, Putman CE: Radiology of the Swan-Ganz catheter and associated pulmonary complications. Radiology 1975, 116: 19-22.CrossRefPubMed McLoud TC, Putman CE: Radiology of the Swan-Ganz catheter and associated pulmonary complications. Radiology 1975, 116: 19-22.CrossRefPubMed
26.
Zurück zum Zitat Page DW, Teres D, Hartshorn JW: Fatal hemorrhage from Swan-Ganz catheter [letter]. N Engl J Med 1974, 291: 260.PubMed Page DW, Teres D, Hartshorn JW: Fatal hemorrhage from Swan-Ganz catheter [letter]. N Engl J Med 1974, 291: 260.PubMed
27.
Zurück zum Zitat Lopez-Sedon J, Lopez E, Maqueda IG, Coma-Canella I, Ramos F, Dominguez F, Jadraque LM: Right ventricular infarction as a risk factor for ventricular fibrillation during pulmonary artery catheterisation using Swan-Ganz catheters. Am Heart J 1990, 119: 207-209.CrossRef Lopez-Sedon J, Lopez E, Maqueda IG, Coma-Canella I, Ramos F, Dominguez F, Jadraque LM: Right ventricular infarction as a risk factor for ventricular fibrillation during pulmonary artery catheterisation using Swan-Ganz catheters. Am Heart J 1990, 119: 207-209.CrossRef
28.
Zurück zum Zitat Rubin SA, Puckett RP: Pulmonary artery: bronchial fistula. A new complication of Swan-Ganz catherterization. Chest 1979, 75: 515-516.CrossRefPubMed Rubin SA, Puckett RP: Pulmonary artery: bronchial fistula. A new complication of Swan-Ganz catherterization. Chest 1979, 75: 515-516.CrossRefPubMed
29.
Zurück zum Zitat Shimm DS, Rigsby L: Ventricular tachycardia associated with removal of a Swan-Ganz catheter. Postgrad Med 1980, 67: 291-294.PubMed Shimm DS, Rigsby L: Ventricular tachycardia associated with removal of a Swan-Ganz catheter. Postgrad Med 1980, 67: 291-294.PubMed
30.
Zurück zum Zitat Mark JB: Systolic pressure variation: A clinical application of respiratory-circulatory interaction. In Atlas of Cardiovascular monitoring. Churchill Livingstone. Mark JB: Systolic pressure variation: A clinical application of respiratory-circulatory interaction. In Atlas of Cardiovascular monitoring. Churchill Livingstone.
31.
Zurück zum Zitat Reuter DA, Felbinger TW, Kilger E, Schmidt C, Lamm P, Goetz AE: Optimizing fluid therapy in mechanically ventilated patients after cardiac surgery by on-line monitoring of left ventricular stroke variations. Comparison with aortic systolic pressure variations. Br J Anesth 2002, 88: 124-126. 10.1093/bja/88.1.124CrossRef Reuter DA, Felbinger TW, Kilger E, Schmidt C, Lamm P, Goetz AE: Optimizing fluid therapy in mechanically ventilated patients after cardiac surgery by on-line monitoring of left ventricular stroke variations. Comparison with aortic systolic pressure variations. Br J Anesth 2002, 88: 124-126. 10.1093/bja/88.1.124CrossRef
32.
Zurück zum Zitat Avila , et al.: Predicting hypovolemia during mechanical ventilation: a prospective, clinical trial of doppler variations of aorta and axillary arterial velocities to identify systolic pressure variation. Anaesthesiology 2002, 97: B17.CrossRef Avila , et al.: Predicting hypovolemia during mechanical ventilation: a prospective, clinical trial of doppler variations of aorta and axillary arterial velocities to identify systolic pressure variation. Anaesthesiology 2002, 97: B17.CrossRef
33.
Zurück zum Zitat Gunn RS, Pinsky MR: Implications of arterial pressure variation in patients in the intensive care unit. Crit Care 2001, 7: 212-217. Gunn RS, Pinsky MR: Implications of arterial pressure variation in patients in the intensive care unit. Crit Care 2001, 7: 212-217.
34.
Zurück zum Zitat Michard F, Chemla D, Richard C, Wysocki M, Pinsky MR, Lecarpentier Y, Teboul JL: Clinical use of respiratory changes in arterial pulse pressure to monitor the hemodynamic effects of PEEP. Crit Care Med 1999, 159: 935-939.CrossRef Michard F, Chemla D, Richard C, Wysocki M, Pinsky MR, Lecarpentier Y, Teboul JL: Clinical use of respiratory changes in arterial pulse pressure to monitor the hemodynamic effects of PEEP. Crit Care Med 1999, 159: 935-939.CrossRef
35.
Zurück zum Zitat Michard F, Boussat S, Chemla D, Anguel N, Mercat A, Lecarpentier Y, Richard C, Pinsky MR, Teboul JL: Relation between respiratory changes in arterial pulse pressure and fluid responsiveness in septic patients with acute circulatory failure. Crit Care Med 2000, 162: 134-138.CrossRef Michard F, Boussat S, Chemla D, Anguel N, Mercat A, Lecarpentier Y, Richard C, Pinsky MR, Teboul JL: Relation between respiratory changes in arterial pulse pressure and fluid responsiveness in septic patients with acute circulatory failure. Crit Care Med 2000, 162: 134-138.CrossRef
36.
Zurück zum Zitat Michard F, Teboul JL: Using heart-lung interactions to assess fluid responsiveness during mechanical ventilation. Crit Care 2000, 4: 282-289. 10.1186/cc710PubMedCentralCrossRefPubMed Michard F, Teboul JL: Using heart-lung interactions to assess fluid responsiveness during mechanical ventilation. Crit Care 2000, 4: 282-289. 10.1186/cc710PubMedCentralCrossRefPubMed
37.
Zurück zum Zitat Berkenstadt H, Margalit N, Hadani M, Friedman Z, Segal E, Villa Y, Perel A: Stroke volume variation as a predictor of fluid responsiveness in patients undergoing brain surgery. Anesth Analg 2001, 92: 984-989.CrossRefPubMed Berkenstadt H, Margalit N, Hadani M, Friedman Z, Segal E, Villa Y, Perel A: Stroke volume variation as a predictor of fluid responsiveness in patients undergoing brain surgery. Anesth Analg 2001, 92: 984-989.CrossRefPubMed
38.
Zurück zum Zitat Perel A, Pizov R, Cotev S: Systolic blood pressure variation is a sensitive indicator of hypovolemia in ventilated dogs subjected to graded hemorrhage. Anesthesiology 1987, 67: 498-502.CrossRefPubMed Perel A, Pizov R, Cotev S: Systolic blood pressure variation is a sensitive indicator of hypovolemia in ventilated dogs subjected to graded hemorrhage. Anesthesiology 1987, 67: 498-502.CrossRefPubMed
39.
Zurück zum Zitat Harrigan PWJ, Pinsky MR: Heart–lung interactions. Part 1: effects of lung volume and ventilation as exercise. Int J Intensive Care 2001, Spring: 6-13. Harrigan PWJ, Pinsky MR: Heart–lung interactions. Part 1: effects of lung volume and ventilation as exercise. Int J Intensive Care 2001, Spring: 6-13.
40.
Zurück zum Zitat Harrigan PWJ, Pinsky MR: Heart–lung interactions. Part 2: effects of intrathoracic pressure. Int J Intensive Care 2001, Summer: 99-108. Harrigan PWJ, Pinsky MR: Heart–lung interactions. Part 2: effects of intrathoracic pressure. Int J Intensive Care 2001, Summer: 99-108.
41.
Zurück zum Zitat Rooke GA, Schwid HA, Shapira Y: The effect of graded hemorrhage and intravascular volume replacement on systolic pressure variation in humans during mechanical and spontaneous ventilation. Anesth Analg 1995, 80: 925-932.PubMed Rooke GA, Schwid HA, Shapira Y: The effect of graded hemorrhage and intravascular volume replacement on systolic pressure variation in humans during mechanical and spontaneous ventilation. Anesth Analg 1995, 80: 925-932.PubMed
42.
Zurück zum Zitat Marik PE: The systolic blood pressure variation as an indicator of pulmonary capillary wedge pressure in ventilated patients. Anaesth Intensive Care 1993, 21: 405-408.PubMed Marik PE: The systolic blood pressure variation as an indicator of pulmonary capillary wedge pressure in ventilated patients. Anaesth Intensive Care 1993, 21: 405-408.PubMed
43.
Zurück zum Zitat Tavernier B, Makhotine O, Lebuffe G, Dupont J, Scherpereel P: Systolic pressure variation as a guide to fluid therapy in patients with sepsis-induced hypotension [abstract]. Anesthesiology 1998, 89: 1309-1310.CrossRef Tavernier B, Makhotine O, Lebuffe G, Dupont J, Scherpereel P: Systolic pressure variation as a guide to fluid therapy in patients with sepsis-induced hypotension [abstract]. Anesthesiology 1998, 89: 1309-1310.CrossRef
44.
Zurück zum Zitat Pizov , et al.: Positive end-expiratory pressure-induced hemodynamic changes are reflected in the arterial pressure waveform [abstract]. Crit Care Med 1996, 24: 1381-1387. 10.1097/00003246-199608000-00018CrossRefPubMed Pizov , et al.: Positive end-expiratory pressure-induced hemodynamic changes are reflected in the arterial pressure waveform [abstract]. Crit Care Med 1996, 24: 1381-1387. 10.1097/00003246-199608000-00018CrossRefPubMed
45.
Zurück zum Zitat Szold A, Pizov R, Segal E, Perel A: The effect of tidal volume and intravascular volume state on systolic pressure variation in ventilated dogs [abstract]. Intensive Care Med 1989, 15: 368-371.CrossRefPubMed Szold A, Pizov R, Segal E, Perel A: The effect of tidal volume and intravascular volume state on systolic pressure variation in ventilated dogs [abstract]. Intensive Care Med 1989, 15: 368-371.CrossRefPubMed
46.
Zurück zum Zitat Beaussier M, Coriat P, Perel A, Lebret F, Kalton P, Chemla D, Lienhart A, Viars P: Determinants of systolic pressure variation in patients ventilated after vascular surgery. J Cardiothorac Vasc Anesth 1995, 9: 547-551.CrossRefPubMed Beaussier M, Coriat P, Perel A, Lebret F, Kalton P, Chemla D, Lienhart A, Viars P: Determinants of systolic pressure variation in patients ventilated after vascular surgery. J Cardiothorac Vasc Anesth 1995, 9: 547-551.CrossRefPubMed
47.
Zurück zum Zitat Pizov R, Segal E, Kaplan L, Floman Y, Perel A: The use of systolic pressure variation in hemodynamic monitoring during deliberate hypotension in spine surgery [abstract]. J Clin Anesth 1990, 2: 96-100. 10.1016/0952-8180(90)90061-7CrossRefPubMed Pizov R, Segal E, Kaplan L, Floman Y, Perel A: The use of systolic pressure variation in hemodynamic monitoring during deliberate hypotension in spine surgery [abstract]. J Clin Anesth 1990, 2: 96-100. 10.1016/0952-8180(90)90061-7CrossRefPubMed
48.
Zurück zum Zitat Perel A, Pizov R, Cotev S: Systolic blood pressure variation is a sensitive indicator of hypovolemia in ventilated dogs subjected to graded hemorrhage. Anesthesiology 1987, 67: 498-502.CrossRefPubMed Perel A, Pizov R, Cotev S: Systolic blood pressure variation is a sensitive indicator of hypovolemia in ventilated dogs subjected to graded hemorrhage. Anesthesiology 1987, 67: 498-502.CrossRefPubMed
49.
Zurück zum Zitat Weiss YG, Oppenheim-Eden A, Gilon D, Sprung CL, Muggia-Sullam M, Pizov R: Systolic pressure variation in hemodynamic monitoring after severe blast injury [abstract]. J Clin Anesth 1999, 11: 132-135. 10.1016/S0952-8180(99)00006-9CrossRefPubMed Weiss YG, Oppenheim-Eden A, Gilon D, Sprung CL, Muggia-Sullam M, Pizov R: Systolic pressure variation in hemodynamic monitoring after severe blast injury [abstract]. J Clin Anesth 1999, 11: 132-135. 10.1016/S0952-8180(99)00006-9CrossRefPubMed
50.
Zurück zum Zitat Ornstein E, Eidelman LA, Drenger B, Elami A, Pizov R: Systolic pressure variation predicts the response to acute blood loss [abstract]. J Clin Anesth 1998, 10: 137-140. 10.1016/S0952-8180(97)00257-2CrossRefPubMed Ornstein E, Eidelman LA, Drenger B, Elami A, Pizov R: Systolic pressure variation predicts the response to acute blood loss [abstract]. J Clin Anesth 1998, 10: 137-140. 10.1016/S0952-8180(97)00257-2CrossRefPubMed
51.
Zurück zum Zitat Klinzing S, Seeber P, Schiergens V, Sakka S, Reinhart K, Meier-Hellmann A: Stroke volume variation as a predictor of fluid responsiveness for cardiac output in patients undergoing cardiac surgery [abstract]. Crit Care Med 2002,29(suppl):173/M55. Klinzing S, Seeber P, Schiergens V, Sakka S, Reinhart K, Meier-Hellmann A: Stroke volume variation as a predictor of fluid responsiveness for cardiac output in patients undergoing cardiac surgery [abstract]. Crit Care Med 2002,29(suppl):173/M55.
52.
Zurück zum Zitat Reuter DA, Kirchner A, Kilger E, Lamm P, Goetz AE: Left ventricular stroke volume variations for functional preload monitoring after cardiac surgery in high risk patients [abstract]. SCCM 93/M1 Reuter DA, Kirchner A, Kilger E, Lamm P, Goetz AE: Left ventricular stroke volume variations for functional preload monitoring after cardiac surgery in high risk patients [abstract]. SCCM 93/M1
53.
Zurück zum Zitat Roberts PR, Allen S, Robinson S, Tanser SJ, Jonas MM, Morgan JM: Use of lithium dilution assessment of cardiac output to optimise right/left ventricular activation in resynchronisation therapy [abstract]. Heart 2002,87(suppl II):146. Roberts PR, Allen S, Robinson S, Tanser SJ, Jonas MM, Morgan JM: Use of lithium dilution assessment of cardiac output to optimise right/left ventricular activation in resynchronisation therapy [abstract]. Heart 2002,87(suppl II):146.
54.
Zurück zum Zitat Kern JW, Shoemaker WC: Meta-analysis of hemodynamic optimisation in high-risk patients. Crit Care Med 2002, 30: 1686-1692. 10.1097/00003246-200208000-00002CrossRefPubMed Kern JW, Shoemaker WC: Meta-analysis of hemodynamic optimisation in high-risk patients. Crit Care Med 2002, 30: 1686-1692. 10.1097/00003246-200208000-00002CrossRefPubMed
55.
Zurück zum Zitat Bennett ED: Goal-directed therapy is successful in the right patients [editorial]. Crit Care Med 2002, 30: 1909-1910. 10.1097/00003246-200208000-00042CrossRefPubMed Bennett ED: Goal-directed therapy is successful in the right patients [editorial]. Crit Care Med 2002, 30: 1909-1910. 10.1097/00003246-200208000-00042CrossRefPubMed
56.
Zurück zum Zitat Rivers E, Nguyen B, Havstad S, Ressler J, Muzzin A, Knoblich B, Peterson E, Tomlanovich M, Early Goal-Directed Therapy Collaborative Group: Early goal-directed therapy in the treatment of severe sepsis and septic shock. N Engl J Med 2001, 345: 1368-1377. 10.1056/NEJMoa010307CrossRefPubMed Rivers E, Nguyen B, Havstad S, Ressler J, Muzzin A, Knoblich B, Peterson E, Tomlanovich M, Early Goal-Directed Therapy Collaborative Group: Early goal-directed therapy in the treatment of severe sepsis and septic shock. N Engl J Med 2001, 345: 1368-1377. 10.1056/NEJMoa010307CrossRefPubMed
57.
Zurück zum Zitat Singh S, Manji M: A survey of pre-operative optimisation of high-risk surgical patients undergoing major elective surgery. Anaesthesia 2001, 56: 988-1002. 10.1046/j.1365-2044.2001.01974.xCrossRefPubMed Singh S, Manji M: A survey of pre-operative optimisation of high-risk surgical patients undergoing major elective surgery. Anaesthesia 2001, 56: 988-1002. 10.1046/j.1365-2044.2001.01974.xCrossRefPubMed
58.
Zurück zum Zitat Sandham JD, Hull RD, Brant RF, Knox L, Pineo GF, Doig CJ, Laporta DP, Viner S, Passerini L, Devitt H, Kirby A, Jacka M, Canadian Critical Care Clinical Trials Group: A randomized, controlled trial of the use of pulmonary-artery catheters in high-risk surgical patients. N Engl J Med 2003, 348: 5-14. 10.1056/NEJMoa021108CrossRefPubMed Sandham JD, Hull RD, Brant RF, Knox L, Pineo GF, Doig CJ, Laporta DP, Viner S, Passerini L, Devitt H, Kirby A, Jacka M, Canadian Critical Care Clinical Trials Group: A randomized, controlled trial of the use of pulmonary-artery catheters in high-risk surgical patients. N Engl J Med 2003, 348: 5-14. 10.1056/NEJMoa021108CrossRefPubMed
59.
Zurück zum Zitat Jonas M, Bruce R, Knight J, Kelly F: Clinical assessment of cardiac output versus LiDCO indicator dilution measurement: are clinical estimates of cardiac output and oxygen delivery reliable enough to manage critically ill patients? Crit Care Med 2002. Jonas M, Bruce R, Knight J, Kelly F: Clinical assessment of cardiac output versus LiDCO indicator dilution measurement: are clinical estimates of cardiac output and oxygen delivery reliable enough to manage critically ill patients? Crit Care Med 2002.
60.
Zurück zum Zitat Boyd O, Grounds RM, Bennett ED: A randomized clinical trial of the effect of deliberate perioperative increase of oxygen delivery on mortality in high-risk surgical patients. JAMA 1993, 270: 2699-2707. 10.1001/jama.270.22.2699CrossRefPubMed Boyd O, Grounds RM, Bennett ED: A randomized clinical trial of the effect of deliberate perioperative increase of oxygen delivery on mortality in high-risk surgical patients. JAMA 1993, 270: 2699-2707. 10.1001/jama.270.22.2699CrossRefPubMed
61.
Zurück zum Zitat Wilson J, Woods I, Fawcett J, Whall R, Dibb W, Morris C, McManus E: Reducing the risk of major elective surgery: randomised controlled trial of preoperative optimisation of oxygen delivery. BMJ 1999, 318: 1099-1103.PubMedCentralCrossRefPubMed Wilson J, Woods I, Fawcett J, Whall R, Dibb W, Morris C, McManus E: Reducing the risk of major elective surgery: randomised controlled trial of preoperative optimisation of oxygen delivery. BMJ 1999, 318: 1099-1103.PubMedCentralCrossRefPubMed
62.
Zurück zum Zitat Sinclair S, James S, Singer M: Intraoperative intravascular volume optimisation and length of hospital stay after repair of proximal femoral fracture: randomised controlled trial. BMJ 1997, 315: 909-912.PubMedCentralCrossRefPubMed Sinclair S, James S, Singer M: Intraoperative intravascular volume optimisation and length of hospital stay after repair of proximal femoral fracture: randomised controlled trial. BMJ 1997, 315: 909-912.PubMedCentralCrossRefPubMed
63.
Zurück zum Zitat Mythen MG, Webb AR: Perioperative plasma volume expansion reduces the incidence of gut mucosal hypoperfusion during cardiac surgery. Arch Surg 1995, 130: 423-429.CrossRefPubMed Mythen MG, Webb AR: Perioperative plasma volume expansion reduces the incidence of gut mucosal hypoperfusion during cardiac surgery. Arch Surg 1995, 130: 423-429.CrossRefPubMed
64.
Zurück zum Zitat Gan TJ, Soppitt A, Maroof M, el-Moalem H, Robertson KM, Moretti E, Dwane P, Glass PS: Goal-directed intraoperative fluid administration reduces length of hospital stay after major surgery. Anesthesiology 2002, 97: 820-826. 10.1097/00000542-200210000-00012CrossRefPubMed Gan TJ, Soppitt A, Maroof M, el-Moalem H, Robertson KM, Moretti E, Dwane P, Glass PS: Goal-directed intraoperative fluid administration reduces length of hospital stay after major surgery. Anesthesiology 2002, 97: 820-826. 10.1097/00000542-200210000-00012CrossRefPubMed
65.
Zurück zum Zitat Follath F, Cleland JG, Just H, Papp JG, Scholz H, Peuhkurinen K, Harjola VP, Mitrovic V, Abdalla M, Sandell EP, Lehtonen L, Steering Committee and Investigators of the Levosimendan Infusion versus Dobutamine (LIDO) Study: Efficacy and safety of intravenous levosimendan compared with dobutamine in severe low-output heart failure (the LIDO study): a randomised double-blind trial. Lancet 2002, 360: 196-202. 10.1016/S0140-6736(02)09455-2CrossRefPubMed Follath F, Cleland JG, Just H, Papp JG, Scholz H, Peuhkurinen K, Harjola VP, Mitrovic V, Abdalla M, Sandell EP, Lehtonen L, Steering Committee and Investigators of the Levosimendan Infusion versus Dobutamine (LIDO) Study: Efficacy and safety of intravenous levosimendan compared with dobutamine in severe low-output heart failure (the LIDO study): a randomised double-blind trial. Lancet 2002, 360: 196-202. 10.1016/S0140-6736(02)09455-2CrossRefPubMed
66.
Zurück zum Zitat Linton R, Band D, O'Brien T, Jonas M, Leach R: Lithium dilution cardiac output measurement: a comparison with thermodilution. Crit Care Med 1997, 25: 1796-800. 10.1097/00003246-199711000-00015CrossRefPubMed Linton R, Band D, O'Brien T, Jonas M, Leach R: Lithium dilution cardiac output measurement: a comparison with thermodilution. Crit Care Med 1997, 25: 1796-800. 10.1097/00003246-199711000-00015CrossRefPubMed
67.
Zurück zum Zitat Linton RA, Band DM, Haire KM: A new method of measuring cardiac output in man using lithium dilution. Br J Anaesth 1993, 71: 262-266.CrossRefPubMed Linton RA, Band DM, Haire KM: A new method of measuring cardiac output in man using lithium dilution. Br J Anaesth 1993, 71: 262-266.CrossRefPubMed
68.
Zurück zum Zitat Linton RA, Jonas MM, Tibby SM, Murdoch IA, O'Brien TK, Linton NW, Band DM: Cardiac output measured by lithium dilution and transpulmonary thermodilution in patients in a paediatric intensive care unit. Intensive Care Med 2000, 26: 1507-1511. 10.1007/s001340051347CrossRefPubMed Linton RA, Jonas MM, Tibby SM, Murdoch IA, O'Brien TK, Linton NW, Band DM: Cardiac output measured by lithium dilution and transpulmonary thermodilution in patients in a paediatric intensive care unit. Intensive Care Med 2000, 26: 1507-1511. 10.1007/s001340051347CrossRefPubMed
69.
Zurück zum Zitat Stetz CW, Miller RG, Kelly GE, Raffin TA: Reliability of the thermodilution method in the determination of cardiac output in clinical practice. Am Rev Respir Dis 1982, 126: 1001-1004.PubMed Stetz CW, Miller RG, Kelly GE, Raffin TA: Reliability of the thermodilution method in the determination of cardiac output in clinical practice. Am Rev Respir Dis 1982, 126: 1001-1004.PubMed
70.
Zurück zum Zitat JBand D, Linton NW, Kelly F, Burden T, Chevalier S, Thompson R, Birch N, Powell J: The pharmacokinetics of intravenous lithium chloride in patients and normal volunteers. J Trace Microbe Techn 2001, 19: 313-320. 10.1081/TMA-100002220CrossRef JBand D, Linton NW, Kelly F, Burden T, Chevalier S, Thompson R, Birch N, Powell J: The pharmacokinetics of intravenous lithium chloride in patients and normal volunteers. J Trace Microbe Techn 2001, 19: 313-320. 10.1081/TMA-100002220CrossRef
71.
Zurück zum Zitat Linton NW, Linton R: Estimation of changes in cardiac output from arterial blood pressure waveform in the upper limb. Br J Anaesth 2001, 86: 486-496. 10.1093/bja/86.4.486CrossRefPubMed Linton NW, Linton R: Estimation of changes in cardiac output from arterial blood pressure waveform in the upper limb. Br J Anaesth 2001, 86: 486-496. 10.1093/bja/86.4.486CrossRefPubMed
72.
Zurück zum Zitat Michard F, Teboul JL: Predicting fluid responsiveness in ICU patients: a critical analysis of the evidence. Chest 2002, 121: 2000-2008. 10.1378/chest.121.6.2000CrossRefPubMed Michard F, Teboul JL: Predicting fluid responsiveness in ICU patients: a critical analysis of the evidence. Chest 2002, 121: 2000-2008. 10.1378/chest.121.6.2000CrossRefPubMed
Metadaten
Titel
Equipment review: An appraisal of the LiDCO™plus method of measuring cardiac output
verfasst von
Rupert M Pearse
Kashif Ikram
John Barry
Publikationsdatum
01.06.2004
Verlag
BioMed Central
Erschienen in
Critical Care / Ausgabe 3/2004
Elektronische ISSN: 1364-8535
DOI
https://doi.org/10.1186/cc2852

Weitere Artikel der Ausgabe 3/2004

Critical Care 3/2004 Zur Ausgabe

Update AINS

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.