Bench-to-bedside review: The importance of the precision of the reference technique in method comparison studies – with specific reference to the measurement of cardiac output

verfasst von: Maurizio Cecconi, Andrew Rhodes, Jan Poloniecki, Giorgio Della Rocca, R Michael Grounds

Erschienen in: Critical Care | Ausgabe 1/2009

Einloggen, um Zugang zu erhalten

Abstract

Bland-Altman analysis is used for assessing agreement between two measurements of the same clinical variable. In the field of cardiac output monitoring, its results, in terms of bias and limits of agreement, are often difficult to interpret, leading clinicians to use a cutoff of 30% in the percentage error in order to decide whether a new technique may be considered a good alternative. This percentage error of ± 30% arises from the assumption that the commonly used reference technique, intermittent thermodilution, has a precision of ± 20% or less. The combination of two precisions of ± 20% equates to a total error of ± 28.3%, which is commonly rounded up to ± 30%. Thus, finding a percentage error of less than ± 30% should equate to the new tested technique having an error similar to the reference, which therefore should be acceptable. In a worked example in this paper, we discuss the limitations of this approach, in particular in regard to the situation in which the reference technique may be either more or less precise than would normally be expected. This can lead to inappropriate conclusions being drawn from data acquired in validation studies of new monitoring technologies. We conclude that it is not acceptable to present comparison studies quoting percentage error as an acceptability criteria without reporting the precision of the reference technique.
Anhänge
Nur mit Berechtigung zugänglich
Literatur
1.
Zurück zum Zitat Bland JM, Altman DG: Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986, 1: 307-310.CrossRefPubMed Bland JM, Altman DG: Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986, 1: 307-310.CrossRefPubMed
2.
Zurück zum Zitat Bland JM, Altman DG: Comparing methods of measurement: why plotting difference against standard method is misleading. Lancet. 1995, 346: 1085-1087. 10.1016/S0140-6736(95)91748-9.CrossRefPubMed Bland JM, Altman DG: Comparing methods of measurement: why plotting difference against standard method is misleading. Lancet. 1995, 346: 1085-1087. 10.1016/S0140-6736(95)91748-9.CrossRefPubMed
3.
Zurück zum Zitat Bland JM, Altman DG: Measuring agreement in method comparison studies. Stat Methods Med Res. 1999, 8: 135-160. 10.1191/096228099673819272.CrossRefPubMed Bland JM, Altman DG: Measuring agreement in method comparison studies. Stat Methods Med Res. 1999, 8: 135-160. 10.1191/096228099673819272.CrossRefPubMed
4.
Zurück zum Zitat Hamilton TT, Huber LM, Jessen ME: PulseCO: a less-invasive method to monitor cardiac output from arterial pressure after cardiac surgery. Ann Thorac Surg. 2002, 74: S1408-1412. 10.1016/S0003-4975(02)04059-6.CrossRefPubMed Hamilton TT, Huber LM, Jessen ME: PulseCO: a less-invasive method to monitor cardiac output from arterial pressure after cardiac surgery. Ann Thorac Surg. 2002, 74: S1408-1412. 10.1016/S0003-4975(02)04059-6.CrossRefPubMed
5.
Zurück zum Zitat Lichtenthal PR, Gordan D: Testing the safety of Baxter continuous cardiac output monitoring system. J Clin Monit. 1996, 12: 243-249. 10.1007/BF00857646.CrossRefPubMed Lichtenthal PR, Gordan D: Testing the safety of Baxter continuous cardiac output monitoring system. J Clin Monit. 1996, 12: 243-249. 10.1007/BF00857646.CrossRefPubMed
6.
Zurück zum Zitat Munro HM, Wood CE, Taylor BL, Smith GB: Continuous invasive cardiac output monitoring – the Baxter/Edwards Critical-Care Swan Ganz IntelliCath and Viligance system. Clin Intensive Care. 1994, 5: 52-55.PubMed Munro HM, Wood CE, Taylor BL, Smith GB: Continuous invasive cardiac output monitoring – the Baxter/Edwards Critical-Care Swan Ganz IntelliCath and Viligance system. Clin Intensive Care. 1994, 5: 52-55.PubMed
7.
Zurück zum Zitat Della Rocca G, Costa MG, Pompei L, Coccia C, Pietropaoli P: Continuous and intermittent cardiac output measurement: pulmonary artery catheter versus aortic transpulmonary technique. Br J Anaesth. 2002, 88: 350-356. 10.1093/bja/88.3.350.CrossRefPubMed Della Rocca G, Costa MG, Pompei L, Coccia C, Pietropaoli P: Continuous and intermittent cardiac output measurement: pulmonary artery catheter versus aortic transpulmonary technique. Br J Anaesth. 2002, 88: 350-356. 10.1093/bja/88.3.350.CrossRefPubMed
8.
Zurück zum Zitat Goedje O, Hoeke K, Lichtwarck-Aschoff M, Faltchauser A, Lamm P, Reichart B: Continuous cardiac output by femoral arterial thermodilution calibrated pulse contour analysis: comparison with pulmonary arterial thermodilution. Crit Care Med. 1999, 27: 2407-2412. 10.1097/00003246-199911000-00014.CrossRefPubMed Goedje O, Hoeke K, Lichtwarck-Aschoff M, Faltchauser A, Lamm P, Reichart B: Continuous cardiac output by femoral arterial thermodilution calibrated pulse contour analysis: comparison with pulmonary arterial thermodilution. Crit Care Med. 1999, 27: 2407-2412. 10.1097/00003246-199911000-00014.CrossRefPubMed
9.
Zurück zum Zitat Button D, Weibel L, Reuthebuch O, Genoni M, Zollinger A, Hofer CK: Clinical evaluation of the FloTrac/Vigileo system and two established continuous cardiac output monitoring devices in patients undergoing cardiac surgery. Br J Anaesth. 2007, 99: 329-336. 10.1093/bja/aem188.CrossRefPubMed Button D, Weibel L, Reuthebuch O, Genoni M, Zollinger A, Hofer CK: Clinical evaluation of the FloTrac/Vigileo system and two established continuous cardiac output monitoring devices in patients undergoing cardiac surgery. Br J Anaesth. 2007, 99: 329-336. 10.1093/bja/aem188.CrossRefPubMed
10.
Zurück zum Zitat Mayer J, Boldt J, Schollhorn T, Rohm KD, Mengistu AM, Suttner S: Semi-invasive monitoring of cardiac output by a new device using arterial pressure waveform analysis: a comparison with intermittent pulmonary artery thermodilution in patients undergoing cardiac surgery. Br J Anaesth. 2007, 98: 176-182. 10.1093/bja/ael341.CrossRefPubMed Mayer J, Boldt J, Schollhorn T, Rohm KD, Mengistu AM, Suttner S: Semi-invasive monitoring of cardiac output by a new device using arterial pressure waveform analysis: a comparison with intermittent pulmonary artery thermodilution in patients undergoing cardiac surgery. Br J Anaesth. 2007, 98: 176-182. 10.1093/bja/ael341.CrossRefPubMed
11.
Zurück zum Zitat Sander M, Spies CD, Grubitzsch H, Foer A, Muller M, von Heymann C: Comparison of uncalibrated arterial waveform analysis in cardiac surgery patients with thermodilution cardiac output measurements. Crit Care. 2006, 10: R164-10.1186/cc5103.PubMedCentralCrossRefPubMed Sander M, Spies CD, Grubitzsch H, Foer A, Muller M, von Heymann C: Comparison of uncalibrated arterial waveform analysis in cardiac surgery patients with thermodilution cardiac output measurements. Crit Care. 2006, 10: R164-10.1186/cc5103.PubMedCentralCrossRefPubMed
12.
Zurück zum Zitat Barin E, Haryadi DG, Schookin SI, Westenskow DR, Zubenko VG, Beliaev KR, Morozov AA: Evaluation of a thoracic bioimpedance cardiac output monitor during cardiac catheterization. Crit Care Med. 2000, 28: 698-702. 10.1097/00003246-200003000-00016.CrossRefPubMed Barin E, Haryadi DG, Schookin SI, Westenskow DR, Zubenko VG, Beliaev KR, Morozov AA: Evaluation of a thoracic bioimpedance cardiac output monitor during cardiac catheterization. Crit Care Med. 2000, 28: 698-702. 10.1097/00003246-200003000-00016.CrossRefPubMed
13.
Zurück zum Zitat Neviere R, Mathieu D, Riou Y, Chagnon JL, Wattel F: Non-invasive measurement of cardiac output in patients with acute lung injury using the carbon dioxide rebreathing method. Clin Intensive Care. 1994, 5: 172-175.PubMed Neviere R, Mathieu D, Riou Y, Chagnon JL, Wattel F: Non-invasive measurement of cardiac output in patients with acute lung injury using the carbon dioxide rebreathing method. Clin Intensive Care. 1994, 5: 172-175.PubMed
14.
Zurück zum Zitat Jakobsen CJ, Melsen NC, Andresen EB: Continuous cardiac output measurements in the perioperative period. Acta Anaesthesiol Scand. 1995, 39: 485-488. 10.1111/j.1399-6576.1995.tb04104.x.CrossRefPubMed Jakobsen CJ, Melsen NC, Andresen EB: Continuous cardiac output measurements in the perioperative period. Acta Anaesthesiol Scand. 1995, 39: 485-488. 10.1111/j.1399-6576.1995.tb04104.x.CrossRefPubMed
15.
Zurück zum Zitat Critchley LA, Critchley JA: A meta-analysis of studies using bias and precision statistics to compare cardiac output measurement techniques. J Clin Monit Comput. 1999, 15: 85-91. 10.1023/A:1009982611386.CrossRefPubMed Critchley LA, Critchley JA: A meta-analysis of studies using bias and precision statistics to compare cardiac output measurement techniques. J Clin Monit Comput. 1999, 15: 85-91. 10.1023/A:1009982611386.CrossRefPubMed
16.
Zurück zum Zitat Pittman J, Bar-Yosef S, SumPing J, Sherwood M, Mark J: Continuous cardiac output monitoring with pulse contour analysis: a comparison with lithium indicator dilution cardiac output measurement. Crit Care Med. 2005, 33: 2015-2021. 10.1097/01.CCM.0000179021.36805.1F.CrossRefPubMed Pittman J, Bar-Yosef S, SumPing J, Sherwood M, Mark J: Continuous cardiac output monitoring with pulse contour analysis: a comparison with lithium indicator dilution cardiac output measurement. Crit Care Med. 2005, 33: 2015-2021. 10.1097/01.CCM.0000179021.36805.1F.CrossRefPubMed
17.
Zurück zum Zitat Bein B, Worthmann F, Tonner PH, Paris A, Steinfath M, Hedderich J, Scholz J: Comparison of esophageal Doppler, pulse contour analysis, and real-time pulmonary artery thermodilution for the continuous measurement of cardiac output. J Cardiothorac Vasc Anesth. 2004, 18: 185-189. 10.1053/j.jvca.2004.01.025.CrossRefPubMed Bein B, Worthmann F, Tonner PH, Paris A, Steinfath M, Hedderich J, Scholz J: Comparison of esophageal Doppler, pulse contour analysis, and real-time pulmonary artery thermodilution for the continuous measurement of cardiac output. J Cardiothorac Vasc Anesth. 2004, 18: 185-189. 10.1053/j.jvca.2004.01.025.CrossRefPubMed
18.
Zurück zum Zitat Pearse R, Dawson D, Fawcett J, Rhodes A, Grounds RM, Bennett ED: Early goal-directed therapy after major surgery reduces complications and duration of hospital stay. A randomised, controlled trial [ISRCTN38797445]. Crit Care. 2005, 9: R687-10.1186/cc3887.PubMedCentralCrossRefPubMed Pearse R, Dawson D, Fawcett J, Rhodes A, Grounds RM, Bennett ED: Early goal-directed therapy after major surgery reduces complications and duration of hospital stay. A randomised, controlled trial [ISRCTN38797445]. Crit Care. 2005, 9: R687-10.1186/cc3887.PubMedCentralCrossRefPubMed
19.
Zurück zum Zitat Boyd O, Grounds RM, Bennett ED: A randomized clinical trial of the effect of deliberate perioperative increase of oxygen delivery on mortality in high-risk surgical patients. JAMA. 1993, 270: 2699-2707. 10.1001/jama.270.22.2699.CrossRefPubMed Boyd O, Grounds RM, Bennett ED: A randomized clinical trial of the effect of deliberate perioperative increase of oxygen delivery on mortality in high-risk surgical patients. JAMA. 1993, 270: 2699-2707. 10.1001/jama.270.22.2699.CrossRefPubMed
20.
Zurück zum Zitat Roeck M, Jakob SM, Boehlen T, Brander L, Knuesel R, Takala J: Change in stroke volume in response to fluid challenge: assessment using esophageal Doppler. Intensive Care Med. 2003, 29: 1729-1735. 10.1007/s00134-003-1720-4.CrossRefPubMed Roeck M, Jakob SM, Boehlen T, Brander L, Knuesel R, Takala J: Change in stroke volume in response to fluid challenge: assessment using esophageal Doppler. Intensive Care Med. 2003, 29: 1729-1735. 10.1007/s00134-003-1720-4.CrossRefPubMed
Metadaten
Titel
Bench-to-bedside review: The importance of the precision of the reference technique in method comparison studies – with specific reference to the measurement of cardiac output
verfasst von
Maurizio Cecconi
Andrew Rhodes
Jan Poloniecki
Giorgio Della Rocca
R Michael Grounds
Publikationsdatum
01.02.2009
Verlag
BioMed Central
Erschienen in
Critical Care / Ausgabe 1/2009
Elektronische ISSN: 1364-8535
DOI
https://doi.org/10.1186/cc7129

Weitere Artikel der Ausgabe 1/2009

Critical Care 1/2009 Zur Ausgabe

Update AINS

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.